Dacell DN-10W, DN-20W, DN-25W, DN-30W, DN-50W Digital Indicator Manual
Dacell DN-10W, DN-20W, DN-25W, DN-30W, DN-50W Digital Indicator Manual
USER’S MANUAL
DN-10W (VER 5.0)
DN-20W (VER 5.0)
DN-25W (VER 1.0)
DN-30W (VER 1.0)
DN-50W (VER 1.5)
We thank you for using our product. Please refer to this manual or contact our office if
you find any problems during using our product.
This product is an indicator that emplifies the micro voltage signals from each sensor
and display in digital. It is mainly used to measure physical quantities of the items using
strain guage such as loadcell, pressure sensor, displacement sensor and torque sensor.
This product has the following special features.
You can do set-up the maxium and lower limit value by the keys on the front panel
and output contact point signal of the rear panel.
All the set-up values will be memorized on the flash memory so the inputted data can
be saved and no need to do the re-setting even in case of the interruption of the
electric power or power disconnection.
This function is for automatic reset in case the system is stopped due to the external
factors such as noise.
2
2. Attention
For the efficient and safe use, please carefully read and be fully aware of the
following details before using this product.
It is strictly forbidden to use this product for any other purpose of use or to attempt to
make any alteration on this product.
During calibration, Do not input free-load state and real-weight load until it becomes
stable. Pressing Enter Key in unstable condition leads to calibration error.
Do not press any Key in use at one's discretion. Please refer to 7. Setting-up mode for
the function and method of Key.
3
3. Specification
Specification model DN-50W DN-10W DN-20W DN-25W DN-30W
0~10V,
SIGNAL 1 ~ 3mv/V 0~2Vrms pulse
4~20mA
4
4. Pront Panel (common feature)
5
5. Rear Panel (common feature)
Please check the location of terminal and its use. While the projected button is being
pressed, please insert the cable into the lower hole completely. As soon as you release
the button, the connection will be completed. At the point, please slightly pull the calbe
and check whether the cable is come off or not.
(The most suitable calbe is Φ 0.5~1. Linking cable must be brazing or used with I
terminal)
(DN-10W,DN-25W,DN-30W,DN-50W)
SW1 : 10Hz
SW2 : 100Hz
SW3 : 1kHz
○
20 DN-20W Sensor supply voltage & sensor output select switch
7
6. Wiring Diagram
1) DN-10W, DN-50W
2) DN-20W
8
3) DN-25W
4) DN-30W
9
7. How to use Hold mode
Hold mode is largely divided into Peak Hold and Sample Hold. Please select Analog,
Digital or Display Hold as per your purpose of use (Hi or Low Speed)
To input Hold, you can use Hold key on the front panel or external input. For the
operation method, please refer to the drawing below.
DN-30W
1) Peak hold : To Hold the maxium value among the measured values.
· Peak hold mode : It is to hold and display the maximum value of (+) direction.
2) Sample hold : It is to hold and display the value at the time of Hold signal input
among the measured values.
DN-50W
1) Peak Hold : To Hold the maxium value among the measured values.
Analog peak hold mode : It is a high-speed Peak hold. It is to hold and display the
maximum value of (+) direction.
Display peak hold mode : It is a low-sped Peak hold. It is to hold and display the
maximum value of (+) direction.
Absolute peak hold mode : It is a low-speed Peak hold. It is to hold and display the
maximum value of the absolute value (+/-).
2) Sample hold : It is to hold and display the value at the time of Hold signal input
among the measured values.
Digital sample hold mode : It is operated by high-speed sample hold.
· Display sample hold mode : It is operated by low-speed sample hold.
10
M M
e e
② ①
a a
s s
u
r u
e r
e
① − : Sensor input value ②
d d
v v
a a
② − : Display value & ①
l
u l
u
e e
Analog ouput value
Time Time
On
Front Panel HOLD LED
Off
On
Front Panel HOLD Key off
11
7-3. How to use comparison ouput function
For comparison output function, there are 3 different modes such as Decision, High
limit, Low limit. It displays through the relay of rear panel comparing each setup value.
On High limit and Low limit mode, Hysteresis can be used.
To set up upper limit (High) and lower limit(Low), please use the key on the front panel.
