Health Santé

Canada Canada

Radiation Protection
and Safety for
Industrial X-ray Equipment

Safety Code 34
Errata

Appendix I, page 52:

c. bears the words "CAUTION, X-RAYS: NO UNAUTHORIZED


USE" and « ATTENTION, RAYONS X : AUCUNE
UTILISATION NON AUTORISÉE »; and
Radiation Protection
and Safety for
Industrial X-ray Equipment

Safety Code 34

Safe Environments Programme


Healthy Environments and Consumer Safety Branch

03-HECS-333
Our mission is to help the people of Canada Foreword
maintain and improve their health. Ionizing radiations are increasingly used in a variety of applica-
Health Canada tions in medicine, research and industry because of their known bene-
fits to society. One important industrial application utilizes various
sources of ionizing radiations to investigate the integrity of structures
or components through radiographic images, a practice generally re-
ferred to as nondestructive radiography or industrial radiography. In-
dustrial radiography to date is an established practice that provides
benefits concurrent with radiation risks. The radiation protection ob-
jective, therefore, is to keep the risks as low as reasonably achievable
(ALARA) while maximizing benefits.
Guidelines for ionizing radiation protection have been in exis-
tence for more than 60 years and have been subject to numerous revi-
sions in that time. One authoritative international standard-setting
organization is the International Commission on Radiological Pro-
tection (ICRP). In 1990, the ICRP reaffirmed its system of radiologi-
cal protection aimed at reducing the radiation risks associated with
ionizing radiation use, and recommended lower dose limits for radia-
tion workers and the public(1). The objectives of that system with re-
spect to radiation exposure are twofold: to prevent the occurrence of
deterministic effects, and to reduce the incidence of stochastic ef-
Published by authority of the fects. Sharing these same goals, the International Atomic Energy
Minister of Health Agency (IAEA) revised and updated its guidance documents(2,3) that
not only reflect the 1990 recommendations of the ICRP, but also aim
to lower ionizing radiation risks while not limiting the beneficial uses
of ionizing radiation. The ICRP and IAEA publications are influen-
Également disponible en français sous le titre : tial internationally and are used by a large number of countries to de-
Les appareils de radiologie industriels : velop national regulations for radiation protection and safety.
radioprotection et sécurité This document is one in a series of Safety Codes published by
Health Canada in accordance with Treasury Board Hazardous Sub-
stances Directive(4) to promote radiation protection and safety in the
federal workplace. This Safety Code draws on the ICRP and IAEA
This publication can be made available in/on computer diskette/large objectives, concepts and recommendations, and provides advisory
print/audio-cassette/braille, upon request. information and guidance on radiation protection and safety specific
to industrial x-ray equipment. It supersedes Safety Code 27 and is in-
tended for federal facilities. Thus, facilities subject to the Canada La-
bor Code Part II, Occupational Health and Safety Legislation(5) shall
© Her Majesty the Queen in Right of Canada, 2003 comply with the requirements of this Safety Code.
Cat. H46-2/03-333E
ISBN 0-662-35570-9
This Safety Code comprises an Introduction, the Management of
Radiation Protection and Safety, Requirements for Industrial X-ray
Equipment, Additional Guidance, References, Glossary and Appen-
dices. The Introduction provides a background on industrial x-ray

3
Our mission is to help the people of Canada Foreword
maintain and improve their health. Ionizing radiations are increasingly used in a variety of applica-
Health Canada tions in medicine, research and industry because of their known bene-
fits to society. One important industrial application utilizes various
sources of ionizing radiations to investigate the integrity of structures
or components through radiographic images, a practice generally re-
ferred to as nondestructive radiography or industrial radiography. In-
dustrial radiography to date is an established practice that provides
benefits concurrent with radiation risks. The radiation protection ob-
jective, therefore, is to keep the risks as low as reasonably achievable
(ALARA) while maximizing benefits.
Guidelines for ionizing radiation protection have been in exis-
tence for more than 60 years and have been subject to numerous revi-
sions in that time. One authoritative international standard-setting
organization is the International Commission on Radiological Pro-
tection (ICRP). In 1990, the ICRP reaffirmed its system of radiologi-
cal protection aimed at reducing the radiation risks associated with
ionizing radiation use, and recommended lower dose limits for radia-
tion workers and the public(1). The objectives of that system with re-
spect to radiation exposure are twofold: to prevent the occurrence of
deterministic effects, and to reduce the incidence of stochastic ef-
Published by authority of the fects. Sharing these same goals, the International Atomic Energy
Minister of Health Agency (IAEA) revised and updated its guidance documents(2,3) that
not only reflect the 1990 recommendations of the ICRP, but also aim
to lower ionizing radiation risks while not limiting the beneficial uses
of ionizing radiation. The ICRP and IAEA publications are influen-
Également disponible en français sous le titre : tial internationally and are used by a large number of countries to de-
Les appareils de radiologie industriels : velop national regulations for radiation protection and safety.
radioprotection et sécurité This document is one in a series of Safety Codes published by
Health Canada in accordance with Treasury Board Hazardous Sub-
stances Directive(4) to promote radiation protection and safety in the
federal workplace. This Safety Code draws on the ICRP and IAEA
This publication can be made available in/on computer diskette/large objectives, concepts and recommendations, and provides advisory
print/audio-cassette/braille, upon request. information and guidance on radiation protection and safety specific
to industrial x-ray equipment. It supersedes Safety Code 27 and is in-
tended for federal facilities. Thus, facilities subject to the Canada La-
bor Code Part II, Occupational Health and Safety Legislation(5) shall
© Her Majesty the Queen in Right of Canada, 2003 comply with the requirements of this Safety Code.
Cat. H46-2/03-333E
ISBN 0-662-35570-9
This Safety Code comprises an Introduction, the Management of
Radiation Protection and Safety, Requirements for Industrial X-ray
Equipment, Additional Guidance, References, Glossary and Appen-
dices. The Introduction provides a background on industrial x-ray

3
equipment and radiation hazards, and indicates the intent and scope Canadian Police College
of the Safety Code. The Management of Radiation Protection and Cancer Care Manitoba, Radiation Protection Service
Safety Section specifies the responsibilities of stakeholders which CanSpec Group Inc.
are key in the management of radiation protection and safety relevant Conan Quantum Inspection and testing (Division of
to industrial x-ray equipment. The section on Requirements for in- Stavely Services Canada Inc.)
dustrial X-ray Equipment specifies what is needed for regulatory Department of National Defence
compliance of equipment, for permanent and temporary work sites, Health Canada
and for on-site evaluations. Additional Guidance information is in- Insight Inspections Inc.
cluded for non-radiography use of industrial x-ray equipment, per- Labcan(1989) Ltée
sonnel monitoring, survey meters, emergency procedures, resale and Ontario Ministry of Labour, Radiation Protection Service
disposal. References are provided as well as a Glossary and Appen- Pratt and Whitney Canada Ltd.
dices. The information and guidance provided in this Safety Code Rolls-Royce Canada Ltd.
would be of interest to regulatory authorities, industrial x-ray equip- Royal Canadian Mounted Police
ment manufacturers, owners, operators, organizations or
radiographers carrying out industrial radiography, and clients re-
sponsible for hiring such organizations.
This Safety Code may be adopted for use elsewhere. Facilities
under provincial or territorial jurisdiction are advised to consult their
appropriate agency for information and guidance on radiation protec-
tion and safety, since statutes and requirements could differ from one
jurisdiction to the other. Relevant (provincial/territorial) contact in-
formation is available at a Health Canada website(6).
Given the diverse applications of industrial radiography, it is not
possible to foresee all work situations; therefore, it is plausible that
unexpected problems may occasionally occur and the guidance in
this Safety Code may be insufficient to address them. For such situa-
tions, consultation with the appropriate regulatory authority is ad-
vised.
This Safety Code was drafted, prepared and finalized by
H.P. Maharaj in accordance with Health Canada review and publish-
ing criteria. Interpretation of requirements or guidance in this docu-
ment should be directed to the attention of: Head, Nonmedical
X-Rays, PL 6301A, 775 Brookfield Road, Ottawa, Ontario, K1A 1C1
(Facsimile number: (613) 941-1734).
Acknowledgment is extended to the following contributors for
their comments and/or suggestions towards improvement and clarity
of this document:
Atlas Testing Laboratories and Services Ltd.
Atlantic NDT
Bowe and Landry NDT
British Columbia Centre for Disease Control
Canadian Nuclear Safety Commission

4 5
equipment and radiation hazards, and indicates the intent and scope Canadian Police College
of the Safety Code. The Management of Radiation Protection and Cancer Care Manitoba, Radiation Protection Service
Safety Section specifies the responsibilities of stakeholders which CanSpec Group Inc.
are key in the management of radiation protection and safety relevant Conan Quantum Inspection and testing (Division of
to industrial x-ray equipment. The section on Requirements for in- Stavely Services Canada Inc.)
dustrial X-ray Equipment specifies what is needed for regulatory Department of National Defence
compliance of equipment, for permanent and temporary work sites, Health Canada
and for on-site evaluations. Additional Guidance information is in- Insight Inspections Inc.
cluded for non-radiography use of industrial x-ray equipment, per- Labcan(1989) Ltée
sonnel monitoring, survey meters, emergency procedures, resale and Ontario Ministry of Labour, Radiation Protection Service
disposal. References are provided as well as a Glossary and Appen- Pratt and Whitney Canada Ltd.
dices. The information and guidance provided in this Safety Code Rolls-Royce Canada Ltd.
would be of interest to regulatory authorities, industrial x-ray equip- Royal Canadian Mounted Police
ment manufacturers, owners, operators, organizations or
radiographers carrying out industrial radiography, and clients re-
sponsible for hiring such organizations.
This Safety Code may be adopted for use elsewhere. Facilities
under provincial or territorial jurisdiction are advised to consult their
appropriate agency for information and guidance on radiation protec-
tion and safety, since statutes and requirements could differ from one
jurisdiction to the other. Relevant (provincial/territorial) contact in-
formation is available at a Health Canada website(6).
Given the diverse applications of industrial radiography, it is not
possible to foresee all work situations; therefore, it is plausible that
unexpected problems may occasionally occur and the guidance in
this Safety Code may be insufficient to address them. For such situa-
tions, consultation with the appropriate regulatory authority is ad-
vised.
This Safety Code was drafted, prepared and finalized by
H.P. Maharaj in accordance with Health Canada review and publish-
ing criteria. Interpretation of requirements or guidance in this docu-
ment should be directed to the attention of: Head, Nonmedical
X-Rays, PL 6301A, 775 Brookfield Road, Ottawa, Ontario, K1A 1C1
(Facsimile number: (613) 941-1734).
Acknowledgment is extended to the following contributors for
their comments and/or suggestions towards improvement and clarity
of this document:
Atlas Testing Laboratories and Services Ltd.
Atlantic NDT
Bowe and Landry NDT
British Columbia Centre for Disease Control
Canadian Nuclear Safety Commission

4 5
Table of Contents

1. Introduction 9
1.1 Background 9
1.2 Objective 11
1.3 Scope 12
1.4 Definitions 12

2. Management of Radiation
Protection and Safety:
Relevant Parties and Responsibilities 13
2.1 Regulatory authority 13
2.2 Industrial x-ray equipment owner 13
2.2.1 Radiation safety officer 14
2.3 Certified industrial radiographer 18
2.4 Manufacturer of industrial x-ray equipment 20
2.4.1 Service personnel 21
2.5 Clients hiring organizations or individuals
to perform radiography 22
2.6 Unauthorized individuals in radiography areas 23

3. Requirements for
Industrial X-ray Equipment 24
3.1 Industrial x-ray equipment design standards 24
3.2 Legal registration of industrial x-ray equipment
in federal facilities 27
3.3 Job site requirements for radiography 28
3.3.1 Permanent installation 29
3.3.2 Temporary job site 31
3.4 Facility on-site evaluation 33

4. Additional Guidance 34
4.1 Industrial radiography accelerators 34
4.2 Industrial radiography underwater 34
4.3 Non-radiography uses of industrial x-ray equipment 34
4.4 Personnel monitoring 35

7
4.5 Survey meter 37 1. Introduction
4.5.1 Information relevant to x-ray tube based systems
and electron beam welders 37
4.6 Emergency response for
unintentional radiation exposure 37
4.7 Resale of x-ray equipment 39
4.8 Disposal of x-ray equipment 40 1.1 Background
Radiation Equipment
References 41
Machines have been specifically designed to generate x rays for
the purpose of investigating the integrity of structures or components
Glossary 44
through radiographic images; this application is called nondestruc-
tive radiography or industrial radiography. Three types of ionizing
Appendix I : X-ray warning sign 52
radiation sources are typically available for such purposes: (1) Radio-
active materials that are gamma ray emitters such as iridium-192
Appendix II: Recommended dose limits 53 (192Ir), cesium-137(137Cs) and cobalt-60(60Co). (2) Neutrons that
Table 1. Effective dose limits 53 are produced in reactors or by other means (particle accelerators,
Table 2. Equivalent dose limits 53 radionuclides), and this application of neutrons is specifically re-
Table 3. Organ or tissue weighting factors 54 ferred to as neutron radiography. (3) X-ray tubes that are characteris-
Table 4. Radiation weighting factors 55 tic of conventional x-ray machines.
Industrial radiography machines which are x-ray tube based can
produce dose rates in air of about 2 Gy per minute at one metre(7).
They may be highly portable or mobile and convenient for use at tem-
porary job sites. Sometimes they can be operated by a single worker
in a wide range of conditions such as at aircraft hangers, pipeline con-
struction and deployment, fabrication facilities, offshore platform
operations, bridges, or construction sites. At temporary work sites,
the working conditions coupled with frequent manipulation of such
high-intensity radiation sources present much potential for radiation
exposure to occur. Both the worker(s) and other persons proximal to
the work area can be exposed to high radiation fields which, poten-
tially, can result in radiation accidents that could lead to serious per-
sonal injuries or death. In other work situations, x-ray tube based
devices may be installed in a shielded enclosure equipped with safety
components which significantly reduce radiation risks. To date in-
dustrial radiography is an established practice that provides benefits
concurrent with radiation risks.
In another industrial application there are systems specifically
designed to focus intense beams of high-energy electrons that melt
and bond metals under vacuum conditions, and these electron-metal
interactions can produce x rays as a byproduct of the bonding pro-
cess. Such systems are called electron beam welders. By design,

8 9
4.5 Survey meter 37 1. Introduction
4.5.1 Information relevant to x-ray tube based systems
and electron beam welders 37
4.6 Emergency response for
unintentional radiation exposure 37
4.7 Resale of x-ray equipment 39
4.8 Disposal of x-ray equipment 40 1.1 Background
Radiation Equipment
References 41
Machines have been specifically designed to generate x rays for
the purpose of investigating the integrity of structures or components
Glossary 44
through radiographic images; this application is called nondestruc-
tive radiography or industrial radiography. Three types of ionizing
Appendix I : X-ray warning sign 52
radiation sources are typically available for such purposes: (1) Radio-
active materials that are gamma ray emitters such as iridium-192
Appendix II: Recommended dose limits 53 (192Ir), cesium-137(137Cs) and cobalt-60(60Co). (2) Neutrons that
Table 1. Effective dose limits 53 are produced in reactors or by other means (particle accelerators,
Table 2. Equivalent dose limits 53 radionuclides), and this application of neutrons is specifically re-
Table 3. Organ or tissue weighting factors 54 ferred to as neutron radiography. (3) X-ray tubes that are characteris-
Table 4. Radiation weighting factors 55 tic of conventional x-ray machines.
Industrial radiography machines which are x-ray tube based can
produce dose rates in air of about 2 Gy per minute at one metre(7).
They may be highly portable or mobile and convenient for use at tem-
porary job sites. Sometimes they can be operated by a single worker
in a wide range of conditions such as at aircraft hangers, pipeline con-
struction and deployment, fabrication facilities, offshore platform
operations, bridges, or construction sites. At temporary work sites,
the working conditions coupled with frequent manipulation of such
high-intensity radiation sources present much potential for radiation
exposure to occur. Both the worker(s) and other persons proximal to
the work area can be exposed to high radiation fields which, poten-
tially, can result in radiation accidents that could lead to serious per-
sonal injuries or death. In other work situations, x-ray tube based
devices may be installed in a shielded enclosure equipped with safety
components which significantly reduce radiation risks. To date in-
dustrial radiography is an established practice that provides benefits
concurrent with radiation risks.
In another industrial application there are systems specifically
designed to focus intense beams of high-energy electrons that melt
and bond metals under vacuum conditions, and these electron-metal
interactions can produce x rays as a byproduct of the bonding pro-
cess. Such systems are called electron beam welders. By design,

8 9
electrons are emitted from heated filaments and accelerated to seve- survive the radiation insult, but are misrepaired and are able to un-
ral hundred kilovolts before impacting on metallic materials which dergo subsequent divisions. These cells, with the progression of time,
are generally placed in a vacuum. In some designs, however, the elec- may be transformed by external agents (e.g., chemicals, diet, radia-
tron beam passes through a series of orifices, each of which is indivi- tion exposure, lifestyle habits, etc.). After a latency period of years,
dually evacuated while the target material is positioned within close they may develop into leukemia or a solid tumor (cancer). Such latent
range (few centimeters) of the last orifice. Beam currents and high effects are called stochastic (or random).
voltages are typically in the range of 20-200 mA and 120- 450 kV, Germ cells are present in the ovaries and testes and are responsi-
respectively. A device operating at 150 kV and 50 mA, for example, ble for reproduction. Should they be modified by radiation, heredi-
would yield an estimated electron dose rate in air of ~ 0.5 Gy per se- tary effects may occur in the progeny of the individuals exposed to
cond at 50 cm, while the scattered x-ray radiation field at 1 metre radiation. Radiation-induced hereditary effects have not been ob-
would approximate 1 Gy per hour for a 1-cm primary electron beam served in human populations yet they have been demonstrated in ani-
incident on a copper target(8). The welding process requires the use of mals. Exposure of the embryo or fetus to ionizing radiation could
highly focused beams, a requirement that not only reduces the num- increase the risk of leukemia in infants and, during certain periods in
ber of electrons striking the metals being bonded but also lowers the early pregnancy, may lead to mental retardation and congenital mal-
contribution of byproduct x rays potentially scattered. The operation formations if the amount of radiation is sufficiently high.
of electron beam welders presents a potential for exposure to x rays Exposure to ionizing radiations has the potential to cause early or
and electrons. late adverse health effects. This is why the radiation risks associated
Collectively, in this document, x-ray tube based machines used with industrial x-ray equipment need to be managed.
for industrial radiography and electron beam welders are classed as
“industrial x-ray equipment.” They present potential risks of expo-
sure to x rays and electrons. The radiation protection objective, there- 1.2 Objective
fore, is to keep the risks ALARA while maximizing benefits. The objective of this Safety Code is to present information for
the radiation protection and safety of individuals operating, using and
Radiation hazard servicing industrial radiography x-ray equipment at permanent in-
X rays and electrons are types of ionizing radiation. In general stallations or at temporary job sites, and of persons proximal to such
when ionizing radiation traverses matter the interaction is probabilis- work areas. The owners of industrial x-ray equipment, the organiza-
tic, that is, there may or may not be an interaction. In the case of a me- tions or individuals carrying out industrial radiography, and clients
dium composed of cells of living organisms, the interaction with who hire such organizations or individuals are responsible for ensur-
individual cells may be direct or indirect. At the cellular level, direct ing that all safety procedures are followed and that the work is done in
interaction with DNA or other constituents can cause damage. In the a manner that does not pose undue risks to any person.
indirect mechanism, reactive ions are formed due to the breakdown of This document provides basic requirements and guidance in-
the water molecules present in the cells; such ions can interact with tended for the radiation protection and safety of industrial
any cellular constituent thereby leading to potential damage. Various radiographers, other users, service personnel and the public. It does
possibilities exist for the fate of cells exposed to ionizing radiation: not discuss industrial radiography techniques or electron beam weld-
(1) Damaged cells are completely repaired by the body’s inherent ing processes or other requirements (e.g., electrical or explosive).
repair mechanisms. (2) Damaged cells die during their attempt to re-
produce. Thus tissues and organs in which there is substantial cell
loss may become functionally impaired. There is a “threshold” dose
for each organ and tissue above which functional impairment will
manifest as a clinically observable adverse outcome. Exceeding the
threshold dose increases the level of harm. Such outcomes are called
deterministic effects and occur at high doses. (3) Damaged cells

