AbraPol 20 15887001 en Revision 0
AbraPol 20 15887001 en Revision 0
Always state Serial No and Voltage/frequency if you have technical questions or when ordering spare parts.
You will find the Serial No. and Voltage on the type plate of the machine itself. We may also need the Date
and Article No of the manual. This information is found on the front cover.
The following restrictions should be observed, as violation of the restrictions may cause cancellation of
Struers legal obligations:
Instruction Manuals: Struers Instruction Manual may only be used in connection with Struers equipment
covered by the Instruction Manual.
Service Manuals: Struers Service Manual may only be used by a trained technician authorised by Struers.
The Service Manual may only be used in connection with Struers equipment covered by the Service Manual.
Struers assumes no responsibility for errors in the manual text/illustrations. The information in this manual is
subject to changes without notice. The manual may mention accessories or parts not included in the present
version of the equipment.
Original instructions. The contents of this manual is the property of Struers. Reproduction of any part of
this manual without the written permission of Struers is not allowed.
Struers A/S
Pederstrupvej 84
DK-2750 Ballerup
Denmark
Telephone +45 44 600 800
Fax +45 44 600 801
AbraPol-20
Instruction Manual
AbraPol-20
Safety Precaution Sheet
To be read carefully
before use
1. The operator should be fully aware of the use of the machine according
to the instruction manual.
4. Before lifting the machine, ensure that the supplied lifting boom is
securely bolted to the machine.
5. When lifting the machine using a forklift, lift from front - never lift the
machine from the side or the rear.
6. When lifting the machine using lifting straps, ensure that the straps are
crossed and do not press on the sides of the machine.
8. Check that the actual voltage corresponds to the voltage stated on the
side of the machine and that the connections comply with local
regulations.
The machine must be earthed.
9. Make sure that the water connections are without leaks. The main
water supply should be turned off or disconnected if you leave the
machine unattended.
10. Alcohol based consumables: follow the current safety rules for
handling, mixing, filling, emptying and disposal of the
alcohol-based liquids.
Struers recommend the use of an external exhaust system.
AbraPol-20
Instruction Manual
11. Make sure that the specimens in the specimen holder are securely
fixed when operating.
12. If you observe malfunctions or hear unusual noises - stop the machine
and call technical service.
13. The machine must be disconnected from the mains prior to any service.
Wait until residual potential on the capacitors is discharged.
14. Do not cycle mains power more than once every three minutes.
Damage to the drive will result.
15. To achieve maximum safety and lifetime of the machine, use only
original Struers consumables.
The equipment should only be used for its intended purpose and as detailed in the Instruction Manual.
The equipment is designed for use with consumables supplied by Struers. If subjected to misuse, improper
installation, alteration, neglect, accident or improper repair, Struers will accept no responsibility for
damage(s) to the user or the equipment.
Dismantling of any part of the equipment, during service or repair, should always be performed by a qualified
technician (electromechanical, electronic, mechanical, pneumatic, etc.).
AbraPol-20
Instruction Manual
User’s Guide
Table of Contents Page
2. Operation
Front Panel ....................................................................................... 16
Front Panel Controls ........................................................................ 17
Software Settings ............................................................................. 18
Setting the Language ................................................................ 21
Setup Bottle Configuration ........................................................ 23
Reading the Display ......................................................................... 26
Sleep Mode ............................................................................... 27
Changing/Editing Values .................................................................. 28
Numeric Values ......................................................................... 28
Alphanumeric Values ................................................................ 29
Text Values ............................................................................... 31
Programming a Preparation Step ..................................................... 32
Mounting a Preparation Disc (300 or 350 mm) ................................ 32
Inserting the Specimen Holder ......................................................... 32
Adjusting the Specimen Holder Position .......................................... 33
Starting the Preparation Process (Struers Method) ......................... 35
Stopping the Preparation Process ................................................... 35
Manual Functions ............................................................................. 36
1
AbraPol-20
Instruction Manual
3. Maintenance
General Cleaning ............................................................................. 38
Cooling Tank ............................................................................. 38
Painted Surfaces ....................................................................... 38
Weekly Service ................................................................................ 38
Checking the Recirculation Cooling Unit ................................... 38
Monthly service ................................................................................ 38
Replacing the Cooling Water .................................................... 38
Cleaning of Tubes ..................................................................... 39
Yearly Service .................................................................................. 43
Inspection of Cover ................................................................... 43
2
AbraPol-20
Instruction Manual
1. Getting Started
Checking the Contents of the In the packing box you should find the following parts:
Packing Box 1 AbraPol-20 (Grinding polishing machine)
1 Outlet hose 2.5 m with straight PVC pipe,
for connection to external cooling unit
1 Hose clamp, 32-50 mm
1 Inlet hose for tap water 3/4" (2m)
1 Filter gasket
1 Gasket ø11/ø24 (1.5 mm)
1 Reduction ring with gasket
1 Hose for compressed air
1 Hose connection for compressed air
2 Hose clamps, 12 mm
2 Silicone tubes for use with alcohol based lubricants
2 Bottles, ½ l, with lids
(4 Bottles if delivered with Additional Dosing Unit)
1 Bottle, 1 l, with lid
(2 Bottles if delivered with Additional Dosing Unit)
1 Fork spanner (24 mm)
1 Set of Instruction Manuals
Additional Dosing Unit (accessory) Three additional pumps for dosing of OP-Suspension, lubricants and
suspensions.
Outlet Kit (accessory) Shift valve for recirculation cooling unit or drain.
Accurate Material Removal Unit For accurate removal of material. Stops process after required
(accessory) amount of material has been removed.
Unpacking and Placing AbraPol-20 should be placed on a plane and horizontal floor, capable
AbraPol-20 of bearing the weight of this machine, please refer to the Technical
Data section.
The machine must be placed close to the power supply, compressed
air, water mains and water outlet facilities.
Unscrew the nuts from the transport brackets fixing the machine
to the pallet.
Lift the machine from the pallet by means of a forklift truck from
the front, and place in a suitable location.
Remove the safety-springs from the front crossbar, and remove
the bar.
Important
If necessary, turn the adjustable feet
so that the machine stands firmly and is level.
3
AbraPol-20
Instruction Manual
Getting Acquainted with Take a moment to familiarise yourself with the location and names of
AbraPol-20 the AbraPol-20 components.
Front view
Safety guard
Control panel,
(see 2. Basic Operations)
Emergency Stop
Nozzle block
Bottle unit
Specimen holder motor
Handle for adjusting specimen holder position
Specimen quick coupling
LED-light to illuminate preparation disc
4
AbraPol-20
Instruction Manual
Side views
Rear view
5
AbraPol-20
Instruction Manual
IMPORTANT
AbraPol-20 must be connected with an electric cable and an additional wire
for earth connection which MUST be hard-wired into the mains
(ref. EN 50178 / 5.2.11.1)
Important:
For Electrical Installations with Residual Current Circuit Breakers
For AbraPol-20 connected to electrical installations with residual current
circuit breakers, a residual current circuit breaker,
type B, 30 mA is REQUIRED
(ref. EN 50178 / 5.2.11.1).
Please contact a qualified electrician to verify which option is suitable for the
local installation setup.
Allen screws
6
AbraPol-20
Instruction Manual
Supplying Compressed Air Connect the compressed air supply with the inlet on the rear, left
side of the machine by means of the air hose and the hose
connection delivered with the machine.
Fasten the air hose with a hose clamp.
Connection to an External An exhaust system should be connected when using alcohol based
Exhaust System suspensions or lubricants.
Connect a 50 mm pipe to the outlet at the rear of the machine,
on the left and connect to the exhaust system.
Recommended capacity for exhaust system: 180m3/h at 0mm water
gauge.
7
AbraPol-20
Instruction Manual
Water Outlet AbraPol-20 is delivered with a short piece of hose (0.8 m) attached to
the water outlet. This is sufficient to connect the machine to a
recirculation cooling unit placed underneath the machine.
8
AbraPol-20
Instruction Manual
Mounting a Recirculation Please refer to the manual for the Recirculation Cooling Unit for
Cooling Unit (accessory) details.
IMPORTANT
Always maintain the correct concentration of Struers Additive in the cooling
water (percentage stated on the Additive container).
Remember to top up with Struers Additive each time you refill with water.
Tube holder
9
AbraPol-20
Instruction Manual
Mains Water Connection When the AbraPol-20 is used without a Recirculation Cooling Unit,
with the second dosing unit for dosing of OP-Suspensions or when
disc cooling is required, AbraPol-20 must be connected to water
directly from the mains.
Mount the pressure hose onto the water inlet tube on the back of
the AbraPol-20.
− Insert the filter gasket in the coupling nut with the flat side
against the pressure hose.
− Tighten the coupling nut completely.
Mount the other end of the pressure hose on the water mains tap
for cold water:
− Mount the reduction ring with gasket on the water mains tap,
if necessary.
− Introduce the gasket and tighten the coupling nut completely.
Placing Bottles in the Place the filled bottles in the dosing unit and connect the tubing.
Dosing Unit Enter the bottle details in the Bottle Configuration menu to make
them available for preparation methods. See “Setup Bottle
Configuration”.
The tubes can easily go through the bottle unit to reach larger
containers placed on the floor e.g. Lubricants.
Nb!
OP-Suspensions can not be placed on the floor.
10
AbraPol-20
Instruction Manual
X8
A8
X1 X2 X3 X4 X5 X6 X7
RD RD RD RD
1 1 1 1
2 2 2 2
BK BK BK BK
A9
M6 M7 M8
RD RD
RD
+ + + BU10
- - -
BK BK BK PLUG
&
PUMP4 PUMP5 PUMP6 PLAY
NB
Remember the earth connections.
NB
OP-suspensions can only be used in bottle 4.
11
AbraPol-20
Instruction Manual
Mounting the Outlet Kit The shifting valve can be mounted from the rear of the AbraPol-20
(accessory) (recommended) or from the front.
Switch off the machine
To install from the rear:
Remove the lower back cover plate
To install from the front:
Remove lower cover plate and the Recirculation Cooling Unit, if
there is one.
Take the entire unit, and place it in position by letting the two
Allen bolts slide into the slots located in the left rear side.
Allen bolts
in slots
Pipe A: Pipe C:
Connect to Outlet hose
drain hose to the drain
Pipe B:
Connect to
Recirculation Cooling Unit
12
AbraPol-20
Instruction Manual
Cut the water outlet hose to the required length and mount onto
pipe A of the Shifting Valve, clamp using a hose clamp.
Take another piece of drain hose and check that it is long
enough to reach the Recirculation Cooling Unit (Ø40 mm and
0.5m long). Mount onto pipe B and clamp using a hose clamp.
Connect to the Recirculation Cooling Unit.
Connect a piece of drain hose from the outlet at the rear (pipe C)
and lead to the drain. Check that the outlet hose slopes
downwards towards the drain along the whole of its length.
Reposition the Recirculation Cooling Unit.
Make sure that the hoses do not interfere with the belts.
13
AbraPol-20
Instruction Manual
Locate the electrical cable for the air solenoid valve (located on
the inner left side of the machine) and remove the cap covering
the stripped wires.
Open the cover of the DIN terminal by removing the screw and
connect the wires to the terminals.
Replace the cover for the DIN terminal.
Note:
It is not important to which terminal the wires are connected,
each wire can be connected to either terminal.
Find the air supply hose and cut off the plug (stopper). Fix the
cut end into the air supply inlet.
Replace the front cover and reconnect main power.
Adjusting Disc Cooling Should the rate of disc cooling require adjustment:
Loosen the contra nut and turn the regulator screw to the
required rate.
Re-tighten the contra nut.
14
AbraPol-20
Instruction Manual
Connection of Removal
measurement unit A7
A7
GN
W20
BN WH
2 3
1 2 3 4 5
X22
WHBN GN YE YE
W20
15
AbraPol-20
Instruction Manual
2. Operation
Front Panel For more information also see page on Front Panel Controls.
16
AbraPol-20
Instruction Manual
Esc
aborts functions/changes.
ABRASIVE
Starts the preparation process.
START EMERGENCY
STOP
- Push the red button to stop.
- Turn the red button clockwise
to release.
STOP MAIN
SWITCH
the right side of the machine.
17
AbraPol-20
Instruction Manual
Software Settings Switch on the power at the main switch located at the right hand side
of the machine.
The following display will appear briefly:
The display will then change to the last screen shown before
AbraPol-20 was switched off, usually a preparation method.
18
AbraPol-20
Instruction Manual
When switching on AbraPol-20 for the first time, the following screen
will appear for a short while:
19
AbraPol-20
Instruction Manual
The MAIN MENU is the highest level in the menu structure. Pre-
defined Struers methods, your own methods, manual functions or the
configuration menu can then be selected.
20
AbraPol-20
Instruction Manual
21
AbraPol-20
Instruction Manual
22
AbraPol-20
Instruction Manual
23
AbraPol-20
Instruction Manual
Turn and push knob to select type of
Lubricant/Suspension.
24
AbraPol-20
Instruction Manual
Esc Press Esc twice to return to the Main Menu.
