Update the Dilumat Expert firmware
I. FIRMWARE UPDATE
1. Turn on the DILUMAT and plug the USB key containing the firmware update.
2. If you have enabled the user management option, log on with the Administrator
account.
3. Backup the configuration:
a. Press the key then the key.
b. Press .
c. Press to access the Import / Export screen.
d. Press .
e. Select your chosen external device in the "Export to" screen.
f. In the right-hand column, select where you want to store the configuration
file.
g. Press to confirm.
4. On the control unit, Press .
5. Select Support , then Update.
6. Press on the update screen.
2 DIL_TECDOC005_EN_D
7. Select Generic mass storage in the CONTENT pane (or the USB key specific
name).
The USB key appears in the FILE MANAGER pane.
8. Select the update file dilumat-fig-1.2.5821.adf in the CONTENT pane and
press .
The update begins and the DILUMAT restarts.
9. A message Firmware Update completed is displayed. Press OK.
10. The update is ended. Log on as usual.
II. FIRMWARE UPDATE VERIFICATION
1. On the control unit, press .
2. Select Support , then .
Verify that the Firmware version is: fig-1.2.5821
3 DIL_TECDOC005_EN_D
Update the Dilumat Start firmware
I. FIRMWARE UPDATE
1. Turn on the DILUMAT and plug the USB key containing the firmware update.
2. Press on and enter the administrator password 35170 to activate the SUPPORT
button.
3. Backup the configuration:
a) Press the key then the key.
b) Press .
c) Press to access the Import / Export screen.
d) Press .
e) Select where you want to store the configuration file.
f) Press to confirm.
4. On the control unit, press .
5. Select Support , then Update.
6. Press on the update screen.
4 DIL_TECDOC005_EN_D
7. Select Generic mass storage in the CONTENT pane (or the USB key specific
name).
The USB key appears in the FILE MANAGER pane.
8. Select the update file dilumat-fig-1.2.5821.adf in the CONTENT pane and
press .
The update begins and the DILUMAT restarts.
9. A message Firmware Update completed is displayed. Press OK.
10. The update is ended. Log on as usual.
II. FIRMWARE UPDATE VERIFICATION
11. On the control unit, Press .
12. Select Support , then .
Verify that the Firmware version is: fig-1.2.5821
5 DIL_TECDOC005_EN_D