BM400V2 PDF
BM400V2 PDF
AU400 e ®/AU400 ®
Chemistry Analyzer
BM400V2 AB
JUNE 2013
Beckman Coulter, Inc.
250 S. Kraemer Blvd.
Brea, CA 92821 U.S.A.
AU400e® /AU400® Chemistry Analyzer
User’s Guide
PN BM400V2AB (JUNE 2013)
製造販売業者: ベックマン・コールター株式会社
東京都江東区有明三丁目5番7号
TOC有明ウエストタワー
Beckman Coulter KK
贝克曼库尔特株式会社
东京都江东区有明三丁目5 番7 号
邮编:135-0063
Table of Contents Volume 2
Table of Contents Volume 2........................................................................................iii
Chapter F Maintenance
1. Maintenance Schedules and Preparation........................................................ F-3
2. Daily Maintenance.............................................................................................. F-9
2.1 Inspect Sample and Reagent Syringes for Leaks.................................................................. F-10
2.2 Check the Wash Solution Rolling Pump for Leaks................................................................. F-11
2.3 Inspect the Concentrated Wash Solution Level...................................................................... F-12
2.4 Inspect and Clean Sample Probe, Reagent Probe, and Mix Bars.......................................... F-13
2.5 Inspect the Stability of the Upper Cover............................................................................... F-13a
2.6 Inspect the Printer and Paper................................................................................................. F-14
2.7 Prepare for a Sample Probe Wash......................................................................................... F-15
2.8 Change DI water in the Pre-dilution Bottle............................................................................. F-16
3. Weekly Maintenance........................................................................................ F-17
3.1 Perform a W2 (Washes Cuvettes, Mix Bars, Reagent Probes, and Waste Lines)................. F-18
3.2 Perform a Photocal (after the W2).......................................................................................... F-20
3.3 Perform a Photometer Check................................................................................................. F-21
3.4 Wash the Sample Pre-dilution Bottle...................................................................................... F-22
4. Monthly Maintenance....................................................................................... F-23
4.1 Clean the Sample Probe and Reagent Probe Wash Wells.................................................... F-24
4.2 Clean the Mix Bar Wash Wells............................................................................................... F-25
4.3 Clean the Wash Nozzle Unit, Deionized-Water Tank and Filter, and Sample Probe Filter..... F-26
5. Maintenance Required Every Three Months.................................................. F-33
5.1 Replace the Wash Solution Rolling Tube............................................................................... F-34
5.2 Clean Air Filters...................................................................................................................... F-35
6. As Needed Maintenance.................................................................................. F-37
6.1 Replace the Sample and Reagent Probes............................................................................. F-38
6.2 Replace Mix Bars.................................................................................................................... F-40
6.3 Replace the Wash Nozzle Joint Tubes................................................................................... F-41
6.4 Replace Sample and Reagent Syringes................................................................................. F-44
6.5 Replace the Photometer Lamp............................................................................................... F-46
6.6 Clean or Replace Individual Cuvettes.................................................................................... F-48
6.7 Clean the Cuvettes and the Cuvette Wheel (Part of 6 month PM)......................................... F-49
6.7a Clean the Cuvettes and Cuvette Wheel after a Cuvette Wheel Flood.................................. F-50a
6.8 Perform a W1 Procedure........................................................................................................ F-51
6.9 Clean Belts and Rack Feed areas.......................................................................................... F-52
6.10 Clean the Wash Nozzle Unit................................................................................................... F-53
6.11 Clean or Replace the Sample Probe Filter............................................................................. F-56
6.12 Clean the Deionized-Water Tank............................................................................................ F-58
6.13 Clean or Replace the Deionized-Water Filter......................................................................... F-60
6.14 Clean or Replace the Static Discharge Brushes.................................................................... F-62
7. ISE Maintenance .............................................................................................. F-63
7.1 Daily ISE Maintenance........................................................................................................... F-64
7.1.1 Inspect the ISE Reagent (Buffer) Syringe for Leaks................................................... F-65
7.1.2 ISE Cleaning............................................................................................................... F-66
7.2 Weekly ISE Maintenance........................................................................................................ F-68
7.2.1 Perform a Selectivity Check for the Na/K Electrodes................................................. F-69
7.3 Maintenance Every Two Weeks............................................................................................. F-70
7.3.1 Clean the mix bars, liquid level sensors, sample pot and sample pot tubing. ........... F-71
7.4 ISE Maintenance Required Every Three Months................................................................... F-75
7.4.1 Replace the Mixture and Mid-Standard Pump Roller Tubing..................................... F-76
7.4.2 Replace Valve Tubing................................................................................................. F-78
7.5 As Needed ISE Maintenance................................................................................................. F-80
7.5.1 Replace the Reference Electrode............................................................................... F-81
7.5.2 Add Reference Electrode Solution............................................................................. F-83
Automated Chemistry Analyzer
AU400e/AU400 User's Guide iii
June, 2013
7.5.3 Replace Reagents...................................................................................................... F-84
7.5.4 Replace the Na, K, or Cl Electrode............................................................................. F-86
7.5.5 ISE Cleaning Procedure............................................................................................. F-88
7.5.6 Replace the reagent buffer syringe............................................................................. F-88
Chapter H Troubleshooting
To Begin Troubleshooting, Answer the Following Questions............................................................H-3
1. Troubleshooting Using On-Line Help...............................................................H-5
2. Troubleshooting the Analyzer...........................................................................H-7
2.1 Data Problems..........................................................................................................................H-8
2.1.1 Checking Abnormal Data in the Software Screens.......................................................H-9
2.1.2 Troubleshooting Software........................................................................................... H-11
2.1.3 Troubleshooting Reagents and Samples....................................................................H-13
2.1.4 Troubleshooting Mechanical Problems.......................................................................H-16
2.2 System Problems....................................................................................................................H-21
2.3 Data Processor Problems.......................................................................................................H-25
2.4 Recovering from an Emergency Stop or Power Loss.............................................................H-28
2.5 Recovering from a Cuvette Wheel Overflow..........................................................................H-29
2.5.1 What Causes a Overflow?..........................................................................................H-29
2.5.2 Recognizing a Overflow..............................................................................................H-29
2.5.3 Recovering from a Overflow......................................................................................H-30
2.5.4 After the Overflow Problem Is Fixed...........................................................................H-30
3. Troubleshooting the ISE .................................................................................H-31
To Begin Troubleshooting the ISE, Answer the Following Questions.....................................H-32
3.1 ISE Sample Requirements...................................................................................................H-34
3.2 Dispensing System................................................................................................................H-35
3.2.1 Sample Probe and ISE Reagent Syringe...................................................................H-36
3.2.2 Pump Tubing:..............................................................................................................H-37
3.2.3 Mix Bar........................................................................................................................H-38
3.2.4 Sample Pot.................................................................................................................H-39
3.2.5 Pinch Valve Tubing.....................................................................................................H-40
3.3 Measuring Components........................................................................................................H-41
3.3.1 O-rings........................................................................................................................H-42
3.3.2 Electrodes...................................................................................................................H-43
3.3.3 Mixture Pump Tubing..................................................................................................H-45
3.3.4 Thermistor...................................................................................................................H-46
3.3.5 Flowcell Block.............................................................................................................H-47
3.4 Calibration Errors...................................................................................................................H-48
3.5 Selectivity Check....................................................................................................................H-50
3.6 Sequential Sample Measure................................................................................................H-51
3.6.1 Shifts & Trends...........................................................................................................H-52
3.6.2 How to Check Reagent Integrity.................................................................................H-54
Chapter I Specifications
1. System Operation During Analysis................................................................... I-3
Operation Process..................................................................................................................... I-3
Operation Modes....................................................................................................................... I-4
2. Analyzer Hardware Configuration..................................................................... I-5
Operation Switches................................................................................................................... I-5
Rack Feeder Unit....................................................................................................................... I-7
Sample Probe and Reagent Probe Units.................................................................................. I-9
The Mix Unit............................................................................................................................. I-10
The Wash Nozzle Unit..............................................................................................................I-11
iv
Automated Chemistry Analyzer
AU400e/AU400 User's Guide
June, 2013
The STAT Table....................................................................................................................... I-12
Sample Syringe and Reagent Syringes................................................................................... I-14
Rolling Pump Unit.................................................................................................................... I-15
Reagent Refrigeration Unit...................................................................................................... I-16
The Incubator.......................................................................................................................... I-18
The Photometer Unit............................................................................................................... I-19
Tank Storage........................................................................................................................... I-20
Breakers and Fuse.................................................................................................................. I-21
The ISE Unit ........................................................................................................................... I-22
The ISE Reagent Syringe ....................................................................................................... I-23
The ISE Reagent Bottles......................................................................................................... I-24
3. System Connections......................................................................................... I-25
4. Major System Specifications............................................................................ I-26
Specifications Related to System Installation.......................................................................... I-26
Sampling Specifications.......................................................................................................... I-27
Reaction Unit Specifications.................................................................................................... I-28
Photometer Unit Specifications............................................................................................... I-29
Data Processor Unit Specifications......................................................................................... I-29
Analysis Processing Specifications......................................................................................... I-29
Input/Output Specifications...................................................................................................... I-30
ISE Unit Specifications............................................................................................................ I-34
5. Calculations....................................................................................................... I-35
5.1 Reagent Blank (Zero Adjustment)........................................................................................... I-35
5.2 Endpoint Assay........................................................................................................................ I-37
5.3 Rate Assay............................................................................................................................... I-39
5.4 Fixed Point Assay.................................................................................................................... I-40
5.5 Sample Blank........................................................................................................................... I-41
5.6 LIH........................................................................................................................................... I-42
6. Quality Control.................................................................................................. I-43
7. AU400 Terminology........................................................................................... I-49
Introduction
This chapter includes detailed information about the maintenance required for each system
component. To ensure data is accurate, perform maintenance on schedule, and keep a record on
the maintenance/inspection check lists provided in this chapter.
Caution
Follow laboratory safety guidelines when performing maintenance procedures. Observe all warning and
caution instructions listed in the following procedures. Also, do not place any body parts in the path
of moving equipment, unless you are certain that the analyzer is in Stop or Standby and that diagnostic
functions are not activated.
Chapter F Maintenance
Introduction............................................................................................................... F-1
1. Maintenance Schedules and Preparation........................................................ F-3
2. Daily Maintenance.............................................................................................. F-9
2.1 Inspect Sample and Reagent Syringes for Leaks.................................................................. F-10
2.2 Check the Wash Solution Rolling Pump for Leaks................................................................. F-11
2.3 Inspect the Concentrated Wash Solution Level...................................................................... F-12
2.4 Inspect and Clean Sample Probe, Reagent Probe, and Mix Bars.......................................... F-13
2.5 Inspect the Stability of the Upper Cover............................................................................... F-13a
2.6 Inspect the Printer and Paper................................................................................................. F-14
2.7 Prepare for a Sample Probe Wash......................................................................................... F-15
2.8 Change DI water in the Pre-dilution Bottle............................................................................. F-16
3. Weekly Maintenance........................................................................................ F-17
3.1 Perform a W2 (Washes Cuvettes, Mix Bars, Reagent Probes, and Waste Lines)................. F-18
3.2 Perform a Photocal (after the W2).......................................................................................... F-20
3.3 Perform a Photometer Check................................................................................................. F-21
3.4 Wash the Sample Pre-dilution Bottle...................................................................................... F-22
4. Monthly Maintenance....................................................................................... F-23
4.1 Clean the Sample Probe and Reagent Probe Wash Wells.................................................... F-24
4.2 Clean the Mix Bar Wash Wells............................................................................................... F-25
4.3 Clean the Wash Nozzle Unit, Deionized-Water Tank and Filter, and Sample Probe Filter..... F-26
5. Maintenance Required Every Three Months.................................................. F-33
5.1 Replace the Wash Solution Rolling Tube............................................................................... F-34
5.2 Clean Air Filters...................................................................................................................... F-35
Automated Chemistry Analyzer
AU400/AU400e User's Guide Chapter F - Maintenance F-1
April 14, 2010
6. As Needed Maintenance.................................................................................. F-37
6.1 Replace the Sample and Reagent Probes............................................................................. F-38
6.2 Replace Mix Bars.................................................................................................................... F-40
6.3 Replace the Wash Nozzle Joint Tubes................................................................................... F-41
6.4 Replace Sample and Reagent Syringes................................................................................. F-44
6.5 Replace the Photometer Lamp (Part of 6 month PM)............................................................ F-46
6.6 Clean or Replace Individual Cuvettes.................................................................................... F-48
6.7 Clean the Cuvettes and the Cuvette Wheel (Part of 6 month PM)......................................... F-49
6.7a Clean the Cuvettes and Cuvette Wheel after a Cuvette Wheel Flood.................................. F-50a
6.8 Perform a W1 Procedure........................................................................................................ F-51
6.9 Clean Belts and Rack Feed areas.......................................................................................... F-52
6.10 Clean the Wash Nozzle Unit................................................................................................... F-53
6.11 Clean or Replace the Sample Probe Filter............................................................................. F-56
6.12 Clean the Deionized-Water Tank............................................................................................ F-58
6.13 Clean or Replace the Deionized-Water Filter......................................................................... F-60
6.14 Clean or Replace the Static Discharge Brushes.................................................................... F-62
7. ISE Maintenance .............................................................................................. F-63
7.1 Daily ISE Maintenance........................................................................................................... F-64
7.1.1 Inspect the ISE Reagent (Buffer) Syringe for Leaks................................................... F-65
7.1.2 ISE Cleaning............................................................................................................... F-66
7.2 Weekly ISE Maintenance........................................................................................................ F-68
7.2.1 Perform a Selectivity Check for the Na/K Electrodes................................................. F-69
7.3 Maintenance Every Two Weeks............................................................................................. F-70
7.3.1 Clean the mix bars, liquid level sensors, sample pot and sample pot tubing. ........... F-71
7.4 ISE Maintenance Required Every Three Months................................................................... F-75
7.4.1 Replace the Mixture and Mid-Standard Pump Roller Tubing..................................... F-76
7.4.2 Replace Valve Tubing................................................................................................. F-78
7.5 As Needed ISE Maintenance................................................................................................. F-80
7.5.1 Replace the Reference Electrode............................................................................... F-81
7.5.2 Add Reference Electrode Solution............................................................................. F-83
7.5.3 Replace Reagents...................................................................................................... F-84
7.5.4 Replace the Na, K, or Cl Electrode............................................................................. F-86
7.5.5 ISE Cleaning Procedure............................................................................................. F-88
7.5.6 Replace the reagent buffer syringe............................................................................. F-88
Consumable Management
ALARMCLEAR
Additional Instruction
• Please see additional instruction on maintenance procedures in the maintenance videos
available within online help.
Bleach
For the purposes of maintenance procedures on Beckman Coulter Chemistry-Immuno systems,
undiluted bleach is defined as 5-10% Sodium hypochlorite, and may also contain up to 0.5-1%
Sodium hydroxide. Many brands of liquid household bleach contain impurities or lack sufficient
concentration of sodium hypochlorite for cleaning. Consequently, only the following should be
used. Do not use scented formulations of these products:
• Clorox®
• Javex-5™
• Reagent grade or other purified preparations of sodium hypochlorite
Prepare working concentrations by mixing bleach and DI water in the following proportions:
Concentration Volume of Clorox Volume of DI Water
10% 10 + 90
20% 20 + 80
NOTE: Sodium hypochlorite quickly loses its germicidal action upon exposure to light,
therefore, all dilutions should be made daily and in an area away from direct light. A 24-hour
preparation of this solution under an average fluorescent light environment is not sufficient to
cause degradation of the solution, provided the solution is placed into a closed container. Clear
containers are suitable for 24-hour storage of this solution.