1) Decision mode : Measured Value ≤ Lower limit setup value ⇒ RY1 ON (HI)
Measured Value ≥ Upper limit setup value ⇒ RY2 ON (LO)
Lower limit setup value < Measured value < Upper limit setup
value ⇒ RY3 ON (OK)
2) High limit mode : Measured Value ≥ RY1 setup value ⇒ RY1 ON
Measured Value ≥ RY2 setup value ⇒ RY2 ON
Measured Value ≥ RY3 setup value ⇒ RY3 ON
Measured Value ≥ RY4 setup value ⇒ RY4 ON
Measured Value < RY1 setup value – Hysteresis value⇒ RY1 OFF
Measured Value< RY2 setup value - Hysteresis value⇒ RY2 OFF
Measured Value < RY3 setup value – Hysteresis value⇒ RY3 OFF
Measured Value< RY4 setup value - Hysteresis value⇒ RY4OFF
3) Low limit mode : Measured Value ≤ RY1 setup value ⇒ RY1 ON
Measured Value ≤ RY2 setup value ⇒ RY2 ON
Measured Value ≤ RY3 setup value ⇒ RY3 ON
Measured Value ≤ RY4 setup value ⇒ RY4 ON
Measured Value > RY1 setup value + Hysteresis value⇒ RY1 OFF
Measured Value > RY2 setup value + Hysteresis value⇒ RY2 OFF
Measured Value > RY3 setup value + Hysteresis value⇒ RY3 OFF
Measured Value > RY4 setup value + Hysteresis value⇒ RY4 OFF
4) Low & High limit mode
Measured Value ≤ RY1 setup value ⇒ RY1 ON
Measured Value ≤ RY2 setup value ⇒ RY2 ON
Measured Value ≥ RY3 setup value ⇒ RY3 ON
Measured Value ≥ RY4 setup value ⇒ RY4 ON
Measured Value > RY1 setup value + Hysteresis value ⇒ RY1 OFF
Measured Value > RY2 setup value + Hysteresis value ⇒ RY2 OFF
Measured Value < RY3 setup value - Hysteresis value ⇒ RY3 OFF
Measured Value < RY4 setup value - Hysteresis value ⇒ RY4 OFF
12
13
8. Setting Modes
1) DN-10W, DN-50W
For Setting mode, there are 4 different types of mode such as Function mode, Digital
calibration mode, Actual load calibration mode and SPAN constant calibration mode
Measuring Mode
Function mode
①
END
Digital calibration
②
mode
END
Actual load
③
calibration mode
END
SPAN constant
④
calibration mode
END
① Function mode: Access to each function setup mode. Please refer 8-2 Function
Mode.
② Digital calibration mode: It is to calibrate into the sensor’s output value. No need
to prepare for the actual load (standard weight). Please refer 8-3 How to calibrate.
③ Actual load calibration mode: It is to calibrate by adding the actual load (Standard
weight or the load you know). Please refer 8-4 How to calibrate.
④ SPAN constant calibration mode: It is to calibrate with the S.CAL value written
down for load calibration. Please refer 8-5 How to calibrate.
14
2) DN-20W
For setting mode, there are 3 different types of mode such as Function mode, Digital
calibration mode, Actual load calibration mode.
Mesuring mode
Function mode
①
END
Digital calibration
②
mode
END
Actual load
③
calibration mode
END
① Function mode
Acess to each function setup mode. Please refer 8-2 function mode.
② Digital calibration mode
It is to calibrate into the sensor’s output value. No need to prepare for the actual
load (stand weight). Please refer 8-3 how to calibrate.
③ Actual load calibration mode
It is to calibrate by adding the actual load (standard weight or the load you know).
Please refer 8-4 how to calibrate.
15
3) DN-25W
For setting mode, there are 2 different types of mode such as Function mode, Actual
load calibration mode.
Measuring mode
Function mode
①
END
Actual load
②
calibration mode
END
① Function mode
Acess to each function setup mode. Please refer 8-2 function mode.
② Actual load calibration mode
It is to calibrate by adding the actual load (standard weight or the load you
know). Please refer 8-3 how to calibrate.