10 11
electrons are emitted from heated filaments and accelerated to seve- survive the radiation insult, but are misrepaired and are able to un-
ral hundred kilovolts before impacting on metallic materials which dergo subsequent divisions. These cells, with the progression of time,
are generally placed in a vacuum. In some designs, however, the elec- may be transformed by external agents (e.g., chemicals, diet, radia-
tron beam passes through a series of orifices, each of which is indivi- tion exposure, lifestyle habits, etc.). After a latency period of years,
dually evacuated while the target material is positioned within close they may develop into leukemia or a solid tumor (cancer). Such latent
range (few centimeters) of the last orifice. Beam currents and high effects are called stochastic (or random).
voltages are typically in the range of 20-200 mA and 120- 450 kV, Germ cells are present in the ovaries and testes and are responsi-
respectively. A device operating at 150 kV and 50 mA, for example, ble for reproduction. Should they be modified by radiation, heredi-
would yield an estimated electron dose rate in air of ~ 0.5 Gy per se- tary effects may occur in the progeny of the individuals exposed to
cond at 50 cm, while the scattered x-ray radiation field at 1 metre radiation. Radiation-induced hereditary effects have not been ob-
would approximate 1 Gy per hour for a 1-cm primary electron beam served in human populations yet they have been demonstrated in ani-
incident on a copper target(8). The welding process requires the use of mals. Exposure of the embryo or fetus to ionizing radiation could
highly focused beams, a requirement that not only reduces the num- increase the risk of leukemia in infants and, during certain periods in
ber of electrons striking the metals being bonded but also lowers the early pregnancy, may lead to mental retardation and congenital mal-
contribution of byproduct x rays potentially scattered. The operation formations if the amount of radiation is sufficiently high.
of electron beam welders presents a potential for exposure to x rays Exposure to ionizing radiations has the potential to cause early or
and electrons. late adverse health effects. This is why the radiation risks associated
Collectively, in this document, x-ray tube based machines used with industrial x-ray equipment need to be managed.
for industrial radiography and electron beam welders are classed as
“industrial x-ray equipment.” They present potential risks of expo-
sure to x rays and electrons. The radiation protection objective, there- 1.2 Objective
fore, is to keep the risks ALARA while maximizing benefits. The objective of this Safety Code is to present information for
the radiation protection and safety of individuals operating, using and
Radiation hazard servicing industrial radiography x-ray equipment at permanent in-
X rays and electrons are types of ionizing radiation. In general stallations or at temporary job sites, and of persons proximal to such
when ionizing radiation traverses matter the interaction is probabilis- work areas. The owners of industrial x-ray equipment, the organiza-
tic, that is, there may or may not be an interaction. In the case of a me- tions or individuals carrying out industrial radiography, and clients
dium composed of cells of living organisms, the interaction with who hire such organizations or individuals are responsible for ensur-
individual cells may be direct or indirect. At the cellular level, direct ing that all safety procedures are followed and that the work is done in
interaction with DNA or other constituents can cause damage. In the a manner that does not pose undue risks to any person.
indirect mechanism, reactive ions are formed due to the breakdown of This document provides basic requirements and guidance in-
the water molecules present in the cells; such ions can interact with tended for the radiation protection and safety of industrial
any cellular constituent thereby leading to potential damage. Various radiographers, other users, service personnel and the public. It does
possibilities exist for the fate of cells exposed to ionizing radiation: not discuss industrial radiography techniques or electron beam weld-
(1) Damaged cells are completely repaired by the body’s inherent ing processes or other requirements (e.g., electrical or explosive).
repair mechanisms. (2) Damaged cells die during their attempt to re-
produce. Thus tissues and organs in which there is substantial cell
loss may become functionally impaired. There is a “threshold” dose
for each organ and tissue above which functional impairment will
manifest as a clinically observable adverse outcome. Exceeding the
threshold dose increases the level of harm. Such outcomes are called
deterministic effects and occur at high doses. (3) Damaged cells

10 11
1.3 Scope 2. Management of
This Safety Code applies specifically to industrial x-ray equip- Radiation Protection and Safety:
ment operating at energies up to 6 MeV for use in industrial radiogra- Relevant Parties and Responsibilities
phy and in material melting-and-bonding applications. It therefore
covers x-ray tube based equipment, electron beam welders and low
energy (#6 MV) accelerators.
It may have limited application in x-ray photon-based scanning
systems used for cargo surveillance purposes. The characteristics of
such systems include: (i) large enclosures or modular configurations Mitigation of radiation risks associated with the use of industrial
which comprise shielded walls or panels, which can facilitate human x-ray equipment requires the collective action of various stake-
intrusion, and which may be used in temporary work sites; (ii) intense holders: the regulatory authority, the equipment manufacturers,
x-ray beams produced by high voltage generators (greater than equipment owners, equipment operators, equipment servicing per-
450 kVp) or by accelerators; and (iii) various safety mechanisms to sonnel, organizations carrying out industrial radiography, clients
limit exposures to the operators and the public. responsible for hiring such organizations, and the public. The stake-
This Safety Code excludes: (i) cabinet x-ray systems; (ii) radio- holders’ responsibilities and functions are indicated below.
active-based equipment that emits gamma rays or neutrons for use in
industrial radiography; (iii) systems specifically designed for the pro-
duction of neutrons which may be used in industrial radiography; and
2.1 Regulatory Authority
(iv) facilities that utilize high-intensity gamma ray or electron The primary functions of the regulatory authority (Health
sources for food irradiation or material sterilization or material Canada) are: to develop standards and provide guidance; to verify, en-
property-modification purposes, applications which require a dose sure and enforce compliance with the standards and safety require-
on the order of kilogray. ments; and to conduct evaluations of industrial x-ray equipment and
of organizations carrying out radiography.
In accord with its health protection mandate, the regulatory au-
1.4 Definitions thority has the powers to enforce compliance with this Safety Code,
Some of the terms used in this document are defined in the to make or authorize measurements that would facilitate evaluation
glossary. for compliance with the requirements of this Safety Code or to re-
quire an operating organization or facility to modify or correct any
procedure, practice, system or component to ensure safety. It can
issue directives prohibiting use of industrial x-ray equipment, de-
pending on the hazards and risks presented by deviations of proce-
dures or by safety violations, and can annul the confirmation of
registration.

2.2 Industrial X-ray Equipment Owner


Only products that are compliant with the RED Act(9) shall be
procured. The owner of industrial x-ray equipment is ultimately
responsible for the radiation safety of that equipment, and for
ensuring that it is used in accordance with all legislation and regula-
tions, including conformance with other safety standards (e.g.,

12 13
1.3 Scope 2. Management of
This Safety Code applies specifically to industrial x-ray equip- Radiation Protection and Safety:
ment operating at energies up to 6 MeV for use in industrial radiogra- Relevant Parties and Responsibilities
phy and in material melting-and-bonding applications. It therefore
covers x-ray tube based equipment, electron beam welders and low
energy (#6 MV) accelerators.
It may have limited application in x-ray photon-based scanning
systems used for cargo surveillance purposes. The characteristics of
such systems include: (i) large enclosures or modular configurations Mitigation of radiation risks associated with the use of industrial
which comprise shielded walls or panels, which can facilitate human x-ray equipment requires the collective action of various stake-
intrusion, and which may be used in temporary work sites; (ii) intense holders: the regulatory authority, the equipment manufacturers,
x-ray beams produced by high voltage generators (greater than equipment owners, equipment operators, equipment servicing per-
450 kVp) or by accelerators; and (iii) various safety mechanisms to sonnel, organizations carrying out industrial radiography, clients
limit exposures to the operators and the public. responsible for hiring such organizations, and the public. The stake-
This Safety Code excludes: (i) cabinet x-ray systems; (ii) radio- holders’ responsibilities and functions are indicated below.
active-based equipment that emits gamma rays or neutrons for use in
industrial radiography; (iii) systems specifically designed for the pro-
duction of neutrons which may be used in industrial radiography; and
2.1 Regulatory Authority
(iv) facilities that utilize high-intensity gamma ray or electron The primary functions of the regulatory authority (Health
sources for food irradiation or material sterilization or material Canada) are: to develop standards and provide guidance; to verify, en-
property-modification purposes, applications which require a dose sure and enforce compliance with the standards and safety require-
on the order of kilogray. ments; and to conduct evaluations of industrial x-ray equipment and
of organizations carrying out radiography.
In accord with its health protection mandate, the regulatory au-
1.4 Definitions thority has the powers to enforce compliance with this Safety Code,
Some of the terms used in this document are defined in the to make or authorize measurements that would facilitate evaluation
glossary. for compliance with the requirements of this Safety Code or to re-
quire an operating organization or facility to modify or correct any
procedure, practice, system or component to ensure safety. It can
issue directives prohibiting use of industrial x-ray equipment, de-
pending on the hazards and risks presented by deviations of proce-
dures or by safety violations, and can annul the confirmation of
registration.

2.2 Industrial X-ray Equipment Owner


Only products that are compliant with the RED Act(9) shall be
procured. The owner of industrial x-ray equipment is ultimately
responsible for the radiation safety of that equipment, and for
ensuring that it is used in accordance with all legislation and regula-
tions, including conformance with other safety standards (e.g.,

12 13
electrical, explosive) that might be applicable for the safe use of the 1. To register all industrial x-ray equipment with the regulatory
equipment in the intended workplace. This responsibility requires a authority and obtain a valid confirmation of registration which
strong commitment to safety by the owner of the industrial x-ray must be posted conspicuously at that facility. For equipment in-
equipment, demonstrated by the establishment and use of a radiation stalled inside a permanent installation or stored in a designated
safety program. key-locked room, the wall beside the main access door to the
The owner shall appoint a Radiation Safety Officer (RSO) to inside of such a structure would suffice for posting the confirma-
oversee and implement the radiation safety program at each facility in tion of registration. Where a facility has portable industrial x-ray
which industrial radiography is carried out, and shall provide ade- equipment for use off site, the industrial radiographer tasked
quate financial and physical resources for the incumbent duties and with operating the equipment at the off-site location shall have
actions. A facility-designated RSO, by appointment to the position, appropriate documentation attesting that the equipment is duly
is empowered to undertake necessary supervision of radiation registered. Non-registered equipment shall not be used.
protection activities, stop unsafe practices, implement safety direc- 2. To prepare or review all safety, ALARA and radiation accident
tives and enforce prohibitions issued by the regulatory authority pur- or overexposure response procedures and submit them to the
suant to this Safety Code. An RSO shall be a certified industrial regulatory authority for review.
radiographer at the NDT Level 1, or higher, as described in Section
3. To ensure that all protective and safety equipment and materials
2.2.1.12. of this Safety Code.
required in the safety, ALARA, and radiation accident or
The owner of industrial x-ray equipment that is to be operated in
overexposure response procedures are available.
a facility under federal jurisdiction shall ensure that the equipment is
duly registered as described in Section 3.2 of this Safety Code and 4. To ensure that a program of regular maintenance is in place for all
that there is a valid confirmation of registration before the equipment industrial x-ray equipment used at the facility.
is used. A federal facility that operates industrial x-ray equipment 5. To provide written notification to the regulatory authority of in-
without a valid confirmation of registration is in contravention of leg- dustrial x-ray equipment that does not comply with the require-
islation(5). ments of Section 3.1 of this Safety Code in the event that
noncompliance was identified after the equipment has been in-
2.2.1 Radiation Safety Officer stalled or in use, and clearly detail the item(s) of noncompliance.
Prior to procurement of industrial x-ray equipment, the RSO 6. To ensure that the proposed design plans for a new permanent
must receive written confirmation from the equipment manufacturer installation or for modifications of an existing permanent instal-
or its authorized marketing agent that the product complies with the lation are reviewed by the regulatory authority, before any con-
RED Act (refer to Section 4.1 of this Safety Code). The onus is on the struction or modification begins.
equipment manufacturer to show proof of or demonstrate regulatory
7. To consult and liaise with the appropriate personnel responsible
compliance. Only products that are compliant with the RED Act shall
for the construction of new permanent installations or of modifi-
be procured. The RSO must ensure that (i) the industrial x-ray equip-
cations to existing installations in order to ensure compliance
ment and the facility(ies) in which it will be in use meet all applicable
with regulatory requirements and incorporation of recommenda-
radiation safety and regulatory requirements in this Safety Code, in-
tions of the regulatory authority.
cluding conformance with other applicable standards (e.g., electrical,
explosive); and (ii) industrial radiography is performed in accordance 8. To arrange for on-site evaluations associated with the
with safety procedures consistent with those of the practice. In a case pre-commissioning tests of permanent installations.
for which noncompliance is identified after the equipment has left the 9. To identify the controlled and supervised areas at the facility.
manufacturer’s premises, guidance to ensure regulatory compliance 10. To acquire a calibrated and operable survey meter at each facility
of the equipment is given in Section 2.4 of this Safety Code. where industrial x-ray equipment is in use. Such meters shall be
The specific and incumbent actions required of an RSO are: calibrated by a national calibration ionizing radiation laboratory

14 15
electrical, explosive) that might be applicable for the safe use of the 1. To register all industrial x-ray equipment with the regulatory
equipment in the intended workplace. This responsibility requires a authority and obtain a valid confirmation of registration which
strong commitment to safety by the owner of the industrial x-ray must be posted conspicuously at that facility. For equipment in-
equipment, demonstrated by the establishment and use of a radiation stalled inside a permanent installation or stored in a designated
safety program. key-locked room, the wall beside the main access door to the
The owner shall appoint a Radiation Safety Officer (RSO) to inside of such a structure would suffice for posting the confirma-
oversee and implement the radiation safety program at each facility in tion of registration. Where a facility has portable industrial x-ray
which industrial radiography is carried out, and shall provide ade- equipment for use off site, the industrial radiographer tasked
quate financial and physical resources for the incumbent duties and with operating the equipment at the off-site location shall have
actions. A facility-designated RSO, by appointment to the position, appropriate documentation attesting that the equipment is duly
is empowered to undertake necessary supervision of radiation registered. Non-registered equipment shall not be used.
protection activities, stop unsafe practices, implement safety direc- 2. To prepare or review all safety, ALARA and radiation accident
tives and enforce prohibitions issued by the regulatory authority pur- or overexposure response procedures and submit them to the
suant to this Safety Code. An RSO shall be a certified industrial regulatory authority for review.
radiographer at the NDT Level 1, or higher, as described in Section
3. To ensure that all protective and safety equipment and materials
2.2.1.12. of this Safety Code.
required in the safety, ALARA, and radiation accident or
The owner of industrial x-ray equipment that is to be operated in
overexposure response procedures are available.
a facility under federal jurisdiction shall ensure that the equipment is
duly registered as described in Section 3.2 of this Safety Code and 4. To ensure that a program of regular maintenance is in place for all
that there is a valid confirmation of registration before the equipment industrial x-ray equipment used at the facility.
is used. A federal facility that operates industrial x-ray equipment 5. To provide written notification to the regulatory authority of in-
without a valid confirmation of registration is in contravention of leg- dustrial x-ray equipment that does not comply with the require-
islation(5). ments of Section 3.1 of this Safety Code in the event that
noncompliance was identified after the equipment has been in-
2.2.1 Radiation Safety Officer stalled or in use, and clearly detail the item(s) of noncompliance.
Prior to procurement of industrial x-ray equipment, the RSO 6. To ensure that the proposed design plans for a new permanent
must receive written confirmation from the equipment manufacturer installation or for modifications of an existing permanent instal-
or its authorized marketing agent that the product complies with the lation are reviewed by the regulatory authority, before any con-
RED Act (refer to Section 4.1 of this Safety Code). The onus is on the struction or modification begins.
equipment manufacturer to show proof of or demonstrate regulatory
7. To consult and liaise with the appropriate personnel responsible
compliance. Only products that are compliant with the RED Act shall
for the construction of new permanent installations or of modifi-
be procured. The RSO must ensure that (i) the industrial x-ray equip-
cations to existing installations in order to ensure compliance
ment and the facility(ies) in which it will be in use meet all applicable
with regulatory requirements and incorporation of recommenda-
radiation safety and regulatory requirements in this Safety Code, in-
tions of the regulatory authority.
cluding conformance with other applicable standards (e.g., electrical,
explosive); and (ii) industrial radiography is performed in accordance 8. To arrange for on-site evaluations associated with the
with safety procedures consistent with those of the practice. In a case pre-commissioning tests of permanent installations.
for which noncompliance is identified after the equipment has left the 9. To identify the controlled and supervised areas at the facility.
manufacturer’s premises, guidance to ensure regulatory compliance 10. To acquire a calibrated and operable survey meter at each facility
of the equipment is given in Section 2.4 of this Safety Code. where industrial x-ray equipment is in use. Such meters shall be
The specific and incumbent actions required of an RSO are: calibrated by a national calibration ionizing radiation laboratory

