Service Manual: VIO 300 D VIO 200 D
Service Manual: VIO 300 D VIO 200 D
All rights to this manual, in particular rights of duplication, dissemination and translation, are reserved. No part of this man-
ual may be reproduced in any form (by photocopying, microfilming or other methods) or processed, duplicated or dissem-
inated by the use of electronic systems without the written consent of Erbe Elektromedizin GmbH.
The information contained in this manual may be amended or supplemented without prior notice and represents no obli-
gation on the part of Erbe Elektromedizin GmbH.
Printed in Germany
Copyright © Erbe Elektromedizin GmbH, Tübingen 2018
                                                                                                                                 Table of Contents
              Table of Contents
                         1   Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
                             Classification of the safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                     7
                             Knowledge of the User Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                    7
                             Protection from the risk of electric shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                      7
                             Electrostatically sensitive components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .                      8
                             Liability and warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .          8
                         2   Modifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
                             From VIO version 2.2.x               .............................................. 9
                             From VIO version 2.3.x               .............................................. 9
                             From VIO version 2.7.x               . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
                             From VIO version 2.8.x               . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
                         3   Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
                             Controls at the front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
                             Controls at the rear. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4 Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
                         5   Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
                             General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .           15
80116-287 
2018-09
                         6   Remedying malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
                             Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
                             ERROR list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
                                    Abbreviations of Error displays . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
                                    A/E-Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
                                    B-Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
                                    C-Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
                                    D-Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
                                    F-Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
                                    2,3,5,6-Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
                                    4 (NE)-Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
                                    9-Errors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
                                                                                                                                                    5 / 84
Table of Contents
                                                                                                                                                      80116-287 
                                                                                                                                                      2018-09
6 / 84
                                                                                        1 • Safety information
              CHAPTER 1
              Safety information
                     ATTENTION!   The ATTENTION! safety indication refers to a risk which can cause
                                  equipment to become unserviceable.
                                  miliarity with the user manuals, in particular the procedures for setting up,
2018-09
                                  For safety reasons, multiple outlets and extension cords should not be
                                  used. If their use is unavoidable, they also must be provided with proper
                                  grounding.
                      WARNING!    Unplug the power cord from the outlet before exchanging parts of the
                                  unit or cleaning it.
WARNING! Do not plug a wet power cord into the unit or into an outlet.
                                                                                                         7 / 84
1 • Safety information
                         WARNING!   Do not touch any unprotected wires or conductive surfaces while the
                                    unit is opened and under voltage.
                         WARNING!   Blown line fuses may only be replaced by a competent technician. Only
                                    replacement fuses of the rating specified on the unit's name plate may
                                    be used. Before resuming operation the unit must be subjected to a
                                    performance test by a competent technician.
                                                                                                                80116-287 
                                                                                                                2018-09
                                    It is recommended that the technical safety check also be performed by
                                    Erbe or persons trained by Erbe.
                    ATTENTION!      Only use original Erbe spare parts. The manufacturer accepts no liability
                                    and the warranty rights becomes void if original spare parts are not
                                    used.
8 / 84
                                                                                                    2 • Modifications
              CHAPTER 2
              Modifications
                                             Hardware
                                             No changes
                                             Software
                   Component affected             Description of the modification
                 Bipolar resection adapter        The bipolar resection adapter can be connected to the multifuc-
                                                  tion socket on the VIO. Supplement to the safety check to test the
                                                  bipolar resection adapter.
                     Setup 2 settings             New setup setting ° modes
                  Generator adjustment            Generator adjustment is possible without APC 2.
80116-287 
2018-09
                                             Hardware
                                             No changes
                                             Software
                   Component affected             Description of the modification
                                                                                                                9 / 84
2 • Modifications
                               Hardware
                               No changes
                               Software
          Component affected        Description of the modification
                               Hardware
                               No changes
Software
                                                                                                   80116-287 
          Component affected        Description of the modification
                                                                                                   2018-09
               VIO module           New mode/new upgrade (only relevant for VIO 300 D):
                                    BiTherm -> enables the use of the HYBRIDtherm application (=
                                    combination of VIO 300 D + ERBECRYO 2 + HYBRIDtherm instru-
                                    ment).
               ERROR list           B error now includes HYBRIDtherm messages.
10 / 84
                                                                                                              3 • Controls
              CHAPTER 3
              Controls
                              IMPORTANT!       This chapter contains an overview of the controls of the unit(s). The rel-
                                               evant User Manual for the unit(s), knowledge of which is assumed for
                                               servicing work, provides detailed information about how to use the
                                               unit(s).
                                                                                             12
              ERBE                                                                                     16
              VIO 300 D
                                                                                                       17
                          2                                            6                     13
                          3                                            7
                                                                                             14
                          4                                            8             11
80116-287 
2018-09
                          5                                            9                     15
              1
                                                                                              F
                                                                               10
                                                                                                       18
Fig. 3-1
                                           1    Power Switch
                                     2–9        Selection buttons
                                       10       Up/Down buttons
                                       11       Enter button
                                   12–15        Focus buttons
                                       16       Pilot lamps for footswitches
                                       17       Pilot lamp for AUTO START
                                       18       Pilot lamps for neutral electrodes
                                                                                                                   11 / 84
3 • Controls
ECB
1 2 3 4
Fig. 3-2
               1    Footswitch sockets
               2    ECB socket (Erbe Communication Bus)
               3    Potential equalization terminal
                                                                        80116-287
                                                                        2018-09
               4    Power supply module with fuses
12 / 84
                                                                                                        4 • Technical Data
              CHAPTER 4
              Technical Data
Power connection
Operating mode
                                                                                                                   13 / 84
4 • Technical Data
Acclimatizing
 If the unit has been stored or transported at temperatures below +10 °C or above +40 °C, the unit will
 require approx. 3 hours to acclimatize at room temperature.
Standards
                                                                                                          80116-287 
                                                                                                          2018-09
14 / 84
                                                                                                                 5 • Setup
              CHAPTER 5
              Setup
                                             General information
                                             This unit has two Setup levels. The first level is accessible to users and ser-
                                             vice staff. The second level is only for use by the service staff.
                                                                                                                    15 / 84
5 • Setup
 AUTO               V 2.1.x        Input of start delay for the AUTO START function.
 START 2                           The start delay value for AUTO START 2 depends on the value entered
                                   for AUTO START 1 but is always above the start delay value of AUTO
                                   START 1. 
                                   A start delay between 0.1 and 10 s is possible.
 Service pro-       V 2.1.x        This menu item leads to the second Setup level.
 gram
                  IMPORTANT!      This setup menu is only available in English depending on the language
                                  setting selected on the unit.
                                                                                                              80116-287 
                                                                                                              2018-09
 Neutral elec-      V 2.1.x        single surface, dual surface, either way, dynamic.
 trode                             On delivery, the unit is set to neutral electrode “dual surface”.
 AUTO START         V 2.1.x        Setting for whether AUTO START is permitted as an activation type.
