Data sheet
testo 435
Multi-function
measuring instrument
testo 435 – The allrounder for
ventilation and indoor air quality
Large selection of probes (optional):
IAQ probe for evaluating Indoor Air Quality                                                                          °C
Thermal probes with integrated temperature and air
humidity measurement
                                                                                                                    %RH
Vane and hot wire probes
Integrated differential pressure probe for Pitot tube
measurement (see versions)                                                                                          m/s
Wireless probe for temperature and humidity (see versions)
Easy operation with user profiles                                                                                   hPa
PC software for the analysis, archiving and documentation
                                                                                                                    ppm
of measurement data (see versions)                                                                                  CO2
                                                                                                                    Lux
All HVAC measurement parameters                                   determination of volume flow, you can use all possiblities
The multi-function measuring instrument testo 435 is your         for the measurement of flow velocity – such as thermal
reliable partner for analyzing Indoor Air Quality. Indoor Air     probes, vanes and Pitot tubes.
Quality has a crucial influence on the well-being of humans
at their workplaces, and is also a determining and important      The right instrument for every application
factor in storage and production processes.                       The new testo 435 is available in four versions. Depending
In addition to this, the Indoor Air Quality signals whether the   on the application, you can choose from versions with
HVAC system is working at optimum energy level, or needs          built-in differential pressure measurement as well as
to be adjusted using the testo 435. The parameters CO2,           versions with additional instrument functions such as
relative humidity and room temperature are available for the      instrument memory, PC software and an extended range of
evaluation of Indoor Air Quality.                                 probes.
In addition to this, absolute pressure, draught, Lux,
U-value and surface temperature can be measured. For the
                                                                                                     www.testo.com
testo 435
Technical data
  testo 435-1                                                   testo 435-2
  testo 435-1 multi-functional measurement                      testo 435-2 multi-function measuring instrument for
  instrument for A/C, ventilation and Indoor Air                air conditioning, ventilation and Indoor Air Quality with
  Quality, battery and calibration protocol included            data memory, PC software and USB data transfer cable,
                                                                includes battery and calibration protocol
  Part no. 0560 4351                                            Part no. 0563 4352
  testo 435-3                                                   testo 435-4
                                                                testo 435-4 multi-function measuring instrument
  testo 435-3 multi-functional measurement                      with built-in differential pressure measurement for air
  instrument with built-in differential pressure                conditioning, ventilation and Indoor Air Quality, with
  measurement for A/C, ventilation and Indoor Air               data memory, PC software and USB data transfer
  Quality, battery and calibration protocol included            cable, includes battery and calibration protocol
  Part no. 0560 4353                                            Part no. 0563 4354
Improved convenience of operation thanks to user profiles
The operation of testo 435 is easy and efficient: User
profiles are stored for the typical applications duct
measurement and IAQ measurement. This makes
complicated programming of the measuring instrument
unnecessary.
                                                              Measure temperature and                 2 connections for external probes
                                                              humidity values wirelessly over
Secure measurement data monitoring                            a distance of up to 20m without
                                                              obstruction
The measurement data reports provide the customer with
the data from the duct, long-term and degree of turbulence
measurement. The company logo can be integrated into
the form. The testo 435-1 and testo 435-3 allow the
measurement values to be printed cyclically on the Testo
fast printer.
Flexibility thanks to wireless probes                           General technical data
In addition to classical probes on wires, a wireless
                                                                Operating temperature     -20 to +50 °C
measurement up to a distance of 20 m (without obstruction)
                                                                Storage temperature       -30 to +70 °C
is possible. Damage to the wire or hindrances in usage are
                                                                Dimensions                220 x 74 x 46 mm
thus eliminated. A maximum of three wireless probes can
                                                                Battery type              Alkali manganese, mignon, Type AA
be recorded and displayed with testo 435. The wireless
                                                                Battery life              200 h (typical vane measurement)
probes are for the measurement parameters temperature
                                                                Weight                    428 g
and, depending on the instrument type, humidity. The
                                                                Housing material          ABS/TPE/Metal
optional, easily plugged-in radio module can be retrofitted
                                                                Protection class          IP54
at any time.
