1088HDCameraOperatingManual PDF
1088HDCameraOperatingManual PDF
Table of Contents
Table of Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1
Cleaning the Camera Head . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Disinfecting the Camera Head. . . . . . . . . . . . . . . . . . . . . . . . 26
Sterilizing the Camera Head . . . . . . . . . . . . . . . . . . . . . . . . . 27
100% Ethylene Oxide Gas . . . . . . . . . . . . . . . . . . . . . . . . 27
Steris System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Sterrad System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Using Sterile Drapes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Maintaining the 1088 Camera . . . . . . . . . . . . . . . . . . . . . . . . 29
Replacing the Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Periodic Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . 29
Disposing of the 1088 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
Troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . 37
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
International Service . . . . . . . . . . . . . . . . . . . . . . . . . 41
2
Warnings and Cautions
3
8. Never sterilize the camera console.
9. Disconnect the control unit from the electrical outlet when
inspecting the fuses.
10. Read the entire instruction manual before assembling or
connecting the unit.
11. Before each use, check the outer surface of the endoscope
to ensure that there are no rough surfaces, sharp edges, or
protrusions.
The warranty is void if any of these warnings is disregarded.
Stryker Endoscopy accepts full responsibility for the effects on safety,
reliability, and performance of the equipment only if:
• Readjustments, modifications, and/or repairs are carried out
exclusively by Stryker Endoscopy.
• The electrical installation of the relevant operating room complies
with the applicable IEC, CEC, and NEC requirements.
Stryker Endoscopy reserves the right to make improvements to the
product(s) described herein. Product(s), therefore, may not agree in detail
to the published design or specifications. All specifications are subject to
change without notice. Please contact the local Stryker Endoscopy
Distributor listed in the Other Service section, or phone your local Stryker
Endoscopy sales representative or agent for information on changes and
new products.
4
Warnings and Cautions
Symbol Definitions
In addition to the cautionary symbols already listed, other symbols found
on the 1088 Camera and in this manual have specific meanings that
clarify the proper use and storage of the 1088 Camera. The following list
defines the symbols associated with this product.
Equipotentiality
5
6
Product Description and Intended Use
7
1 2 3 4 5 6 7
8 9 10 11 12 13 14 15
16 17 18 19 20 21 22 23
8
Product Description and Intended Use
9
The Camera Head
The camera head connects to the camera console and captures video
and photographic images, which it relays to the camera console. It
features several controls that are accessible through a button control pad
located on the top of the camera head (see the Operating the Camera
Head section of this manual).
Figure 2 below lists the features of the camera head.
1
2
10
Product Description and Intended Use
The Coupler
The coupler threads on to the face of the camera head, enabling a scope
to be attached to the camera. It provides a focusing ring to adjust image
sharpness.
The features of the coupler are listed in Figure 3 below.
3
4
2
1
11
Setting Up the 1088 Camera
12
Setting Up the 1088 Camera
13
Wiring Diagram 1: Camera, Light Source and Flat-Panel Monitor
Flat-Panel Monitor
DVI-I / VGA
Adapter
DVI
1088 Camera
DVI
BNC
14
Setting Up the 1088 Camera
Wiring Diagram 2: Camera, Light Source, SDC Pro HD, Sidne™, and
Flat-Panel Monitor
StrykerVision 1 Monitor
USB
Sidne™
DVI
DVI
BNC
Remote
DVI
SDC HD
15
Wiring Diagram 3: Camera, Light Source, Flat-Panel Monitor, and
CRT Monitor
StrykerVision 1 Monitor
DVI-I / VGA
Adapter
VGA
1088 Camera
DVI
DVI
S-VHS
BNC
X7000 Light Source
CRT Monitor
Note If you are using any device with unterminated analog video
inputs, you must connect a cable from the VIDEO OUT of
that device to the VIDEO IN on the monitor.
Note A color bar pattern will appear on the monitor when the
camera head is not connected to the camera console.
Follow the instructions in the “Setting Up the Camera
Head” section of this manual to connect the camera head
to the console.