14 15
(e.g., National Research Council, Ottawa, Canada; NIST, USA) pregnancy, and on the need for special restrictions to limit
or, a laboratory that has been accredited by a national laboratory exposure of female employees in the work area. (Consultation
to perform such calibrations. Calibrations shall be done at sev- with the regulatory authority is suggested.)
eral photon energies to cover the range of photon energies, gen- 17. To arrange for training, where applicable, and briefing of indus-
erated by the x-ray equipment that will be in use. Calibration trial radiographers, other users and allied personnel on radiation
frequency is once a year or after the meter has been serviced or safety issues or procedures, and maintain appropriate record
repaired. Records of the results of survey meter calibrations shall keeping of training material and a roster of trained personnel.
be kept at the facility. Each facility shall have a back up radiation
18. To supervise safety procedures for industrial radiography at tem-
protection survey meter that is also calibrated and functional as
porary job sites or delegate such tasks to a safety officer or an
indicated in this subsection.
equivalent alternative as described in Section 3.3.2.3. of this
11. To undertake reviews and modifications, as necessary, of opera- Safety Code.
tional procedures relevant to protection and safety to ensure radi-
19. To ensure that individuals training for industrial radiographer
ation exposure to industrial radiographers is ALARA and public
certification are supervised by an appropriately certified indus-
dose limits (in Appendix II, Tables 1 and 2 of this Safety Code)
trial radiographer, and are subject to all protection, safety and
are not exceeded.
dose monitoring procedures commensurate with those applica-
12. To ensure that any person operating industrial x-ray equipment ble to an industrial radiographer.
for industrial radiography purposes is an industrial radiographer,
20. To develop and maintain a system of personnel monitoring that
with certification at the NDT Level 1, or higher, in accordance
includes periodic reviews of personnel monitoring data, investi-
with the requirements of the Canadian General Standards
gation of inconsistencies, implementation of remedial actions in
Board(10). Persons operating industrial x-ray equipment specifi-
a timely manner, and retention of all such documents and re-
cally for police and security radiography must be certified at
cords.
Level I or higher in accordance with the Canadian General Stan-
dards Board requirements(11). 21. To ensure that there are available sufficient passive dosimeters
for the industrial radiographers, other users and allied personnel
13. To ensure that individuals operating electron beam welders and
as applicable, and that passive personnel dosimeters assigned to
other users receive radiation safety education as outlined in Sec-
an individual are not shared by others.
tion 4.3 of this Safety Code.
22. To acquire personnel dosimetry records from industrial
14. To ensure that both industrial radiographers and other users have
radiographers who are engaged in service contractual agree-
received practical training on the proper use of the specific x-ray
ments with multiple employers, then assess and ensure that their
equipment and of the survey meter(s) present at the facility, and
annual permissible occupational dose limit has not been ex-
have received instruction on the operational, ALARA, and safety
ceeded, prior to undertaking the industrial radiography work at
procedures applicable to the facility.
hand.
15. To conduct quarterly reviews of the industrial radiographer’s
23. To investigate radiation accidents or abnormal exposures or sus-
signed and dated log reflecting periodic safety checks done on
pected exposures to personnel and file an appropriate report with
the industrial x-ray equipment and its safety-related compo-
the regulatory authority within 5 calendar days of the incident.
nents, and on the safety accessories comprising the permanent
installation as described in Section 3.3.1. of this Safety Code. 24. To ensure that individuals accidentally exposed to ionizing radi-
ation receive proper medical attention.
16. To make decisions on reassignment of duties for declared or
confirmed pregnant industrial radiographers to ensure that the 25. To ensure that service personnel are qualified and authorized by
pregnancy dose limit is not exceeded for the duration of the the industrial x-ray equipment manufacturer to service the

16 17
(e.g., National Research Council, Ottawa, Canada; NIST, USA) pregnancy, and on the need for special restrictions to limit
or, a laboratory that has been accredited by a national laboratory exposure of female employees in the work area. (Consultation
to perform such calibrations. Calibrations shall be done at sev- with the regulatory authority is suggested.)
eral photon energies to cover the range of photon energies, gen- 17. To arrange for training, where applicable, and briefing of indus-
erated by the x-ray equipment that will be in use. Calibration trial radiographers, other users and allied personnel on radiation
frequency is once a year or after the meter has been serviced or safety issues or procedures, and maintain appropriate record
repaired. Records of the results of survey meter calibrations shall keeping of training material and a roster of trained personnel.
be kept at the facility. Each facility shall have a back up radiation
18. To supervise safety procedures for industrial radiography at tem-
protection survey meter that is also calibrated and functional as
porary job sites or delegate such tasks to a safety officer or an
indicated in this subsection.
equivalent alternative as described in Section 3.3.2.3. of this
11. To undertake reviews and modifications, as necessary, of opera- Safety Code.
tional procedures relevant to protection and safety to ensure radi-
19. To ensure that individuals training for industrial radiographer
ation exposure to industrial radiographers is ALARA and public
certification are supervised by an appropriately certified indus-
dose limits (in Appendix II, Tables 1 and 2 of this Safety Code)
trial radiographer, and are subject to all protection, safety and
are not exceeded.
dose monitoring procedures commensurate with those applica-
12. To ensure that any person operating industrial x-ray equipment ble to an industrial radiographer.
for industrial radiography purposes is an industrial radiographer,
20. To develop and maintain a system of personnel monitoring that
with certification at the NDT Level 1, or higher, in accordance
includes periodic reviews of personnel monitoring data, investi-
with the requirements of the Canadian General Standards
gation of inconsistencies, implementation of remedial actions in
Board(10). Persons operating industrial x-ray equipment specifi-
a timely manner, and retention of all such documents and re-
cally for police and security radiography must be certified at
cords.
Level I or higher in accordance with the Canadian General Stan-
dards Board requirements(11). 21. To ensure that there are available sufficient passive dosimeters
for the industrial radiographers, other users and allied personnel
13. To ensure that individuals operating electron beam welders and
as applicable, and that passive personnel dosimeters assigned to
other users receive radiation safety education as outlined in Sec-
an individual are not shared by others.
tion 4.3 of this Safety Code.
22. To acquire personnel dosimetry records from industrial
14. To ensure that both industrial radiographers and other users have
radiographers who are engaged in service contractual agree-
received practical training on the proper use of the specific x-ray
ments with multiple employers, then assess and ensure that their
equipment and of the survey meter(s) present at the facility, and
annual permissible occupational dose limit has not been ex-
have received instruction on the operational, ALARA, and safety
ceeded, prior to undertaking the industrial radiography work at
procedures applicable to the facility.
hand.
15. To conduct quarterly reviews of the industrial radiographer’s
23. To investigate radiation accidents or abnormal exposures or sus-
signed and dated log reflecting periodic safety checks done on
pected exposures to personnel and file an appropriate report with
the industrial x-ray equipment and its safety-related compo-
the regulatory authority within 5 calendar days of the incident.
nents, and on the safety accessories comprising the permanent
installation as described in Section 3.3.1. of this Safety Code. 24. To ensure that individuals accidentally exposed to ionizing radi-
ation receive proper medical attention.
16. To make decisions on reassignment of duties for declared or
confirmed pregnant industrial radiographers to ensure that the 25. To ensure that service personnel are qualified and authorized by
pregnancy dose limit is not exceeded for the duration of the the industrial x-ray equipment manufacturer to service the

16 17
equipment, and have read and understood all radiation safety 3. provide the facility RSO with a record of any radiation exposure
procedures applicable to the facility, before they can proceed history, especially if the industrial radiographer has been or is
with the impending servicing tasks. engaged in service contractual agreements with multiple em-
26. To ensure that persons proximal to industrial radiography work ployers at any time;
sites are briefed on the radiography work being done, the poten- 4. ensure that his or her annual permissible occupational equivalent
tial hazards and radiations risks, the applicable regulations, the dose limit (20 mSv, refer to Appendix II, Table 1 of this Safety
safety procedures that will be followed, the duration of the work, Code) has not been exceeded, prior to undertaking industrial
and are provided with contact information on the RSO to whom radiography at a facility;
ionizing radiation-related issues or concerns may be directed. 5. receive safety instructions and training specific to the industrial
(This action is perceived as promoting a positive safety culture in x-ray equipment and survey meter(s) to be used at the permanent
the workplace.) installation or temporary job site, and confirm with the facility
27. To initiate revisions of safety procedures and ensure that the RSO that this action has been taken;
facility is in compliance with the requirements of this Safety 6. keep radiation exposure to himself or herself and others
Code for as long as the facility uses, owns or stores industrial ALARA, and use one of the facility’s functional and calibrated
x-ray equipment. survey meter to ensure radiation levels are within dose limits that
28. To ensure that the industrial x-ray equipment is returned to the would not present an undue risk to any person;
manufacturer for disposal or, if the manufacturer is no longer in 7. wear (i) a personnel radiation dosimeter (passive dosimeter) sup-
business, is delivered to a company specifically contracted by the plied by a dosimetry service provider(12, 13) which is approved by
equipment owner to undertake the actions indicated in Section the Canadian Nuclear Safety Commission (CNSC) in accor-
4.8 of this Safety Code in order to ensure safe disposal. dance with CNSC regulations(14), and (ii) an electronic alarm do-
simeter (active dosimeter) which shall emit an audible signal
when the equivalent dose rate reaches or exceeds 5 mSv/h or
2.3 Certified Industrial Radiographer
where the total equivalent dose reaches or exceeds 2 mSv;
A certified industrial radiographer is an individual meeting the 8. not wear another individual’s passive radiation dosimeter;
requirements as described in Section 2.2.1.12. of this Safety Code.
9. perform preliminary checks on all safety devices (interlocks,
The individual has a responsibility to carry out the radiography work
warning lights, timers, shields, survey meters, personnel alarm
in accordance with instructions and defined operating and safety pro-
dosimeters, etc.) for proper functioning and resolve any inade-
cedures to ensure his or her protection and that of others. It is there-
quacies identified, before undertaking industrial radiography,
fore imperative that there be strict adherence of the safety
and maintain an appropriately signed and dated log of such
instructions, procedures and precautions provided in the operating
checks;
manual of the specific industrial x-ray equipment to be used. The
individual shall adopt an overall safety philosophy when doing indus- 10. check the inside of a permanent installation or controlled area to
trial radiography and, in this context, exercise vigilance with respect confirm that no person is present before initiating the generation
to the interlocks, alarms, warning indicators and signs. A certified of x-radiation;
industrial radiographer shall: 11. follow all protection and safety rules, including procedures
1. carry on his or her person appropriate proof of identification and established for the pending radiography work at the facility;
certification as an industrial radiographer whenever radiography 12. notify the facility RSO of any known or suspected abnormal
is being conducted; radiation exposure to any person, and maintain a record of such
2. provide the facility RSO with appropriate evidence affirming notification;
identification, certification and training;

18 19
equipment, and have read and understood all radiation safety 3. provide the facility RSO with a record of any radiation exposure
procedures applicable to the facility, before they can proceed history, especially if the industrial radiographer has been or is
with the impending servicing tasks. engaged in service contractual agreements with multiple em-
26. To ensure that persons proximal to industrial radiography work ployers at any time;
sites are briefed on the radiography work being done, the poten- 4. ensure that his or her annual permissible occupational equivalent
tial hazards and radiations risks, the applicable regulations, the dose limit (20 mSv, refer to Appendix II, Table 1 of this Safety
safety procedures that will be followed, the duration of the work, Code) has not been exceeded, prior to undertaking industrial
and are provided with contact information on the RSO to whom radiography at a facility;
ionizing radiation-related issues or concerns may be directed. 5. receive safety instructions and training specific to the industrial
(This action is perceived as promoting a positive safety culture in x-ray equipment and survey meter(s) to be used at the permanent
the workplace.) installation or temporary job site, and confirm with the facility
27. To initiate revisions of safety procedures and ensure that the RSO that this action has been taken;
facility is in compliance with the requirements of this Safety 6. keep radiation exposure to himself or herself and others
Code for as long as the facility uses, owns or stores industrial ALARA, and use one of the facility’s functional and calibrated
x-ray equipment. survey meter to ensure radiation levels are within dose limits that
28. To ensure that the industrial x-ray equipment is returned to the would not present an undue risk to any person;
manufacturer for disposal or, if the manufacturer is no longer in 7. wear (i) a personnel radiation dosimeter (passive dosimeter) sup-
business, is delivered to a company specifically contracted by the plied by a dosimetry service provider(12, 13) which is approved by
equipment owner to undertake the actions indicated in Section the Canadian Nuclear Safety Commission (CNSC) in accor-
4.8 of this Safety Code in order to ensure safe disposal. dance with CNSC regulations(14), and (ii) an electronic alarm do-
simeter (active dosimeter) which shall emit an audible signal
when the equivalent dose rate reaches or exceeds 5 mSv/h or
2.3 Certified Industrial Radiographer
where the total equivalent dose reaches or exceeds 2 mSv;
A certified industrial radiographer is an individual meeting the 8. not wear another individual’s passive radiation dosimeter;
requirements as described in Section 2.2.1.12. of this Safety Code.
9. perform preliminary checks on all safety devices (interlocks,
The individual has a responsibility to carry out the radiography work
warning lights, timers, shields, survey meters, personnel alarm
in accordance with instructions and defined operating and safety pro-
dosimeters, etc.) for proper functioning and resolve any inade-
cedures to ensure his or her protection and that of others. It is there-
quacies identified, before undertaking industrial radiography,
fore imperative that there be strict adherence of the safety
and maintain an appropriately signed and dated log of such
instructions, procedures and precautions provided in the operating
checks;
manual of the specific industrial x-ray equipment to be used. The
individual shall adopt an overall safety philosophy when doing indus- 10. check the inside of a permanent installation or controlled area to
trial radiography and, in this context, exercise vigilance with respect confirm that no person is present before initiating the generation
to the interlocks, alarms, warning indicators and signs. A certified of x-radiation;
industrial radiographer shall: 11. follow all protection and safety rules, including procedures
1. carry on his or her person appropriate proof of identification and established for the pending radiography work at the facility;
certification as an industrial radiographer whenever radiography 12. notify the facility RSO of any known or suspected abnormal
is being conducted; radiation exposure to any person, and maintain a record of such
2. provide the facility RSO with appropriate evidence affirming notification;
identification, certification and training;

18 19
13. supervise, as delegated by the owner of the industrial x-ray imported, leased or sold in Canada. To facilitate demonstration of
equipment in collaboration with the facility RSO, any individual product compliance, manufacturers of industrial x-ray equipment or
in training for industrial radiographer certification; their marketing agents shall:
14. exercise vigilance and check, at least once a month,: (i) the x-ray (i) provide written notification to the regulatory authority,
tube and cables for visible damage and wear, and (ii) the warning Health Canada, of the particular product intended for com-
labels and signs on the control console and radiation source merce in Canada;
assembly, including the exposure factors on the control console (ii) include with that written notification: a copy of the promo-
for legibility; tional literature, technical specifications, including instal-
lation, operational, safety, maintenance and disposal
15. not use any defective industrial x-ray equipment or survey
instructions in respect of that particular product, if these
meters or radiation dosimeters or shielding; and
materials have not been previously submitted to the regu-
16. where applicable, when industrial radiography work is to be latory authority; and
carried out at a client’s facility which is off site from the facility (iii) provide supportive evidence that the particular product
in which the industrial radiographer is registered, brief the client conforms with the requirements as specified in Section 3.1
as described in Section 2.5 of this Safety Code. of this Safety Code.
Due consideration shall be given to female industrial The regulatory authority may evaluate the product and provide a
radiographers of reproductive capacity. Female industrial radiog- response to the manufacturer accordingly. (Evaluation and responses
raphers, whether certified or in training, who are confirmed or may take place 4-6 weeks following receipt of all required documen-
declared pregnant shall discuss with the facility RSO their potential tation and supportive evidence from the manufacturer.) The importa-
occupational doses and radiation hazards for the remainder of the tion, lease or sale of industrial x-ray equipment that does not conform
pregnancy. (Should private or sensitive information about the radiog- with the RED Act is unlawful. Any violation of the RED Act is a crimi-
rapher’s pregnancy be divulged in the discussion, such information nal offence.
shall be treated in a confidential manner.) Having been made knowl- In a case for which noncompliance is revealed after procurement
edgeable of the potential occupational doses and radiation hazards or installation of the industrial x-ray equipment, the equipment man-
associated with the work, should the pregnant industrial radiographer ufacturer and the importer, upon written or verbal notification of the
choose to continue radiography work, personnel monitoring shall be noncompliance by the regulatory authority or by the equipment
implemented on a monthly basis and the badge readings shall be owner, shall take the necessary actions to bring the equipment into
promptly reviewed by the facility RSO to ensure that the pregnancy compliance within 30 days from the date of noncompliance notifica-
dose limit (in Appendix II, Table 1 of this Safety Code) is not ex- tion. In such situations, the equipment owner is further advised to
ceeded. Records of these actions shall be properly documented and cease using the equipment until the noncompliance issues are re-
retained by the facility RSO. No pregnant industrial radiographer solved to the satisfaction of the regulatory authority.
shall be re-assigned to other duties simply on the basis of information The manufacturer is solely responsible for providing training on
that the individual is pregnant. the servicing, safety and radiation protection specific to the industrial
x-ray equipment, and for issuing appropriately written authorizations
to individuals so trained to carry out servicing.
2.4 Manufacturer of
Industrial X-ray Equipment 2.4.1 Service Personnel
Manufacturers of industrial x-ray equipment shall acknowledge Personnel responsible for servicing industrial x-ray equipment
that such equipment is federally regulated in Canada (refer to Sec- in a facility or other workplace shall:
tion 3.1 of this Safety Code). It is the manufacturers’ responsibility to
ensure that their products comply with the RED Act(9) before they are

20 21
13. supervise, as delegated by the owner of the industrial x-ray imported, leased or sold in Canada. To facilitate demonstration of
equipment in collaboration with the facility RSO, any individual product compliance, manufacturers of industrial x-ray equipment or
in training for industrial radiographer certification; their marketing agents shall:
14. exercise vigilance and check, at least once a month,: (i) the x-ray (i) provide written notification to the regulatory authority,
tube and cables for visible damage and wear, and (ii) the warning Health Canada, of the particular product intended for com-
labels and signs on the control console and radiation source merce in Canada;
assembly, including the exposure factors on the control console (ii) include with that written notification: a copy of the promo-
for legibility; tional literature, technical specifications, including instal-
lation, operational, safety, maintenance and disposal
15. not use any defective industrial x-ray equipment or survey
instructions in respect of that particular product, if these
meters or radiation dosimeters or shielding; and
materials have not been previously submitted to the regu-
16. where applicable, when industrial radiography work is to be latory authority; and
carried out at a client’s facility which is off site from the facility (iii) provide supportive evidence that the particular product
in which the industrial radiographer is registered, brief the client conforms with the requirements as specified in Section 3.1
as described in Section 2.5 of this Safety Code. of this Safety Code.
Due consideration shall be given to female industrial The regulatory authority may evaluate the product and provide a
radiographers of reproductive capacity. Female industrial radiog- response to the manufacturer accordingly. (Evaluation and responses
raphers, whether certified or in training, who are confirmed or may take place 4-6 weeks following receipt of all required documen-
declared pregnant shall discuss with the facility RSO their potential tation and supportive evidence from the manufacturer.) The importa-
occupational doses and radiation hazards for the remainder of the tion, lease or sale of industrial x-ray equipment that does not conform
pregnancy. (Should private or sensitive information about the radiog- with the RED Act is unlawful. Any violation of the RED Act is a crimi-
rapher’s pregnancy be divulged in the discussion, such information nal offence.
shall be treated in a confidential manner.) Having been made knowl- In a case for which noncompliance is revealed after procurement
edgeable of the potential occupational doses and radiation hazards or installation of the industrial x-ray equipment, the equipment man-
associated with the work, should the pregnant industrial radiographer ufacturer and the importer, upon written or verbal notification of the
choose to continue radiography work, personnel monitoring shall be noncompliance by the regulatory authority or by the equipment
implemented on a monthly basis and the badge readings shall be owner, shall take the necessary actions to bring the equipment into
promptly reviewed by the facility RSO to ensure that the pregnancy compliance within 30 days from the date of noncompliance notifica-
dose limit (in Appendix II, Table 1 of this Safety Code) is not ex- tion. In such situations, the equipment owner is further advised to
ceeded. Records of these actions shall be properly documented and cease using the equipment until the noncompliance issues are re-
retained by the facility RSO. No pregnant industrial radiographer solved to the satisfaction of the regulatory authority.
shall be re-assigned to other duties simply on the basis of information The manufacturer is solely responsible for providing training on
that the individual is pregnant. the servicing, safety and radiation protection specific to the industrial
x-ray equipment, and for issuing appropriately written authorizations
to individuals so trained to carry out servicing.
2.4 Manufacturer of
Industrial X-ray Equipment 2.4.1 Service Personnel
Manufacturers of industrial x-ray equipment shall acknowledge Personnel responsible for servicing industrial x-ray equipment
that such equipment is federally regulated in Canada (refer to Sec- in a facility or other workplace shall:
tion 3.1 of this Safety Code). It is the manufacturers’ responsibility to
ensure that their products comply with the RED Act(9) before they are

