Deltabar S PMD70/75, FMD76/77/78: Technical Information
Deltabar S PMD70/75, FMD76/77/78: Technical Information
TI382P/00/en/12.07
No. 71060097
Deltabar S
Table of contents
2 Endress+Hauser
Deltabar S
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . 67
PMD70 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
PMD70 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
PMD75 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
PMD75 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
PMD75 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
FMD76 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
FMD76 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
FMD76 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 74
FMD77 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
FMD77 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
FMD77 (continued) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
FMD78 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
3 Endress+Hauser
Deltabar S
+ –
+ –
With ceramic measuring With metallic measuring With ceramic With metallic With metallic measuring
diaphragms diaphragms measuring diaphragms measuring diaphragms diaphragms and
and diaphragm seal capillary diaphragm
mounted on one side seals
Field of – Flow – Flow – Level – Level – Level
application – Level – Level – Differential pressure
– Differential pressure – Differential pressure
Process – 1/4 – 18 NPT – 1/4 – 18 NPT Low-pressure side (–): Low-pressure side (–): – Wide range of
connections – RC 1/4 – RC 1/4 – 1/4 – 18 NPT – 1/4 – 18 NPT diaphragm seals,
– RC 1/4 – RC 1/4 → see page 5, section
"Overview of diaphragm
High-pressure side (+): High-pressure side (+):
– DN 80 – DN 100 – DN 50 – DN 100 seal FMD78"
– ANSI 3" – 4" – ANSI 2" – 4"
– JIS 80A – 100A – JIS 80A – 100A
Measuring ranges from –25...+25 mbar from –10...+10 mbar from –100...+100 mbar from –100...+100 mbar from –100...+100 mbar
to –3...+3 bar to –40...+40 bar to –3...+3 bar to –16 bar...+16 bar to –40...+40 bar
Overload 1 on one side: max. 100 bar on one side: max. 420 bar on one side: max. 100 bar on one side: max. 160 bar on one side: max. 160 bar
on both sides: max. 150 bar on both sides: max. 630 bar on both sides: max. 240 bar
Process –20...+85°C –40...+120°C –20...+85°C up to + 400°C (+752°F) up to +400°C (+752°F)
temperature range (–4...+185°F) (–40...+248°F) (–4...+185°F)
Ambient –20...+85°C –40...+85°C –20...+85°C –40...+85°C –40...+85°C
temperature range (–4...+185°F) (–40...+185°F) 2 (–4...+185°F) (–40...+185°F) 2 (–40...+185°F) 2
Ambient –40 to +60°C (–40 to +140°F)
temperature range
separate housing
Reference – Up to ±0.075% of the set span – Up to ±0.075 % of the set span
Accuracy – PLATINUM version: up to ±0.05% of the set span
Supply voltage – For non-hazardous areas: 10.5...45 V DC
– EEx ia: 10.5...30 V DC
Output 4...20 mA with superimposed HART protocol, PROFIBUS PA or FOUNDATION Fieldbus
Options – High-pressure version up to pstat 700 bar
– PMD75, FMD77, FMD78: Gold-Rhodium-coated diaphragm, NACE-compliant materials
Specialities – Metal-free measurement – pstat up to 420 bar – Abrasion-resistant and – For high media – Wide range of
(options) with PVDF flange – Diaphragm: tantalum corrosion-resistant temperatures diaphragm seals
– Available with Deltatop/ – Available with Deltatop/ – No diaphragm-seal
Deltaset as flow compact Deltaset as flow compact temperature effects
device device – Metal-free
measurement possible
with ECTFE-coated
process connection
– HistoROM®/M-DAT memory module
1) dependent on the lowest-rated element, with regard to pressure, of the selected components
2) lower temperature on request
4 Endress+Hauser
Deltabar S
P01-FMD78xxx-03-xx-xx-xx-011
P01-FMD78xxx-03-xx-xx-xx-006
Endress+Hauser 5
Deltabar S
Design Diaphr. seal Connection Version Standard Nominal diameter Nom. press./Class
Flange with Membrane ANSI flange ANSI B 16.5 – 3" with 2"/4"/6"/ 150 lbs
extended diaphragm seal 8" extended
diaphragm seal (MDM) diaphragm seal
– 4" with 2"/4"/6"/
P01-FMD78xxx-03-xx-xx-xx-002
8" ext. diaphr. seal
6 Endress+Hauser
Deltabar S
Measuring principle
Ceramic measuring diaphragms used for PMD70 Metallic measuring diaphragms used for PMD75, FMD77 and FMD78
and FMD76
➀
➀ ➁
➁ ➆
➅
p1 p2
➂ ➇
➂ p1 ➃ p2 p1 p2
➈
➃ ➄
➄
P01-xMD7xxxx-03-xx-xx-xx-000 P01-xMD7xxxx-03-xx-xx-xx-002 P01-xMD7xxxx-03-xx-xx-xx-003
Ceramic measuring cell PMD70 and FMD76 Metal measuring cell 10 mbar and 30 mbar Metal measuring cell as of 100 mbar
1 Meter body 1 Sensing element 6 Sensing element
2 Diaphragm 2 Silicon diaphragm 7 Overload diaphragm/Middle diaphragm
3 Electrodes 3 Separating diaphragm 8 Filling oil
4 Glass frit fixes the diaphragm onto the meter 4 Filling oil 9 Separating diaphragm
body 5 Integrated overload protection
5 Temperature sensor
Endress+Hauser 7
Deltabar S
+ – + –
Q ~ Δp p1 p2 Q ~ Δp p1 p2
Q Q
P01-PMD7xxxx-15-xx-xx-xx-000
Flow measurement with Deltabar S and primary element, left: Orifice plate and right: Pitot tube
Q Flow
Δp Differential pressure, Δp = p1 – p2
Your benefits
• Choice of four flow modes of operation: volume flow, norm volume flow (European norm conditions),
standard volume flow (American standard conditions) and mass flow.
• Choice of diverse flow units with automatic unit conversion.
• A customised unit can be specified
• Low flow cut off: when activated, this function suppresses small flows which can lead to large fluctuations
in the measured value.
• Contains two totalizers as standard. One totalizer can be reset to zero.
• The totalizing mode and unit can be individually set for each totalizer. This allows independent daily and
annual quantity totalizing.
• With the product families Deltatop and Deltaset, Endress+Hauser is offering two universal and reliable
solutions for flow measurement:
– Deltatop, the compact, ready-to-use flow measuring unit including differential pressure transmitter
Deltabar S
– Deltaset with differential pressure transmitter Deltabar S, the modular flow measuring unit suitable for
medium temperatures up to 1000°C (1832°F) and pressures up to 400 bar
Note: For more information about flow measurement with the Deltabar S differential pressure transmitter
and orifice plate or pitot tube, see Technical Information TI297P Deltatop/Deltaset.
8 Endress+Hauser
Deltabar S
– – Δp
h= ρ g
➀
h
➂
➁ +
+
P01xMD7xxxx-15-xx-xx-xx-000
Your benefits
• Choice of three level operating modes
• Volume and mass measurements in any tank shapes by means of a freely programmable characteristic curve
• Choice of diverse level units with automatic unit conversion
• A customised unit can be specified
• Has a wide range of uses, e.g.
– for level measurement in tanks with superimposed pressure
– in the event of foam formation
– in tanks with agitators of screen fittings
– in the event of liquid gases
– for standard level measurement
Endress+Hauser 9
Deltabar S
Human interface
On-site display (optional) A 4-line liquid crystal display (LCD) is used for display and operation. The on-site display shows measured
values, dialog text as well as fault and notice messages in plain text, thereby supporting the user in every stage
of operation.
Header line
Main line
Unit
Information
line
– + E Symbol
Bargraph
Editing modes
Operating keys
Selection
options
Value that
can be edited
FOUNDATION Fieldbus
Functions:
• 8-digit measured value display including sign and decimal point, bargraph for current display
• Option for configuring the display according to individual requirements and desires, such as language,
alternating display, display of other measured values such as sensor temperature, contrast setting
• Comprehensive diagnostic functions (fault and warning message)
Header line
Main line
Unit
Information
line
Symbol
Bargraph
P01-xMD7xxxx-07-xx-xx-xx-001
10 Endress+Hauser
Deltabar S
Operating elements With regard to T14 and T15 housings, the operating keys are located either outside the device under the
protection cap or inside on the electronic insert. In T17 housings, the operating keys are always located inside
on the electronic insert.
In addition, devices with an on-site display and a 4 to 20 mA HART or PROFIBUS PA electronic insert have
operating keys on the on-site display.
4...20 mA HART
PROFIBUS PA/
FOUNDATION Fieldbus
0%
Zero
P01-xxxxxxxx-19-xx-xx-xx-038
The operating keys located externally on the device work on the Hall sensor principle. As a result, no additional
openings are required in the device. This guarantees:
• Complete protection against environmental influences such as moisture and contamination
• Simple operation without any tools
• No wear.
➃➄ ➅➆ ➄➅
t ➅ t
Sim.
on
➀ on
➀ ➁ ➂ on
➀ ➁ ➂
➁ ➃
off off 0% off 1 2
➃
0%
Sensor
Zero
Display ➄ Sensor
Zero
Display
Simulation
➂ on on ON CK S D A 0 8
on
HW
1 2 1 2 off off 1 2 3 4 5 6 7 8
PC
1 2 off
PC
PC
1 2 3 4 5 6 7 8
Address
Histo ROM
SW Histo ROM
HW
HART
R
250002271-–
TM
250002272-–
250002273-–
HW-Version: 1 FOUNDATION
SW-Version: 2 HW-Version: 1 HW-Version: 1
SW-Version: 2 SW-Version: 2
Electronic insert HART Electronic insert PROFIBUS PA Electronic insert FOUNDATION Fieldbus
1 Operating keys 1 Green LED to indicate value being accepted 1 Green LED to indicate value being accepted
2 Slot for optional display 2 Key for position calibration and device reset 2 Key for position calibration
3 Slot for optional HistoROM®/M-DAT 3 DIP-switch for bus address 3 Slot for optional display
4 DIP-switch for locking/unlocking measured- 4 Slot for optional display 4 Slot for optional HistoROM®/M-DAT
value-relevant parameters 5 Slot for optional HistoROM®/M-DAT 5 DIP-switch for locking/unlocking measured-
5 DIP-switch for damping on/off 6 DIP-switch for locking/unlocking measured- value-relevant parameters
6 Green LED to indicate value being accepted value-relevant parameters 6 DIP-switch for simulation mode on/off
7 DIP-switch for damping on/off
Endress+Hauser 11
Deltabar S
HistoROM®/M-DAT HistoROM®/M-DAT is a memory module, which is attached to the electronic insert. The HistoROM®/M-DAT
(optional) can be retrofitted at any stage (Order number: 52027785).