F-5
DAILY ANALYZER 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Chapter F - Maintenance
2.1 Inspect Sample and
Reagent Syringes for
leaks
2.2 Check Wash Solution
Rolling Pump for leaks
2.3 Inspect concentrated
wash solution level
2.4 Inspect and clean
Sample & Reagent
probes and Mix Bars
2.5 Inspect the Stability of
the Upper Cover
2.6 Inspect printer and
paper
2.7 Prepare for a Sample
Probe wash
2.8 Change DI water in the
pre-dilution bottle
DAILY ISE 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
7.1.1 Inspect the ISE Reagent
Syringe for leaks
7.1.2 ISE Cleaning
WEEKLY ANALYZER 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Chapter F - Maintenance
7.4.1 Replace Mixture & Mid-
standard Pump Roller
Tubing (Part of 6 month
PM. User must perform
procedure alternately.)
7.4.2 Replace Valve Tubing
(Part of 6 month PM.
User must perform
procedure alternately.)
F-6
Month _______________ Year ________
F-7
AS NEEDED ANALYZER 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
Chapter F - Maintenance
6.1 Replace Cuvettes
6.2 Replace the Sample
and Reagent Probes
6.3 Replace Mix Bars
6.4 Replace the Wash
Nozzle Joint Tubes
6.5 Replace Sample and
Reagent Syringes
6.6 Replace DI Water Filter
(Part of 6 month PM)
6.7 Replace Sample Probe
Filter (Part of 6 month
PM)
6.8 Replace Photometer
Lamp (Part of 6 month
PM)
6.9 Clean the Cuvettes and
the Cuvette Wheel (Part
of 6 month PM)
6.10 Clean or Replace
Individual Cuvettes
6.11 Perform a W1
Procedure
6.12 Clean Rack Feed Areas
Caution
Do not place any body parts in the path of moving equipment, unless you are certain that the analyzer is
in Stop or Standby and that diagnostic functions are not activated.
Contents
Caution
Do not place any body parts in the path of moving
equipment, unless you are certain that the analyzer is in
R S ISE R Stop or Standby and that diagnostic functions are not
activated.
Daily Procedure 3. Check the two fixing screws on the case head.
Prepare the following: If necessary, tighten them by turning them
OLP2061E
• Dry, clean cloth
clockwise.
4. Verify that the bottom screw fits securely against
1. Open the right front cover. the piston.
2. Check the following areas for leaks: bottom of 5. Visually check for leaks inside the syringe case.
the syringe cases, the case head, syringe case, Replace any damaged component.
area around the fixing screws and the tubing.
(Also make sure the tubing is not crimped.) To perform syringe verification for
troubleshooting, perform the following steps:
ISE reagent syringe
6. Loosen the bottom fixing screw first, then
Sample syringe
Mounting groove the top fixing nut. Pull the syringe and case
Reagent syringe
forward.
Fixing nut 7. Verify that the syringe provides a smooth,
Case head resistant pull by pulling on the piston.
FR 8. Turn the syringe case by hand. If the syringe
ON
T case is loose, turn it clockwise toward the case
Tubing head to tighten.
connectors
9. Visually check each case head for cracks. If
Syringe case
there are cracks on the case head, replace the
Piston fixing screw case head.
10. Pull the syringe from the case head and verify
See Also that there is one O-ring, and that it is not
For information about replacing and troubleshooting damaged.
syringes, refer to the“ As Needed Maintenance”
section of this Chapter and the Troubleshooting 11. Re-install the syringe and secure the top fixing
Chapter. nut first, and then the bottom fixing screw.
Note: Verify that the correct size syringe
(reagent or sample) is placed in the appropriate
Caution
position. To replace a syringe, refer to the “As
If your skin comes in contact with liquid, immediately Needed” section of this chapter.
rinse it with water.
12. Close the right front cover.
AU400e
Wash solution
rolling pump
Rolling tube
AU400
Connectors
1. Place the analyzer in standby or stop. 7. Check the tube connector. If it is loose, turn
clockwise and make sure it is finger tight.
2. Open the front left cover. Check for leaks again. If the tube still leaks,
replace it.
Caution
8. Close the left front cover.
If your fingers come into contact with any liquid,
immediately wash them with water. The wash solution
pump dispenses concentrated wash solution. Note
AU400 analyzers contain two Wash Solution Rolling
3. Blot the pump with the dry clean cloth. If the Pumps. AU400e analyzers contain one pump.
pump is wet, blot until completely dry.
4. Visually check the wash solution rolling tube for
cracks.
5. If cracks or other damage is found, replace the
wash solution rolling tube.
6. Verify that the labels on the tube match the
labels on the pump.
Master detergent
tank A
(factory optional) Master detergent
Master detergent tank A
NT tank B (factory optional)
FRO
Master detergent
tank B
NT
FRO
Daily Procedure
Prepare the following:
Warning
• •
• If the wash solution spills, wipe it up immediately
Wash Solution (#OSR0001) for tanks A (factory with a dry cloth. If the wash solution is not wiped
optional) & B up, a toxic gas may be generated and system
components could corrode.
1. Open the right front cover. • Be sure to use the correct wash solution
(OSR0001).
Warning
When removing the cap from the concentrated wash
solution tank, do not allow your hands and clothing to
come into contact with the wash solution. If contact
occurs, immediately wash your hands with water for
at least fifteen minutes and refer to the MSDS sheet.
If wash solution comes into +contact with your eyes
or it is ingested, immediately rinse with water, avoid
vomiting and consult the MSDS sheet.
Stream
2. Visually check the sample probe, reagent probe,
and mix bars for bends or scratches.
If they are bent or scratched, replace them.
Caution
Replace the mix bar if there are chips in the teflon
coating. Chipped Mix Bars can cause carry-over and
inaccurate results.
7. Close the upper cover.
8. Optional Step: For inspection and cleaning
procedures, refer to the following. Chips
See Also
For replacement procedures, refer to the "As
Needed Maintenance" section of this chapter.
■ Inspecting and Cleaning the Sample Probes,
Reagent Probes, and Mix Bars
1. Inspect the probes and mix bars. If there is
staining or crystallization on the outside of the See Also
sample probe, reagent probe, or mix bars, clean For information about how to replace the sample or
the part with an Alcohol Prep. reagent probe or mix bars refer to the "As Needed
Maintenance” section of this chapter.
Automated Chemistry Analyzer
AU400/AU400e User's Guide Chapter F - Maintenance F-13
January 28, 2005
2.5 Inspect the Stability of the Upper Cover
Before starting daily analysis, check the stability of the upper cover of the analyzer to verify that
it is stable and remains in the upright position when raised. If the upper cover starts to descend
when opened, have the cover supports inspected and replaced by Beckman Coulter authorized
personnel.
Upper Cover
See Also
For information about how to use the printer, refer to the operator's manual supplied with the
printer.
Daily Procedure
Prepare the following:
• Printer paper
1. Check that the printer is on. If not, turn on the power to the printer. Make sure that the
printer is online.
2. Check that the paper is loaded correctly and that a sufficient amount of paper remains. If the
paper is not loaded correctly, reload it.
See Also
For information about remedies for printer problems and how to load the paper, refer to the
operator's manual supplied with the printer.
Note:
By following good lab practice and ensuring a clean, sealable container is used, diluted OSR0001 can be
prepared ahead of time. Diluted OSR0001 is good for 6 months from the date of preparation.
W1 Position
OLP2017f
Pre-dilution bottle
T
ON Pre-dilution position
FR
Caution
Place the bottle in the compartment so it does not protrude above the surface of the analyzer top. If it is
not placed properly, reagent probe crashes could occur. Also, do not place the cap on the bottle when it
is on the analyzer, This also causes probe crashes.
Contents
3.1 Perform a W2 (Washes Cuvettes, Mix Bars, Reagent Probes, and Waste Lines)................. F-18
3.2 Perform a Photocal (after the W2).......................................................................................... F-20
3.3 Perform a Photometer Check................................................................................................. F-21
3.4 Wash the Sample Pre-dilution Bottle...................................................................................... F-22
OLP2122
The W2 can be performed in conjunction with an End Process. Start the W2, wait until W2 Positioningdisplays
pin
in
the mode area. Verify that the system does not indicate “insufficient detergent? by allowing the timer to
reach 24 minutes. Exit System Status to Main Menu, then select the End Process STAT tablekey. The system can
5
20
S-
K-
L
ON
SE
6
after a W2.
L-N
a
19
FR
7
S-H
18
8
17
9
CLE W1
16
AN
W2 10
15
11
14
13 12
S
TA
T
E
N
S
D
TA
T
S
E
T
IS
E
P
R
IM
E
STAT table top view
STAT table cover
(small) 22
1 2
3
21
STAT table
4
U-H U-L
covers (large)
20
5
S-
L
K-S
EL
-N
a
19
6
S-H
18
Positioning pin
7
17
8
STAT table
16
CL
1 2
3
1
22
EA
9
W
4
N
21
U-H U-L
5
W2
T
20
S-L
W2
K-
ON
SE
15
L-
Na
10
19
FR
7
S-H
18
14
8
11
Position
17
13
9
CLE W1 12
16
AN
W2 10
15
11
Weekly Procedure
14
13 12
S
TA
T
E
N
S
D
TA
T
S
E
T
IS
E
P
R
IM
E
bottle.)
4
U-H U-L
5
S-
(part #ZM0062).
S-H
N W
W2
10
process.
• A W2 requires approximately 25 minutes to 1. Fill a sample tube (ZM0062) for the W2 with at
be performed. Alternate the cleaning solution least 3 mLs of HCl or bleach. If HCl was used
each week using a freshly prepared 10% bleach previously, then use bleach this time.
solution one week, and 1N HCl the next week. 2. Remove the two STAT table covers.
Cleaning Solution
Bottle
T
ON Pre-dilution position
FR
See Also
For information about how to perform a photocal,
refer to the "Weekly Maintenance” section in this
chapter.
7. Make sure the system power is on.
8. Select the [System Status] icon. The System
Status screen will appear.
9. Press function key F6 (W2 start). The W2 Start
window will appear. This window allows ISE
cleaning during the W2. Place a container of
ISE cleaning solution in the "Clean" position on
Weekly Procedure
Prepare the following: 5. Select a cuvette number (1-88) to perform the
• DI water check.
6. Add approximately 500 ul of DI water to the
Caution cuvette number selected with a pipet.
To obtain accurate results, do not start a photometer 7. Select "Yes" at the Start Measure prompt.
check until the photometer lamp has stabilized after
system start-up (approximately 20 minutes). 8. The 0% and 100% values will display on the
screen.
Acceptable Ranges: 0% = -5 to 25
Instruction
The analyzer must be in the Stop mode to enter
100% = 0.01 to 1.7
diagnostics. If the numbers exceed these ranges, replace
the photometer lamp and perform a photocal.
1. Transfer the unit to the Stop Mode, then select After the lamp is replaced, the operator must
[Maintenance], [Maker Maintenance], [ANL repeat a photocal with the new lamp. Refer to
DIAG]. the replace Lamp procedure in the As Needed
2. Select the "Combination" tab located at the top Maintenance section.
of the screen.
Instruction
3. Select "Photometer Check A/0-100%."
The water will automatically be removed from the
4. Select the "Print" check box to print the results. cuvettes during the next wash or start process.
Pre-dilution bottle
T
ON Pre-dilution position
FR
Caution
Do not leave any bleach residue in the bottle! Residue
will adversely affect analysis results.
Contents
4.1 Clean the Sample Probe and Reagent Probe Wash Wells.................................................... F-24
4.2 Clean the Mix Bar Wash Wells............................................................................................... F-25
4.3 Clean the Wash Nozzle Unit, Deionized-Water Tank and Filter, and Sample Probe Filter..... F-26
OLP2018E
Reagent probe
Reagent probe
wash well
Sample probe
T
ON
FR Sample probe wash well
Monthly Procedure
Caution
Prepare the following:
If bleach spills, immediately wash the affected area
• Squirt bottle with freshly prepared 10% bleach with water.
solution
• Cotton swab 6. Squirt the 10% bleach solution into each wash
well while cleaning the inside with a cotton
1. Open the upper cover. swab. During cleaning, be careful not to
2. Verify that the analyzer is in the Warm-up or damage the sample and reagent probe tips.
Standby Mode. 7. Press the stat rotation/diag button. The sample
3. Select [Maintenance], [ANL Maintenance]. and reagent probe will return to its home
position over the wash well.
4. Select “D/Cleaning Wash Wells.”
8. Select “F/Prime Washing-line” on the
5. Press the STAT ROTATION/DIAG button. The [Maintenance], [ANL Maintenance] screen.
sample and reagent probes move to the cuvette
wheel position. 9. Press the STAT ROTATION/DIAG button. The
DI water is dispensed from the sample and
reagent probes into the wash wells.
9. Close the upper cover.
Mixing bar
Mixing unit
T
ON
FR
Monthly Procedure
Prepare the following:
• Squirt bottle with freshly prepared 10% bleach
solution
• Cotton swab
1. Open the upper cover.
2. Verify that the analyzer is in the Warm-up or
Standby Mode. Caution
3. Manually turn the mix bar units so the mix bars Do not spill the bleach solution outside the mix bar
are not over the wash wells. wash well. If it spills, immediately wipe it up.
4. Squirt the 10% bleach solution into each wash
well while cleaning the inside with a cotton 6. Select “F/Prime Washing-line” on the
swab. [Maintenance], [ANL Maintenance] screen.
5. Realign the mix bar unit as closely as possible 7. Press the STAT ROTATION/DIAG button. The
to avoid errors. mix bars will go through one wash sequence.
8. Close the upper cover.
Deionized-Water Tank
Regular cleaning prevents the accumulation of deposits or bacteria that could affect analysis.
Deionized-Water Filter
If the deionized-water filter is dirty, the system may generate abnormal analysis data.
Monthly Procedure
Prepare the following:
• Dry, clean cloth • Freshly prepared 20% bleach
• Sonicator filled with DI water • Extra DI tank (In start-up kit)
• Basin
Tube mounting
joint manifold
Knob
Wash nozzle
Wash nozzle FR
ON
station T
Caution
After placing the wash nozzle unit in the sonicator with
DI water, verify that the tips of the wash nozzle do not
rest on the bottom of the sonicator. When the sonicator
is running the wash nozzle to vibrate and can damage
the tips.
Instruction
• When cleaning the wash nozzle unit using a
sonicator, be careful not to scratch the wash
Remove and Clean the Wash Nozzle Unit nozzles.
1. Open the upper cover. • Do not scratch or tear the joints and tubes. A leak
may result from the scratched part and the cuvette
2. Verify that the analyzer is in the Warm-up or wheel may overflow.
Standby Mode.
3. Select [Maintenance], [ANL Maintenance]. Tips
4. Select “E/Replacing Wash Nozzle.” The sonicator is the optimal device for cleaning the
wash nozzles. If a sonicator is not available, use DI
5. Press the STAT ROTATION/DIAG button. The water. While pouring water into the wash nozzles, clean
liquid in the tubing on the wash nozzle unit is each nozzle hole using the supplied stylet. Rinse the
drained. nozzles in DI water, then dry them with a soft cloth.
6. Open the rear cover.
9. Clean the tips of the nozzles for 15 minutes
7. Remove the four tube mounting joint manifolds using DI water in a sonicator. No wash solution
from the wash nozzle unit. is required.
Do not submerge the springs on top of these
Instruction nozzles - only submerge the tips of the wash
• When handling the wash nozzles, be careful not to nozzles, including the shortest nozzles tips.
scratch them.
• When loosening the knob on the wash nozzle unit,
do not loosen the positioning screws on both sides
of the knob. These screws are used for positioning
the wash nozzle unit.
Joint
Connecting hose
Button Fixture
Top
Filter case
Button
NT Joint
FRO Basin
Instruction TOP
Perform an End Process before removing the Sample
Probe Filter, DI-Water Filter, or DI-Water Tank. If this
Button
procedure is performed with the system sub-power on,
water could spill from tubing and tanks.
Case
6. Unscrew the Sample Probe Filter Case and
remove the sample probe filter.
