Manual de Servicio Extractora Ardo Carum PDF
Manual de Servicio Extractora Ardo Carum PDF
Languages:Page:
de Serviceanleitung2–27
Explanation
a. Carum Breastpump i.
b. Connection opening for PumpSet
c. Control panel with LCD
d. Connection for mains cable h.
e. Bottle holder
f. EasyFreeze holder
g. Mains cable
h. Carum Container
i . Address field
j. Bottle holder j.
k. EasyFreeze holder
l. Carum
m. Mains cable
n. PumpSet
k.
f. d.
e.
a. l.
c.
b. m.
g.
n.
29 Ardo medical AG
en en
TABLE OF CONTENTS
1. IMPORTANT SAFETY DETAILS 31 8. REPAIR / REPLACEMENT 40
1.1. Warnings, instructions and advisory symbols 31 8.1. General Note: maintenance / repair 40
2. PRODUCT DESCRIPTION 32 8.2. Tools Required 40
2.1. Suckling Curve 32 8.3. Testing Equipment Required 40
8.4. Opening the Pump Assembly 40
3. APPLICATION FOR USE 32
8.5. Removal of the Pump Assembly 41
3.1. Indications 32
8.6. Installing the Pump Assembly 41
3.2. Contra- indications 32
8.7. Closing the Pump Assembly 43
4. START-UP / OPERATION 33 8.8. Replacing the Power supply complete
4.1. Icons 33 Art. Nr. 99.03.776 43
4.2. Installation 34 8.9. Replacing the Display (with 3 .5 « TFT) PCB
4.2.1. Electrical connection 34 Art. No. 99.03.496 44
4.2.2. Connecting the PumpSet 34 8.10. Replacing the Lithium Ion Battery Pack
4.2.3. Installing the bottle holder 34 Art. No. 99.03.498 45
4.2.4. Installing the EasyFreeze holder 34 8.11. Replacing the Control PCB
4.2.5. Installing the Mobile stand (optinal) 35 Art. No. 99.03.475 or Art. No. 99.03.770 46
4.3. Operation 35 8.12. Replacing the Pump Assembly
4.3.1. Switch on / switch off 35 Art. No. 99.03.782 47
4.3.2. Regulation of the vacuum and display 35 8.13. Replacing the Left and Right Cylinder
4.3.3. Cycle regulation and display 35 Art. No. 99.03.781 48
4.3.4. Switch stimulation mode / 8.14. Replacing the Housing Cover Art. No. 99.03.777 49
expression mode 36 8.15. Replacing the Housing Base Art. No. 99.03.570 50
4.3.5. Using express mode, 9. ACCESSORIESE / SPARE PARTS 52
skipping stimulation mode 36 9.1. Spare parts for repairs 52
4.3.6. Timing / Time Button 36
10. GUARANTEE 53
4.3.7. Backlight 36
11. TECHNICAL INFORMATION 53
5. ERROR ANALYSIS / ERROR DISPLAY /
11.1. Version 53
SERVICE INDICATOR 37
11.2. Technical data 53
5.1. Further possible errors when using the unit 37
11.3. Compliance 54
6. MAINTENANCE / CARE / CLEANING 37 11.4. EMV Test records 54
6.1. Visual inspection before switch-on 37
12. TRANSPORT / STORAGE / DISPOSAL 54
6.2. Cleaning / disinfection 38
12.1. Transport and storage conditions 54
6.3. General Information: Maintenance 38
12.2. Disposal 54
6.3.1. Checklist: Maintenance inspections
for rentals 38 13. ANNEX 55
6.3.2. Checklist: Maintenance inspection 13.1. EMC (Electromagnetic compatibility) –
for hospitals 38 Records of Compliance 55
30 Ardo medical AG
en en
Indicates a possible dangerous situation which, unless it is avoided, may lead to slight or
CaUtion moderate injury.
Indicates possible non-injury threatening situation,. which may however lead to the risk of, for
notE
example, property damage, or have a negative influence on functionality or reliability.
