Itemp Hart TMT182: Application
Itemp Hart TMT182: Application
TI078R/09/en
51002073
TMT182
Measuring system The iTEMP® HART® TMT182 temperature head transmitter is a 2-wire transmitter with analog output. It has
measurement input for resistance thermometers (RTD) in 2-, 3- or 4-wire connection, thermocouples and
voltage transmitters. Setting up of the TMT182 is done using the HART® -Protocol with hand operating
module (DXR275, DXR375) or PC (e.g. configuration software ReadWin® 2000 or FieldCare).
Input
Measured variable Temperature (temperature linear transmission behaviour), resistance and voltage
Measuring range Depending upon the sensor connection and input signal. The transmitter evaluates a number of different
measurement ranges.
Type of input
10 to 400 10
Resistance transmitter Resistance
10 to 2000 100
2 Endress+Hauser
TMT182
Output
Output signal Analog 4 to 20 mA, 20 to 4 mA
! Note!
Guaranteed values for setting "high alarm" ( 21 mA):
• Standard model: > 21.5 mA
• Advanced diagnostic model: 22.5 mA
Current limit 23 mA
Power supply
Electrical connection
T09-TMT182ZZ-04-06-XX-en-000
For the unit operation via HART® protocol (terminals 1 and 2) a minimum load resistance of 250 is necessary
in the signal circuit!
Endress+Hauser 3
TMT182
Performance characteristics
Response time 1 s (TC), 1.5 s (RTD)
! Note!
The accuracy data are typical values and correspond to a standard deviation of ± 3 (normal distribution), i.e.
99.8% of all the measured values achieve the given values or better values.
1) % is related to the adjusted measurement range. The value to be applied is the greater.
Influence of ambient Total temperature drift = input temperature drift + output temperature drift
temperature (temperature
drift) Effect on the accuracy when ambient temperature changes by 1 K (1.8 °F):
4 Endress+Hauser
TMT182
Influence of cold junction Pt100 DIN IEC 60751 Cl. B (internal cold junction with thermocouples TC)
Installation conditions
Installation instructions • Installation angle:
no limit
• Installation area:
Terminal head accord. to DIN 43 729 Form B; TAF10 field housing
Environment conditions
Ambient temperature -40 to +85 °C (-40 to 185 °F) for Ex-area, see Ex-certificate
limits
Condensation Permitted
Endress+Hauser 5
TMT182
Electromagnetic Interference immunity and interference emission as per IEC 61326 and NAMUR NE 21
compatibility (EMC)
Mechanical construction
Design, dimensions
R09-TMT182ZZ-06-06-xx-en-001
Material • Housing: PC
• Potting: PUR
Human interface
Display elements No display elements are present directly on the temperature transmitter.
The measured value display can be called up using the ReadWin® 2000 or FieldCare PC software.
Operating elements At the temperature transmitter no operating elements are available directly. The temperature transmitter will
be configured by remote operation with the PC software ReadWin® 2000 or FieldCare.
6 Endress+Hauser
TMT182
Hazardous area approvals For further details on the available Ex versions (ATEX, CSA, FM, etc.), please contact your nearest
Endress+Hauser sales organisation. All relevant data for hazardous areas can be found in separate Ex
documentation. If required, please request copies from us or your Endress+Hauser sales organisation.
Endress+Hauser 7
TMT182
Ordering information
Questionnaire
8 Endress+Hauser
TMT182
Product structure This information provides an overview of the order options available. The information is not exhaustive,
however, and may not be fully up to date. More detailed information is available from your local
Endress+Hauser representative.
Approval
A Non-hazardous area
B ATEX II1G EEx ia IIC T4/T5/T6
C FM IS, Class I, Div. 1+2, Group A, B, C, D
D CSA IS, Class I, Div. 1+2, Group A, B, C, D
E ATEX II3G Ex nA IIC T4/T5/T6
F ATEX II3D
G ATEX II1G EEx ia IIC T6, II3D
H ATEX II3G Ex nA IIC T4/T5/T6, II3D
I FM+CSA IS, NI, Class I, Div. 1+2, Group A, B, C, D
J CSA General Purpose
K IECEx Ex ia IIC T6/T5/T4
L TIIS Ex ia IIC T4
M TIIS Ex ia IIC T6
1 NEPSI Ex ia IIC T4...T6
2 NEPSI Ex nA II T4...T6
Configuration connection
A Factory setup Pt100 3-wire 0...100 °C
1 Thermocouple TC
2 RTD 2-wire
3 RTD 3-wire
4 RTD 4-wire
Configuration
A Factory setup Pt100 3-wire 0...100 °C)
B Measuring range, see additional specification
C TC configuration range, see questionnaire
D RTD configuration range, see questionnaire
Additional option
A Standard - DIN mounting set
B Works calibration certificate 6-point, DIN mounting set
C Diagnostic, advanced, DIN mounting set
K US - M4 mounting screws
L Diagnostic, advanced, US - M4 mounting screws
Endress+Hauser 9
TMT182
Accessories
Commubox FXA191 (RS232) or FXA195 (USB)
Order code: FXA191-... or FXA195-...
PC-operating software: ReadWin® 2000 or FieldCare
ReadWin® 2000 can be downloaded free of charge from the internet from the following address:
www.endress.com/readwin
Hand operating module ’HART® Communicator DXR375’, Order code: DXR375-...
DIN rail clip according to IEC 60715 (TH35) for head transmitter mounting
Order code: 51000856
Field housing TAF10 for Endress+Hauser head transmitter, aluminum, IP 66,
dimensions W x H x D: 100 x 100 x 60 mm (3.94" x 3.94" x 2.36")
Order code: TAF10-...
Documentation
Operating short manual iTEMP® HART® TMT182 (KA142R/09/a3)
Additional documentation for use in explosion-hazardous areas:
ATEX II1G: XA006R/09/a3
ATEX II3G: XA011R/09/a3
ATEX II3D: XA027R/09/a3
Operating short manual TAF10 Field housing (KA093R/09/a2)
Functional safety manual TMT182 (SD006R/09/en)
10 Endress+Hauser
TMT182
Endress+Hauser 11
TMT182
Instruments International
Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland
TI078R/09/en/13.10
51002073
FM+SGML 6.0