NB
The different suspensions and lubricants used must always be defined here
so that AbraPol-20 can locate the correct suspension or lubricant.
25
AbraPol-20
Instruction Manual
Reading the Display The display can show various kinds of information, for example
parameters of a preparation method, or the ongoing process.
An example of the screen for a preparation method is shown below:
A Heading.
B Inverted text: cursor position.
C Function key options
26
AbraPol-20
Instruction Manual
Sleep Mode To increase the lifetime of the display, the backlight is dimmed
automatically if AbraPol-20 has not been used for 15 min. Press any
key to re-activate the backlight.
Please Note
The sample screens in this Instruction Manual show a number of
possible texts. The actual display screen may differ from the
samples in the Manual.
27
AbraPol-20
Instruction Manual
Changing/Editing Values Depending on the type of value, there are two different ways of
editing.
STRUERS METHODS can not be edited or changed in any way, this
is only valid for USER METHODS and CONFIGURATION.
Numeric Values
Turn knob to select the value to be changed, e.g. Force:
A scroll box appears around the value.
28
AbraPol-20
Instruction Manual
A pop-up menu appears.
29
AbraPol-20
Instruction Manual
30
AbraPol-20
Instruction Manual
31
AbraPol-20
Instruction Manual
Inserting the Specimen Holder Place the specimen holder under the quick coupling.
Press and hold the black flange of the column down with your
thumbs while guiding the pressure tap of the specimen holder
into the coupling.
Release the black flange.
Turn the specimen holder until the three pins engage with the
corresponding holes.
32
AbraPol-20
Instruction Manual
Adjusting the Specimen Holder The position of the specimen holder has to be adjusted correctly in
Position relation to the preparation disc to get the best possible preparation
results.
Loosen the two handles located on each side of the specimen
holder motor.
The specimen holder motor can now be moved manually to the
sides. The eccentricity can be read on the scale on the specimen
holder motor.
When the correct position is found, fasten the handles again.
33
AbraPol-20
Instruction Manual
34
AbraPol-20
Instruction Manual
Stopping the Preparation When the time has elapsed the preparation disc will automatically
Process stop and the specimen holder will return to its initial position. The
machine is now ready for the next step.
A preparation step can be stopped at any time during the
process by pressing the key STOP .
A process can also be stopped using the emergency stop button.
The emergency stop button must be pulled outwards again to
release it before the machine can be restarted. The sample
mover will return to its initial position before the process can be
started again.
35
AbraPol-20
Instruction Manual
Manual Functions Manual preparation can also be carried out from the Manual
Functions menu.
36
AbraPol-20
Instruction Manual
Press Start.
The preparation disc will start rotating at the pre-set speed
and the selected suspension/lubricant will automatically be
added in the pre-set level.
Press Stop.
The preparation disc will stop rotating and dosing will stop.
37
AbraPol-20
Instruction Manual
3. Maintenance
General Cleaning The machine and the Recirculation Cooling System should be kept
as clean as possible in order to avoid contamination of the
specimens.
Cooling Tank Each time the water is changed, the cooling tank should be cleaned
carefully to remove all grinding waste.
IMPORTANT
Always maintain the correct concentration of Struers Additive in the cooling
water (percentage stated on the container of the Additive). Remember to
add Struers Additive each time you refill with water.
IMPORTANT
Do not use benzene or petroleum for cleaning when the additive is used.
Painted Surfaces Painted surfaces and keyboard should be cleaned with a moist cloth
and common household detergents.
IMPORTANT
Never use alcohol, acetone or similar solvents.
Weekly Service
Checking the Recirculation The cooling unit should be checked for cooling water after 8 hours
Cooling Unit use or at least every week. The unit must be refilled if the flushing
pump cannot reach the cooling water or if the cooling water is too
contaminated.
Please refer to the manual for the Recirculation Cooling Unit for
details.
Monthly service
Replacing the Cooling Water Replace the cooling water in the Recirculation Cooling Unit at
least once a month.
Please refer to the manual for the Recirculation Cooling Unit for
details.
38
AbraPol-20
Instruction Manual
Cleaning of Tubes The cleaning procedure for CLEANING OF TUBES can be used
weekly or when necessary.
Clean the tubes every time bottles are changed or replaced, thus
avoiding lubricant/suspension left in tubes interfering with the
preparation procedures.
39
AbraPol-20
Instruction Manual
40
AbraPol-20
Instruction Manual
41
AbraPol-20
Instruction Manual
Cleaning of All Used Tubes The tubes in use should be cleaned from time to time.
F1
Press F1 to start the cleaning process and follow the pop-
up instructions.
42
AbraPol-20
Instruction Manual
Yearly Service
Inspection of Cover Visually inspect the cover and the glass for signs of wear or
damage.
Important
Carry out inspection at more regular intervals if AbraPol-20
is used for more than one 7 hour shift a day.
43
AbraPol-20
Instruction Manual
Reference Guide
Table of Contents Page
44
AbraPol-20
Instruction Manual
45
AbraPol-20
Instruction Manual
1. Advanced Operations
46
AbraPol-20
Instruction Manual
Bottle Configuration Please see the User’s Guide section for detailed information on
Bottle Configuration.
Setup User Consumables The menus User Surface Configuration, User Suspension
Configuration and User Lubricant Configuration allow the user to add
ten grinding/polishing surfaces, ten suspensions and ten lubricants.
47
AbraPol-20
Instruction Manual
Defining a User Surface Using this option, it is possible to define up to 10 new user-defined
Configuration surfaces. The name, the abrasive rule and lubricant rule for each
surface can also be defined.
48
AbraPol-20
Instruction Manual
49
AbraPol-20
Instruction Manual
50
AbraPol-20
Instruction Manual
Defining a User Suspension Using this option, it is possible define up to 10 new user-defined
Configuration suspensions. The name, the abrasion type and lubricant rule for
each suspension can then be defined.
51
AbraPol-20
Instruction Manual
Note
There are only 2 types of abrasives available.
Be sure to select the correct one.
52
AbraPol-20
Instruction Manual
53
AbraPol-20
Instruction Manual
Note
If the suspension used requires the use of a lubricant, select 1.
If the suspension doesn’t require a lubricant, e.g oxide polishing
suspensions or DiaPro, select 2.
If the abrasive is only to be used with water-free lubricants, select 3.
54
AbraPol-20
Instruction Manual
Defining a User Lubricant Using this option, it is possible to define up to 10 new user-defined
Configuration lubricants. The name and lubricant type can be defined for each
lubricant.
55
AbraPol-20
Instruction Manual
56
AbraPol-20
Instruction Manual
Note
If the lubricant contains water, select 1.
If the lubricant is water-free and is to be used when preparing water-
sensitive materials, select 2.
57
AbraPol-20
Instruction Manual
Options Select Options and push the knob to activate the OPTIONS menu
58
AbraPol-20
Instruction Manual
Display Contrast The contrast settings of the display can be adjusted for maximum
readability.
Units Select between µm or mils (mils??) for the removal rate settings.
Acoustic signal Select to have each activation of a key on the touchpad confirmed by
an acoustic signal.
Pause between steps Several preparation steps can be combined to a single step with
several phases.
Normally steps with the same surface, suspension and lubricant will
be run without interruption. Pause between steps can be used to
program things differently.
F1, Default Value The factory setting of some parameters can be restored by pressing
the function key F1 when the appropriate value is highlighted.
59
AbraPol-20
Instruction Manual
Operation Modes:
Process: Methods can be selected and viewed but no editing is
possible.
Development: Methods can be selected, viewed and edited
Configuration: Methods can be selected, viewed and edited and bottles
can be configured.
60
AbraPol-20
Instruction Manual
61
AbraPol-20
Instruction Manual
62
AbraPol-20
Instruction Manual
Important!
Remember to make a note of the new Pass code
as settings can no longer be changed without the Pass code.
63
AbraPol-20
Instruction Manual
Method Options Methods can be adjusted to suit the particular requirements of the
user.
Struers Methods The AbraPol-20 software includes 10 Metalog Guide Methods. The
AbraPol-20 will automatically calculate the correct preparation time
and dosing level depending on if a 300 mm disc and 350 mm disc is
used.
As the Struers methods can not be changed or deleted, the first step
will often be to copy a Struers method into the User Methods
database and then adjust it to suit individual requirements.
Storing Capacity The new/adapted methods can be stored in 10 groups; each group
can contain 20 methods resulting in an overall storage capacity of
200 methods.
F1 Select the correct method and press
F1:COPY.
64
AbraPol-20
Instruction Manual
65
AbraPol-20
Instruction Manual
Renaming a Method The names of the methods in the USER METHODS database can be
edited and changed. After copying a method from the Struers
Methods the name can be changed to a name of your choice.
NB!
If a method name starts with “Copy of ……” the text “Copy of”
can be deleted by pressing F4 once.
66
AbraPol-20
Instruction Manual
Name Editing Principles Place the auxiliary cursor on the character you want to select, using
F1: LEFT or F2: RIGHT. Turn the knob to move to the character
cursor to select a character.
A Auxiliary cursor
B Character cursor
C Character set
67
AbraPol-20
Instruction Manual
68
AbraPol-20
Instruction Manual
Saving a Method While working with a user method, changes made in the database
can be saved.
F4 Press F4: SAVE to save the changes after all necessary
parameters have been altered.
69
AbraPol-20
Instruction Manual
Important
When saving changes, the original method will be overwritten.
To preserve the original method, a copy should be made and renamed. The
new method can then be changed as required.
For details, see Copying a Method.
70
AbraPol-20
Instruction Manual
When asked for confirmation,
NB!
Struers Methods can not be deleted.
71
AbraPol-20
Instruction Manual
Step Options Single steps can also be copied to customise preparation methods.
Copying a Step Select the method and the preparation step to be copied.
72
AbraPol-20
Instruction Manual
Inserting a Step
F4 Press F4: Save.
The changed method is saved.
73
AbraPol-20
Instruction Manual
F3 Press F3 Delete.
A pop-up will appear.
F4 Press F4 Save
To save the changed method.
74
AbraPol-20
Instruction Manual
Method Parameters For every method step a series of parameters can be changed:
Surface Pushing the knob allows you to select between all the surfaces in the
AbraPol-20 database:
75
AbraPol-20
Instruction Manual
76
AbraPol-20
Instruction Manual
Level The dosing level can be set. The first number is the pre-dosing time
in seconds. It can be set between 0-10
The second number is the dosing level that can be adjusted between
0-20
Process Time When Time is selected in Process mode the preparation time can be
adjusted between 5 seconds and 60 minutes.
Rotation speed The speed of the disc can be set between 50 – 500 rpm. When
SynchroSpeed is set to Yes the max. setting is 300 rpm.
The speed of the holder can be set between 50 – 300 rpm.
Sample holder direction The direction can be set to either co- or counter-rotation.
Co-rotation will give the best preparation result and the most uniform
polishing. Counter-rotation can be used when polishing using oxide
polishing suspensions to keep the suspension on the polishing cloth.
77
AbraPol-20
Instruction Manual
300 rpm on Sample Mover When using 300 rpm on the sample mover and 300 rpm on the disc
and Disc please note the following recommendations:
− Use disc cooling.
− Pre-dose sufficiently when working with new preparation
surfaces.
− Use a 200 mm sample holder on a 350 mm surface to
distribute the lubricant and abrasive adequately.
− Use DP-Lubricant, Green or DP-Lubricant, Red.
DP-Lubricant, Blue is alcohol-based and will evaporate too
quickly.
78
AbraPol-20
Instruction Manual
Manual Functions Several manual functions are included in the software of AbraPol-20.
Press Esc until you reach the Main Menu.
79
AbraPol-20
Instruction Manual
Manual Preparation Manual functions are possible as the disc can run independently of
the specimen holder.
Disc Speed The speed can be set between 50 and 300 rpm. in steps of 10 rpm.
Suspension The different suspensions that are configured in the bottle set-up can
be selected here.
Dosing Level The dosing level can be set to a value between 0 and 20.
80
AbraPol-20
Instruction Manual
Dosing Level The dosing level can be set to a value between 0 and 20.
Starting Manual Preparation After defining all parameters the manual preparation can be started
by pressing Start.
NB!
If no suspension or lubricant is selected or the dosing levels are set to 0
the manual preparation can not be started.
Sample mover UP/DOWN This function is described earlier in the manual in the section
Adjusting the Specimen Holder Position.
Adjusting the Speed of the The speed at which the Sample Mover head moves up and down
Sample Mover Head can be adjusted by turning the adjustment screw using a screwdriver:
Increase speed: turn the screw head Counter-clockwise
Reduce speed: turn the screw head Clockwise
Adjustment
screw
Tip!
Take care not to set the speed too high,
this could result in damage to the samples or the
grinding/polishing consumable.
81
AbraPol-20
Instruction Manual
Changing Tubes When working with alcohol based lubricants, the Novoprene tubes
mounted in the pumps supplied with AbraPol-20 will harden over
time. Therefore a piece of silicone tubing is supplied with the
AbraPol-20 as Silicone has a better resistance against alcohol.
Tabs
Press the two tabs on the pump and remove the bottom cover.