Your benefits
• Quick and safe commissioning of the same measuring points by copying the configuration data of one
transmitter to another transmitter
• Reliable process monitoring thanks to cyclical recording of pressure and sensor temperature measured values
• Simple diagnosis by recording diverse events such as alarms, configuration changes, counters for measuring
range undershoot and overshoot for pressure and temperature as well as user limit overshoot and undershoot
for pressure and temperature etc.
• Analysis and graphic evaluation of the events and process parameters via software (contained in scope of
supply).
HistoROM®/M-DAT can be ordered via feature 100 "Additional options 1" or feature 110 "Additional
options 2" or as spare parts. → See also page 67 ff. A CD with the Endress+Hauser operating program is also
included in the scope of delivery.
You can copy data from one transmitter to another transmitter when operating a FOUNDATION Fieldbus
device via an FF configuration program. You need the Endress+Hauser FieldCare operating program and the
Commubox FXA291 service interface and the ToF Adapter FXA291 to be able to access the data and events
saved in the HistoROM®/M-DAT.
Functional Safety SIL/ The Deltabar S differential pressure transmitters with 4...20 mA output signal have been developed to
IEC 61508 Declaration of IEC 61508 standard and have been certified by TÜV SÜD. These devices can be used for flow and differential
conformity (optional) pressure monitoring up to SIL3.
→ For a detailed description of the safety functions with Deltabar S, settings and characteristic quantities for
functional safety, please refer to the "Manual for Safety Manual - Deltabar S" SD189P.
→ For devices with SIL2/IEC 61508 declaration of conformity, see page 68 ff, Feature 100 "Additional
option 1" and Feature 110 "Additional option 2", version E "SIL2/IEC 61508, Declaration of Conformity".
Handheld terminals – HART With a handheld terminal, all the parameters can be configured anywhere along the 4...20 mA line via menu
operation.
Handheld terminal DXR375 – With a handheld terminal DXR375, all the parameters can be configured via menu operation.
FOUNDATION Fieldbus
12 Endress+Hauser
Deltabar S
FieldCare FieldCare is an Endress+Hauser asset management tool based on FDT technology. With FieldCare, you can
configure all Endress+Hauser devices as well as devices from other manufacturers that support the FDT
standard. The following operating systems are supported: WinNT4.0, Win2000 and Windows XP.
FieldCare supports the following functions:
• Configuration of transmitters in offline and online operation
• Loading and saving device data (upload/download)
• HistoROM®/M-DAT analysis
• Documentation of the measuring point
Connection options:
• HART via Commubox FXA191 and the RS 232 C serial interface of a computer
• HART via Commubox FXA195 and the USB port on a computer
• PROFIBUS PA via segment coupler and PROFIBUS interface card
• Service interface with adapter Commubox FXA291 and ToF Adapter FXA291 (USB).
Remote operation – An FF configuration program is required to integrate a device with "FOUNDATION Fieldbus signal" into an FF
FOUNDATION Fieldbus network or to set the FF-specific parameters. Please contact your local Endress+Hauser Sales Center for more
information.
Commubox FXA291 The Commubox FXA291 connects Endress+Hauser field instruments with CDI interface (= Endress+Hauser
Common Data Interface) to the USB interface of a personal computer or a notebook. For details refer to
TI405C/07/en.
! Note!
For the following Endress+Hauser instruments you need the "ToF Adapter FXA291" as an additional accessory:
• Cerabar S PMC71, PMP7x
• Deltabar S PMD7x, FMD7x
• Deltapilot S FMB70
ToF Adapter FXA291 The ToF Adapter FXA291 connects the Commubox FXA291 via the USB interface of a personal computer or
a notebook. For details refer to KA271F/00/a2.
Endress+Hauser 13
Deltabar S
Input
Measured variable Differential pressure, from which flow (volume or mass current) and level (level, volume or mass) are derived
Nominal Measurement Smallest MWP 1 Overload 2 Min. operating pressure 3 Versions in the
value limit calibratable order code 4
span 5
Nominal Measurement Smallest MWP 1 Overload 2 Min. operating pressure 3 Versions in the
value limit calibratable order code 4
span 5
1) The MWP (maximum working pressure; MWP = PN) for the measuring device depends on the weakest element of the components selected with regard to
pressure, i.e. the process connection (→ see page 34 ff) has to be taken into consideration in addition to the measuring cell (→ see table above). Also observe
pressure-temperature dependency. For the appropriate standards and further information, see page 33, "Pressure specifications" section.
2) The maximum pressure for the measuring device is dependent on the lowest-rated element, with regard to pressure, of the selected components. → See also
page 33, section "Pressure specifications".
3) The minimum operating pressure indicated in the table applies to silicone oil under reference operating conditions.
Min. operating pressure at 85°C (185°F) for silicone oil: 10 mbarabs.
FMD77 and FMD78: Min. operating pressure: 50 mbarabs; observe also the pressure and temperature application limits of the selected filling oil on page 58.
For vacuum applications, please observe the installation instructions on Page 65 ff.
4) Versions in the order code → See also page 70 ff, feature 40 "Nominal range; PN"
5) Turn down > 100:1 on request
6) PN 160 versions with stainless steel A2 screws, PN 420 versions with stainless steel A4 M12 screws
PN 420 versions for PMD75 only.
7) PMD75 only
8) minimum span that can be calibrated for PMD75: 1 mbar; minimum span that can be calibrated for FMD77 and FMD78: 5 mbar
9) For PMD75 with CRN-approved process connections, the MWP is 315 bar.
10) "–" side: 100 bar
14 Endress+Hauser
Deltabar S
Explanation of terms Explanation of the terms: Turn down (TD), set span and zero based span
Case 1:
• ⏐Lower range value⏐ ≤ ⏐Upper range value⏐
Example:
• Lower range value (LRV) = 0 mbar
• Upper range value (URV) = 100 mbar
• Nominal value (URL) = 500 mbar
Turn down:
• TD = URL / ⏐URV⏐ = 5:1
set span:
• URV – LRV = 100 mbar
This span is based on the zero point.
P01-xxxxxxxx-05-xx-xx-xx-001
Case 2:
• ⏐Lower range value⏐ ≥ ⏐Upper range value⏐
Example:
• Lower range value (LRV) = –300 mbar
• Upper range value (URV) = 0 bar
• Nominal value (URL) = 500 mbar
Turn down:
• TD = URL / ⏐(LRV)⏐ = 1,67:1
set span:
• URV – LRV = 300 mbar
This span is based on the zero point.
P01-xMD7xxxx-05-xx-xx-xx-007
Endress+Hauser 15
Deltabar S
Output
Output signal • 4...20 mA with superimposed digital communication protocol HART 5.0, 2-wire
• Digital communication signal PROFIBUS PA (Profile 3.0)
• Digital communication signal FOUNDATION Fieldbus
847 804
➃ ➃
413
➂ 369
➂
10.5 20 30 40 45 U 11.5 20 30 40 45 U
[V] [V]
U – 10.5 V U – 11.5 V
RLmax ≤ RLmax ≤
23 mA 23 mA
P01-xMD7xxxx-05-xx-xx-xx-000
Load diagram, observe the position of the jumper and the explosion protection (→ See also page 22, section "Taking 4...20
mA test signal".)
1 Jumper for 4...20 mA test signal inserted in "Non-test" position
2 Jumper for 4...20 mA test signal inserted in "Test" position
3 Supply voltage 10,5 (11,5)...30 V DC for 1/2 D, 1 GD, 1/2 GD, FM IS, CSA IS, IECEx ia, NEPSI Ex ia
4 Supply voltage 10,5 (11,5)...45 V DC for device for non-hazardous areas, 1/2 D, 1/3 D, 2 G EEx d, 3 G EEx nA,
FM XP, FM DIP, FM NI, CSA XP, CSA Dust-Ex, NEPSI Ex d
RLmax Maximum load resistance
U Supply voltage
Note!
When operating via a handheld terminal or via PC with an operating program, a minimum communication
resistance of 250 Ω must exist within the loop.
16 Endress+Hauser
Deltabar S
100 %
63 %
t1 t2 t
P01-xxxxxxxx-05-xx-xx-xx-007
Reading cycle
• HART commands: on average 3 to 4 per second
Response time
≤ 250 ms
Endress+Hauser 17
Deltabar S
Reading cycle
cyclic:
• max.: 100/s
• typical value: 20/s
acyclic:
• max.: 20/s
• typical value: 10/s
Response time
• cyclic: approx. 10 ms per request
• acyclic: < 50 ms
All values are typical values.
18 Endress+Hauser
Deltabar S
Reading cycle
• cyclic: up to 5/s, dependent on the number and type of function blocks used in a closed-control loop
• acyclic: 10/s
Response time
• cyclic: < 80 ms
• acyclic: < 40 ms
All values are typical values.
Endress+Hauser 19
Deltabar S
Power supply
Electrical connection Note!
• When using the measuring device in hazardous areas, installation must comply with the corresponding
national standards and regulations and the Safety Instructions or Installation or Control Drawings. → See
also page 81, section "Safety Instructions" and "Installation/Control Drawings".