                                                                Warranty                  2 years
testo 435
Technical data
 Overview of versions testo 435
 The table provides a fast overview of the connectable probes and the instrument equipment per version.
 Connectable probes (optional)                                                     testo 435-1     testo 435-2      testo 435-3           testo 435-4
 IAQ probe for the measurement of CO2, air temperature, indoor air humidity              X                 X                X                 X
 and absolute pressure
 Thermal flow velocity probe with integrated temperature and air humidity                X                 X                X                 X
 measurement
 Vane and hot wire probes                                                                X                 X                X                 X
 Temperature probe for immersion/penetration, air and surface measurements               X                 X                X                 X
 Wireless probes for temperature measurements                                            X                 X                X                 X
 Ambient CO probe                                                                        X                 X                X                 X
 Absolute pressure probe                                                                 X                 X                X                 X
 Integrated differential pressure measurement for flow velocity measurement                                                 X                 X
 with Pitot tube and for monitoring filters (not retro-fittable)
 Comfort probe for degree of turbulence measurement for the objective                                      X                                  X
 evaluation of the air velocity in a room
 Humidity probe for air temperature and air humidity measurements                                          X                                  X
 Wireless probe for air temperature and air humidity measurement                                           X                                  X
 Lux probe for the measurement of light intensity                                                          X                                  X
 Temperature probe for determining U-value                                                                 X                                  X
 Instrument equipment
 Easy operation with user profiles                                                       X                 X                X                 X
 Illuminated display                                                                     X                 X                X                 X
 Testo fast printer for documenting measurement data (optional)                          X                 X                X                 X
 Instrument store for 10000 measurement values (not retro-fittable)                                        X                                  X
 PC software for the analysis, archiving and documentation of the readings                                 X                                  X
                                     testo 435-1/-2/-3/-4
 Sensor types                        NTC                                Type K (NiCr-Ni)          Type T (Cu-CuNi)              Testo humid. sensor,
                                                                                                                                cap.
 Measuring range                     -50 to +150 °C                     -200 to +1370 °C          -200 to +400 °C               0 to +100 %RH
 Accuracy                            ±0.2 °C(-25 to +74.9 °C)           ±0.3 °C (-60 to +60 °C)   ±0.3 °C (-60 to +60 °C)       See probe data
 ±1 digit                            ±0.4 °C (-50 to -25.1 °C)          ±(0.2 °C +0.5% of m.v.)   ±(0.2 °C +0.5% of m.v.)
                                     ±0.4 °C (+75 to +99.9 °C)          (remaining range)         (remaining range)
                                     ±0.5% of m.v. (remaining range)
 Resolution                          0.1 °C                             0.1 °C                    0.1 °C                        0.1 %RH
                                     Vane                               Hot wire                  Absolute pressure probe       CO2 (IAQ probe)
 Measuring range                     0 to +60 m/s                       0 to +20 m/s              0 to +2000 hPa                0 to +10000 ppm CO2
 Accuracy ±1 digit                   See probe data                     See probe data            See probe data                See probe data
 Resolution                          0.01m/s (60 + 100 mm Vane)         0.01 m/s                  0.1 hPa                       1 ppm CO2
                                     0.1 m/s (16 mm Vane)
                                     testo 435-2/-4                                          testo 435-3/-4
 Sensor types                        Lux                                                     Differential pressure probe, internal
 Measuring range                     0 to +100000 Lux                                        0 to +25 hPa
 Accuracy                            See probe data                                          ±0.02 hPa (0 to +2 hPa)
 ±1 digit                                                                                    ±1% of m.v. (remaining range)
 Resolution / Overload               1 Lux; 0.1 Hz                                           0.01 hPa / 200 hPa
testo 435
Accessories
 Transport and Protection                                                                                                    Part no.
