Industrial Gas Thermometer Guide
Industrial Gas Thermometer Guide
Main Features
For corrosive gasses and
Wetted parts and case material
liquids stainless steel
Class 1 according to
Option : case material 1.4404
EN 13190 (316L) for aggressive atmos-
Capillary 0.5 to 30 m
phere
Mechanical or inductive
contacts
Applications
Food & Beverage
Oil & Gas / Chemical
Laboratory & Medical
Energy
Technical Data
Nominal size 100, 160 mm ATEX Only for inductive contacts (type 8xxx)
Temperature ranges -200 ... 800 °C Type examination certificate
DEKRA 14 ATEX 0034
Accuracy Class 1 (according to EN 13190)
EN 60079-0 : 2009
Protection rating IP 65 (EN 60529) EN 60079-11 : 2007
Immersion tube Stainless steel 1.4541 (AISI 321) EN 61241-11 : 2006
Capillary Stainless steel 1.4541 (AISI 321) Marking
max. length 30 m 0344
II 2 G
Bezel ring Stainless steel 1.4301 (AISI 304)
Ex ia IIC T4 Gb
Window Laminated safety glass II 2 D
Movement Brass Ex ia IIIC T135 °C ... T85 °C Db
Window gasket Elastomer Electrical data
Umax = 20 Vdc
Dial Aluminium, white Imax = 60 mA
Pointer Aluminium, black Pmax = 130 mW
Accessories Sliding connections and thermowells, see data- Sensor type: SI2-K08-Y1 (Turck)
sheets: B51.01 (AGF/AGW) and T6.215 (Ø 8) This certification is valid for connection to a certified intrinsically safe
Temperature limit 130 % F.S. circuit with maximum values as listed above and in the instructions
and certificate of the sensor. Design and specifications subject to change without notice
Options
Stainless steel housing 1.4404 (316L) Code 0110 Customer specific immersion length (1) Code 9003_xxxx
Movement stainless steel Code 0651 Red mark Code 9700
Process connections (see table on page 4/5) Code 73xx Green mark Code 9701
Oil filling (paraffin oil, Ondina) Code 0777 Stainless steel tag plate and wire Code 9941
Screws secured with loctite Code 0799
(1)
xxxx = L in mm.
2014-12-09
Dimensions in mm
120
A
°
K2
20
Ø15
PG 9
60
Ø12
K1
Diameter
Dimensions
100 160
Diameter A 101 161
Dimensions
100 160 B 80 83
A 101 161 C 13 13
B 104 108 D 132 196
K 92 122 H 5.5 6
P 116 178
K1 92 122
K2 15 0
C2 10 0
°
A
C2
D 132 196
K2
H 5.5 6
P 116 178
PG 9 K1 K1 123 126
K2 15 0
Design and specifications subject to change without notice
Diameter
Dimensions
100 160
A 101 161
120
°
B 110 114
A
D
C 19 19
D 132 196
H 5.5 6
P 116 178
K 92 122
2014-12-09
Dimensions in mm
Diameter
Dimensions
100 160
51 A 101 161
120
B 110 114
°
D 132 196
H 5.5 6
P 116 178
PG 9
K 92 122
Diameter
Dimensions
100 160
A
A 101 161
C
B 84 88
C 25 0
K2
D 112 180
E 109 113
P 75 139
K1 128 131
K2 15 54
Diameter
Dimensions
100 160
Design and specifications subject to change without notice
A 101 161
A
B 104 108
K 92 122
Ø7 (3x)
56
2014-12-09
Types of capillary
Process connections
Immersion tube
Models Code Connection HEX-1 HEX-2 T (mm)
731A G¼ 22 - 7 Ø 6 to 8
731B G⅜ 27 - 9 Ø 6 to 11
731C G½ 27 - 9 Ø 6 to 15
731D G¾ 32 - 9 Ø 6 to 20
731E G1 41 - 13 Ø 6 to 20
731P M18 x 1.5 27 - 8 Ø 6 to 12
731Q M20 x 1.5 27 - 9 Ø 6 to 14
Coupling nut (standard)
731R M24 x 1.5 32 - 9 Ø 6 to 18
732C G½ 27 - 14 Ø 6 to 15
732D G¾ 32 - 16 Ø 6 to 20
732E G1 41 - 18 Ø 6 to 20
Coupling nut (long model)
733A G¼ 17 - 12 Ø 6 to 8
733B G⅜ 22 - 12 Ø 6 to 11
Design and specifications subject to change without notice
733C G½ 22 - 14 Ø 6 to 15
733D G¾ 30 - 16 Ø 6 to 20
733E G1 36 - 18 Ø 6 to 20
733P M18 x 1.5 22 - 12 Ø 6 to 12
733Q M20 x 1.5 22 - 14 Ø 6 to 14
Rotating nipple (standard)
