Unit 7 Voltammetry: Structure
Unit 7 Voltammetry: Structure
UNIT 7 VOLTAMMETRY
Structure
7.1 Introduction
Objectives
7.2 Electrodes and Processes
7.3 Common Voltammetric Methods
7.4 Direct Methods
Linear Sweep Voltammetry
Sampled DC Polarography
Hydrodynamic Voltammetry
7.5 Pulse Methods
Normal Pulse Voltammetry
Differential Pulse Voltammetry
Square Wave Voltammetry
Cyclic Voltammetry
7.6 Stripping Methods
Anodic Stripping Voltammetry
Cathodic Stripping Voltammetry
Adsorptive Stripping Voltammetry
7.7 Alternating Current Methods
Voltammetry-Instrument
7.8 Summary
7.9 Terminal Questions
7.10 Answers
7.1 INTRODUCTION
This is the first unit of the electroanalytical methods course. We will start our
discussion by defining what is voltammetry and then continue our discussion with the
different methods of voltammetry. We will discuss the sampled DC polarography and
hydrodynamic voltammetry. We will also discuss in detail the pulse methods viz.
Normal pulse polarography, Differential pulse polarography and Square wave
polarography/voltammetry.
Then cyclic voltammetry (CV) will be discussed along with a diagnostic test to decide
reversibility of a redox reaction. Finally the instrumentation involving CV will be
explained.
Objectives
After studying this unit, you should be able to:
• define voltammetry,
• discuss the electrodes and processes of voltammetry,
• discuss the dc methods in voltammetry,
• explain the pulse methods of voltammetry,
• explain anodic, cathodic and adsorptive stripping methods in voltammetry,
• discuss the alternating current methods in voltammetry,
• name the instruments involved in voltammetry, and
• select a suitable method of voltammetry for the analysis of a particular sample.
5
Electroanalytical
Methods-III 7.2 ELECTRODES AND PROCESSES
Voltammetry is the general name given to a group of electroanalytical methods in
which the current is measured as a function of applied potential wherein the
polarization of the indicator or working electrode is enhanced. For this, the working
electrodes used in voltammetry are generally microelectrodes having surface area of
Microelectrodes are
few square micrometers to square millimeters. This field has been developed from
working electrodes that Polarography (developed by Heyrovsky); which is a particular type of voltammetry,
have surface areas smaller which used a dropping mercury electrode as the working micro-electrode.
than a few square Voltammetry includes all techniques in which the current potential measurements are
millimeters. made at stationary and fixed electrodes. After studying this unit, you will be in a
position to understand different voltammetric techniques.
Thus, we can say that the electrochemical method in which the current is measured as
a function of the potential applied to a microelectrode is known as voltammetry. These
electrodes may be solid like gold, platinum and glassy carbon or mercury. If the
electrode is formed by a drop of mercury dropping from a fine glass capillary, the
technique is called “Polarography” (discussed in detail in Unit 8).
Other forms of working electrodes used are hanging mercury dropping electrode
(HMDE), static mercury drop electrode (SMDE), thin mercury film electrode (TMFE),
glassy carbon electrode (GCE), carbon paste electrode (CPE), etc. Working electrodes
made of noble metals are used less frequently. Some electrodes are given in Fig. 7.1.
Processes
Various methods are assigned to the terms voltammetry and polarography which differ
in the measuring techniques and the type of electric potential used to excite the
determination process.
Voltammetry is widely used for the quantitative analysis of trace metal ions at micro
level. This study of redox processes lead to the qualitative and quantitative analysis of
heavy metals and some organic substances in solution. In addition to this aspect this
technique is quite versatile for research purpose in the fields of
i) redox processes
6
Voltammetry
7
Electroanalytical
Methods-III 7.3 COMMON VOLTAMMETRIC METHODS
The most widely used voltammetric methods are
B) Pulse methods
v) Normal Pulse voltammetry
vi) Differential Pulse voltammetry
vii) Square wave voltammetry
viii) Cyclic voltammetry
C) Stripping voltammetry
ix) Anodic stripping voltammetry
x) Cathodic stripping voltammetry
xi) Adsorptive stripping voltammetry
A + e− = B E o = 0.00V
which obeys Nernst equation.