82
AbraPol-20
Instruction Manual
Remove the 3 loose rollers and replace the Novoprene tube with
the silicon tube.
83
AbraPol-20
Instruction Manual
Remember...
Struers offers a comprehensive range of consumables for
grinding and polishing.
84
AbraPol-20
Instruction Manual
Service Information Struers recommends that a regular service check be carried out after
every 1500 hours of use. Information on total operation time and
servicing of the machine is displayed on the screen at start-up:
85
AbraPol-20
Instruction Manual
Metalog Guide™
A complete guide to materialographic specimen preparation.
www.struers.com/KNOWLEDGE/Metalog Guide.
86
AbraPol-20
Instruction Manual
4. Trouble shooting
87
AbraPol-20
Instruction Manual
88
AbraPol-20
Instruction Manual
89
AbraPol-20
Instruction Manual
90
AbraPol-20
Instruction Manual
91
AbraPol-20
Instruction Manual
5. Technical Data
Subject Specifications
Metric/ US
International
Disc Rotational speed 50-500 rpm 50-500 rpm
in steps of 10 in steps of 10
Size 300/350 mm 11.8”/13.7”
Power Consumption 2.2 kW 3 Hp
Specimens Speed 50-300rpm 50-300rpm
in steps of 10 in steps of 10
Direction CW/CCW
Force 0-700 N 0-150 lbf
Motor power consumption 0.55 kW 0.75 Hp
Dosing Unit Level 0-20
No. of pumps 3 (6 with Additional Dosing Unit)
Software and LC Display 320x240 pixels
electronics Controls Touch pads/Push-turn knob
Memory EPROM/RAM/NV-RAM
Dimensions Width 840 mm 33.1”
and weight Depth 980 mm 38.6”
Height 1560 mm 61.4”
Weight 400 kg 880 lbs
Environment Safety standards Please refer to the Declaration of Conformity
Noise level (idle 300/300) 77 dbA
Surrounding temperature 5-40°C
Humidity Max. 95%RH
Water mains Water supply max. 10 bar max. 10 bar
Compressed Compressed air supply 6-10 bar
air Recommended quality, ISO 8573-1,
class 5.6.4 for normal use
92
AbraPol-20
Instruction Manual
Subject Specifications
Electrical Data
Power consumption 3 kW
No. of phases 3 (3L+PE)
Output, main motor 2.2 kW
Voltage/frequency: Max. load
3 x 200 - 210 V / 50/60Hz 16.5 A
3 x 220 - 240 V / 50/60 Hz 15.6 A
3 x 380 - 415 V / 50/60 Hz 9.0 A
3 x 460 - 480 V / 50/60 Hz 7.2 A
Mains Cable Voltage/frequency: Min. Minimum cable size Max. Minimum cable size
Recommend- Fuse @ Min. fuse Fuse @ Max. fuse
ation size size
3 x 200 - 210 V / 50/60Hz 25 A 3xAWG12 / 4 mm² + PE 50 A 3x AWG10 / 4 mm² + PE
3 x 220 - 240 V / 50/60 Hz 25 A 3x AWG12 / 4 mm² + PE 50 A 3x AWG10 / 4 mm² + PE
3 x 380 - 415 V / 50/60 Hz 20 A 3x AWG12 / 4 mm² + PE 50 A 3x AWG10 / 4 mm² + PE
3 x 460 - 480 V / 50/60 Hz 20 A 3x AWG12 + PE 50 A 3xAWG10 + PE
Important:
Local standards may overrule the recommendations for the main supply cable.
If necessary, please contact a qualified electrician to verify which option is suitable for the
local installation setup.
Residual type B, 30 mA (or better) is REQUIRED.
Current Circuit
Breaker
93
AbraPol-20
Instruction Manual
6. Menu Overview
MAIN MENU
e Disc. e Accoustic
VIEW EDIT Cooling Signal
METHOD STEP METHOD STEP
e Options e Pause
Between steps
s s
e Operation e Disc
Mode Cooling
PROCESS PROCESS
START START e Synchro
Speed
94
AbraPol-20
Instruction Manual
Main Controls
Main Switch The main switch is located at the right side of the machine.
Emergency Stop Red push button on the front of the machine. Stops all ongoing
operations. Release the emergency stop by pulling the button.
Double Start Buttons Start the actual preparation by pressing the two buttons
simultaneously.
Function Keys
F1 F2 F3 F4
Controls for various purposes.
See the bottom line of the individual screens.
95
AbraPol-20
Instruction Manual
Editing process For editing the process, push knob during the process.
Copying a Preparation Method Turn the knob to select the preparation method you want to copy.
Press F1 for copying a Preparation Method
Press F1 to accept copying
Enter the screen USER METODS press F2 to insert the method
96
588 AbraPol-20
588F_2016
The above has been declared according to the global method, module A
Dansk Overensstemmelseserklæring
Fabrikant, Struers ApS
ansvarlig for Teknisk Pederstrupvej 84
Dossier DK-2750 Ballerup, Danmark
Telefon 44 600 800
erklærer herved, at Produktnavn: AbraPol-20
Type nr.: 588
Maskintype: Slibe og poliermaskine
Bankers:
Nordea Bank Denmark A/S
Account no.: 2191-0126305060
Letter of conformity: Safety hood/window for the following equipment VAT no.: 15683309
The safety hoods/window primary function is to protect against samples not clamped properly.
The safety hood is made from PETG, which is generally used in all Struers machines as “safety glass”.
PETG is resistant to scratches and chemicals and provides the highest degree of safety for the
operator.
This is to certify that the product is produced according to regulations from the Commodity Supplier.
Yours faithfully
Flemming Perret-Gentil
R&D Manager
R&D Department, Equipment
Struers assumes no responsibility for errors in the manual text/illustrations. The information in this manual is
subject to change without notice. The manual may mention accessories or parts not included in the present
version of the equipment.
Original instructions. The contents of this manual are the property of Struers. Reproduction of any part of
this manual without the written permission of Struers is not allowed.
Struers A/S
Pederstrupvej 84
DK-2750 Ballerup
Denmark
Telephone +45 44 600 800
Fax +45 44 600 801
AbraPol-20
Spare Parts and Diagrams
AbraPol-20
Drawings
AbraPol-20, complete ....................................................... 15880001O
Casing with electrical, assembled. ................................... 15880005E
Casing with motors, assembly ......................................... 15880007C
Casing, assembly ............................................................. 15880010I
Main mechanism, assembly ............................................. 15880020F
Box for tub, assembly ....................................................... 15880050C
Tub, assembled ................................................................ 15880058D
Sample motor, assembly .................................................. 15880045E
Plate with PCB and pneumatic distr. assembled. ............ 15880034B
Control box AbraPol-20, assembled ................................ 15880082I
Front, assembly ................................................................ 15880025B
Dosing Arm, assembled ................................................... 15090055E
Safety guard, assembly .................................................... 15890080D
Top water valve ................................................................ 15880095D
Window, assembled ......................................................... 15090004
Quick-release coupling, complete .................................... 15090009C
Disc, assembled ............................................................... 15090005
Air connection, assembled ............................................... 15090032E
Pumps module, assembly ................................................ 15880051F
Valves for Cooli water ...................................................... 15880090C
Diagrams
Transformer Connections (2 pages) ................................ 15093452A
Air diagram ....................................................................... 15882000C
Water connections ........................................................... 15882005B
Block diagram .................................................................. 15883050E
Circuit diagram main voltage ............................................ 15883100E
CPU Board A1 (5 pages) ................................................. 15883120F
Doser connection board ................................................... 15883200A
Wiring diagram (10 pages) ............................................... 15883450E
Overview, variant parts AbraPol-20.................................. 15887600D
1
AbraPol-20
Spare Parts and Diagrams
2
AbraPol-20
Spare Parts and Diagrams
3
AbraPol-20
Spare Parts and Diagrams
4
AbraPol-20
Spare Parts and Diagrams
5
AbraPol-20
Spare Parts and Diagrams
6
AbraPol-20
Spare Parts and Diagrams
7
AbraPol-20
Spare Parts and Diagrams
8
AbraPol-20
Spare Parts and Diagrams
9
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
22 0
390
21 0
40
20 0 600 610 62 0 63 0 64 0
B B
10
140
150
180
13 0
190 80
B D
SECTION B-B
47 0 SC ALE 1 : 2
C
B C
3 40 2 50 480
D D
180
1 90 490
320
240 40
570 Locked by Omnifit 50M
400 50
E E
H H
2 80 1
370 35 0 21 0 22 0
2 90 1
3 00 8
A4 3 10 4
1030 F1+F2+F3 Variant
F3 F2 F 1
A3 3 20 1
3 30 2
3 40 1
1040 F4+F5 Variant
3 50 1
DETAIL I F6 F5F4 3 60 1
SC ALE 1 : 4 3 70 6
J
XPE8 J
670 30 0 630 6 20 300 61 0 T1 3 80 1
3 90 4
4 00 1
4 10 1
680
Jumper on T1: Variant
1020 4 20 1
4 30 1
4 40 1
660 600 4 50 1
4 60 1
K
575 K 4 70 1
4 80 1
I 650 4 90 6
44 0
4 95 6
41 0
5 00 1
410 42 0 48 0 49 0 Antisize grease
(Parting Lubricant 785 FG) 5 10 6
5 20 1
Smøres ind i high
vacuum grease 5 30 1
5 40 1
640 5 60 1
L
5 70 1
5 75 1
5 90 1
5 95 2
6 00 1
45 0 430 42 0 O 2013-06-11 H4: Pos. 700 new item. JJO 6 10 1
A 14.12.2006 BMJ 6 20 2
Revision Crea. date Revision description Draw. Init Appr. date Appr. Init 6 30 2
yydydy--m
mmm--ydyd yyd
yyd-m m-dyd
y
Material: Scale: Format: Tolerance: DS/ISO 2768- mK
6 40 1
SECTION D-D
M
40 595 590 SCALE 1 : 1.5 A 2:1 A1 Surface treat.: None 6 50 1
P ed e rs t ru p ve j 8 4
ID: Description: Rev : 6 60 2
DK -2 7 50 B a lle ru p/C o pe n ha ge n
6 70 2
1 2 3 4 5 6 7 8 9 10 11 12
De nm a rk
P h o n e :+ 4 5 4 4 6 0 0 8 0 0
Fa x: + 4 5 4 4 6 0 0 8 0 4
15880001 AbraPol-20, Complete O
6 90 1
7 00 1
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
A A
1 20
41 0
D ETAIL A
S C A LE 2 : 5
1 10
3 30 l = 5 5m m
X120 90
1 40
B 60 B
1 50
1 30
10
1 60
3 30 l = 330 m m
1 70 5 00
A
1 80
C C
49 0
42 0
D D
70 43 0
10 0
E E
31 0
23 0
22 0
190
F F
27 0
2 00
2 10
POS. NO. AMOUNT DR AW . N O.
10 1 15880007 C asing w
G G
24 0 20 1 15 89 05 7 6 L ef t s upp
30 9 2 Z A 1 0 0 0 6 S k iv e 6 D
40 2 2T R 5 06 12 M C S k r u
50 1 15 880 03 4 W . P C B a
2 50 60 5 2T R 50 5 10 M C s k r u
70 8 2 Z A 1 0 0 0 5 S k iv e M 5
90 1 15881200P late for m
10 0 2 2T R 50 5 12 M C s k r u
44 0
11 0 1 2T I 10 82 0 M S P s k r u
H
30 0 H 12 0 1 2T A20080 K ontram
13 0 2 2Z I20503 C rinkle wa
14 0 1 2X M 1 15 27 C O N N E
20 30 40 15 0 2 2Z I20302 C rinkle wa
16 0 2 2 T A 1 0 0 3 0 M o t r ik M
17 0 2 2T X 90 30 8 U N C s k r
18 0 1 2 X M 9 0 0 0 9 H a e t te fo
70 60
19 0 1 1 5 8 8 0 0 1 6 P la t e f o r
20 0 4 2Z I20604 C rinkle wa
XPE16 21 0 4 2T R 80 6 12 U LS S k r
J J
22 0 1 2g k20045 U niversa
39 0 40 0 23 0 1 2T D20109 MS -kont
D ET AI L E
S C A LE 1 : 2 24 0 1 7 7 2 0 5 6 5 3 P a a tr y k t e
25 0 1 10160180 Fares kilt,
50
26 0 7 2T A4 00 60 L A AS E M
27 0 2 2G K 8 20 10 T Y- R A P
28 0 1 2 N S 4 1 9 1 2 C la m p p
l = 40m m 29 0 1 2G X 20004 K lem ryk
29 0
30 0 2 2T X 90 40 5 U N C s k r
K K 31 0 1 2G K 90 1 47 N e op r e n
32 0 1 15880082 C ontrol b
60 46 0 33 0 2 2ip10151 T aetning s
33 0 2 2ip10151 T aetning s
34 0 4 2 Z A 2 0 0 0 5 S k iv e 5 D
2 80 2 60 c. 2 0 c . 20 35 0 4 2T A 10050 Moetrik M
4 50 47 0 36 0 2 15890810 H olding w
37 0 2 2T R 80 5 10 U L S M 5x
P O S. 470 S T I C K T O P O S . 4 50 39 0 4 2 Z I4 0 4 0 5 S K I V E M .