733R M24 x 1.5 27 - 14 Ø 6 to 18
734C G½ 22 - 20 Ø 6 to 15
734D G¾ 30 - 20 Ø 6 to 20
2014-12-09
734E G1 36 - 25 Ø 6 to 20
Rotating nipple (long model)
Process connections
Immersion tube
Models Code Connection HEX-1 HEX-2 T (mm)
735A G¼ 27 22 12 Ø 6 to 8
735B G⅜ 27 22 12 Ø 6 to 11
735C G½ 27 22 14 Ø 6 to 15
735D G¾ 27 27 16 Ø 6 to 20
735E G1 27 36 18 Ø 6 to 20
735H ¼ NPT 27 22 14 Ø 6 to 8
735J ½ NPT 27 22 20 Ø 6 to 15
735K ¾ NPT 27 27 20 Ø 6 to 20
735L 1" NPT 27 36 25 Ø 6 to 20
735P M18 x 1.5 27 22 12 Ø 6 to 12
735Q M20 x 1.5 27 22 14 Ø 6 to 14
Coupling nut + double nipple
735R M24 x 1.5 27 27 14 Ø 6 to 18
736A G¼ 22 27 12 Ø 6 to 11
736B G⅜ 22 27 12 Ø 6 to 14
736C G½ 22 27 14 Ø 6 to 18
736D G¾ 22 32 16 Ø 6 to 23
736E G1 22 36 18 Ø 6 to 30
736H ¼ NPT 22 27 14 Ø 6 to 13
736J ½ NPT 22 27 20 Ø 6 to 20
736K ¾ NPT 22 27 20 Ø 6 to 25
736L 1" NPT 22 36 25 Ø 6 to 32
736P M18 x 1.5 22 27 12 Ø 6 to 16
736Q M20 x 1.5 22 27 14 Ø 6 to 18
Sliding connection on the extension tube
736R M24 x 1.5 22 27 14 Ø 6 to 22
737A G¼ 22 27 12 Ø 6 to 11
737B G⅜ 22 27 12 Ø 6 to 14
737C G½ 22 27 14 Ø 6 to 18
737D G¾ 22 32 16 Ø 6 to 23
737E G1 22 36 18 Ø 6 to 30
737H ¼ NPT 22 27 14 Ø 6 to 13
737J ½ NPT 22 27 20 Ø 6 to 20
737K ¾ NPT 22 27 20 Ø 6 to 25
737L 1" NPT 22 36 25 Ø 6 to 32
737P M18 x 1.5 22 27 12 Ø 6 to 16
737Q M20 x 1.5 22 27 14 Ø 6 to 18
Sliding connection on the capillary
Design and specifications subject to change without notice
Temperature ranges
Code Range Code Range Code Range Code Range (double scale)
°C °C °F °C / °F
69T -200 ... 50 12T 0 ... 100 68U -100 ... 120 51V -30 ... 70 / -40 ... 160
77T -120 ... 40 20T 0 ... 120 09U -50 ... 120 68V -70 ... 50 / -100 ... 100
76T -100 ... 50 13T 0 ... 160 08U -40 ... 160 55V -30 ... 170 / 0 ... 350
74T -80 ... 40 22T 0 ... 200 02U 0 ... 200 11V 0 ... 60 / 30 ... 140
68T -70 ... 50 14T 0 ... 250 03U 0 ... 250 20V 0 ... 120 / 30 ... 250
50T -60 ... 40 23T 0 ... 300 04U 0 ... 300 13V 0 ... 160 / 30 ... 320
65T -40 ... 40 15T 0 ... 400 05U 0 ... 400 14V 0 ... 250 / 30 ... 500
55T -30 ... 170 25T 0 ... 500 06U 0 ... 500 15V 0 ... 400 / 30 ... 750
51T -30 ... 70 16T 0 ... 600 20U 30 ... 250 16V 0 ... 600 / 100 ... 1100
61T -30 ... 30 82T 0 ... 800 13U 30 ... 320
84T -20 ... 100 30T 100 ... 500 22U 30 ... 400
54T -20 ... 60 23U 30 ... 580
52T -20 ... 40 15U 30 ... 750
11T 0 ... 60 28U 100 ... 800
27T 0 ... 80 29U 200 ... 1000
Diameter of the Immersion tube (mm) Capillary length (m) How to use this table:
1. Select in the right table the first column, which
Ø6 Ø8 Ø9 Ø 10 Ø 11 Ø 12 Ø 14 Ø 16 ≤2 ≤5 ≤ 10 ≤ 20 ≤ 30
covers your capillary length.
70 35 30 25 20 18 15 13 160 160 160 160 160 2. Move down in this column to the first value,
120 60 45 35 30 25 20 17 80 80 80 80 100 which is not higher than your temperature span.
190 90 60 55 45 35 30 25 60 - - - -
3. Move horizontal to the left table into the column
←
-
135
170
100
130 100
←
80 65
80
55
65
40
50
30
40
-
-
60
-
-
60
-
-
80
-
4. Read the minimum length of the sensitive part of
the immersion tube for your selection.
- - 190 150 120 95 70 55 - - - 60 - If this is too long, choose a larger diameter, a larger
range or a shorter capillary.
Length of sensitive part Ls (mm) Min. temperature span (°C)
Example:
Standard version with plain immersion tube (without process connection) 4 m capillary, -20 … 40 °C (span = 60 °C),
Immersion tube Ø 10 mm:
Minimum length of the sensitive part: Ls ≥ 80 mm