• Devices with integrated overvoltage protection must be earthed. → See also page 31.
• Protective circuits against reverse polarity, HF influences and overvoltage peaks are installed.
4...20 mA HART
➀ ➅ 10.5 V DC
➆ 11.5 V DC
4…20 mA
Test
Test
➂ ➄
➇
4... 20mA Test
➃
P01-xMx7xxxx-04-xx-xx-xx-001
PROFIBUS PA
The two-wire cable must be connected to the "PA+" and "PA–" terminals.
FOUNDATION Fieldbus
The two-wire cable must be connected to the "FF+" and "FF–" terminals.
Deltabar S
+
7
+ – – 8 6
1
2 5
3 4
Han7D
+ –
P01-xMD7xxxx-04-xx-xx-xx-000
20 Endress+Hauser
Deltabar S
Deltabar S
+ –
– +
M12
+ –
P01-xMD7xxxx-04-xx-xx-xx-008
Endress+Hauser offers for devices with M12 plug the following accessories:
Plug-in jack M 12x1, straight
• Material: Body PA; coupling nut CuZn, nickel-plated
• Degree of protection (fully locked): IP67
• Order number: 52006263
Plug-in jack M 12x1, elbowed
• Material: Body PBT/PA; coupling nut GD-Zn, nickel-plated
• Degree of protection (fully locked): IP67
• Order number: 51006327
Cable 4x0.34 mm2 with M12 socket, elbowed, screw plug, 5 m length
• Material: Body PUR; coupling nut CuSn/Ni; cable PVC
• Degree of protection (fully locked): IP67
• Order number: 52010285
Deltabar S
–
+ –
7/8" +
+ –
P01-xMD7xxxx-04-xx-xx-xx-009
Endress+Hauser 21
Deltabar S
4...20 mA HART
• Version for non-hazardous areas, jumper for 4...20 mA test signal in "Test" position (delivery status):
11.5...45 V DC
• Version for non-hazardous areas, jumper for 4...20 mA test signal in "Non-test" position: 10.5...45 V DC
PROFIBUS PA
• Version for non-hazardous areas: 9...32 V DC
FOUNDATION Fieldbus
• Version for non-hazardous areas: 9...32 V DC
Current consumption • PROFIBUS PA: 11 mA ± 1 mA, switch-on current corresponds to IEC 61158-2, Clause 21
• FOUNDATION Fieldbus: 14 mA ± 1 mA, switch-on current corresponds to IEC 61158-2, Clause 21
Cable entry → See also page 67 ff, feature 30 "Housing, Cable entry, Protection".
Residual ripple Without influence on 4...20 mA signal up to ± 5 % residual ripple within the permitted voltage range [according
to HART hardware specification HCF_SPEC-54 (DIN IEC 60381-1)]
22 Endress+Hauser
Deltabar S
Note!
Position-dependent zero shift can be corrected. → See also page 27, section "General installation instructions"
and page 63 ff, section "Installation instructions, Diaphragm seal systems".
Endress+Hauser 23
Deltabar S
PMD75
10 mbar, 30 mbar measuring cells:
• TD 1:1: ±0.15% of the set span
• TD > 1:1: ±0.15% of the set span x TD
100 mbar measuring cell:
• TD 1:1 to TD 4:1: ±0.075% of the set span
• TD > 4:1: ±(0.012 x TD + 0.027)% of the set span
Measuring cells ≥ 500 mbar:
• TD 1:1 to TD 15:1: ±0.075% of the set span
• TD > 15:1: ±(0.0015 x TD + 0.053)% of the set span
Platinum version,
Measuring cells ≥ 100 mbar:
• TD 1:1: ±0.05 % of the set span
Total performance – PMD75 The "Total performance" specification comprises the non-linearity including hysteresis, non-reproducibility, the
thermal change of the zero point as well as the influence of the line pressure (pst = 70 bar).
• AISI 316L/1.4435, Alloy, Gold-Rhodium coated or Monel membrane:
±0.15 % of the set span 1, 2
• Tantal membrane:
±0.30 % of the set span 1, 2
Total Error The total error comprises the long-term stability and the total performance:
AISI 316L/1.4435, Alloy, Gold-Rhodium coated or Monel membrane:
• 10 mbar, 30 mbar and 100 mbar measuring cell = 0,33% of URL/year
• ≥ 500 mbar measuring cell = 0,20% of URL/year
Tantal membrane:
• 10 mbar, 30 mbar and 100 mbar measuring cell = 0,48% of URL/year
• ≥ 500 mbar measuring cell = 0,35% of URL/year
24 Endress+Hauser
Deltabar S
Influence of the operating Measuring AISI 316L/1.4435, Alloy, Gold-Rhodium Tantal membrane
pressure on zero point and cell coated or Monel membrane
span – PMD75, FMD77,
FMD78 Influence of the Influence of the Influence of the Influence of the
operating pressure on operating pressure on operating pressure on operating pressure on
the zero point the span the zero point the span
10 mbar ±0.15 % of URL/7 bar ±0.035 % of URL/7 bar ±0.28 % of URL/7 bar ±0.28 % of URL/7 bar
30 mbar ±0.35 % of URL/70 bar ±0.14 % of URL/70 bar ±0.70 % of URL/70 bar ±0.70 % of URL/70 bar
100 mbar ±0.15 % of URL/70 bar ±0.14 % of URL/70 bar ±0.42 % of URL/70 bar ±0.42 % of URL/70 bar
500 mbar ±0.075 % of URL/70 bar ±0.14 % of URL/70 bar ±0.14 % of URL/70 bar ±0.14 % of URL/70 bar
3 bar ±0.075 % of URL/70 bar ±0.14 % of URL/70 bar ±0.14 % of URL/70 bar ±0.14 % of URL/70 bar
16 bar ±0.075 % of URL/70 bar ±0.14 % of URL/70 bar ±0.14 % of URL/70 bar ±0.14 % of URL/70 bar
40 bar ±0.075 % of URL/70 bar ±0.14 % of URL/70 bar ±0.14 % of URL/70 bar ±0.14 % of URL/70 bar
Note!
The influence of the operating pressure on the zero point can be calibrated out.
Endress+Hauser 25
Deltabar S
Total performance – The "Total performance" specification comprises the non-linearity including hysteresis, non-reproducibility, the
PMD70, FMD76 thermal change of the zero point as well as the influence of the line pressure (pst = 70 bar).
• ±0.15% of the set span 1, 2
Total Error The total error comprises the long-term stability and the total performance:
• 25 mbar and 100 mbar measuring cell = 0,33% of URL/year
• 500 mbar and 3 bar measuring cell = 0,20% of URL/year
Note!
The influence of the operating pressure on the zero point can be calibrated out.
26 Endress+Hauser
Deltabar S
Level measurement
• PMD70, PMD75, FMD76 and FMD77 are best suited to level measurement in open tanks. All Deltabar S
devices are suitable for level measurement in closed tanks.
Measuring arrangement level measurement in closed tanks and closed tanks with superimposed vapour
• PMD70, PMD75: Mount device below the lower measuring connection. Always connect the negative side
above the maximum level.
• FMD76, FMD77: Mount device direct on the tank. Always connect the negative side above the maximum
level.
• In the case of level measurement in closed tanks with superimposed vapour, a condensate trap ensures
pressure which remains constant on the minus side.
Pressure measurement
• The PMD70 and PMD75 are best suited to differential pressure measurement.
• Measuring arrangement for gases: Mount device above the measuring point.
• Measuring arrangement for liquids and steams: Mount device below tapping point.
• For differential pressure measurement in vapour, mount the condensate traps at the same level as the same
the tapping point and at the same distance from Deltabar S.
Endress+Hauser 27
Deltabar S
Heat insulation – FMD77 The FMD77 must only be insulated up to a certain height. The maximum permitted insulation height is labelled
on the devices and applies to an insulation material with a heat conductivity
≤ 0.04 W/(m x K) and to the maximum permitted ambient and process temperature (→ see table below). The
data were determined under the most critical application "quiescent air".
TA λ ≤ 0.04 W
m•K
TP
P01-FMD77xxxx-11-xx-xx-xx-000
FMD77
Wall- and pipe-mounting Endress+Hauser offers a mounting bracket for installing the device on pipes or walls. → See also page 67 ff,
feature 110, "Additional options 2".
Note!
If a valve block is used, its dimensions should also be taken into consideration.
➀
37.5
97.5
135
74
12
4
74
106
P01-xMD7xxxx-11-xx-xx-xx-008
28 Endress+Hauser
Deltabar S
"Separate housing" version With the "separate housing" version, you are able to mount the housing with the electronics insert at a distance
from the measuring point. This facilitates trouble-free measurement:
• Under particularly difficult measuring conditions (at installation locations that are cramped or difficult to
access)
• If rapid cleaning of the measuring point is required
• If the measuring point is exposed to vibrations.
IP 69K applies for the process connection with sensor if FEP cable are used.
You can choose between different cable versions:
• PE (2 m, 5 m and 10 m)
• FEP (5 m).
→ See also Page 68 ff, Feature 110, "Additional options 2", Version "G".
→ For the dimensions, see Page 54.
r ³ 120 mm
➀
➁
P01-xMD7xxxx-11-xx-xx-xx-010
In the case of the "separate housing" version, the sensor is delivered with the process connection and cable ready
mounted. The housing and a mounting bracket are enclosed as separate units. The cable is provided with a socket at both
ends. These sockets are simply connected to the housing and the sensor.
1 Process connection with sensor
2 Cable, both ends are fitted with a socket
3 Mounting bracket provided, suitable for pipe and wall mounting
4 Housing with electronic insert
Turn the housing The housing can be rotated up to 380° by loosening the Allen screw.
Your benefits
• Simple mounting by optimally aligning the housing
• Good, accessible device operation
• Optimum readability of the on-site display (optional).