 AUTO START         V 2.3.x        On: AUTO START function with power limitation of 50 Watt 
 Limit                             (factory setting)
                                   Off: AUTO START function with no power limitation
 Time limit         V 2.1.x        Setting the time period after which activation is automatically ended:
                                   1 to 99 s or OFF
 Display time       V 2.1.x        Setting the length of time for which indicator window and error mes-
                                   sages appear on the display:
                                   1 to 15 s or OFF.
 Automatic          V 2.1.x        Setting the length of time for which an input window appears on the
 time                              display:
                                   3 to 29 s or Not automatic.
 Start screen       V 2.1.x        Selection of start screen: 
                                   Guide or List of Programs.
16 / 84
                                                                                                              5 • Setup
              Expert mode       V 2.1.x       If the expert mode is activated, the following settings are also avail-
                                              able:
                                              – Temperature monitoring for neutral electrodes
                                              – Advanced setting options for ENDO CUT I and Q
                                              – Advanced setting options for BiClamp
                                              – Advanced setting options for PRECISE APC
              Language          V 2.1.x       Self-explanatory.
              APC              V 2.1.x       Self-explanatory.
              supply
              APC Auto-         V 2.1.x       The instrument is purged with gas automatically when it is plugged
              Purge                           into the APC socket and an instrument that is already plugged into the
                                              APC socket is purged with gas automatically when the unit is started
                                              up.
              APC Purge         V 2.1.x       Selection of time for which the instrument is purged with gas auto-
              Duration                        matically:
                                              0 to 10 s.
              APC              V 2.1.x       Selection of purge flow (in %) at which the instrument is purged with
              PurgeFlow                       gas automatically. Purge flow relates to the default COAG-Flow setting
                                              stored in the instrument.
              max. APC          V 2.1.x       Setting of maximum cylinder pressure in the argon cylinder used.
80116-287 
2018-09
              cyl.pressure                    Correct reporting on the cylinder level display on the HF surgical unit
                                              depends upon the maximum cylinder pressure setting of the actual
                                              argon gas bottle used:
                                              100 to 240 bar.
              Sound sam-        V 2.1.x       Selection of type of warning signals.
              ple
              Decoupling C      V 2.1.x       Setting on the decoupling capacitor.
              (C = capaci-
              tor)                            MAX:
                                              Decoupling capacitor with maximum capacity.
                                              MIN:
                                              Decoupling capacitor with minimized capacity.
                                              At this setting, neuromuscular stimuli are reduced in the PULSED APC
                                              mode.
              BiCision SC       V 2.7.x       Extended short circuit detection for BiCision:
              Info                            In case an electrical connection is established between the instrument
                                              branches (e.g. through the advanced instrument blade) during activa-
                                              tion.
                                              On: extended short circuit detection is activated, advisory message is
                                              shown.
                                              Off: extended short circuit detection is deactivated, no advisory mes-
                                              sage is shown.
                                                                                                                 17 / 84
5 • Setup
                                                                                                            80116-287 
                  V 2.1.x       No. APC errors:
                                                                                                            2018-09
                                Records the frequency of errors detected and signaled by the APC.
                  V 2.1.x       Operating time: 
                                Runtime meter or ON counter for add-on modules, if these modules
                                have a corresponding counter.
                  V 2.1.x       Loudsp. test:
                                Unit checks the loudspeaker function. Three different tones must be
                                heard.
                  V 2.1.x       CAN looped mem.: 
                                Displays the CAN messages before the last error occurred. The test
                                program was integrated for product development for debugging.
                  V 2.1.x       ScreenTest:
                                Tests the function of the screen. A total of 12 rectangular color seg-
                                ments have to be displayed, each of the four in the color ranges red,
                                green and blue.
                  V 2.1.x       Error list IIF/NE:
                                Stores all errors detected and signaled by the IIF (instrument interface)
                                and the NE (Nessy2).
                  V 2.1.x       Hardware TP:
                                Branching to the hardware test programs.
18 / 84
                                                                                                                                                           5 • Setup
                                                                                                                                                               19 / 84
5 • Setup
                                       Call up Setup
                       IMPORTANT!      Within a menu there are different ways to scroll forwards: 
                                       (a) with the Down button
                                       or
                                       (b) with the selection button next to the "next" menu item. 
                                       Only variant (a) is used in this manual.
VIO 300 D
1 5
2 6
               3                                               7
                                                                      9
                                                               8           12
               4
                                                                      10             F
Fig. 5-1
                                                                                                                  80116-287 
                                       2. Select menu item "Other functions".
                                                                                                                  2018-09
                                       3. Select menu item "Setup". The unit switches to Setup level 1. See
                                          above table for settings that can be changed here.
20 / 84
                                                                            5 • Setup
              Change settings
              1. Select the setting to be changed using the adjacent selection
                 button (1...8). The setting is highlighted.
              2. Change the setting with the Up/Down buttons (9/10).
              3. Confirm the changed setting with the Enter button (12).
80116-287 
2018-09
                                                                                 21 / 84
22 / 84
                        5 • Setup
          80116-287 
          2018-09
                                                                               6 • Remedying malfunctions
              CHAPTER 6
              Remedying malfunctions
                                  Safety information
                     ATTENTION!   Adjustments, technical tests, modifications, maintenance and repair
                                  work may only be performed by Erbe or persons trained by Erbe. If the
                                  work is not performed by trained persons, Erbe accepts no liability and
                                  warranty rights become void.
ERROR list
                                                                                                    23 / 84
6 • Remedying malfunctions
                                  A/E-Errors
 Recognizing module: A = APC 2, E = Extension module VEM 2
                                                                                                              80116-287 
                                                                                                              2018-09
 A/E        B                               Error in test mode.              Information in test mode
                                                                             (adjustment).
 A/E        C+D                                                              Restart the unit. If the error
                                                                             occurs again, notify Erbe Ser-
                                                                             vice.
 A          10                              Underpressure at selected        Check the gas supply (tank,
                                            gas input.                       pressure regulator). If the
                                                                             error persists, notify Erbe
                                                                             Service.
 A          11                              Overpressure at selected gas     Check the gas supply (tank,
                                            input.                           pressure regulator). If the
                                                                             error persists, notify Erbe
                                                                             Service.
 A          12                              Caloric and differential pres-   Check the gas supply (type of
                                            sure sensor do not agree         gas). If the error persists,
                                            (wrong gas).                     notify Erbe Service.
 A          13                                                               Restart the unit. If the error
                                                                             occurs again, notify Erbe Ser-
                                                                             vice.
24 / 84
                                                                                            6 • Remedying malfunctions
                                                                                                                   25 / 84
6 • Remedying malfunctions
                                                                                                          80116-287 
                                                                         error persists, notify Erbe
                                                                                                          2018-09
                                                                         Service.
 A          91                             Low pressure at cylinder 2.   Check the gas supply (tank,
                                                                         pressure regulator). If the
                                                                         error persists, notify Erbe
                                                                         Service.
 A          A0                                                           Restart the unit. If the error
                                                                         occurs again, notify Erbe Ser-
                                                                         vice.
