Smart Balance User Manual Keypad Version
Smart Balance User Manual Keypad Version
EC40011
1 lb = 453.59237g
1 oz = 28.349523125g
25 Error information
Content Display Error solution
ADO--- AD value over flow Take off the weights or send to
1 Notice ......................................................................Error! Bookmark not defined.
repair
2 Specification .....................................................Error! Bookmark not defined.
ADL--- AD value over flow Take off the weights or send to
3 Main parameter ...................................................Error! Bookmark not defined.
repair
4 Port .........................................................................Error! Bookmark not defined.
OVR--- Over load Check if the weight of weighs is
5 Keypad ..................................................................Error! Bookmark not defined.
+9e more that the Max. Cap.
6 ON/OFF ..................................................................Error! Bookmark not defined.
Batter sign red Low power Charging
7 Charging ...............................................................Error! Bookmark not defined.
Low battery, Battery ‘ voltage is Stop using and charging
8 Display ...................................................................Error! Bookmark not defined.
please very low
9 Zero ........................................................................Error! Bookmark not defined.
charging
10 Tare .....................................................................Error! Bookmark not defined.
11 Choose units ....................................................Error! Bookmark not defined.
12 Counting (Option) ...........................................Error! Bookmark not defined.
13 Select and edit goods....................................Error! Bookmark not defined.
14 Print ....................................................................Error! Bookmark not defined.
15 Check the record ............................................Error! Bookmark not defined.
16 Export file to U disk ...................................Error! Bookmark not defined.
17 Parameter setting ...........................................Error! Bookmark not defined.
18 Communication ...............................................Error! Bookmark not defined.
19 Time setting .....................................................Error! Bookmark not defined.
20 User management(Option) ...........................Error! Bookmark not defined.
21 Units conversion.............................................Error! Bookmark not defined.
22 Error information ............................................Error! Bookmark not defined.
1 30
Notice
1.1 Please charging before first use.
When low power sign shows, please use the original adapter to
charging. Red Charging light means scale is in charging mode.
When charging light turn to green, it means charging finished. The
charging time is about 5-7 hours.
1.2 Put and balance the scale on the stable and flat table with four
adjustable feet. Please notice the air bubble should be kept in
the center of circle.
Input the old password, the input the new password twice, click [ok] to complete 1.3 Please avoid the water.
the setting. 1.4 Please use in suitable temperature work place.
23.3 Exit 1.5 If mobile or wireless machines are in use nearby when turning
on the scale, the display will flash once, and then become
normal. If it doesn't work, please power on again.
1.6 The original adaptor should be used when charging.
Press to select in the pop-up window then enter
1.7 After a long-time use (more than one year), the rechargeable
password setting. The window as follow: battery can't keep using for long after more than 5-hours
charging, which means new one should be replaced. If you
have no idea how to replace, please ask a nearby agent to
give you a helping hand.
1.8 Please turn off the scale when it won't be used for a long time.
Please pack up the scale if do not use for a long time,
charging every three month. Charging to full before start to
use.
1.9 Avoid the knocking, please do not throw the weight on the scale.
1.10 Please do not exceed the Maximum Capacity.
Use or to choose the user name, enter the right 2 Specification
password, then press ok key. After login succeed, the upper left corner 1.1 5 inch TFT color display, easy to read.
will show the login user name.
1.2 Adjustable brightness of back light, fit to all time use.
1.3 Solid frame design to insure the stable weighing.
24 Units conversion
1.4 With data memory, can store many weighting date.
1 kg = 1000g
1.5 Fast reading, stable, accurate, easy using.
29 2
1.6 Low power alarm, charging light show.
1.7 Counting, Hi/Lo, target value, …
3 Main parameter
3.1 Available accuracy: 1/30000
3.2 Power adapter: built –in or outer AC 12V/3000mA adapter.
3.3 Work current: 200mA
3.4 Display: 5 inch TFT
3.5 Sensor sensitivity range: 1 - 3mV/V
3.6 Power: DC6V/10.0AH(Low power sign appears when voltage
lower than 5.6V)
3.7 Work temperature: 0-40 ℃, lower than 90%RH
3.8 Storage temperature: -25~55℃ Choose the responding user, click [edit] to edit user information.
To start using one user, user must set the user name and password, and tick it.