Endress+Hauser 29
Deltabar S
max. 380°
P01-xMD7xxxx-17-xx-xx-xx-001
Oxygen applications Oxygen and other gases can react explosively to oils, grease and plastics, such that, among other things, the
following precautions must be taken:
– All components of the system, such as measuring devices, must be cleaned in accordance with the BAM
(DIN 19247) requirements.
– Dependent on the materials used, a certain maximum temperature and a maximum pressure for oxygen
applications must not be exceeded.
The devices suitable for gaseous oxygen applications are listed in the following table with the specification pmax.
Order code for devices cleaned for pmax for oxygen applications Tmax for oxygen
oxygen applications applications
Ultra pure gas applications Endress+Hauser also offers devices for special applications, such as ultra pure gas, cleaned from oil and grease.
No special restrictions regarding the process conditions apply to these devices.
→ See also page 67 ff, PMD70 and PMD75: feature 80 "Seal", FMD76 and FMD77: feature 70 "Process
connection low-pressure side, material, seal".
Diaphragms for materials with With regard to materials in which hydrogen build-up takes place, hydrogen atoms can diffuse through the metal
hydrogen build-up (Gold- diaphragms. This can result in incorrect measurement results.
Rhodium coating) Endress+Hauser offers diaphragms with Gold-Rhodium coating for this application.
→ See also page 70 "Ordering information PMD75", page 76 "Ordering information FMD77" or page 79
"Ordering information FMD78", feature 60 "Membrane material".
30 Endress+Hauser
Deltabar S
Ambient temperature range • PMD75, FMD77, FMD78: –40...+85°C (–40...+185°F), devices for lower temperatures on request
• PMD70, FMD76: –20...+85°C (–4...+185°F)
• On-site display: –20...+70°C (–4...+158°F)
Extended temperature application range with restrictions in optical properties such as display speed and
contrast: –40...+85°C (–40...+185°F)
• Separate housing: –40 to +60°C (–40 to +140°F)
For devices for use in hazardous areas, see Safety instructions, Installation or Control Drawing (→ See also page
81, sections "Safety Instruction" and "Installation/Control drawings").
The device can be used in this temperature range. The values of the specification, such as thermal change, may
be exceeded. → See also DIN 16086.
Degree of protection • → See page 67 ff, feature 30 "Housing, Cable entry, Protection".
• Degree of protection IP 68 for T17 housing: 1.83 mH2O for 24 h
• Degree of protection IP 69K for process connection with sensor with separate housing and FEP cable.
Climate class Class 4K4H (air temperature: –20...55°C/–4...+131F, relative humidity: 4...100%) fulfilled as per
DIN EN 60721-3-4 (condensation possible)
Electromagnetic compatibility • Electromagnetic compatibility to EN 61326 and NAMUR recommendation EMC (NE21). For details refer
to the declaration of conformity.
• With enhanced immunity against electromagnetic fields as per EN 61000-4-3:
30 V/m with closed cover (for devices with T14 or T15 housing)
• Maximum deviation: < 0.5% of span 1
• All EMC measurements were performed with a turn down (TD) = 2:1.
1) Larger deviations possible with PMD70 with 25 mbar or 100 mbar sensor
Endress+Hauser 31
Deltabar S
Versions for feature 70 in the order code Seal Temperature operating range
32 Endress+Hauser
Deltabar S
Versions for feature 70 in the Seal on the LP side (–) Process temperature range 1
order code
Pressure specifications • The maximum pressure for the measuring device is dependent on the lowest-rated element with regard to
pressure, see the following sections for this:
– → page 14 ff, section "Measuring range"
– → chapter "Mechanical construction".
The MWP (maximum working pressure) is specified on the nameplate. This value refers to a reference
temperature of 20°C (68°F) or 100°F for ANSI flanges and may be applied to the device for an unlimited
time. Observe pressure-temperature dependency.
• The pressure values permitted at higher temperatures can be found in the following standards:
– EN 1092-1: 2001 Tab. 18 1
– ASME B 16.5a – 1998 Tab. 2-2.2 F316
– ASME B 16.5a – 1998 Tab. 2.3.8 N10276
– JIS B 2220
• For PMD70 and PMD75, the MWP applies for the temperature ranges specified in the "Ambient temperature
range" (→ page 31) and "Process temperature limits" (→ page 32) sections.
• The test pressure corresponds to the over pressure limit of the measuring instrument (Over pressure limits
OPL = 1.5 x MWP) and may fit only temporally limited, so that no permanent damage develops.
• The Pressure Equipment Directive (EC Directive 97/23/EC) uses the abbreviation "PS". The abbreviation
"PS" corresponds to the MWP (maximum working pressure) of the measuring device.
• In the case of sensor range and process connections where the OPL (Over Pressure Limit) of the pressure
connection is smaller than the nominal value of the sensor, the device is set at the factory, at the very
maximum, to the OPL value of the process connection. If you want to use the entire sensor range, select a
process connection with a higher OPL value (1.5 x PN; PN = MWP).
• In oxygen applications, the values for "pmax and Tmax for oxygen applications" as per page 30, "Oxygen
applications" may not be exceeded.
1) With regard to its stability property, the material 1.4435 is identical to 1.4404 which is grouped under 13EO
in EN 1092-1 Tab. 18. The chemical composition of the two materials can be identical.
Endress+Hauser 33
Deltabar S
Mechanical construction
Housing dimensions T14,
optional display on the side 152
111 111
FIELD TERMINALS
H
P01-xMD7xxxx-06-00-xx-xx-000
P01-xMD7xxxx-06-00-xx-xx-001
P01-xMD7xxxx-06-00-xx-xx-002
34 Endress+Hauser
Deltabar S
+ – – +
7/16-20 UNF +
M10
41.3
108
H
1/4-18 NPT
RC1/4
54 96
84
P01-PMD70xxx-06-09-xx-xx-000
B 1/4-18 NPT IEC 61518 7/16-20 UNF Steel C 22.8 2 vent valves (AISI 316L/1.4404) 4.0 kg
2
included
D 1/4-18 NPT IEC 61518 7/16-20 UNF AISI 316L 4.0 kg
F 1/4-18 NPT IEC 61518 7/16-20 UNF Alloy C276 3 Vent valves (Alloy C276/2.4819), 4.2 kg
see page 68 feature 110
"Additional options 2".
U RC 1/4 7/16-20 UNF AISI 316L 2 2 vent valves (AISI 316L/1.4404) 4.0 kg
included
1 1/4-18 NPT IEC 61518 PN 160: M10 Steel C 22.8 4.0 kg
2
2 1/4-18 NPT IEC 61518 PN 160: M10 AISI 316L 4.0 kg
3 1/4-18 NPT IEC 61518 PN 160: M10 Alloy C276 3 Vent valves (Alloy C276/2.4819), 4.2 kg
see page 68, feature 110
"Additional options 2".
Endress+Hauser 35
Deltabar S
H
108
41.3
1/4-18 NPT
54 96
84
P01-PMD70xxx-06-09-xx-xx-001
Process connection PMD70, version G, PVDF inlay, PN = 10 bar, process temperature T = –10...+60°C (14...+140°F)
Device height H
36 Endress+Hauser
Deltabar S
7/16-20 UNF
M 10
H
41.3
+ – – +
85
1/4-18 NPT
RC1/4
54 106
85
100
7/16-20 UNF
M10 (M12)
H
41.3
72
+– 1/4-18 NPT –+
RC1/4
54 100
87
98
P01-PMD75xxx-06-09-xx-xx-005
B 1/4-18 NPT IEC 61518 7/16-20 UNF Steel C 22.8 2 vent valves (AISI 316L/1.4404) 4.2 kg
2
included
D 1/4-18 NPT IEC 61518 7/16-20 UNF AISI 316L 4.2 kg
3
F 1/4-18 NPT IEC 61518 7/16-20 UNF Alloy C276 Vent valves (Alloy C276/2.4819), 4.5 kg
see page 71, feature 110
"Additional options 2".
U RC 1/4 7/16-20 UNF AISI 316L 2 2 vent valves (AISI 316L/1.4404) 4.2kg
included
1 1/4-18 NPT IEC 61518 – PN 160: M10 Steel C 22.8 4.2 kg
– PN 420: M12
3 1/4-18 NPT IEC 61518 – PN 160: M10 Alloy C276 3 Vent valves (Alloy C276/2.4819), 4.5 kg
– PN 420: M12 see page 71, feature 110
"Additional options 2".
1) Weight of process connections without vent valves with 10 mbar or 30 mbar sensors, process connections without
vent valves with sensors ≥ 100 mbar weght approx. 800 g less. Housing weight see page 55.
2) AISI 316L/1.4435 or 1.4404
3) Alloy C276/2.4819
Endress+Hauser 37
Deltabar S
Process connections PMD75 Oval flange, connection 1/4-18 NPT or RC 1/4, with side vent
with metallic measuring
diaphragms (continued)
41.3
+ – – +
85
42
1/4-18 NPT
RC1/4
54 106
85
100
P01-PMD75xxx-06-09-xx-xx-004
7/16-20 UNF
1/4-18 NPT
H
41.3
28
72
+– 1/4-18 NPT –+
RC1/4
54 100
87
98
P01-PMD75xxx-06-09-xx-xx-003
C 1/4-18 NPT IEC 61518 7/16-20 UNF Steel C 22.8 4 locking screws and 4.2 kg
2 vent valves included
E 1/4-18 NPT IEC 61518 7/16-20 UNF AISI 316L 2 4.2 kg
(AISI 316L/1.4404)
H 1/4-18 NPT IEC 61518 7/16-20 UNF Alloy C276 3 Vent valves (Alloy C276/2.4819), 4.5 kg
see page 71, feature 110
"Additional options 2".