16
Setting Up the 1088 Camera
Note To unplug the camera from the control unit, grasp the
knobbed portion of the connector and pull straight out. This
releases the connector latching mechanism.
17
Setting Up the Coupler
• Twist the camera end of the coupler into the camera head
(clockwise) until it forms a tight seal (see Figure 5).
18
Setting Up the 1088 Camera
(c) (a)
(b)
3. Attach a light cable from the light source to the light post on
the endoscope (c) (see Figure 6).
19
Operating the 1088 Camera
1. Point the scope at several stacked 4x4 white gauze pads, a white
laparoscopic sponge, or any clean white surface.
2. Look at the monitor and make sure that no glare is visible off of
the white surface.
3. Press and hold the WHITE BALANCE button until “WB” begins
flashing on the video monitor.
4. Continue pointing the scope at the white surface until the video
monitor indicates that white balance is “OK.” The video picture
will change color.
• If the video monitor indicates that white balance is “NOT
GOOD,” repeat the process.
• If you cannot achieve a white balance that is “OK,” refer to the
“Troubleshooting” section of this manual.
20
Operating the 1088 Camera
21
Using the Shutter Switch
The SHUTTER switch on the rear console panel activates or deactivates
the automatic shutter.
• Select “ON” to enable the automatic shutter. The camera will
automatically adjust picture brightness in response to varying
light levels.
• Select “OFF” to disable the automatic shutter. You will adjust
picture brightness by using an automatic light source.
Note You may use an automatic light source and the camera
auto shutter at the same time, but the camera auto shutter
should normally be used alone.
22
Operating the 1088 Camera
23
Using the W Button
The W button activates the white-balance function or the light/zoom
function.
• Press the W button for more than one second to activate the
white-balance function. (The white-balance function may also
be initiated using the WHITE BALANCE switch on the front
console panel. See the “Using the White Balance Switch”
section of this manual for details.)
• Press the W button for less than one second to increase the
light or zoom. Each press will raise the light or zoom level one
out of four total increments. When the light or zoom level has
reached its maximum level, pressing the W button again will
cycle the level back to the lowest setting.
24
Cleaning and Maintaining the 1088 Camera
Should the camera console need cleaning, wipe it down with a sterile
cloth and mild cleaning solution.
2. Rinse the camera head and cable in lukewarm tap water. Use a
mild enzymatic detergent and a soft brush to remove resistant
debris and bioburden from all surfaces.
3. Rinse off all soap residue with water.
4. Clean glass surfaces with alcohol.
5. Dry the camera head and cable thoroughly with a soft towel or
gauze surgical sponge.
6. Prepare the camera head and cable for disinfection by coiling the
camera cable into a loop about ten inches in diameter, avoiding
any kinks or twists in the cable.
25
Disinfecting the Camera Head
Caution Before soaking, inspect the camera cable for cuts and
breaks. Never soak a cut or a broken cable because
moisture will seep into the camera head and cause
damage. Return any camera with a damaged cable to
the Stryker Endoscopy Repair Department for service.
Never soak the camera and cable in the same tray with
sharp instruments.
26
Cleaning and Maintaining the 1088 Camera
27
Steris System
Sterrad System
28
Cleaning and Maintaining the 1088 Camera
1. Unplug the power cord from the wall outlet and remove the cord
from the camera console.
2. Unlatch the fuse holder above the AC inlet and remove it. (You
may need to press the tab on the fuse holder with a slender
screwdriver to release the latch.)
3. Replace the fuse with the same value and rating.
Warning To avoid the risk of fire, only use fuses of the value
specified on the fuse label located on the rear panel of
the camera control unit.
29
Troubleshooting
Picture is too • Adjust the light level with the camera head.
dark • Increase the light-source output.
• Check the fiberoptic light cable for excessive
broken fibers.
• Press the GAIN switch to increase illumination.
Picture is too • Adjust the light level with the camera head.
bright • Decrease the light-source output.
• Ensure GAIN is OFF.
30
Troubleshooting
Image is not • Release the scope from the coupler and then
centered reconnect it. Make sure the scope is seated
correctly in the coupler.