Here, E is the potential of the working electrode, E˚ is the standard reduction potential
of A.
8
As E is lowered further, all the A is converted to B, as soon as it reaches the electrode. Voltammetry
In normal DC polarography, the electrode area is not constant and hence the current
fluctuates with the life time of each drop within some range as shown in Fig.7.3. This
problem can be alleviated by measuring the current once at the end of the life time of
each drop and this method is called Sampled DC Polarography or Tast Polarography.
9
Electroanalytical The major advantage of this current sampling is that it substantially reduces large
Methods-III current fluctuations due to continuous growth and fall of the drops at dropping
electrode.
From Fig. 7.3, it can be seen that the current near the end of the life of a drop is nearly
constant and it is this current only that is measured in the current sampled technique
resulting in a smooth curve with very little current fluctuations.
SAQ 1
a) Why does the current fluctuate with the life time of each drop as shown in
Faradaic Current: Fig.7.3?
Current due to
electrochemical reaction …………………………………………………………………………………………...
in a cell.
…………………………………………………………………………………………..
Cathodic Current: A
faradaic current due to a b) What do you mean by Tast Polarography?
reduction reaction.
…………………………………………………………………………………………...
Anodic Current: A
faradaic current due to an …………………………………………………………………………………………..
oxidation reaction.
SAQ 2
What are the major advantages of current sampling in Sampled DC polarography?
…………………………………………………………………………………………...
…………………………………………………………………………………………..
10
ii) convection resulting from stirring or vibration. Voltammetry
iii) diffusion due to concentration gradient established near electrode during the
reaction.
This is the current that is proportional to the concentration of electroactive substance
and is given by
nFAD
il = ca … (7.2)
δ
where
ca = Concentration of analyte, n = number of electrons, F = Faraday constant,
A = Area of the electrode surface, D = Diffusion coefficient of the analyte,
il = Limiting current, δ = thickness of Nernst layer
Hydrodynamic
voltammetry:
Voltammetry in which the
Fig. 7.5: A three electrode cell for hydrodynamic voltammetry
solution is mechanically
Here, stirring is accomplished using a magnetic stirrer but in some other cases the stirred.
electrode is rotated at a constant high speed in the solution thus providing a stirring
action. The obtained voltammogram is a wave and the current voltage plot is
characterized by an initial residual current that grows up with increasing potential till
the E½ and then slowly increases till a stationary level is reached. At this value of
potential the ions reaching the electrode discharge immediately thus the concentration
of the analyte in the diffusion layer is almost nil. The continuous flowing of ions into
the diffusion layer due to stirring maintain a constant current.
Another way of carrying out this technique involves the flowing of analyte solution
through a tube in which the microelectrode is fixed. This technique has been used as a
voltammetric detector in HPLC.
Some important applications of this technique include the detection and determination
of oxygen and chemical species like glucose, lactose and sucrose. This detector has
also been used in fundamental studies in electrochemistry and detection of end points
in electrometric and volumetric titrations.
11
Electroanalytical SAQ 3
Methods-III
a) What do you mean by Hydrodynamic voltammetry?
…………………………………………………………………………………………...
…………………………………………………………………………………………...
b) In Hydrodynamic voltammetry, how the flow of analyte solution is carried out?
…………………………………………………………………………………………...
…………………………………………………………………………………………...
The potential excitation signals used in different voltammetric methods are given in
Fig.7.6 and the individual methods are discussed in the following sections.
The peak current in differential pulse voltammetry increases linearly with the
concentration of the analyte. It also increases with the pulse amplitude but in a
complicated linear fashion. In practice amplitudes greater than 100 mv are not used,
since the peaks broaden and the resolution is significantly inhibited.
12
Voltammetry
13
Electroanalytical
Methods-III
Fig. 7.7: Pulse application measuring procedure and current potential curves for
(a) Normal pulse voltammetry; and (b) Differential pulse voltammetry
The higher detection limit in this technique is due to superior discrimination against
charging current and impurity Faradaic current.
n2 F 2 D
ip = A.C a ∆EA tp … (7.3)
4 RT π
where, ip = Peak current, n = Number of exchanged electrons, F = Faraday constant,
R = Gas constant, T = Temp. in Abs. Scale, A = Electrode surface area,
Ca = Concentration of the analyte, ∆ EA = Pulse amplitude, D = Diffusion coefficient
of the analyte, and tp = Pulse duration.