26 / 84
                                                                                           6 • Remedying malfunctions
                                              B-Errors
              Recognizing module: B = Control panel
                                                                                                                27 / 84
6 • Remedying malfunctions
                                                                                                                 80116-287 
                                               the event or error list.
                                                                                                                 2018-09
 B          14 + 15                                                             Restart the unit. If the error
                                                                                occurs again, notify Erbe Ser-
                                                                                vice.
 B          16                                 Insufficient EEPROM mem-         Delete unnecessary pro-
                                               ory:                            grams.
                                               Occurs if the memory for
                                               user programs is full.
 B          17         CAN ID of second        Double activation:              Check the accessories. If the
                       activation signal.      Occurs if two activation sig-    error persists, notify Erbe
                                               nals (e.g. both pedals of a      Service.
                                               footswitch) are present
                                               simultaneously (within
                                               100 ms).
28 / 84
                                                                                             6 • Remedying malfunctions
                                                                                                                  29 / 84
6 • Remedying malfunctions
                                                                                                                 80116-287 
                                             by the Nessy symmetry
                                                                                                                 2018-09
                                             monitoring during activa-
                                             tion.
 B          21                                                                Restart the unit. If the error
                                                                              occurs again, notify Erbe Ser-
                                                                              vice.
 B          22                               Occurs if activation was ter-    Check the accessories. If the
                                             minated automatically (e.g.      error persists, notify Erbe
                                             by AutoStop) and tissue is       Service.
                                             still touched.
 B          23                                                                Restart the unit. If the error
                                                                              occurs again, notify Erbe Ser-
                                                                              vice.
 B          24                               Occurs if an incorrect status    Remove IES 2 from the unit
                                             message is received from         and notify Erbe Service.
                                             the IES 2.
 B          25                               Occurs if a modified socket      Check and save the socket
                                             configuration was found          configuration in the "Version
                                             (e.g. by replacement or ret-     List" test program. If the error
                                             rofitting of a socket).          occurs again, notify Erbe Ser-
                                                                              vice.
30 / 84
                                                                                           6 • Remedying malfunctions
                                                                                                                   31 / 84
6 • Remedying malfunctions
                                                                                                           80116-287 
                                                                                                           2018-09
 B          89                             Single-pedal footswitch has    Information, no fault condi-
                                           been disconnected from the     tion.
                                           system: 
                                           Occurs if a single-pedal
                                           footswitch is disconnected
                                           from the system.
 B          8A                             APC socket 2 recognized:      Information, no fault condi-
                                           Occurs if an APC socket 2 is   tion.
                                           detected.
 B          8B                                                            Restart the unit. If the error
                                                                          occurs again, notify Erbe Ser-
                                                                          vice.
 B          8C                             APC module recognized:         Information, no fault condi-
                                           Occurs if an APC module is     tion.
                                           detected.
 B          8D                             No status message from         Check connectors. Check APC.
                                           APC module: 
                                           Occurs if an APC module is
                                           disconnected from the sys-
                                           tem.
32 / 84
                                                                                          6 • Remedying malfunctions
                                                        system: 
                                                        Occurs if a multifunction
                                                        footswitch is disconnected
                                                        from the system.
              B          95        Number of output     New instrument recognized      Information, no fault condi-
                                   socket *0X10000 +    by system:                    tion.
                                   instrument number.   Occurs if an instrument with
                                                        instrument recognition is
                                                        connected.
              B          97 + 98                                                       Restart the unit. If the error
                                                                                       occurs again, notify Erbe Ser-
                                                                                       vice.
                                                                                                               33 / 84
6 • Remedying malfunctions
 B          99         Activation type:      This activation type is not      Assign the correct activation
                                             available:                      type to the instrument.
                       1 => Dual-pedal        Occurs if the user assigns an
                       footswitch both        activation signal that is not
                       pedals                currently available. (e.g.
                       2 => Dual-pedal        unconnected footswitch).
                       footswitch only
                       Coag
                       3 => Dual-pedal
                       footswitch only Cut
                       4 => Single-pedal
                       footswitch
                       5 => AutoStart 1
                       6 => AutoStart 2
 B          9A                                Please check the clock:          Set time.
                                              Occurs if the VIO was
                                              switched off for such a long
                                              time that the supply current
                                              to the built-in realtime clock
                                              was no longer sufficient.
 B          9B                                Master remote control rec-       Information, no fault condi-
                                                                                                               80116-287 
                                              ognized: 
                                                                                                               2018-09
                                                                               tion.
                                              Occurs if a master remote
                                              control is recognized by the
                                              VIO.
 B          9C                                Master remote control has        Information, no fault condi-
                                              been disconnected from the       tion.
                                              system: 
                                              Occurs if the master remote
                                              control is disconnected from
                                              the system.
 B          9D                                Remote control recognized:       Information, no fault condi-
                                              Occurs if a remote control is    tion.
                                              recognized by the VIO.
 B          9E                                Remote control has been          Information, no fault condi-
                                              disconnected from the sys-       tion.
                                              tem: 
                                              Occurs if the remote control
                                              is disconnected from the
                                              system.
34 / 84
                                                                                             6 • Remedying malfunctions
                                   gle-pedal
                                   footswitch)
                                   + 0x200 (with dual-
                                   pedal footswitch)
                                   + 0x400 (with mul-
                                   tifunction
                                   footswitch)
              B          A4        CAN ID of second        Two footswitches are con-      Remove the duplicate
                                   footswitch or 2 if it   nected: Occurs if two          footswitch from the unit.
                                   cannot be assigned      footswitches of the same
                                   to any CAN ID.          type are connected.
              B          A6                                EEPROM is updated: Occurs      Information, no fault condi-
                                                           if the unit is programmed      tion.
                                                           using a connected PC.
              B          A8                                                               Restart the unit. If the error
                                                                                          occurs again, notify Erbe Ser-
                                                                                          vice.
                                                                                                                  35 / 84
6 • Remedying malfunctions
                                                                                                                80116-287 
                                              tion is assigned while con-
                                                                                                                2018-09
                                              tact has already been
                                              recognized.
 B          AD – AF                                                            Restart the unit. If the error
                                                                               occurs again, notify Erbe Ser-
                                                                               vice.
 B          B0         NESSY symmetry         NESSY symmetry warning:         Check the NE accessories.
                       value in %.            Occurs if the Nessy symme-
                                              try monitoring signals a
                                              value between 20 % and
                                              50 %.
 B          B1         NESSY current          NESSY current density warn-      Check the NE accessories.
                       value.                 ing: 
                                              Occurs if the Nessy current
                                              density monitoring signals a
                                              value above the limit curve.
 B          B2                                IES footswitch detected:         Information, no fault condi-
                                              Occurs if an IES footswitch      tion.
                                              is connected.
 B          B3                                                                 Restart the unit. If the error
                                                                               occurs again, notify Erbe Ser-
                                                                               vice.
36 / 84
                                                                                          6 • Remedying malfunctions
                                                                                                                 37 / 84
6 • Remedying malfunctions
                                                                                                          80116-287 
                                           footswitch nor AUTO START
                                                                                                          2018-09
                                           is assigned.