Optics are dirty • Rotate the scope. If dust particles in the picture
rotate, the dust is located on the scope itself.
Follow the manufacturer’s instructions for
cleaning the eyepiece and negative lens.
• If particles in the picture do not move when you
rotate the scope, the particles are located
behind the scope. Remove the scope and clean
the window on the front of the coupler with a dry
or alcohol-tipped cotton swab.
• If dust particles lie between the coupler and
camera, remove the coupler and clean the
coupler and camera windows.
31
Problem Possible Solution
32
Electromagnetic Compatibility Tables
Electromagnetic Compatibility
Like other electrical medical equipment, the 1088 Camera requires
special precautions to ensure electromagnetic compatibility with other
electrical medical devices. To ensure electromagnetic compatibility
(EMC), the 1088 Camera must be installed and operated according to the
EMC information provided in this manual.
Note The 1088 Camera has been designed and tested to comply
with IEC 60601-1-2:2001 requirements for EMC with other
devices.
1088 Camera is intended for use in the electromagnetic environment specified below. The customer
or the user of 1088 Camera should ensure that it is used in such an environment.
RF emissions CISPR
Class B
11
1088 Camera is suitable for use in all
establishments, including domestic
Harmonic emissions
Class A establishments and those directly connected to
IEC61000-3-2
the public low-voltage power supply network
that supplies buildings used for domestic
Voltage Fluctuations/ purposes.
flicker emissions Complies
IEC61000-3-3
33
Guidance and Manufacturer's Declaration: Electromagnetic Immunity
1088 Camera is intended for use in the electromagnetic environment specified below. The customer
or
the user of 1088 Camera should ensure that it is used in such an environment.
Electromagnetic
IEC 60601 Test
Immunity Test Compliance Level Environment:
Level
Guidance
Floors should be
wood, concrete, or
Electrostatic Discharge ceramic tile. If floors
±6kV contact ±2,4,6kV contact are covered with
(ESD)
±8kV air ±2,4,8kV air synthetic material,
IEC61000-4-2 the relative humidity
should be at least
30%.
Power-frequency
magnetic fields
Power frequency (50/60Hz) should be at levels
magnetic field characteristic of a
3 A/m N/A
typical location in a
IEC 61000-4-8 typical commercial
or hospital
environment.
NOTE: Ut is the a.c. mains voltage prior to application of the test level.
34
Electromagnetic Compatibility Tables
1088 Camera is intended for use in the electromagnetic environment specified below. The customer or
the user of 1088 Camera should ensure that it is used in such an environment.
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
35
Guidance and Manufacturer's Declaration: Electromagnetic Immunity
1088 Camera is intended for use in the electromagnetic environment specified below. The customer or
the user of 1088 Camera should ensure that it is used in such an environment.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast,
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the 1088 Camera system is used exceeds the applicable RF
compliance level above, the 1088 Camera system should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the 1088 Camera unit.
(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
The 1088 Camera system is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The user of the 1088 Camera system can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communications
equipment (transmitters) and the 1088 Camera system as recommended below, according to the
maximum output power of the communications equipment.
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
36
Technical Specifications
Technical Specifications
Imaging System
1/3” Progressive Scan CCDs
High Definition
Scanning System
Horizontal: 47.536 kHz
Vertical: 60.02 Hz
Video Outputs
Digital/Analog: Two Digital Video Interface (DVI)/RGBHV
1280 x 1024 (SXGA)resolution
Connector: 29-pin DVI-I
37
Operating Conditions
Temperature: 0° to 40° C
Relative Humidity:10% to 75%
Transport & Storage Conditions
Temperature: -20° to 60° C
Relative Humidity: 10% to 75%
Atmospheric Pressure: 700 hPa to 1060 hPa
Input Electrical Ratings
100-240V ~ 50/60 Hz 0.6A
Total Shipping Weight
11 lbs (5.0 kg)
Dimensions
Camera Control Unit:
12.5”w x 3.8”h x 15.25”d
(31.8cm w X 9.7cm h X 38.7cm d)
Camera Head Cable to Camera Control Unit:
10.3 feet (3.15m) sealed cable
or 20.7 foot (6.30m) sealed cable (sensed by CCU)
Enhancement
16 levels (switchable)
Classification & Approvals
Complies with medical safety standards:
IEC 60601-1: 1988 + A1: 1991 + A2: 1995
IEC 60601-2-18: 1996 + A1: 2000
CAN/CSA C22.2 No 601.1-M90
UL2601-1: 2nd Edition
AS/NZS 3200.1.0:1998
Complies with medical EMC standard:
IEC 60601-1-2:2001
Class I Equipment
Type BF Applied Part
Water Ingress Protection, IPX0-Ordinary Equipment
Continuous Operation
38
Warranty
Warranty
Stryker Endoscopy warrants the Model 1088 Medical Video Camera
against defects in both materials and workmanship to the registered
owner at the time of purchase. All components including the charge
coupled device (CCD) located in the camera head are covered by the
warranty for a period of one year from the date of purchase.