1) Weight of process connections without vent valves with 10 mbar or 30 mbar sensors, process connections without
vent valves with sensors ≥ 100 mbar weght approx. 800 g less. Housing weight see page 55
2) PN 160 bar measuring cells: AISI 316L/1.4435, PN 420 bar measuring cells: AISI 316L/1.4435 or 1.4404
3) Alloy C276/2.4819
38 Endress+Hauser
Deltabar S
Process connections PMD75 Oval flange, prepared for diaphragm seal mount
with metallic measuring
diaphragms (continued)
7/16-20 UNF
H
➁
➀
➂
41.3
72
+– M6 –+
54 54
87
98
P01-PMD75xxx-06-09-xx-xx-002
Left: Process connection PMD75, version W, prepared for diaphragm seal mount
Right: Position of the copper ring seal
H Device height → see the following section "Device height H"
1 Diaphragm seal attachment
2 Copper ring seal
3 Cup diaphragm
Device height H
T15 housing, optional display on the top, flat cover 223 mm (236 mm)
T15 housing, optional display on the top, high cover 234 mm (247 mm)
1) Values for devices with 10 mbar and 30 mbar measuring cell in brackets
Endress+Hauser 39
Deltabar S
41
41.3
7/16-20 UNF
H
h
g2
1/4-18 NPT
2
ø46
b
g
k
D
P01-FMD76xxx-06-09-xx-xx-000
Flange Boltholes
1
Version Material Nominal Shape Nominal Diameter Thickness Raised Quantity Diameter Hole Flange
diameter pressure face circle weight 2
D b g g2 k
40 Endress+Hauser
Deltabar S
Process connection FMD76 ANSI flanges, connection dimensions as per ANSI B 16.5, raised face RF and
with ceramic measuring JIS flanges, connection dimensions as per JIS B 2220, raised face RF
diaphragms (continued)
41
1.61 in
41.3
1.63 in
7/16-20 UNF
H
h
g2
1/4-18 NPT
ø46
b
ø1.81 in
2 (0.078 in)
g
k
ANSI
JIS
D
P01-FMD76xxx-06-09-xx-xx-001
Process connection FMD76, high-pressure side: ANSI or JIS flange (see table below),
low-pressure side: connection 1/4-18 NPT
H Device height → see page 42, section "Device height H, devices with flange"
h Height of the device without flange thickness b
Flange Boltholes
Version Material Nominal Class/ Diameter Thickness Raised face Quantity Diameter Hole circle Flange
diameter Nominal weight 1
pressure
D b g g2 k
R ECTFE 3 4.9
U ECTFE 3 7.1
V Alloy C276 8
JIS flanges
Endress+Hauser 41
Deltabar S
T14 housing, optional display on the side 175 mm + flange thickness b (see tables)
T15 housing, optional display on the top, flat cover 181 mm + flange thickness b (see tables)
T15 housing, optional display on the top, high cover 192 mm + flange thickness b (see tables)
T17 housing, optional display on the side 191 mm + flange thickness b (see tables)
41.3
+– –+
1/4-18 NPT
150
103.7
P01-FMD77xxx-06-xx-xx-xx-001
42 Endress+Hauser
Deltabar S
H
h
g2
b
dM
1.6 mm
0.06 in
L
g
k d3
D g
P01-FMD77xxx-06-09-xx-xx-002
Process connection FMD77, high-pressure side EN/DIN flange with and without extended diaphragm seal, material
AISI 316L
H Device height see page 45, → section "Device height H"
h Height of the device without flange thickness b
Ver- No- Nominal Shape Dia- Thick- Raised Exten- Exten- Quan- Dia- Hole max. TK TK Flange
1
sion minal pressure meter ness face sion sion tity meter circle Dia- Am- Pro- weight 2
dia- length dia- phragm bient cess
meter meter dia-
meter
D b g L d3 g2 k dM
100 6.7
200 7.8
Endress+Hauser 43
Deltabar S
Process connections FMD77 ANSI flanges, connection dimensions as per B 16.5, raised face RF
with metallic measuring
diaphragms, high-pressure
side (continued)
H
h
g2
b
1.6 mm dM
0.06 in
L
g
k d3
D g
P01-FMD77xxx-06-09-xx-xx-000
Process connection FMD77, high-pressure side ANSI flange with and without extended diaphragm seal, material
AISI 316/316L
H Device height → see page 45, section "Device height H"
h Height of the device without flange thickness b
D b g L d3 g2 k dM
[lb./sq.in] [in] [in] [in] [in] [in] [in] [in] [in] [mbar/10 K] [kg]
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm]
4 6.6
101.6
6 7.1
152.4
8 7.7
203.8
44 Endress+Hauser
Deltabar S
Process connections FMD77 JIS flanges, connection dimensions as per JIS B 2220, raised face RF
with metallic measuring
diaphragms, high-pressure
side (continued)
H
h
0.0787"
g2
2 mm
b
dM
g
k
D
P01-FMD77xxx-06-09-xx-xx-001
Process connection FMD77, high-pressure side, JIS flange, material AISI 316L/1.4435
H Device height → see the following section "Device height H"
h Height of the device without flange thickness b
Ver- Nominal Nominal Diameter Thick- Raised Quan- Diameter Hole max. Dia- TK TK Flange
sion diameter pressure ness face tity circle phragm Ambient Process weight 1
diameter
D b g g2 k dM
Device height H
T14 housing, optional display on the side 325 mm + flange thickness b (see tables)
T15 housing, optional display on the top, flat cover 331 mm + flange thickness b (see tables)
T15 housing, optional display on the top, high cover 342 mm + flange thickness b (see tables)
T17 housing, optional display on the side 341 mm + flange thickness b (see tables)
Endress+Hauser 45
Deltabar S
P01-FMD78xxx-06-xx-xx-xx-000
Device height H
46 Endress+Hauser
Deltabar S
b
dM
P01-FMD78xxx-06-09-xx-xx-000
D b dM
D b dM
1) The specified nominal pressure applies to the diaphragm seal. The maximum pressure for the measuring device is dependent on the lowest-rated element,
with regard to pressure, of the selected components. → See also page 33, section "Pressure specifications".
Endress+Hauser 47
Deltabar S
Process connection FMD78 Threaded ISO 228 G 1/2 B and ANSI 1/2 MNPT, separator with PTFE seal
with metallic measuring
diaphragms (continued)
100 100
90
89
➀ ➀
SW22 SW22
22 AF 22 AF
6 1/2 NPT
G½
P01-FMD78xxx-06-09-xx-xx-012
Process connection FMD78, left: with threaded connection ISO 228 G 1/2 B, right: with threaded connection
ANSI 1/2 MNPT
1 PTFE seal as standard max. 260 °C/500 °F (higher temperatures on request)
[mbar/10 K] [kg]
dM
C7
P01-FMD78xxx-06-09-xx-xx-005
Process connection FMD78, Material: AISI 316L/1.4435, surface roughness of the wetted surfaces ≤ 0.8 μm as standard.
Lower surface roughness on request.
Version Nominal Nominal Nominal Diameter max. Height TK Process Weight of two
diameter diameter diameter Diaphragm diaphragm
ISO 2852 DIN 32676 diameter seals
C7 dM h
TD DN 51 DN 50 2 64 48 30 +1.00 2.2
48 Endress+Hauser
Deltabar S
d1
D
d2
L
P01-FMD78xxx-06-09-xx-xx-001
Process connection FMD78, Material: AISI 316L, surface roughness of the wetted surfaces ≤ 0.8 μm as standard. Lower
surface roughness on request.
Version Nominal Nominal Diameter Diameter Diameter Height Face-to-face TK Process Weight of two
diameter diameter length diaphragm
ISO 2852 seals
D d1 d2 h L
[in] [mm] [mm] [mm] [mm] [mm] [mbar/10 K] [kg]
1) Including 3.1 and pressure test as per Pressure Equipment Directive, category II
dM max. = 50
ø68
P01-FMD78xxx-06-09-xx-xx-006
Process connection FMD78, surface roughness of the wetted surfaces ≤ 0.8 μm as standard. Lower surface roughness on
request.
[mbar/10 K] [kg]
Endress+Hauser 49
Deltabar S
41
23.6
dM max. =50 4 x ø11.5
ø65–1.2
ø84
ø105
P01-FM78xxx-06-09-xx-xx-002
Process connection FMD78, surface roughness of the wetted surfaces ≤ 0.8 μm as standard. Lower surface roughness on
request.
[mbar/10 K] [kg]
53.5
ø91.45
ø100
ø132.7
P01-FMD78xxx-06-09-xx-xx-011
Process connection FMD78, surface roughness of the wetted surfaces ≤ 0.8 μm as standard. Lower surface roughness on
request.
[mbar/10 K] [kg]
50 Endress+Hauser
Deltabar S
Process connection FMD78 Taper adapter with coupling nut, DIN 11851 (dairy fitting)
with metallic measuring
diaphragms (continued)
40
f
m
k
dM
D
G
P01-FMD78xxx-06-09-xx-xx-007
Process connection FMD78, material AISI 316L/1.4435, surface roughness of the wetted surfaces ≤ 0.8 μm as standard.
Lower surface roughness on request.
Ver- Nominal Nominal Diameter Adapter Thread Height Height max. TK Process Weight of
sion diameter pressure height Diaphragm two
diameter diaphragm
seals
D f G k m dM
dM
d1
G
P01-FMD78xxx-06-09-xx-xx-008
Process connection FMD78, material AISI 316L/1.4435, surface roughness of the wetted surfaces ≤ 0.8 μm as standard.
Lower surface roughness on request.
