KIKKOMANS BIOTECHNOLOGY
ATP SWAB TESTS
USING
LUMITESTER PD-10 & LUCIPAC-W
SUPPLIER :
BIOCHEMICAL DIVISION, KIKKOMAN CORPORATION
2-1-1, NISHI-SHINBASHI
MINATO-KU, TOKYO 105-8428
TEL 03-5521-5490, FAX 03-5521-5498
C 2001 Kikkoman Corporation
PD-10 HYGIENE MONITORING - 2
Exist as energy source in all living things on
earth
Battery for living things
Evidence for living activity
During the Apollo program, the existence
of living organisms were determined by
the presence of ATP.
Does not exist in non-living things
Index for degree of cleanliness
Rapid inference of quantity of microbes
C 2001 Kikkoman Corporation
PD-10 HYGIENE MONITORING - 3
MERITS OF USING KIKKOMANS ATP SWAB TEST
SYSTEM FOR HYGIENE MONITORING
1 DEGREE OF CONTAMINATION CAN BE NUMERICALLY EXPRESSED
RAPIDLY ON SITE.
- ABLE TO KNOW RAPIDLY IF CLEANING IS INSUFFICIENT
RECLEANING POSSIBLE.
- HYGIENE MONITORING CAN BE DONE NUMERICALLY
- VERY SUITABLE AS ON-SITE INSTRUCTION TOOL
2 DETECTS NOT ONLY MICROORGANISMS BUT ALSO FOOD RESIDUE
FOOD RESIDUE SOURCE OF NUTRIENTS FOR MICROBES
3 THE LIGHTEST PORTABLE LUMINOMETER IN THE WORLD
4 EASY TO USE
2-3 days
TRADITIONAL Check
Check
TRADITIONAL
CULTURE
CULTURE
METHOD Checking for
METHOD microbes only
Within 30 sec
LUCIPAC-W
LUCIPAC-W Check
Check
Checking for
microbes +
food residues
C 2001 Kikkoman Corporation
PD-10 HYGIENE MONITORING - 4
POSSIBLE TO DETERMINE IF CLEANING IS
SUFFICIENT USING ATP METHOD
TRADITIONAL MICROORGANISM
ATP METHOD
METHOD
(PLATE COUNTING)
FOOD RESIDUE
JUDGED DIRTY CHOPPING BOARD JUDGED DIRTY
BOTH WASHING AND STERILIZATION INSUFFICIENT
FOOD RESIDUE
JUDGED CLEAN CHOPPING BOARD JUDGED DIRTY
INSUFFICIENT WASHING BUT ENOUGH STERILIZATION
AS TIME PASSES, MICROBES MULTIPLY TO DANGEROUSLY HIGH LEVELS
JUDGED CLEAN CHOPPING BOARD JUDGED CLEAN
BOTH WASHING AND STERILIZATION SUFFICIENT
WHEN FOOD RESIDUE REMAINS DUE TO INSUFFICIENT WASHING ...
BECOMES HOTBED FOR MULTIPLICATION OF MICROBES
BECOMES REASON FOR LOSS OF STERILIZATION EFFECT IN SOME CASES
CAN THEREFORE BE CAUSE OF FOOD POISONING THROUGH
SECONDARY CONTAMINATION
USE LUMITESTER PD-10 & LUCIPAC-W FOR EDUCATING STAFF ON
HYGIENE MATTERS AND MONITORING HYGIENE CONDITIONS OF
EQUIPMENT & FACILITIES, WHICH ARE GENERAL HYGIENE MONITORING
CRITERIA AND ARE PRECONDITIONS FOR HACCP IMPLEMENTATION.
C 2001 Kikkoman Corporation
PD-10 HYGIENE MONITORING - 5
RELATIONSHIP BETWEEN RLU VALUES (FROM ATP
SWAB TESTS) AND CELL COUNTS
(EXAMPLE OF DATA FROM KITCHEN)
(CFU / ml)
10,000,000
A LOT OF MICROBES DETECTED
1,000,000
CELL COUNTING DATA
100,000
10,000
1,000
100
LATENT RISK OF CONTAMINATION
N.D.