Other permissions depend on the requirement of user.
Pin Define Note Note: The default user name is Admin, all the permissions of it is opened.
1 E+ Please change its password in time.
2 E- 23.2 Password setting
3 S+
4 S-
5 Ground Connect to the shielded wire Press to select in the pop-up window then enter
1.1 Fix password setting. The window as follow:
Put the load cell cable through the pillar, then insert the pillar into base,
fasten the fixed screw. After put the cable through the holder. Insert the
holder into the bracket of the indicator. Fix the load cell connector.
Adjust the holder to right position, and then fasten the plastic screw on
the display.
3 28
5 Port
5.1 When indicator crashed, please press the reset button which on the left
of back housing to reset.
5.2 When the charging light is red, it means the indicator is charging. When
the charging light is green, it means the indicator charging finished.
5.3 USB1 can connect to U-disk and scanning gun.
5.4 USB2 is option.
5.5 RS232 1# can connect to PC, download the print format and so on.
5.6 RS232 2# can connect to the outer printer.
After choose the right date and time, click [ok].
Note: After click [ok], the second clock will be set to 0.
6 Keypad
23 User management(Option)
6.1 Common weighing version
23.1 Edit user information
27 4
Pin Define Pin Define
2 RXD 3 TXD
3 TXD 2 RXD
5 GND 5 GND
Header1 Header2
ST, Stable NT, Net weight
US, Unstable GS, Gross weight
6.3 Function instruction
OV, Over load TR, Tare weight
6.3.1 Number key + letter key, use to input number, letter.
Weight value data is ASCII character, may have follow characters:
“ 0” ~ “9” number “ ” blank “ . ”Decimal point “ - ” negative sign
6.3.3 Delete key, use to delete content. In goods Press to select in the pop-up window then enter
time setting. The window as follow:
showing and record showing window, use to delete the chose
item.
6.3.5 Use to get the stored PLU value. E.g. input number 12,
then press [PLU], the No.12 stored PLU values will show.
(Include unit weight, SKU id…) Press this key and hold 3s, user
can enter to the goods choosing interface.
5 26
21.1.4 USB
When insert USB, scanner gun or keypad, user can define these
equipment’s function. 6.3.6 Get sample’s unit weight by enter weights quantity.
21.1.4.1.1 Analog keypad
Input letter and number will be shown in the input box.
21.1.4.1.2 Common scan 6.3.7 Enter the unit weight value. Use number to input the unit
The scanned content shows in the scan item window weight, then press this key to set unit weight.
which in the upper of the goods window. User can print
out that content..
21.1.4.1.3 Scan and print 6.3.8 Use to change the setting item.
Based on the common print, print at the same time.
21.1.4.1.4 Scan goods
Input letter and number will not be shown in the input 6.3.9 In weighing mode, use this key to change the
box. When software checks out the Enter, the scale units. In setting and input mode, use this key to leftward
will judge if there has goods serial number in the input move.
content. If it has, the scale will call out the goods and
print out the corresponding content.
21.1.4.1.5 Scan and print 6.3.10 In weighing mode, use this key to check the print
Input letter and number will not be shown in the input records. In setting and input mode, use this key to upward
box. When software checked out the Enter, the scale movement or change the option’s parameter.
will print out the corresponding content.
21.2 Wire connection
Scale with DB9 female connector (factory default) 6.3.11 In weighing mode, use this key to tare. In
Connect to Indicator (male Connect to PC port(female setting and input mode, use this key to downward movement
connector) connector) or change the option’s parameter.
Pin Define Pin Define
2 TXD 2 RXD
6.3.12 In weighing mode, use this key to turn back
3 RXD 3 TXD
5 GND 5 GND to zero. In setting and input mode, use this key to
rightward move.
scale with DB9 male connector
Connect to Indicator (female Connect to PC port(female
6.3.13 In weighing mode, use this key to enter the
connector) connector)
25 6
user defined parameter setting. In setting and input mode,
this key use to confirm the select or shift the input box.