20 21
1. provide proof to the facility RSO or workplace manager, as others on site. The client shall, therefore, designate a safety officer to
applicable, that they have received service training specific to the ensure that all safety procedures are being adhered to at the facility.
industrial x-ray equipment that is to be serviced; Where different industrial radiography jobs on site are required, the
2. provide written evidence to the facility RSO or workplace man- issuance of work permits in conjunction with client supervision is
ager, as applicable, that they are authorized by the industrial necessary to reduce the likelihood of radiation accidents or abnormal
x-ray equipment manufacturer to service the industrial x-ray exposures. Based on the foregoing, the client in conjunction with the
equipment; radiography service provider shall prepare and retain a document
(e.g., a checklist) which incorporates appropriate signatures and
3. follow all safety rules, including ALARA and accident response
dates in order to demonstrate that the client was briefed on the safety
procedures established at the facility;
issues relevant to the job, and is satisfied that appropriate safety ac-
4. wear personnel monitoring devices as specified in Section 2.3.7. tions were taken at the work site. A copy of all documents so prepared
of this Safety Code; shall be kept on file at the radiographer’s facility and at the client’s
5. acknowledge that, for the application of this Safety Code, ‘radia- premises where the radiography was carried out.
tion worker’ dose limits apply;
6. ensure that their annual permissible occupational dose limits of
2.6 Unauthorized Individuals in
20 mSv (Appendix II, Table 1 of this Safety Code) is not ex-
ceeded; and Industrial Radiography Areas
7. specify any unsafe conditions that could arise during operation Unauthorized individuals, other workers and staff in an organi-
of the industrial x-ray equipment in a written, signed and dated zation or facility, and members of the general public, who might be in
report which shall be submitted to the facility RSO or workplace areas where industrial radiography is being carried out must adhere to
manager as applicable. all safety warnings, instructions and notices posted in such areas, and
apply “common sense thinking and action” for their own protection
and safety. An individual who violates this ‘standard of care’ and in-
2.5 Clients Hiring Organizations or curs harm is deemed to have acted voluntarily in a negligent manner.
Individuals to Perform
Industrial Radiography
Before any radiography services are undertaken at a client’s
facility, the client shall be briefed on the industrial radiography, the
potential risks and hazards that are intrinsic to the work, the safety
procedures that will be followed for the client’s specific work re-
quired, applicable regulations, and proof of radiography personnel
certification and training. This briefing shall be done by the certified
industrial radiographer hired or authorized to perform the radio-
graphic services at the client’s premises, before any such services are
carried out.
The client shall acknowledge that contractual conditions must
not hinder the operating organization or individuals from performing
the radiographic work safely, that is, regulatory and safety require-
ments take precedence in industrial radiography. The client must en-
sure that industrial radiography is conducted in a safe manner to
minimize the risk of ionizing radiation exposure to the operators and

22 23
1. provide proof to the facility RSO or workplace manager, as others on site. The client shall, therefore, designate a safety officer to
applicable, that they have received service training specific to the ensure that all safety procedures are being adhered to at the facility.
industrial x-ray equipment that is to be serviced; Where different industrial radiography jobs on site are required, the
2. provide written evidence to the facility RSO or workplace man- issuance of work permits in conjunction with client supervision is
ager, as applicable, that they are authorized by the industrial necessary to reduce the likelihood of radiation accidents or abnormal
x-ray equipment manufacturer to service the industrial x-ray exposures. Based on the foregoing, the client in conjunction with the
equipment; radiography service provider shall prepare and retain a document
(e.g., a checklist) which incorporates appropriate signatures and
3. follow all safety rules, including ALARA and accident response
dates in order to demonstrate that the client was briefed on the safety
procedures established at the facility;
issues relevant to the job, and is satisfied that appropriate safety ac-
4. wear personnel monitoring devices as specified in Section 2.3.7. tions were taken at the work site. A copy of all documents so prepared
of this Safety Code; shall be kept on file at the radiographer’s facility and at the client’s
5. acknowledge that, for the application of this Safety Code, ‘radia- premises where the radiography was carried out.
tion worker’ dose limits apply;
6. ensure that their annual permissible occupational dose limits of
2.6 Unauthorized Individuals in
20 mSv (Appendix II, Table 1 of this Safety Code) is not ex-
ceeded; and Industrial Radiography Areas
7. specify any unsafe conditions that could arise during operation Unauthorized individuals, other workers and staff in an organi-
of the industrial x-ray equipment in a written, signed and dated zation or facility, and members of the general public, who might be in
report which shall be submitted to the facility RSO or workplace areas where industrial radiography is being carried out must adhere to
manager as applicable. all safety warnings, instructions and notices posted in such areas, and
apply “common sense thinking and action” for their own protection
and safety. An individual who violates this ‘standard of care’ and in-
2.5 Clients Hiring Organizations or curs harm is deemed to have acted voluntarily in a negligent manner.
Individuals to Perform
Industrial Radiography
Before any radiography services are undertaken at a client’s
facility, the client shall be briefed on the industrial radiography, the
potential risks and hazards that are intrinsic to the work, the safety
procedures that will be followed for the client’s specific work re-
quired, applicable regulations, and proof of radiography personnel
certification and training. This briefing shall be done by the certified
industrial radiographer hired or authorized to perform the radio-
graphic services at the client’s premises, before any such services are
carried out.
The client shall acknowledge that contractual conditions must
not hinder the operating organization or individuals from performing
the radiographic work safely, that is, regulatory and safety require-
ments take precedence in industrial radiography. The client must en-
sure that industrial radiography is conducted in a safe manner to
minimize the risk of ionizing radiation exposure to the operators and

22 23
3. Requirements for (iii) a red illuminated failsafe ‘x-ray on’ indicator;
(iv) indicators of the operational high voltage and beam cur-
Industrial X-ray Equipment rent in respect of the ionizing radiation source;
(v) a lock of a type that requires the insertion of a key before
ionizing radiation can be produced and for which the
removal of the key terminates the production of ionizing
radiation;
This part of the Safety Code outlines the regulatory require- (vi) a device that controls the duration of ionizing radiation
ments for industrial x-ray equipment, respecting design, compliance generation;
evaluation and enforcement; registration; and work site practice of (vii) a radiation ON/OFF control that requires activation by the
industrial radiography. operator for ionizing radiation production;
(viii) an emergency switch which, when activated, de-energizes
the ionizing radiation source;
3.1 Industrial X-ray Equipment (ix) appropriate provisions for the connections of interlock
Design Standards systems and of remotely positioned warning devices that
provide visible and audible warnings during ionizing radi-
X-ray equipment is federally regulated in Canada under the
ation generation;
Radiation Emitting Devices (RED) Act(9) and companion regula-
(x) a warning sign, next to the key lock, which (a) indicates
tions. The RED Act provides the authority to write regulations for
that hazardous ionizing radiation emissions are produced
specific classes of x-ray equipment. Where specific RED Act regula-
when the device is operating, and (b) prohibits unautho-
tions are not available for a class of x-ray equipment, the general
rized use;
provisions of the RED Act respecting prohibition, deception and noti-
(xi) a radiation warning sign that reflects the x-ray tube logo
fication apply within the scope of ensuring worker and public safety.
shown in Appendix I of this Safety Code; and
This is the case for industrial x-ray equipment. Thus, the importation,
(xii) an identification label, respecting the industrial x-ray
lease or sale of industrial x-ray equipment that does not conform with
equipment, which indicates the name and business address
the RED Act is contrary to Canadian law. It is the manufacturer’s and
of the manufacturer, the model number, the serial number,
the importer’s responsibility to ensure that the x-ray equipment avail-
the date of manufacture and the country of manufacture.
able for commerce in Canada complies with the RED Act. Any viola-
tion of the RED Act is a criminal offence. (2) A radiation-shielded assembly that contains the ionizing radia-
Re-sale industrial x-ray equipment shall also comply with the tion source and is equipped with a simple battery operated laser
RED Act at time of sale. The seller is responsible for ensuring regula- alignment device or a suitable alternative that aids radiography
tory compliance of the equipment, for bearing the associated costs of set-up procedures.
compliance, and for notifying the buyer of his or her safety obliga- (3) Labels affixed on the external surface of the radiation-shielded
tions upon acquisition of the equipment. Additional information is assembly that
provided in Section 4.7 of this Safety Code. (i) identify the electrical connections or means that facilitate
Industrial x-ray equipment that is designed and constructed to activation of audible and visible warning devices posi-
conform with the following minimum requirements would meet the tioned on the periphery of controlled areas during ionizing
general provisions of the RED Act: radiation generation at a temporary site;
(1) a control console that is equipped with: (ii) display the x-ray tube logo shown in Appendix I of this
(i) a power switch; Safety Code; and
(ii) an illuminated ‘power on’ indicator;

24 25
3. Requirements for (iii) a red illuminated failsafe ‘x-ray on’ indicator;
(iv) indicators of the operational high voltage and beam cur-
Industrial X-ray Equipment rent in respect of the ionizing radiation source;
(v) a lock of a type that requires the insertion of a key before
ionizing radiation can be produced and for which the
removal of the key terminates the production of ionizing
radiation;
This part of the Safety Code outlines the regulatory require- (vi) a device that controls the duration of ionizing radiation
ments for industrial x-ray equipment, respecting design, compliance generation;
evaluation and enforcement; registration; and work site practice of (vii) a radiation ON/OFF control that requires activation by the
industrial radiography. operator for ionizing radiation production;
(viii) an emergency switch which, when activated, de-energizes
the ionizing radiation source;
3.1 Industrial X-ray Equipment (ix) appropriate provisions for the connections of interlock
Design Standards systems and of remotely positioned warning devices that
provide visible and audible warnings during ionizing radi-
X-ray equipment is federally regulated in Canada under the
ation generation;
Radiation Emitting Devices (RED) Act(9) and companion regula-
(x) a warning sign, next to the key lock, which (a) indicates
tions. The RED Act provides the authority to write regulations for
that hazardous ionizing radiation emissions are produced
specific classes of x-ray equipment. Where specific RED Act regula-
when the device is operating, and (b) prohibits unautho-
tions are not available for a class of x-ray equipment, the general
rized use;
provisions of the RED Act respecting prohibition, deception and noti-
(xi) a radiation warning sign that reflects the x-ray tube logo
fication apply within the scope of ensuring worker and public safety.
shown in Appendix I of this Safety Code; and
This is the case for industrial x-ray equipment. Thus, the importation,
(xii) an identification label, respecting the industrial x-ray
lease or sale of industrial x-ray equipment that does not conform with
equipment, which indicates the name and business address
the RED Act is contrary to Canadian law. It is the manufacturer’s and
of the manufacturer, the model number, the serial number,
the importer’s responsibility to ensure that the x-ray equipment avail-
the date of manufacture and the country of manufacture.
able for commerce in Canada complies with the RED Act. Any viola-
tion of the RED Act is a criminal offence. (2) A radiation-shielded assembly that contains the ionizing radia-
Re-sale industrial x-ray equipment shall also comply with the tion source and is equipped with a simple battery operated laser
RED Act at time of sale. The seller is responsible for ensuring regula- alignment device or a suitable alternative that aids radiography
tory compliance of the equipment, for bearing the associated costs of set-up procedures.
compliance, and for notifying the buyer of his or her safety obliga- (3) Labels affixed on the external surface of the radiation-shielded
tions upon acquisition of the equipment. Additional information is assembly that
provided in Section 4.7 of this Safety Code. (i) identify the electrical connections or means that facilitate
Industrial x-ray equipment that is designed and constructed to activation of audible and visible warning devices posi-
conform with the following minimum requirements would meet the tioned on the periphery of controlled areas during ionizing
general provisions of the RED Act: radiation generation at a temporary site;
(1) a control console that is equipped with: (ii) display the x-ray tube logo shown in Appendix I of this
(i) a power switch; Safety Code; and
(ii) an illuminated ‘power on’ indicator;

24 25
(iii) identify the name and business address of the manufac- 3.2 Legal Registration of Industrial
turer of the industrial x-ray equipment, the equipment X-ray Equipment in Federal Facilities
model number, the serial number, the date of manufacture
and the country of manufacture. Any activity that might involve an assembled or ready-to-use in-
(4) All marks, labels and signs required on the control console and dustrial x-ray equipment carries an element of radiation risk. To en-
on the radiation-shielded assembly that are securely affixed and sure that the associated risks remain low, a system of control
are clearly visible. necessitating industrial x-ray equipment evaluation and operational
review must be in place. This requirement exists in the Canada
(5) All controls, lights, meters or other indicators on the control con- Labour Code Part II, Occupational Safety and Health Regulations,
sole and on the radiation-shielded assembly that are clearly Part X, Section 10.26(5): essentially, all federal departments and
marked as to function and conform to the following illumination agencies under federal jurisdiction must register their x-ray equip-
and colour codes: ment with the regulatory authority, Health Canada.
State or condition Colour The registration process involves:
Radiation ON Red (i) a submission of (a) a filled-in registration form that re-
Emergency Red quires details on various elements (e.g., the facility, x-ray
Warning (stand-by) Amber or yellow equipment, trained personnel, maintenance provider, per-
Radiation OFF (safe) Green sonnel dosimetry services, radiation survey meters), and
Information Blue (b) supportive documentation on emergency, safety and
(6) Installation, operational, safety, maintenance, and disposal in- disposal procedures;
structions that are provided by the manufacturer of the equip- (ii) an evaluation of the submitted material by the regulatory
ment such that, when followed, will enable the equipment to authority; and
operate within the standards of functioning and performance (iii) the issuance by the regulatory authority of a confirmation
specified for it, and to be disposed of in a safe and environmen- of registration valid for 3 years.
tally responsible manner. Where a confirmation of registration has been granted to a fed-
eral facility in accordance with Section 3.2.(iii) of this Safety Code
(7) Portable and mobile industrial x-ray equipment, in addition to and the facility subsequently makes a material change to the indus-
meeting all of the above requirements, that is designed to be trial radiography operation, safety procedures, permanent installa-
equipped with: tion or to the industrial x-ray equipment, a supplementary report that
(i) means (e.g., cables of sufficient length, radiofrequency clearly describes the change shall be submitted to the regulatory au-
mechanisms or other accessories) for enabling activation thority, Health Canada, for review within 14 calendar days of the said
of the radiation “ON/OFF” control from a location exter- change. In this context, industrial x-ray equipment that has been
nal to a controlled area that is established for temporary re-sold, removed from service or transferred to other users or to an-
sites in accordance with Section 3.3.2. of this Safety Code; other facility shall be reported. Failure to file such a supplementary
(ii) adequate support for the radiation-shielded assembly to report would automatically invalidate the confirmation of registra-
prevent tipping, drifting or vibration during operation of tion.
the equipment; and Re-registration shall occur every three years, unless specified
(iii) an automatic radiation warning alarm system either com- otherwise. A facility that operates industrial x-ray equipment without
prising a beacon light/siren device, or an alternative that a valid confirmation of registration would be in contravention of the
provides an equivalent function. federal Occupational Safety and Health Regulations(5).
All texts respecting radiation warning signs and labels required
in this part shall be in black on a yellow background and be written in
both English and French.

26 27
(iii) identify the name and business address of the manufac- 3.2 Legal Registration of Industrial
turer of the industrial x-ray equipment, the equipment X-ray Equipment in Federal Facilities
model number, the serial number, the date of manufacture
and the country of manufacture. Any activity that might involve an assembled or ready-to-use in-
(4) All marks, labels and signs required on the control console and dustrial x-ray equipment carries an element of radiation risk. To en-
on the radiation-shielded assembly that are securely affixed and sure that the associated risks remain low, a system of control
are clearly visible. necessitating industrial x-ray equipment evaluation and operational
review must be in place. This requirement exists in the Canada
(5) All controls, lights, meters or other indicators on the control con- Labour Code Part II, Occupational Safety and Health Regulations,
sole and on the radiation-shielded assembly that are clearly Part X, Section 10.26(5): essentially, all federal departments and
marked as to function and conform to the following illumination agencies under federal jurisdiction must register their x-ray equip-
and colour codes: ment with the regulatory authority, Health Canada.
State or condition Colour The registration process involves:
Radiation ON Red (i) a submission of (a) a filled-in registration form that re-
Emergency Red quires details on various elements (e.g., the facility, x-ray
Warning (stand-by) Amber or yellow equipment, trained personnel, maintenance provider, per-
Radiation OFF (safe) Green sonnel dosimetry services, radiation survey meters), and
Information Blue (b) supportive documentation on emergency, safety and
(6) Installation, operational, safety, maintenance, and disposal in- disposal procedures;
structions that are provided by the manufacturer of the equip- (ii) an evaluation of the submitted material by the regulatory
ment such that, when followed, will enable the equipment to authority; and
operate within the standards of functioning and performance (iii) the issuance by the regulatory authority of a confirmation
specified for it, and to be disposed of in a safe and environmen- of registration valid for 3 years.
tally responsible manner. Where a confirmation of registration has been granted to a fed-
eral facility in accordance with Section 3.2.(iii) of this Safety Code
(7) Portable and mobile industrial x-ray equipment, in addition to and the facility subsequently makes a material change to the indus-
meeting all of the above requirements, that is designed to be trial radiography operation, safety procedures, permanent installa-
equipped with: tion or to the industrial x-ray equipment, a supplementary report that
(i) means (e.g., cables of sufficient length, radiofrequency clearly describes the change shall be submitted to the regulatory au-
mechanisms or other accessories) for enabling activation thority, Health Canada, for review within 14 calendar days of the said
of the radiation “ON/OFF” control from a location exter- change. In this context, industrial x-ray equipment that has been
nal to a controlled area that is established for temporary re-sold, removed from service or transferred to other users or to an-
sites in accordance with Section 3.3.2. of this Safety Code; other facility shall be reported. Failure to file such a supplementary
(ii) adequate support for the radiation-shielded assembly to report would automatically invalidate the confirmation of registra-
prevent tipping, drifting or vibration during operation of tion.
the equipment; and Re-registration shall occur every three years, unless specified
(iii) an automatic radiation warning alarm system either com- otherwise. A facility that operates industrial x-ray equipment without
prising a beacon light/siren device, or an alternative that a valid confirmation of registration would be in contravention of the
provides an equivalent function. federal Occupational Safety and Health Regulations(5).
All texts respecting radiation warning signs and labels required
in this part shall be in black on a yellow background and be written in
both English and French.

