0 ratings0% found this document useful (0 votes) 1K views13 pagesMedela Basic, Median, Dominant - Service Manual PDF
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content,
claim it here.
Available Formats
Download as PDF or read online on Scribd
Service Instructions
Order Nr. 190.0101b
BASIC 036
MEDIAN 046
DOMINANT 056
medela
ee ier ee Tee nnn tenet ey eatenTable of Contents Page
1. Technical Data 3
2. Construction 4
3. Pneumatic Diagram 5
4. Electric Circurt Diagrams 6-8
5. Trubleshoot ing 3
6. Maintenance Procedures 10 - 13
Servicing 10
Dismat }ing and Cleaning Piston/cy linder unit 4
Reasserb ly Piston/Cy linder Unit n
Replacement F lat- and Toothed Belts 1"
Service Call Indicator 12
Correcting an Electronic Fault 13
7. Spare Parts 14-19
Drawings 14
Spare Parts List / Order Numbers 7
Median Dominant:1, Technical Data
mere antes
Power Supply Standard : 220 Vs $0 He
Options : see sales literature
{ Electric Standard IEC 601.1 / Protection Class I or IT
Power Consumption: 100 W 100 w 125 W
{ vacuum Range * and - i
Displacement :
j Secretion Aspirator and 5 - 90 kPa 5 - 90 kPa 5 - 90 kPa
‘ Surgical Suction Pump 21 / Min 301 / Min 451. / Min
i Vacuum Extractor see above
Uterin Aspirator a see above
Low Vacuum Aspirator 10-80 mbar | 10-80 mbar | —--———-
201/Min | 201 / Min
Hight + / - Stand 267/984 mm | 227/944 nm | 227/944 mm
Width " 240/420 nm | 325/494 nm | 325/494 mn
Depth . 324/480 mm | 324/616 mm | 324/616 mm
Weight ” 8,4/20 kg 10/21 kg 10/21 kg
Basic 036 | Median 046 |oominant: 056
x These figures are based on the altitude of our production facilities in
Baar/Ch, 444 meters above sea-level. According to local conditions resp.
barometric pressure, slight deviations in measurements may be noted. The
table below contains several factors as reference values. The vacuum
Figures given mist be multiplied by one of these factors, according to
the altitude of the location where measurements are caried out.
Location and vacuum reading
Location | sea-level | 444m a.s1.|1000n a.s1.|2000m a.s1.|4000m a.sl.
Factor I 1.06 L 1.0 0.94 0.83 0.64
a2. Construction
Medela suction pumps operate on the tried and tested piston/cylinder system.
The entire unit is designed to be simple and robust in construction.
Two double-acting cylinders powered by a single-phase asynchronous motor with
integral belt drive produce high suction rates. The pump system operates at a
low shaft speed. This means that noise and vibrations - as well as mechanical
wear and tear - are kept to a minimum.
The cylinders pivot on teflon-coated sleeve bearings, which require no
further lubrication. The sealed ball bearings fitted to the piston rods are
permanently lubricated. The use of carefully-selected materials means that the
piston/cylinder units require no lubrication; the pistons work dry, with
almost complete freedom from wear.
The valves are grouped in a block which is easy to dismantle. This greatly
facilitates cleaning operations.
The specially-designed vacuum regulator (patent pending) permits the pump to
work up to the preselected level without by-pass, i.e. with full performance.
This means that the desired suction level is reached very quickly. The range
of adjustment covered by the vacuum regulator is linear. The evacuated air is
allowed to escape via a sound absorber.
The whole low-voltage section of models fitted with electronics is mounted on
a printed circuit board.
The components described above are all easily accessible, When carrying out
maintenance work, it is only necessary to remove the four screws on each side
and the outer casing can then be removed. The pump unit and controls are now
free for access.
The clear layout of the operating and monitoring elements ensures simple,
foolproof operation.
The mobile stand is fitted with four, pivoting castors. Two of these are
fitted with brakes. A rail to which bottle-holders and other accessories can
be attached is fitted at working height.
The safety bottle with lid is directly attached to the suction connection of
the suction pump as a complete unit. The purpose of this vessel is to
intercept droplets of liquid carried by air suction from the secretion bottle,
thereby protecting the suction pump unit from contamination. For reasons of
hygiene, however, it is vital to use a bacteria filter at all times in order
to filter the evacuated air before it enters the pump.
The secretion bottles are equipped with practical quick-locking lids, which
can be fitted to all sizes of bottle. An overflow protection device (Float
valve) is a standard fitting. This prevents fluid from entering the pump,
whenever the secretion bottles have inadvertantly not been emptied in time. If
necessary (e.g. for cleaning purposes) the float valve can easily be removed.