40 0 2 2T A4 00 40 Laa s e m o
L
41 0 1 10160129 A dvarsels
34 0 3 70 42 0 1 10 160 18 1 F C C R U L
3 20 43 0 1 772 06 00 2 PA A T R Y
30 260 44 0 1 101 61 02 4 T a g - Ab r
36 0
45 0 1 2MO90051 Noise filte
E 2012-05-07 Pos.500 tilføjet
B A4: Variant JTV 2012-05-07 JTV 46 0 1 2G K 31112 F ixing s c
A 08-12-2006 BMJ
See AbraPol-20, complete Rev ision Crea. date Rev ision description Draw. Init Appr. date Appr. Init
47 0 2 2ip10152 T aetning li
Se AbraPol-20, komplet 48 0 1 1 5 8 8 0 1 0 0 P la t e fo r
3 50 d d-mm-y y dd -mm- yy
1 2 3 4 5 6 7 8 9 10 11 12
Denmark
Phone:+45 44 600 800
Fax: +45 44 600 804
15880005 Casing with electrical, assembly E
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
A
30 0
20 A A
D E T A IL D
S CA LE 2 : 5
10
240
M1
B B
28 0 29 0 2 60 250 270
D
XPE14
C C
31 0 32 0
E E
230
D D
B1
22 0
XPE12
3 10 33 0
21 0
20 0
E E
15 0
160
170
F F
M2
G G
A
S E CT I O N A -A
S CA LE 1 : 5
12 0 11 0 1 00 1 00 11 0 1 20 13 0 140
H H
G
90 80 70 40 30 60 50
G
J J
X22 X23
K K
S EC T IO N G - G
S CA LE 1 : 2
L
1 2 3 4 5 6 7 8 9 10 11 12
Denm ark
P hone:+4 5 44 600 800
F ax: +4 5 44 600 8 04
15880007 Casing with motors, assembly C
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
A
XPE19 XPE17 A
XPE7
1 90 1 80
B
1 80 1 90
A
XPE15
13 0 1 40 1 50 1 60
10 0 11 0 12 0 3 60
B B
3 30
27 0 2 60
l = 507
2 80 3 10
l = 12 0
l = 510 3 10 280
4 30 44 0 2 80 3 10
C C
26 0 27 0
l = 5 30
31 0 28 0 XPE23
4 50
1 80 1 90
D D
90 80 60 1 70 50 75 80
40
QS1
See wire set
70
X116 l = 625 70
See wire set
30 0 29 0 28 0
E E
XPE11
190 18 0 24 0 2 30 1 90
40
l = 750
XPE13 YS1
31 0 2 80
1 90 1 80 2 00 2 20 25 0
F
See wire set. F
S C A LE 1 : 5 SCALE 1 : 5
60
1 70 4 2T R 10412 C H R sk rue
A 1 80 20 2 Z I4 0 4 0 5 S K IV E M . U D
1 90 14 2 T A 4 0 0 4 0 L a a s e m o tr ik
20 30
2 00 1 15 8 9 0 50 0 S t a nd
2 10 1 2S S 000 19 S C H M E R SA
2 20 2 2Z I20403 C rink le wash
3 20 2 30 4 2Z A 10004 Sk iv e 4 D IN
2 40 1 1 5 8 9 0 5 0 1 C o v e r f o r lo c
J J 2 50 2 2T R 10435 C H R S k rue
3 40 l = 80 2 60 2 15891190 G uide lis t I
2 70 8 2T R 7 05 12 M K s k r ue M
2 80 1 2g o80025 Ledning s kan
3 40 l = 80
2 80 1 2g o80025 Ledning s kan
2 80 1 2g o80025 Ledning s kan
2 80 1 2g o80025 Ledning s kan
34 0 l = 320 2 80 1 2g o80025 Ledning s kan
2 80 1 2g o80025 Ledning s kan
2 90 7 2 Z A 1 0 0 0 5 S k iv e M 5 A 2
K K
34 0 l = 320 3 00 3 2T R 8 05 08 U L S M 5x 8 A
3 10 15 2Z N 03054 P op-nite, alu
3 30 1 2G K90459 N eoprentylle
3 40 2 2ip10152 T aetninglis te
3 40 2 2ip10152 T aetninglis te
3 50 1 15890587 H older for s h
3 60 4 2T A 400 60 LA A SE M O E
1 80 19 0 XPE4
3 70 1 1 5 8 8 0 0 9 0 V a lv e s f o r c o
3 80 4 2 T A 4 0 0 5 0 L a a s e m o tr ik
L 3 90 1 2nu29316 G rifl-S lange
1 80 1 90 4 00 2 2N S 21 220 SN E K KE B A
XPE5 4 10 1 2N F 60000 K obling m . n
30 P os 30 to be adj usted 60 0/+ 5m m from the floor to the edg e of the cabinet (see s ection A-A ).
4 20 1 2 g k 8 0 0 1 0 T y r a p 2 .5 x 1
4 30 1 2Y H 0 00 10 S pe ed C ont
I 2013-06-14 F7: Pos. 210 remov ed. Text added. 4 50 2 2N F 10083 Q uic k c oupl
JJO
A 06-12-2006 BMJ
4 24 2 2N U 12445 N ylonroer, s
Rev ision Crea. date Rev ision description Draw. Init Appr. date Appr. Init 4 51 1 2Y H 00 0 11 N ut M 1 6x 1 +
yy yy- mm-d d yy yy- mm-d d
4 52 1 2S E 20317 Main S w itc h
M aterial: S cale: Format: Tolerance: DS/I SO 2768- mK
M 2:1 A1 Surface treat.: None
Pederstrupvej 84
ID: Description: R ev:
DK-2750 Ballerup/Copenhagen
1 2 3 4 5 6 7 8 9 10 11 12
Denmark
Phone:+45 44 600 800
Fax: +45 44 600 804
15880010 Casing, assembly I
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
A A
24 0
27 5
28 0
3 00
B B
3 20
3 40
33 0 2 70 2 60 2 50
C C
3 10
2 90
D D
20
3 50
4 40 450
SE CT I O N B- B
S C AL E 1 : 2 B B
30
E E
4 20 43 0 40
10
F F
22 0
1 90
G 360 G
P O S . N O . A MO U N T D RAW . NO .
10 1 1 589 00 23 B a s e f or m e c
A A 20 1 15880360 G round plate
2 30 30 4 2ZI61022 F eders kiv e 10
40 4 2T T 51030 MC sk rue M1
H H
50 2 15090380 S oej le
60 8 2 Z C 1 3 0 4 0 F a c e t s k iv e 3 0
70 4 2 T F 0 3 0 1 5 M o e tr ik f . a k s
80 1 2yc 50101 C ylinder C D Q
10 0 2 2IF 00012 Pakning , P V C
11 0 2 2 N F 4 0 0 4 1 N ip p e l 2 5 3 1
12 0 7 2IF 00011 Pakning , P V C
13 0 4 2N F 10034 Lynkobl. banj
14 0 2 2N F 20032 B anj o-hulsk ru
15 0 4 2ZI60616 F edersk iv e 6 D
J 70 3 80 J
16 0 4 2 T T 5 0 6 9 4 M C s k ru e M 6
17 0 1 2JH 50010 K obling flexo
18 0 1 2 T I 51 03 5 M SP s k r ue m .
19 0 1 15099006 G lidefoering,
60 2 70
20 0 4 2 B F 2 0 0 4 0 K u g le b o e s n in
21 0 4 2II04052 IN A-T aetnings
22 0 1 1 5 0 9 0 3 5 0 O e v r e s ty r
39 0
23 0 1 2G S 10315 B uffer A A -31
60 24 0 1 15090040 Lej eopbyg nin
21 0 20 0 2 00 2 10 25 0 4 2T T 51035 MC sk rue M1
K K
26 0 4 2Z I21005 C rink le was he
27 0 6 2 Z A 1 0 0 1 0 S k iv e 1 0 1 2 5
27 5 1 2IP 00601 T aetning slis te
1 80 17 0 15 0 16 0 90 B1
40 0 41 0 3 90 50 70 60 28 0 1 14590017 D istance ring
+2 27 0 37 0
10 0 80 29 0 1 2JE 00315 K ilerem sk ive
30 0 1 2JE 92035 K lem boes ning
31 0 1 2 Z F1 1 22 8 F e de r A 12 x8
32 0 1 14590018 D is c dia12.5_
33 0 1 2ZI61225 F eders kiv e 12
L
34 0 1 2T T 51235 MC sk rue M1
35 0 2 2JD 01320 Sm alk ilerem
1 00 36 0 1 1 58 80 02 1 D o w n m o t or ,
71
5 49, 5
A A
70 80 90
D ET AI L A
SCALE 1 : 2
B XPE29 B
536
14 0 150
A
C C
PO S. N O . AM O
10
20
10 20 40 30
1 60 40 30 30
40
50
D D 55
60
70
A8 80
90
1 00 11 0 120 1 30 100
110
110
130
130
E
140
ISOMETRIC VIEW 150
B 160
D ET AI L B
SCALE 1 : 2
60 C 2009.03.03 Pos.55 added SPE 2009.03.03
A 26.01.2006 JF 26.01.2006 FPG
Revision Crea. date Revision description Draw. Init Appr. date Appr. Init
yyyy-mm-dd yyyy-mm-dd
Material: Scale: Format:
50 55 Tolerance: DS/ISO 2768- mK
F
1:5 A3 Surface treat.: None
ID: Description: Rev:
4 65 Pederstrupvej 84
DK-2750 Ballerup/Copenhagen
Denmark
1 2 3 4
Phone:+45 44 600 800
Fax: +45 44 600 804 15880050 Box for tub, assembly C
1 2 3 4 5 6 7 8
A A
B B
SECTION A-A
C C
10
60
D A A D
20
30
70
40
D 29.10.2008 Pos.70 15880242 added; See 1588LD01 list removed SPE 29.10.2008 BRY
A 04.12.2006 SPE
Revision Crea. date Revision description Draw. Init Appr. date Appr. Init
dd-mm-yy dd-mm-yy
Material: Scale: Format: Tolerance: DS/ISO 2768- mK
F
1:5 A3 Surface treat.: None
Pederstrupvej 84
ID: Description: Rev:
DK-2750 Ballerup/Copenhagen
Denmark
1 2 3 4
Phone:+45 44 600 800
Fax: +45 44 600 804 15880058 Tub, assembled D
1 2 3 4 5 6 7 8
A A
A A
B B
C C
20 30 10 50 80
40 110
60
D
120 D
90
130
A 200 A
2 90 280 330 2 70 300 3 10 250 22 0 230 2 40 2 10 140 260 170 1 90
L=550mm
Fra forsyining
150
A
B
180 B
1 70
130
80 XP E 1 8
150
Y3 1 20
Y4
Y5 160
C
60 70 A2 1 10
C
L=200mm
140
XPE20
110 120
D
X2 1 D
L=1020mm
Til bunden af cylinder
10
A
L=900mm
Til toppen af cylinder SECTION A-A
140
60 70 3 20
340 l = 30mm
X21: See wire set 90 1 00
C
B 12.09.2007 Pos, 140 have changed place with 180. SPE 12.09.2007
A 30-11-2006 YKJ
H
Revision Crea. date Revision description Draw. Init Appr. date Appr. Init
dd-mm-yy dd-mm-yy
Material: Scale: Format: Tolerance: DS/ISO 2768- mK
1:2 A2 Surface treat.: None
Pederstrupvej 84
ID: Description: Rev:
DK-2750 Ballerup/Copenhagen
Denmark
1 2 3 4 5 6 7 8
Phone:+45 44 600 800
Fax: +45 44 600 804
15880034 W. PCB and pneu. distr., assembly B
1 2 3 4 5 6 7 8
10 70
A A
190 220
50
240 POS. NO. AMOUNT DRA
210 10 1 158
40
25 20 1 158
A5
25 1 154
30
70 30 1 2HD
B B
40 4 2TN
XPE22 180
230 A1 50 4 2ZE
XPE21
230 70 70 12 2TA
80 1 2HR
200
BR1 90 1 2TD
100 1 2IO
190 80
110 1 150
135 135
120 1 2TI1
C
90 130 C 130 2 2SA
135 2 2SA
SB2
SB1
1 2 3 4
Phone:+45 44 600 800
Fax: +45 44 600 804 15880082 Control box AbraPol-20, assembly I
1 2 3 4 5 6 7 8
XPE6
A A
20 30 40 50 60 10 20
B B
C C
D D
C , , , , ,
B 05.11.2007 Pos. 20 15881195 changed SPE 05.11.2007
A 27.11.2006 JF 27.11.2006 FPG
Revision Crea. date Revision description Draw. Init Appr. date Appr. Init
dd-mm-yy dd-mm-yy
Material: Scale: Format: Tolerance: DS/ISO 2768-
F
1:4 A3 Surface treat.:
Pederstrupvej 84
ID: Description: Rev:
B
DK-2750 Ballerup/Copenhagen
Denmark
1 2 3 4
Phone:+45 44 600 800
Fax: +45 44 600 804 15880025 Front, assembly
1 2 3 4 5 6 7 8
6
5 210 220 230
4 3
B
10 170 180 240 B
2
1
A 10 A
100
40 30
B B
110
150 90 80 140
C C
D 120 D
130
60 20
POS.