 B          C1                             EIP 2 was recognized by the     Information, no fault condi-
                                           system:                        tion.
                                           Occurs if the EIP 2 is con-
                                           nected to the system.
 B          C2                             EIP 2 was disconnected from     Information, no fault condi-
                                           the system:                    tion.
                                           Occurs if EIP 2 is discon-
                                           nected from the system.
 B          C3                             IES 2 footswitch was discon-    Information, no fault condi-
                                           nected from the system:        tion.
                                           Occurs if the IES 2
                                           footswitch is disconnected
                                           from the system.
 B          C4                             Purge function not assigned:  Information, no fault condi-
                                           Occurs if the purge button     tion.
                                           on the APC is pressed and
                                           no APC socket was selected.
38 / 84
                                                                                           6 • Remedying malfunctions
                                                                                                                39 / 84
6 • Remedying malfunctions
                                                                                                             80116-287 
                                           an input pressure that is too    solar radiation/other heat
                                                                                                             2018-09
                                           high.                            sources.
 B          D9                             The input pressure sensor        Fully open the bottle valve.
                                           on the ERBRCRYO 2 is sig-        Allow the gas bottle to warm
                                           naling (during activation) a     up to room temperature. Con-
                                           low input pressure.              nect a full gas bottle.
 B          DA                             The input pressure sensor        Fully open the bottle valve.
                                           on the ERBRCRYO 2 is sig-        Allow the gas bottle to warm
                                           naling (during activation) an    up to room temperature. Con-
                                           input pressure that is too       nect a full gas bottle.
                                           low.
 B          DB                             The input pressure sensor        Fully open the bottle valve.
                                           on the ERBRCRYO 2 is sig-        Allow the gas bottle to warm
                                           naling (outside of activation)   up to room temperature. Con-
                                           an input pressure that is too    nect a full gas bottle.
                                           low.
 B          DC                             The input pressure sensor        Fully open the bottle valve.
                                           on the ERBRCRYO 2 is sig-        Allow the gas bottle to warm
                                           naling (during activation) an    up to room temperature. Con-
                                           input pressure that is too       nect a full gas bottle.
                                           low.
40 / 84
                                                                                           6 • Remedying malfunctions
                                                                                                                41 / 84
6 • Remedying malfunctions
                                  C-Errors
 Recognizing module: C = CPU + Sensors
                                                                                            80116-287 
                                                           vice.
                                                                                            2018-09
 C          41 – 48                                        Restart the unit. If the error
                                                           occurs again, notify Erbe Ser-
                                                           vice.
 C          51 – 58                                        Restart the unit. If the error
                                                           occurs again, notify Erbe Ser-
                                                           vice.
 C          5A + 5B                                        Restart the unit. If the error
                                                           occurs again, notify Erbe Ser-
                                                           vice.
 C          61 – 66                                        Restart the unit. If the error
                                                           occurs again, notify Erbe Ser-
                                                           vice.
 C          70 – 72                                        Restart the unit. If the error
                                                           occurs again, notify Erbe Ser-
                                                           vice.
 C          75                                             Restart the unit. If the error
                                                           occurs again, notify Erbe Ser-
                                                           vice.
42 / 84
                                                                                          6 • Remedying malfunctions
                                                                                                                43 / 84
6 • Remedying malfunctions
                                 D-Errors
 Recognizing module: D = Smoke evacuation system IES 2
                                                                                                              80116-287 
 D           83                                                              Restart the unit. If the error
                                                                                                              2018-09
                                                                             occurs again, notify Erbe Ser-
                                                                             vice.
 D           86                                                              Restart the unit. If the error
                                                                             occurs again, notify Erbe Ser-
                                                                             vice.
 D           87                             Obstruction.                     Remove the suction tube
                                                                             from the tissue.
 D           88                             The filter is missing or         Insert or replace the filter.
                                            obstructed.
44 / 84
                                                                                           6 • Remedying malfunctions
                                             F-Errors
              Recognizing module: F = Footswitch
                                                                                                               45 / 84
6 • Remedying malfunctions
                                  2,3,5,6-Errors
 Recognizing module: 2,3,5,6 = IIF (Instrument Interface) of corresponding socket slot
                                                                                                                80116-287 
                                                                               vice.
                                                                                                                2018-09
 2,3,5,6     F                                U Coag error:                   Check the accessories. If the
 (socket)                                     +12 V voltage during Coag        error persists, notify Erbe
                                              activation too low.              Service.
 2,3,5,6     70 – 72                                                           Restart the unit. If the error
 (socket)                                                                      occurs again, notify Erbe Ser-
                                                                               vice.
 2,3,5,6     81                               Short circuit instrument rec-    Check the accessories. If the
 (socket)                                     ognition R:                     error persists, notify Erbe
                                              The resistor value deter-        Service.
                                              mined for the instrument
                                              recognition corresponds to a
                                              short circuit.
 2,3,5,6     82                               Short circuit activation rec-    Check the accessories. If the
 (socket)                                     ognition R_NEU:                 error persists, notify Erbe
                                              The resistor value deter-        Service.
                                              mined for the activation rec-
                                              ognition corresponds to a
                                              short circuit.
46 / 84
                                                                                              6 • Remedying malfunctions
                                                                                                                   47 / 84
6 • Remedying malfunctions
                                 4 (NE)-Errors
 Recognizing module: 4 (NE) = Nessy2
                                                                                            80116-287 
                                                                                            2018-09
48 / 84
                                                                                            6 • Remedying malfunctions
                                               9-Errors
              Recognizing module: 9 = Erbe Irrigation Pump EIP 2
                                                                                         vice.
              9          90                                                              Restart the unit. If the error
                                                                                         occurs again, notify Erbe Ser-
                                                                                         vice.
              9          A0                                                              Restart the unit. If the error
                                                                                         occurs again, notify Erbe Ser-
                                                                                         vice.
              9          C0                                                              Restart the unit. If the error
                                                                                         occurs again, notify Erbe Ser-
                                                                                         vice.
                                                                                                                  49 / 84
6 • Remedying malfunctions
                             80116-287 
                             2018-09
50 / 84
                                                                              7 • Maintenance and servicing
              CHAPTER 7
              Maintenance and servicing
                                                                                                      51 / 84
7 • Maintenance and servicing
                                 Safety information
                      WARNING!   For safety reasons (personnel protection) the test specimen should
                                 generally be operated by a suitable isolating transformer. 
                                 An exceptional case is the tests for grounded conductor resistance,
                                 ground leakage current, and patient leakage current, in which the test
                                 specimen is supplied with current via the safety tester.
                    ATTENTION!   In the event of a fault occurring in the test specimen or individual com-
                                 ponents during the technical safety check the test steps taken so far no
                                 longer apply. 
                                 Remedy the defect and repeat the technical safety check from the
                                 beginning.