The detection limit for the determination by this technique is similar to square wave
polarography at 10–7 – 10–8 mol/dm3. But for irreversible reaction it is lowered.
For reversible reactions, a larger pulse is used so that its size is great enough for the
oxidation of the product formed in the forward reaction occurs during the reverse
phase. The forward pulse gives the cathodic current i1, while the reverse phase gives
the anodic current i2 and the difference between these two ∆i = i2 – i1 is plotted to give
the voltammograms. This difference is directly proportional to the concentration, the
potential at the peak corresponds to the E½. It is possible to increase the precision of
14
the analysis by signal averaging data from several voltammetric scans. Detection Voltammetry
limits for square wave voltammetry are reported to be 10–7 to 10–8 M.
SAQ 4
i) Distinguish between
a) Linear-scan polarography and pulse polarography.
b) Differential-pulse polarography and square-wave polarography.
…………………………………………………………………………………………...
…………………………………………………………………………………………...
…………………………………………………………………………………………...
ii) Which one is better and why? NPP(NPV) or DPP(DPV) ?
…………………………………………………………………………………………...
…………………………………………………………………………………………...
…………………………………………………………………………………………...
15
Electroanalytical silver/silver chloride electrode (Ag/AgCl).The controlling potential which is applied
Methods-III across these two electrodes can be considered as excitation signal. The excitation
signal for CV is a linear potential scan with a triangular waveform as given in Fig. 7.9.
This triangular potential excitation signal sweeps the potential of the electrode
between two values, sometimes called the switching potentials. The excitation signal
in Fig. 7.9 causes the potential first to scan negatively from +0.80 (initial potential) to
– 0.20 V (EF final potential) versus SCE at which point the scan direction is reversed,
resulting a positive scan back to the original potential of +0.80 V (Ei). The scan rate as
reflected by the slope is 50 mV per second as shown in Fig. 7.9. A second cycle is
indicated by the darked line. Single or multiple cycles can be used.
Fig. 7.9: Typical excitation signal for cyclic voltammetry-a triangular potential
waveform with switching potentials at 0.8 and –0.2 V versus SCE
Fig. 7.10: Cyclic voltammogram of 6mM K3Fe(CN)6 in 1M KNO3, Scan initiated at 0.8 V
versus SCE in negative direction at 50 mV/s. Platinum electrode area=2.54 mm2.
A cyclic voltammogram is obtained by measuring the current at the working electrode
during the potential scan. The current can be considered the response signal to the
potential excitation signal. The voltammogram is a display of current (vertical axis)
versus potential (horizontal axis). Since the potential varies linearly with time, the
horizontal axis can also be thought of as a time axis. This will be helpful in
understanding the basic principle of the technique.
16
obtain this voltammogram is shown in Fig. 7.9 but with the negative switching Voltammetry
potential of –0.15V. Thus, the vertical axis in Fig.7.9 is now the horizontal axis of
Fig.7.10. The initial potential (Ei) of 0.80 V applied at (a) is chosen to avoid any
electrolysis of [Fe(CN)6]3– when the electrode is switched on. The potential is then
scanned negatively, forward scan, as indicated by the arrow. When the potential is
sufficiently negative to reduce [FeIII(CN)6]3– cathodic current is indicated by (b) due to
electrode process
where E0′ is the formal reduction potential of the sample. An initial value of E which is
sufficiently positive of E0 ′ maintains a ratio in which [FeIII(CN)6]3– greatly
predominates. Thus, application of +0.80 V at the initial potential causes negligible
current. However, as E is scanned negatively, conversion of [FeIII(CN)6]3– to
[FeII(CN)6]4– by reduction is mandatory for satisfaction of the Nernst equation. The
ratio of iron redox states that must exist at the electrode surface at several potentials
during the scan is shown on the lower horizontal axis in Fig.7.10. The logarithmic
relationship between E and [FeIII(CN)6]3– /[FeII(CN)6]4– is reflected by a rapid rate of
change in the region where E = Eo i.e., [FeIII(CN)6]3– /[FeII(CN)6]4– = 1. This causes
the diametric rise in cathodic current (b → d) during the forward scan.