 Service case for measuring instrument, probe and accessories, dimensions 454 x 319 x 135 mm                                0516 1035
 Service case for measuring instrument, probe and accessories, dimensions 518 x 398 x 155 mm                                0516 1435
 Additional accessories and spare parts
 testovent 410, volume flow funnel, Ø 340 mm/330x330 mm, incl. case                                                         0554 0410
 testovent 415, volume flow funnel, Ø 210 mm/210x210 mm, incl. case                                                         0554 0415
 testovent 417 funnel set for plate outlets (Ø 200 mm) and funnel for ventilator (330 x 330 mm) for ingoing and             0563 4170
 outgoing air
 Connection hose; silicone; length 5 m; max. load 700 hPa (mbar)                                                            0554 0440
 Connection hose silicone-free for differential pressure measurement, length 5 m, load up to maximum 700 hPa, (mbar)        0554 0453
 Control and adjustment set for Testo humidity probes, salt solution with 11.3% RH and 75.3% RH, incl. adapter for          0554 0660
 Testo humidity probes, quick checks or calibration of humidity probe
 Sintered PTFE filter, Ø 12 mm, for corrosive media, High humidity range (long-term measurements), high flow velocities.    0554 0756
 Stainless steel sintered filter, pore size 100 µm, probe protection in dusty atmospheres or higher flow velocities         0554 0641
 Plug-in mains adapter, 5 VDC 500 mA with European adapter, 100-250 VAC, 50-60 Hz                                           0554 0447
 Lithium battery button cell, CR2032 AA batteries for radio handle                                                          0515 5028
 Adhesive material for fixing and sealing                                                                                   0554 0761
 Printer and Accessories
 Testo fast printer IrDA with wireless infrared interface; 1 roll thermal paper; 4 AA batteries, for printing out           0554 0549
 measurements on site
 Spare thermal paper for printer (6 rolls), permanent ink                                                                   0554 0568
 measurement data documentation legible for up to 10 years
 External fast charger for 1-4 AA rech. batteries, incl. 4 Ni-MH rech. batteries with individual cell charging and charge   0554 0610
 control display, incl. impulse trickle charging, integrated discharge function, with built-in international mains plug,
 100-240 V, 300 mA, 50/60 Hz
 Calibration Certificates
 ISO calibration certificate/temperature                                                                                    0520 0071
 meas. instr. with surface probe; calibration points +60°C; +120°C; +180°C
 ISO calibration certificate humidity                                                                                       0520 0006
 Calibration points 11.3 %RH and 75.3 %RH at +25 °C
 ISO calibration certificate pressure                                                                                       0520 0005
 differential pressure; 5 points distributed over meas. range
 ISO calibration certificate velocity                                                                                       0520 0024
 hot wire, vane anemometer; calibration points 0.3; 0.5; 0.8; 1.5 m/s
 ISO calibration certificate velocity                                                                                       0520 0004
 hot wire, vane anemometer, Pitot tube; calibration points 1; 2; 5; 10 m/s
 ISO calibration certificate velocity                                                                                       0520 0034
 hot wire, vane anemometer, Pitot tube; calibration points 5; 10; 15; 20 m/s
 ISO calibration certificate/light                                                                                          0520 0010
 Calibration points 0; 500; 1000; 2000; 4000 Lux
 ISO calibration certificate/CO2
 CO2 probes; calibration points 0; 1000; 5000 ppm
testo 435
Probes
 Probe type                                    Dimensions                          Measuring         Accuracy                       t99    Part no.
                                               Probe shaft/probe shaft tip         range
 IAQ probes (testo 435-1/-2/-3/-4)
 IAQ probe to assess Indoor                                                        0 to +50 °C       ±0.3 °C                              0632 1535
                                                                                   0 to +100 %RH     ±2 %RH (+2 to +98 %RH)
 Air Quality, CO2, humidity,                                                       0 to +10000       ±(75 ppm CO2 ±3% of m.v.)
 temperature and absolute pressure                                                 ppm CO2           (0 to +5000 ppm CO2)
 measurement, with desk-top stand                                                  +600 to +1150     ±(150 ppm CO2 ±5% of m.v.)