7. Clean the Sample Probe Filter for 10 minutes,
using DI water in a sonicator. If a sonicator
is not available, rinse in DI water. While
pouring DI water onto the filter, scrub it with a
Button
toothbrush.
8. Inspect the O-ring, Filter Positioning Ring,
and the inside of the case. Remove any
accumulations with a damp towel.
Connector
Deionized-water tank
Float switch
Fixture
Joint
Tubes
Filter case Deionized-water
drain hose
T
ON
FR
Basin
Instruction
Loosen the DI-Water Filter case over the basin. The
OLP2023E
deionized water in the filter case will drain from the
joint. If the deionized water spills onto the system,
immediately wipe it off with a dry, clean cloth.
1. Unplug connector #240.
2. Disconnect the joint of the deionized water
drain hose from the tank by pressing the gray
Quick Disconnect button.
3. Pull the DI water tank forward while removing
the tubes inserted in the tank. Lay the tubes in
Water supply tube
the basin.
4. Remove the deionized-water tank from the Filter case
instrument. Basin
Caution
If the tank is empty, and the pump runs without water, Cap
the pump may be damaged.
Button
Instruction
Be sure to attach the tube mounting joint manifolds 5. Repeat the previous step two or three times
in their original places. If a manifold is attached until the air in the affected tubing is completely
incorrectly, normal analysis cannot be performed. removed.
6. Close the left front cover and the upper cover.
4. Match the two positioning holes on the wash
nozzle unit with the positioning screws, then
secure the station by finger-tightening the knob.
5. Attach the manifolds, hand tighten them firmly.
When attaching the manifolds, match the color
connections and tighten them firmly. For
information about the tube mounting joint
positions, refer to the drawing on the following
page. Make sure the O-rings are in place on
each manifold and not damaged.
OLP2120.eps
Type 1 Type 2
A - Aspirate
B - Overflow
C - Dispense
Contents
Rolling tubes
Wash solution
rolling pump A
Included on AU400
only
Relay tubes
T
ON
FR
Matching
Every Three Months numbers
Prepare the following:
• New tubing MU9623
1. Place the analyzer in Warm-up or Standby.
2. Open the left front cover.
3. Select [Maintenance], [ANL Maintenance].
4. Select "G/Replace Detergent Tube." 7. Connect the new tubing and rotate the
5. Press the STAT ROTATION/DIAG button. The connectors to tighten them.
pump rotates in reverse. The concentrated wash 8. Stretch the tubing around the rolling pump.
solution in the tube returns to the tank. Slip the tube into the groove as shown in the
illustration. Note: The numbers on the tube
Note should match the numbers on the side of the
The AU400 includes two pumps. The AU400e only pumps.
contains one pump.
9. Press the STAT ROTATION/DIAG button.
The pump rotates counterclockwise. The
Caution concentrated wash solution fills the tubing.
Handle the tubing with care or wash solution could 10. Check for leaks in the tubing then close the left
splash on skin. If wash solution splashes on your skin, front cover.
immediately wash the area with water.
FR
ON
T
Air filter(S)
MU9593
Air filter(L)
MU9594
Air filter(S)
MU9593
FR
ON
T
Caution
Do not clean the air filters with water. Moisture could 3. Vacuum the air filters thoroughly.
enter the system and damage the internal electronics. 4. Place the air filters back in their original
locations.
1. Confirm that the system is is in Standby mode.
2. Remove the three air filters as shown in the
figure.
Contents
OLP2053E
T
ON
FR
As Needed
Probe connector
Instruction ON
T
FR Sample probe wash well
Check that the sample probe is positioned over the
wash well before replacing it. Liquid will drip during
probe replacement.
10. Tighten the probe connector to secure the
sample probe. Tighten the connector firmly.
1. Verify that the analyzer is turned on and is in the
Warm-up or Standby Mode. 11. Replace the sample probe cover. First place it
on the arm from the probe side. A click should
2. Open the upper cover.
be heard.
3. Select [Maintenance], [ANL Maintenance].
12. From [Maintenance], [ANL Maintenance] select
4. Select "A/Replace S Probe & Syringe." "F/Prime Washing-line."
5. Press the STAT ROTATION/DIAG button. 13. Press the STAT ROTATION/DIAG button. The
Water will drain from the sample probe tubing. DI water will be dispensed from the probe tip.
Check that the DI water is dispensed in a thin
6. Remove the sample probe cover. The probe
straight stream. It should not spray or dispense
cover has wedges inside the cover. While
at an angle.
spreading both sides of the cover from the inside
by hand, lift the cover to remove it from the 14. Close the upper cover.
probe arm.
7. Disconnect the probe connector.
Instruction
When replacing the sample probe, be careful not to
bend or scratch the probe tip.
Probe connector
Reagent probe
As Needed Instruction
Reagent Probe Replacement Procedure When replacing the reagent probe, be careful not to
bend or scratch the probe tip.
Prepare the following:
• Reagent probe MU9958
8. While holding the probe connector, pull the
reagent probe upward.
Instruction
9. Place a new reagent probe through the top of the
Check that the reagent probe is positioned over the
wash well before replacing it. Liquid will drip during holder.
probe displacement. 10. Tighten the probe connector to secure the probe.
Tighten the connector firmly.
1. Verify that the analyzer is turned on and is in the 11. Replace the probe cover. First place it on the
Warm-up or Standby Mode. arm from the probe side. A click should be
heard.
2. Open the upper cover.
12. From [Maintenance], [ANL Maintenance], select
3. Select [Maintenance], [ANL Maintenance].
"F/Prime Washing-line."
4. Select "B/Replace R Probe & Syringe."
13. Press the STAT ROTATION/DIAG button. The
5. Press the STAT ROTATION/DIAG button. DI water will be dispensed from the probe tip.
Water will drain from the reagent probe tubing. Check that the DI water is dispensed in a thin
straight stream. It should not spray or dispense
6. Remove the reagent probe cover. The probe
at an angle.
cover has wedges inside the cover. While
spreading both sides of the cover from the inside 14. Close the upper cover.
by hand, lift the cover to remove it from the
probe arm.
7. Disconnect the probe connector.
Mix bar
Mix unit
T
ON
FR
As Needed
Prepare the following:
Chips
• New Mix Bar MU9599
Caution
Replace the mix bars while the mix unit drive is not
operating. Personal injury may result if replacement of
a mix bar is attempted during operation of the mix unit.
Tube
Tube mounting
joint
Tube mounting
joint manifold
Knob
Joint tube
Tube
Instruction
• When handling wash nozzles, be careful not to
scratch them.
Joint tube
• When loosening the knob on the wash nozzle unit,
do not loosen the positioning screws on both sides
of the knob. These screws are used to position the
wash nozzle unit.
Wash nozzle
station
8. Loosen the knob on the wash nozzle unit, then
remove the wash nozzle unit along with the
tubing. Put it on a flat working surface.
As Needed
9. Remove the tubes to be replaced.
Prepare the following:
• New joint tube ZM1131 (3 pieces per set) Caution
1. Open the upper cover. Replace tubes one by one. If a wrong nozzle and tube
are connected, problems could occur. Refer to the
2. Verify that the analyzer is turned on and is in the drawing on the next page for correct tube and nozzle
Warm-up or Standby Mode. positions.
Joint tube
Tube
Cross-sectional View
Joint tube
Tube
Joint tube
Position both ends
of the tube and Approx.
nozzle in the center 1mm
Wash nozzle
Wash nozzle of the joint tube.
station
station
Nozzle
A - Aspirate
B - Overflow
C - Dispense
Tubing
Connectors
FR
ON
Syringe case T
Fixing
nut 3. Select [Maintenance], [ANL Maintenance].
4. Select one of the following menu items
Case
head according to the type of syringe to be replaced.
To replace the sample syringe, select “A/Replace
S Probe & Syringe.” To replace the reagent
Syringe syringe, select “B/Replace R Probe & Syringe.”
case
5. Press the STAT ROTATION/DIAG button.
Liquid is drained from the tube of the selected
syringe.
Syringe
Caution
Piston Do not place any body parts in the path of moving
fixing equipment, unless you are certain that the analyzer is in
screw
Stop or Standby and that diagnostic functions are not
activated.
Reagent Sample ISE Reagent
6. Loosen the bottom fixing screw first, then
As Needed the top fixing nut. Pull the syringe and case
Use the following procedure to replace the sample forward.
syringe and reagent syringes.
Caution
Prepare the following: • Do not apply excessive force to the fixing screws
when removing the syringe case. This can cause
• New sample syringe ZM0111 damage to the syringe case. Do not use pliers
when tightening the screws.
• New reagent syringe ZM0112
• If the relay tubes are disconnected from the
• (Same for ISE Reagent [Buffer] syringe.) solenoid valve during syringe case removal, screw
the relay tubes onto the connectors.
Remove and Replace Syringes
7. Tilt the syringe head and case upside down
1. Open the right front cover.
before replacing the syringe. This will prevent
2. Verify that the analyzer is turned on and is in air from entering the tubing lines and keeps the
the Warm-up or Standby Mode. water that collects in the head of the syringe
from leaking.
Case head
O-ring
Syringe
OLP2061E
FR
ON
T
Syringe case
Fixing nut
Case head
FR
ON
T
Tubing
connectors
Syringe case
8. To remove the syringe case, turn it Piston fixing screw
counterclockwise while holding the case head. Note: Top screws must be tightened first to ensure
9. Pull the syringe out from the case head. If proper syringe alignment on the analyzer.
the O-ring remains in the case head, carefully 13. Tighten the piston fixing screw (bottom screw)
remove it with tweezers to prevent scratching finger tight only. Do not use pliers.
the case head.
Check the system
Warning 1. Perform the following as required:
Verify that the correct size syringe is being replaced
(sample or reagent). The S and R syringe have • • If a sample syringe is replaced, select “H/S
different piston shaft diameters. Syringe Prime" on the [ANL Maintenance]
Piston shaft diameter: D screen. This takes approximately 25 minutes
to complete. Selecting F2 Exit stops the
D=5mm: R Syringe
D=2mm: S Syringe prime and exits the menu.
• • If a reagent syringe is replaced, select
Do not pull the piston out of a new syringe. This could
cause the syringe to be inaccurate even if it is replaced. “F/Prime Washing-line" on the [ANL
Maintenance] screen. Repeat this step
five or six times until the air in the tube is
10. Insert the new syringe into the case head completely removed. If the ISE Reagent
without dumping the existing water from the (Buffer) Syringe was replaced, select [System
head of the syringe. Water may escape on the Status], [ISE Status], F6 Prime, D/Buffer
side of the syringe head. Dry the excess water. Prime.
Completely dry the syringe and syringe case to
2. Press the Stat Rotation/Diag button. For Sample
prevent condensation from forming on the case.
or Reagent Syringes, DI water flows through the
11. Tilt the syringe head upward without twisting probe, removing air from the probe. On the ISE
the tubing lines. Screw the syringe case Buffer syringe, buffer flows through the tube and
clockwise onto the case head. “J” nozzle, removing the air.
12. Re-install the syringe and secure the top fixing 3. Close the right front cover.
nut first, and then the bottom fixing screw.
Photometer lamps are under warranty for 1,000 hours. If the photometer lamp deteriorates,
appropriate analysis results will not be obtained. Replace the photometer lamp every six months,
or when the photometer check is out-of-range, or when alarms occur. After replacing the
photometer lamp, perform a photocal to obtain accurate data.
Knobs
Lamp holder
T
ON
FR Lamp cords
Warning
• To prevent electric hazards, be sure to perform an
End Process before replacing the photometer lamp.
• Wait 5 minutes or more after the End Process
is completed. Do not touch the lamp with bare
hands until the photometer lamp has cooled down
completely. The lamp is very hot and can cause
burns.
• Never touch the glass of the photometer lamp with
bare hands or gloves! Oil from skin or fingerprints
changes the light intensity of the lamp and
decreases the measuring accuracy or cause
premature failures.
• If the photometer lamp is stained, perform an End
Process and wait at least 5 minutes. Check that the
photometer lamp has cooled down completely, then
wipe off the stain with a soft cloth dampened with
alcohol.
Instruction
As Needed When removing the lamp unit cover, be careful not to
bump the cover against the sample probe.
Prepare the following:
• New photometer lamp MU9888 3. Remove the lamp unit cover.
1. Check that an End Process was performed and 4. Loosen the two knobs on the terminals, then
allow at least 5 minutes for the lamp to cool disconnect the two lamp cords from the
after system shut down. terminals.
2. Open the upper cover. 5. Loosen the lamp holder by turning it
counterclockwise, then pull the photometer lamp
from the receptacle.
Protrusion
Lamp unit cover
Notch
Knobs
Lamp holder
T
ON
Lamp receptacle FR Lamp cords
Guide key
Caution
Collar notch
To obtain optimal analysis data, do not start a photocal
until the photometer lamp has stabilized after starting
Lamp holder the system. The photometer lamp will stabilize
approximately 20 minutes after system start-up.
6. Remove the lamp from the lamp holder and
collar. Insert the new lamp into the collar and 2. The mode changes from Warm Up to Standby
lamp holder. after 20 minutes.
7. Insert the lamp into the lamp receptacle. Be 3. Optional: Perform a photometer check to make
sure to match the notch on the guide key and the sure the lamp was installed properly.
collar notch of a new photometer lamp unit with 4. Perform the Weekly Photocal procedure.
the protrusion on the lamp receptacle.
Caution
Caution After replacing the photometer lamp, be sure to perform
Tighten the lamp holder securely. If the lamp holder is the photocal measurement to obtain accurate data.
loose, accurate analysis data cannot be obtained.
See Also
For information about how to perform a photocal, refer
to the "Weekly Maintenance" section in this chapter.
manual.
Mixing bar
Hook
Mixing unit
Cuvette wheel cover
Wash nozzle
station
Knob
As Needed (Part of Six Month PM) screws into the two specified holes shown in the
Prepare the following: drawings on the following page.
• This procedure requires using Neutrad. The 7. Use caution when removing the wheel.
wash solution can be ordered by calling Bumping other devices could cause damage.
1‑800‑766‑7000. The catalog number is Fisher Hold the two screws you placed on the wheel
#0435510. The wash solution can be ordered in and lift.
1 liter or 1 gallon bottles. 8. Turn the wheel upside down and use the reverse
• Cotton tipped applicators end of a cotton tipped applicator to gently push
the cuvette out from the bottom of the cuvette
Removing the cuvette wheel & cuvettes wheel.
1. Put the analyzer in Standby. 9. Remove all 88 cuvettes from the wheel.
Replace any cuvettes that have obvious
2. Open the rear cover of the analyzer.
scratches.
3. Loosen the knob on the wash nozzle unit.
Remove the wash nozzle unit and hang it on the Clean and Replace the Cuvettes
hook. 1. Sonicate the cuvettes in 2% Neutrad solution for
4. Turn the mix units clockwise approximately 60 15 minutes, or soak the cuvette in 5% Neutrad
degrees by hand to move the mix bars away solution overnight.
from the cuvette wheel. 2. Thoroughly rinse the cuvettes in deionized
water, or sonicate them in deionized water for
Instruction 10 minutes to remove any residual detergent.
When removing the cuvette wheel cover, be careful not 3. Allow the cuvettes to dry completely. Use one
to scratch or bump the sample probe, reagent probes,
of the following methods:
and mix bars.
• Allow cuvettes to dry naturally.
5. Gently lift the cuvette wheel cover and remove • Use an oven with the heat set under 50
it. degrees Celsius. (A hair dryer works well.)
6. Remove the two screws on the inner • Use a Kimwipe.
circumference of the cuvette wheel. Tighten the
4. Rinse the cuvette wheel with DI water and dry
thoroughly with a clean cloth.