Safety Regulations
• After use, disconnect Carum from the mains.
• In the event of an accidental splash, or the unit falling into water - IMMEDIATELY disconnect it from the mains.
Do not touch the unit until it has been disconnected.
• Disconnection from the mains is only guaranteed if the mains plug is pulled out of the socket.
• The unit is designed for indoor use only. Do not use outdoors.
• Use only the original power cord or car battery cable from ARDO.
• Do not open the unit (results in voiding of the warranty).
• Carum is not water-proof. Therefore the unit must never be immersed in water or other fl uids, nor used or
stored in places where it might fall into the water. Do not use while in the bath or while taking a shower.
• Keep the mains cable and all accessories away from sources of heat.
• Do not use Carum in an area where oxygen gas is in use or sprays are being applied.
• Only use a genuine ARDO PumpSet and genuine ARDO parts. See Section 4. (Start-up / Operation) and Section
8 (Maintenance / Care / Cleaning).
• Carum should only be used for the purposes detailed in Section. 2 (Application).
31 Ardo medical AG
en en
2. PRODUCT DESCRIPTION
The Carum Pump System consists of the breast pump and a PumpSet (single or double pumpset).
The Carum breast pump is a quiet, user-friendly double piston pump for hospital use and for rental use.
The unit is equipped with a robust plastic housing (Quality: VO) and a simplified, logically organised keyboard
with a LCD display readout.
The energy-saving electrical equipment is designed for mains- and for car battery-operation, or for (optional)
battery use.
The sophisticated controls allow each mother to enjoy a comfortable and efficient pumping experience (plus a
tested suckling curve) thanks to the flexible system for adjusting vacuum and cycles.
3.1. Indications
Carum • To alleviate engorgement.
• To increase and maintain the ability to breastfeed.
• Ensure the supply of breast milk should mother and baby be separated.
• Carum can be used as both a single pump and as a double pump.
32 Ardo medical AG
en en
4. START-UP / OPERATION
4.1. Icons
Important information
“ON / OFF“ switch
Cleaning
Assembling
Compliance with general guidelines: MDD 93 / 42 EWG and MDD 2009
0123
Manufacturer
The unit is compliant with the stipulations of the WEEE 2002 / 96 / EG guidelines
Mains cable
Express mode
Stimulation mode
Device is recyclable
33 Ardo medical AG
en en
4.2. Installation
4.2.1. Electrical connection
• Mains operation: Insert mains cable (A) into the Carum mains (B) input
socket (Fig.1) and connect the local power plug to the mains power
supply.
B
• Battery operation: Before using the battery for the first time, connect
the unit to the mains power supply for 6 hours.
A
4.2.2. Connecting the PumpSet
Use of the PumpSet is detailed in the separate user manual:
“ARDO PumpSet“. The PumpSet is an applied part of the pump. C
• To use the single PumpSet:
Attach the tube connector (C) to the unit (D). (Fig. 2).
D
34 Ardo medical AG
en en
4.3. Operation
Vacuum and cycles can be adjusted independently of each other. Select
the adjustment which is the most comfortable for you.
35 Ardo medical AG
en en
4.3.7. Backlight
Using the Carum in a darkened room can be made more comfortable
by dimming the intensity of the LCD. To reduce the intensity of the
backlighting, press the ‘light’ button (to the left of the LDC) once.
The dim function can be turned on and off as required.
36 Ardo medical AG
en en
Switch the unit off, and then on again. If the error message comes up again, return the unit to the
supplier. The unit must be repaired by trained service personnel, or by the manufacturer.
Error: the unit does not work on battery power i.e. the pump does not operate.
Solution: Connect the unit to the mains to recharge the battery. The battery charging time is ca.
4 hrs. Advice: The unit may be used while the battery is being charged, but this will extend the
charging time.
Should any of these errors occur, and be confirmed, return the unit to the supplier immediately. The unit must be
repaired by trained service personnel or by the manufacturer
6. MAINTENANCE / CARE / CLEANING
The silicone pot integrated into the PumpSet constitutes a total barrier, and ensures that neither breast milk nor
pathogenic agents can get into the pump unit. The unit is maintenance-free; care of the unit, and its accessories,
is limited to inspection and cleaning.