The version which without menu key, long press this key
to enter the setting menu.
6.3.15 In weighing mode, use this key to show the Press [connect] to use this function directly. Or choose static IP and
details of accumulative weight value and quantity. When input the IP address, mask, gateway, and then click connect key to
showing goods and weight records, this key use to shift connect. If user do not know these parameter, please contact the
to the next page. webmaster.
If user suspect the module crashed, please click [module reset] to
reset.
6.3.16 In weighing mode, use this key to print out 21.1.3 Public setting
the current weight value and quantity. In MR mode, use 21.1.3.1 Print mode
this key to print out the accumulative weight value and 21.1.3.1.1 Manual print
quantity. Press [print] key to print out content once.
21.1.3.1.2 Auto print
Put on the weights, after the weight value stabled, the
6.3.17 In weighing mode, long press this key to enter indicator will print out content once.
the setting menu. 21.1.3.1.3 Qualified print
Put on the weights, after the weight value stabled, if
the weight value is not in the no-alarm range, the
6.3.18 In weighing mode, long press this key to enter indicator will print out content once.
Hi-Lo setting interface. 21.1.3.2 Check hilo (In qualified range print)
7 ON/OFF Choose this function, the indicator will just print in manual
Press ON/OFF key to power on and power off.. mode when weight value in qualified range.
8 charging
7 24
This is communication format. When mode set to print mode, this Connect the scale with original adapter to charging. While charging, the
parameter is for print format. charging light is red. It will turn green when charging finished (the charging
21.1.2.3 Target IP, port length should be 5-7h).
Here use to input the server IP and port, it must be the TCP server. Attention: Do not use up the battery, otherwise the battery will be broken. If
After input, click [Connect]. user do not use the indicator for a long time, user should recharging every 40
21.1.2.4 Network setting days.
Press [Network] to set the network. After charge, please take off the adapter, please do not always connect the
21.1.2.4.1 Wi-Fi version indicator to power supply.
Please do not use other adapters, it may damage the inside battery.
9 Display
9.1 Weighing window
Counting mode
Weighting mode
Press to search the Wi-Fi signal.
All Wi-Fi name will come out in under window, choose and click
the right Wi-Fi, input its password in right side sheet, click
[connect].
21.1.2.4.2 Wire version
9.1.1 The content name of the different area of the window shown on
the top left corner of every area. Such as weight, Hi. Lo., unit
weight, quantity…
9.1.2 ->0<- ---It means Weight is zero.
9.1.3 Tare---It means here has tare weight.
9.1.4 Stable---When this sign appears, it means the weight value
is effective.
23 8
9.1.5 Low power---When voltage is lower than 5.5V, this sign will
appear to remind user charging the indicator.
.
9.2 The status bar
9 22
User can use this function to choose the alarm mode, also can From up and down, the information are goods name, Hi/Lo limit value, the
close alarm function. state of Hi/Lo limit, the number of times of Hi/Lo limit, the total weight of
20.2.7 Alarm in negative or zero weight value Hi/Lo limit.
Choose negative alarm, the alarm function effective when the For an example, when weight is less than the Lo limit, the
weight is negative value.
Choose zero alarm, the alarm function effective when the weight is
zero. will turn to orange color. At this time, press
Use key to tick or not. [Print], the number of times will show under the ,
20.2.8 Alarm after stable and the total weight of Lo limit will show under the number of the times.
Choose this function, the alarm function effective after scale stabled.
21 Communication
21.1 Setting
21 10
20.1 Units setting
Tick the unit which need to use. Remove the tick for the unit
which do not need to use. For an example, user need to use kg, tick it. User
do not need to use g, remove the tick in front of it.
From up and down, the information are goods name, Hi/Lo limit value, the Use and to sliding the slider to set the
status of Hi/Lo limit, the number of times of Hi/Lo limit, the total weight of brightness of the back light. The brightness of backlight can timely
Hi/Lo limit. changed.