The use of standardized tube connectors means that bottles are easy to link
together3. Pneumatic Diagram
equosay punog
LS es Se aS sss
YPOLG @ALEA
4eputt 9 / uoasid4, Electric circuit diagrams
A. Without electronics
ee
| \~4
=)
ty
a Be
J i
es
8, Without electronics
Protection Class II
rota
i]
ih;
= —
: 8 ll 4 aati
6 Ls
0 ve
©. Without electronics, with
connector for foot switch
\ \
Sehatziterl \senuteatse
Seriebegennise “Ba
Fuss-Sehalter tgD. Without electronics, AP
(anaesthet 1c~proot }
/ Vv we
a COs
4 JA F / NC
sertcte utter
seapl. \ute
E. Without electronics, AP
with foot switch
Is
r
o lal
x01 ||
Scrutzeter\
Geinesegetse
_-Foss-Sehater
+
L
t
JF. with electronics
fo
Ngee PANS
A w
sno
om eee S
fy safe sen nr
™ oy {|
el
G. With electronics, AP
fn
]
a [2
ze! |
sctiveaetcse
w/
H. Foot Switch for pumps
with electronics 5G,
to5. Troubleshoot ing
Problem Possible source of trouble [ Check/action to be taken
Mains switch | set at
Power supply — Voltage, power socket
Ee Electrical cord Plug type, plug
3 | connections, | damaged
Bs | cord
22 Fuses pe, condition
2 Drive unit jammed Sunmon technician, or
repair service
| “etectronie faate See page 12/13 —
Leak within the pump or in the Isolate the source oF
external suction circuit
trouble by carrying out
the checks outlined in
Section 6.
External cause: — 7 __
Push-in connector O-rings Condition
Safety bottle O-rings Condition
Rir filter O-rings Condition
Push-in connectors
No leaks at tube ends
Push in as far as stop
Inadequate suction performance
Tubing [“Gracks, brittle areas
Lids: Airtight fit to bottle
Bottles Condition, possible]
cracked edges, hairline
cracks
Bacteria filter housing Condition, absence of
leaks
Fitter paper Condition, position of
housing
Internal cause: —
Tubing or mechanical components Summon technician or
repair service
Blockage or kink tn tubing
within the pump or in the
external suction circuit
| [solate source of
trouble as follows
Remove the safety bottle
from the pump; Teave the
suction connection open.
Tf the vacuum gauge now
indicates zero vacuum =
the cause is external
External cause:
Vacuum gauge indicates a vacuum
Pump does not suck, however
Bacteria filter Investigate the whole
Safety bottle external suction circuit
Overflow valve in the for blockage(s), float
lid, (float valve) valve sticking, kinked
push-in connectors, tubing and rectify.
tubing, etc.
Tf the vacuum gauge
still indicates a vacuum
= the cause is interna)
‘Internal cause
Tubing or mechanical components
echnician or
repair service6. Maintenance procedures
The pump must be given a routine service at intervals of approximately 1,000
operating hours or at least once every year.
To ensure troublefree operation, repairs to the pump may only be carried out
by an agent authorized by Medela. This also includes cylinder cleaning as well
as belt changing.
Servicing
The following figures refer to the drawings and the spare parts‘ list at the
end of this manual.
General:
Disconnect the electrical cord (7) from the mains supply.
Release the eight screws (14) and remove the pump housing (2).
Check the toothed belt (26) after removing the cover plate (43). Also check
the flat belt (17), Wipe ail surfaces free of dust resulting from wear and
tear on the belt. If the belt indicates excessive signs of wear, it must be
replaced according to the special instructions.
Check the sound absorber (80) and remove any dirt if necessary.
Ensure that al] tubing connections are clean, both inside and outside, and
check for discoloration, cracks or defective connectors. Replace if necessary.
All pieces of tubing should be of adequate length and routed so as to prevent
‘them from janming or kinking anywhere.
AI] tubing ends mist extend over the full length of the connector ferrule as
far as the stop, and must be free from cracks. By pulling on them gently,
check that all pieces of tubing are routed freely without any kinks.
It is vital when replacing a piece of tubing:
that only genuine Medela tubing is used for internal connections, end that the
ends are pushed well onto the ferrules as far as the stop. It is important to
ensure that the pieces of tubing (61,66) which come from the pump cylinders
(3,10) are free to move without any kinks when the motor (28) is running.
Valve block
Undo the knurled nuts (4) and remove the valve assembly (62-65). Only remove
the tubing if the inspection measures described above indicate that
replacements or cleaning are required. Clean off all parts of the valves with
Compressed air, Wash off any stubborn dirt with cold water and dry all items
prior to reassembly. Replace any damaged rubber seals (63). Prior to
reassembly, dust all used and new rubber components lightly with talc.
Please note
If the sound absorber (80) and the tubing are contaminated, it must be assumed
that the valves (62 - 65) and the pump cylinder(s) (3/10) require cleaning.
Please take note of the section on "Piston/cylinders"! If the pieces of tubing
(53, 57, 75) to the vacuum regulater (73/74) are dirty, the latter must be
dismantled and cleaned according to the detail drawing.
Electrical connections:
Ensure that all screw- and push-on connections are not loose.
Make certain that all wiring insulation is intact and free from damage.
In the case of AP models (non-electronic), check that the seals on the power
switch housing are not damaged, i.e. in perfect condition.
Check fuses (86),
10Piston/cylinder unit: dismantli c
Remove the screw (49) together with spacers (46 - 48). Slacken screw (108).