E
40 Clamping torque: 2Nm.
Use Loctite 243/Omnifit 100M.
1 2 3 4
Phone:+45 44 600 800
Fax: +45 44 600 804 15890080 Safety guard, assembly D
1 2 3 4 5 6 7 8
A A
B B
120
90
95
30
110
(Gul/Grøn)
L=0.3m 100
30
Extension modul
Doser udvidelsesmodul
30
D 90 30 D
95
20 L=0.5m
50 70
1 2 3 4
Phone:+45 44 600 800
Fax: +45 44 600 804 15880095 Tap water valve D
Page 1 of 1
file://\\Dkpdm1\PDMRelease\100\equi\E-FILE-01096_0001.DWG 03-03-2008
1 2 3 4 5 6 7 8
A
A A
70
10
B B
20
80
C
A C
30
60
40
D D
65
50
E 10 20 Let prespasning
SECTION A-A
1 2 3 4
Phone:+45 44 600 800
Fax: +45 44 600 804 15090009 Opspaendingsenhed, monteret C
Page 1 of 1
file://\\Dkpdm1\PDMRelease\100\equi\E-FILE-01097_0001.DWG 03-03-2008
1 2 3 4
40
50
A
90
200
100
190
130
120
90
110
150
½ turn open
C åbnes ½ omgang
210
145
160
140 110
170
1 20 90 160
A
18 0 40 1 00 50 110 60 120 30 150 14 0 16 0 1 70 A
10
XPE30
XPE28
30
B B
20
C C
D D
M3(pump1) M4(pump2) M5(pump3)
E E
F F
POS . N
1
1
JTV 2009-12-06 JTV
1
F 2009-12-06 Pos 180 changed to reduction angle.
BMJ
1
A 28-11-2006
H
Revision Crea. date Revision description Draw. Init Appr. date Appr. Init 1
dd-mm -yy dd-mm -yy 1
Material: Scale: Format: Tolerance: DS/ISO 2768- mK 1
1:2 A2 Surface treat.: None
ID: Description: Rev:
Pederstrupvej 84
DK-2750 Ballerup/Copenhagen 1
1 2 3 4 5 6 7 8
Denmark
Phone:+45 44 600 800
Fax: +45 44 600 804
15880051 Pumps module, assembly F
1 2 3 4 5 6 7 8
A A
50
B To Doser arm B
40
20
POS
30
C C
10
D D
From Recirkulation
1 2 3 4
Phone:+45 44 600 800
Fax: +45 44 600 804 15880090 Valves for cooli water C
5 4 3 2 1
CONNECTION FOR 200V / 50Hz CONNECTION FOR 200-210V / 60Hz CONNECTION FOR 220-230V / 50Hz
D BN BN BN D
YE YE YE
RD RD RD
RD WH RD WH RD WH
VT VT VT
GN GN GN
YE YE YE
OG OG OG
BU BU BU
CONNECTION FOR 220-240V / 60Hz CONNECTION FOR 380-415V / 50Hz CONNECTION FOR 380-415V / 60Hz
C C
BN BN BN
YE YE YE
RD RD RD
RD WH RD WH RD WH
VT VT VT
GN GN GN
YE YE YE
OG OG OG
BU BU BU
B B
COLOR CODES:
BK = BLACK
BN = BROWN
RD = RED
BN BN OG = ORANGE
YE = YELLOW
GN = GREEN
YE YE BU = BLUE
RD RD VT = VIOLET
GY = GREY
RD WH RD WH WH = WHITE
VT VT STRUERS A/S
GN GN VALHOEJS ALLE 1176
YE YE Rev. A: Baan PDM
DK-2610 ROEDOVRE
A OG OG DENMARK
PHONE: + 45 3670 3500
A
BU BU Transformer connections.
5 4 3 2 1
5 4 3 2 1
D D
Primær: Sekundær:
0V
BLÅ
P1
200V 24V / 4A
GRØN RØD
C P2 S1 C
GUL HVID
230V 0V
0V 24V / 4A
ORANGE VIOLET
P3 S2
GUL
0V
200V
RØD
B P4 B
BRUN
= intern forbindelse
230V
STRUERS A/S
Rev. A: Baan PDM VALHOEJS ALLE 1176
DK-2610 ROEDOVRE
DENMARK
PHONE: + 45 3670 3500
Y3 Y4 Y5
P A P A P A
R R R
P2
BP1
BP2
P1 6 .0 B a r
Y1
A 3 R
2
P
1
3.4 Bar
R A
P B
S
S k if t e v e n t il
SU3
SA2
Drain
SU1
W ater supply:
Recirculation
system
SU2 SA 1
(Accessories)
Y1 W 15
GY
BN
Water to
OP-S
PK YE
Y1
E
SA1 Throttle-valve (Assy:15880095, part: 2YH00008) for regulation of inflow water for disc cooling (Assy:15880058)
SA2 Throttle-valve (Assy:15880090, part: 2YH10602) for regulation of inflow water for water pistol (Assy15880055)
SU1 Three-way electromagnetic valve (Assy: 15880095, part: 2YH12311)
SU2 Three-way spherical valve (Assy: 15880090, part: 2YH03629)
SU3 Three-way air actuator driven valve: 15880080
QS1 T1
L10
L1 F1-3 3 3 F4-5
L2 3 2 2
U1: 24Vac, 4A
L3 3 x FUSE 3 2 x FUSE
Mains choke
2 F6 2
D Z1
1 x FUSE
U2: 24Vac, 4A D
PE RFI FILTER
3 3 IEC742 SAFETY
PE TRANSFORMER
4 4
TO CPU TO CPU
BOARD, A1 A3 BOARD, A1 A4
SAFE STANDSTILL SAFE STANDSTILL
FREQUENCY FREQUENCY
3 INVERTER 3kW 3 INVERTER 3kW
TO CPU TO CPU
BOARD, A1 BOARD, A1
CAN BUS CAN BUS
3 3
C C
M1 M2
LED LAMP
3 x SOLENOID
DOOR VALVE
INVERTERS INVERTERS COOLI
QUARD tap water
A10 (SAFE (CAN BUS) (REC. SYSTEM)
STANDSTILL) SWITCH disc cooling
water to OP-S
Display
2
SUPPLY IN: 2
U1 (24Vac, 4A)
22
16
6 8 8 3
Inputs: Outputs:
Sample mover top position
Removal measurement unit installed 4 x 24VDC air valve:
START SERVICE
Absolute pressure buttom Sample mover up Incremental STOP Keyboard INTERFACE
Absolute pressure top Ait outlet 1 dial buttom EMRG. STOP
Ait outlet 2 (RS232)
Voltage supply 1 (input)
Pump extension board output Sample mover down
Voltage supply 2 (+15V) 6 x 24VDC pupm motors
Voltage supply 3 (+24V) 1 x 24VAC Shifting valve recirk. / drain
Removal measurement
Dosingunit installed
A STRUERS A/S
A
REV A: OH (2006/10/23) Pederstrupvej 84
REV B: OH (2007/03/29) DK-2750 Ballerup
- L10 added Denmark
REV C: OH (2007/07/27)
- 2nd PE terminal added
REV D: OHO (2008/11/20) ABRAPOL-20, BLOCK DIAGRAM
- Z1 added
REV E: TKO (2009-05-11) - A10 added Size CAGE Code DWG NO Rev
A2
15883050 E
Monday, May 11, 2009 Scale Sheet
OH 1 of 1
5 4 3 2 1
5
SEE REQUIREMENTS FOR MAINS CABLE (EU or UL/CSA)
4 3 2 1
- note #1
MUST BE EXTERNALLY FUSED
Fmax: See note #2
QS1
CONNECTION TO SAFETY
L10 AWG12 TRANSFORMER
W1 (EU) AWG10 AWG10 AWG12
L1.1 BK7
X25
BK15 2MT72034
BN 1 BK1
L1 T1
BK4 F1 BK12 L1
L1
BK8 AWG14
L2 T2 F2 L2 L2.1 F4
D L2 BK 3 BK2 BK5 BK13
BK8
BK10 RD1 U1: 24Vac TO X1-1 D
L3 T3 F3 L3 L3.1 BK9 F5 S2
L3 BK or GY 5 BK3 BK6 BK14
BK9
BK11 RD2 U1: 0Vac TO X1-2
BU (not used) F6
X1 MAINS SWITCH RD3 RD7 U2: 24Vac
GNYE19 GNYE20 TO X21-1
XPE1 AWG12 AWG12 S1
PE RD4 U2: 0Vac
PE GNYE GNYE1 GNYE23 XPE31 XPE32 AWG12
TO X21-2
AWG10 AWG12
4(5) x 4mm² POWER
INPUT MAIN T1 200VA
PLATE MAIN PLATE MAIN PLATE
BOX IEC742 SAFETY
TRANSFORMER
XPE35
GNYE
PE GNYE22 GNYE16 GNYE23
AWG12
BK15
BK8
BK9
4mm² POWER AWG10 AWG12
INPUT MAIN
BOX PLATE
E L1 L2 L3 A3
OR LINE
M1
LO
LO
K32
K31
39
28
20
HI
HI
34
33
7
7
Z1 RFI Suppression Filter W4
QS1 LOAD U1 U
E L1' L2' L3'
X3 X3.1 U
C W1 (UL/CSA) W26 V V2 V Motor
C
BK 1 BK18 X2.1 W 3 Y/D
BK16
BK17
BK18
L1 L1 W W3
550W
GNYE24
BK17
RD 3 L2 GNYE PE
L2 PE
FILTER BK16
L3 X1.1 +t
OG 5 FRAME
L3 PE
AWG12 BR1
WH (not used)
X1 T1
MAINS SWITCH X2.2 BR2
GNYE3
T2
GN (GNYE)
PE AWG12 Screen
PE
4(5) x AWG 10 Frequency Inverter - 3KW TOPSERV 113
Sample mover motor
W26 4 x 1,5 mm2 +
GNYE15 3 x 230Vac - DELTA
XPE2 AWG12 INVERTER 2 x 1 mm2 + screen connected 3 x 400Vac - STAR
screen
BK16
BK17
BK18
GN (GNYE)
PE MAIN
XPE25 FRAME to cable gland
AWG10 PLATE
MAIN PLATE
A4
IMPORTANT!!! M2
For AbraPol-20 connected to electrical installations
LO
LO
K32
K31
39
28
20
HI
HI
34
33
7
7
W9
with residual current circuit breakers, a residual U1 U
X3 X3.1 U
B current circuit breaker type B OR type B time GNYE2
AWG12 V Motor B
MAIN V2
delayed, 300 mA is REQUIRED. Alternatively, COLOR CODE: PLATE
V
BK18 X2.1 W 3 Y/D
another protective measure shall be applied ------------------------ L1 W W3
BK = black 2,2kW
(isolation by a transformer). AbraPol-20 MUST be BN = brown
BK17
L2
PE GNYE PE
hard-wired into the mains. RD = red BK16
L3 X1.1 +t
OG = orange
YE = yellow PE
The second Protective Earth must be earthed by GN = green T1 BR1
X2.2 BR2
connection to a terminal marked or PE, BU = blue
VT = violet
T2
alternatively, use a system whereby there is an GY = grey Screen
F1+F2+F3 F4+F5
VOTAGE / FREQ. VOTAGE / FREQ. note #1 note #2 (fuse size) (fuse size) F6 M1+M2
(from nameplate) (depending on country) W1 - mains cable max. ext. fuse (fuse size) Connection
See note #3 See note #3
3 x 200V / 50Hz 4mm^2 3 x 50AT 3 x 20AT (CC) 2AT (CC) 4AT DELTA
A 3 x 200-210V 50/60Hz
3 x 200-210V / 60Hz AWG = 10 3 x 50AT 3 x 20AT (CC) 2AT (CC) 4AT DELTA A
REV E: OHO (2008/11/20) REV A: OH (2006/10/08) STRUERS A/S
- Z1, GNYE23, GNYE24, W26 added Pederstrupvej 84
3 x 220-230V / 50Hz 4mm^2 3 x 50AT 3 x 20AT (CC) 2AT (CC) 4AT DELTA - BK7,BK8, BK9 for con. L10-A3 removed
REV B: OH (2007/01/24)
DK-2750 Ballerup
- max. value for ext. fuse changed
3 x 220-240V 50/60Hz - connection wiht BK16, BK17, BK18 added - fuses F1-F5 changed Denmark
3 x 220-240V / 60Hz AWG = 10 3 x 50AT 3 x 20AT (CC) 2AT (CC) 4AT DELTA - BK15, BK8, BK9 reconnected REV C: OH (2007/03/29)
- variants of mains cable added
- L10, X25 added --> changed connections ABRAPOL-20, CIRCUIT DIAGRAM - MAIN VOLTAGE
3 x 380-415V / 50Hz 4mm^2 3 x 50AT 3 x 16AT (aM) 1AT (aM) 4AT STAR
3 x 380-415V 50/60Hz Note #3: - BK12, BK13, BK14, BK15 added
- XPE31,32 and GNYE19,20 added
3 x 380-415V / 60Hz AWG = 10 3 x 50AT 3 x 16AT (aM) 1AT (aM) 4AT STAR F1, F2, F3, F4, F5 are time delay fuses REV D: OH (2007/07/27) Size CAGE Code DWG NO Rev
- PE, XPE35 and GNYE22 added
3 x 460-480V 60Hz
CC...Class-CC character istic - XPE26 removed, GNYE16 reconnected
A2
15883100 E
3 x 460-480V / 60Hz AWG = 10 3 x 50AT 3 x 15AT (CC) 1AT (CC) 4AT STAR aM...aM characteristic Scale Sheet
Wednesday, February 11, 2009 OHO 1 of 1
5 4 3 2 1
1
A1
12
11
10
2
1
1
INHIBIT
INHIBIT
GND-S
GND
GND
GND
GND
GND
GND
GND
RX/TX
RX/TX
Output (24V
Output (24V
Output (24V
(NAMUR)
(NAMUR)
(NAMUR)
-
+
connector
Disp.