                                                                                   hPa               (+5001 to +10000 ppm CO2)
                                                                                                     ±10 hPa
 Ambient CO probe, for detecting                                                   0 to +500 ppm     ±5% of m.v. (+100.1 to +500          0632 1235
                                                                                   CO                ppm CO)
 CO in buildings and rooms                                                                           ±5 ppm CO (0 to +100 ppm CO)
 Globe probe Ø 150mm, TC Type K,                                                   0 to +120 °C      Class 1                              0602 0743
 for measuring radiant heat
 Flow velocity probes (testo 435-1/-2/-3/-4)
 Thermal velocity probe with                                                       -20 to +70 °C     ±0.3 °C                              0635 1535
                                                                                   0 to +100 %RH     ±2 %RH (+2 to +98 %RH)
 built-in temperature and humidity                                                 0 to +20 m/s      ±(0.03 m/s +4% of m.v.)
 measurement, Ø 12 mm, with
 telescopic handle (max. 745 mm)
 Vane meas. probe, 16 mm                                                           +0.6 to +40 m/s   ±(0.2 m/s +1.5% of m.v.)             0635 9535
 diameter, with telescopic handle                                                  Operating
 max. 890 mm, e.g. for meas. in                                                    temperature
 ducts, can be used from 0 to +60                                                  0 to +60 °C
 °C
 Vane meas. probe, 60 mm                                                           +0.25 to +20      ±(0.1 m/s +1.5% of m.v.)             0635 9335
                                                                                   m/s
 diameter, with telescopic handle
 max. 910 mm, e.g. for meas. at                                                    Operating
 duct exit, can be used from 0 to                                                  temperature
                                                                                   0 to +60 °C
 +60 °C
 Hot wire probe for m/s and °C,                                                    0 to +20 m/s      ±(0.03 m/s +5% of m.v.)              0635 1025
                                                                                   -20 to +70 °C     ±0.3 °C (-20 to +70 °C)
 Ø probe head 7.5 mm, with
 telescopic handle (max. 820 mm)
 Funnel measurement (testo 435-1/-2/-3/-4)
 Vane meas. probe, 100 mm                                                          +0.3 to +20 m/s   ±(0.1 m/s +1.5% of m.v.)             0635 9435
                                                                                   0 to +50 °C       ±0.5 °C
 diameter, for measurements with
 funnel set 0563 4170
 testovent 417 funnel set for plate                                                                                                       0563 4170
 outlets (Ø 200 mm) and funnel
 for ventilator (330 x 330 mm) for
 ingoing and outgoing air
 Flow straightener testovent 417                                                                                                          0554 4172
 Flow straigthener set testovent 417 consisting of funnel set testovent 417 and flow straigthener testovent 417                           0554 4173
 Absolute pressure probes (testo 435-1/-2/-3/-4)
 Absolute pressure probe 2000 hPa                                                  0 to +2000 hPa    ±5 hPa                               0638 1835
 2) More temperature probes on the internet at www.testo.com
testo 435
Probes
 Probe type                                      Dimensions                                           Measuring         Accuracy                           t99      Part no.
                                                 Probe shaft/probe shaft tip                          range
 Air probes     2)
                     (testo 435-1/-2/-3/-4)
 Efficient, robust NTC air probe,                                                                     -50 to +125 °C    ±0.2 °C (-25 to +80 °C)            60 s    0613 1712
                                                                       115 mm              50 mm                        ±0.4 °C (remaining range)
 Fixed cable 1.2 m
                                                                       Ø 5 mm              Ø 4 mm
 Surface probes         2)
                             (testo 435-1/-2/-3/-4)
 Fast-action surface probe with                                                                       -60 to +300 °C    Class 2 1)                         3s      0602 0393
                                                                       115 mm
 sprung thermocouple strip, also
 for uneven surfaces, measurement
 range short-term to +500°C, TC                                        Ø 5 mm              Ø 12 mm
 Type K, Fixed cable
 Pipe wrap probe for pipe diameter                                                                    -60 to +130 °C    Class 2 1)                         5s      0602 4592
 5 to 65 mm, with exchangeable
 measuring head. Meas. range
 short-term up to +280 °C, TC Type
 K, Fixed cable
 Clamp probe for measurements on                                                                      -50 to +100 °C    Class 2 1)                         5s      0602 4692
 pipes, pipe diameter 15 to 25 mm
 (max. 1“), meas. range short-term
 up to +130°C, TC Type K, Fixed
 cable
 Immers./penetr. probes            2)
                                        (testo 435-1/-2/-3/-4)
 Waterproof immersion/penetration                                                                     -60 to +400 °C    Class 2 1)                         7s      0602 1293
                                                                       114 mm              50 mm
 probe, TC Type K, Fixed cable
 1.2 m
                                                                       Ø 5 mm              Ø 3.7 mm
 IAQ probes (testo 435-2/-4)
 Comfort level probe for degree                                                                       0 to +50 °C       ±0.3 °C                                    0628 0109
                                                                   max. 820 mm                        0 to +5 m/s       ±(0.03 m/s +4% of m.v.)