16
8-1. Function mode
END
17
2) How to set RELAY data
Measuring Mode
press key
Input
END
18
3) Function mode table
F-01 Ο Ο Ο Ο(Pulse) Ο
F-02 Ο Ο Ο Ο Ο
F-03 Ο Ο Ο Ο Ο
F-04 Ο Ο Ο Ο Ο
F-05 Ο Ο Ο Ο Ο
F-06 Ο Ο Ο Ο Ο
F-07 Ο Ο Ο Ο Ο
F-08 Ο Ο Ο Ο Ο
F-09 Ο Ο Ο Ο Ο
F-10 Ο Ο Ο Ο Ο
F-11 Ο - Ο - Ο
F-12 Ο - Ο - Ο
F-13 Ο - - - Ο
F-14 Ο Ο Ο Ο Ο
19
# Function mode list #
F-01
(DN- Pulse/circle 0 ~ 360 60
30W)
F-04 Hold mode Sample hold, Peak hold, Absolute peak hold Peak hold
F-06 Hysteresis 0 ~ 99 0
F-07
(DN- DAC mode Display DAC, High speed DAC Display DAC
50W)
F-09 ID Number 0 ~ 32 0
F-10 Baud rate & PRINT 2400, 4800, 9600, 19200, print 9600
F-14 Key Disabling Zero key, Lo key, Hi key, Hold key 0000
20
F-01
1) Decimal point (Decimal point Set-up): DN-10W, DN-20W, DN-25W, DN-50W
(Standard setup value : 1)
Display data
0 00000 : No decimal point
1 0000.0 : One decimal place
2 000.00 : Two decimal places
3 00.000 : Three decimal places
21
F-04. Hold mode
1) DN-10W, DN-20W, DN-25W
(Standard setup value: 1)
Display data Setting
: To hold the display value at the time of Hold
0 Sample Hold
signal input.
: To hold the maximum value of display values
1 Peak Hold (+)
during Hold signal input.
: To hold the maximum absolute value of display
2 Absolute Peak Hold(+/-)
values during Hold signal input.
2) DN-30W
(Standard setup value: 1)
Display data Setting
0 Sample Hold : To hold the display value at the time of Hold
signal input.
1 Peak Hold (+) : To hold the maximum value of display values
during Hold signal input.
3) DN-50W
Display data Setting
0 Display sample Hold : To hold the display value at the time of Hold signal input.
1 Digital sample Hold : To hold the A/D convertion value at the time of Hold
signal input.
2 Diplay Peak Hold : To hold the maximum value of display values during Hold signal
input.
3 Absolute Peak Hold : To hold the maximum absolute value of display values during
Hold signal input.
4 Analog Peak Hold : To hold the maximum value of sensor input during Hold signal
input.
22
F-06. Hysteresis
(Standard setup value : 00)
Display data Setting
00~99 00 : Hysteresis – not used
01 ~99 : Hysteresis – used (Decision mode is not applied)
23
00~32 01 ~32 : Device number is set-up (Command mode : Transmit data by
command)
24
F-10. Baud rate & Print (Communication Speed and print Setup)
1) DN-10W, DN-20W, DN-25W, DN-30W
(Standard setup value: 9.60)
Display data Setting Stream mode Command mode
2.40 2400 bps O O
4.80 4800 bps O O
9.60 9600 bps O O
19.20 19200 bps O X
PRINT PRINT DATA OUT (PT-100) - -
2) DN-50W
(Standard setup value : 9.60)
Display data Setting Stream mode Command mode
2.40 2400 bps O O
4.80 4800 bps O O
9.60 9600 bps O O
19.20 19200 bps O X
38.40 38400 bps O X
57.60 57600 bps O X
PRINT PRINT DATA OUT (PT-100) - -
F-12. Auto zero tracking time (Auto Zero tracking time setup)
(Standard setup value : 0.0)
Display data Setting
00 0.0 : Auto Zero is not used.
~
5.0 0.1 ~5.0 : Set up the auto zero operation time (0.1 ~ 5.0 sec)
25
F-13. Force unit (Conversion Unit setup)
(Standard setup value : 0)
Display data Setting
0 kg.f kg/㎠
1 N kgf × 9.8 - -
2 ℓb kgf × 2.2 - -
3 - - Bar kg/㎠ × 0.98
4 - - MPa kg/㎠ × 0.098
Note) During calibration, set up the unit you want after calibrating it into Kg.