Input
Input
Input
FPC
14 pole
DC)
DC)
DC)
X105
9 pole D-sub female 9 pole D-sub male 6 pole 3.5 mm connector 12 pole Molex Minifit jr.
StruersNet+Cooli Service interface RS-485 interface 3x24V output. 3xNAMUR inputs
4
X115 X117 X119 X125
IL-G
4 pole
3
Backlight
X106
2
1
10
9 1 BN
24V AC BK
jr.
10 pole Molex Minifit
3 pole Terminal
4
TO X21
24V AC POWER
X127
input
W1
Discrete keys
2 BK BN
24V AC
8
X114
3
Block
3 GND-S SCREEN
INPUT
2 x AWG18 + shield
7
CPU - BOARD, A1
8
7
knop
Incremental
jr.
8 pole Molex Minifit
3 1
X113
GND
6
dial
2
2
5
LIN
1
3
24V unprotected
5 pole Terminal
6
X118
GND
Bus
LIN
Microfit
6 pole Molex
3
Y Protected by : 4
Block
5 emergency stop sw.
Joystick
X111
2
X 24V protected by : 5
4 emergency stop sw.
Z
1
Interlock switch
5V YS1 Door Guard Switch
A A
X101 X107 X108 X121 X110 X109 DOOR DOOR DOOR
3 pole Terminal Block 20 pole Molex Minifit jr. 6 pole Molex Minifit jr. 2 pole Terminal Block 2 pole Molex Minifit jr. 5 pole Molex Minifit jr.
SWITCH SWITCH LOCK
SIL
8 pole
24V DC POWER INPUT Contactor control. 24V DC outputs for contactors. Supervision input from contactors Interlock Motor temp. Motor current Load Cell
Keyboard
X112
K1 - Output (24V
K2 - Output (24V
K3 - Output (24V
K4 - Output (24V
K5 - Output (24V
Power indicator
Green
Monitor
Monitor
Monitor
Monitor
Monitor
GND-S
K1 -
K2 -
K3 -
K4 -
K5 -
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
DC)
DC)
DC)
DC)
DC)
GND
24V
12
11
22
32
21
A2
A1
20
10
3
19
18
17
16
15
14
13
12
11
9
BK1
BK5
BK4
BK2
BK3
BK6
RD
OG
BK
BN
RD2
RD1
BK3
BK6
BK2
BK5
BK1
BK4
BU5
BU6
W10
cable 7 x 0,75mm2
RD2
BU6
RD1
W7 2 x 0,75mm2
X20
1
W6 2 x 0,75mm2
W2
RD1
RD2
BU5
4 x AWG18
RD
RD
RD
BK
BK
BK
BN
BN
BN
W3 W3 W12
TO X3(A4) TO X3(A3) TO X115(A1)
W5
2 x 0,75mm2 INVERTER
2 x STRAP 0,5mm2 INVERTER 2 x STRAP 0,5mm2 FRAME COLOR CODE:
FRAME ------------------------
R1 BK = black
120R BN = brown
RD = red
RD2
RD1
RD2
RD1
BU1
BU2
BU3
BU4
RD
RD
RD
RD
OG = orange
OG
BK
BK
BK
BK
BN
BN
BN
BN
A3 A4
RD1
RD2
YE = yellow
GN = green
BU = blue
1 2
LO
LO
LO
LO
K32
K31
K32
K31
39
28
20
39
28
20
HI
HI
HI
HI
VT = violet
34
33
34
33
7
GY = grey
WH = white
X3 X3.1 U X3 X3.1 U
PK = pink
V V BE = beige
1
5 4 3 2 1
A - RS485
Connector
TO X3(A4) X120
RS485
for service
X116 Y1
B-
PX0412/04S
RD
BK
BN
Connector
10
YE
9
Connector for 9 -pin MALE X3 +24VDC
Tap water
1
Cooli GN
5
GND
BK
BN
WH
BN
BU
D +24VDC D
PK Disc Cooling
GND
W12 GY
+24VDC
3 x AWG20 BN Water to OP-S
W11 W14 GND
+ shield W13
2 x 0,5mm2 2 x AWG18 + shield
3 x AWG20 W15 3 x solenoid valve
+ shield
6 x 0,5mm2
BN
RD
SCREEN
BK
WH
SCREEN
SCREEN
GY
GN
BK
PK
YE
BN
BN
BN
BU
RD
BK
BN
A1
A5
12
11
10
6
1
1
INHIBIT
INHIBIT
GND-S
GND
GND
GND
GND
GND
GND
GND
RX/TX
RX/TX
24V RECIRC
CAN-HIGH
GND
CAN-LOW
LCD Display, 320x240 dots
Output (24V
Output (24V
Output (24V
(NAMUR)
(NAMUR)
(NAMUR)
-
+
14
connector
Disp.
Input
Input
Input
FPC
14 pole
DC)
DC)
DC)
X105
9 pole D-sub female 9 pole D-sub male 6 pole 3.5 mm connector 12 pole Molex Minifit jr.
StruersNet+Cooli Service interface RS-485 interface 3x24V output. 3xNAMUR inputs
4
X115 X117 X119 X125
IL-G
4 pole
3
Backlight
X106
2
1
C
4
W16 10
C
BN BN
BK 5
START SB1 RD 9
24V AC
1
jr.
10 pole Molex Minifit
3 pole Terminal
4
24V AC POWER
3 BK OG
X127
input
Discrete keys
2
24V AC
4 x AWG18 8
X114
4 RD + shield 3
Block
3 GND-S
INPUT
W17
STOP SB2 BN 7
BK 2
3 OG
RD 6
1 BN 1
OG
4 x AWG18
SB3-1 + shield 8 CPU BOARD, A1
2 BK BR1 W18
4
EMERGENCY 7
Incremental dial knop
knop
Incremental
jr.
8 pole Molex Minifit
STOP 1 RD Vcc 6 GY
BN 3 1
X113
CH.A 5
GND
SB3-2 CH.B 4 WH 6
dial
2 OG 3 PK 2
2
PUSH LIN
BUTTON2 YE 5
1 GN 1
GND
3
B cable 6 x 0,5mm2 24V unprotected B
5 pole Terminal
6
X118
GND
Bus
LIN
Microfit
6 pole Molex
3
Y Protected by : 4
Block
5 emergency stop sw.
Joystick
X111
2
X 24V protected by : 5
4 emergency stop sw.
Z
1
Interlock switch
5V
A6
X101 X107 X108 X121 X110 X109
Keyboard 8 3 pole Terminal Block 20 pole Molex Minifit jr. 6 pole Molex Minifit jr. 2 pole Terminal Block 2 pole Molex Minifit jr. 5 pole Molex Minifit jr.
SIL
8 pole
Keyboard
24V DC POWER INPUT Contactor control. 24V DC outputs for contactors. Supervision input from contactors Interlock Motor temp. Motor current Load Cell
X112
K1 - Output (24V
K2 - Output (24V
K3 - Output (24V
K4 - Output (24V
K5 - Output (24V
Power indicator
Green
Monitor
Monitor
Monitor
Monitor
Monitor
GND-S
DC)
K1 -
K2 -
K3 -
K4 -
K5 -
GND
GND
GND
GND
GND
GND
GND
GND
GND
GND
DC)
DC)
DC)
DC)
GND
24V
COLOR CODE:
------------------------
20
10
3
19
18
17
16
15
14
13
12
11
9
1
BK = black
BN = brown
RD = red
OG = orange
YE = yellow
GN = green
BU = blue
VT = violet
GY = grey
A WH = white A
PK = pink
BE = beige
RO = rose
GNYE = yellow-green
5 4 3 2 1
5 4 3 2 1
D D
MAIN MAIN MAIN MAIN MAIN MAIN MAIN MAIN MAIN FRONT
FRAME FRAME FRAME FRAME FRAME FRAME FRAME FRAME FRAME PANEL
XPE2
MAIN XPE35 MAIN MAIN GNYE4 GNYE5 GNYE6 GNYE7 GNYE8 GNYE9 GNYE10 GNYE11 GNYE12 GNYE13
PLATE PLATE PLATE AWG10 AWG14 AWG14 AWG10 AWG14 AWG14 AWG14 AWG14 AWG14 AWG14
XPE3 XPE6 XPE8 XPE10 XPE12 XPE14 XPE16 XPE18 XPE20 XPE22
GNYE1 GNYE22 GNYE2
AWG10 AWG10 AWG12
C C
POWER POWER MAIN FRONT TOP POWER INPUT BOX FOR BASE FRAME LEFT SUPPORT PLATE FOR COVER FOR PCB
PE
PE
INPUT INPUT PLATE PANEL BOX TUBE BOX PLATE TOP PCB TOP PCB PLATE
BOX BOX
INVERTER A4
GNYE GNYE
MAIN MAIN BOX FOR BOX FOR MAIN MAIN MAIN
FRAME PLATE TUBE TUBE PLATE PLATE PLATE
Power inlet 2nd ground wire
5 4 3 2 1
5 4 3 2 1
X127
3
CPU - BOARD, A1
LED LAMP, A10
BK
BN
CPU - BOARD, A1
X119 W1
2 x AWG18 + shield
D 6
5
D
4 X21
3
2 BK
WH BK
RD4
T1
2
B - RS485
BN AWG 18
A - RS485 1 WH BN
RD7
F6 RD3
-- 0Vac
S1
1
-- 24Vac, 4A
AWG 18 AWG 18
-- NC
RD2 -- 0Vac
RD1 S2
-- 24Vac, 4A
W14
Transformer
RD6
RD5
RD2
RD1
2 x AWG18 + shield
4 x AWG 18
A2
OPTION: X4 X1
Removal measurement unit
10
1
1
A7 GN
W20 2
X22 3
1
WH WH BN 4
2 BN 3 X 0.25mm2 1 A - RS485
RP1 BN BN BK 5 B - RS485
2
GN GN 6
3
YE GY
3
4 7
WH
YE YE 8
5
9
0.25mm2
10
W19
X3
C 6 X 0.5mm2 WH 1
4
C
X8
BN 2
3
OPTION:
GN 3
2
EXTENSION PUMP MODULE Pumps 4 to 6
GY 4
1
YE 5
MACHINE CONTROL - BOARD, A2 A9
X14
6
A8 PP 2
PP 1
X7 X207
BU10 STRAP 0,75mm2
PLUG & PLAY
PLUG & PLAY
DOSER CONNECTION BOARD
2
W25 2 pol Molex Mini-Fit Jr 2 pol Molex Mini-Fit Jr
X9
16 X6 X206
1
RD BK
2
8
15
WH
GY
6 X 0.5mm2 PUMP 6 (-)
PUMP 6 (+) M8
+ -
M8
2
7 YE PUMP 6
3 14
6 GN
PK
2 pol Molex Mini-Fit Jr 2 pol Molex Mini-Fit Jr
PUMP 6
4 13 BN
X12
5 X5 X205
12 BE RD BK
4
11
BN
PUMP 5 (-)
PUMP 5 (+) M7
+ -
M7
2
PUMP 5
1
3
10
2 BK 2 pol Molex Mini-Fit Jr 2 pol Molex Mini-Fit Jr
PUMP 5
2 9 X4 X204
X8
1
W24 RD BK
3
17
BU 26 X24 26 GND-S PUMP 4 (-)
PUMP 4 (+) M6
+ -
M6
2
4 13 13 PUMP 4
5
15 POL D-SUB
GN
4 X 0.5mm2
GN
25
12
25
12 PUMP 6 (+)
2 pol Molex Mini-Fit Jr 2 pol Molex Mini-Fit Jr
PUMP 4
6 24 24
X7
6 BN 11 11 PUMP 6 (-)
5 BN 23 23
7 YE 10 10 PUMP 5 (+)
X3 X203
4 YE 22 22
RD BK
8
3 PK
PK 9
21
9
21
PUMP 5 (-) PUMP 3 (-)
PUMP 3 (+) M5
+ -
M5
2
9 X13 X10 X6 X5 WH 8 8 PUMP 4 (-) PUMP 3
10
2 WH
GY
20
7
20
7 PUMP 4 (+) PUMP 3
X11
8
6 6 PUMP 3 (-) X2 X202
GN 18 18 RD BK
BN
5
17
5
17
PUMP 3 (+)
PP 2
PUMP 2 (-)
PUMP 2 (+) M4
+ -
M4
2
BU PUMP 2
YE BK
4
16
4
16
PUMP 2 (-)
PP 1