Automated Chemistry Analyzer
AU400/AU400e User's Guide Chapter F - Maintenance F-49
August 29, 2008
OLP2042E
Cuvette wheel
Cuvette
Photometric
Yellow positioning mark face
Cuvette wheel
T
ON
FR Green positioning mark
Frosted glass
face
5. Insert the cuvettes back into the wheel. Ensure 9. Turn the mix bar unit back to its original
the cuvette is gently pushed down completely position.
into the wheel, or problems with analysis could 10. Remove the wash nozzle unit from the hook
occur. and screw it back into position.
Green positioning
mark
T Positioning pin on the
ON
FR cuvette wheel cover
Flat screws
Flat screw
Flat screw
Phillips head
screw
Note:
Do not touch the photometric side of the cuvette.
Handle by the frosted edge only.
• Use an oven with the heat set under 50 Alignment Positioning pin
indicators (one of two pins)
degrees Celsius. (A hair dryer works well.)
• Use a Kimwipe. Cuvette wheel
installed
4. Rinse the cuvette wheel with DI water only. Do
not use any cleaners or detergents as this may
ruin the finish on the metal. Dry the wheel
thoroughly with a lint free cloth.
Note:
In the event of a cuvette wheel overflow, dry the
incubation bath with a lint free cloth. Do not use any
cleaners or detergents to clean the incubation bath.
6. Replace the cuvette wheel cover by matching the 2. Select "F/Prime Washing-line."
green dot on the cover with the green dot on the 3. Press the STAT/Rotation Diag button. The wash
analyzer. nozzle tubing is primed with DI water.
7. Remove the wash nozzle station from the hook 4. Repeat the previous step several times until air
and place it on the base. Support the wash is removed.
nozzle station when replacing it on the base so
that it does not hit the cuvettes. Align the holes 5. Close the upper cover.
on the wash nozzle unit over the two positioning 6. Perform the photocal procedure prior to running
screws on the base and tighten the knob. Verify the analyzer. All cuvettes must pass the photocal
that all tubing and tube mounting joints are before analysis can take place.
connected to the nozzles and joint manifolds.
Refer to the “Replace the Wash Nozzle Joint See Also
Tubes” in the As Needed Maintenance section For information about how to perform a photocal, refer
for correct tubing placement in case any lines to the "Weekly Maintenance" section in this chapter.
are disconnected.
8. Replace any mix bars that were removed. Insert
the mix bars into the top of the mix unit. While
inserting the mix bar, rotate it slightly to engage
the flat portion of the mix bar with the gear of
the unit. Do not force the mix bar into the unit!
Damage to the gears may occur.
As Needed
This procedure takes 9 minutes. View the upper left corner of the screen for procedure progress.
The system counts down the remaining time.
Perform this procedure when either of the following is true:
a. Stop is selected in the middle of a run, and analysis is not started again immediately.
b. Power was lost to the system during analysis, and not recovered in a short period of time.
1. From Standby, select the [System Status] icon.
2. Select function key F5 (W1 Start). The system counts down the remaining time.
Rack Collection
Area
Rack Feeder
Belt
Tube
Tube mounting
joint
Tube mounting
joint manifold
Knob
Wash nozzle
Wash nozzle FR
ON
station T
Instruction
• When handling the wash nozzles, be careful not to
scratch them.
• When loosening the knob on the wash nozzle unit
station, do not loosen the positioning screws on
both sides of the knob. These screws are used for
positioning the wash nozzle unit.
As Needed Procedure
8. Loosen the knob on the wash nozzle unit, then
Prepare the following: remove the wash nozzle unit along with the
• Dry, clean cloth (KIMWIPE) tubing.
• Sonicator filled with DI water Clean and replace the Wash nozzle unit
Remove the Wash nozzle unit Caution
1. Open the upper cover. After placing the wash nozzle unit in the sonicator,
2. Verify that the analyzer is in the Warm-up or verify that the tips of the wash nozzle do not rest on the
bottom of the sonicator. When the sonicator is running
Standby Mode. the wash nozzle to vibrate and can damage the tips.
3. Select [Maintenance], [ANL Maintenance].
Automated Chemistry Analyzer
AU400/AU400e User's Guide Chapter F - Maintenance F-53
January 28, 2005
Instruction Caution
• When cleaning the wash nozzle unit using a If O-rings are used without cleaning for an extended
sonicator, be careful not to scratch the wash period of time, or if the cover on the jount unit is closed
nozzles. without any O-ring set in the groove, detergent crystals
• Do not scratch or tear the joints and tubes. A leak will collect and could scratch the O-ring. Check the
may result from the scratched part and the cuvette O-rings during Monthly mainteance. For more reliable
wheel may overflow. maintenance, replace the O-rings once a year.
Tips
The sonicator is the optimal device for cleaning the Instruction
wash nozzles. If a sonicator is not available, use DI
water. While pouring water into the wash nozzles, clean Be sure to attach the tube mounting joint manifolds
each nozzle hole using the supplied stylet. Rinse the in their original places. If a manifold is attached
nozzles in DI water, then dry them with a soft cloth. incorrectly, normal analysis cannot be performed.
9. Clean the wash nozzle unit along with the tubing 13. Attach the manifolds, tighten them firmly.
for 15 minutes, using DI water in a sonicator. When attaching the manifolds, match the color
No wash solution is required. connections and tighten them firmly. For
information about the tube mounting joint
10. Take out the wash nozzle unit from the positions, refer to the drawing on the following
sonicator, then wipe up the water drops using a page.
soft cloth.
14. Match the two positioning holes on the wash
11. Using a gauze dampened with water, wipe any nozzle unit with the positioning screws, then fix
foreign matter from O-rings such as dust or the station by tightening the knob.
detergent residue. If foreign matter remains,
remove the O-ring with tweezers, wash with 15. Close the rear cover.
water, and re-insert.
Check the system
12. If any O-ring is scratched or damaged, replace
1. Select “F/Prime Washing-line” on the
with O-ring MU9638.
[Maintenance], [ANL Maintenance] screen.
2. Press the Stat Rotation/DIAG button. The air
in the tubing of the wash nozzle unit will be
released.
3. Close the upper cover.
OLP2120.eps
Type 1 Type 2
A - Aspirate
B - Overflow
C - Dispense
Joint
Connecting hose
Button Fixture
Top
Filter case
Button
T Joint
ON
FR Basin
As Needed Procedure 6. Unscrew the case and remove the sample probe
Prepare the following: filter.
• Dry, clean cloth
• Basin
• Sonicator filled with DI water Button
• New water filter ZM3079 (if needed)
Instruction Cap
Perform an End Process before performing this procedure.
If this is performed with the sub-power to the system
turned on, water could spill from tubing and tanks. Filter
Positioning
1. Confirm that an End Process was performed. Ring
O-ring
Filter case
T
ON OLP2024E
FR
Joint
Connecting hose
Button Fixture
Filter case
Button
T Joint
ON
FR Basin
2. If necessary, replace the filter with part number 2. After initialization, select [Maintenance], [ANL
ZM3079. Maintenance].
9. Refer to the illustrations and verify that the top 6. Close the left front cover and the upper cover.
of the filter case is pointed upward and connect
the two hoses to the filter case fittings. Push
each hose connector into the filter case until a
click is heard.
10. Push the Sample Probe Filter Case into the
metal clip.
Automated Chemistry Analyzer
AU400/AU400e User's Guide Chapter F - Maintenance F-57
January 28, 2005
6.12 Clean the Deionized-Water Tank
Regular cleaning prevents the accumulation of deposits or bacteria that could affect analysis.
This procedure is also included in the monthly procedure, Clean the Wash nozzle unit,
ÊÊOLP2033E
Deionized-Water Tank and Filter, and Sample Probe Filter.
Tips
If low-quality deionized water is obtained from the main water source, the two diluted wash solution tanks may need
to be cleaned in addition to the deionized-water tank. For detailed information, contact Beckman Coulter Technical
Services.
Fixture
Joint
T
ON
FR
Basin
• Dry, clean cloth 7. Pull the DI water tank forward while removing
the tubes inserted in the tank. Lay the tubes in
• Basin the basin
• Freshly prepared 20% bleach 8. Remove the deionized-water tank from the
• Extra DI tank (In start-up kit) instrument.
9. Loosen the water tank cap and remove the float
Instruction switch.
Perform an End Process before performing this 10. Rinse the inside of the tank and tubes with a
procedure. If this procedure is performed with the fresh 20% bleach solution and rinse thoroughly
sub-power to the system turned on, the float switch in
the deionized-water tank will activate and excess water with deionized water.
will drain from the tubing. 11. Let the tank air dry.
Cap
Float switch
T
ON
FR
Deionized-water tank
Tubes
2. Insert the float switch and the water supply Check the system
tubes into the deionized-water tank. 1. Press the Green power button to apply power to
3. Replace the DI-water Tank in the system. Verify to the system.
that the tubing is fully inserted in the DI-water 2. After initialization, select [Maintenance], [ANL
Tank. Maintenance].
4. Reconnect the joints on the deionized-water tank 3. Select "F/Prime Washing-line."
and Sample Probe Filter Case. Push each joint
until a click is heard. 4. Press the STAT ROTATION/DIAG button.
5. Tighten the cap. • Observe the DI-Water and filter tubing. The
deionized water will flow through the tube
6. Reconnect connector #240. When reconnecting and then the air in the DI-Wiater tubing
the #240 connector, do not force the connection will be released. This prime may need to be
or the pins can be damaged. Turn the connector performed 2-3 times to ensure that all of the
gently until it slides into the connection easily air is out of the tubing.
and without force.
5. Repeat the previous step two or three times
7. Push the Sample Probe Filter Case into the until the air in the affected tubing is completely
metal clip. removed.
8. Remove the basin. 6. Close the left front cover and the upper cover.
Filter case
Basin
Note
Part numbers MU8424 and MU8423 include one static discharge brush, order one of each if
replacing both brushes. Analyzers should have two static discharge brushes installed on each side
of the rack transport. If the analyzer is equipped with only one static discharge brush contact
your Field Service Engineer.
Caution
To avoid infection always wear latex gloves to clean or replace the static discharge bushes. If your hands
come in contact with the static discharge brushes, wash the affected area thoroughly.
1. Select the Stop key or icon to place the analyzer in STOP mode.
2. Remove the dark acrylic cover from the rack transport area over the cup detector and
barcode reader area.
3. Loosen the static discharge brush screws by turning them gently counter clockwise until
they stop turning. The screws will not come off the analyzer.
4. Lift the brush assembly out of the instrument by tilting the bottom of the static discharge
brush away from the static brush housing, then, lift it up and out clearing the mounting
hole for the flat mounting pin located on the side of the static discharge screw.
5. Follow the same procedure with the brush assembly on the other side of the rack
transport.
6. Clean any stain on the brushes with an alcohol prep from the base to the end of the bristle
tips.
7. If the static discharge brushes are still stained after cleaning or show signs of wear,
replace them.
8. Dispose of the old brushes in a receptacle for biohazard waste.
9. Reinstall the clean or replaced brushes, making sure the flat mounting pins on the static
discharge housing fit into the mounting holes on the brush assembly.
10. Tighten the brush screws, on each assembly.
11. Replace the dark acrylic cover from the rack transport above the cup detector and barcode
reader area.
12. Select the Stop key or icon to place the analyzer in warm up or standby mode.
Contents
OLP2069E
Connector #142
Connector #143
Mount plate
FR
ON
T
Case head
Syringe case
FR
ON
T
Daily Maintenance 5. Visually check for leaks inside the syringe case.
Prepare the following: Replace any damaged component.
• ISE Cleaning Solution (AUH1019) 7. Select [ISE Status] on the screen. The ISE
Status screen will appear.
• One ISE cup (MU9232)
8. Press function key F5 (ISE unit start) on the
1. Check that the system and the ISE unit are on [ISE Status screen]. The window for selecting
and the system is in Warm-up or Standby. the kind of ISE single operations will appear.
U-H
1 2
U-L
3
5
20
S-L
K-
SE
6
L-N
• (The set contains bottles for Na and K.)
a
19
7
S-H
18
8
17
9
CLE W1
16
AN
W2 10
15
11
14
N T 13 12
S
RO
TA
T
E
N
S
D
TA
F
S
E
T
Note
IS
E
P
R
IM
STAT ROTATION/DIAG
E
switch
Before performing this procedure a valid ISE STAT Table Top View
calibration is necessary. 11 12
10 13
9 14
15
AN
8
E
16
7
2. Pour Na and K electrode check solution into
17
6
different sample containers.
S-H
18
5
a
-N
EL
19
K-S
L
S-
4
U-L
U-H
20
3
Note:
The AU400 ISE unit automatically primes the mid-standard reagent every hour of inactivity to keep the
electrodes conditioned. Normally this procedure takes 20 minutes and requires removing the sample pot
and tubing. To prevent a spill if the unit is apart for more than one hour, turn off the ISE switch.
Connector #142
Connector #143
Mount plate
FR
ON
T
J-Nozzles
Mix Bar
9. Carefully wipe the Mix Bar and Liquid Level Sensors with an alcohol prep using a
downward stroke.
10. Set the Mix Unit aside.
Sample pot
Connector
FR
ON
T
12. Hold the Sample Pot and remove the Sample Pot tubing from the inlet of the flowcell.
Remove the Pinch Valve tubing from the Pinch Valve.
Disconnect the Pinch Valve tubing #5 at the Y-connector near the Mixture Aspiration Pump.
The Mixture Aspiration Pump is labeled 3 and 4.
ISE Diagram Pinch valve Pinch valve tubing
Sample Pot
Bypass tubing Y-connector
Mixture Aspiration
Pump
Tubing
T-connector
Flow Cell inlet Flow Cell outlet
Note:
Perform this procedure over a sink. Use caution when flushing the tubing. Position the tubing
downward to avoid splashing the eyes or face.
a. b.
14. Place the Sample Pot and tubing in a beaker containing 1% Wash Solution.
15. Place the beaker in the sonicator (filled with water) and sonicate for 10 minutes.
Note:
Perform this procedure over a sink. Use caution when flushing the tubing. Position the tubing
downward to avoid splashing the eyes or face.
J-nozzles
1. Select “E/MID/REF Prime.” Press the Stat/Rotation Diag button. The unit begins priming
for eight cycles.
2. Observe the Mix Unit J-nozzle. Liquid should flow in a straight line from the J-nozzle into
the center of the Sample Pot.
3. Observe the tubing at the outlet of the flowcell (#6). Verify that all bubbles are removed. If
bubbles are present, check all tubing placement and connections and repeat steps 1 and 2.
Do not calibrate the ISE until all air is removed from the flowcell.
4. Select “D/Buffer Prime”. Press the Stat/Rotation Diag button. The unit begins priming for
eight cycles.
5. Verify that the Sample Pot fills and drains with buffer. Liquid should flow in a straight line
from the J-nozzle into the center of the Sample Pot.
6. Close the ISE cover.
7. Use fresh standards to perform an ISE calibration and verify that a valid slope is obtained.
See Chapter C Basic Operations.
T
ON
FR Rolling pump
Maintenance Every Three Months 5. The MID solution and mixture roller tubes
Prepare the following: are stretched over the rolling pump cylinders.
S
Stretch and lift tubing up and away from the
TA
T
E
N
S
D
• Roller tube MU9623
TA
T
rolling pump. Allow it to rest on top of the
S
E
T
IS
E
cylinders.
P
R
1. Select [System Status], [ISE Status], F6 Prime.
IM
E
6. Unscrew each roller tube connector to remove
NOTE the roller tube.
Remain in the "F6 Prime" menu. 7. Connect a new roller tube. Secure the
connectors at both ends of the tube by screwing
2. Select "A/Replace Electrode." them hand tight.
3. Press the STAT ROTATION/DIAG button.
Liquid will be drained from the tubing. Instruction
3. Open the upper cover. Verify the tubing numbers match the labeled pump
connectors. If connected incorrectly, liquid will not be
4. Open the ISE cover. supplied or drained correctly.
Instruction
The roller tubing consists of the MID solution roller
tube and the mixture roller tube. Replace these two
roller tubes at the same time.