1 10 100 1,000 10,000 100,000 (RLU)
BIOLUMINESCENCE (MEASURED BY PD-10)
POINTS WITH HIGH BIOLUMINESCENCE COINCIDE WITH POINTS WITH HIGH CELL
COUNTS
WITH ATP METHOD, IT IS POSSIBLE TO DETECT LOCATIONS WHERE FOOD RESIDUE IS
PRESENT AS A RESULT OF INSUFFICIENT CLEANING, EVENTHOUGH FEW MICROBES MAY
EXIST
USING PD-10 & LUCIPAC-W, IT IS POSSIBLE TO MONITOR CLEANLINESS IN
A RAPID AND EFFECTIVE MANNER
C 2001 Kikkoman Corporation
PD-10 HYGIENE MONITORING - 6
INTRODUCTION OF ATP SWAB TESTS USING
LUMITESTER PD-10 & LUCIPAC-W
STEP 1
CHOOSE POTENTIAL SAMPLING POINTS FOR CONDUCTING SWAB TESTS
LOCATIONS WHERE CONTAMINATION WILL HAVE HIGH POSSIBILITY OF DIRECTLY AFFECTING
FINISHED PRODUCTS, SUCH AS FILLING MACHINES AFTER HEATING, PACKING LINES ETC
LOCATIONS THAT ARE NOT NORMALLY USED BUT ARE ONLY OCCASSIONALLY USED
LOCATIONS WHICH ARE ESPECIALLY EFFECTIVE FOR STAFF EDUCATION
OTHER LOCATIONS WHICH ARE DIFFICULT TO CLEAN AND ARE SOURCES OF CONCERN
STEP 2
CHECK EXISTING CONDITIONS OF THE SAMPLING POINTS BEING
CONSIDERED FOR SWAB TESTS
USING LUCIPAC-W, CHECK THE CONDITIONS OF EACH POTENTIAL SAMPLING POINT
AFTER CLEANING
C 2001 Kikkoman Corporation
PD-10 HYGIENE MONITORING - 7
STEP 3
DECIDING ON ACTUAL SAMPLING POINTS FOR SWAB TESTS, RECONSIDERATION
OF CLEANING METHODS FOR PROBLEMATIC LOCATIONS.
Decide the actual sampling points for swab tests. Choose important places such as
filling nozzles or locations with high RLU readings (it is more practical to concentrate
on 20 sampling points).
Reconsidering washing methods for sampling points where RLU readings are always
high
Taking RLU readings again from sampling points where washing methods were
changed
STEP 4
DECIDING THE STANDARD VALUES FOR SAMPLING POINTS SELECTED
FOR SWAB TESTS.
It is practical to conduct swab tests on 20 sampling points in one day.
(Using the following table as reference , please create your own in-house standard values)
STANDARD VALUES (RLU)
1st Check Result 2nd Check Result
TEST POINTS Action
Pass Fail (RLU) (RLU)
Hands <3000 >6000 4000 B (CAUTION) Advise 1600 A (PASS)
Fridge door knob <2000 >4000 3000 B (CAUTION) Reclean 1500 A (PASS)
Sink <2000 >4000 640 A (PASS) Reclean 26 A (PASS)
Vat <200 >400 100 A (PASS)
Beer Server <2000 >4000 1500 A (PASS)
Cutting board <500 >1000 2500 C (FAIL) 1500 C (FAIL)
Knife <500 >1000 250 A (PASS)
Bowl <200 >400 250 (CAUTION)
Weighing scale <200 >400 1000 C (FAIL) Reclean 180 A (PASS)
STEP 5
ACTUAL DETERMINATION OF CLEANLINESS USING LUCIPAC-W
References:
ATP Swap Tests Research Group
The role and practice of ATP swap tests in HACCP
Food and Development Vol. 35 No. 8, P38-43
C 2001 Kikkoman Corporation