                    ATTENTION!   The specifications made in this chapter for test programs / test pro-
                                 gram settings allow the service technician to perform the technical
                                 safety check. In addition, the test program mode provides a number of
                                 diagnostic and inspection options for the service technician. They are
                                 described in the repair manual. Erbe Elektromedizin expressly states
                                 that the full functional range of the test program mode may not be
                                 used without precise knowledge of the individual test programs.
                                                                                                                80116-287 
                                                                                                                2018-09
                    IMPORTANT!   It is assumed that the user knows how to operate the test specimen,
                                 the test equipment, the measuring equipment, and auxiliary test equip-
                                 ment. The test instructions only apply in conjunction with the relevant
                                 test steps.
                    IMPORTANT!   The test report for the technical safety check can be requested from
                                 Erbe Technical Service Tübingen. For the address see address sheet on
                                 last page.
52 / 84
                                                                                           7 • Maintenance and servicing
                                                                                                                 53 / 84
7 • Maintenance and servicing
                                                                                                                  80116-287 
  Ground leakage current mea-       • Ground leakage current, normal condition (N.C.).
                                                                                                                  2018-09
                     surement       • Ground leakage current, single-fault condition (S.F.C.).
   Patient leakage current mea-     • Patient leakage current, normal condition (N.C.).
                       surement     • Patient leakage current, single-fault condition (S.F.C.).
                                    IMPORTANT! Close the output relay of the unit for testing the patient leak-
                                    age current. Procedure:
                                    1. Call up Setup level 2 (see page 20).
                                    2. Select the "Test programs" setting.
                                    3. Use the Up button to select "Hardware TP" and press Enter to con-
                                       firm. The test specimen switches to the test program mode.
                                    4. Use the Up/Down buttons to select the "TP relay" test program and
                                       press Enter to call it up.
54 / 84
                                                             7 • Maintenance and servicing
5. Close all relays in the test program via „All switch on“:
DC resistance
                                                                                   55 / 84
7 • Maintenance and servicing
Test setup
                      WARNING!   Across the measuring lines there is the DC voltage of 500 V! In order to
                                 avoid injuries, only switch on the test specimen and safety tester when
                                 all the electrical connections have been made.
Fig. 7-1
                                 • The test specimen is connected to the power supply via the power
                                   cord.
                                 • The NE socket of the test specimen is connected to the safety tester
                                   via the patient cable NE with the adapter cord.
                                 • The AE socket of the test specimen is connected to the safety tester
                                   via the patient cable AE and the electrode handle with the laboratory
                                   measuring cable.
                                                                                                            80116-287 
                                                                                                            2018-09
                                 Test procedure
                                 1. Start safety tester in the "Insulation resistance" function The mea-
                                    sured value displayed should be >200 MOhm.
                                 2. Start test specimen and select the test progam "TP relay".
                                 3. In the "TP relay" test program, connect all output relays on the test
                                    specimen with "All switch on". The measurement value displayed on
                                    the safety tester has to drop considerably from >200 MOhm.
                                 4. Determine insulation resistance using the safety tester. The measured
                                    value must be >2 MOhm.
                                 5. Document the measured value.
                                 6. Switch the test specimen off.
56 / 84
                                                                                          7 • Maintenance and servicing
Performance tests
                                              Test setup
                                              • The test specimen is connected to the power supply via the power
                                                cord.
                                              • The dual-pedal footswitch with ReMode is connected.
                                              • The NE patient cable with the shorted adapter cable is connected to
                                                the NE socket on the test specimen.
Test procedure
                             Power switch     1. Check power switch for smooth operation. The power switch must be
                                                 easy to operate and must neither stick nor scrape.
80116-287 
2018-09
                                              2. Press power switch. The power switch must snap into the "ON" posi-
                                                 tion and the test specimen must perform a system start.
                          Start routine /     1. The test specimen must perform the system start without error mes-
                   acknowledgement tone          sage(s).
                                              2. The test specimen must emit an acoustic signal (acknowledgement
                                                 tone) during the self-test.
                        Control buttons /     1. Check all selection buttons on the control panel for functionality.
              acknowledgement tone when          Press each button at least twice. An acoustic signal (=acknowledge-
                                 pressed         ment tone) must be emitted each time the buttons are pressed.
                          Pushbuttons /       1. Check all the pushbuttons (focus buttons, up and down buttons, enter
              acknowledgement tone when          button) on the test specimen to make sure they are operating prop-
                                 pressed         erly. Press each button at least twice. There must be an acoustic sig-
                                                 nal (acknowledgement tone) every time a button is pressed.
                  Neutral electrode setting   1. In Setup level 2 obtain and document the current setting of the neu-
                                                 tral electrode.
Date and time 1. In Setup level 2 check date and time and correct if necessary.
                                                                                                                  57 / 84
7 • Maintenance and servicing
                                   LEDs:
                                   1. When starting the system, all LEDs must briefly light up at the same
                                      intensity.
                Activation tones   1. Press CUT pedal on the dual-pedal footswitch at least twice. When
                                      pressing the pedal there must be an acoustic signal every time 
                                      (=acknowledgement tone).
                                   2. Press COAG pedal on the dual-pedal footswitch at least twice. When
                                      pressing the pedal there must be an acoustic signal every time 
                                      (=acknowledgement tone).
Footswitch activation
                                                                                                             80116-287 
                                                                                                             2018-09
                                   Test setup
                                   • The test specimen is connected to the power supply via the power
                                     cord.
                                   • The single-pedal footswitch is connected.
                                   • The dual-pedal footswitch with ReMode is connected.
58 / 84
                                                                                      7 • Maintenance and servicing
                                        Test procedure
                                        1. Call up Setup level 2 (see page 20).
                                        2. Select the "Test programs" setting.
                                        3. Use the Up button to select "Hardware TP" and press Enter to con-
                                           firm. The test specimen switches to the test program mode.
                                        4. Use the Up/Down buttons to select the "TP activation signal" test
                                           program and press Enter to call it up.
              Single-pedal footswitch   1. Press COAG pedal on the single-pedal footswitch. The test program
                      activation COAG      must confirm activation by lighting up the "Single-pedal" icon in
                                           color.
80116-287 
2018-09
               Dual-pedal footswitch    1. Press CUT pedal on the dual-pedal footswitch. The test program must
               activation CUT / COAG       confirm activation lighting up the relevant field in color (yellow) in the
                                           "MF" icon.
                                        2. Press COAG pedal on the dual-pedal footswitch. The test program
                                           must confirm activation lighting up the relevant field in color (blue) in
                                           the "MF" icon.
               Dual-pedal footswitch    1. Press ReMode button on the dual-pedal footswitch. The test program
                            ReMode         must confirm switchover by lighting up the relevant field in color in
                                           the "MF" icon.
                                        2. Remain in the Test program.
                                                                                                              59 / 84
7 • Maintenance and servicing
Fingerswitch activation
                                     Test setup
                                     • The test specimen is connected to the power supply via the power
                                       cord.
Test procedure
20190-045 1. Connect the electrode handle with patient cable AE to the Monopolar
                                                                                                                  80116-287 
          Activation CUT / COAG         socket of the test specimen.