S
TA
T
E
N
S
D
TA
T
S
E
T
Nos.
IS
E
P
R
IM
E
T
ON Nos.
FR
S
TA
T
E
N
S
D
TA
T
9. Press the STAT ROTATION/DIAG button. The
S
E
T
IS
E
P
two rolling pumps are turned on to move the
R
IM
E
solution in each roller tube. Check that the
liquid is completely aspirated from the sample
pot to the electrode unit. If liquid remains in
the sample pot, check the pinch valve tubing
connection. Refer to the diagram located on the
ISE unit door.
10. Select Close.
11. Close the ISE unit cover.
12. Close the upper cover.
Pinch valve
4
T
ON
FR
Mixture Aspiration
Pump
Tubing
#3 & #4
#6
Flow Cell inlet Flow Cell outlet
Pinch valve
T
ON
FR
While pressing the green button on top of the and insert the top tube (#5 in the drawing) by
pinch valve, pull the lower tube from the slot pushing it all the way into the slot.
(#6 in the drawing), then release the button.
10. Press the STAT ROTATION/DIAG button.
S
TA
Instruction
T
The two rolling pumps are turned to move the
E
N
S
D
TA
T
S
E
solution in each roller tube. Check that the
T
Do not mount each branch tube in a different groove on
IS
E
liquid is completely aspirated from the sample
P
the pinch valve. If it is mounted backwards, the liquid
R
IM
E
will not be supplied and drained correctly. pot to the electrode unit. If liquid remains in
the sample pot, check the pinch valve tubing
8. Mount a new section of pinch valve tubing connection. Refer to the diagram located on the
by connecting the branches at positions 4, 5, ISE unit door.
and 6, as described above. Connect the longer
11. Select Close.
Y-branched tube at position 5.
12. Close the ISE unit cover.
Note: There is a diagram inside the ISE Unit
cover indicating the correct positions in the 13. Close the upper cover.
pinch valve for tubing 5 and 6.
9. Replace the lower tube (#6 in the drawing) by
pressing the green button on top of the pinch
valve. Make sure the tube is inserted all the
way into the slot. Release the green button
Lock lever
(Electrodes are locked in this state.)
Thermistor connector
Thermistor
T
ON
FR
S
TA
drained.
IS
As Needed Maintenance
Prepare the following: Tips
If the STAT ROTATION/DIAG button is pressed again,
• REF electrode MU9197
MID-solution will be supplied to the tube.
• REF electrode packing MU9202
• Internal Reference Solution AUH1017 9. Disconnect all electrode cords. Reference is
green, Na, K, and Cl are color-coded.
Instruction 10. Move the lock lever to unlock the electrodes.
The reference electrode may break if excessive force is 11. Remove the Na, K, and Cl electrodes.
applied when replacing the electrode.
12. Gently lift the block on which the REF electrode
is mounted.
1. Check that the system and the ISE unit are on
and the system is in Warm-up or Standby. Instruction
2. Open the upper cover. When loosening the set screw, do not push or twist the
thermistor cord, otherwise it may break.
3. Open the ISE unit cover.
4. Press the [System Status] key. The System 13. Loosen the set screw, then gently remove the
Status screen appears. REF electrode along with the set screw. Gently
remove the packing with a pair of hemostats.
5. Select [ISE Status] on the screen. The ISE
Status screen appears.
6. Press function key F6 (prime) on the [ISE
Status] screen. The window for selecting prime
operations appears.
Automated Chemistry Analyzer
AU400/AU400e User's Guide Chapter F - Maintenance F-81
January 28, 2005
Connector
REF electrode
insertion hole
Packing
Set screw
REF electrode Block
T
ON
FR
REF electrode
14. Check that no air bubbles are introduced to the electrodes are properly placed, the lock lever
electrode tip on a new REF electrode. is secured, and the O-rings are in the proper
position.
If air bubbles reside in the electrode tip of
the new REF electrode, remove the bubbles 25. Close the ISE unit cover.
by orienting the electrode bar downward and 26. Close the upper cover.
tapping it with a finger.
S
27. Wait approximately 5 minutes before performing
TA
T
15. Replace the packing in the block.
E
N
S
D
calibration analysis.
TA
T
S
E
T
16. Mount the new REF electrode along with the set
IS
E
P
screw on the block. See Also
R
IM
E
For information about how to perform calibration, refer
17. Tighten the set screw to secure the REF to the Basic Operations Chapter.
electrode.
18. Replace Na, K, and Cl electrodes based on the Caution
color-coded diagram. To obtain the best possible analysis data, perform two
successive calibration measurements to confirm the
19. Remount the block. electrode stability. If the difference between the first
and second calibrations is within the following values,
20.Move the lock lever to lock the electrodes.
the electrodes are stable:
21. Connect all electrode cords. Reference is green; Na 0.020, K 0.045, Cl 0.025
Na, K, Cl are color-coded. If each difference is not within the above values, the
electrode membrane may not be stabilized. Open the
22. Press the STAT ROTATION/DIAG button.
upper cover and the ISE unit cover. Press function key
MID-standard is supplied to the electrodes. F6 (prime) on the [ISE Status] screen. Select mid-prime
23. Select "E/MID/REF Prime" on the [ISE Status] then repeat steps 16 to 20.
screen.
24. Press the STAT ROTATION/DIAG button again.
The two rolling pumps move the solution in
each roller tube. Press the STAT ROTATION/
DIAG button. The rolling pumps activate eight
times. Visually check tube #6 to verify that no
bubbles pass through the electrodes. Repeat this
operation until there are no air bubbles in the
tube. If air bubbles are present, verify that the
OLG4085E.AI
REF electrode
Reference line
On board stability
Check your package insert for on board stability claims.
Aspiration tube
MID solution Cap
tank Buffer solution
ISE
tank ON
F
OF
FR
ON
T
REF solution tank
Caution
• Never replace reagents during analysis.
• The density of the REF solution is higher than
those of others. If another solution is mixed with
the REF solution, analysis data will be incorrect.
Never add new reagent to the remaining reagent.
K electrode
Na electrode
Cl electrode
Cord
S
TA
T
E
N
S
D
TA
T
S
E
T
IS
E
P
R
IM
E
8. Press the STAT ROTATION/DIAG button. The
liquid that remains in the region of electrodes is
drained.
Tips
Select "A/Replace Electrode" and press the STAT
As Needed Maintenance ROTATION/DIAG button once. The liquid in the tube
is drained. If the STAT ROTATION/DIAG button is
Prepare the following: pressed again, Mid-standard solution is supplied to the
tube.
• New Na electrode MU9194
• New K electrode MU9195 9. Disconnect the cords from the Cl, Na, or K
• New Cl electrode MU9196 electrode.
1. Check that the system and the ISE are on and 10. Move the lock lever to unlock the electrodes.
that the system is in Warm-up or Standby. 11. Remove the three electrodes.
2. Open the upper cover.
3. Open the ISE unit cover.
Instruction
A total of four O-rings are attached to one side of three
4. Press the [System Status] key. The System electrodes and to the metal part above the Cl electrode.
Status screen appears. Exercise care so the O-rings are not lost when removing
the electrodes.
5. Select [ISE Status] on the screen, then press
enter. The [ISE Status] screen appears.
6. Press function key F6 (prime) on the [ISE
Status] screen. The window for selecting prime
operations appears.
7. Select "A/Replace Electrode."
K electrode
Na electrode
Cl electrode
Tube
Blue cord
Yellow cord
Red cord
S
corresponding electrode.
E
N
To obtain the best possible analysis data, perform two
D
TA
T
15. Press the STAT ROTATION/DIAG button. The
S
E
successive calibration measurements to confirm the
T
MID solution is supplied to the tube.
IS
electrode stability. If the difference between the first
E
P
R
IM
and second calibrations is within the following values,
E
16. Select "E/MID/REF Prime" on the [ISE Status] the electrodes are stable:
screen. Na 0.020, K 0.045, Cl 0.025
17. Press the STAT ROTATION/DIAG button. The If each difference is not within the above values, the
rolling pumps activate eight times. Visually electrode membrane may not be stabilized. Open the
check tube #6 to verify that no bubbles pass upper cover and the ISE unit cover. Press function key
F6 (prime) on the [ISE Status] screen. Select mid-prime
through the electrodes. Repeat this operation then repeat steps 16 to 20.
until there are no air bubbles in the tube. If air
bubbles are present, verify that the electrodes
are properly placed, the lock lever is secured,
and the O-rings are in the proper position.
18. Close the ISE unit cover.
19. Close the upper cover.
20. Wait at least 5 minutes before performing
calibration analysis.
• 10% bleach Solution (10mL Clorox bleach + 9. Pipette 10mL of ISE Mid-Standard Solution
90mL DI H20) into the sample pot and manually turn the roller
pump to clear the Mid- Standard Solution.
1. Select [System Status], [ISE Status]. Repeat 3-5 times.
2. Select F6/Prime. 10. Replace the mix assembly and the pinch valve
3. Select A)Replace Electrode then press the STAT tubing.
Rotation/DIAG button. The flow cell will drain. 11. Press the STAT Rotation/Diag button on the
4. From the current menu, remove the mix analyzer.
assembly from the sample pot. 12. Perform three to four mid-standard primes.
5. From the current menu, remove both pinch 13. Perform a Total prime then exit the current
valve tubing’s (#5 and #6) from the pinch valve. menu.
6. For the first 2 minutes, pipette the 10% bleach 14. Calibrate the ISE.
solution into the ISE sample pot and manually
turn the left-hand roller pump assembly 15. Run QC Material.
clockwise until most of the bleach empties from
the sample pot into the ISE tubing. Continue
filling the sample pot with the bleach solution
while turning the left roller pump assembly.
Introduction
If an error flag is detected during sample analysis, it is logged in the analysis data. Error flags
are printed to the right of the result. Flags are also displayed on the data display screen. On the
DPR display, data with a flag appears in a different color (red) than data without flags (black).
The following table gives a cause and corrective action for each error flag.
Contents
Introduction...............................................................................................................G-1
1. Checking Results...............................................................................................G-4
2. Data Error Flags.................................................................................................G-5
FLAG: i (Manual dilution not calculated for ISE tests)................................................ G-5
FLAG: d (Flags QC data that was excluded from calculation)..................................... G-5
FLAG: e (Edited Data).................................................................................................. G-5
FLAG: ( (Shortage of detergent for contamination parameters in user defined detergent
bottles R1 & R2)............................................................................................. G-5
FLAG: R (Reagent is empty)......................................................................................... G-5
FLAG: # (Sample Level Detection Error)...................................................................... G-5
FLAG: % (Clot detected)............................................................................................... G-6
FLAG: ? (Unable to perform calculations).................................................................... G-6
FLAG: U (Reagent blank absorbance at last photometric point is low)........................ G-7
FLAG: u (Reagent blank/routine absorbance at first photometric point is low)............ G-7
FLAG: Y (Reagent blank absorbance at last photometric point is high)...................... G-8
FLAG: y (Reagent blank/routine absorbance at first photometric point is high)........... G-8
FLAG: @ (Test results are too high).............................................................................. G-9
FLAG: $ (Not enough data to determine linearity of reaction)...................................... G-9
FLAG: D (Absorbance of reaction is greater than maximum OD range).................... G-10
FLAG: B (Absorbance of reaction is less than minimum OD range).......................... G-10
FLAG: * (Linearity error in rate methods)....................................................................G-11
FLAG: & (Prozone test data is abnormal)....................................................................G-11
FLAG: Z (Prozone error)..............................................................................................G-11
FLAG: ! (Unable to calculate concentration)............................................................. G-12
FLAG: ) (Cannot convert from OD to CONC)............................................................ G-12
1. Look at the printout and check it for error flags. If error flags appear on the printout,
perform the corrective action corresponding to that flag. For information on correcting
flags, see the following pages.
2. Perform corrective actions for any alarms generated during the run. For information on
correcting alarms, use the alarm on-line help function.
See Also
For information about checking results, refer to the Basic Operations
OLP4077E
Chapter, "Checking Results.”
Index date/time
System date/time
22:37
Report operator name Report operator name Report operator name
Print date
Print time
S.NO. U0005
S.ID
1995-11-21
10:34
sample type urine
Print date
Print time
S.NO. U0006
S.ID
1995-11-21
10:34
sample type urine
Print date
Print time
S.NO. P0001
S.ID
1995-11-21
10:34
Index time 22:37 Index time 22:37 Index time
Operator name Operator name Operator name
Sex Age Month Sex Age Month Sex
Patient information-3
Patient information-4
Patient information-3
Patient information-4
Patient information-3
Patient information-4
OLYMPUS BECKMAN BECKMAN
Report operator name Report operator name Report operator name
Patient information-5 Patient information-5 Patient information-5
S.NO. U0005 sample type urine S.NO. U0006 sample type urine S.NO. P0001
S.ID S.ID S.ID
Sex Age Month Sex Age Month Sex
Error Flag
Patient information-3 Patient information-3 Patient information-3
The system allows two error flags to appear on the data printout per analyte. If more than two events
have occurred, the flags will be listed in priority order. The priority order of all flags are shown on the
table of contents in this chapter.
Important: After lamp installation, allow the lamp to stabilize, then a photocal must be
performed. Repeat the Photometer Check with the new lamp installed to verify integrity.
Recalibrate all tests before starting specimen analysis.
4. Verify that the correct settings were programmed in [Parameters], [Specific Test
Parameters].
5. Check the reaction data including those processed immediately before and after the flagged
data result. If any abnormality is found, check the cuvettes and cuvette wash station for an
overflow, then recheck the results processed before and after the flagged data results. If the
issue persists contact Beckman Coulter Technical services.
6. Check syringes.
7. Check probes.
8. Verify calibrator material.
9. Verify reagent integrity and position.
See Also
For detailed procedures on checking syringes, probes and calibrator material, refer to the section in this
chapter called "Troubleshooting for Data Flags ?, @, $, D, F, G, !."
FLAG: $ (Not enough data to determine linearity of reaction)
Cause: Less than three read points of a rate reaction fall within the acceptable OD range
specified. In order to properly calculate a rate reaction, at least three readings must
be taken prior to reaching the maximum or minimum OD limits. If the OD limits are
exceeded, the reaction may have gone into substrate depletion due to either a high
result, or a problem with the integrity of the reagent. The nonlinearity calculations are
not made.
Action:
1. The sample may be severely lipemic, icteric, hemolytic or may contain excessively large
amounts of the analyte being tested. Dilute the sample and run it again.
2. Check syringes.
3. Check probes.
4. Verify calibrator material.
5. Verify reagent integrity and position.
6. Check the reaction data including those processed immediately before and after the flagged
data result. If any abnormality is found, check the cuvettes and cuvette wash station for an
overflow, then recheck the results processed before and after the flagged data results. If the
issue persists contact Beckman Coulter Technical services..
See Also
For detailed procedures on checking syringes, probes and calibrator material, refer to the section in this
chapter called "Troubleshooting for Data Flags ?, @, $, D, F, G, !."
See Also
For detailed procedures on performing a calibration, refer to the Basic Operations Chapter.
2. The operator can choose to accept the results and remove the flag by selecting [Routine],
[Data Management], [Data Edit]. Select function key F5 (Data Display), then F5 (Edit).
Check the Flag box and delete the "b" flag.
See Also
For detailed procedures on checking syringes, probes and calibrator material, refer to the section in this
chapter called "Troubleshooting for Data Flags ?, @, $, D, F, G, !," and Chapter H Troubleshooting.
FLAG: G (Results lower than the dynamic range)
Cause: The concentration of the sample is below the Dynamic Range Low (linearity of the
reagent) limit.
Action:
1. Verify that the correct Dynamic Range limits are programmed in the [Parameters], [Specific
Test Parameters] screen, then repeat analysis.