26 27
3.3 Job Site Requirements for the primary beam using overhead cranes, robots or conveyor; to pro-
Industrial Radiography vide in the enclosure access for personnel entry and exit; to facilitate
passage of utility pipes and x-ray control or imaging cables; to affix
Radiography may be carried out in a permanent installation site lights; and to provide essential heating and ventilation duct systems.
and/or at a temporary job site. These sites are characterized by differ- Inadequate shielding at such discontinuities might lead to unaccept-
ences in work conditions and therefore hazard reduction strategies able levels of stray radiation, which potentially could expose person-
must integrate both the industrial x-ray equipment and human nel or the public, thereby increasing radiation risks. Guidance for
factors. effective shielding at such discontinuities is available(15). Though in-
frequent, operational demands may involve the replacement of an ex-
3.3.1 Permanent installation isting radiation source with one of a higher beam penetration
In the majority of cases industrial radiography is carried out in a capability (that is, higher energy x rays), and this may require
shielded enclosure. When a shielded enclosure is properly designed changes to the installation shielding or re-consideration of primary
and constructed using sound engineering principles and operated and secondary barriers or to occupancy of surrounding areas.
within the design limitations, no ‘Controlled Area’ external to it is Based on the foregoing, it is essential that a radiation protection
created. However, if conditions do not merit a controlled area, the oc- adviser be consulted (or hired) to advise on matters relevant to radia-
cupational exposure conditions need to be kept under review by the tion safety. The expert may be a consultant who is competent in the
RSO. This scenario would constitute a ‘Supervised Area’ for which field of ionizing radiation protection and safety and has several years
the following actions are required: (i) delineate the supervised area by experience in the design of ionizing radiation shielded facilities.
suitable means; (ii) display suitable and sufficient signs in appropri- However, responsibility for compliance with the relevant regulations
ate positions of the area to warn of the radiation sources and the asso- and standards remains with the organization or agency having owner-
ciated risks; and (iii) conduct periodic review of conditions to ship of the permanent installation.
determine a need for protective measures or safety provisions or The minimum requirements for a permanent installation are:
changes to the boundaries defining the area.
1. shielding of sufficient thickness that assures protection of indi-
The planning of a permanent installation should follow
viduals surrounding the installation;
life-cycle management concepts. Though not exhaustive, a number
of factors need to be considered: siting, location, construction, com- 2. warning signs of the type shown in Appendix I of this Safety
missioning, operation, maintenance and decommissioning based on Code that are affixed on all doors or panels or openings which
present and foreseeable future needs. Drawings of the enclosure and provide partial- or whole-body entry into the permanent installa-
its surroundings should be prepared; entrances should be identified; tion;
and dimensions and personnel occupancy areas need to be specified 3. a radiation control station that (i) is suitably designed and posi-
as well. Heating, lighting, ventilation and utility requirements also tioned external to the permanent installation, (ii) contains the
merit attention. Shielding, interlocks, warning indicators, emergency ionizing radiation source control console, (iii) provides for the
activators, and radiation control systems and stations require in-depth operator direct visual or electronic surveillance of the interior of
consideration because of their importance to radiation safety. Gen- the permanent installation during radiography, and (iv) facili-
erally, limitation of external radiation exposure to personnel, due to tates real-time imaging display and assessment capabilities;
the use of ionizing radiation sources present inside the installation, 4. two independent interlocks that are affixed on the main door
requires shielding. Estimates of shielding thickness may be deter- which is nearest the control console and which provides
mined using established methods(15); users of this reference should whole-body entry to the permanent installation, and they shall be
ensure appropriate dose limits are used when determining shielding designed in such a manner that, when activated, they shall
estimates. Often, depending on the nature of the work or location promptly terminate x-ray production and require a manual reset
of the installation, penetration of the shielding will be necessary for at the control console to resume x-radiation generation;
various reasons, such as, to place or remove heavy test objects in

28 29
3.3 Job Site Requirements for the primary beam using overhead cranes, robots or conveyor; to pro-
Industrial Radiography vide in the enclosure access for personnel entry and exit; to facilitate
passage of utility pipes and x-ray control or imaging cables; to affix
Radiography may be carried out in a permanent installation site lights; and to provide essential heating and ventilation duct systems.
and/or at a temporary job site. These sites are characterized by differ- Inadequate shielding at such discontinuities might lead to unaccept-
ences in work conditions and therefore hazard reduction strategies able levels of stray radiation, which potentially could expose person-
must integrate both the industrial x-ray equipment and human nel or the public, thereby increasing radiation risks. Guidance for
factors. effective shielding at such discontinuities is available(15). Though in-
frequent, operational demands may involve the replacement of an ex-
3.3.1 Permanent installation isting radiation source with one of a higher beam penetration
In the majority of cases industrial radiography is carried out in a capability (that is, higher energy x rays), and this may require
shielded enclosure. When a shielded enclosure is properly designed changes to the installation shielding or re-consideration of primary
and constructed using sound engineering principles and operated and secondary barriers or to occupancy of surrounding areas.
within the design limitations, no ‘Controlled Area’ external to it is Based on the foregoing, it is essential that a radiation protection
created. However, if conditions do not merit a controlled area, the oc- adviser be consulted (or hired) to advise on matters relevant to radia-
cupational exposure conditions need to be kept under review by the tion safety. The expert may be a consultant who is competent in the
RSO. This scenario would constitute a ‘Supervised Area’ for which field of ionizing radiation protection and safety and has several years
the following actions are required: (i) delineate the supervised area by experience in the design of ionizing radiation shielded facilities.
suitable means; (ii) display suitable and sufficient signs in appropri- However, responsibility for compliance with the relevant regulations
ate positions of the area to warn of the radiation sources and the asso- and standards remains with the organization or agency having owner-
ciated risks; and (iii) conduct periodic review of conditions to ship of the permanent installation.
determine a need for protective measures or safety provisions or The minimum requirements for a permanent installation are:
changes to the boundaries defining the area.
1. shielding of sufficient thickness that assures protection of indi-
The planning of a permanent installation should follow
viduals surrounding the installation;
life-cycle management concepts. Though not exhaustive, a number
of factors need to be considered: siting, location, construction, com- 2. warning signs of the type shown in Appendix I of this Safety
missioning, operation, maintenance and decommissioning based on Code that are affixed on all doors or panels or openings which
present and foreseeable future needs. Drawings of the enclosure and provide partial- or whole-body entry into the permanent installa-
its surroundings should be prepared; entrances should be identified; tion;
and dimensions and personnel occupancy areas need to be specified 3. a radiation control station that (i) is suitably designed and posi-
as well. Heating, lighting, ventilation and utility requirements also tioned external to the permanent installation, (ii) contains the
merit attention. Shielding, interlocks, warning indicators, emergency ionizing radiation source control console, (iii) provides for the
activators, and radiation control systems and stations require in-depth operator direct visual or electronic surveillance of the interior of
consideration because of their importance to radiation safety. Gen- the permanent installation during radiography, and (iv) facili-
erally, limitation of external radiation exposure to personnel, due to tates real-time imaging display and assessment capabilities;
the use of ionizing radiation sources present inside the installation, 4. two independent interlocks that are affixed on the main door
requires shielding. Estimates of shielding thickness may be deter- which is nearest the control console and which provides
mined using established methods(15); users of this reference should whole-body entry to the permanent installation, and they shall be
ensure appropriate dose limits are used when determining shielding designed in such a manner that, when activated, they shall
estimates. Often, depending on the nature of the work or location promptly terminate x-ray production and require a manual reset
of the installation, penetration of the shielding will be necessary for at the control console to resume x-radiation generation;
various reasons, such as, to place or remove heavy test objects in

28 29
5. one interlock, which trips a safety relay thereby removing power State or condition Colour
from the x-ray generator, shall be affixed on all remaining doors Radiation ON Red
that provide whole-body entry to the installation, including any Emergency Red
panel that permits partial-body entry to the permanent installa- Warning (stand-by) Amber or yellow
tion; Radiation OFF (safe) Green
6. clearly visible red ‘radiation-on’ warning indicators that are Information Blue
failsafe and that illuminate when x-radiation is generated: one All texts respecting radiation warning signs and labels required
indicator shall be prominently positioned inside the permanent in this part shall be in black on a yellow background and be written in
installation on a fixed vertical structure at a height of 2 metres both English and French. Before a newly constructed or modified
above the installation floor and, at least one indicator on the out- permanent installation is commissioned for use, an on-site evaluation
side of the permanent installation, near each entry door or panel must be conducted.
that provides access to the interior of that installation;
7. clearly visible yellow or amber ‘stand-by’ warning indicators 3.3.2 Temporary job site
that illuminate when ionizing radiation is not generated: one The use of engineering controls at temporary job sites is limited.
indicator shall be prominently positioned inside the permanent Consequently, strict adherence to administrative measures is neces-
installation on a fixed vertical structure at a height of 2 metres sary to reduce the likelihood of radiation accidents and unintentional
above the installation floor and, at least one indicator on the out- exposures. Individuals must therefore receive training and instruc-
side of the permanent installation, near each entry door or panel tions commensurate with the work procedures, and they must be
that provides access to the interior of the permanent installation; committed to safety. To the extent practical, radiography should be
performed at times when personnel occupancy of the ambient area is
8. inside the permanent installation, an audible warning signal that
low or zero. Use should be made of collimators and secondary shield-
(i) is distinct and loud enough to gain the attention of an individ-
ing (walls, shielded enclosures, vehicles unoccupied by humans,
ual, and (ii) is initiated for at least 5 seconds preceding ionizing
mobile lead screens), where practical, to reduce dose levels.
radiation generation;
The minimum requirements for radiography at a temporary job
9. clearly labelled red ‘emergency’ activators that (i) are installed site are:
inside the permanent installation at 1 metre above the floor in
1. A controlled area must be established to facilitate the radio-
such a way that they can be activated without passing through the
graphic work required. It shall be clearly designated by
primary beam, and (ii) function in a manner that when any such
(i) written warnings and safety instructions that are strategi-
activator is activated it shall: (a) promptly terminate ionizing
cally posted external to the controlled area to warn against
radiation generation; (b) automatically open the main door,
unauthorized entry; and
referred to in Section 3.3.1.4. of this Safety Code, which is
(ii) barriers that prevent entry into any area within which the
nearest the control console and which provides whole-body
equivalent dose rate exceeds 0.1 mSv per hour(14) as a re-
entry to the permanent installation; and (c) require a manual re-
sult of use of the industrial x-ray equipment.
set within the permanent installation, before ionizing radiation
generation can be resumed at the control console; and 2. Appropriate signals (visible from at least 10 metres) and audible
warnings (loudspeaker, horns) that alert unauthorized personnel
10. all controls, lights, warning signals or other indicators associated
of radiography in the designated controlled area.
with the permanent installation must: (i) be clearly labelled,
(ii) be imbedded with the appropriate warning text, and (iii) be 3. At least two individuals, authorized by the RSO, are required to
in conformance with the following illumination and colour monitor or patrol the boundary of the established controlled area
coding: and to ensure unauthorized individuals do not enter inside the
controlled area during ionizing radiation generation. One of the
authorized personnel shall be a certified industrial radiographer

30 31
5. one interlock, which trips a safety relay thereby removing power State or condition Colour
from the x-ray generator, shall be affixed on all remaining doors Radiation ON Red
that provide whole-body entry to the installation, including any Emergency Red
panel that permits partial-body entry to the permanent installa- Warning (stand-by) Amber or yellow
tion; Radiation OFF (safe) Green
6. clearly visible red ‘radiation-on’ warning indicators that are Information Blue
failsafe and that illuminate when x-radiation is generated: one All texts respecting radiation warning signs and labels required
indicator shall be prominently positioned inside the permanent in this part shall be in black on a yellow background and be written in
installation on a fixed vertical structure at a height of 2 metres both English and French. Before a newly constructed or modified
above the installation floor and, at least one indicator on the out- permanent installation is commissioned for use, an on-site evaluation
side of the permanent installation, near each entry door or panel must be conducted.
that provides access to the interior of that installation;
7. clearly visible yellow or amber ‘stand-by’ warning indicators 3.3.2 Temporary job site
that illuminate when ionizing radiation is not generated: one The use of engineering controls at temporary job sites is limited.
indicator shall be prominently positioned inside the permanent Consequently, strict adherence to administrative measures is neces-
installation on a fixed vertical structure at a height of 2 metres sary to reduce the likelihood of radiation accidents and unintentional
above the installation floor and, at least one indicator on the out- exposures. Individuals must therefore receive training and instruc-
side of the permanent installation, near each entry door or panel tions commensurate with the work procedures, and they must be
that provides access to the interior of the permanent installation; committed to safety. To the extent practical, radiography should be
performed at times when personnel occupancy of the ambient area is
8. inside the permanent installation, an audible warning signal that
low or zero. Use should be made of collimators and secondary shield-
(i) is distinct and loud enough to gain the attention of an individ-
ing (walls, shielded enclosures, vehicles unoccupied by humans,
ual, and (ii) is initiated for at least 5 seconds preceding ionizing
mobile lead screens), where practical, to reduce dose levels.
radiation generation;
The minimum requirements for radiography at a temporary job
9. clearly labelled red ‘emergency’ activators that (i) are installed site are:
inside the permanent installation at 1 metre above the floor in
1. A controlled area must be established to facilitate the radio-
such a way that they can be activated without passing through the
graphic work required. It shall be clearly designated by
primary beam, and (ii) function in a manner that when any such
(i) written warnings and safety instructions that are strategi-
activator is activated it shall: (a) promptly terminate ionizing
cally posted external to the controlled area to warn against
radiation generation; (b) automatically open the main door,
unauthorized entry; and
referred to in Section 3.3.1.4. of this Safety Code, which is
(ii) barriers that prevent entry into any area within which the
nearest the control console and which provides whole-body
equivalent dose rate exceeds 0.1 mSv per hour(14) as a re-
entry to the permanent installation; and (c) require a manual re-
sult of use of the industrial x-ray equipment.
set within the permanent installation, before ionizing radiation
generation can be resumed at the control console; and 2. Appropriate signals (visible from at least 10 metres) and audible
warnings (loudspeaker, horns) that alert unauthorized personnel
10. all controls, lights, warning signals or other indicators associated
of radiography in the designated controlled area.
with the permanent installation must: (i) be clearly labelled,
(ii) be imbedded with the appropriate warning text, and (iii) be 3. At least two individuals, authorized by the RSO, are required to
in conformance with the following illumination and colour monitor or patrol the boundary of the established controlled area
coding: and to ensure unauthorized individuals do not enter inside the
controlled area during ionizing radiation generation. One of the
authorized personnel shall be a certified industrial radiographer

30 31
and the other should be a safety officer or an equivalent designate All texts respecting radiation warning signs and labels required
who have read and understood the safety procedures relevant to in this part shall be in black on a yellow background and be written in
industrial radiography at temporary job sites. The RSO can also both English and French.
function as the other authorized individual.
4. A calibrated and functional survey meter(s), as outlined in Sec-
3.4 Facility On-site Evaluation
tion 2.2.1.10. of this Safety Code, shall be used to ensure that the
equivalent dose rate at the boundary of the designated controlled On-site evaluation is intended to ensure compliance with the
area, where authorized personnel might be present does not ex- requirements of this Safety Code. It must be conducted by the regula-
ceed 0.1 mSv per hour(14). tory authority; equivalents may be permitted provided that prior con-
5. Protocols that specify work set-up procedures, such as reposi- sent has been obtained from the regulatory authority. Such an
tioning of the radiography test object or the radiation source or its evaluation should confirm that
support assembly or the image receptor procedures, are com- (i) permanent installations are adequately constructed;
pleted before x-radiation is generated. (ii) safety systems and components function as intended;
(iii) warning signs are appropriate and are correctly deployed;
6. A temporary radiation control station that is positioned external
(iv) emergency procedures are clearly posted at the control
to the designated controlled area for the initiation, generation or
console and are understood by all industrial radiographers,
termination of ionizing radiation and for real-time image acqui-
other users and allied personnel associated with the indus-
sition and assessment. Every effort must be made to establish
trial radiography operation;
such a control station.
(v) stray radiation levels external to the permanent installation
7. The attendant industrial radiographer must have appropriate are within limits to ensure that no undue risks are pre-
documentation which substantiates that the portable x-ray sented;
equipment that is used on site has been duly registered. In cases (vi) radiographers, other users and allied personnel are trained
where portable industrial x-ray equipment registered at a spe- and are competent to use the radiation survey meter(s);
cific facility (resident facility) is used elsewhere, the resident (vii) radiographers and other users are trained to conduct
facility RSO shall provide the attendant industrial radiographer checks that establish the proper functioning of safety com-
with written confirmation that the portable x-ray equipment has ponents associated with the ionizing radiation producing
been duly registered in conformity with the validated confirma- equipment or the permanent facility or the temporary job
tion of registration. site; and
8. All controls, lights, warning signals or other indicators associ- (viii) personnel monitoring and control are in accordance with
ated with industrial radiography at a temporary site must: (i) be the requirements of Sections 2.2.1.16.; 2.2.1. 20-24.; and
clearly labelled, (ii) be imbedded with the appropriate warning 2.3.6-8. of this Safety Code.
text, and (iii) be in conformance with the following illumination On-site evaluations may be conducted periodically or at any
and colour coding: time.
State or condition Colour
Radiation ON Red
Emergency Red
Warning (stand-by) Amber or yellow
Radiation OFF (safe) Green
Information Blue

32 33
and the other should be a safety officer or an equivalent designate All texts respecting radiation warning signs and labels required
who have read and understood the safety procedures relevant to in this part shall be in black on a yellow background and be written in
industrial radiography at temporary job sites. The RSO can also both English and French.
function as the other authorized individual.
4. A calibrated and functional survey meter(s), as outlined in Sec-
3.4 Facility On-site Evaluation
tion 2.2.1.10. of this Safety Code, shall be used to ensure that the
equivalent dose rate at the boundary of the designated controlled On-site evaluation is intended to ensure compliance with the
area, where authorized personnel might be present does not ex- requirements of this Safety Code. It must be conducted by the regula-
ceed 0.1 mSv per hour(14). tory authority; equivalents may be permitted provided that prior con-
5. Protocols that specify work set-up procedures, such as reposi- sent has been obtained from the regulatory authority. Such an
tioning of the radiography test object or the radiation source or its evaluation should confirm that
support assembly or the image receptor procedures, are com- (i) permanent installations are adequately constructed;
pleted before x-radiation is generated. (ii) safety systems and components function as intended;
(iii) warning signs are appropriate and are correctly deployed;
6. A temporary radiation control station that is positioned external
(iv) emergency procedures are clearly posted at the control
to the designated controlled area for the initiation, generation or
console and are understood by all industrial radiographers,
termination of ionizing radiation and for real-time image acqui-
other users and allied personnel associated with the indus-
sition and assessment. Every effort must be made to establish
trial radiography operation;
such a control station.
(v) stray radiation levels external to the permanent installation
7. The attendant industrial radiographer must have appropriate are within limits to ensure that no undue risks are pre-
documentation which substantiates that the portable x-ray sented;
equipment that is used on site has been duly registered. In cases (vi) radiographers, other users and allied personnel are trained
where portable industrial x-ray equipment registered at a spe- and are competent to use the radiation survey meter(s);
cific facility (resident facility) is used elsewhere, the resident (vii) radiographers and other users are trained to conduct
facility RSO shall provide the attendant industrial radiographer checks that establish the proper functioning of safety com-
with written confirmation that the portable x-ray equipment has ponents associated with the ionizing radiation producing
been duly registered in conformity with the validated confirma- equipment or the permanent facility or the temporary job
tion of registration. site; and
8. All controls, lights, warning signals or other indicators associ- (viii) personnel monitoring and control are in accordance with
ated with industrial radiography at a temporary site must: (i) be the requirements of Sections 2.2.1.16.; 2.2.1. 20-24.; and
clearly labelled, (ii) be imbedded with the appropriate warning 2.3.6-8. of this Safety Code.
text, and (iii) be in conformance with the following illumination On-site evaluations may be conducted periodically or at any
and colour coding: time.
State or condition Colour
Radiation ON Red
Emergency Red
Warning (stand-by) Amber or yellow
Radiation OFF (safe) Green
Information Blue