This warranty does not apply to any unit which has been subject to
misuse, neglect, improper installation or that which has been altered,
adjusted, or tampered with by any person other than Stryker Endoscopy
authorized personnel.
If upon examination by authorized service personnel, it is determined that
the malfunction is due to misuse, or abuse, warranty provisions will not
apply. An estimate of the cost of repair work will be given to the customer
prior to servicing and repairing the unit.
The customer is responsible for returning the defective equipment to the
factory at his or her own expense. Stryker Endoscopy or its
representative will service the unit, repair or replace any defective parts
thereof, and return the unit.
If, upon examination, it is determined that the fault has been caused by
misuse or abnormal conditions of operation, the repairs will be billed to
the customer as out-of-warranty repairs.
Instruments repaired under Stryker Endoscopy’s standard repair program
will be issued a thirty day warranty against defects in both materials and
workmanship, provided the original warranty period has passed.
Instruments submitted due to defects in materials and workmanship
during the warranty period will be repaired at no charge to the customer.
The warranty as set forth herein is exclusive and in lieu of all other
warranties, remedies, obligations and liabilities of Stryker Endoscopy
Inc., expressed or implied, including the implied warranties of
merchantability and fitness for use and of consequential damages. These
products are being sold only for the purpose described herein, and such
warranty only runs to the purchaser. In no event shall Stryker Endoscopy
be liable for any breach of warranty in any amount exceeding the
purchase price of the product.
No agent, employee or representative of Stryker Endoscopy has the
authority to bind the Company to any other warranty, affirmation, or
representation concerning this instrument.
This warranty is valid only to the original purchaser of Stryker Endoscopy
products directly from Stryker Endoscopy or from a Stryker Endoscopy
authorized agent. The warranty cannot be transferred or assigned by the
original purchaser.
39
Service and Claims
Caution Never open the camera head or its cable or attempt any
service not described in this manual. These units have
been factory sealed to prevent moisture from entering
the electronic components. If the camera head or the
cable seal is intentionally broken, the equipment
warranty will be void.
40
International Service
International Service
For service in the USA, call your Stryker Endoscopy representative or
Stryker Endoscopy Customer Service at 1-800-624-4422. Outside of the
USA, please contact your Stryker Endoscopy distributor at one of the
following locations:
Stryker Corporation Stryker European Rep - RA/QA
2725 Fairfield Road Manager
Kalamazoo, MI 49002 ZAC Satolas Green Pusignan
USA Av. De Satolas Green
Phone:1-269-385-2600 69881 MEYZIEU Cedex, France
Telex:224464 STRYKER KMZ Phone:33-1-48175000
Fax:1-269-385-1996 Fax:33-1-48632175
41
Stryker Finland Stryker Middle East / Africa
PL 80 (Makelankatuz) Via Della Posta
FIN 00501 Helsinki 6934 Bioggio
FINLAND Switzerland
Phone:358 (0) 9 7744 680 Phone:(4021) 212-1122
Fax:358 (0) 9 7744 6820 Fax:(4021) 212-1133
42
5900 Optical Court
San Jose, CA 95138
1-800-624-4422
1-800-729-2917
1000-400-675 Rev. A