2. Check syringes.
3. Check probes.
4. Verify calibrator material.
5. Verify reagent integrity and position. Refer to "Troubleshooting for Data Flags" in this
chapter for procedures.
6. Follow laboratory protocol for samples with values less than the Dynamic Range low limit.
See Also
• See Chapter D, "Repeat Specific," in volume 1 of the User's Guide.
• For detailed procedures on checking syringes, probes and calibrator material, refer to the section
in this chapter called "Troubleshooting for Data Flags ?, @, $, D, F, G, !," and Chapter H
Troubleshooting.
Note
The AU400/AU400e will generate an alarm in addition to the "p" flag for an ISE test and other tests if
the panic value range is exceeded.
Action:
1. Verify the correct panic range is programmed for your lab.
2. Follow your laboratory protocol for panic results.
FLAG: T (Abnormality found in inter-chemistry check)
Cause: This flag will be attached to the data which exceeds the High/Low limits specified for
the Check Parameters. The check parameters are designed to test the relationship
between different analyte results on the same sample. i.e. Direct Bilirubin should be
less than Total Bilirubin.
Action:
1. Review data. Verify that the correct inter-item range for your lab is programmed.
2. Follow your laboratory protocol for abnormal test results.
FLAG: P (Positive: Any value above flag level H)
Cause: The value is above flag level H in [Specific Test Parameters].
Note: This flag is not a panic value. This flag is used mainly in drug screenings to define the
positive limit of a result.
Action:
1. None.
FLAG: N (Negative: Any value below flag level L)
Cause: The value is below flag level L in [Specific Test Parameters].
Action:
1. None.
FLAG: H (Result higher than normal value range)
Cause: The data has exceeded the High limit on the normal value range. The normal range is
set in [Parameters], [Specific Parameters], "Normal Ranges".
Action:
1. Review data.
?
STANDBY 1997/09/03 10:10
(R) Routine (P) Parameters (A) Auxiliary (M)Maintenance (U) User (L) Logout
(S)Start Condition
(T)Test Requisition Errors found as below.Start Measure?
(R)Reaction Monitor Yes No
(C)Calibration Monitor
(P)Photocal Monitor
(Q)QC Monitor Serum Urine Others
(D)Data Management
Start Routine S.No.:
(L)Data Report
Operator Name:
ALARMCLEAR
See Also
For start-up procedures, refer to the Basic Operations Chapter.
Cal. expired:
Calibration frequency has expired. Check the calibration stability date in [System Status], [Reagent
Status]. Requisition for the expired tests in the [Routine], [Test Requisition], [Calibration] screen.
Run the reagent blank and calibrate the expired tests.
Cuvette temperature:
Temperature has exceeded standards. Call Beckman Coulter Technical Services.
Error cuvette:
A cuvette may be dirty or scratched. Check the error in [System Status], [Cuvette. Status]. Refer
to the Maintenance Chapter for procedures on cleaning and/or replacing cuvettes. Perform a
photocal. Also refer to the Maintenance Chapter for procedures on performing a photocal.
ISE no cal.:
1. The ISE was not calibrated for serum or urine. OR 2. An End Process was performed and
"ISE no cal" occurs again when the instrument is on. You can start analysis and use the last
calibration slopes obtained.
ISE stop:
The ISE unit has stopped. Reset the ISE unit in [System Status], [ISE Status].
On board expired:
Reagent on-board stability has expired. Select [System Status], [Reagent Status] and check the
on-board stability, if it has expired, replace the reagent. Perform a reagent volume check to
update the information.
Unset reagent:
A reagent assigned to the round in use (on-board) is missing from the refrigerator. Determine
what reagent is missing. Use [Reagent Status], function key F7 (Test Display) and compare with
the reagents in the refrigerator. Perform a reagent volume check.
?
STANDBY 1997/09/03 10:10
(R) Routine (P) Parameters (A) Auxiliary (M)Maintenance (U) User (L) Logout
(S)Start Condition
(T)Test Requisition Errors found as below.Start Measure?
(R)Reaction Monitor Yes No
(C)Calibration Monitor
(P)Photocal Monitor
(Q)QC Monitor Serum Urine Others
(D)Data Management
Start Routine S.No.:
(L)Data Report
Operator Name:
ALARMCLEAR
See Also
For start-up procedures, refer to the Basic Operations Chapter.
DI water shortage:
The deionized water tank is low.
Incorrect Parameters:
A message appears on screen indicating the incorrect parameter. Open the screen containing
the incorrect parameter, determine the parameter error and correct it. The parameters are saved
automatically.
ISE Busy:
The ISE unit is operating. After the ISE operation is complete start analysis.
Unchecked Reagent:
A reagent check was not performed or the [Parameter] menu was accessed. This will change
the reagent status from "checked" to "unchecked." Select the [System Status], [Reagent Status]
screen and perform a reset.
Washer A shortage:
Replenish fluid in concentrated detergent tank A.
Washer B shortage:
Replenish fluid in concentrated detergent tank B.
Other Errors:
A specific error name and description displays. Clear the error and restart analysis.
Introduction
This Chapter is designed for the operator to obtain quick solutions to common problems
that could occur with the AU400. The following pages provide information on general
troubleshooting procedures for the analyzer and the ISE unit. Keep a history of problems and
corrective actions that occur with the instrument. This helps correct a problem quickly, should it
become necessary to contact Beckman Coulter Technical Services.
Chapter H Troubleshooting
Introduction...............................................................................................................H-1
To Begin Troubleshooting, Answer the Following Questions............................................................H-3
1. Troubleshooting Using On-Line Help...............................................................H-5
2. Troubleshooting the Analyzer...........................................................................H-7
2.1 Data Problems..........................................................................................................................H-8
2.1.1 Checking Abnormal Data in the Software Screens.......................................................H-9
2.1.2 Troubleshooting Software........................................................................................... H-11
2.1.3 Troubleshooting Reagents and Samples....................................................................H-13
2.1.4 Troubleshooting Mechanical Problems.......................................................................H-16
2.2 System Problems....................................................................................................................H-21
2.3 Data Processor Problems.......................................................................................................H-25
2.4 Recovering from an Emergency Stop or Power Loss.............................................................H-28
2.5 Recovering from a Cuvette Wheel Overflow..........................................................................H-29
2.5.1 What Causes a Overflow?..........................................................................................H-29
2.5.2 Recognizing a Overflow..............................................................................................H-29
2.5.3 Recovering from a Overflow......................................................................................H-30
2.5.4 After the Overflow Problem Is Fixed...........................................................................H-30
See Also
For information about how to interpret the data printout, refer to the Basic Operations Chapter.
See Also
For information about data flags, refer to the Error Flags Chapter.
See Also
For information about how to perform a calibration, refer to the Basic Operations Chapter.
Is QC out-of-range?
See Also
For information about how to perform a QC, refer to the Basic Operations Chapter.
See Also
For information about schedule maintenance, refer to the Maintenance Chapter.
For Alarms
An alarm usually indicates a mechanical problem or a communication failure. Follow the
corrective action provided in the on-line help function.
See Also
For information about how to use on-line help, refer to "Troubleshooting Using On-line Help" located in
this chapter.
See Also
For information on start-up procedures, refer to the Basic Operations Chapter.
Is the reagent stable?
See Also
For information on checking reagent stability, refer to the "Troubleshooting for Data Flags" procedure in
the Error Flags Chapter.
Is there enough reagent in the bottles?
Select the [System Status] icon. Select [Reagent Status]. Press function key F5 (check start).
See Also
For information on reagents, refer to "Changing Reagents” in the Basic Operations Chapter.
Has the reagent, calibrator, and QC material been properly stored?
See Also
For information on proper storage, refer to the package insert. The "Troubleshooting for Data
Flags" procedure in the Error Flags Chapter also provides general information on handling these
materials.
Has the analyzer been calibrated correctly?
See Also
For information about performing calibrations, refer to "Performing Calibrations” in the Basic
Operations Chapter.
Has scheduled maintenance been performed?
See Also
For information about schedule maintenance, refer to the Maintenance Chapter.
See Also
For information about scheduled maintenance, refer to the Maintenance Chapter.
?
STANDBY 1997/09/03 10:10
(R) Routine (P) Parameters (A) Auxiliary (M) Maintenance (U) User (L) Logout
Help
ALARMCLEAR
Alarm Help
Displays information about system alarms. Press the alarm icon (as shown in the illustration
above) while the alarm message is displayed. If using the mouse, move the pointer onto the
alarm help icon, then press the left mouse button.
1. The following information will be displayed on alarm help:
• Alarm Number – An alarm number is associated with each alarm.
• Alarm Name – Name of alarm that occurred on the system.
• System Status – Describes the current state of the system or how the system
responds to the alarm.
• User Action – Provides steps to correct the problem and resume normal operations.
• Cause – Explains why the error occurred.
• Additional Information – Provides more corrective actions for advanced operators
or engineers. Do not perform these functions unless you are qualified!
2. Follow the corrective actions provided on the screen. If the alarm is still generated after
performing all recommended procedures, contact Beckman Coulter Technical Services.
3. To close the help window, select the Exit Help bookmark in the bookmark column or use
the mouse to move the pointer onto the “X” enclosed in a box located at the top of the
screen, then press the left mouse button.
Menu Help
Describes the purpose of the specific software screen and how to use it. Select a screen
that you would like to know more about, press function key F1. If using the mouse, move
the pointer onto the menu help icon that indicates function key F1 then click the left mouse
button.
?
STANDBY 1997/09/03 10:10
Report Format
Common Sample Results(Fix) Results(Seq.) Comment Line
Operation Help
Analyzer Alarm
ALARMCLEAR
Contents
See Also
For information about how to interpret the data printout, refer to "Checking Results" located in the Basic
Operations Chapter.
See Also
For information about data flags, refer to the Error Flags Chapter.
See Also
For information about how to perform a calibration, refer to the Basic Operations Chapter.
Is QC out-of-range?
See Also
For information about how to perform a QC, refer to the Basic Operations Chapter.
See Also
For information about schedule maintenance, refer to the Maintenance Chapter.
See Also
For information about the [QC Monitor] screen, refer to the Software Chapter.
See Also
For information about error flags, refer to the Error Flags Chapter and the "Quality Control" section in
the Specifications Chapter.
See Also
For more information about the [Reaction Monitor] screen, refer to the Software Chapter.
See Also
For more information about the [Calibration Monitor] screen, refer to the Software Chapter.
See Also
For information about the [Photocal Monitor] screen, refer to the Software Chapter. For information
about how to perform a photocal, refer to the Maintenance Chapter.
See Also
For information about the [Data Statistics] screen, refer to the Software Chapter.
See Also
For information about the [Histogram] screen, refer to the Software Chapter.
See Also
For information about the [Calibration Monitor] screen, refer to the Software Chapter.
If abnormal data is found in some tests:
Identify the same tests that include abnormal data between calibrators. If all abnormalities
are derived from the same calibrator, the calibrator may be the cause of the abnormal data.
Identify if the reagent position for the tests in question are related.
See Also
For detailed procedures on verifying calibrator material, refer to the Error Flags Chapter.
If abnormal data is found in all tests:
There is a high possibility that the calibration analysis itself may result in abnormal data.
Check the mix bars, R1 probe/syringe, sample probe/syringe, water, calibration material and
common hardware.
See Also
For detailed procedures on checking the syringes, checking the probes and verifying calibrator material,
refer to the Maintenance Chapter and Flags Chapter.
See Also
For information about how to check reaction progress, refer to the [Reaction Monitor] screen in the
Software Chapter.
See Also
For information about how to check reaction progress, refer to the [Photocal Monitor] screen in the
Software Chapter.
See Also
For information about photometer lamp replacement, refer to the Maintenance Chapter.
Instruction
See the package insert provided by the reagent manufacturer for a list of acceptable anticoagulants.
See Also
For information about sampling, refer to "Sampling Specifications" in the Specifications Chapter.
• To prevent sample evaporation, do not leave samples unsealed for an extended
period of time. If samples evaporate, correct analysis cannot be obtained.
• The serum is hemolyzed, lipemic, or icteric (LIH). Check the serum for the
extent of LIH. See "Specific Test Parameters" in the Software Chapter for more
information.
• Liquid-level sensor malfunction due to bubbles on the serum or urine surface.
Remove the bubbles from the surface and repeat analysis.
• The sample cups and racks were not placed properly on the system. Place the
sample cups and racks properly on the system.
See Also
For more information on sample requirements refer to the following: "NCCLS document
(GP16-A), Urinalysis and Collection, Transportation, and Preservation of Urine Specimens:
Approved Guideline," and "College of American Pathologists document, Patient Preparation &
Specimen Handling, Chemistry/Clinical Microscopy (Fascicle VI)."
See Also
For information about setting or replacing reagent bottles, refer to the Basic Operations Chapter.
• The liquid-level sensor malfunctioned during reagent aspiration due to bubbles in
the reagent bottle. Remove the bubbles in the reagent bottle.
See Also
For detailed information about maintenance, refer to the Maintenance Chapter.
• Insecticide was used in the vicinity of the system.
Insecticides can affect the cholinesterase (CHE) levels. If contamination is suspected,
replace the sample cups, reagents, and reagent bottles with new ones. Also wash the
sample probes, reagent probes, mix bars and cuvettes. Never use insecticide in the
vicinity of the system.
See Also
For detailed information about maintenance, refer to the Maintenance Chapter.
• Water purity, electrical specifications, environmental conditions Check major system
specifications, located in the Specifications Chapter.
Syringe(s) Problems
• Water may leak from the syringes. Hand tighten the syringe cases and case heads of the
sample and reagent syringes.
• General Syringe Troubleshooting.
• Verify the top and bottom screws are hand tight.
• Verify that probes are not clogged.
• Verify the bottom screw is tight up against the piston.
• Verify there is a smooth, resistant pull.
• Verify the correct size syringe is in use (reagent or sample).
• Verify there is one O-ring being used, and that it is not damaged.
• Verify the syringe is on the analyzer correctly.
• Check the syringe tubing for crimps or leaks.
• Check the teflon tip of the syringe for wear.
• Bubbles were generated in the syringe tubing
Select [ANL Maintenance] from the [Maintenance] menu. After selecting “F/Prime
Washing-line” press the STAT ROTATION/DIAG Switch. Air will be removed from the
tubing.
• The syringe tubing is clogged.
Remove the relay tubes, then clean each tube inside using a stylet. If the syringe tubing
remains clogged after cleaning, contact Beckman Coulter Technical Services.
See Also
For information about syringe cleaning and replacement, refer to the Maintenance Chapter.
Probe Problems
• Leaks from the sample probe and reagent probe due to loose probe connectors. Remove the
covers on the sample probe and reagent probe, then tighten the probe connectors.
• Leaks from the dispense tubing Remove the covers on the sample probe and reagent probe,
then tighten the probe connectors. If this does not work, contact Beckman Coulter Technical
Services.
• A clogged sample or reagent probe. Drain the DI water from the sample probe or reagent
probe and check the way the DI water drains.
• The sample probe tip or reagent probe tip was bent or deformed. Replace the damaged
sample probe or reagent probe.
• The sample aspiration position of the sample probe is incorrect. The sample probe tip
is not positioned at the center of the sample cup. Visually inspect the sample probe for
abnormalities. If the sample probe is bent, replace it. If the sample aspiration position is
abnormal and the sample probe is not bent, contact Beckman Coulter Technical Services.
See Also
For information about how to replace the sample probe or reagent probe, refer to the Maintenance
Chapter.
See Also
For information about cleaning the mix bars, refer to the Maintenance Chapter.
• The coating on mix bar was removed.
Replace the mix bar.
See Also
For information about how to replace mix bars, refer to the Maintenance Chapter.
• While the mix bars are rotating, abnormal sounds such as rubbing, gear contact noise, or
other abnormal operating sounds were generated.
Contact Beckman Coulter Technical Services.