                                                                                                                  2018-09
                                     2. Press CUT button on the electrode handle. The test program must
                                        confirm activation by lighting up the relevant field in color in the
                                        "Handpiece" icon.
                                     3. Press COAG button on the electrode handle. The test program must
                                        confirm activation by lighting up the relevant field in color in the
                                        "Handpiece" icon.
                                     4. Remove the electrode handle.
                      20190-115      1. For an MF socket: Connect the VIO ReMode electrode handle to the
          Activation CUT / COAG         MF socket of the test specimen.
              (only if there is an   2. For an MF-2 socket: Plug the bipolar adapter (art. no. 20183-101)
            multifuction socket)        into the MF-2 socket. Connect the VIO ReMode electrode handle to
                                        the bipolar adapter.
                                     3. Press CUT button on the VIO ReMode electrode handle. The test pro-
                                        gram must confirm activation by lighting up the relevant field in color
                                        in the "Handpiece" icon.
                                     4. Press COAG button on the VIO ReMode electrode handle. The test pro-
                                        gram must confirm activation by lighting up the relevant field in color
                                        in the "Handpiece" icon.
60 / 84
                                                                                        7 • Maintenance and servicing
                            20190-115       1. Press ReMode button on the electrode handle. The test program must
                                ReMode         confirm switchover by lighting up the relevant field in color in the
                      (only if there is a      "Handpiece" icon.
                    multifuction socket)    2. Remove the VIO ReMode electrode handle.
                                            3. Remove the bipolar adapter if present.
                Exit Test program mode      1. Press the Up button until "Restart" appears.
                                            2. Press Enter to confirm the setting. The test specimen exits the test
                                               program mode.
                           via EEPROM
2018-09
                                            • The test specimen is connected to the power supply via the power
                                              cord.
                                            Test procedure
                                            1. For an MF socket: Connect the measuring cable (art. no. 20100-152)
                                               to the MF socket of the test specimen. The test specimen must dis-
                                               play instrument detection with an advisory message. The setting for
                                               COAG mode must switch to BiClamp.
                                            2. For an MF-2 socket: Connect the measuring cable (art. no. 20100-
                                               189) to the MF-2 socket of the test specimen. The test specimen
                                               must display instrument detection with an advisory message. The
                                               setting for COAG mode must switch to BiClamp.
                                                                                                               61 / 84
7 • Maintenance and servicing
Test setup
                                                                                                   T1   T2
                                                                                        BP1   1                  2   BP2
X X
X X
Testbox 20100-102
                                                                                                                           80116-287 
                                                                                                                           2018-09
                                    Fig. 7-2
                                    • The test specimen is connected to the power supply via the power
                                      cord.
                                    • The Bipolar socket of the test specimen is connected to the VIO Test-
                                      box via the bipolar cable with the adapter cable.
Test procedure
                    1st test step   1. In the Setup of the test specimen establish the set start delay for the
                                       setting AUTO START 1. Record the value.
                                    2. Set test specimen to:
                                       BIPOLAR SOFT, Effect 1, 50 watts
                                       AUTO START 1
                                    3. On the VIO Testbox press button T1. The test specimen must start
                                       activation after the set start delay.
                                    4. Press button T2. The test specimen must terminate activation.
62 / 84
                                                                         7 • Maintenance and servicing
                              Test setup
80116-287 
2018-09
1 X
                                                                                            T3
                                                                                 BP3   3           4   BP4
X X
Testbox 20100-102
Fig. 7-3
                              • The test specimen is connected to the power supply via the power
                                cord.
                              • The Bipolar socket of the test specimen is connected to the VIO Test-
                                box via the bipolar cable with the adapter cable.
                              • The single-pedal footswitch is connected.
                                                                                                        63 / 84
7 • Maintenance and servicing
                                Test procedure
                                1. Set test specimen to:
                                   BIPOLAR SOFT with AutoStop, Effect 4, 50 watts
                                2. On the VIO Testbox hold down button T3.
                                3. Activate BIPOLAR SOFT with the footswitch, keeping the pedal
                                   pressed.
                                4. Hold down button T3 for approx. 5 s, then let go. Within another 9 s
                                   at the latest the test specimen must terminate activation and emit
                                   two brief signal tones.
Spark monitor
                                                                                                           80116-287 
                                                           or
                                                                                                           2018-09
                                       20194-075           Patient cable NE, international
                                Test setup
                                The test setup depends on the respective socket equipment of the test
                                specimen and can therefore vary. This test setup arises from a socket
                                equipment bipolar socket, monopolar socket, monopolar socket, NE socket.
                                Measurements are taken on the monopolar socket in the second socket
                                slot.
64 / 84
                                                         7 • Maintenance and servicing
                                                                Power    S1 DC 70 V
                                                                              +    -
                                                         AE
NE
Fig. 7-4
              • The test specimen is connected to the power supply via the power
                cord.
              • The test specimen is connected to the testbox.
              Test procedure
              1. Call up Setup level 2 (see page 20).
              2. Select the "Test programs" setting.
              3. Use the Up button to select "Hardware TP" and press Enter to con-
                 firm. The test specimen switches to the test program mode.
              4. Use the Up/Down buttons to select the "TP generator" test program
                 and press Enter to call it up.
80116-287 
                                                                                  65 / 84
7 • Maintenance and servicing
                                                                                                         80116-287 
                                                            socket is available)
                                                                                                         2018-09
                                            20100-190       BIPOLAR CUT+ / BIPOLAR CUT++ MF-2
                                                            measuring cable (only if an MF-2 socket is
                                                            available)
                                                 –          HF power meter
                                            20189-303       Dual-pedal footswitch with ReMode
AE
NE
HF power meter
Fig. 7-5
66 / 84
                                                                                        7 • Maintenance and servicing
                          • The test specimen is connected to the power supply via the power
                            cord.
                          • The levels of power are determined with the HF power meter. The
                            measuring cables are plugged into the HF power meter direct.
                          Test procedure
                          1. Set test specimen to:
                             AUTO CUT, Effect 8, 300 watts
                          2. Set HF power meter to:
                             RL = 500 ohms
                          3. Activate test specimen via CUT button on the electrode handle.
                          4. Determine and document measured value. The tolerance range is 240
                             to 360 watts.
Bipolar
HF power meter
Fig. 7-6
                          • The test specimen is connected to the power supply via the power
                            cord.
                          • The levels of power are determined with the HF power meter. The
                            measuring cables are plugged into the HF power meter direct.
                          • The dual-pedal footswitch with ReMode is connected.
                          1. The DRY CUT is only standard scope of supply with VIO 300 D electrosurgical units. With VIO 200 D
                             electrosurgical units it can be purchased and installed as an upgrade. For details of how to test DRY
                             CUT with VIO 200 D electrosurgical units see chapter "Performance test upgrades".
                                                                                                                        67 / 84
7 • Maintenance and servicing
                                      Test procedure
                                      1. Set test specimen to:
                                         BIPOLAR CUT, Effect 8, 100 watts
                                      2. Set HF power meter to:
                                         RL = 500 ohms
                                      3. Activate test specimen via CUT pedal on the footswitch.
                                      4. Determine and document measured value. The tolerance range is 80
                                         to 120 watts.
