Characteristics of the Panasonic UD-8xx series Thermoluminescent Dosimeters
UD-802A Element 1 Element 2 Element 3 Element 4
n n
Phosphor Li2B4O7 Li2B4O7 CaSO4 CaSO4
2
Front Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm2 Plastic – 160 mg/ cm2 Lead – 0.7 mm
Rear Filtration Plastic – 14 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Lead – 0.7 mm
Remarks
The most commonly used Panasonic dosimeter in the US. Use of natural LiBO makes neutron dosimetry
possible if the appropriate energy correction factor for neutrons is known.
UD-802A2 Element 1 Element 2 Element 3 Element 4
n n
Phosphor Li2B4O7 Li2B4O7 CaSO4 CaSO4
2
Front Filtration Plastic – 14 mg/cm Plastic – 75 mg/cm2 Plastic – 160 mg/cm2 Lead – 0.7 mm
Rear Filtration Plastic – 14 mg/cm2 Plastic – 75 mg/cm2 Plastic – 160 mg/cm2 Lead – 0.7 mm
Remarks
The most commonly used Panasonic dosimeter in the US. Use of natural LiBO makes neutron dosimetry
possible if the appropriate energy correction factor for neutrons is known.
UD-804A Element 1 Element 2 Element 3 Element 4
Phosphor NONE CaSO4 CaSO4 CaSO4
Front Filtration Plastic – 14 mg/cm2 Lead – 0.7 mm Lead – 0.7 mm Lead – 0.7 mm
Rear Filtration Plastic – 14 mg/cm2 Lead – 0.7 mm Lead – 0.7 mm Lead – 0.7 mm
Remarks
Intended for environmental measurements. Contains no first element. Remaining three elements are
replicate CaSO4. Can measure low doses (about 10 mR/month). Addition of an element 1 of Li2B4O7 with
14 mg/cm2 makes environmental beta measurement possible. When E1 is added, the dosimeter is
designated as a UD-814.
UD-806A Element 1 Element 2 Element 3 Element 4
n n n n
Phosphor Li2B4O7 Li2B4O7 Li2B4O7 Li2B4O7
Front Filtration Plastic – 14 mg/cm2 Plastic – 160 mg/cm2
Plastic – 160 mg/cm2
Plastic – 160 mg/cm2
Rear Filtration Plastic – 14 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
Remarks
Four elements of LiBO
UD-807 Element 1 Element 2 Element 3 Element 4
n
Phosphor Li2B4O NONE NONE NONE
Front Filtration
Rear Filtration
Remarks
A single element of LiBO not set in a regular Panasonic dosimeter. Used for extremity monitoring. Must
be manually inserted into a special dosimeter in order to be read by the UD-702 manual TLD reader or by
the UD-710, UD-716, or UD-7900 automatic TLD readers.
UD-808A Element 1 Element 2 Element 3 Element 4
7
Phosphor Li211B4O7 7
Li211B4O7 CaSO4 7
Li211B4O7
2 2 2
Front Filtration Plastic – 14 mg/cm Plastic – 60 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2
2 2 2
Rear Filtration Plastic – 14 mg/cm Plastic – 60 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2
Remarks
Intended for use together with the UD-809. Measures beta particles and photons, but is insensitive to
neutrons. An algorithm must be used to separate photons from neutrons in the UD-809.
UD-809A Element 1 Element 2 Element 3 Element 4
7
Phosphor Li211B4O7 6
Li210B4O7 6
Li210B4O7 6
Li210B4O7
Front Filtration Cadmium – 0.7 mm Tin – 0.7 mm Cadmium – 0.7 mm Cadmium – 0.7 mm
Rear Filtration Cadmium – 0.7 mm Cadmium – 0.7 mm Cadmium – 0.7 mm Tin – 0.7 mm
Remarks
A primary neutron dosimeter, intended to be used with the UD-808. An algorithm is used to determine
the contribution of thermal, epithermal, and fast neutrons.
UD-810A1 Element 1 Element 2 Element 3 Element 4
7
Phosphor Li211B4O7 7
Li211B4O7 6
Li210B4O7 CaSO4
2
Front Filtration Plastic – 14 mg/cm Plastic – 160 mg/ cm2 Plastic – 160 mg/cm2 Lead – 0.7 mm
2
Rear Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Lead – 0.7 mm
Remarks
A specially designed dosimeter not intended for general use.
UD-810A2 Element 1 Element 2 Element 3 Element 4
7
Phosphor Li211B4O7 7
Li211B4O7 6
Li210B4O7 CaSO4
2
Front Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
2
Rear Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
Remarks
A specially designed dosimeter not intended for general use.
UD-812-A5 Element 1 Element 2 Element 3 Element 4
7
Phosphor Li211B4O7 7 7
Li211B4O7 CaSO4 CaSO4
2 2
Front Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2 Lead – 0.7 mm
2 2
Rear Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2 Lead – 0.7 mm
Remarks
Not a standard Panasonic dosimeter. Available for special design and use.
UD-812-A7 Element 1 Element 2 Element 3 Element 4
Phosphor CaSO4 CaSO4 CaSO4 CaSO4
Front Filtration Lead – 0.7 mm Lead – 0.7 mm Lead – 0.7 mm Lead – 0.7 mm
Rear Filtration Lead – 0.7 mm Lead – 0.7 mm Lead – 0.7 mm Lead – 0.7 mm
Remarks
Not a standard Panasonic dosimeter. Available for special design and use.