 of turbulence measurement with
 telescopic handle (max. 820
 mm) and stand, meets EN 13779
 requirements
 Lux probe, for measuring light                                                                       0 to 100.000      Accuracy acc. to DIN 13032-1:              0635 0545
                                                                                                      Lux               f1 = 6% = V(Lambda)
 intensity                                                                                            0 to 300 Hz       adjustment
                                                                                                                        f2 = 5% = cos-like weighting,
                                                                                                                        Class C
 Humidity probes (testo 435-2/-4)
 Humidity/temperature probe                                                                           -20 to +70 °C     ±0.3 °C                                    0636 9735
                                                                                                      0 to +100 %RH     ±2 %RH (+2 to +98 %RH)
                                                                       Ø 12 mm
 Surface probes         2)
                             (testo 435-2/-4)
 Temperature probe to determine                                                                       -20 to +70 °C     Class 11)                                  0614 1635
                                                                                                                        U-value: ±0.1 ±2% of fsv*
 U-value, triple sensor system
 for measuring wall temperature,                                                                      Note: An additional probe for measuring outer temperatures
 modelling clay included                                                                              is required when determining the U-value e.g. 0602 1793 or
                                                                                                      0613 1002.
                                                                                                      *when used with an NTC or wireless humidity probe for
                                                                                                      measuring outside temperature and 20 K difference between
                                                                                                      the air inside and outside
 Prandtl‘s Pitot tubes (testo 435-3/-4)
 Pitot tube, 350 mm long                                                                                                Operating temperature                      0635 2145
                                                               350 mm / 500 mm / 1000 mm                                0 to +600 °C
 Pitot tube, 500 mm long                                                                                                                                           0635 2045
                                                                       Ø 7 mm
 Pitot tube, 1000 mm long                                                                                                                                          0635 2345
 1) According to standard EN 60584-2, the accuracy of Class 2 refers to -40 to +1200 °C.
 2) More temperature probes on the internet at www.testo.com
testo 435
Radio probes
 Radio handles with probe head for air-/ immersion-penetration-meas.                                                                     Part no.
 Radio handle for plug-in probe heads, incl. T/C adapter, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT,               0554 0189
 DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO; Radio freq. 869.85 MHz FSK
 T/C probe head for air/immersion/penetration measurement (T/C Type K)                                                                  0602 0293
 Radio handle for plug-in probe heads, incl. T/C adapter, approval for USA, CA, CL; Radio freq. 915.00 MHz FSK                          0554 0191
 T/C probe head for air/immersion/penetration measurement (T/C Type K)                                                                  0602 0293
 Dimensions                                 Measuring         Accuracy                                    Resolution          t99
 Probe shaft/probe shaft tip                range
                    100 mm      30 mm       -50 to +350 °C    Radio handle:                               0.1 °C (-50 to      t99 (in
                                            Short-term to     ±(0.5 °C +0.3% of m.v.) (-40 to +500 °C)    +199.9 °C)          water)
                    Ø 5 mm      Ø 3,4 mm    +500 °C           ±(0.7 °C +0.5% of m.v.) (remaining range)   1.0 °C (remaining   10 s
                                                              T/C probe head: Class 2                     range)
 Radio handles with probe head for surface measurement                                                                                   Part no.