MF
HF power meter
Fig. 7-7
                                      • The test specimen is connected to the power supply via the power
                                        cord.
                                                                                                             80116-287 
                                                                                                             2018-09
                                      • The levels of power are determined with the HF power meter. The
                                        measuring cables are plugged into the HF power meter direct.
                                      • The dual-pedal footswitch with ReMode is connected.
                                      Test procedure
                                      1. Set test specimen to:
                                         BIPOLAR CUT+, Effect 8
                                      2. Set HF power meter to:
                                         RL = 500 ohms
                                      3. Activate test specimen via CUT pedal on the footswitch.
                                      4. Determine and document measured value. The tolerance range is 296
                                         to 400 watts.
68 / 84
                                                                                  7 • Maintenance and servicing
Bipolar
HF power meter
Fig. 7-8
                                    • The test specimen is connected to the power supply via the power
                                      cord.
                                    • The levels of power are determined with the HF power meter. The
                                      measuring cables are plugged into the HF power meter direct.
                                    • The dual-pedal footswitch with ReMode is connected.
                                                                                                           69 / 84
7 • Maintenance and servicing
                                Test procedure
                                1. Set test specimen to:
                                   BIPOLAR FORCED COAG, Effect 2, 90 watts
                                2. Set HF power meter to:
                                   RL = 200 ohms
                                3. Activate test specimen via COAG pedal on the footswitch.
                                4. Determine and document measured value. The tolerance range is 72
                                   to 108 watts.
AE
NE
HF power meter
Fig. 7-9
• The test specimen is connected to the power supply via the power
                                                                                                       80116-287 
                                  cord.
                                                                                                       2018-09
                                • The levels of power are determined with the HF power meter. The
                                  measuring cables are plugged into the HF power meter direct.
                                Test procedure
                                1. Set test specimen to:
                                   SOFT COAG, Effect 8, 200 watts
                                2. Set HF power meter to:
                                   RL = 50 ohms
                                3. Activate test specimen via COAG button on the electrode handle.
                                4. Determine and document measured value. The tolerance range is 160
                                   to 240 watts.
                                1. Set test specimen to:
                                   FORCED COAG, Effect 4, 120 watts
                                2. Set HF power meter to:
                                   RL = 500 ohms
                                3. Activate test specimen via COAG button on the electrode handle.
                                4. Determine and document measured value. The tolerance range is 96
                                   to 144 watts.
70 / 84
                                                           7 • Maintenance and servicing
                                                                                 71 / 84
7 • Maintenance and servicing
AE
NE
HF power meter
Fig. 7-10
                                      • The test specimen is connected to the power supply via the power
                                        cord.
                                      • The levels of power are determined with the HF power meter. The
                                        measuring cables are plugged into the HF power meter direct.
                                      Test procedure
                                      1. Set test specimen to:
                                         DRY CUT, Effect 8, 200 watts
                                      2. Set HF power meter to:
                                         RL = 500 ohms
                                                                                                                  80116-287 
                                                                                                                  2018-09
                                      3. Activate test specimen via CUT button on the electrode handle.
                                      4. Determine and document measured value. The tolerance range is 160
                                         to 240 watts.
                                      Test procedure
                                      1. Activate the BiClamp function with the footswitch. After approx.
                                         8 seconds the test specimen must interrupt activation and emit a
                                         warning message "C-84".
72 / 84
                                                                             7 • Maintenance and servicing
IMPORTANT! For these tests in Setup level 2 set the Expert mode to "ON".
Oscilloscope
                                                                                High Voltage
                                                                                Differential probe
AE
NE
HF power meter
                                 Fig. 7-11
80116-287 
2018-09
                                 Test procedure
                                 1. Set test specimen to:
                                    EndoCut I, Effect 1, Cutting duration 1, Cutting interval 1
                                 2. Set HF power meter to:
                                    RL = 1000 ohms
                                 3. Connect the probe of an oscilloscope to AE and NE.
                                 4. Set oscilloscope to:
                                    200 V / Div, 20 ms
                                 5. Activate test specimen via the CUT button on the electrode handle.
                                 6. Determine the duration of the cutting pulse. The tolerance range is 90
                                    to 110 ms.
                                                                                                        73 / 84
7 • Maintenance and servicing
Bipolar
HF power meter
Fig. 7-12
                                                                                                          80116-287 
                                                                                                          2018-09
                                 • The test specimen is connected to the power supply via the power
                                   cord.
                                 • The levels of power are determined with the HF power meter. The
                                   measuring cables are plugged into the HF power meter direct.
                                 • The dual-pedal footswitch with ReMode is connected.
                                 Test procedure
                                 1. Set test specimen to:
                                    BI PRECISE COAG, Effect 8, 50 watts
                                 2. Set HF power meter to:
                                    RL = 75 ohms
                                 3. Activate test specimen via COAG pedal on the footswitch.
                                 4. Determine and document measured value. The tolerance range is 40
                                    to 60 watts.
74 / 84
                                                                                            7 • Maintenance and servicing
Monitor circuits
Testbox 20100-101
1 2 3 4 5
6 7 8
9 10 11 12
Fig. 7-13
                                                 • The test specimen is connected to the power supply via the power
                                                   cord.
                                                 • The test is performed without a load.
                                                 • The NE socket of the test specimen is connected to the VIO Testbox via
                                                   the patient cable NE with the adapter cable.
                                                 • The dual-pedal footswitch with ReMode is connected.
                                                 1. On the test specimen the (single surfaced) neutral electrode lamp
                                                    must light up red.
                                                 2. Activate AUTO CUT via the footswitch. The test specimen must inhibit
                                                    activation and emit or display an optical and acoustic warning.
                                                                                                                         75 / 84
7 • Maintenance and servicing
Testbox 20100-101
1 2 3 4 5
6 7 8
9 10 11 12
Fig. 7-14
                                     • The test specimen is connected to the power supply via the power
                                       cord.
                                     • The test is performed without a load.
                                     • The NE socket of the test specimen is connected to the VIO Testbox via
                                       the patient cable NE with the adapter cable.
                                     • The dual-pedal footswitch with ReMode is connected.
                                     1. On the test specimen the (single surfaced) neutral electrode lamp
                                        must light up green.
                                     2. Activate AUTO CUT via the footswitch. It must be possible to activate
                                        the test specimen without error or warning signals.
                                                                                                                   80116-287 
              critical resistance
                                                                                                                   2018-09
                               for           Erbe Art. No.      Description
                   dual surfaced
                                                 20194-070      Patient cable NE
              return electrodes
                                                                or
76 / 84
                                                                         7 • Maintenance and servicing
Testbox 20100-101
1 2 3 4 5
6 7 8
9 10 11 12
Fig. 7-15
                              • The test specimen is connected to the power supply via the power
                                cord.
                              • The test is performed without a load.
                              • The NE socket of the test specimen is connected to the VIO Testbox via
                                the patient cable NE with the adapter cable.