Version Nominal Nominal Diameter Height Thread max. TK Process Weight of two
diameter pressure Diaphragm diaphragm
diameter seals
d1 h G dM
Endress+Hauser 51
Deltabar S
Process connection FMD78 EN/DIN flanges, connection dimensions as per EN 1092-1/DIN 2527
with metallic measuring JIS flanges, connection dimensions as per JIS B 2220
diaphragms (continued)
g2
b
dM
2
g
k
D
P01-FMD78xxx-06-09-xx-xx-009
Ver- Nominal Nominal Shape 1 Dia- Thick- Raised Quan- Dia- Hole max. TK Weight of
sion diameter pressure meter ness face tity meter circle Diaphragm Process two
diameter diaphragm
seals
D b g g2 k dM
Ver- Nominal Nominal Diameter Thickness Raised Quantity Diameter Hole max. TK Weight of
sion diameter pressure face circle Diaphragm Process two
diameter diaphragm
seals
D b g g2 k dM
52 Endress+Hauser
Deltabar S
Process connection FMD78 ANSI flanges, connection dimensions as per ANSI B 16.5, raised face RF
with metallic measuring
diaphragms (continued)
g2
b
dM
1.6 mm
0.06 in
L
g
k d3
D g
P01-FMD78xxx-06-09-xx-xx-010
Process connection FMD78, ANSI flange with and without extended diaphragm seal, material AISI 316/AISI 316L
Ver- Nominal Class Dia- Thick- Raised Exten- Exten- Quan- Dia- Hole max. Dia- TK Weight of
sion dia- meter ness face sion sion tity meter circle phragm Process two dia-
meter length dia- diameter phragm
meter seals
D b g L d3 g2 k dM
[lb/sq.in] [in] [in] [in] [in] [in] [in] [in] [in] [mbar/ [kg]
[mm] [mm] [mm] [mm] [mm] [mm] [mm] [mm] 10 K]
6 14.2
152.4
8 15.4
203.6
AH 4 150 9 0.94 6.19 – – 8 0.75 7.5 3.50 +0.25 14.4
228.6 23.9 157.2 19.1 190.5 89
6 22.3
152.4
8 24.8
203.6
Endress+Hauser 53
Deltabar S
131
45.5
ø51.7
P01-xxxxxxxx-06-xx-xx-xx-000
Dimensions T14 housing, optional display on the side. Housing weight see page 55.
101 76 118
147
45.5
ø51.7
P01-xxxxxxxx-06-xx-xx-xx-001
Dimensions T17 housing, optional display on the side. Housing weight see page 55.
T14
55
55
ø54.1
➀ ➀ ➀
71
+– +– +– +– +–
P01-xxxxxxxx-06-xx-xx-xx-002
Reduction of the mounting height of the process connection, for application of the separate housing.
1 Process connection adapter.
If the separate housing is used, the mounting height of the process connection is reduced by approx. 55 mm
as compared to the dimensions of the standard version.
The minimum bending radius (r) for the cable is 120 mm (4.7").
54 Endress+Hauser
Deltabar S
Weight Housing
Process connections
• Process connections PMD70 with ceramic measuring diaphragms: ab Seite 35 ff
• Process connections PMD75 with metallic measuring diaphragms: ab Seite 37 ff
• Process connection FMD76 with ceramic measuring diaphragms: ab Seite 40 ff
• Process connections FMD77 with metallic measuring diaphragms, low-pressure side: ab Seite 42 ff
• Process connection FMD78 with metallic measuring diaphragms: ab Seite 47 ff
DIN/EN flanges
Endress+Hauser supplies DIN/EN flanges made of stainless steel AISI 316L with the material number 1.4435
or 1.4404. With regard to its stability property, the material 1.4435 is identical to 1.4404 which is grouped
under 13EO in EN 1092-1 Tab. 18. The chemical composition of the two materials can be identical.
Endress+Hauser 55
Deltabar S
56 Endress+Hauser
Deltabar S
Design and operation mode Diaphragm seals are separating equipment between the measuring system and the process medium.
A diaphragm seal system consists of:
• A diaphragm seal in a one-sided system, e.g. FMD77 or two diaphragm seals, in a two-sided system, e.g.
FMD78
• One capillary tube or two capillary tubes
• Fill fluid and
• A differential pressure transmitter.
The process pressure acts via the diaphragm seal membrane on the liquid-filled system, which transfers the
process pressure via the capillary tube onto the sensor of the differential pressure transmitter.
Endress+Hauser delivers all diaphragm seal systems as welded versions. The system is hermetically sealed,
which ensures the highest reliability.
Note!
The correlations between the individual diaphragm seal components are presented in the following section.
For further information and comprehensive diaphragm seal system designs, please contact your local
Endress+Hauser Sales Center.
Diaphragm seal
The diaphragm seal determines the application range of the system by
• the diaphragm diameter
• the diaphragms: stiffness and material
• the design (oil volume).
Diaphragm diameter
The larger the diaphragm diameter (less stiffness), the smaller the temperature effect on the measurement
result.
Note: To keep the temperature effect in practice-oriented limits, you should select diaphragm seals with a
nominal diameter of ≥ DN 80, in as far as the process connection allows for it.
Diaphragm stiffness
The stiffness is dependent on the diaphragm diameter, the material, any available coating and on the diaphragm
thickness and shape. The diaphragm thickness and the shape are defined constructively. The stiffness of a
diaphragm seal membrane influences the temperature operating range and the measuring error caused by
temperature effects.
Capillary
Diaphragm seals are used with the following capillary internal diameters as standard:
• ≤ DN 50: 1 mm
• > DN 50: 2 mm
The capillary tube influences the TK zero point, the ambient temperature operating range and the response time
of a diaphragm seal system as a result of its length and internal diameter.
→ See also page 59 ff, sections "Influence of the temperature on the zero point", "Ambient temperature range"
and "Response time".
→ Observe the installation instructions regarding capillary tubes. See page 64 ff, section "Installation
instructions".
Endress+Hauser 57
Deltabar S
Filling oil
When selecting the filling oil, fluid and ambient temperature as well as the operating pressure are of crucial
importance. Observe the temperatures and pressures during commissioning and cleaning. A further selection
criterion is the compatibility of the filling oil with the requirements of the process medium. For this reason,
only filling oils that are harmless to health are used in the food industry, such as vegetable oil or silicone oil. →
See also the following section "Diaphragm seal filling oils".
The filling oil used influences the TK zero point and the temperature operating range of a diaphragm seal system
and the response time. → See also page 59 ff, sections "Influence of the temperature on the zero point" and
"Response time".
FMD77: F Inert oil –40...+80°C –40...+175°C 1.87 27 0.000876 0.91 Oil for ultra pure gas
FMD78: D, 4 (–40...+176°F) (–40...+347°F) and oxygen
applications
FMD77: D Vegetable oil –10...+120°C –10...+200°C 0.94 9.5 0.00101 1.05 suitable for foods
FMD78: B, 2 (+14...+248°F) (+14...+392°F) FDA 21 CFR 172.856
58 Endress+Hauser
Deltabar S
Influence of the temperature A temperature change results in a volume change of the filling oil. The volume change is dependent on the
on the zero point coefficient of thermal expansion of the filling oil and on the volume of the filling oil at calibration temperature
(constant in the range: +21 to +33°C (+69.8 to 91.4°F)). → See also page 46, section "Diaphragm seal filling
oils".
For example, the filling oil expands in the event of a temperature increase. The additional volume presses
against the diaphragm seal membrane. The stiffer a diaphragm is, the greater its return force, which counteracts
a volume change and acts on the measuring cell together with the operating pressure, thus shifting the zero
point. For the "TK Process", see page 47 ff, section "Process connections FMD78".
The following diagrams display the temperature coefficient "TK Ambient" dependent on the capillary length.
The following application is displayed: capillary temperature and transmitter temperature (ambient
temperature) change, the process temperature corresponds to the calibration temperature.
The temperature coefficients obtained from the diagrams apply to silicone oil and the membrane material
AISI 316L/1.4435. For other filling oils, these temperature coefficients must be multiplied by the TK correction
factor of the corresponding filling oil. For the TK correction factors, see page 58, section "Diaphragm seal filling
oils".
TK Ambient 12
[mbar/10 K] 7
6
10
5
6
4
3
4
2
2
mbar
1.4 1
10 K
0
0 1 2 3 4 5 6 7 8 9 10
Capillary length
[m]
P01-FMD78xxx-05-xx-xx-xx-004
Example for:
– Diaphragm seal versions "B5, EN/DIN Flange DN 80 PN 10-40 B1, AISI 316L"
– Capillary length: 5 m
– Ambient temperature, capillary/transmitter: 45°C
– Filling oil: silicone oil
1. Select characteristic curve type for the diaphragm seal versions "B5" in accordance with the following
table.
Result: characteristic curve type 2
2. Obtain value for TK Ambient from the diagram.
Result: 1.4 mbar/10 K
3. TAmbient – TCalibration = 45°C – 25°C = 20°C ⇒ 1.4 mbar/10 K x 20 K = 2.8 mbar
Result: In this application, the zero point is shifted by 2.8 mbar.
Note!
• The influence of temperature on the zero point can be corrected with position calibration.
• The temperature influence can be minimised by using a filling oil with a smaller coefficient of thermal
expansion, shorter capillaries, diaphragm seals with larger diaphragm diameter or by using a smaller capillary
internal diameter.
Endress+Hauser 59
Deltabar S
TK Ambient 90
[mbar/10 K]
80
70
11
60
50
40
10
30
9
20
8
10
0
0 1 2 3 4 5 6 7 8 9 10
Capillary length
[m]
P01-FMD78xxx-05-xx-xx-xx-005
60 Endress+Hauser
Deltabar S
Ambient temperature range The filling oil, capillary length, capillary internal diameter, process temperature and the oil volume of the
diaphragm seal determine the ambient temperature operating range of the diaphragm seal system.
The following diagrams display the permitted ambient temperature operating range in relation to the capillary
length. The diagrams apply to a process temperature of +25°C (+77°F) and to silicone oil.
The operating range can be extended by using a filling oil with a smaller coefficient of expansion and by using
shorter capillaries.
Note!
• Endress+Hauser recommends using low temperature oil for applications that require short response times or
that are near the lower temperature limit. → See also page 58, "Diaphragm seal filling oils" section.
• For further information, comprehensive diaphragm seal system designs and measuring technology solutions
acting close to the operating limits, please contact your local Endress+Hauser Sales Center.