The Carum breast pump can be repaired, but only by trained service personnel or by the manufacturer. For
this reason, only the bottle holder, the EasyFreeze holder, the connector cable, and the car charging cable, are
available for order as spare parts. See Section: Accessories / Spare Parts.
Ignoring these checks, or the cleaning instructions, may lead to risk of infection, electric-shock or damage to the unit.
37 Ardo medical AG
en en
6.2. Cleaning / disinfection
• Carum breast milk pump: Important: unplug the unit from the mains connection before cleaning.
• The Carum is not waterproof. It conforms to enclosure IP (Ingress Protection) rating IP20.
• Cleaning at home: As needed, wipe with a damp cloth, commercial detergent may be used.
• Cleaning in the hospital: Disinfect by wiping with a non-phenol based disinfectant.
• Do not put the unit into the dish-washer or into a washing-up bowl. It is not possible to clean the pump or the
power unit. Attempting to do so may cause the unit to malfunction or to become contaminated – requiring
suitable disposal. (See section 11: Transport / Storage / Disposal).
• Accessory PumpSet: Please read the separate User Manual, Section: Cleaning.
38 Ardo medical AG
en
Determination
necessary repair work.
of fault
Pump runs
Battery ok? a power source? Battery replaced
no yes O
yes
no
yes
39
Pumpset Keyboard Carum Other errors, first
Cover replaced functional?
right? yes ok? no O no O determine
no
7. DETERMINING THE CAUSE OF FAULTS
yes yes
Pumpset replace / Cylinder replaced Display ok? Display replaced Case back
O train O no O ok? no
yes yes
Display
shows error? no
yes
Pump unit Pump unit replaced Cover replaced Replace case back
ok? no O O O
yes
Control Print
O replaced?
End
The ‘fault tree’, below, should allow service personnel to diagnose the possible source of faults, and carry out the
en
Ardo medical AG
en en
8. REPAIR / REPLACEMENT
Repairs may only be carried out by suitably trained personnel.
The described checks or controls may not be changed or varied in any respect.
WARNING Defective Carum parts may only be replaced with parts recommended by the
manufacturer.
40 Ardo medical AG
en en
41 Ardo medical AG
en en
– Secure the mounting plate (Art. No. 99.03.442) onto the pump
by means of the 4 protrusions (Art. No. 99.03.444).
– Insert the left and right plugs (99.03.493) into the intended
socket positions in the housing base.
– Take care when plugging in. The Vacuum hose must not
NOTE be kinked. The hose is fitted with a loop, as shown in the
illustration.
42 Ardo medical AG
en en
– Turn the Carum over so that the pump stands on its feet.
– Connect the Carum to the mains and carry out a functionality
test.
43 Ardo medical AG
en en
– Slide the power supply unit into the correct position in the base
of the housing.
– Screw the power supply unit tightly onto the base of the
housing with the 2 screws (Art. No. 99.03.751).
8.9. Replacing the Display (with 3 .5 « TFT) PCB Art. No. 99.03.496
44 Ardo medical AG
en en
– Install the new display (with 3.5“ TFT) PCB (Art. No. 99.03.496)
into the housing.
– Position the 4 mounting plates onto the PCB (Art. No. 99.03.495)
– Secure the LCD PCB with the 4 screws (Art. No. 99.03.752) using
the TORX® screwdriver TX 10.
– Connect the membrane keyboard ribbon cable
(Art. No. 99.03.496).
8.10. Replacing the Lithium Ion Battery Pack Art. No. 99.03.498
– Insert the new Lithium Ion battery pack (Art. No. 99.03.498) into
the housing cover.
– Secure the battery pack in place with the battery holder
(Art. No. 99.03.443) and the 4 screws (Art. No. 99.03.571) using
the TORX® screwdriver TX 20.