The cylinders (3,10) can be removed by gentie pressure applied simeltaneously
to the cylinder end covers (91,92) and the cranks (105). The lower cylinder
cover (104) can now be removed after the screws (93) have been released. [t is
not usually necessary to carry out any further dismantling operations.
Dirt and deposits can be removed with Tukewarm water {do not use any cleaning
agents!). The ball bearing in the integral piston rod/big end unit (105) is
sealed for life and must on no account come into contact with any cleaning
agents.
Black discolorations on the cylinder wall (96) or on the piston rod (105) are
a sign of satisfactory lubrication and should not be removed. The presence of
black powder is the result of extremely small, but sharp ridges on the rubbing
surfaces. Using emery cloth, the corresponding areas should be very carefully
polished in the direction of piston travel.
Warning: The cylinder wall and the piston rod have been very carefully
surface-treated and must be handled accordingly.
Piston/cylinder unit: reassembly
Warning: No lubricants may be used in the course of the following operations!
Reassemble the cylinders (3,10) in the reverse order to that described above.
When cold, it must be possible to slide the piston rods (105) up and down with
a force of between 6 - 10 kg. The cylinders must be refitted with great care
to ensure that the fine coating on the sleeve bearings is not damaged! Ensure
that the two cranks are offset by 90°.
Replacement of flat- and toothed belts
If, during servicing, excessive wear and tear (fine rubber particles) are
discovered and/or the inner belt face shows clear signs of wear, the belt in
question must. be replaced.
Flat belt:
Rotate the flat belt pulley (16) by hand and withdraw the belt (17) sideways.
Fit a new belt using the same method. The belt should centre itself when the
motor is running, i.e, it must not run over the edge of the drive shaft. If
it, does not do so, adjust the position of the motor via the screws (12). if
the belt centres itself during running, belt tension is automatically correct.
Toothed belt:
To change the toothed belt (26), first remove the cylinder (3). (Refer to the
instructions above.) Then unscrew (40) the bearing cover (41). Remove the
Flange ring (38) via the screws (39) and change the toothed belt (26). Belt
tension is determined by the distance and position of the toothed pulleys and
cannot be altered.
The drive mechanism is reassembled in the reverse order. Please pay careful
attention to the above instructions on cylinder/piston reassembly.
1Service call indicator
A. Models without a liquid crystal display:
‘The alarm indicates that moisture has gained access to the suction unit
(bottles not emptied/mechanical overflow valve disassembled).
~ Return the pump to the technical department. or a Madela service
agent. for dismantling and cleaning. The work to be done is
described on pages 10/11.
- he internal liquid sensor (see item 121 / pages 14/19) must
also be Cleaned and dried, otherwise the same service alarm will be
triggered off again.
The fuse on the back of the circuit board mst be replaced. This will
also cancel the visual service alarm.
B, Models with a liquid crystal display: (see illustration p. 13)
Several reasons for this alarn are possible. The cause can be traced
with the aid of the integral diagnostic program (*).
~ Press keys (7) and (2) on the low vacuum pump resp. (1) and (3) on
the vacuum extractor simultaneously. An "S" for "Service" will now
appear on the display. When released, a symbolic "G" (reason for the
alarm) will appear, together with a nunber with the following
meaning:
°
1
No service required (as there is no alam condition)
‘The battery on the PCB (for memory and timer) mst be
Moisture has gained access to the suction unit, e.g.
(bottles not enptied/mechanical overflow valve
disassenbled).
~ Return the pump to the technical department or a Medela
service agent for desmantling and cleaning. The work to be
done is described on pages 10/11.
~ the internal liquid sensor (see item 121 / pages
14/19) mast also be cleaned and dried, otherwise the
sane service alarm will be triggered off again.
- For resetting the electronics, switch on the pump (K)
and briefly short-circuit connections 39 and 54 on the
circuit board. This resetting operation is acknowledged by
neans of an acoustic signal and the electronic system is
now ready for operation again.
* Note:
During the diagnostic phase, the "Start/Stop" key cannot he used.
Retum to normal operation is effected by pressing key 7 (low vacuum
Pump) resp. key 1 (vacuum extractor) three tines.
12Correcting an electrical fault
A. The "Start/stop" key does not funtion:
= Check fuses (1). The main switch (x) must be set to "cn",
~ Probably, keys (7) and (2) on the low vacuum pump resp. (1) and (3)
on the vacuum extractor were pressed simtancously by mistake. This
Programs the electronics to a help setting for servicing purposes and
eliminates normal contro} functions.
Press the key (7) resp. key (1) three times. The pump can now be
controlled again using the "start/stop" key.
A Oa
(SG) &
B. Electronics indicate other malfunctions:
~ Tt is possible that the programming has been affected by a voltage
peak in the mains supply.
~ Connect the unit to the mains supply. Set switch (K) set to "ON"
- Briefly short-circuit PCB-connections 39 and 54, This resetting
operation is acknowledged by means of an acoustic signal.
~ If the electronic system does not return to normal operation even
after this has been done, the circuit board mst be replaced. Prior
to restarting, the resetting operation described above mist also be
carried out first (shorting PCB- connections 39/54).