Group A Group B
80 80
Ambient temperature [°C]
Group C Group D
80 80
Ambient temperature [°C]
60 60
40 40
20 permitted 20 permitted
operating range operating range
0 0
–20 –20
–40 –40
0 1 2 3 4 5 6 7 8 9 10 0 1 2 3 4 5 6 7 8 9 10
Capillary length [m] Capillary length [m]
P01-FMD78xxx-05-xx-xx-xx-009 P01-FMD78xxx-05-xx-xx-xx-008
Group E Group F
80 80
Ambient temperature [°C]
60 60
40 40
permitted 20 permitted
20 operating range operating range
0 0
–20 –20
–40 –40
0 1 2 3 4 5 6 7 8 9 10 0 1 2 3 4 5 6 7 8 9 10
Capillary length [m] Capillary length [m]
P01-FMD78xxx-05-xx-xx-xx-015 P01-FMD78xxx-05-xx-xx-xx-013
Endress+Hauser 61
Deltabar S
Group G Group H
80 80
Ambient temperature [°C]
Group I Group J
80 80
Ambient temperature [°C]
62 Endress+Hauser
Deltabar S
Response time The viscosity of the filling oil, the capillary length and the capillary internal diameter influence the frictional
resistance. The greater the frictional resistance, the longer the response time.
Furthermore, the volume change of the measuring cell influences the response time. The lower the
volume change of the measuring cell is, the less filling oil has to be shifted in the diaphragm seal system.
The following diagram shows typical response times (T90) for the various filling oils dependent on the
measuring cell and the capillary internal diameter. The values given are in seconds per metre of capillary length
and must be multiplied by the actual length of the capillary. The response time of the transmitter must also be
taken into consideration.
6
Silicone oil
4 High temperature oil
Vegetable oil
Inert oil
2
0
DN 50 DN 80 DN 50 DN 80 Nominal diameter
(1 mm) (2 mm) (1 mm) (2 mm) Capillary internal diameter
500 mbar 3 bar Measuring cell
P01-FMD78xxx-05-xx-xx-xx-000
100 %
➀
90 %
T90 t
P01-xxxxxxxx-05-xx-xx-xx-006
Larger capillary internal diameter The temperature effect increases with increasing diameter.
Shorter capillaries –
Filling oil with lower viscosity – Observe compatibility of the filling oil with the process
fluid.
– Observe the filling oil operating limits.
Endress+Hauser 63
Deltabar S
Selecting the measuring cell (observe the hydrostatic pressure of the filling fluid column in the capillaries!)
pi
–
ΔH = 1 m Hv = 1.8 m
min. +
Hu = 0.2 m
H1 = 0.3 m
+ –
Pressure on the negative side of the differential pressure transmitter (p–) when the tank is empty (min. level)
kg kg
= 1.8 m • 0.96 • 9.81 m m
s + 0.3 m • 0.96 dm3 • 9.81 s + pi
dm3
= 197.77 mbar + p i
Pressure on the positive side of the differential pressure transmitter (p+) when the tank is empty (min. level)
kg kg
= 0.2 m • 1 • 9.81 m m
s + 0.3 m • 0.96 dm3 • 9.81 s + pi
dm3
= 47.87 mbar + pi
64 Endress+Hauser
Deltabar S
Result:
If the tank were full, a differential pressure of –51.80 mbar would be present at the differential pressure transmitter. When
the tank is empty, a differential pressure of –149.90 mbar is present. Therefore, a 500 mbar measuring cell is required for
this application.
Installation instructions
In order to obtain more precise measurement results and to avoid a defect in the device, mount the capillaries
as follows:
• vibration-free (in order to avoid additional pressure fluctuations)
• not in the vicinity of heating or cooling lines
• insulate if the ambient temperature is below ore above the reference temperature
• with a bending radius of ≥ 100 mm.
Vacuum applications
For applications under vacuum, Endress+Hauser recommends mounting the pressure transmitter underneath
the lower diaphragm seal. A vacuum load of the diaphragm seal caused by the presence of filling oil in the
capillaries is hereby prevented.
When the pressure transmitter is mounted above the lower diaphragm seal, the maximum height difference
H1 in accordance with the following illustration on the left must not be exceeded. The maximum height
difference is dependent on the density of the filling oil and the smallest ever pressure that is permitted to occur
at the diaphragm seal on the positive side (empty tank), see the following illustration, on the right.
12.0
Low temperature oil
10.0
Height difference H1 [m]
Vegetable oil
8.0
– Silicone oil
6.0
High temperature
– + oil
+ –
4.0
H1 Inert oil
2.0
+
0.0
50 100 200 300 400 500 600 700 800 900 1000
Pressure, diaphragm seal positive side [mbarabs]
P01-FMD7xxxx-11-xx-xx-xx-001 P01-FMD7xxxx-05-xx-xx-xx-002
Installation above the lower diaphragm seal Diagram of maximum installation height above the lower diaphragm
seal for vacuum applications dependent on the pressure at the
diaphragm seal on the positive side
Endress+Hauser 65
Deltabar S
Ex approvals • ATEX
• FM
• CSA
• NEPSI
• IECEx
• GOST on request
• also combinations of different approvals
All explosion protection data are given in separate documentation which is available upon request. The Ex
documentation is supplied as standard with all devices approved for use in explosion hazardous areas.
→ See also page 81 ff, sections "Safety Instructions" and "Installation/Control Drawings".
Overspill protection WHG (German Water Resources Act). See "Ordering information" → ä 67 (see also ZE259P/00/de).
CRN approvals Some device versions have CRN approval. For a CRN-approved device, a CRN-approved process connection
(→ see page 68, feature 70 "Process connection") has to be ordered with a CSA approval (→ see page 67,
feature 10 "Approval"). These devices are fitted with a separate plate bearing the registration number
0F10524.5C.
Pressure Equipment Directive The devices PMD70, PMD75, FMD76, FMD77 und FMD78 corresponds to Article 3 (3) of the EC directive
(PED) 97/23/EC (Pressure Equipment Directive) and has been designed and manufactured according to good
engineering practice.
Additionally applies:
– FMD78 with pipe diaphragm seal ≥ 1.5"/PN40:
Suitable for stable gases in group 1, category II
– PMD75, PN 420
Suitable for stable gases in group 1, category I
66 Endress+Hauser
Deltabar S
Ordering information
PMD70 This overview does not mark options which are mutually exclusive.
10 Approval:
A For non-hazardous areas
1 ATEX II 1/2 G EEx ia IIC T6
6 ATEX II 1/2 G EEx ia IIC T6, overspill protection WHG
➀ 2 ATEX II 1/2 D
4 ATEX II 1/3 D
8 ATEX II 1 GD EEx ia IIC T6
3 ATEX II 1/2 GD EEx ia IIC T6
– + E
➁ 7 ATEX II 3 G EEx nA II T6
S FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
Q FM DIP, Class II, III Division 1, Groups E – G
R FM NI, Class I, Division 2, Groups A – D
U CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
W CSA Class II, III Division 1, Groups E – G (Dust-Ex)
H NEPSI Ex ia IIC T6
I IECEx Zone 0/1 Ex ia IIC T6
20 Output; Operation:
➂ A 4...20 mA HART, SIL operation outside, LCD (→ see Fig. ➀, o)
B 4...20 mA HART, SIL operation inside, LCD (→ see Fig. ➀, p)
➃ C 4...20 mA HART, SIL operation inside (→ see Fig. p)
M PROFIBUS PA, operation outside, LCD (→ see Fig. ➀, o)
1 2
N PROFIBUS PA, operation inside, LCD (→ see Fig. ➀, p)
PC
HART
R
O PROFIBUS PA, operation inside (→ see Fig. p)
250002271-–
Endress+Hauser 67
Deltabar S
80 Seal:
A FKM Viton
B EPDM
D Kalrez
E Chemraz
1 FKM Viton, cleaned from oil and greace
2 FKM Viton, cleaned for oxygen service
Note application limits pressure/temp.
68 Endress+Hauser
Deltabar S
PMD75 This overview does not mark options which are mutually exclusive.