PUMP 2 (+) 2 pol Molex Mini-Fit Jr 2 pol Molex Mini-Fit Jr
PUMP 2
B B
3 3
PK 15 15
WH
WH
BN X1 X201
BN
BN
2 2 PUMP 1 (-)
RD
RD
RD
RD BK
BK
BK
BK
W23
WH BE 14
1
14
1 PUMP 1 (+)
PUMP 1 (-)
PUMP 1 (+) M3
+ -
M3
2
GY PUMP 1
27 27 GND-S
2 pol Molex Mini-Fit Jr 2 pol Molex Mini-Fit Jr
PUMP 1
6 X 0.5mm2
25 POL D-SUB FEMALE 25 POL D-SUB MALE
W21
RD24
BK36
2 X 0.25mm2
W22
2 X 0.25mm2 2 X 1mm2
WH
BN
X23
1
2
BN
BU
COLOR CODE:
------------------------
BK = black
BN = brown
X7
WH
BN
RD BK RD BK RD BK RD = red
OG = orange
YE = yellow
GN = green
3 BU = blue
VT = violet
1 2
GY = grey
A Air valve
WH = white
PK = pink A
BE = beige
B1 Y2 Y3 Y4 Y5 RO = rose
GNYE = yellow-green
Top position Sample mover up Air outlet 1 Air outlet 2 Sample mover down
STRUERS A/S
REV A: OH (2006/10/08)
REV B: OH (2007/01/09) Pederstrupvej 84
- S1 and S2 on T1 changed DK-2750 Ballerup
Drain/recirk. - see changes on other pages Denmark
REV C: OH (2007/05/03)
REV D: OH (2007/07/27)
ABRAPOL-20, CIRCUIT DIAGRAM
Y6 - see changees on page 3
REV E: OHO (2008/11/20)
- see changees on page 3
MACHINE CONTROL - BOARD, A2
OPTION: REV F: TKO (2009/04/02) - A10 added Size CAGE Code DWG NO Rev
A1
Shifting valve: drain/recirk. 15883120 F
Friday, April 17, 2009 Scale Sheet
OH 4 of 5
5 4 3 2 1
5 4 3 2 1
A4
Frequency Inverter - 3kW, safety class 3 CPU - BOARD, A1
(Safe standstill part only)
X3.1
X107-1
K31 RD1
5V
D D
34 RD2
X20-2
RD1 1
SB3-3
2 RD2
X20-1
BU6
X107-2
GND
X127-2 X21-2
T1
24V BK RD4
24V -- 0Vac
RD3 -- 24Vac, 4A
S1
INVERTER A4 CONTROL PSU 24VDC -- NC
33 RD1
X20-3
BU5
X107-12 X127-1
BN
X21-1
RD7
F6 -- 0Vac
S2
-- 24Vac, 4A
GND GND
Transformer
C C
YS1
A3 X108-4
BK4 12
TO MICROPROCESSOR (P63)
Frequency Inverter - 3kW, safety class 3
(Safe standstill part only) GND DOOR
X3.1 SWITCH
X107-9 X108-1
K31 OG
5V
BK1 11
5V
5V
A EMERGENCY STOP A
5 4 3 2 1
5 4 3 2 1
X7
PP 2
PP 1 PLUG & PLAY
2
D D
2 pol Molex Mini-Fit Jr
X6
PUMP 6 (-)
PUMP 6 (+) PUMP 6, M8
2
2 pol Molex Mini-Fit Jr
X5
PUMP 5 (-)
PUMP 5 (+) PUMP 5, M7
2
2 pol Molex Mini-Fit Jr
X8 X4
26 GND-S PUMP 4 (-)
C
PUMP 4 (+) PUMP 4, M6 C
2
13
25
12 PUMP 6 (+)
2 pol Molex Mini-Fit Jr
24
11 PUMP 6 (-)
23
10 PUMP 5 (+)
X3
22 MOUNTING HOLES
9 PUMP 5 (-) PUMP 3 (-)
21 PUMP 3 (+) PUMP 3, M5
2
8 PUMP 4 (-)
20
7 PUMP 4 (+)
2 pol Molex Mini-Fit Jr
19
6 PUMP 3 (-) X2 PCB1
18 PUMP 2 (-)
5 PUMP 3 (+)
17 PP 2
PUMP 2 (+) PUMP 2, M4
2
12050428-1
4 PUMP 2 (-)
GND-S
16 PP 1
B PUMP 2 (+) 2 pol Molex Mini-Fit Jr B
3
15 X1
2 PUMP 1 (-)
14 PUMP 1 (-)
1 PUMP 1 (+) PUMP 1 (+) PUMP 1, M3
1
27 GND-S
2 pol Molex Mini-Fit Jr
25 POL D-SUB MALE
STRUERS A/S
REV A: OH (2006/11/09) PEDERSTRUPVEJ 84
DK-2750 BALLERUP
DENMARK
PHONE: +45 44 600 800
5 4 3 2 1
5 4 3 2 1
PCB plate
COLOR CODE:
------------------------
BK = black
BN = brown
W11 W12 RD = red
OG = orange
W14 YE = yellow
W13
SCREEN
GN = green
D W15 BU = blue D
GY
YE
WH
BK
BN
VT = violet
RD
BN
GY = grey
1 2 3 4 5 6 WH = white
6 5 4 3 2 1
A1 1
6 7 8 9
2 3 4 5
9 8 7 6
5 4 3 2 1 7 8 9
10
11 12
PK = pink
BE = beige
RO = rose
X105 GNYE = yellow-green
SCREEN
SCREEN
X125
RD
GN
BK
BK
PK
BN
BU
BN
BN
1
X115 X117 X119
A1:X105 to A5 (page 2)
14 X127
BN
1
BK
2
SCREEN
3
X106
4 3 2 1
A1:X106 to A5 (page 2)
W1
C W16 C
A1:X114 to SB1, SB2 (page 2)
X114
BK BN
10
1 2 3 4 5
OG RD
6 7 8 9
BK BN
OG RD
W17 A1:X114 to SB3-1, SB3-2 (page 2)
X113
GY
8
1 2 3 4
BN
6 7
PK WH
GN YE
W18
5
5 4 3 2 1
X111
5 6
1 2 3
B X112 B
W11 A1:X115 to X116 (page 6)
1
W12 A1:X115 to A4:X3 (page 7)
RD2
BK6
BK5
BK4
BU5
BK
BN
W1
RD1
RD
OG
BK3
BK2
BK1
BU6
RD1 RD2
W5 X20 to SB3-3 (page 2) BU5 RD1 W2 W2 A1:X107 to A3:X3.1 (page 4)
A A
STRUERS A/S
REV A: OH (2007/05/02)
Pederstrupvej 84
REV B: OH (2007/07/27) DK-2750 Ballerup
- See changees on other pages Denmark
REV C: AKR (2007/09/19)
- See changes on page 8.
REV D: OHO (2008/12/03) ABRAPOL-20, WIRING DIAGRAM
- See changees on other pages PCB plate
REV E: TKO (2009/04/02) Size CAGE Code DWG NO Rev
- See changees on other pages A2
15883450 E
Thursday, April 02, 2009 Scale Sheet
OHO 1 of 10
5 4 3 2 1
5 4 3 2 1
Front panel
D A5 D
A5 to A1:X105 (page 1)
A5 to A1:X106 (page 1)
C C
BR1 to A1:X113 (page1)
W18
A6
A6 to A1:X112 (page 1)
XPE21
GNYE13 XPE21 to XPE22 (page 1)
W18
W5
BR1 SB3-3 to X20 (page 1)
1 2 3 4 5 6
GY
GN
YE
PK
BN
WH
W16
B B
OG
BK
3
3
SB1
SB2
4
4
RD
BN
W17
BK BN
2
SB3-1 1
COLOR CODE:
XPE24 OG
2
SB3-2 1
RD ------------------------
BK = black
BN = brown
RD1 RD2
1
SB3-3 2
W5
RD = red
OG = orange
YE = yellow
GNYE14 GN = green
BU = blue
VT = violet
GY = grey
A WH = white
PK = pink
A
BE = beige
RO = rose
XPE24 to XPE23 (page 9) GNYE = yellow-green
5 4 3 2 1
5 4 3 2 1
COLOR CODE:
------------------------
Plate for top PCB W24
BK = black
BN = brown
W25 RD = red
OG = orange
YE = yellow
GY
BK
PK
BN
BN
WH
W23
GY
GN
YE
PK
BN
X9 GN = green
BU = blue
VT = violet
A2 9 10 11 12 13 14 15 GY = grey
D 4 3 2 1 1 2 3 4 5 6 7 8 6 5 4 3 2 1
WH = white D
PK = pink
X8 X7 BE = beige
GN
BE
YE
BU
WH
RO = rose
X5 GNYE = yellow-green
BK
1 2 3 4 5 6 7 8
BK36 RD
X1 BK
RD1
Y2
5 4 3 2 1
RD2 RD
RD24 BK
RD5 RD
RD6 BK36
RD24
Y2 to Y5 are placed from the backside of the Plait for top PCB
BK
Y3
RD
C X4 BK C
X6
10 9 8 7 6 5 4 3 2 1
Y4 WH
1 2
RD BN
W22
BK
Y5
RD X10
XPE18
1 2 3 4 5 6 7 8
WH
BN
GNYE11
W21
1 2
W24 A2:X9 to A8:X8 (page 6)
W21
GY
GN
BK
YE
BN
BN
WH
W14
W14 A2:X3 to A1:X119 (page 1)
1 2 W1 X21 to A1:X127 (page 1)
X21
A A
X21:1 to F6 (page
X21:2 to T1 (page
A2:X1 to T1 (page
STRUERS A/S
REV A: OH (2007/05/02)
Pederstrupvej 84
REV B: OH (2007/07/27) DK-2750 Ballerup
- See changees on other pages Denmark
REV C: AKR (2007/09/19)
- See changes on page 8.
REV D: OHO (2008/12/03) ABRAPOL-20, WIRING DIAGRAM
- See changees on other pages Plate for top PCB
REV E: TKO (2009/04/02)
- A10 connectoed to X21 Size CAGE Code DWG NO Rev
4)
4)
4)
5 4 3 2 1
5 4 3 2 1
T1 to X21:2 (page
F6 to X21:1 (page
T1 to A2:X1 (page
Connection of W3 is represented from the side view
39
A3
28
39
BU2
BU1
28
3)
3)
3)
7
D R1
LO D
W3 HI X3
XPE4 7
RD4
RD7
RD
LO
BK
HI
BN
7
7
XPE3 20
20
GNYE4
BR2 OG
T2 K31
BR1
T1
X2.2 K32
BN
XPE25 X3.1 33
BK
GNYE15 RD
34
W4 U1 W2
U GNYE3
A3:X2.1, X2.2 to M1 (page 7) V2 PE
V BK18
W3 L1
W BK17
C BR1 X2.1 X1.1
L2
BK16 C
L3
GNYE BR2
+UG
SCREEN PE
-UG
W4 W26
Connection to
main plate
RD7
RD3
GNYE22
XPE35
BK14 BK6
T1 RD7 RD3 F3
See diagram 15093452 F6
for wiring
BK9
BK11 BK13 BK5
P1 F5 BK9 F2
OUT
OUT
IN
BU 0V
B B
IN
P2
OG 0V
RD2 YE BK8
RD1
0V
S2 24V
VT
GN
P1
BK10
F4 BK8
BK12
F1
BK4
RD4 WH 200V
0V
RD3 RD P2 BK15 GNYE23 GNYE1
S1 24V RD 200V X25 XPE1 COLOR CODE:
------------------------
GNYE2 GNYE3
P1
XPE2
XPE35 to PE (page
XPE1 to PE (page
VT = violet
GY = grey
WH = white
F1, F2, F3 to L10 (page 5) PK = pink
BE = beige
10)
7)
1)
RO = rose
W3
GNYE = yellow-green
10)
10)
7)
7)
7)
7)
A A
REV D: OHO (2008/12/03)
STRUERS A/S
- Connection A3:X1.1 to L10 removed REV A: OH (2007/05/02)
Pederstrupvej 84
- X25, F4, F5 reconnected to Z1 REV B: OH (2007/07/27) DK-2750 Ballerup
- Connection W26 A3:X1.1 to Z1 added - XPE35 and GNYE22 added
Denmark
- GNYE23 added - See changees on other pages
- See changees on other pages REV C: AKR (2007/09/19)
REV E: TKO (2009/04/02) - See changes on page 8.