45 Ardo medical AG
en en
8.11. Replacing the Control PCB Art. No. 99.03.475 or Art. No. 99.03.770
46 Ardo medical AG
en en
– Plug the motor cable (Art. No. 99.03.473) into the control PCB
(Art. No. 99.03.475 or Art. No. 99.03.770).
– Plug the connecting cable (Art. No. 99.03.774), control PCB to
display PCB, to the control PCB (Art. No. 99.03.475 or
Art. No. 99.03.770).
– Unscrew the two screws (Art. No. 99.03.759) in the pump unit
housing using the TORX® screwdriver TX 8.
– Release the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770)
from the holder and remove.
47 Ardo medical AG
en en
– Slide the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770)
into the new pump assembly holder and install.
– Secure the control PCB (Art. No. 99.03.475 or Art. No. 99.03.770)
with the two screws (Art. No. 99.03.759) to the pump assembly
housing, using the TORX® screwdriver TX 8.
8.13. Replacing the Left and Right Cylinder Art. No. 99.03.781
48 Ardo medical AG
en en
– Release the locking washers (Art. No. 99.03.758) from each side
using the appropriate tool.
– Remove the complete vacuum cylinder units
(Art. No. 99.03.486) from both sides.
49 Ardo medical AG
en en
– If the optional Lithium Ion battery pack (Art. No. 99.03.498) was
installed, reinstall the battery pack in the housingcover.
– Insert the battery pack into the battery pack holder
(Art. No. 99.03.443) and re-attach the holder with the 4 screws,
(Art. No. 99.03.752) using the TORX® screwdriver TX 20.
– Install the display (with 3.5“ TFT) PCB (Art. No. 99.03.496) in the
housing cover.
– Position the 4 fixing plates on the PCB (Art. No. 99.03.495).
– Secure the LCD PCB board with the 4 screws (Art. No. 99.03.752).
– Connect the membrane keyboard ribbon cable
(Art. No. 99.03.496).
50 Ardo medical AG
en en
– Replace the old housing base with the new housing base.
51 Ardo medical AG
en en
9. ACCESSORIESE / SPARE PARTS
52 Ardo medical AG
en en
10. GUARANTEE
The Carum is guaranteed for 3 years from the date of purchase. The purchase invoice serves as proof of guarantee.
The vacuum level generated is sensitive to prevailing atmospheric pressure. The higher the prevailing
atmospheric pressure, the more intense the generated vacuum – and vice versa. The vacuum values (including
tolerances) specified in the user manual are based on an atmospheric pressure of 880hPa.
53 Ardo medical AG
en en
11.3. Compliance
In accordance with guidelines: MDD 93 / 42 CEE of 14.7.1993, and 2010, in respect of medical products
and 2010. 2007 / 47 / CE, 2006 / 42 / CE, EN 60601-1, EN60601-1-2, EN ISO 10079-1, EN 60601-1-4, 0123
EN 60601-1-6, EN 60601-1-11
12. TRANSPORT / STORAGE / DISPOSAL
For the transport and storage of your Carum, we recommend using the original packaging. We further
recommend separate storage for the unit is protected from UV (e.g. sunlight / halogen lamps), heat and dust.
12.2. Disposal
The unit may be returned to the manufacturer, or be dismantled and recycled (as metal or plastic) in an
environmentally-friendly manner, through a recycling firm specialising in electrical goods. There are no special or
increased risk factors.
Manufacturer:
Ardo medical AG
Gewerbestrasse 19
6314 Unterägeri
Switzerland
Homepage: www.ardomedical.com
Email: [email protected]
54 Ardo medical AG
de de
13. ANLAGEN / ANNEX
13.1. EMV Prüfaufzeichnungen / EMC (Electromagnetic compatibility) – Records of Compliance
55 Ardo medical AG
99.03.596 / 1.2012
Ardo medical AG
Gewerbestrasse 19
6314 Unterägeri
Switzerland
T +41(0)41 754 70 70
F +41(0)41 754 70 71
[email protected]
www.ardomedical.com