32 33
4. Additional Guidance for non-radiography purposes have the appropriate education, train-
ing and competence; and (b) determine the individuals’ suitability for
the particular job intended.
Radiation safety education should be consistent with that of a
Radiation Safety Course, such as the one administered by the NDE
Institute of Canada(19) or equivalent. The training and education
This section provides information for guidance purposes. should include: (i) basic atomic and radiation physics; (ii) a knowl-
edge of x-ray production and interaction of ionizing radiation with
matter; (iii) radiation detection and measurement methods, including
4.1 Industrial Radiography Accelerators survey meters; (iv) basic understanding of biological effects of ioniz-
Accelerators used for industrial radiography produce photons ing radiation; (v) personnel monitoring devices; (vi) fundamentals of
in the MeV range at high dose rates. For example, a 3 MV-Linatron radiation protection: time, distance, shielding; (vii) ICRP principles
accelerator can produce a dose rate of 3 Gy per minute at 1 metre and and applicable radiation dose limits to workers and the public;
a similar device operating at 9 MV can yield a dose rate of 30 Gy per (viii) applicable regulations and operational standards for the specific
minute at the same distance(16). industrial x-ray machine; (ix) instructions on the operation, radiation
High-energy (MeV) photons can interact with atomic nuclei, hazards and safety specific to the industrial x-ray machine to be used;
causing nuclei transformation and release of energy in the form of (x) discussion of the relevant radiation surveys and results specific to
photons or energetic particles or a combination of both. One such the industrial x-ray machine(s); and (xi) emergency procedures.
particle can be a neutron. The interaction process is called The RSO shall identify the subset of industrial x-ray equipment
photodisintegration. This reaction is energetically feasible whenever users who are engaged in non-radiography work, and ensure that they
the photon energy exceeds the binding energy of a proton or neutron have the necessary education and training and have demonstrated
in the atom nucleus. For materials heavier than hydrogen, except competence before undertaking the job in question. Except for the
beryllium, the photon threshold energy for such reactions is generally certification criterion, the requirements of Section 2.3 of this Safety
>5 MeV(17). For accelerators operating at <6 MV, the neutron yield Code shall apply to that subset of industrial x-ray machine users.
from photodisintegration reactions would be extremely low and
therefore would not be a concern as a radiation hazard(18).
4.4 Personnel Monitoring
Personnel monitoring devices are needed to record and control
4.2 Industrial Radiography Underwater whole-body exposures to ensure occupational limits are not exceeded
This procedure normally employs radioactive-based sources. in accordance with the ICRP recommendations (Appendix II of this
The procedure, practice or radiation source is beyond the scope of this Safety Code). For external dose monitoring purposes, devices should
Safety Code. be worn on the clothing closest to the body either at the waist or chest
level. Patented and state-of-the-art personnel monitoring sys-
tems, capable of registering and reporting dose levels as low as
4.3 Non-Radiography Uses of 0.01 mSv, are readily available(12); older technology also exists(13).
Industrial X-ray Equipment As a general guide to users of ionizing radiation sources, it is ex-
Industrial x-ray equipment may occasionally be used for pur- tremely important to notify your dosimetry service provider of the
poses other than industrial radiography. In a facility in which this is radiation sources you are using or likely to use, seek the appropri-
the case, it is the responsibility of the industrial x-ray machine owner ate passive dosimeters, and ensure that your occupational doses
in that facility to (a) ensure that individuals using the x-ray equipment reported, correctly reflect the contribution from the various
radiation sources you are using.

34 35
4. Additional Guidance for non-radiography purposes have the appropriate education, train-
ing and competence; and (b) determine the individuals’ suitability for
the particular job intended.
Radiation safety education should be consistent with that of a
Radiation Safety Course, such as the one administered by the NDE
Institute of Canada(19) or equivalent. The training and education
This section provides information for guidance purposes. should include: (i) basic atomic and radiation physics; (ii) a knowl-
edge of x-ray production and interaction of ionizing radiation with
matter; (iii) radiation detection and measurement methods, including
4.1 Industrial Radiography Accelerators survey meters; (iv) basic understanding of biological effects of ioniz-
Accelerators used for industrial radiography produce photons ing radiation; (v) personnel monitoring devices; (vi) fundamentals of
in the MeV range at high dose rates. For example, a 3 MV-Linatron radiation protection: time, distance, shielding; (vii) ICRP principles
accelerator can produce a dose rate of 3 Gy per minute at 1 metre and and applicable radiation dose limits to workers and the public;
a similar device operating at 9 MV can yield a dose rate of 30 Gy per (viii) applicable regulations and operational standards for the specific
minute at the same distance(16). industrial x-ray machine; (ix) instructions on the operation, radiation
High-energy (MeV) photons can interact with atomic nuclei, hazards and safety specific to the industrial x-ray machine to be used;
causing nuclei transformation and release of energy in the form of (x) discussion of the relevant radiation surveys and results specific to
photons or energetic particles or a combination of both. One such the industrial x-ray machine(s); and (xi) emergency procedures.
particle can be a neutron. The interaction process is called The RSO shall identify the subset of industrial x-ray equipment
photodisintegration. This reaction is energetically feasible whenever users who are engaged in non-radiography work, and ensure that they
the photon energy exceeds the binding energy of a proton or neutron have the necessary education and training and have demonstrated
in the atom nucleus. For materials heavier than hydrogen, except competence before undertaking the job in question. Except for the
beryllium, the photon threshold energy for such reactions is generally certification criterion, the requirements of Section 2.3 of this Safety
>5 MeV(17). For accelerators operating at <6 MV, the neutron yield Code shall apply to that subset of industrial x-ray machine users.
from photodisintegration reactions would be extremely low and
therefore would not be a concern as a radiation hazard(18).
4.4 Personnel Monitoring
Personnel monitoring devices are needed to record and control
4.2 Industrial Radiography Underwater whole-body exposures to ensure occupational limits are not exceeded
This procedure normally employs radioactive-based sources. in accordance with the ICRP recommendations (Appendix II of this
The procedure, practice or radiation source is beyond the scope of this Safety Code). For external dose monitoring purposes, devices should
Safety Code. be worn on the clothing closest to the body either at the waist or chest
level. Patented and state-of-the-art personnel monitoring sys-
tems, capable of registering and reporting dose levels as low as
4.3 Non-Radiography Uses of 0.01 mSv, are readily available(12); older technology also exists(13).
Industrial X-ray Equipment As a general guide to users of ionizing radiation sources, it is ex-
Industrial x-ray equipment may occasionally be used for pur- tremely important to notify your dosimetry service provider of the
poses other than industrial radiography. In a facility in which this is radiation sources you are using or likely to use, seek the appropri-
the case, it is the responsibility of the industrial x-ray machine owner ate passive dosimeters, and ensure that your occupational doses
in that facility to (a) ensure that individuals using the x-ray equipment reported, correctly reflect the contribution from the various
radiation sources you are using.

34 35
Industrial radiographers and other users must wear photon- 4.5 Survey Meter
sensitive passive dosimeters as well as instantaneous reading elec-
tronic alarm dosimeters. Each passive personal dosimeter must be 4.5.1 Information relevant to x-ray tube based systems
worn by only one individual. It is advised that passive detectors be and electron beam welders
stored in a secure, properly shielded location between periods of use To avoid underestimation of the exposure rate from industrial
to avoid registering exposures from extraneous sources. Personnel x-ray equipment, the cross-sectional area of the radiation beam must
monitoring data need to be retained as a permanent record and be be larger than the sensitive area of the survey meter detector. For
made readily available for review by industrial radiography person- routine survey work the meter should yield readings that are accurate
nel, other users of the equipment and the regulatory authority. to ± 20%.The energy response should be flat to within ± 20% over the
Computation of occupational doses for purposes of assessment desired photon energy range to be encountered. Survey meters that
against worker dose limits recommended by the ICRP (refer to utilize an ionization chamber must be calibrated over the energy
Appendix II in this Safety Code) must be based on the summation of range for which they are to be used; calibration factors shall be used
all exposures incurred by the individual from the ionizing radiation to convert the meter readings to appropriate dose values. A meter
sources used in industrial radiography. Photon-sensitive passive de- having several measurement ranges provides greater flexibility. A
tectors collectively record x-ray and gamma-ray components. Natu- meter that provides maximum scale readings in the range of 10 µSv/h
ral background radiation or medical radiation contributions are not to 10 mSv/h (or the equivalent in exposure or dose units) is suggested.
computed as occupational exposures. The detector response time must be short enough for an accurate ra-
An electronic personnel dosimeter not only provides direct dose diation measurement. The meter must not exhibit “fold-back” under
reading capability, but also is designed to emit an audible signal any circumstances. Fold-back is said to occur when an instrument is
intended to provide instantaneous feedback to its wearer about the exposed to an exposure (or dose) rate or cumulative exposure (or ab-
radiation conditions prevailing in an area. The alarm set point may be sorbed dose) in excess of its display range and shows a lower or zero
for a dose rate or an integrated dose. Electronic alarm dosimeters reading. Ideally, a meter that is designed against fold-back should
shall (i) be checked to ensure proper functioning before use; (ii) be set show an off-scale deflection or, in the case of a digital display, show
to give an alarm at a preset dose equivalent rate of 5 mSv/h or an inte- an overload condition when it is exposed to ionizing radiation fields
grated dose of 2 mSv(14), with an accuracy of ± 20 % of the true radia- that exceed the measurement range. In work areas where radio fre-
tion dose; (iii) be calibrated at periods not exceeding 12 months; and quency (RF) fields might be present ensure that the survey meter is
(iv) require special means to change the preset alarm function. Fol- not sensitive to the ambient RF fields (by wrapping the meter in
lowing activation of the alarm, the dosimeter wearer is obligated to metallic gauze).
exercise appropriate radiation control measures to reduce his or her Several web sites may be consulted regarding radiation protec-
occupational dose. Such dosimeters work reasonably well for con- tion instrumentation(21).
ventional x-ray (and gamma-ray) sources. However, there are radia-
tion sources that produce very high x-ray levels in a time frame in the 4.6 Emergency Response for
order of microseconds (e.g., flash x-ray sources(20)) and the elec-
tronic personnel dosimeter may not have sufficient time to respond
Unintentional Radiation Exposures
within that time frame; this sub-optimal dosimeter response could be Unintentional exposure to radiation may be caused by equip-
falsely interpreted to mean the absence of x rays. Thus, an alternative ment failure or human error or a combination of both. Radiation acci-
dosimeter that provides direct dose reading display capability should dent victims must receive prompt medical attention by a physician.
be used, provided that it has been tested to provide reliable results and In addition, the root cause of the incident must be investigated and
the wearer must be aware of any inherent limitations thereof. Direct remedial measures taken to prevent recurrence at the facility.
dose-reading dosimeter results shall be recorded for individual wear- To address such situations, the facility is responsible for develop-
ers. ing an emergency response plan and having the capabilities to

36 37
Industrial radiographers and other users must wear photon- 4.5 Survey Meter
sensitive passive dosimeters as well as instantaneous reading elec-
tronic alarm dosimeters. Each passive personal dosimeter must be 4.5.1 Information relevant to x-ray tube based systems
worn by only one individual. It is advised that passive detectors be and electron beam welders
stored in a secure, properly shielded location between periods of use To avoid underestimation of the exposure rate from industrial
to avoid registering exposures from extraneous sources. Personnel x-ray equipment, the cross-sectional area of the radiation beam must
monitoring data need to be retained as a permanent record and be be larger than the sensitive area of the survey meter detector. For
made readily available for review by industrial radiography person- routine survey work the meter should yield readings that are accurate
nel, other users of the equipment and the regulatory authority. to ± 20%.The energy response should be flat to within ± 20% over the
Computation of occupational doses for purposes of assessment desired photon energy range to be encountered. Survey meters that
against worker dose limits recommended by the ICRP (refer to utilize an ionization chamber must be calibrated over the energy
Appendix II in this Safety Code) must be based on the summation of range for which they are to be used; calibration factors shall be used
all exposures incurred by the individual from the ionizing radiation to convert the meter readings to appropriate dose values. A meter
sources used in industrial radiography. Photon-sensitive passive de- having several measurement ranges provides greater flexibility. A
tectors collectively record x-ray and gamma-ray components. Natu- meter that provides maximum scale readings in the range of 10 µSv/h
ral background radiation or medical radiation contributions are not to 10 mSv/h (or the equivalent in exposure or dose units) is suggested.
computed as occupational exposures. The detector response time must be short enough for an accurate ra-
An electronic personnel dosimeter not only provides direct dose diation measurement. The meter must not exhibit “fold-back” under
reading capability, but also is designed to emit an audible signal any circumstances. Fold-back is said to occur when an instrument is
intended to provide instantaneous feedback to its wearer about the exposed to an exposure (or dose) rate or cumulative exposure (or ab-
radiation conditions prevailing in an area. The alarm set point may be sorbed dose) in excess of its display range and shows a lower or zero
for a dose rate or an integrated dose. Electronic alarm dosimeters reading. Ideally, a meter that is designed against fold-back should
shall (i) be checked to ensure proper functioning before use; (ii) be set show an off-scale deflection or, in the case of a digital display, show
to give an alarm at a preset dose equivalent rate of 5 mSv/h or an inte- an overload condition when it is exposed to ionizing radiation fields
grated dose of 2 mSv(14), with an accuracy of ± 20 % of the true radia- that exceed the measurement range. In work areas where radio fre-
tion dose; (iii) be calibrated at periods not exceeding 12 months; and quency (RF) fields might be present ensure that the survey meter is
(iv) require special means to change the preset alarm function. Fol- not sensitive to the ambient RF fields (by wrapping the meter in
lowing activation of the alarm, the dosimeter wearer is obligated to metallic gauze).
exercise appropriate radiation control measures to reduce his or her Several web sites may be consulted regarding radiation protec-
occupational dose. Such dosimeters work reasonably well for con- tion instrumentation(21).
ventional x-ray (and gamma-ray) sources. However, there are radia-
tion sources that produce very high x-ray levels in a time frame in the 4.6 Emergency Response for
order of microseconds (e.g., flash x-ray sources(20)) and the elec-
tronic personnel dosimeter may not have sufficient time to respond
Unintentional Radiation Exposures
within that time frame; this sub-optimal dosimeter response could be Unintentional exposure to radiation may be caused by equip-
falsely interpreted to mean the absence of x rays. Thus, an alternative ment failure or human error or a combination of both. Radiation acci-
dosimeter that provides direct dose reading display capability should dent victims must receive prompt medical attention by a physician.
be used, provided that it has been tested to provide reliable results and In addition, the root cause of the incident must be investigated and
the wearer must be aware of any inherent limitations thereof. Direct remedial measures taken to prevent recurrence at the facility.
dose-reading dosimeter results shall be recorded for individual wear- To address such situations, the facility is responsible for develop-
ers. ing an emergency response plan and having the capabilities to

36 37
implement the plan. Personnel must be trained to handle emergency The facility shall prepare a written report that contains: a de-
equipment and to follow written procedures. The plan shall be tested scription of the accident; methods used to protect other workers and
and validated, and deficiencies shall be identified and corrected. The the public; assessments of exposures to the accident victims, workers,
facility needs to liaise with the various personnel identified in the emergency services personnel and members of the public; cause of
emergency procedures. the accident and corrective actions. That report shall be submitted (by
As a guide, the generic emergency response plan should include: the RSO) to the regulatory authority for review and follow up within
(1) Response initiator – a person who initiates the response and per- 5 calendar days after the incident (see Section 2.2.1.23. of this Safety
forms actions to mitigate the accident at the scene. Usually this is Code).
the attendant radiographer or the first responder on the scene.
(2) Emergency response manager – a person who is in charge of the 4.7 Resale of X-ray Equipment
overall plan, manages the priorities and ensures protection of
other workers, emergency workers and the public. This person X-ray equipment intended for resale shall also comply with the
could be a safety officer or manager or senior staff member in the RED Act and regulations at time of sale. The seller is responsible for
facility. (i) ensuring regulatory compliance of the product and for
bearing the associated costs, and
(3) Radiological assessor – a person who is responsible for conduct-
(ii) notifying the purchaser that it is the purchaser’s responsi-
ing radiation surveys and dose assessment, and for providing
bility to make certain that the following requirements are
radiation protection support to emergency workers and advice to
met before the x-ray machine can be used:
the facility. This person is usually the RSO or a hired consultant
(a) the x-ray machine must be installed by trained and
with relevant expertise.
authorized maintenance personnel;
Emergency procedures should incorporate the following crite-
(b) the x-ray machine must be inspected by a radiation
ria:
safety inspector authorized by the federal regulatory
(i) be concise and easy to follow;
authority, Health Canada, and a corresponding report
(ii) include what situations are indicative of requiring emer-
prepared;
gency action;
(c) all operators must receive radiation safety training
(iii) specify the immediate action to be taken to deliver prompt
specific to the x-ray machine, prior to its use; and
medical attention to radiation accident victims or those
(d) the x-ray machine shall conform to the operating rules
suspect; it is advised that the attendant physician be a radi-
commensurate with those of the appropriate regula-
ation oncologist or be knowledgeable in the biological
tory authority that has jurisdiction of the facility in
effects of ionizing radiation on humans;
which the x-ray machine will be in use.
(iv) specify the immediate action to be taken to minimize radi-
If the purchaser’s facility is under federal jurisdiction, the
ation exposure to persons in the vicinity of the radiation
Nonmedical X-Ray Unit, Consumer and Clinical Radiation Protec-
source;
tion Bureau, Health Canada shall be contacted and the facility shall
(v) specify the names and telephone numbers of the RSO, the
adopt this Safety Code. Otherwise, the appropriate provincial or terri-
physician or relevant medical institution, the equipment
torial radiation protection authority shall be contacted to determine
manufacturer, the qualified expert, emergency services
the applicable operational requirements for the x-ray machine.
and the regulatory authority, and ensure that such informa-
tion is up to date;
(vi) notify the regulatory authority of the incident as soon as
possible, and provide accurate and complete information.