UD-812-A11 Element 1 Element 2 Element 3 Element 4
Phosphor CaSO4 CaSO4 CaSO4 CaSO4
Front Filtration Lead/Tin 0.7 mm Lead/Tin 0.7 mm Lead/Tin 0.7 mm Lead/Tin 0.7 mm
Rear Filtration Lead/Tin 0.7 mm Lead/Tin 0.7 mm Lead/Tin 0.7 mm Lead/Tin 0.7 mm
Remarks
Not a standard Panasonic dosimeter. Available for special design and use.
UD-812-A14 Element 1 Element 2 Element 3 Element 4
Phosphor CaSO4 CaSO4 CaSO4 CaSO4
2 2
Front Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2 Lead - 0.7 mm
2 2
Rear Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2 Lead - 0.7 mm
Remarks
Not a standard Panasonic dosimeter. Available for special design and use.
UD-813-A1 Element 1 Element 2 Element 3 Element 4
7
Phosphor Li211B4O7 7
Li211B4O7 n
Li2B4O7 CaSO4
2 2
Front Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2 Lead - 0.7 mm
Rear Filtration Plastic – 14 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Lead - 0.7 mm
Remarks
Not a standard Panasonic dosimeter. Available for special design and use.
UD-813-A4 Element 1 Element 2 Element 3 Element 4
6
Phosphor Li210B4O7 6
Li210B4O7 7
Li211B4O7 7
Li211B4O7
2 2
Front Filtration Plastic – 160 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
2 2
Rear Filtration Plastic – 160 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
Remarks
Not a standard Panasonic dosimeter. Available for special design and use.
UD-813-A6 Element 1 Element 2 Element 3 Element 4
6
Phosphor Li210B4O7 7
Li211B4O7 7
Li211B4O7 6
Li210B4O7
2 2
Front Filtration Plastic – 75 mg/cm Plastic – 75 mg/cm Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
2 2
Rear Filtration Plastic – 160 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
Remarks
Not a standard Panasonic dosimeter. Available for special design and use.
UD-813-A9 Element 1 Element 2 Element 3 Element 4
n n n
Phosphor Li2B4O7 Li2B4O7 CaSO4 Li2B4O7
2
Front Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 75 mg/cm2
Rear Filtration Plastic – 14 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
Remarks
Not a standard Panasonic dosimeter. Available for special design and use.
UD-813-A14 Element 1 Element 2 Element 3 Element 4
n n n
Phosphor Li2B4O7 Li2B4O7 CaSO4 Li2B4O7
2
Front Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
Rear Filtration Plastic – 14 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
Remarks
Not a standard Panasonic dosimeter. Available for special design and use.
UD-814-A1 Element 1 Element 2 Element 3 Element 4
n
Phosphor Li2B4O7 CaSO4 CaSO4 CaSO4
Front Filtration Plastic – 14 mg/cm2 Lead – 0.7 mm Lead – 0.7 mm Lead – 0.7 mm
Rear Filtration Plastic – 14 mg/cm2 Lead – 0.7 mm Lead – 0.7 mm Lead – 0.7 mm
Remarks
Not a standard Panasonic dosimeter. Available for special design and use. Often used like the UD-804
environmental dosimeter except with the addition of a LiBO element in position 1, encapsulated in 14
mg/cm2 plastic to enable monitoring for beta particles in the environment.
UD-814-A4 Element 1 Element 2 Element 3 Element 4
n n n
Phosphor Li2B4O7 Li2B4O7 CaSO4 Li2B4O7
2 2 2
Front Filtration Plastic – 14 mg/cm Plastic – 60 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2
Rear Filtration Plastic – 14 mg/cm2 Plastic – 60 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
Remarks
Not a standard Panasonic dosimeter. Available for special design and use. Often used like the UD-804
environmental dosimeter except with the addition of a LiBO element in position 1, encapsulated in 14 mg/
cm2 plastic to enable monitoring for beta particles in the environment.
UD-814-A6 Element 1 Element 2 Element 3 Element 4
7
Phosphor Li211B4O7 7
Li211B4O7 7
Li211B4O7 CaSO4
2 2
Front Filtration Plastic – 14 mg/ cm Plastic – 160 mg/cm Plastic – 160 mg/cm2 Lead – 0.7 mm
Rear Filtration Plastic – 14 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Lead – 0.7 mm
Remarks
Not a standard Panasonic dosimeter. Available for special design and use. Often used like the UD-804
environmental dosimeter except with the addition of a LiBO element in position 1, encapsulated in 14 mg/
cm2 plastic to enable monitoring for beta particles in the environment.
UD-814-A9 Element 1 Element 2 Element 3 Element 4
7
Phosphor Li211B4O7 7
Li211B4O7 7
Li211B4O7 6
Li210B4O7
2 2 2
Front Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2
2 2 2
Rear Filtration Plastic – 14 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm Plastic – 160 mg/cm2
Remarks
Not a standard Panasonic dosimeter. Available for special design and use. Often used like the UD-804
environmental dosimeter except with the addition of a LiBO element in position 1, encapsulated in 14
mg/cm2 plastic to enable monitoring for beta particles in the environment.
UD-815 Element 1 Element 2 Element 3 Element 4
Phosphor NONE NONE CaSO4 CaSO4
Front Filtration Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
Rear Filtration Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2 Plastic – 160 mg/cm2
Remarks
Originally intended for automatic calibration of the UD-710 automatic TLD reader. The design of this
dosimeter is similar to the design of the UD-811 dosimeter except for encapsulation. Most users prefer to
use the same type dosimeter for calibration as is used for routine monitoring of personnel and/or the
environment.