The important parameters of a cyclic voltammogram are the magnitudes of the anodic
peak current ( I p a ) and cathodic peak current ( I p c ), and anodic peak potential ( Ep a )
17
Electroanalytical and cathodic peak potential ( Ep ). These parameters are labeled in Fig.7.10. One
Methods-III c
method of measuring Ip involves extrapolation of a baseline current as shown in
Fig.7.10. The establishment of correct baseline is essential for the accurate
measurement of peak currents. It is generally difficult for more complicated systems.
The peak current for a reversible system is described by the Randles-Sevcik equation
for the forward sweep of the first cycle.
ip = ( 2.686 × 105 ) n 3 2 AD1 2C0v1 2 … (7.8)
Ip
c
= 1 … (7.9)
Ip
a
According to Eq. (7.8), Ip increases with v1/2. . However, the ratio of peak currents can
be significantly effected by chemical reactions coupled to the electrode process.
A redox couple in which both species rapidly exchange electrons with the working
electrode is called an electrochemically reversible couple. The formal reduction
potential (E0′) for a reversible couple is centered between Ep a and Ep c .
Ep + Ep
E 0′ = a c
… (7.10)
2
The number of electrons transferred in the electrode reaction (n) for a reversible
couple can be determined from the separation between peak potentials.
Peak separation greater than the predicted theoretical value (59/n mV) is typically
caused by IR drop due to high resistance between working and reference electrodes.
This is especially a problem at fast scan rates where ‘I’ can be large and in non-
18
aqueous solvent where ‘R’ can be large. Also, slow heterogeneous electron exchange Voltammetry
with the electrode can increase Ep .
The various criteria applied for the diagnosis of reversibility, irreversibility or quasi-
reversibility are summarized in Table 7.1.
Table 7.1: Diagnostic test to decide the reversibility of a redox reaction by cyclic
voltammetry
Criteria for Criteria for quasi–reversible Criteria for total
reversible system system irreversible system
(i) E = Epa– Epc (i) Ip increases with v1/2 (i) No reverse peak.
= (0.0591/n) V but is not proportional to
it.
(ii) [Ep –Ep/2] = (ii) [ I p a / I p c ] = 1 provided (ii) I p c ∝ v1/2
(0.0591/n) V
αc = αa = 0.5 * .
(iii) [ I p a / I p c ] = 1 (iii) Ep is greater than 0.059/n (iii) Ep c shift -30/ αc nα *
V and increases with
mV for each decade
increasing v.
increase in v.
(iv) Ip ∝ v1/2 (iv) Ep c shifts negatively with (iv) (Ep–Ep/2) = 48/ αc nα
increasing v. mV.
(v) EP is dependent
of v
(vi) At potential
beyond Ep, I– 2 ∝
time.
In some cases, the anodic current is low as compared to the cathodic current. This is
because throughout most of the experiment there is a concentration difference which
drives back the reduced species away from the electrode, most of the reduced species
produced in forward reduction process, therefore, diffuse into the bulk solution and
cannot be re-oxidized on the time scale of a cyclic voltammetric experiment during
reverse direction, i.e. anodic process. Hence, it gives reduced anodic peak current as
compared to cathodic peak current.
Instrumentation
Cyclic voltammetry requires a waveform generator to produce the excitation signal, a
potentiostat to apply this signal to an electrochemical cell, similar to a polarographic
cell, a current to voltage converter to measure the resulting current and an X-Y
recorder or oscilloscope to display the voltammogram. The first three items are
normally incorporated into a single electronic device, though modular instruments
19
Electroanalytical may also be used. The potentiostat ensures that the working electrode potential will
Methods-III not be influenced by the reactions which take place. Data may be obtained with X-Y
recorder at slow scan i.e. less than 500 mV/second while storage oscilloscope at faster
rates having scan rate upto 20,000 V/second have been used, however ratio faster than
100 V/second are rarely practical because of IR Drop and changing current.
Waveform
Generator Potentiostat
Cell
Current to
X Y voltage
converter
A large variety of working electrodes have been used in voltammetry. The dropping
mercury electrode is used especially in polarography and the hanging mercury drop
electrode (HMDE) in cyclic voltammetry. A thin coat of mercury can be deposited on
a substrate such as graphite to form a mercury film electrode (MFE). A significant
advantage of mercury is its good negative potential range. Solid electrode such as
platinum, gold, glossy carbon, wax impregnated graphite and carbon-paste are also
used in cyclic voltammetry. Such electrodes have a better positive potential range than
mercury.