 Radio handle for plug-in probe heads, incl. T/C adapter, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT,               0554 0189
 DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO; Radio freq. 869.85 MHz FSK
 T/C probe head for surface measurement (T/C Type K)                                                                                    0602 0394
 Radio handle for plug-in probe heads, incl. T/C adapter, approval for USA, CA, CL; Radio freq. 915.00 MHz FSK                          0554 0191
 T/C probe head for surface measurement (T/C Type K)                                                                                    0602 0394
 Dimensions                                 Measuring         Accuracy                                    Resolution          t99
 Probe shaft/probe shaft tip                range
                    120 mm      40 mm       -50 to +350 °C    Radio handle:                               0.1 °C (-50 to      5s
                                            Short-term to     ±(0.5 °C +0.3% of m.v.) (-40 to +500 °C)    +199.9 °C)
                    Ø 5 mm      Ø 12 mm     +500 °C           ±(0.7 °C +0.5% of m.v.) (remaining range)   1.0 °C (remaining
                                                              T/C probe head: Class 2                     range)
 Radio probes incl. humidity probe head                                                                                                  Part no.
 Radio handle for plug-in probe heads, incl. T/C adapter, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT,               0554 0189
 DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO; Radio freq. 869.85 MHz FSK
 Humidity probe head                                                                                                                    0636 9736
 Radio handle for plug-in probe heads, incl. T/C adapter, approval for USA, CA, CL; Radio freq. 915.00 MHz FSK                          0554 0191
 Humidity probe head                                                                                                                    0636 9736
 Dimensions                                 Measuring         Accuracy                                    Resolution
 Probe shaft/probe shaft tip                range
                                            0 to +100 %RH     ±2 %RH (+2 to +98 %RH)                      0.1 %RH
                                            -20 to +70 °C     ±0.3 °C                                     0.1 °C
 Radio handles for attachable T/C probes                                                                                                 Part no.
 Radio handle for plug-in probe heads, incl. T/C adapter, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT,               0554 0189
 DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO ; Radio freq. 869.85 MHz FSK
 Radio handle for plug-in probe heads, incl. T/C adapter, approval for USA, CA, CL; Radio freq. 915.00 MHz FSK                          0554 0191
 Illustration                               Measuring         Accuracy                                    Resolution
                                            range
                                            -50 to +1000 °C   ±(0.7 °C +0.3% of m.v.) (-40 to +900 °C)    0.1 °C (-50 to +199.9 °C)
                                                              ±(0.9 °C +0.5% of m.v.) (remaining range)   1.0 °C (remaining range)
testo 435
Radio probes
 Radio module for upgrading measuring instrument with radio option                                                        Part no.
 Radio module for measuring instrument, 869.85 MHz, approval for the countries: DE, FR, UK, BE, NL, ES, IT, SE, AT,     0554 0188
 DK, FI, HU, CZ, PL, GR, CH, PT, SI, MT, CY, SK, LU, EE, LT, IE, LV, NO
 Radio module for measuring instrument, 915.00 MHz FSK, approval for USA, CA, CL                                        0554 0190
                                                                                                                                            0981 9614/msp/I/06.2017
 Technical data Radio probes
 Radio immersion/penetration probe, NTC                                     Common Technical Data
 Battery type             2 x 3V button cell (CR 2032)                      Measuring rate           0.5 s or 10 s, adjustable on handle
 Battery life             150 h (meas. rate 0.5 s)
                          2 months (meas. rate 10 s)
                                                                            Radio coverage           Up to 20 m (without obstructions)
 Radio handle                                                               Radio transmission       Unidirectional
 Battery type             2 AAA micro batteries                             Operating temperature    -20 to +50 °C
 Battery life             215 h (meas. rate 0.5 s)                          Storage temperature      -40 to +70 °C
                          6 months (meas. rate 10 s)
                                                                            Protection class         IP54
                                                                                                                                           Subject to change without notice.
                                                                                                                      www.testo.com