38 39
implement the plan. Personnel must be trained to handle emergency The facility shall prepare a written report that contains: a de-
equipment and to follow written procedures. The plan shall be tested scription of the accident; methods used to protect other workers and
and validated, and deficiencies shall be identified and corrected. The the public; assessments of exposures to the accident victims, workers,
facility needs to liaise with the various personnel identified in the emergency services personnel and members of the public; cause of
emergency procedures. the accident and corrective actions. That report shall be submitted (by
As a guide, the generic emergency response plan should include: the RSO) to the regulatory authority for review and follow up within
(1) Response initiator – a person who initiates the response and per- 5 calendar days after the incident (see Section 2.2.1.23. of this Safety
forms actions to mitigate the accident at the scene. Usually this is Code).
the attendant radiographer or the first responder on the scene.
(2) Emergency response manager – a person who is in charge of the 4.7 Resale of X-ray Equipment
overall plan, manages the priorities and ensures protection of
other workers, emergency workers and the public. This person X-ray equipment intended for resale shall also comply with the
could be a safety officer or manager or senior staff member in the RED Act and regulations at time of sale. The seller is responsible for
facility. (i) ensuring regulatory compliance of the product and for
bearing the associated costs, and
(3) Radiological assessor – a person who is responsible for conduct-
(ii) notifying the purchaser that it is the purchaser’s responsi-
ing radiation surveys and dose assessment, and for providing
bility to make certain that the following requirements are
radiation protection support to emergency workers and advice to
met before the x-ray machine can be used:
the facility. This person is usually the RSO or a hired consultant
(a) the x-ray machine must be installed by trained and
with relevant expertise.
authorized maintenance personnel;
Emergency procedures should incorporate the following crite-
(b) the x-ray machine must be inspected by a radiation
ria:
safety inspector authorized by the federal regulatory
(i) be concise and easy to follow;
authority, Health Canada, and a corresponding report
(ii) include what situations are indicative of requiring emer-
prepared;
gency action;
(c) all operators must receive radiation safety training
(iii) specify the immediate action to be taken to deliver prompt
specific to the x-ray machine, prior to its use; and
medical attention to radiation accident victims or those
(d) the x-ray machine shall conform to the operating rules
suspect; it is advised that the attendant physician be a radi-
commensurate with those of the appropriate regula-
ation oncologist or be knowledgeable in the biological
tory authority that has jurisdiction of the facility in
effects of ionizing radiation on humans;
which the x-ray machine will be in use.
(iv) specify the immediate action to be taken to minimize radi-
If the purchaser’s facility is under federal jurisdiction, the
ation exposure to persons in the vicinity of the radiation
Nonmedical X-Ray Unit, Consumer and Clinical Radiation Protec-
source;
tion Bureau, Health Canada shall be contacted and the facility shall
(v) specify the names and telephone numbers of the RSO, the
adopt this Safety Code. Otherwise, the appropriate provincial or terri-
physician or relevant medical institution, the equipment
torial radiation protection authority shall be contacted to determine
manufacturer, the qualified expert, emergency services
the applicable operational requirements for the x-ray machine.
and the regulatory authority, and ensure that such informa-
tion is up to date;
(vi) notify the regulatory authority of the incident as soon as
possible, and provide accurate and complete information.

38 39
4.8 Disposal of X-ray Equipment References
For the disposal of an x-ray machine, the RSO shall observe the
instructions provided by the manufacturer in the product manual or
contact the manufacturer for information and guidance. In a case
where a manufacturer is no longer in the business of manufacturing,
selling or servicing industrial x-ray equipment, the following proce- 1. International Commission on Radiological Protection. 1990
dures shall be followed: Recommendations of the International Commission on Radio-
(i) the vacuum in the x-ray tube must be breached; logical Protection. Oxford: Pergamon Press; ICRP Publication
(ii) the x-ray tube window should be investigated to determine 60;1991.
whether or not it contains beryllium, and if it does, special
disposal procedures must apply since beryllium presents a 2. International Atomic Energy Agency. International Basic Safety
toxic ingestion or inhalation hazard; Standards for Protection against Ionizing Radiation and for the
(iii) the transformer oil, if this exists, must be disposed of in Safety of Radiation Sources. Vienna, IAEA; Safety Series
accordance with pertinent environmental legislation; and 115;1996.
(iv) the lead must be recycled accordingly.
3. International Atomic Energy Agency. Radiation Protection and
Safety in Industrial Radiography. Vienna, IAEA; Safety Series
13;1999.

4. Treasury Board of Canada Secretariat. Occupational Safety and


Health: Hazardous Substances Directive. Available at:
http://www.tbs-sct.gc.ca/Pubs_pol/hrpubs/TBM_119/hazl_e.html
Accessed 19 April 2002.

5. Human Resources Development Canada. Canada Occupational


Safety and Health Regulations. Ottawa: Queen’s Printer; Canada
Gazette; Revised Statutes of Canada, SOR/2000-374; 2000.
Available at:
http://laws.justice.gc.ca/en/L-2/sor-86-304/text.html
Accessed 26 February 2002.

6. Listing of provincial and territorial contacts. Information on


radiation protection and safety. Available at:
http://www.hc-sc.gc.ca/hecs-sesc/rpb/fptrpc_members.html
Accessed 23 April 2002.

7. Maharaj, H.P. Radiation output and beam quality measurements


of industrial x-ray tubes. CSNDT Journal, 19:36-41; 1998.

40 41
4.8 Disposal of X-ray Equipment References
For the disposal of an x-ray machine, the RSO shall observe the
instructions provided by the manufacturer in the product manual or
contact the manufacturer for information and guidance. In a case
where a manufacturer is no longer in the business of manufacturing,
selling or servicing industrial x-ray equipment, the following proce- 1. International Commission on Radiological Protection. 1990
dures shall be followed: Recommendations of the International Commission on Radio-
(i) the vacuum in the x-ray tube must be breached; logical Protection. Oxford: Pergamon Press; ICRP Publication
(ii) the x-ray tube window should be investigated to determine 60;1991.
whether or not it contains beryllium, and if it does, special
disposal procedures must apply since beryllium presents a 2. International Atomic Energy Agency. International Basic Safety
toxic ingestion or inhalation hazard; Standards for Protection against Ionizing Radiation and for the
(iii) the transformer oil, if this exists, must be disposed of in Safety of Radiation Sources. Vienna, IAEA; Safety Series
accordance with pertinent environmental legislation; and 115;1996.
(iv) the lead must be recycled accordingly.
3. International Atomic Energy Agency. Radiation Protection and
Safety in Industrial Radiography. Vienna, IAEA; Safety Series
13;1999.

4. Treasury Board of Canada Secretariat. Occupational Safety and


Health: Hazardous Substances Directive. Available at:
http://www.tbs-sct.gc.ca/Pubs_pol/hrpubs/TBM_119/hazl_e.html
Accessed 19 April 2002.

5. Human Resources Development Canada. Canada Occupational


Safety and Health Regulations. Ottawa: Queen’s Printer; Canada
Gazette; Revised Statutes of Canada, SOR/2000-374; 2000.
Available at:
http://laws.justice.gc.ca/en/L-2/sor-86-304/text.html
Accessed 26 February 2002.

6. Listing of provincial and territorial contacts. Information on


radiation protection and safety. Available at:
http://www.hc-sc.gc.ca/hecs-sesc/rpb/fptrpc_members.html
Accessed 23 April 2002.

7. Maharaj, H.P. Radiation output and beam quality measurements


of industrial x-ray tubes. CSNDT Journal, 19:36-41; 1998.

40 41
8. Information on health risks to persons using electron beam 16. Varian Medical Systems. Industrial Products: Field Portable
welders. Available at: X-ray Systems.2001. Available at :
http://www.hps.org/publicinformation/ateq143.html http://varian.com/ind/lnm000a.html
Accessed 26 July 2001.
9. Department of National Health and Welfare. Radiation Emitting
Devices Act, Chapter 34 (1st Supplement), Revised Statutes of 17. Choppin, G.R., Rydberg, J. Nuclear Chemistry. Pergamon Press:
Canada. Ottawa, Queen’s Printer, Canada Gazette; 1984. Avail- Toronto;1980.
able at: http://www.canlii.org/ca/regu/crc1370
Accessed 10 March 2003. 18. National Council on Radiation Protection and Measurements.
Neutron Contamination from Medical Accelerators. Bethesda.
10. Canada General Standards Board. Non-Destructive Testing – MD; NCRP, Report 79, 1984.
Qualification and Certification of Personnel (ISO 9712:
1999,MOD). Ottawa: Canada General Standards Board; 19. Web site that provides information on radiation safety course.
National Standards of Canada, CAN/CGSB – 48.9712-2000; Available at:
2000. http://www.ndeinst.org/courses/rad-safety.shtml
Accessed 17 December 2002.
11. Canada General Standards Board. Certificate of Nondestructive
Testing Personnel (Police and Security Radiography Method). 20. Physics International Company. Selection Guide for Flash X-ray
Ottawa: Canada General Standards Board; National Standards Systems. San Leandro, California. Technical Bulletin
of Canada, CAN/CGSB – 48.20-94; 1994. 5952-6701, 1995. Available from: Olin Corporation Aerospace
Division, 2700 Merced St., San Leandro, Ca 94577-0599, USA.
12. Website on radiation dosimetry services.
http://www.landauerinc.com 21. Websites that provide information on Radiation Protection In-
Accessed 15 April 2002. strumentation.
http://www. inovision.com; http://www. eberline.com
13. Websites that provide information on radiation dosimetry http://www.ludlums.com
services. http://www.thermo.com/rpm
http://www.dosimetry.com; http://www.hc-sc.gc.ca http://www.cardinal.com
Accessed 15 April 2002. Accessed 27 February 2002.

14. Canadian Nuclear Safety Commission. Nuclear Substances and


Radiation Devices Regulations. Ottawa; Nuclear Safety and
Control Act. SOR/2000-207. 2000. Available at:
http://laws.justice.gc.ca/en/N-28.3/SOR-2000-207/144272.html
Accessed 02 July 2001.

15. National Council on Radiation Protection and Measurements.


Structural Shielding Design and Evaluation for Medical Use of
X rays and Gamma rays of Energies Up to 10 MeV. Bethesda.
MD; NCRP, Report 49, 1976.

42 43
8. Information on health risks to persons using electron beam 16. Varian Medical Systems. Industrial Products: Field Portable
welders. Available at: X-ray Systems.2001. Available at :
http://www.hps.org/publicinformation/ateq143.html http://varian.com/ind/lnm000a.html
Accessed 26 July 2001.
9. Department of National Health and Welfare. Radiation Emitting
Devices Act, Chapter 34 (1st Supplement), Revised Statutes of 17. Choppin, G.R., Rydberg, J. Nuclear Chemistry. Pergamon Press:
Canada. Ottawa, Queen’s Printer, Canada Gazette; 1984. Avail- Toronto;1980.
able at: http://www.canlii.org/ca/regu/crc1370
Accessed 10 March 2003. 18. National Council on Radiation Protection and Measurements.
Neutron Contamination from Medical Accelerators. Bethesda.
10. Canada General Standards Board. Non-Destructive Testing – MD; NCRP, Report 79, 1984.
Qualification and Certification of Personnel (ISO 9712:
1999,MOD). Ottawa: Canada General Standards Board; 19. Web site that provides information on radiation safety course.
National Standards of Canada, CAN/CGSB – 48.9712-2000; Available at:
2000. http://www.ndeinst.org/courses/rad-safety.shtml
Accessed 17 December 2002.
11. Canada General Standards Board. Certificate of Nondestructive
Testing Personnel (Police and Security Radiography Method). 20. Physics International Company. Selection Guide for Flash X-ray
Ottawa: Canada General Standards Board; National Standards Systems. San Leandro, California. Technical Bulletin
of Canada, CAN/CGSB – 48.20-94; 1994. 5952-6701, 1995. Available from: Olin Corporation Aerospace
Division, 2700 Merced St., San Leandro, Ca 94577-0599, USA.
12. Website on radiation dosimetry services.
http://www.landauerinc.com 21. Websites that provide information on Radiation Protection In-
Accessed 15 April 2002. strumentation.
http://www. inovision.com; http://www. eberline.com
13. Websites that provide information on radiation dosimetry http://www.ludlums.com
services. http://www.thermo.com/rpm
http://www.dosimetry.com; http://www.hc-sc.gc.ca http://www.cardinal.com
Accessed 15 April 2002. Accessed 27 February 2002.

14. Canadian Nuclear Safety Commission. Nuclear Substances and


Radiation Devices Regulations. Ottawa; Nuclear Safety and
Control Act. SOR/2000-207. 2000. Available at:
http://laws.justice.gc.ca/en/N-28.3/SOR-2000-207/144272.html
Accessed 02 July 2001.

15. National Council on Radiation Protection and Measurements.


Structural Shielding Design and Evaluation for Medical Use of
X rays and Gamma rays of Energies Up to 10 MeV. Bethesda.
MD; NCRP, Report 49, 1976.

42 43
Glossary Cabinet x-ray system. An x-ray system designed to contain an x-ray
tube and generator, detectors and space for the examination of mate-
rials within an enclosure that is independent of existing architectural
structures of a building except the floor on which it may be placed,
provides radiation attenuation, and excludes personnel from its inte-
rior during generation of ionizing radiation.
The following definitions apply to this Safety Code.
Cancer. A general term used to indicate any of the various types of
Accident. An event which leads to or could lead to unplanned or malignant neoplasms.
inadvertent irradiations of humans.
Certified industrial radiographer. An industrial radiographer qual-
ALARA. Acronym for As Low As Reasonably Achievable. It is a ified in accordance with the national standards of Canada or equiva-
concept that means the design and use of radiation sources and the lent; the Canadian General Standards Board standards are national.
practices associated therewith must be such as to ensure that expo- An individual who is a certified exposure device operator in accor-
sures to radiation are kept as low as reasonably achievable with social dance with Canadian Nuclear Safety Commission regulations is also
and economic factors taken into account. considered a certified industrial radiographer for the purposes of this
Safety Code.
Allied personnel. Safety personnel or other staff in the vicinity of
radiography work areas of a facility. Collimator. A shield placed in the path of a primary beam to restrict
the size of that radiation beam.
Alpha particle. A helium nucleus made up of two neutrons and two
protons. An alpha particle can be stopped in a piece of paper a few Congenital. Existing at birth or acquired during in-utero develop-
microns thick. It produces substantially more ions than an x ray or ment and not through heredity.
gamma ray or an electron and, therefore, the potential for cellular
damage is greater. Controlled area. A defined area within which industrial radiography
is carried out and where dose levels are likely to exceed 0.3 mSv in a
Atom. Basic component of matter. It is the smallest part of an element week. It shall be under the supervision of a radiation safety officer
that has all the chemical properties of that element. It consists of a (RSO).
nucleus and electrons around the nucleus.
DNA. Deoxyribonucleic acid. The genetic material of cells.
Background radiation. Radiation arising from natural sources pres-
ent in the environment, solar and cosmic radiation in outer space, and Deterministic effect. An adverse biological effect for which a
naturally occurring radioactive elements present in the atmosphere, threshold dose generally exists. The severity of the effect increases
the ground, building materials and the human body. The background with dose. Below the threshold dose the effect is not detected
radiation level varies from place to place. An individual residing in clinically.
North America would incur an average radiation dose value of about
2.4 mSv (240 mrem) per year. Dose. The quotient, expressed in gray, of the energy absorbed
through exposure to ionizing radiation and the mass of the body or
Binding energy. A general term that refers to the minimum energy that part of the body that absorbed the radiation. (1 gray (Gy) =
required to remove a proton or a neutron or an electron from an atom. 100 rad.)

44 45
Glossary Cabinet x-ray system. An x-ray system designed to contain an x-ray
tube and generator, detectors and space for the examination of mate-
rials within an enclosure that is independent of existing architectural
structures of a building except the floor on which it may be placed,
provides radiation attenuation, and excludes personnel from its inte-
rior during generation of ionizing radiation.
The following definitions apply to this Safety Code.
Cancer. A general term used to indicate any of the various types of
Accident. An event which leads to or could lead to unplanned or malignant neoplasms.
inadvertent irradiations of humans.
Certified industrial radiographer. An industrial radiographer qual-
ALARA. Acronym for As Low As Reasonably Achievable. It is a ified in accordance with the national standards of Canada or equiva-
concept that means the design and use of radiation sources and the lent; the Canadian General Standards Board standards are national.
practices associated therewith must be such as to ensure that expo- An individual who is a certified exposure device operator in accor-
sures to radiation are kept as low as reasonably achievable with social dance with Canadian Nuclear Safety Commission regulations is also
and economic factors taken into account. considered a certified industrial radiographer for the purposes of this
Safety Code.
Allied personnel. Safety personnel or other staff in the vicinity of
radiography work areas of a facility. Collimator. A shield placed in the path of a primary beam to restrict
the size of that radiation beam.
Alpha particle. A helium nucleus made up of two neutrons and two
protons. An alpha particle can be stopped in a piece of paper a few Congenital. Existing at birth or acquired during in-utero develop-
microns thick. It produces substantially more ions than an x ray or ment and not through heredity.
gamma ray or an electron and, therefore, the potential for cellular
damage is greater. Controlled area. A defined area within which industrial radiography
is carried out and where dose levels are likely to exceed 0.3 mSv in a
Atom. Basic component of matter. It is the smallest part of an element week. It shall be under the supervision of a radiation safety officer
that has all the chemical properties of that element. It consists of a (RSO).
nucleus and electrons around the nucleus.
DNA. Deoxyribonucleic acid. The genetic material of cells.
Background radiation. Radiation arising from natural sources pres-
ent in the environment, solar and cosmic radiation in outer space, and Deterministic effect. An adverse biological effect for which a
naturally occurring radioactive elements present in the atmosphere, threshold dose generally exists. The severity of the effect increases
the ground, building materials and the human body. The background with dose. Below the threshold dose the effect is not detected
radiation level varies from place to place. An individual residing in clinically.
North America would incur an average radiation dose value of about
2.4 mSv (240 mrem) per year. Dose. The quotient, expressed in gray, of the energy absorbed
through exposure to ionizing radiation and the mass of the body or
Binding energy. A general term that refers to the minimum energy that part of the body that absorbed the radiation. (1 gray (Gy) =
required to remove a proton or a neutron or an electron from an atom. 100 rad.)

44 45
Dose equivalent. The product of dose and a radiation weighting fac- Gamma ray. High-energy, short wavelength electromagnetic radia-
tor. For x rays and gamma rays the radiation weighting factor is 1. tion emitted in the radioactive decay of an unstable atom.
Dose equivalent can be expressed in rem. 1 rem ~ 1 rad for soft tissue.
In 1990, the ICRP proposed using the term ‘equivalent dose’ instead Gray. International Standard (SI) unit for dose. 1 gray (Gy) = 100 rad.
of dose equivalent.
High radiation field. A qualitative indication of the equivalent dose
Dose limit: Applies in relation to persons and refers to the limit on rates that exist in a particular area. Areas with equivalent dose rates in
effective dose or equivalent dose specified in Appendix II. excess of 1 mSv per hour should be regarded as high radiation fields.

Effective dose. It is the sum of the products, obtained by multiplying IAEA. International Atomic Energy Agency. An agency formed in
the equivalent dose of radiation received by each tissue or organ listed 1954 within the United Nations to create and apply international safe-
in Table 3 of Appendix II and the corresponding tissue or organ guards consistent with the peaceful uses of atomic energy. It has
weighting factor given in column 2 of the same Table. The unit is the issued internationally agreed radiation protection standards based on
seivert (Sv). the recommendations of the ICRP.

Electromagnetic radiation. Radiation considered as a wave of ICRP. International Commission on Radiological Protection. An in-
electric and magnetic energy traveling through a medium. In space it dependent group of experts from a wide range of scientific disciplines
travels at the speed of light. It includes x rays and gamma rays. which have published recommendations for the protection of radia-
tion workers and the public against ionizing radiation for more than
Electron beam welder. A machine that produces high-intensity 50 years.
focused electron beams for melting and bonding materials in vacuum
conditions. Ionizing radiation. Radiation that has enough energy to dislodge an
electron of an atom. It includes x rays, gamma rays, alpha particles,
Embryo. A developing human organism from conception to the end neutrons, protons and electrons.
of the sixth week of gestation.
Industrial radiographer. A person having the academic and profes-
Ensure. To make certain. sional training to use radiation equipment for industrial radiography.
A certified industrial radiographer is within the scope of this defini-
Equivalent dose. It is the product of dose and a radiation weighting tion.
factor expressed in seivert. It is the term recommended by the ICRP in
1990 to replace the dose equivalent. (One seivert (Sv) = 100 rem.) Industrial radiography. Examination of the structure of materials
by nondestructive methods in which ionizing radiation is used to
Facility. A site where industrial radiography is carried out. make radiographic images.