20
SAQ 5 Voltammetry
…………………………………………………………………………………………...
ii) What is the difference between voltammetry and polarography?
…………………………………………………………………………………………...
…………………………………………………………………………………………...
iii) What is the advantage of solid electrode (like platinum) over mercury electrode
in CV?
…………………………………………………………………………………………...
…………………………………………………………………………………………...
iv) Why CV has been recognized as the most versatile electrochemical technique?
…………………………………………………………………………………………...
…………………………………………………………………………………………...
v) In a system that uses a three electrode potentiostat, between which two electrode
is the potential measured?
…………………………………………………………………………………………...
…………………………………………………………………………………………...
The first step is basically preconcentration or electrodeposition while the second one is
the stripping step. Three different types of this technique are available.
This may be explained in detail taking the example of Cu+2, the deposition reaction at
HMDE (Hanging Mercury Drop Electrode) held at a cathodic potential sufficient to
deposit the metal ion on the electrode
Cu 2 + (aq ) + 2e − = Cu (Hg ) … (7.12)
21
Electroanalytical where Cu(Hg) is copper amalgamated with mercury. This step is basically a
Methods-III preconcentration step where the metal ion from a large volume of the solution is
concentrated on a small volume of the electrode. The solution is stirred during the
electrolysis to increase the rate of deposition. Near the end of the deposition time
stirring is stopped and usually 1-30 mins are used for deposition. In the second step,
the potential is scanned anodically towards more positive potential. When the potential
becomes positive, the analyte is stripped from the electrode returning to its oxidized
form
And the current during this step is monitored as a function of potential giving rise to
the voltammogram, the peak current of which is proportional to the analyte
concentration in the solution. This technique is sensitive to experimental conditions
like area of mercury film electrode or size of the Hg drop, deposition time, rest time,
rate of stirring and scan rate during the stripping step. If properly controlled this
method will be quite sensitive and accurate and is best used for metals that form
amalgams.
Stripping voltammetry:
A voltammetric
technique in which the
analyte is first deposited
on the electrode from a
large volume of the
analyte and then stripped
electrochemically
monitoring the current as
a function of the
potential. Fig. 7.12: Details of stripping voltammetry
22
In anodic stripping voltammetry the determination of the analyte is based on the Voltammetry
anodic dissolution. This process is followed voltammetrically and the peak current
thus obtained is proportional to the potential scan rate and radius of mercury drop
(HMDE) or scan rate and surface area of the mercury film (TMFE). In both the cases
the peak current ip is given by
Where, ip = peak current, n = electron transfer during the oxidation of the analyte,
D = diffusion coefficient, K = constant, CM˚ = concentration of metal in the amalgam,
AF = surface area, υ = scan rate, and tace = accumulation time.
However in both the cases the peak current depends on the accumulation time tace and
therefore, is proportional to the concentration of analyte cM˚(Hg) in the amalgam and
indirectly proportional to the concentration of the analyte in the sample solution.
Peak currents in this technique are a linear function of concentration and hence this
technique can be used for the determination of many metal ions in trace quantities in
environmental samples.
SAQ 6
i) Which of the following is better and why?
a) Linear sweep voltammetry and cyclic voltammetry.
b) Anodic stripping voltammetry and Adsorptive stripping voltammetry.
…………………………………………………………………………………….
……...……………………………………………………………………………..
23
Electroanalytical ii) What are the two basic steps involved in stripping voltammetry?
Methods-III
…………………………………………………………………………………….
……...……………………………………………………………………………..
iii) What is so special about the peak currents in Anodic stripping voltammetry?
…………………………………………………………………………………….
……...……………………………………………………………………………..
……...……………………………………………………………………………..
……...……………………………………………………………………………..