                              • The dual-pedal footswitch with ReMode is connected.
                              1. On the test specimen the (dual surfaced) neutral electrode lamp must
                                 light up red.
                              2. Activate AUTO CUT via the footswitch. The test specimen must inhibit
                                 activation and emit or display an optical and acoustic warning.
              2nd test step
80116-287 
2018-09
Testbox 20100-101
1 2 3 4 5
6 7 8
9 10 11 12
Fig. 7-16
                              • The test specimen is connected to the power supply via the power
                                cord.
                              • The test is performed without a load.
                              • The NE socket of the test specimen is connected to the VIO Testbox via
                                the patient cable NE with the adapter cable.
                              • The dual-pedal footswitch with ReMode is connected.
                              1. On the test specimen the (dual surfaced) neutral electrode lamp must
                                 light up green.
                              2. Activate AUTO CUT via the footswitch. It must be possible to activate
                                 the test specimen without error or warning signals.
                                                                                                       77 / 84
7 • Maintenance and servicing
Testbox 20100-101
1 2 3 4 5
6 7 8
9 10 11 12
Fig. 7-17
                                   • The test specimen is connected to the power supply via the power
                                     cord.
                                   • The test is performed without a load.
                                   • The NE socket of the test specimen is connected to the VIO Testbox via
                                     the patient cable NE with the adapter cable.
                                   • The dual-pedal footswitch with ReMode is connected.
                                   1. On the test specimen the (dual surfaced) neutral electrode lamp must
                                      light up red.
                                   2. Activate AUTO CUT via the footswitch. The test specimen must inhibit
                                      activation and emit or display an optical and acoustic warning.
                                                                                                               80116-287 
                                                                                                               2018-09
                                           Erbe Art. No.      Description
78 / 84
                                                                          7 • Maintenance and servicing
                                                                              AE
                                                                                                NE
1 2 3 4 5
6 7 8
9 10 11 12
Testbox 20100-101
Fig. 7-18
                              • The test specimen is connected to the power supply via the power
                                cord.
                              • The NE socket of the test specimen is connected to the VIO Testbox via
                                the patient cable NE with the adapter cable.
                              • The AE socket of the test specimen is connected to the VIO test box via
                                the patient cable AE and the electrode handle with the laboratory
                                measuring cable.
                              1. Activate test specimen via COAG button on the electrode handle for
                                 approx. 10 seconds. During the entire activation time there must be
                                 no warning signal.
              2nd test step
80116-287 
2018-09
                                                                             AE
                                                                                                      NE
1 2 3 4 5
6 7 8
9 10 11 12
Testbox 20100-101
Fig. 7-19
                              • The test specimen is connected to the power supply via the power
                                cord.
                              • The NE socket of the test specimen is connected to the VIO Testbox via
                                the patient cable NE with the adapter cable.
                              • The AE socket of the test specimen is connected to the VIO test box via
                                the patient cable AE and the electrode handle with the laboratory
                                measuring cable.
                              1. Use the COAG button of the electrode handle to activate the test
                                 specimen for approx. 10 seconds . A visual warning must appear
                                 approx. 2 seconds after activation. The test specimen must not inter-
                                 rupt activation in this case.
                                                                                                           79 / 84
7 • Maintenance and servicing
AE
NE
1 2 3 4 5
6 7 8
9 10 11 12
Testbox 20100-101
Fig. 7-20
                                   • The test specimen is connected to the power supply via the power
                                     cord.
                                   • The NE socket of the test specimen is connected to the VIO Testbox via
                                     the patient cable NE with the adapter cable.
                                   • The AE socket of the test specimen is connected to the VIO test box via
                                     the patient cable AE and the electrode handle with the laboratory
                                     measuring cable.
                                   1. Activate the test specimen with the COAG button on the electrode
                                      handle for approx. 10 seconds. Within 2 seconds after activation at
                                      the latest an optical warning must be emitted and within another
                                      2 seconds an acoustic warning must be emitted. The test specimen
                                      must interrupt activation.
                                                                                                                    80116-287 
                                                                                                                    2018-09
                                   Bipolar resection adapter (VIO 300 D from V 2.2.x only)
80 / 84
                                                                                            7 • Maintenance and servicing
RESECTOSCOPE
                                                                                       MF
                                                                                               VIO MF
Fig. 7-21
                                             • The power cord connects the VIO 300 D to the power supply.
                                             • The VIO 300 D is switched off.
                                             • The bipolar resection adapter is connected to the MF socket of the
                                               VIO 300 D.
                                             • The bipolar resection adapter is connected to the safety tester (mea-
                                               suring input for insulation testing) via the measuring cable.
80116-287 
2018-09
                                             Test procedure
                                             1. Switch the VIO 300 D on.
                                             2. Start insulation testing. A charging current briefly flows via the
                                                decoupling capacitor. A resistance > 200 MOhm must then be dis-
                                                played.
                                                                                                                  81 / 84
7 • Maintenance and servicing
Oszilloscope
                                                                                              High Voltage
                                                                                              Differential probe
RESECTOSCOPE
                                                                          MF
                                                                                  VIO MF
Fig. 7-22
                                                                                                                      80116-287 
                                                                                                                      2018-09
                                    • The test setup is configured as illustrated above. IMPORTANT! The test
                                      is carried out without a connected load.
                                    • The power cord connects the VIO 300 D to the power supply.
                                    • The bipolar resection adapter is connected to the MF socket of the
                                      VIO 300 D.
                                    • The dual-pedal footswitch with ReMode is connected.
                                    Test procedure
                                    1. Call up Setup level 2 (see page 20).
                                    2. Select the "Test programs" setting.
                                    3. Use the Up button to select "Hardware TP" and press Enter to con-
                                       firm. The test specimen switches to the test program mode.
                                    4. Use the Up/Down buttons to select the "TP generator" test program
                                       and press Enter to call it up.
                                    5. Perform the following settings in the test program:
82 / 84
                                                        7 • Maintenance and servicing
                                                                              83 / 84
7 • Maintenance and servicing
RESECTOSCOPE
                                                                     MF
                                                                             VIO MF
HF power meter
Fig. 7-23
                                • The power cord connects the VIO 300 D to the power supply.
                                • The bipolar resection adapter is connected to the MF socket of the
                                  VIO 300 D.
                                • The bipolar resection adapter is connected to the HF power meter via
                                  the measuring cable.
                                • The dual-pedal footswitch with ReMode is connected.
Test procedure
                                                                                                          80116-287 
                                1. Set HF power meter to:
                                                                                                          2018-09
                                   RL = 50 ohms
                                2. Keep the previous settings in the "TP generator" test program:
                                   – Test mode: AUTO
                                   – Output socket: 3 BIPOLAR (3 = sockets slot, BIPOLAR = arrange-
                                      ment with multifuction socket, to which the bipolar resection
                                      adapter is connected)
                                3. Use the footswitch to activate the VIO 300 D.
                                4. Determine the power delivered on the HF power meter. The value
                                   must be > 100 watts.
84 / 84