Failsafe. Having the property that any failure causes an action or Industrial x-ray equipment. X-ray machines used for industrial
actions which always result in a safe situation. radiography and electron beam welders used for melting and bonding
materials.
Flash x-ray source. A device capable of producing extremely high
levels of x rays within very short time frames in the order of micro-
seconds or less. Such sources are used for studying dynamic events
associated with ballistics, explosives, implosions or detonators.

46 47
Dose equivalent. The product of dose and a radiation weighting fac- Gamma ray. High-energy, short wavelength electromagnetic radia-
tor. For x rays and gamma rays the radiation weighting factor is 1. tion emitted in the radioactive decay of an unstable atom.
Dose equivalent can be expressed in rem. 1 rem ~ 1 rad for soft tissue.
In 1990, the ICRP proposed using the term ‘equivalent dose’ instead Gray. International Standard (SI) unit for dose. 1 gray (Gy) = 100 rad.
of dose equivalent.
High radiation field. A qualitative indication of the equivalent dose
Dose limit: Applies in relation to persons and refers to the limit on rates that exist in a particular area. Areas with equivalent dose rates in
effective dose or equivalent dose specified in Appendix II. excess of 1 mSv per hour should be regarded as high radiation fields.

Effective dose. It is the sum of the products, obtained by multiplying IAEA. International Atomic Energy Agency. An agency formed in
the equivalent dose of radiation received by each tissue or organ listed 1954 within the United Nations to create and apply international safe-
in Table 3 of Appendix II and the corresponding tissue or organ guards consistent with the peaceful uses of atomic energy. It has
weighting factor given in column 2 of the same Table. The unit is the issued internationally agreed radiation protection standards based on
seivert (Sv). the recommendations of the ICRP.

Electromagnetic radiation. Radiation considered as a wave of ICRP. International Commission on Radiological Protection. An in-
electric and magnetic energy traveling through a medium. In space it dependent group of experts from a wide range of scientific disciplines
travels at the speed of light. It includes x rays and gamma rays. which have published recommendations for the protection of radia-
tion workers and the public against ionizing radiation for more than
Electron beam welder. A machine that produces high-intensity 50 years.
focused electron beams for melting and bonding materials in vacuum
conditions. Ionizing radiation. Radiation that has enough energy to dislodge an
electron of an atom. It includes x rays, gamma rays, alpha particles,
Embryo. A developing human organism from conception to the end neutrons, protons and electrons.
of the sixth week of gestation.
Industrial radiographer. A person having the academic and profes-
Ensure. To make certain. sional training to use radiation equipment for industrial radiography.
A certified industrial radiographer is within the scope of this defini-
Equivalent dose. It is the product of dose and a radiation weighting tion.
factor expressed in seivert. It is the term recommended by the ICRP in
1990 to replace the dose equivalent. (One seivert (Sv) = 100 rem.) Industrial radiography. Examination of the structure of materials
by nondestructive methods in which ionizing radiation is used to
Facility. A site where industrial radiography is carried out. make radiographic images.

Failsafe. Having the property that any failure causes an action or Industrial x-ray equipment. X-ray machines used for industrial
actions which always result in a safe situation. radiography and electron beam welders used for melting and bonding
materials.
Flash x-ray source. A device capable of producing extremely high
levels of x rays within very short time frames in the order of micro-
seconds or less. Such sources are used for studying dynamic events
associated with ballistics, explosives, implosions or detonators.

46 47
Industrial x-ray equipment owner. A person, organization or Occupational. The exposure of an individual to ionizing radiation
institution having title to or administrative control of one or more incurred in the course of employment by virtue of the normal duties
permanent installations or radiation sources used for industrial of the individual in the workplace.
radiography.
Other user. An individual age 18 or older who uses industrial x-ray
Isotope. Atoms of the same element but different mass numbers. The equipment for a purpose other than industrial radiography or material
mass number refers to the sum of the number of protons and neutrons fusion (melting and bonding).
in the nucleus of the atom.
Permanent installation. A shielded enclosure in a fixed location.
Interlock. A device that precludes radiation exposure to an individ-
ual by preventing entry to a hazardous area or by automatically re- Photon. A quantum of electromagnetic radiation equal to a constant
moving the hazard. (Planck’s constant which is: 6.61 ´10-34J s ) times the frequency of
radiation (unit is the hertz), applicable to x rays and gamma rays.
Ion. An atom or group of atoms that carries a positive or negative
charge as a result of having lost or gained one or more electrons. Primary radiation. Radiation that emanates from a radiation source
or target and passes through the radiation source assembly via a
Irradiation. Exposure of a living being or matter to x-radiation or collimator or other beam shaping device. A synonym for primary
electrons. beam.

Leakage radiation. All radiation, except the primary radiation, Proton. One of the three basic particles of an atom. Protons have a
coming from an ionizing radiation source. positive electrical charge, are located in the nucleus of the atom and
each has a mass approximately equal to a neutron.
Leukemia. A disease in humans characterized by an abnormal
increase in the number of white blood cells in tissue or blood. Rad. Unit of absorbed dose. One rad = 100 erg of energy absorbed per
gram of material. 1 rad equals 1 centigray.
Malignant. Resistant to treatment and occurring in severe form and
frequently fatal. Radiation equipment owner. A person, organization, or institution
having title to or administrative control over one or more facilities
MeV. Million electron volts. The kinetic energy acquired by an elec- having source(s) of ionizing radiation.
tron in traveling through a potential difference of 1 million volts.
Radiation protection adviser. A person having the specialized
Muon. A particle that has a mass about one-tenth that of a proton or knowledge, training and experience necessary to perform the task of
about 200 times heavier than an electron. giving advice and guidance in ionizing radiation safety in a compe-
tent and professional manner.
MV. The potential to which accelerating particles is subjected.
RSO. Radiation Safety Officer. An individual having the responsi-
Neutron. One of the three basic particles of all atoms heavier than hy- bility for the overall radiation safety program on behalf of the owner
drogen. Neutrons have no electrical charge, are located in the nucleus of the industrial x-ray equipment and meeting the requirements of
of the atom and each has a mass approximately equal to a proton. Section 2.2 of this Safety Code. ( An individual recognized as an RSO
in accordance with CNSC requirements would suffice for purposes of
Nuclide. A general term used to describe a family of isotopes. this Safety Code.)

48 49
Industrial x-ray equipment owner. A person, organization or Occupational. The exposure of an individual to ionizing radiation
institution having title to or administrative control of one or more incurred in the course of employment by virtue of the normal duties
permanent installations or radiation sources used for industrial of the individual in the workplace.
radiography.
Other user. An individual age 18 or older who uses industrial x-ray
Isotope. Atoms of the same element but different mass numbers. The equipment for a purpose other than industrial radiography or material
mass number refers to the sum of the number of protons and neutrons fusion (melting and bonding).
in the nucleus of the atom.
Permanent installation. A shielded enclosure in a fixed location.
Interlock. A device that precludes radiation exposure to an individ-
ual by preventing entry to a hazardous area or by automatically re- Photon. A quantum of electromagnetic radiation equal to a constant
moving the hazard. (Planck’s constant which is: 6.61 ´10-34J s ) times the frequency of
radiation (unit is the hertz), applicable to x rays and gamma rays.
Ion. An atom or group of atoms that carries a positive or negative
charge as a result of having lost or gained one or more electrons. Primary radiation. Radiation that emanates from a radiation source
or target and passes through the radiation source assembly via a
Irradiation. Exposure of a living being or matter to x-radiation or collimator or other beam shaping device. A synonym for primary
electrons. beam.

Leakage radiation. All radiation, except the primary radiation, Proton. One of the three basic particles of an atom. Protons have a
coming from an ionizing radiation source. positive electrical charge, are located in the nucleus of the atom and
each has a mass approximately equal to a neutron.
Leukemia. A disease in humans characterized by an abnormal
increase in the number of white blood cells in tissue or blood. Rad. Unit of absorbed dose. One rad = 100 erg of energy absorbed per
gram of material. 1 rad equals 1 centigray.
Malignant. Resistant to treatment and occurring in severe form and
frequently fatal. Radiation equipment owner. A person, organization, or institution
having title to or administrative control over one or more facilities
MeV. Million electron volts. The kinetic energy acquired by an elec- having source(s) of ionizing radiation.
tron in traveling through a potential difference of 1 million volts.
Radiation protection adviser. A person having the specialized
Muon. A particle that has a mass about one-tenth that of a proton or knowledge, training and experience necessary to perform the task of
about 200 times heavier than an electron. giving advice and guidance in ionizing radiation safety in a compe-
tent and professional manner.
MV. The potential to which accelerating particles is subjected.
RSO. Radiation Safety Officer. An individual having the responsi-
Neutron. One of the three basic particles of all atoms heavier than hy- bility for the overall radiation safety program on behalf of the owner
drogen. Neutrons have no electrical charge, are located in the nucleus of the industrial x-ray equipment and meeting the requirements of
of the atom and each has a mass approximately equal to a proton. Section 2.2 of this Safety Code. ( An individual recognized as an RSO
in accordance with CNSC requirements would suffice for purposes of
Nuclide. A general term used to describe a family of isotopes. this Safety Code.)

48 49
Radiation source. A sealed gamma ray source or other source emit- Stochastic effects. An adverse biological effect for which the
ting ionizing radiation. probability of occurrence increases with dose but the severity does
not depend on the magnitude of the absorbed dose. There is no dose
Radioactive. Release of energy in the form of particles or gamma threshold for such effects to occur.
rays emitted by the disintegration of the nuclei of atoms.
Stray radiation. The sum of secondary and leakage radiations.
Radioactive isotope. Element that emits ionizing radiation when it
decays. Supervised area. Any area not designated a controlled area but
which requires the occupational exposure conditions to be kept under
Radiography. Use of ionizing radiation to produce radiographic review.
images on film, fluorescent material or an image display device.
Survey meter. A portable instrument that measures exposure or dose
Radionuclide. Any species of an atom that is radioactive. or equivalent dose rate.

Regulatory authority. An authority designated or otherwise recog- Target. Part of an x-ray tube or material surface onto which is
nized by a government for regulatory purposes in connection with directed a beam of accelerated particles to produce x rays.
protection and safety.
X rays. Electromagnetic radiation that is more energetic than visible
Roentgen. Unit of exposure to x rays or gamma rays having energies light.
< 3 MeV. One roentgen = 2.58 ´ 10-4 coulomb per kilogram of air.
X-radiation. A beam of x rays.
Secondary radiation. Ionizing radiation emitted by matter as a result
of the interaction of primary radiation with that matter.

Sell. An offer for sale, or advertising a product for the purpose of sale,
or delivering or distributing a product for sale or lease.

Service personnel. Specifically trained persons, authorized by the


industrial x-ray equipment manufacturer or its agent(s), to undertake
service functions.

Shielded enclosure. An enclosed space that contains a radiation


source and is engineered with safety components to provide adequate
shielding and protection from ionizing radiations.

Shielding. Material used to protect people or living things or items


from ionizing radiation.

Sievert. Unit of equivalent dose. One sievert (Sv) = 100 rem.

50 51
Radiation source. A sealed gamma ray source or other source emit- Stochastic effects. An adverse biological effect for which the
ting ionizing radiation. probability of occurrence increases with dose but the severity does
not depend on the magnitude of the absorbed dose. There is no dose
Radioactive. Release of energy in the form of particles or gamma threshold for such effects to occur.
rays emitted by the disintegration of the nuclei of atoms.
Stray radiation. The sum of secondary and leakage radiations.
Radioactive isotope. Element that emits ionizing radiation when it
decays. Supervised area. Any area not designated a controlled area but
which requires the occupational exposure conditions to be kept under
Radiography. Use of ionizing radiation to produce radiographic review.
images on film, fluorescent material or an image display device.
Survey meter. A portable instrument that measures exposure or dose
Radionuclide. Any species of an atom that is radioactive. or equivalent dose rate.

Regulatory authority. An authority designated or otherwise recog- Target. Part of an x-ray tube or material surface onto which is
nized by a government for regulatory purposes in connection with directed a beam of accelerated particles to produce x rays.
protection and safety.
X rays. Electromagnetic radiation that is more energetic than visible
Roentgen. Unit of exposure to x rays or gamma rays having energies light.
< 3 MeV. One roentgen = 2.58 ´ 10-4 coulomb per kilogram of air.
X-radiation. A beam of x rays.
Secondary radiation. Ionizing radiation emitted by matter as a result
of the interaction of primary radiation with that matter.

Sell. An offer for sale, or advertising a product for the purpose of sale,
or delivering or distributing a product for sale or lease.

Service personnel. Specifically trained persons, authorized by the


industrial x-ray equipment manufacturer or its agent(s), to undertake
service functions.

Shielded enclosure. An enclosed space that contains a radiation


source and is engineered with safety components to provide adequate
shielding and protection from ionizing radiations.

Shielding. Material used to protect people or living things or items


from ionizing radiation.

Sievert. Unit of equivalent dose. One sievert (Sv) = 100 rem.

50 51
Appendix I Appendix II
X-ray Warning Sign Recommended Dose Limits for
Ionizing Radiation

The X-ray warning sign referred to in this Safety Code is a sign


that: In its 1990 Publication 60(1), the International Commission on
a. has a yellow background and black text; Radiological Protection (ICRP) recommended dose limits for ioniz-
b. is of a size appropriate to the size of the device or of the area, ing radiation. The dose limits do not include medical and natural
door, wall, installation or other item in respect of which it is background ionizing radiation exposures and are indicated below.
posted;
Table 1. Effective dose limits
c. bears the words “CAUTION, X-RAYS : NO UNAUTHORIZED
USE ” and “ATTENTION, RAYONS X: UTILISATION NON Person Period Effective Dose (mSv)
AUTORISEE”; and
Radiation worker (a) One year (a) 20 (average)
d. is designed in accordance with the following diagram: 50 (special circumstances)

(b) 5-year period (b) 100

Pregnant radiation Remainder of 2


worker§ pregnancy

Public One year 1

§
In the case of a female radiation worker who is pregnant, the fetus must
be protected from radiation exposure for the remainder of the pregnancy
once pregnancy has been diagnosed.

Table 2. Equivalent dose limits

Tissue or Person Period Equivalent


Organ dose (mSv)

Skin (a) Radiation worker One year 500

(b) Public One year 50

Lens of (a) Radiation worker One year 150


the eye
(b) Public One year 15

Hands (a) Radiation worker One year 500


and feet
(b) Public One year 50

52 53
Appendix I Appendix II
X-ray Warning Sign Recommended Dose Limits for
Ionizing Radiation

The X-ray warning sign referred to in this Safety Code is a sign


that: In its 1990 Publication 60(1), the International Commission on
a. has a yellow background and black text; Radiological Protection (ICRP) recommended dose limits for ioniz-
b. is of a size appropriate to the size of the device or of the area, ing radiation. The dose limits do not include medical and natural
door, wall, installation or other item in respect of which it is background ionizing radiation exposures and are indicated below.
posted;
Table 1. Effective dose limits
c. bears the words “CAUTION, X-RAYS : NO UNAUTHORIZED
USE ” and “ATTENTION, RAYONS X: UTILISATION NON Person Period Effective Dose (mSv)
AUTORISEE”; and
Radiation worker (a) One year (a) 20 (average)
d. is designed in accordance with the following diagram: 50 (special circumstances)

(b) 5-year period (b) 100

Pregnant radiation Remainder of 2


worker§ pregnancy

Public One year 1

§
In the case of a female radiation worker who is pregnant, the fetus must
be protected from radiation exposure for the remainder of the pregnancy
once pregnancy has been diagnosed.

Table 2. Equivalent dose limits

Tissue or Person Period Equivalent


Organ dose (mSv)

Skin (a) Radiation worker One year 500

(b) Public One year 50

Lens of (a) Radiation worker One year 150


the eye
(b) Public One year 15

Hands (a) Radiation worker One year 500


and feet
(b) Public One year 50

52 53
Table 3. Organ or tissue weighting factors Table 4. Radiation weighting factors

Organ or Tissue Weighting Factor, wT


Radiation type and energy Radiation weighting
Gonads (testes or ovaries) 0.2 factor, wR

Red bone marrow 0.12 Photons, all energies 1

Colon 0.12 Electrons and muons, all energies§ 1

Lung 0.12 Neutrons of energy < 10 keV 5

Stomach 0.12 Neutrons of energy 10 keV to 100 keV 10

Bladder 0.05 Neutrons of energy >100 keV to 2 MeV 20

Breast 0.05 Neutrons of energy > 2 MeV to 20 MeV 10

Liver 0.05 Neutrons of energy > 20 MeV 5

Oesophagus 0.05 Protons, other than recoil protons, 5


of energy > 2 MeV
Thyroid 0.05
Alpha particles, fission fragments, heavy nuclei 20
Skin 0.01
§
Bone surface 0.01 Refer to Table S-1 in ICRP 1990 Publication 60(1) for specific notes on
radiation types and energy.
Remainder organs mean all organs
and tissues, excluding the twelve
listed above, and include: adrenals, 0.05
brain, upper large intestine, small
intestine, kidney, muscle, pancreas,
spleen, thymus and uterus¶

Whole body 1

¶ In those rare cases in which one of these remainder organs receives an


equivalent dose that exceeds the equivalent dose received by any one of
the preceding 12 organs listed, a weighting factor of 0.025 shall be applied
to that organ or tissue and a weighting factor of 0.025 shall be applied to
the average equivalent dose received by the remainder organs and
tissues.

54 55
Table 3. Organ or tissue weighting factors Table 4. Radiation weighting factors

Organ or Tissue Weighting Factor, wT


Radiation type and energy Radiation weighting
Gonads (testes or ovaries) 0.2 factor, wR

Red bone marrow 0.12 Photons, all energies 1

Colon 0.12 Electrons and muons, all energies§ 1

Lung 0.12 Neutrons of energy < 10 keV 5

Stomach 0.12 Neutrons of energy 10 keV to 100 keV 10

Bladder 0.05 Neutrons of energy >100 keV to 2 MeV 20

Breast 0.05 Neutrons of energy > 2 MeV to 20 MeV 10

Liver 0.05 Neutrons of energy > 20 MeV 5

Oesophagus 0.05 Protons, other than recoil protons, 5


of energy > 2 MeV
Thyroid 0.05
Alpha particles, fission fragments, heavy nuclei 20
Skin 0.01
§
Bone surface 0.01 Refer to Table S-1 in ICRP 1990 Publication 60(1) for specific notes on
radiation types and energy.
Remainder organs mean all organs
and tissues, excluding the twelve
listed above, and include: adrenals, 0.05
brain, upper large intestine, small
intestine, kidney, muscle, pancreas,
spleen, thymus and uterus¶

Whole body 1

¶ In those rare cases in which one of these remainder organs receives an


equivalent dose that exceeds the equivalent dose received by any one of
the preceding 12 organs listed, a weighting factor of 0.025 shall be applied
to that organ or tissue and a weighting factor of 0.025 shall be applied to
the average equivalent dose received by the remainder organs and
tissues.