Target mode set the Lo limit value to target value. For an example, when the 20.2.2 Sleep time
weight value is half of the Lo limit, the percent state line will show 50%. At This function use to set the length of duration to control the
this time, press [Print], the number of times will show under the backlight turns dark. For an example, this parameter set to 12, it
means the backlight turn to dark after 12s. But need the indicator
without operate and the weight value do not change. If the indicator
, and the total weight of Lo limit will show under the operated or the weight value changed, the backlight will recover.
number of the times. 20.2.3 Power on zero range
It means the allowed difference of power on zero and calibration
value. The default difference is 20%.
20.2.4 Manual zero rang
It means the allowed difference of manual zero and power on zero.
In this range, manual zero is effective. The default difference is 4%.
20.2.5 Stable Operation
It means if zero and tare operate effective after scale stabled.
11 20
11 Zero
When the pan is empty, click key to let the display weight
value back to zero. If the scale is broken or weights heavy than zero range,
this key is invalid. If the display has rest weight value when the pan is empty,
please use this function. Zero rang= Max. Cap. X4%.
12 Tare
12.1 Tare weight is known
then enter to parameter setting. Put the container on the pan, click key, after stable, the weight
window will show net weight and tare symbol.
13 Choose units
If the units is not the suitable one, click key to choose the right
units. The new units will be stored after correct power off.
14 Counting (Option)
14.1 Unit weight of goods is known
Follow the article 15, set the unit weight in goods edit step.
Choose the unit weight by select goods.
14.2 Common counting function
14.2.1 Unit weight is known
19 12
Input the known weight of weights, press [UNIT W.T], the display will
show the counts. At this time, the indicator enter into the counting mode
automatically. function will be ticked. press key to cancel tick.
14.2.2 Unit weight is unknown
Put the sample weights on the pan, use the number keys to input the
counts of sample, press [SMPL] key. After the weight value is stable, Press [Screening] to show the records which been chose.
the quantity window will show the counts of weight.
In the main interface, long press , goods select window will come
out, as follow: Press [U disk], choose the suitable folder, input the name of the file, then press
[Save]. After “store succeed” shown, user can pull out the U disk.
20 Parameter setting
13 18
file can be opened by Excel. Select the goods which need to edit, press edit to edit
18.3 Details goods information.
Choose the option, press [Details], the indicator will show more details
of this goods. As follow picture:
Also can select the goods which need to edit, press
key to edit.
15.1.6 Delete
Select the goods which need to delete, press [Delete] to delete the
goods. After delete, this recode of goods cannot restored anymore..
17 14
Press number keys to input high limit. Please take attention on the right unit
show area. After input, press [OK], the Low limit window will pop-up. Use
number keys to input low limit, then press [OK]. Note: The low limit value must
lower than high limit value.
Number: Goods number. After add, this number cannot be edited If need to cannel the input content, use to switch, then press
again.
Name: goods name, user can edit again.
Specification: Specification of goods, user can edit again. .
Remark: Remark of goods, user can edit again.
Unit weight: Unit weight of goods, use in counting mode. 17 Print
Hi limit: In counting mode, the unit is PCS. Otherwise the unit is weight
unit, please take attention on the right unit show area.
Low Limit: In counting mode, the unit is PCS. Otherwise the unit is Press can print out the current data and store this data into the scale.
weight unit, please take attention on the right unit show area.
A-B, B-C, C-D, D-E: Use to select mode, use to show the weight/quantity 18 Check the record
difference of A&B, B&C, C&D, D&E. Please take attention on the right
unit show area.
A-B, B-C, C-D, D-E: Option function. Press can check the print record, see follow picture:
16 Input the Hi/Lo limit value directly
Not only call in Hi/Lo limit by select goods, also can use keypad to input Hi/Lo
limit.
Long press key, Hi/Lo limit input window will come out.
18.1 Press and can show the previous page and next
page.
18.2 Export
Insert the U disk, user can export the current data into the U disk. This
15 16