Thus, the height of AC polarographic peak can be used to determine the concentration
in a way similar to DC polarographic method. AC polarographic peak currents are
widely dependent on the kinetics of charge transfer reaction and the signals of strongly
irreversible processes are suppressed. For reversible processes the half width of the
90
peak b½ = (mv) depends on the electron exchange as a result of which, for large
n
value of n, b will be narrower and with slow processes, b will be broader.
24
AC polarographic determinations are limited to a sensitivity of about 10–5 mol/dm3 Voltammetry
because of high capacitance current contribution which is the main source for periodic
charging and discharging of double layer.
The above sensitivity for reversible redox processes can be improved to about (5-10)–7
mol/dm3 if the AC is measured at a particular phase shift with reference to the
excitation signal in order to separate the Faradaic and capacitive current. This is the 1st
harmonic AC polarography (AC-1, polarography). The measurement of the harmonic
of the AC resulting from the nonlinearity of the Faradaic resistance, e.g. 2nd harmonic
with the aid of phase selective rectification (AC - 2 polarography) reduces the
capacitance current contribution thus increasing the sensitivity and the selectivity of
the determination. The market available instruments are equipped to measure both the
types of currents.
SAQ 7
How can the height of AC polarographic peak be utilized?
…………………………………………………………………………………………...
…………………………………………………………………………………………...
…………………………………………………………………………………………...
Computer
25
Electroanalytical ii) Reference electrode: Normally a standard electrode is used whose potential is
Methods-III constant throughout the experiment (saturated calomel electrode or Ag/AgCl
electrode).
iii) Counter electrode, which is a coil of platinum wire or a pool of mercury that
simply serves to conduct electricity from the signal source through the solution
to the microelectrode.
The potentiostat allow to impose the potential scanning between the working and
reference electrodes while the current flowing through the circuit is recorded between
the working and auxiliary electrodes.
The computer CVA Trace analyzer from Metrohm Ltd. can be used as an integrated
instrument with keyboard or a commercially available personal computer interface to
measuring system with software can be used:
i) to send the scanning parameters to the potentiostatic analyzer and check its
correctness and accuracy,
ii) handle the current potential output data.
7.8 SUMMARY
In this unit, you learnt about voltammetry. Various electrodes, especially the common
microelectrodes used in voltammetry were discussed. Then we discussed the common
voltammetry methods. The linear sweep voltammetry, sampled D.C. polarography and
hydrodynamic voltammetry were discussed under the direct current methods. A
detailed discussion of the different pulse methods was done including normal pulse
voltammetry, differential pulse voltammetry, square wave voltammetry and cyclic
voltammetry. Also you have learnt about the various stripping methods including
anodic, cathodic and adsorptive stripping voltammetry. Alternating current methods
were also discussed. In the last section of the unit we learnt about the instrumental
details and steps involved in practical voltammetry.
3. Why are the residual currents less with current sampled polarography?
4. Indium and Cadmium ions in 0.1 M HCl give peaks at potentials – 0.557 V and
– 0.597 V by differential pulse polarographic analysis. When a standard solution
containing 0.4 ppm of indium is analyed, it gave peak current (in arbitrary units)
corresponding to 100.25 at – 0.557 V and 43.75 at – 0.597 V. A standard
solution containing 0.4 ppm of Cd gave ips 29.52 at – 0.557 V and 64.8 at –
0.597 V. What is the concentration of indium and cadmium in a sample if the
peak current is 190 at & 0.557 V and 120 at & 0.597V.
5. Anodic stripping voltammetry is used for the analysis of copper in a sea water
sample using standard addition method. When a 50 ml sample is analysed it
gave a peak current of 0.886 µA. A 5.00 µdm3 spike of 10 ppm Cu++ is added to
the sample and the resulting peak current is 3.52 µA. Calculate the
concentration of Cu++ in ppm in the given sample of sea water.
6. Suggest how Eq. (7.11) could be employed to determine the number of electrons
n involved in a reversible reaction at a microelectrode.
26
7. Draw the DME and label its parts. Voltammetry
8. Why are stripping methods more sensitive than other voltammetric procedures?
14. Why would the HMDE and the DME be unsuitable for use in flow systems?
15. What is usually meant by the terms anodic and cathodic and when is an
electrode an anode or a cathode?