<Set-up example>
1001 : Zero & Hold key Lock, Hi and Lo key Release
26
8-2. Digital calibration (Calibration by sensor output value)
1) DN-10W, DN-50W
At the time of purchasing sensor, the rated capacity (R.C) and rated output (R.O)
declared on the calibration sheet can be used for the calibration for easier calibration.
Calibration Completed.
27
2) DN-20W
It will be able to calibration with output type and rated capacity setting.
Measuring mode
: 0 ~ 5V
: 0 ~ 10V
: 4 ~ 20mA
28
8-3. Actual load calibration
1) DN-10W, DN-50W
This is a calibration method by adding actual load on the sensor. Standard weight is Needed.
Input the rated output value by using and key. And then, press
key. (Input 3.0000 if donot know the rated output of sensor)
After adding the actual load (standard weight) on the sensor, press key.
This is a calibration method by adding actual load on the sensor. Standard weight is needed.
And this can calibrate 1~7 stages, and improve linearity of sensor.
Measuring mode
Calibration end
30
3) DN-25W
This is a calibration method by adding actual load on the sensor. Standard weight is needed.
And this can calibrate 1~7 stages, and improve linearity of sensor.
Measuring mode
Calibration end
31
8-4. SPAN constant claibration (Calibratuib by Shunt CAL Value)
1) DN-10W, DN-50W
It is to calibrate with the S.CAL value written down for load calibration. You can
calibrate without any standard weight.
Measuring Mode State
Input the rated output value by using and key. And then, press
key. (Input 3.0000 if donot know the rated output of sensor)
Input S.CAL value by using and key. And then press key.
Calibration Completed.
32
8-5 Lock Set-up
You can prevent any accidental operation due to the unnecessary key control by Lock
set-up. After finishing calibration, it is recommended to set the Lock.
At the first stage, please start while the power is OFF.
Related Function when Lock is set : Function related to calibration.
Power is OFF
or display
Set-up is completed.
33
9. Product Inspection
34
10. OPTION
#Option-02 (RS232C)
Since RS232C Interface is very sensitive of electric noise. So please do the wiring
from AC Power and electric wires separately. Also you must use the shield calbe
always.
Indicator Host PC
TX(Transmission Data) RXD(Receive Datea), No.2 Pin
RX(Receive Datea) TXD(Transmission Data), No.3 Pin
GN(Ground) GND(Ground), 5 Pin
1. TYPE : EIA-232C
Method : Half-duplex, asynchronous method.
2. Baud-rate : Select one of 2400, 4800, 9600, 19200, 38400, 57600bps
3. Parity : No Parity
4. Data bit : 8 bit
5. Stop bit : 1bit
6. Stream mode (Ex. Data +1234.5 transmission )
1) BYTE1, BYTE2
. DATA Stable :ST . DATA Unstable :US
. DATA OVERFLOW :OL . DATA UNDERFLOW :UL
35
2) BYTE3 ~ BYTE6 : fixed character (, N T ,)
3) BYTE7 ~ BYTE14 : DATA 8 BYTE (including +/- )
4) BYTE15 : CARRIAGE RETURN
5) BYTE16 : LINE FEED
7. Command mode : Please setup as RS485 mode and use.( Refer to OP-
03:RS485 )
36
#Option-03 (RS485)
Since RS485 Interface is very sensitive of electric noise. So please do the wiring from
AC Power and electric wires separately. Also you must use the shield calbe always.
1. TYPE : RS485
2. Method : Half-duplex, asynchronous method
3. Baud-rate : Select one of 2400, 4800, 9600bps
4. Parity : No Parity
5. Data bit : 8 bit
6. Stop bit : 1bit
8. Command Chart
Command
Description
ASCII HEX
Transmit the current value of order
P 50H
equipment.
H 48H Hold for order equipment.
R 52H Release hold for order equipment.
Operate the current value of order
Z 5AH
equipment as ZERO.
37
9. Transmission Data Form (INDICATOR -> PC)
38