• The mix bar mounting position was abnormal. The mix bars came into contact with the mix
bar wash well and/or cuvettes.
Contact Beckman Coulter Technical Services.
• The wash water and detergent are not properly drained from the mix bar wash well.
Contact Beckman Coulter Technical Services.
• Mix bars were not properly installed on the unit and the sample and reagents were not mixed
properly.
Install the mix bars again.
Scratches, fingerprints, or foreign matter is noticed on the cuvettes, or the cuvettes were
stained.
See Also
For information about washing or replacing cuvettes, refer to the Maintenance Chapter.
Moisture is detected outside the cuvette(s) and the cuvette wheel.
See Also
For information about cuvette wheel floods, refer to "Recovering from a Cuvette Wheel Flood" in this
chapter.
The wash water and/or detergent spills from the wash nozzles because proper draining
did not occur, or the nozzles did not aspirate correctly.
Are the tube joints on the wash nozzles loose? Tighten the loose tube joints.
The wash nozzles may be clogged. Clean the wash nozzles.
See Also
For information about how to clean the wash nozzles, refer to the Maintenance Chapter.
After cleaning cuvettes, a large amount of water remains in the cuvettes.
Are the tube joints on the wash nozzles loose. Tighten the loose tube joints.
The wash nozzles may be clogged. Clean the wash nozzles.
• The tube in the concentrated detergent tank floats.
Straighten the tube, and insert it toward the bottom of the tank so that it does not come
into contact with the opening.
• The float switch in the concentrated detergent tank or diluted detergent tank malfunctioned.
Check the float switch connector. Do not bring the tube in the tank into contact with
the float switch. Check that the float switch is not directly subjected to detergent from
the tube. If the float switch malfunction is not corrected by the above action, the switch
must be replaced. Contact Beckman Coulter Technical Services.
See Also
For information about cleaning cuvettes, refer to the Maintenance Chapter.
For information about cuvettes replacement, refer to the Maintenance Chapter.
For information about cuvette wheel floods, refer to the Troubleshooting Chapter.
See Also
For information about replacing the photometer lamp, refer to the Maintenance Chapter. Always perform
a photocal after changing the photometer lamp.
See Also
For information about cleaning the deionized water tank, refer to the Maintenance Chapter.
• After cleaning the deionized water tank, detergent remains.
• Clean the deionized water tank again. Rinse thoroughly with deionized water.
• Verify laboratory deionized water meets specifications.
See Also
For information about water specifications, refer to the Specifications Chapter.
See Also
For information about how to clean or replace the deionized water filter or sample probe filter, refer to
the Maintenance Chapter.
Rack Problems
• General Rack Troubleshooting.
• Check barcode placement.
• Verify sample position according to rack color.
• Verify the correct number of magnets are present in the bottom of the rack.
• Verify the rack was loaded correctly.
See Also
For information about racks, refer to the Basic Operations Chapter.
See Also
For information about replacing the deionized-water filter, refer to the Maintenance Chapter.
• Other causes
• The circulation pump, radiator fan, lamp cooling fan, air pump, 24V power supply
fan, or drying pump may be defective.
• Contact Beckman Coulter Technical Services.
Rolling tubes
Detergent rolling
pumps
Connectors
Relay tubes
NT
FRO
See Also
For information about rolling tube replacement, refer to the Maintenance Chapter.
Barcode Errors
• The barcode labels on the sample cups or reagent bottles may be stained.
• Check for water drops or stains on the barcode labels. If water drops are found,
wipe them off. If a sample ID label is stained or damaged, replace it with a new
one. If a reagent ID label is damaged, the reagent position can be fixed so the
barcode will not be used.
• Barcode labels on sample cups or reagent bottles may have fallen off or were not placed
properly. Refer to the Basic Operations Chapter for proper placement procedures.
• The barcode reader may be stained.
• Wipe the window using an alcohol prep.
See Also
For information about placing barcodes, refer to the Basic Operations Chapter.
See Also
For information about how to clean the wells, refer to the Maintenance Chapter.
• The waste line may be obstructed or it was installed improperly.
• Does the waste line meet the system specifications?
• Can you see any obstruction in the lines? If so, contact Beckman Coulter Technical
Services.
See Also
For information about how to clean the deionized water filter, refer to the Maintenance Chapter.
See Also
For information about how to replace the ribbon, refer to the printer manual supplied with the printer.
See Also
For information about how to replace the reagent probe, refer to Maintenance Chapter.
See Also
For information about error flags, refer to the Error Flags Chapter. For information about sample
volume, refer to "Sampling Specifications" in the Specifications Chapter.
See Also
For information about cleaning nozzles, refer to the Maintenance Chapter.
See Also
For information about executing [Retrieve Data Base], refer to the Software Chapter.
See Also
For information about the floppy disk format, refer to the Software Chapter.
• The floppy disk was write-protected when the system attempted to write data to the disk.
Slide the tab on the diskette to “write-enable.”
• The floppy disk is damaged.
• Replace the floppy disk with a new one.
• The floppy disk drive may be damaged.
• If the problem is not corrected after taking the above actions, contact Beckman
Coulter Technical Services.
See Also
For information about real-time output setup, refer to the Software Chapter.
• The paper is empty.
• Load more paper. Select [Routine], [Report], then try printing again.
See Also
For information about loading paper, refer to the printer manual supplied with the printer.
See Also
For detailed information about the [Report] screen, refer to the Software Chapter.
On-line Auto-output by Host Computer not Executed
• The I/F cable to the host computer is disconnected.
• Connect the I/F cable correctly.
• The host I/O parameters were incorrectly modified.
• Set the appropriate I/O parameters by selecting the [Online] screen.
See Also
For information about the [Online] screen, refer to [Routine], [Data Report], [Online] located in
the Software Chapter.
No Data Stored Even Though There Is Sufficient Space on the Hard Disk
• The database in the hard disk has been destroyed.
• Select [Maintenance], [Data Operation], [Retrieve Data Base]. Restore the analysis
data that has been backed up on the floppy disk into the hard disk. If the system
will not start up normally after executing [Retrieve Data Base] (the hard disk may
be damaged), contact Beckman Coulter Technical Services.
Note
The analyzer will stay in a Warm Up mode for 1 1/2 hours after an EM Stop is performed. To bypass
this mode, select [Auxiliary], [Standby Set].
Caution
If the nature of the emergency compromises operator or instrument safety, press the EM Stop Switch
immediately. Please be aware that pressing the EM Stop Switch without performing the other steps
listed below could cause damage to the computer hard drive. Also the operating system and application
files could become corrupted.
The following steps are the preferred method of shutting down during an emergency:
1. Press the following keys on the keyboard: CTRL + ALT + Delete. The Task Manager
appears.
2. From the Task Manager screen, select the Shutdown button. Windows NT will close all
open files and databases.
3. When the message "It is now safe to turn off your computer" appears, press the EM Stop
Switch on the front of the analyzer. All power to the system turns off.
Important
In the event of a stop or an emergency stop, it is not possible to use the data. Analysis must be repeated.
If analysis was in progress and a stop or an emergency stop was performed, reagent still remains in the
cuvettes. This can cause damage or deterioration in the cuvettes, or may cause abnormal results. If
reagent was in the cuvettes for a lengthy period of time, perform a W1 prior to restarting analysis.
Instruction
It is very important that the probes are centered over the cuvettes.
See Also
For information about replacing cuvettes, refer to the Maintenance Chapter.
• Verify the wash nozzle tubing connections are secure.
• Verify the O-rings on the manifolds are in place and not damaged.
Please refer to “Cleaning Cuvettes and the Cuvette Wheel after a Cuvette Wheel Overflow”
located in the As Needed Section of this maintenance chapter.
Contents
See Also
For information about checking results, refer to “Checking Results” located in the Basic
Operations Chapter.
Is the Data Flagged?
See Also
For information about data flags, refer to the Troubleshooting Chapter.
Is the Calibration Out-of-Range?
See Also
For information about how to perform a calibration, refer to the Basic Operations Chapter.
Is QC Out-of-Range?
See Also
For information about how to perform a QC, refer to the Basic Operations Chapter.
Is Data Erratic? (Have you performed scheduled maintenance?)
See Also
For information about schedule maintenance, refer to the Maintenance Chapter.
For Alarms
An alarm usually indicates a mechanical problem or a communication failure. Follow the
corrective action provided in the on-line help function.
See Also
For information about using on-line help, refer to the Troubleshooting Chapter.
See Also
For information on start-up procedures, refer to the Basic Operations Chapter.
See Also
For information on checking reagent stability, refer to the "Troubleshooting for Data Flags"
procedure in the Error Flags Chapter.
See Also
For information about performing calibrations, refer to the Basic Operations Chapter.
See Also
For information about scheduled maintenance, refer to the Maintenance Chapter.
Instruction
See the package insert provided by the reagent manufacturer for a list of acceptable anticoagulants.
See Also
For more information on sample requirements refer to the following: "NCCLS document
(GP16-A), Urinalysis and Collection, Transportation, and Preservation of Urine Specimens:
Approved Guideline," and "College of American Pathologists document, Patient Preparation &
Specimen Handling, Chemistry/Clinical Microscopy (Fascicle VI)."
Fixing nut
Fixing screw
FR
ON
T
Case head
Syringe case
Piston fixing screw
T
ON
FR Rolling pump
S
TA
T
E
N
S
D
TA
T
S
E
T
IS
E
P
R
IM
E
OLP2070E
Sample pot
Connector
FR
ON
T
Problem: Worn or bent mixture aspiration tubing. This condition could also generate
a sample pot overflow error.
Action: Observe sample pot, flowcell tubing, and bypass tubing during aspiration of
calibrator or control. There must not be any air in the flowcell during the measurement
period. Air will be present in the bypass tubing. If leaks occur or the tubing is worn
or flat in appearance, replace the mixture pump tubing.
See Also
For procedures on replacing mixture pump tubing, refer to the Maintenance Chapter.
Connector
#142
Connector
#143
FR
ON
T
Sample pot
Connector
FR
ON
T
T connector
Pinch valve
4
T
ON
FR
E
S
TA
T
IS
E
P
R
IM
K electrode
Na electrode
Cl electrode
Cord
S
TA
T
E
N
S
D
TA
T
S
E
T
IS
E
P
R
IM
E
Problem: Old or obstructed Na, K, Cl, or reference electrodes
Action:
1. Remove obstructions in the flowcell path. Perform the daily cleaning procedure two or
three times. Prime with mid-standard.
2. Perform a calibration.
See Also
For information about performing calibrations, refer to the Basic Operations Chapter.
3. Perform a “Selectivity Check” to verify Na and K membrane selectivity.
See Also
OLP2077
For information about performing a selectivity check, refer to the Troubleshooting Chapter.
4. Try changing electrodes.
See Also
For information about replacing electrodes, refer to the "ISE Maintenance" section of the Maintenance
Chapter.
Connector
REF electrode
insertion hole
Packing
Set screw
REF electrode Block
T
ON
FR
REF electrode
See Also
For information about adding reference electrode solution, refer to the Maintenance Chapter.
Instruction
Electrode replacement should be performed if the slope values are out-of-range, or if other
troubleshooting measures don’t resolve the problem. If possible, remove the suspected electrode and
place it into another AU400 and perform a calibration to verify the performance of the electrode.
T
ON
FR Rolling pump
S
TA
T
E
N
S
D
TA
T
S
E
T
IS
E
P
R
IM
Problem: Old, worn, or crimped tubing or leaking
E
Instruction
Old or worn tubing will be flat (no longer round in diameter) in the area where it is pressed
against the pump.
Action:
1. The mixture pump pulls the samples, calibrators, and mid-standard through the flowcell.
Replace the mixture pump tubing.
See Also
For information about replacing mixture pump tubing, refer to the "ISE Maintenance" section of the
Maintenance Chapter.
3.3.4 Thermistor
Lock lever
(Electrodes are locked in this state.)
Thermistor connector
Thermistor
T
ON
FR
If the following problems occur with the thermistor, perform the action listed below:
Problem:
• Defective packing
S
TA
T
E
N
S
D
TA
• No contact with liquid in the flowcell
S
E
T
IS
E
P
R
IM
• Other thermistor malfunction
E
Action:
1. Visually check the thermistor for liquid contact. Verify the thermistor was not forced in,
and that the screw is tightened. No bubbles should be observed around the thermistor.
2. Call Beckman Coulter Technical Services if a malfunction of the thermistor is suspected.
K electrode
Na electrode
Cl electrode
Cord
S
TA
T
E
N
S
D
TA
T
S
E
T
IS
E
P
R
IM
E
If the following problems occur with the flowcell block, perform the action sequence stated
below:
Problem:
• Obstruction in the flowcell inlet block
• Obstruction in the flowcell outlet block
• Defective port of flowcell inlet block
•
Action:
1. Check the expiration date of the electrodes. The expiration date is located on the box. If
the electrodes have expired, do not use them.
2. Replace defective flowcell inlet or outlet blocks.
3. Perform an ISE cleaning procedure. Call Beckman Coulter Technical Services if the
obstruction is not removed.
Instruction
To determine electrode integrity and on-board life, review the calibration, selectivity check, and QC
results. The following calibration ranges are acceptable:
Na 38 - 65
K 38 - 65
Cl -38 - -65
Mid-standard solution factor ranges:
Na 0.80-1.20
K 0.75-1.25
Cl 0.75-1.25
Replace electrodes when results are out-of-range and other troubleshooting procedures have been
performed. Electrodes are under warranty for 20,000 samples or 6 months.
See Also
For information about performing calibrations, refer to the Basic Operations Chapter. To perform a
calibration from diagnostics, refer to the Software Chapter.
1. Verify calibration results.
2. Review the data daily by using the following criteria:
A. The slopes must be in range:
Na 38 - 65
K 38 - 65
Cl -38 - -65
Mid-standard solution factor ranges:
Na 0.80-1.20
K 0.75-1.25
Cl 0.75-1.25
B. The day-to-day slope values should be consistent for each electrode. Slopes may
decrease with time. This however, does not necessarily indicate a faulty electrode.
C. Check for trends in slope values.
3. Verify the ISE Reagent Integrity.
Instruction
A calibration failure will occur if the buffer or standard solutions become contaminated with the
reference solution. The reference solution has a very high concentration of KCl and even a small
amount will affect ISE results.
See Also
For information about replacing reagents, refer to the Maintenance Chapter.
Instruction
A successful calibration must be achieved before performing a selectivity check.
See Also
For information about performing a selectivity check and replacing electrodes, refer to the "ISE
Maintenance" section in the Maintenance Chapter.
See Also
For information about replacing the electrodes, refer to the Maintenance Chapter.
If all three values are out-of-range in the same direction:
• See “3.2 Dispensing Systems”
• See “3.3 Measuring Components”
• Replace the sample probe
If all three values are randomly out-of-range:
• See “3.3 Measuring Components”
• Check all electrodes
• Check reference electrode packing and internal solution
See Also
For information about replacing the solution, refer to the Maintenance Chapter.
2. Repeat the sequential sample measure until SDs are in the acceptable range.
3. Out-of-Range QC results are flagged with H or L on the printout. Verify that the range
for the QC lot number is properly entered in QC parameters.
Instruction
Check the high and low range against the package insert values. Verify that the lot number in use
corresponds to these values. This range generates H and L flags on the printout.
8. If using an external calibrator (ACAL), verify that the correct calibrator concentration is
entered.
9. Verify that the correct Cal No. is being used.
10. Select [Parameters], [Specific Test Parameters]. Check the factor and offset values. The
factor should be 1.0 and the offset 0.0 unless results are being corrected.
Instruction
Prime the reference solution approximately five (5) times to completely fill the lines.
Introduction
This Chapter provides an outline of system operation during analysis and hardware configuration
descriptions of the analyzer. It shows the piping diagram and major specifications of this
system. Also, terminology specific to the AU400 is described here.