ABRAPOL-20, WIRING DIAGRAM
- See changees on other pages Main plate - part 1
Size CAGE Code DWG NO Rev
A2
15883450 E
Thursday, April 02, 2009 Scale Sheet
OHO 4 of 10
5 4 3 2 1
5 4 3 2 1
D XPE32 XPE31 D
GNYE20 GNYE19
C C
BK12
L3.1
L1
BK9
BK13
L2.1
L10
L2
BK8
L1.1
BK14
L3
BK7
4)
4)
COLOR CODE:
------------------------
BK = black
BN = brown
RD = red
OG = orange
YE = yellow
GN = green
BU = blue
VT = violet
GY = grey
WH = white
PK = pink
BE = beige
RO = rose
GNYE = yellow-green
A A
STRUERS A/S
REV A: OH (2007/05/02)
Pederstrupvej 84
REV B: OH (2007/07/27) DK-2750 Ballerup
- See changees on other pages Denmark
REV C: AKR (2007/09/19)
- See changes on page 8.
REV D: OHO (2008/12/03) ABRAPOL-20, WIRING DIAGRAM
- Connection L10 to A3:X1.1 removed Main plate - part 2
- See changees on other pages
REV E: TKO (2009/04/02) Size CAGE Code DWG NO Rev
- See changees on other pages A2
15883450 E
Thursday, April 02, 2009 Scale Sheet
OHO 5 of 10
5 4 3 2 1
5 4 3 2 1
13 12 11 10 9 8 7 6 5 4 3 2 1
BE
25 24 23 22 21 20 19 18 17 16 15 14
WH
BN X1 X2 X3 X4 X5 X6 X7
PK
BK
YE
D D
RD
RD
RD
BU
BN
GN
1 1 1 1 1 1 1
GY
WH 2 2 2 2 2 2 2
PK
BK
BK
BK
YE
BN for connection of Extension pump module
GN A9 see page 8 with accesories
W21
W24 A8:X8 to A2:X9 (page 3)
W19
BN
WH
GY
GN
YE
BN
RD
RD
RD
for connection of Removal Measurement unit + + +
C A7 see page 8 with accesories W21 X23 to A2:X10 (page 3) C
BU
BN
BK
BK
BK
PUMP1 PUMP2 PUMP3
Connection of Tap water valve Y1
W15
Y1 to A1:X125 (page 1)
Connection of X7 (for Shifting valve Y6) Connection of Door Guard Switch YS1
X7 to A2:X6 (page 3) W22
YS1
GY
BN
BK4
Water to OP-S
12 32 BK6 A2
BK3
WH
BN
3 11 22 21 A1
BK5
1 2
B BK1 BK2 B
Disc Cooling Tap water
for connection of Shifting valve Y6
X7 see page 8 with accesories
YS1 to A1:X108 (page 1) W10
WH GN COLOR CODE:
------------------------
BK = black
BN = brown
RD = red
Connection of Connector for service X120 Connection of Connector for Cooli X116 OG = orange
YE = yellow
PK YE GN = green
X116 BU = blue
SCREEN
VT = violet
3 4 BN GY = grey
X120
RD
BK
BN
Y1 BU
X116 to A1:X115(page 1) W11
WH = white
PK = pink
BE = beige
W13 X120 to A1:X117 (page2) 5 4 3 2 1 2 1 RO = rose
9 8 7 6 GNYE = yellow-green
A A
STRUERS A/S
REV A: OH (2007/05/02)
Pederstrupvej 84
REV B: OH (2007/07/27) DK-2750 Ballerup
- See changees on other pages Denmark
REV C: AKR (2007/09/19)
- See changes on page 8.
REV D: OHO (2008/12/03) ABRAPOL-20, WIRING DIAGRAM
- See changees on other pages Various connections - part 1
REV E: TKO (2009/04/02) Size CAGE Code DWG NO Rev
- See changees on other pages A2
15883450 E
Friday, April 17, 2009 Scale Sheet
OHO 6 of 10
5 4 3 2 1
5 4 3 2 1
Connection of Inverter A4 and Mans switch QS1
W6 A4:X3.1 to X20 (page 1)
GNYE16
GNYE1
W26
W3
RD1
RD2
RD1
RD2
A4:X1.1 to Z1 (page 10)
BK18
BK17
BK16
A4:X1.1 to XPE2 (page 4)
K31
K32
GNYE2
33
34
D D
+UG
-UG
PE
L3
L1
L2
X3.1
X1.1
A4
A4:X3 to A1:X115 (page
QS1 to F1, F2, F3 (page
4)
4)
4)
4)
4)
1)
C C
LO
39
28
39
28
20
20
HI
HI
7
7
7
RD
A10
RD
BK
BN
BK
BN
BU3
3)
BU4
X2.1
BR1
BR2
PE
X2.2
U
W
V
T2
T1
W12
W3
SCREEN
GNYE16
GNYE
BR2
BR1
W3
U1
V2
Mains
GNYE22
GNYE1
control
BK1
BK2
BK3
X1
box
GNYE7
GNYE22
GNYE1
PE
PE
1 3 5
XPE10
BK4
BK5
BK6
T1 T2 T3
A4:X2.1, X2.2 to M2 (page
BK or GY
GNYE
(GNYE)
RD
OG
BK
BK
BN
GN
OR
PE to XPE35 (page
PE to XPE1 (page
COLOR CODE:
------------------------
QS1 W1 W1
BK = black (EU) (UL/CSA)
A BN = brown
RD = red
OG = orange
A
YE = yellow
2nd ground Power inlet
L1 L2 L3 wire
4)
7)
4)
GN = green
BU = blue
VT = violet REV A: OH (2007/05/02)
STRUERS A/S
BK1
BK2
BK3
W9
5 4 3 2 1
5 4 3 2 1
D D
Connection of Removal Connection of Extension pump module A9 Connection of Shifting valve Y6
measurement unit A7
for other connection of X7
see page 6
A7 X8
A8
Y6
X1 X2 X3 X4 X5 X6 X7
RD
RD
RD
BU10
X7
1 1 1 1
2 2 2 2
BU10
BK
BK
BK
C C
A9
1 M6 M7 M8
RD
RD
RD
GN
W20 + + +
BU10
WH
BN
- - - PLUG
BK
BK
BK
2 3
&
PUMP4 PUMP5 PUMP6 PLAY
B B
COLOR CODE:
------------------------
for other connection of X22 BK = black
see page 6 BN = brown
RD = red
OG = orange
1 2 3 4 5
X22 YE = yellow
GN = green
BU = blue
VT = violet
GN
YE
YE
BN
WH
GY = grey
W20 WH = white
PK = pink
BE = beige
RO = rose
GNYE = yellow-green
A STRUERS A/S
A
REV A: OH (2007/05/02)
Pederstrupvej 84
REV B: OH (2007/07/27) DK-2750 Ballerup
- See changees on other pages Denmark
REV C: AKR (2007/09/19)
- RD and BK at X7 changed to BU10.
REV D: OHO (2008/12/03) ABRAPOL-20, WIRING DIAGRAM
- See changees on other pages Accessories
REV E: TKO (2009/04/02) Size CAGE Code DWG NO Rev
- See changees on other pages A2
15883450 E
Thursday, April 02, 2009 Scale Sheet
OHO 8 of 10
5 4 3 2 1
5 4 3 2 1
D
LEFT SUPPORT PLATE MAIN FRAME COVER FOR TOP D
GNYE6
MAIN FRAME
XPE7
C
FRONT PANEL MAIN FRAME C
XPE6 XPE5
GNYE5
XPE29 XPE27
BOX FOR MAIN FRAME
TUBE BOX
B B
XPE12
GNYE8
XPE11
GNYE18 GNYE17
GNYE9
XPE13 PUMPS MODULE DOSER UNIT / COLOR CODE:
------------------------
COVER PLATE BK = black
BASE FRAME BN = brown
RD = red
OG = orange
XPE30 XPE28 YE = yellow
GN = green
XPE14 BU = blue
VT = violet
GY = grey
WH = white
PK = pink
BE = beige
RO = rose
GNYE = yellow-green
A A
STRUERS A/S
REV B: OH (2007/07/27)
Pederstrupvej 84
- New page 9 created
DK-2750 Ballerup
- See changees on other pages
Denmark
REV C: AKR(2007/09/19)
- See changes on page 8.
REV D: OHO (2008/12/03) ABRAPOL-20, WIRING DIAGRAM
- See changees on other pages
REV E: TKO (2009/04/02) Protective bonding circuit (the rest)
- See changees on other pages
Size CAGE Code DWG NO Rev
A2
15883450 E
Thursday, April 02, 2009 Scale Sheet
OHO 9 of 10
5 4 3 2 1
5 4 3 2 1
Z1 to A4:X1.1 (page
Z1 to A3:X1.1 (page
Motor coupling for Motor coupling for
3 x 380V-480V 50/60Hz 3 x 200V-240V 50/60Hz
W26
D
M1 STAR DELTA D
BR1
BR2
BR1
BR2
BR1
BR2
7)
4) S1 S2 S1 S2 S1 S2
GNYE24
BK18
BK18
BK17
BK17
BK16
BK16
W2 U2 V2 W2 U2 V2 W2 U2 V2
W3
W3
W3
U1
U1
U1
LOAD
V2
V2
V2
GNYE
GNYE
GNYE
C C
BR1
BR2
BR1
BR2
BR1
BR2
Z1 S1 S2 S1 S2 S1 S2
B B
W2 U2 V2 W2 U2 V2 W2 U2 V2
U1 V1 W1 U1 V1 W1 U1 V1 W1
COLOR CODE:
GNYE
GNYE
GNYE
------------------------
W3
W3
W3
U1
U1
U1
V2
V2
V2
BK = black
BN = brown
RD = red
OG = orange
YE = yellow
GN = green
BU = blue
screen connected VT = violet
to cable gland GY = grey
WH = white
W9 M2 to A4:X2.1, X2.2 (page 7) PK = pink
BE = beige
LINE
RO = rose
GNYE = yellow-green
A A
L3 L2 L1
REV D: OHO (2008/12/03) STRUERS A/S
GNYE23 - New page created Pederstrupvej 84
- Connection of M1, M2 updated DK-2750 Ballerup
- See changees on other pages Denmark
REV E: TKO (2009/04/02)
- See changees on other pages
ABRAPOL-20, WIRING DIAGRAM
Connection of Z1, M1, M2
BK15
BK9
BK8
Country Transformer T1
Gear Motor M1 Freq. converter Motor M2 Freq. Converter Fuses Fuses Fuse Ordering
nom. Voltage
nameplate A3 nameplate A4 F1,F2,F3 F4, F5 F6 number
/frequency Type Connections
05886129
Japan Connection:
3x200V / 50Hz 200 ÷ 210 V
3x200-
CSA-norm.
210V/50 or
Imax = 0,65 A
60Hz
Japan + USA + Item nr.: Item nr.:
Item nr.: Item nr.:
Canada 2MD10000 2ME60000 Item No.:
2PU82303 2PU82303 Mains Fuses
3x200-210V / 60 Hz Voltage: 3 x 230 V Voltage: 2x
Voltage: Voltage: 20 AT 2 AT Max 50 AT
Power: 0,55 kW 3 x 230V 2XL30402
3 x 230 V 3 x 230 V (Class-CC) (Class-CC)
Power: 2,2 kW
Power: Power: 2FC11200 2FC11020
Connection:
3 kW 3 kW
DELTA Connection: Connection: 05886135
Norway, France, DELTA 220 ÷ 240 V
Belgium
Item No.: 3x220-240V
3x220-230V / 50 Hz
2MT72034 Imax = 0,65 A /50 or 60 Hz
USA, Mexico,
Ulveco No.: Item No.: 4 AT Mains Fuses
Brasil, S.Korea.
AA-72034 2x 2FU14200 Max 50 AT
3x220-240V / 60 Hz
2XL30402
See diagram: Connection: 05886146
15093452 380 ÷ 415 V
Europe
3x380-415V
3x380-415V / 50Hz 16 AT 1 AT
Imax = 0,33 A /50 or 60Hz
Item nr.: (aM) (aM)
Item nr.: 2FC10160 2FC10010
South America 2ME60000
2MD10000 Item nr.: Item nr.: Item No.: Mains Fuses
3x380-415V / 60Hz Voltage:
Voltage: 3 x 400 V 2PU84303 2PU84303 1x Max 50 AT
3 x 400V 2XL30402
Power: 0,55 kW Voltage: Voltage:
Power: 2,2 kW
3 x 400 V 3 x 400 V Connection: 05886154
Connection: Power: Power: 460 ÷ 480 V
Connection:
STAR 3 kW 3 kW 3x460-480V
CSA-norm. STAR 15 AT 1 AT
Imax = 0,33 A /60Hz
USA + Canada (Class-CC) (Class-CC)
3x460-480V / 60 Hz 2FC11150 2FC11010
Item No.: Mains Fuses
1x 50 AT
2XL30402