17. Why can stripping voltammetry be applied to only a very small number of
organic compounds?
18. What are the three most important roles played by the separation step?
7.10 ANSWERS
Self Assessment Questions
27
Electroanalytical 4. i) a)
Methods-III
Linear Scan Polarography Pulse Polarography
b)
5. i) The pulse methods are better than the direct methods as the sensitivity is
improved by about 6.5 times by eliminating charging current. The current
in these methods is measured at a time when the difference between the
Faradaic current and the interfering charging current is large.
ii) Polarography is voltammetry at controlled potential using a DME as the
working electrode.
iii) Solid electrode (like platinum) is advantageous over mercury electrode in
CV as they have a better positive potential range than mercury.
iv) CV has become increasingly popular in all fields of chemistry as a means
of studying of redox states. The technique enables a wide potential range
to be scanned rapidly for reducible or oxidizable substances. This
capability together with its variable time scale and good sensitivity make
this the most versatile electrochemical technique.
v) The potential is measured between a working electrode and reference
electrode.
28
Voltammetry
6. i) a) The cyclic voltammetry is better than linear sweep voltammetry as
CV has the capability for rapidly observing the redox behaviour
over a wide range of potential. The resulting voltammogram is
similar to a conventional spectrum in that it conveys information as
a function of an energy scan.
b) The Adsorptive stripping voltammetry is better than the Anodic
stripping voltammetry. In Adsorptive stripping voltammetry
technique analyte is adsorbed on the electrode surface and there is
no electrolysis step and after complete deposition the potential is
scanned in anodic or cathodic direction depending on whether the
analyte is to be oxidized or reduced. The deposition step in this
method amounts to an electrochemical pre-concentration of the
analyte, i.e. the concentration of the analyte on the surface of the
microelectrode is greater than the bulk of solution and hence these
methods are extremely sensitive and the detection limit can be
improved by factors of 106 for selected species.
Terminal Questions
29
Electroanalytical techniques are quite sensitive to experimental conditions which are to be
Methods-III carefully controlled to get accurate and precise results.
3. The residual currents are less with current sampled polarography as current is
measured near the end of the drop.
where k is a constant
Thus for the sample 0.886 = k (con. Cu++) … (1)
and for the standard solution
3.52 = k
0.05 dm3 5.00×10−6 dm3
3 −6 3
(Con . Cu )
++
+ 3 −6 3
(10 ppm ) …(2)
0.05 dm + 5 x 10 dm 0.05 dm + 5×10 dm
Solving the Eq. (1) for k and substituting in Eq. (2) the concentration of Cu++ in
the given sample of sea water comes out to 3.36 × 10– 4 ppm.
6. The number of electrons transferred in the electrode reaction (n) for a reversible
couple can be determined from the separation between peak potentials.
∆Ep = Ep − Ep = 0.059
a c n
30
electrons n involved in a reversible reaction at a microelectrode can be Voltammetry
determined.
7. Fig. 7.1 c
11. Electrode products can often build up on the surface of a solid or stationary
electrode. Species from the solution may adsorb on the surface. Some solid
electrode materials may even develop oxide layers on the surface. All of these
produce changes in the nature of the electrode and often completely distort the
electrochemical response. By regular polishing back to a clean metal or graphite
surface, some attempt can be made to achieve a reproducible surface. Better
results are usually obtained from a smooth rather than a rough surface.
14. They are much too mechanically unstable. The drop can be knocked off far too
easily. They have in fact been used in flow systems but great care and much
patience is needed. Further it is difficult to construct very small cells to hold a
mercury drop electrode − a major disadvantage.
Even in a very evenly stirred solution a slight displacement of the electrode can
change the flow pattern. Reproducible stirring requires careful cell design and
experimentation. The best stirring is a rotating electrode in a reproducible
position in the cell.
31
Electroanalytical 17. Most organic compounds do not form insoluble electrode products. Most
Methods-III organic compounds are soluble at these high dilution trace levels. Many of those
organic electrode products that are insoluble cannot be reoxidised or reduced to
form a stripping peak, or the insoluble layer is insufficiently coherent. The most
common type of organic compound to which stripping analysis can be applied
are those containing a thiol group.
32