10 Approval:
A For non-hazardous areas
1 ATEX II 1/2 G EEx ia IIC T6
6 ATEX II 1/2 G EEx ia IIC T6, overspill protection WHG
2 ATEX II 1/2 D
4 ATEX II 1/3 D
8 ATEX II 1 GD EEx ia IIC T6
3 ATEX II 1/2 GD EEx ia IIC T6
5 ATEX II 2 G EEx d IIC T6
7 ATEX II 3 G EEx nA II T6
S FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
T FM XP, Class I Division 1, Groups A – D; AEx d
Q FM DIP, Class II, III Division 1, Groups E – G
➀
R FM NI, Class I, Division 2, Groups A – D
U CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
V CSA XP, Class I Division 1, Groups B – D; Ex d
W CSA Class II, III Division 1, Groups E – G (Dust-Ex)
– + E
➁ G
H
NEPSI Exd IIC T6
NEPSI Ex ia IIC T6
I IECEx Zone 0/1 Ex ia IIC T6
B Combined certificates: ATEX II 1/2 G EEx ia IIC T6 + II G EEx d IIC T6
C Combined certificates: FM IS and XP Class I Division 1, Groups A – D
D Combined certificates: CSA IS and XP Class I Division 1, Groups A – D
E Combined certificates: FM/CSA IS and XP Class I Division 1, Groups A – D
F Combined certificates: ATEX II EEx ia / EEx d + FM/CSA IS + XP; ATEX II 1/2G EEx ia IIC T6+; ATEX II 2G EEx d IIC
T6+; FM/CSA IS + XP Cl.I Div.1 Gr.A-D
20 Output; Operation:
➂ A 4...20 mA HART, SIL operation outside, LCD (→ see Fig. ➀, o)
B 4...20 mA HART, SIL operation inside, LCD (→ see Fig. ➀, p)
➃ C 4...20 mA HART, SIL operation inside (→ see Fig. p)
M PROFIBUS PA, operation outside, LCD (→ see Fig. ➀, o)
1 2
N PROFIBUS PA, operation inside, LCD (→ see Fig. ➀, p)
PC
HART
R
O PROFIBUS PA, operation inside (→ see Fig. p)
250002271-–
T17 N Aluminium T15 housing, optional display on the top, IP 66/67/NEMA 6P, 7/8" FF plug
P Aluminium T15 housing, optional display on the top, IP 65/NEMA 4X, Han7D plug, 90°
1 AISI 316L T14 housing, optional display on the side, IP 66/67/NEMA 6P, Gland M 20x1.5
2 AISI 316L T14 housing, optional display on the side, IP 66/67/NEMA 6P, Thread G 1/2
3 AISI 316L T14 housing, optional display on the side, IP 66/67/NEMA 6P, Thread 1/2 NPT
4 AISI 316L T14 housing, optional display on the side, IP 66/67/NEMA 6P, M 12x1 PA plug
5 AISI 316L T14 housing, optional display on the side, IP 66/67/NEMA 6P, 7/8" FF plug
6 AISI 316L T14 housing, optional display on the side, IP 65/NEMA 4X, Han7D plug, 90°
R AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, Gland M 20x1.5
S AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, Thread G 1/2
T AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, Thread 1/2 NPT
U AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, M 12x1 PA plug
V AISI 316L T17 housing, optional display on the side, IP 66/68/NEMA 6P, 7/8" FF plug
Z Housing: see additional specifications
Endress+Hauser 69
Deltabar S
50 Calibration; Unit:
1 Nominal range; mbar/bar
2 Nominal range; kPa/MPa
3 Nominal range; mmH2O/mH2O
4 Nominal range; inH2O/ftH2O
6 Nominal range; psi
8 Adjusted for Deltatop/Deltaset; see additional specification
B Customised; see additional specification
C Factory certificate 5-point; see additional specification
D DKD certificate; see additional specification
K Platinum; see additional specification
L Platinum and factory certificate 5-point; see additional specification
M Platinum and DKD certificate; see additional specification
60 Membrane material:
1 AISI 316L
2 Alloy C
3 Monel
5 Tantal
6 AISI 316L with Gold-Rhodium coating
70 Endress+Hauser
Deltabar S
Endress+Hauser 71
Deltabar S
FMD76 This overview does not mark options which are mutually exclusive.
10 Approval:
A For non-hazardous areas
➀ 1 ATEX II 1/2 G EEx ia IIC T6
6 ATEX II 1/2 G EEx ia IIC T6, overspill protection WHG
2 ATEX II 1/2 D EEx ia IIC T6
8 ATEX II 1 GD EEx ia IIC T6
– + E
➁ 3 ATEX II 1/2 GD EEx ia IIC T6
7 ATEX II 3 G EEx nA II T6
S FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
R FM NI, Class I, Division 2, Groups A – D
U CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
H NEPSI Ex ia IIC T6
I IECEx Zone 0/1 Ex ia IIC T6
20 Output; Operation:
➂ A 4...20 mA HART, SIL operation outside, LCD (→ see Fig. ➀, o)
B 4...20 mA HART, SIL operation inside, LCD (→ see Fig. ➀, p)
➃ C 4...20 mA HART, SIL operation inside (→ see Fig. p)
M PROFIBUS PA, operation outside, LCD (→ see Fig. ➁, o)
1 2
N PROFIBUS PA, operation inside, LCD (→ see Fig. n, p)
PC
HART
R
O PROFIBUS PA, operation inside (→ see Fig. p)
250002271-–
72 Endress+Hauser
Deltabar S
Endress+Hauser 73
Deltabar S
74 Endress+Hauser
Deltabar S
FMD77 This overview does not mark options which are mutually exclusive.
10 Approval:
A For non-hazardous areas
1 ATEX II 1/2 G EEx ia IIC T6
6 ATEX II 1/2 G EEx ia IIC T6, overspill protection WHG
2 ATEX II 1/2 D
4 ATEX II 1/3 D
8 ATEX II 1 GD EEx ia IIC T6
3 ATEX II 1/2 GD EEx ia IIC T6
5 ATEX II 2 G EEx d IIC T6
7 ATEX II 3 G EEx nA II T6
S FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
T FM XP, Class I Division 1, Groups A – D; AEx ia
Q FM DIP, Class II, III Division 1, Groups E – G
➀ R FM NI, Class I, Division 2, Groups A – D
U CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
V CSA XP, Class I Division 1, Groups B – D; Ex d
W CSA Class II, III Division 1, Groups E – G (Dust-Ex)
– + E
➁ G NEPSI Ex d IIC T6
H NEPSI Ex ia IIC T6
I IECEx Zone 0/1 Ex ia IIC T6
B Combined certificates: ATEX II 1/2 G EEx ia IIC T6 + II G EEx d IIC T6
C Combined certificates: FM IS and XP Class I Division 1, Groups A – D
D Combined certificates: CSA IS and XP Class I Division 1, Groups A – D
E Combined certificates: FM/CSA IS and XP Class I Division 1, Groups A – D
F Combined certificates: ATEX II EEx ia / EEx d + FM/CSA IS + XP; ATEX II 1/2G EEx ia IIC T6+; ATEX II 2G EEx d IIC
T6+; FM/CSA IS + XP Cl.I Div.1 Gr.A-D
20 Output; Operation:
➂ A 4...20 mA HART, SIL operation outside, LCD (→ see Fig. ➀, o)
B 4...20 mA HART, SIL operation inside, LCD (→ see Fig. ➀, p)
➃ C 4...20 mA HART, SIL operation inside (→ see Fig. p)
M PROFIBUS PA, operation outside, LCD (→ see Fig. ➀, o)
1 2
N PROFIBUS PA, operation inside, LCD (→ see Fig. ➀, p)
PC
HART
R
O PROFIBUS PA, operation inside (→ see Fig. p)
250002271-–
Endress+Hauser 75
Deltabar S
50 Calibration, Units:
1 Calibration: nominal range, mbar/bar
2 Calibration: nominal range, kPa/MPa
3 Calibration: nominal range, mmH2O/mH2O
4 Calibration: nominal range, inH2O/ftH2O
6 Calibration: nominal range, psi
B Custom calibration: see additional specification
C Factory calibration: see additional specification, Factory calibration certificate, 5-point
D DKD calibration: see additional specification
76 Endress+Hauser
Deltabar S
Endress+Hauser 77
Deltabar S
FMD78 This overview does not mark options which are mutually exclusive.
10 Approval:
A For non-hazardous areas
1 ATEX II 1/2 G EEx ia IIC T6
6 ATEX II 1/2 G EEx ia IIC T6, overspill protection WHG
2 ATEX II 1/2 D
4 ATEX II 1/3 D
8 ATEX II 1 GD EEx ia IIC T6
3 ATEX II 1/2 GD EEx ia IIC T6
5 ATEX II 2 G EEx d IIC T6
7 ATEX II 3 G EEx nA II T6
S FM IS, Class I, II, III Division 1, Groups A – G; NI Class I Division 2, Groups A – D; AEx ia
T FM XP, Class I Division 1, Groups A – D; AEx ia
Q FM DIP, Class II, III Division 1, Groups E – G
➀ R FM NI, Class I, Division 2, Groups A – D
U CSA IS, Class I, II, III Division 1, Groups A – G; Class I Division 2, Groups A – D, Ex ia
V CSA XP, Class I Division 1, Groups B – D; Ex d
W CSA Class II, III Division 1, Groups E – G (Dust-Ex)
– + E
➁ G NEPSI Ex d IIC T6
H NEPSI Ex ia IIC T6
I IECEx Zone 0/1 Ex ia IIC T6
B Combined certificates: ATEX II 1/2 G EEx ia IIC T6 + II G EEx d IIC T6
C Combined certificates: FM IS and XP Class I Division 1, Groups A – D
D Combined certificates: CSA IS and XP Class I Division 1, Groups A – D
E Combined certificates: FM/CSA IS and XP Class I Division 1, Groups A – D
F Combined certificates: ATEX II EEx ia / EEx d + FM/CSA IS + XP; ATEX II 1/2G EEx ia IIC T6+; ATEX II 2G EEx d IIC
T6+; FM/CSA IS + XP Cl.I Div.1 Gr.A-D
20 Output; Operation:
➂ A 4...20 mA HART, SIL operation outside, LCD (→ see Fig. ➀, o)
B 4...20 mA HART, SIL operation inside, LCD (→ see Fig. ➀, p)
HW-Version: 1
SW-Version: 2
78 Endress+Hauser
Deltabar S
50 Calibration, Units:
1 Calibration: nominal range, mbar/bar
2 Calibration: nominal range, kPa/MPa
3 Calibration: nominal range, mmH2O/mH2O
4 Calibration: nominal range, inH2O/ftH2O
6 Calibration: nominal range, psi
B Custom calibration: see additional specification
C Factory calibration: see additional specification, Factory calibration certificate, 5-point
D DKD calibration: see additional specification, DKD-Certificate
60 Membrane material:
1 AISI 316L
2 Alloy C
3 Monel
5 Tantalum
6 AISI 316L with Gold-Rhodium coating
7 AISI 316L with 0.09 mm PTFE foil (not for vacuum applications)
Endress+Hauser 79
Deltabar S
80 Endress+Hauser
Deltabar S
Further documentation
Innovation • Cerabar S/Deltabar S/Deltapilot S,
For process pressure, differential pressure, flow and level measurement: IN010P/00/en
Field of Activities • Pressure measurement: Powerful instruments for process pressure, differential pressure, level and flow:
FA004P/00/en
Endress+Hauser 81
Deltabar S
82 Endress+Hauser
Deltabar S
Endress+Hauser 83
Deltabar S
Instruments International
Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland
TI382P/00/en/12.07
No. 71060097