Contents
1. System Operation During Analysis................................................................... I-3
Operation Process..................................................................................................................... I-3
Operation Modes....................................................................................................................... I-4
2. Analyzer Hardware Configuration..................................................................... I-5
Operation Switches................................................................................................................... I-5
Rack Feeder Unit....................................................................................................................... I-7
Sample Probe and Reagent Probe Units.................................................................................. I-9
The Mix Unit............................................................................................................................. I-10
The Wash Nozzle Unit..............................................................................................................I-11
The STAT Table....................................................................................................................... I-12
Sample Syringe and Reagent Syringes................................................................................... I-14
Rolling Pump Unit.................................................................................................................... I-15
Reagent Refrigeration Unit...................................................................................................... I-16
The Incubator.......................................................................................................................... I-18
The Photometer Unit............................................................................................................... I-19
Tank Storage........................................................................................................................... I-20
Breakers and Fuse.................................................................................................................. I-21
The ISE Unit ........................................................................................................................... I-22
The ISE Reagent Syringe ....................................................................................................... I-23
The ISE Reagent Bottles......................................................................................................... I-24
3. System Connections......................................................................................... I-25
Cuvette wheel
Mixing
Washing and drying
Reagent transfer
Grating
Photometer sensor
Light
source Auto repeat run
Sample transfer
ISE unit
Sample transfer
Mixing bar
Color CRT Standard
Report CPU
monitor printer
Sample pot
Optional
Report printer
Electrode
Hard
Record and data storage disk
Keyboard
Power ON Initialize
Maintenance Warm-up
W1 W2 Photocal measurement
Power OFF
Measure 1 Measure 1 :
Sample analysis in progess.
Pause
Measure 2 :
No more racks detected on the feeder belt.
Measure 3 :
Measure 2 All R1 dispensing is completed.
Pause Measure 4 :
Final Curvette wash.
Measure 5 :
(Not displayed) When a nonfatal mechanical error
occurs during analysis with one of more samples
Measure 3 having the R2 mixed prior to the nonfatal
Pause mechanical error Processing will finish for the
samples that have R2 mixed.
Measure 4
Pause
Measure 5 Stop
: Automatic transition
: User operated transition
Measure 1: Sample analysis is in progress. Racks are moving from the feeder belt to the
transfer belt.
Measure 5: (Not displayed) When a nonfatal mechanical error occurs during analysis with
one or more samples having R2 mixed prior to the nonfatal mechanical error.
One Touch Mode: Preprogrammed STAT analysis mode. Sample requisition is not required.
See Also
For more information on the One Touch Mode, refer to the Basic Operations Chapter.
Operation Switches
ST
AT
STAT ROTATION/DIAG
Button
ISE
ON
F
OF
ISE Power Switch
FR
ON
T
Sample/Rack
barcode reader
Reflecting mirror
Rack feeder
plates
Protective
cover
LED
Rack
T
ON
FR
Caution
Do not look directly at the laser beam emitted from the sample/rack barcode reader. Staring at the beam
may damage your eyes.
Reflecting Mirror
Reflects the sample barcode reader laser beam onto the sample tube.
Instruction
Do not touch the mirror with your fingers or any hard materials. If the mirror is smudged or scratched,
barcode reading may not be performed properly.
Probe arm
Sample probe cover
Reagent probe
Relay tube
Probe connector
Sample probe
Reagent probe
wash well
Sample probe
wash well
T
ON
FR
Probe arm
Sample probe
Dispenses a given volume of sample in a cuvette.
Reagent probe
Dispenses a given volume of reagent in a cuvette.
Probe connector
Connects the sample probe and reagent probe to the relay tube.
Probe arm
Supports the sample probe and reagent probe.
Sample probe wash well
Washes the sample probe.
Reagent probe wash well
Washes the reagent probe.
Mix unit
Mix bar
Mix bar
wash well
T
ON
FR
Mix Unit
This unit controls the mix bars. The sample and reagent inside a cuvette are mixed by
a rotating mix bar. The mix unit also rotates the bars to the wash well. The mix unit is
comprised of one mixer. There are six mix bars in the mixer.
Mix Bar
After the sample and reagent have been dispensed into a cuvette, the mix bar rotates and
mixes the mixture in the cuvette.
Mix Bar Wash Well
This is used for washing the mix bar after mixing.
Knob
Joint tube
Tube mounting joints
Wash nozzles
Wash nozzle
station FR
Aspiration nozzle ON
T
Drying nozzle
Wash nozzles
Wash Nozzles
Used for washing the cuvette after analysis is complete. Each nozzle is made up of three
small nozzles. The longest nozzle aspirates the reaction mixture, detergent, and wash water.
The next longest nozzle dispenses the detergent or wash water. The shortest nozzle aspirates
any detergent or wash water that exceeds a predetermined amount.
Aspiration nozzle
Aspirates the dispensed wash water.
Drying nozzle
Aspirates the remaining drops of water inside the cuvette.
Tube Mounting Joints
Connects each of the nozzles to the internal drain tubing.
Knob
Secures the wash nozzle station.
U-L
2
3
4
cover (large)
U-H
5
20
S-L
K-S
6
EL-
Na
19
7
S-H
18
8
17
W1
9
CLEA
16
N
W2 10
15
T
11
14
ON
13 12
S
FR
TA
T
E
N
S
D
TA
T
S
E
T
IS
E
P
R
IM
E
Top view of STAT table
3 4
2
5
1
6
U-L
22
7
U-
H
21
8
S-L
K-SEL-Na
20
9
19
S-H
W1
10
18
2
W
11
CLEAN
17
12
16
13
15 14
STAT table
Used to perform urgent analysis, automatic QC and automatic calibration. There are cup
positions around the inside and outside of the STAT table. The samples to be analyzed
are set in the cup positions around the outside circumference of the table. Detergent or
reagents used for ISE unit calibration or selectivity checks are set in the positions around
the inside circumference of the table. It is possible for the system to read only the barcodes
on the sample cups that are set in the outer positions on the table. Note: The STAT table is
refrigerated, after placing samples, be sure to replace the table covers. Removing the table
covers frequently may cause the temperature on the table to become unstable.
“WASH 1” Position
During normal analysis, the sample cup (containing 2% Washing Solution for washing the
sample probe) is placed in this position.
“WASH 2” Position
The bleach or HCl for washing the sample probe is poured in a sample cup and placed in this
position.
“CLEAN” Position
When automatically washing the ISE sample pot and electrode line, the sample cup containing
the bleach is placed in this position.
Instruction
Do not place barcoded sample cups in the inner positions on the STAT table. Barcode read errors may
occur with the sample cups placed in the inner positions.
Syringe case
FR
ON
T
Piston fixing screw
Sample Syringe
Used to supply a small fixed amount of sample.
Reagent Syringe
Used to supply a small fixed amount of reagent.
Fixing Nut
Attaches the syringe to the syringe unit.
Syringe Head
Secures the syringe together with the syringe case.
Syringe Case
This case holds the syringe.
Piston Fixing Screw
Attaches the syringe piston to the syringe drive assembly.
Rolling tubes
Rolling pumps
for detergents
Connectors
Relay tubes
ONT
FR
Rolling Pump
Used to aspirate concentrated detergent and dispense it into the diluted detergent tank. There
are two rolling pumps; one for detergent tank A (factory optional) and one for detergent tank
B.
Rolling Tube
The rubber tube on the rolling pump. Detergent is supplied or drained through the tubing.
There are two rolling tubes; one for detergent tank A (factory optional) and one for detergent
tank B.
Relay Tube
The relay tubes are used to connect the rolling tubes through the connectors as shown in the
illustration above.
Reagent
refrigerator cover
Positioning pin
Sample pre-dilution
Reagent refrigerator bottle position
T
ON
FR
Partition
Positioning pin
Partition
Partition
Caution
Place reagent bottles with the barcode label facing the outside of the reagent tray. If the reagent bottles
are set in any other way, the reagent probe may be damaged.
Cuvette wheel
Incubator
T
ON
FR
Cuvette
Incubator
The cuvette wheel is attached to the incubator. The mixture in each of the cuvettes in the
cuvette wheel is incubated in the incubator. During analysis, the temperature in the incubator
is kept at 37°C.
Cuvette Wheel
The cuvette wheel contains 88 cuvettes. As the cuvette wheel rotates, a series of analyses are
performed from dispensing of the sample to performing photometry.
Cuvettes
These are square shaped containers made of glass used in analysis. The sample and reagent
are dispensed into the cuvettes. The light path of a cuvette is 6 mm, and the capacity is 750
µl.
Lamp cover
Knobs
Photometer
lamp
T
ON
FR Lamp cords
Photometer Lamp
A white light source used in performing photometry on the mixture in the cuvettes.
Lamp Cover
The cover is for safety since the photometer lamp becomes very hot. Also, it prevents electric
shock due to touching the terminal board of the photometer lamp, and shields the inside of the
photometer unit.
Knobs
The knobs connect the lamp cords to the terminal board. A DC voltage of approximately +12
V is output.
Lamp Cords
The lamp cords supply power to the photometer lamp.
Concentrated Concentrated
detergent tank A detergent tank B
(factory optional)
Rear view
DPR (PC) breaker
Reagent refrigerator and
incubator breaker
Analyzer breaker
Fuse Heater breaker
FUSE
Pilot lamp 250V 2A
TYPE-F
DPR (PC)
service outlets OP.1 SIGNAL OP.2 SIGNAL
K electrode
Pinch valve Na electrode
Thermistor Cl electrode
Mixing unit
T
ON
FR MID solution dispense rolling pump
Mixture aspiration rolling pump Sample pot
S
TA
T
E
N
S
D
TA
T
S
E
Sample Pot
T
IS
E
P
R
IM
E
The sample and buffer solution are dispensed into the sample pot and mixed. The
amount of dispensed sample is 20 µl of serum (fixed), 25 µl of urine (fixed), and 10
µl of DI water (fixed). The amount of dispensed buffer solution is 618 µl for serum
(fixed) and 750 µl for urine (fixed).
Mixing Unit
This unit mixes sample and buffer solution dispensed into the sample pot. It is
equipped with two liquid-level sensors to detect clogged tubing.
Cl Electrode, Na Electrode and K Electrode
These electrodes are used for measuring the potentials of Cl, Na and K ions in the
mixture and MID solution. The concentrations of individual ions in the mixture can
be calculated from the potential differences between each ion in the mixture and in
the MID solution.
REF Electrode
This is the reference electrode with respect to the Cl, Na and K electrodes.
Thermistor
The potentials of the Cl, Na and K electrodes change depending on the temperature
of solutions even though the concentrations are identical. To remove the effects of
temperature, the output potentials are corrected by the thermistor.
Pinch Valve
This valve is used for alternately discharging the analyzed mixture and the mixture
remaining in the sample pot.
Rolling Pumps
There are two pumps; a rolling pump which aspirates the mixture solution, and a
rolling pump which dispenses the MID solution.
Rolling Tubes
These tubes are made of rubber and wrap around the rolling pump. As the rolling
pump rotates, the tubes are squeezed by the rollers on the pump, and solution is
supplied or removed.
Syringe case
Fixing Nut
Attaches the syringe to the analyzer.
Case Head
Attaches the syringe together with the syringe case.
Syringe Case
Holds the syringe.
Piston Fixing Screw
Attaches the syringe piston to the syringe drive assembly.
ISE
ON
F
OF
FR
ON
T
MOUS
POWER
PRINTER
MONITOR MONITOR
KYB
MONITOR
PERSONAL
COMPUTER
AU400 ANALYZER
MOUSE
KEY BOARD
See Also
For more information on installation environments, refer to the "Precautions on Use and Installation
Environments" Chapter. Note: Check the rating plate on the rear face of the unit for the unit type.
Analyzer: (Inches) Width 57" x Depth 30" x Height 48" Weight 924 lbs
Data processor: Width 27.6" x Depth 27.6" x Height 62" Weight 146 lbs The data processor
(DPR) includes the rack, personal computer, CRT monitor and printer.
Operating Environment
Temperature: 18 to 32 °C
Humidity: 40 to 80% RH (with no condensation)
Temperature fluctuation: ±2 °C or less during analysis
Conductivity of deionized water: 2.0 µs/cm or less (0.5 MΩ or greater)
Temperature of deionized water: 5 to 28 °C
Maximum altitude: 6,500 feet (2,000 M)
Pollution degree: 2
Installation category: II
Power Supply
Sampling Specifications
Sampling Mechanism, Micro-Syringe Type
Clot detection function added: If a clot is detected during sample aspiration and/or sample
dispense, an alarm occurs. The system cancels the sample aspiration and moves to next
sample.
Crash detection function added: If the probe tip bumps against a sample cup while it is
moving down, the system automatically stops the probe operation. The probe may not be able
to dispense accurately in the future if it was damaged. If the probe is damaged, refer to the
Maintenance Chapter for replacement procedures.
Sample Quantity Setting
2 to 50 µl/test: Can be set in increments of 0.5 µl.
Sample Dilution Quantity Setting
0 or 10 µl/test
Sample ID
Read from the barcode. One of the following barcode types can be selected: NW7, Code 39,
Code 128, 2 of 5 standard, 2 of 5 interleaved, and mixed. ISBT-CODE128 can be read only
if no other types are mixed.
Types of Sample Racks
Sample Rack Type The racks are visually classified according to color. The analyzer
recognizes the type of rack by the placement of magnets located on the bottom of the racks.
*1: The maximum diameter of a blood collection tube should be less than 17.5mm. An
adapter fitting to the outside diameter is required.
*2: Another adapter is required for tubes with a 12mm outside diameter.
*3: If the sample source is a blood collection tube, use only tubes with a total length
< 100mm.
340 to 800 nm 13 wavelengths: 340, 380, 410, 450, 480, 520, 540,
570, 600, 660, 700, 750 and 800
nm
Light Source
Resolution of Photometry
0.0001OD
Data Processor Unit Specifications
Configuration of the Data Processor Unit
Input/Output Specifications
Data Display
Display characters: Alphanumerics
Data Input/Output to External Devices
Data exchange by RS232C or floppy disk
OLP4052E
• [Maintenance] menu
(O)Offline Output
(M)PC Anywhere
(A)ANL Maintenance
(C)Consumable Management
(P)Periodic Maintenance
(A)ANL Diag
(I)ISE Diag
(L)Alarm Log
(F)File Management
R001: R is the prefix for reagent blank. R001 is the first blue rack through the analyzer. The
second blue rack through the analyzer prints results labeled 1R002, 2R002. The maximum
number of reagent blanks per index is 999.
1R001: The OD at the first read point of the test (as set in [Parameters], [Specific Test
Parameters]). Generates y (over) or u (under) flags based on an acceptable OD range entered
in the individual parameter screen. For example: Data number 1 OD for serum TP is -0.2330.
Note that a negative sign follows the result. Data number 1 OD for serum AST is under the
acceptable range and therefore has generated a "u" flag.
2R001: The OD at the last read point of the test (as set in [Parameters], [Specific Test
Parameters]). Generates Y (over) and U (under) flags based on the acceptable OD range entered
in the individual parameter screen. For example: Data number 2 OD for urine CREA is 0.0107.
Data number 2 OD for serum AST is under the acceptable range and therefore has generated a
"U" flag.
1-point Assay
This is a general end point assay that determines the reaction mixture OD from the OD measured
at a specified photometric position. The diagram below illustrates the case for one reagent only
with reagent volume equal to or greater than 150 µL. In this case the photometric position is 27,
positions 1-27 may be used. OLP4090E
OD
R1 0 S 1 2 3 4 5 6 7 8 27
P0 Px
R1 0 S 1 2 .... 9 10 R2 11 .... 27
R1.V S.V Pz R2.V Px
Instruction
Twenty-seven is the last read point. There are no OD values or variables after this point.
Reaction mixture OD = Color reaction channel OD (ODC) - Serum blank channel OD (ODB)
ODC
0 0
R1 S R2 27
Instruction
To link the color reaction test and serum blank test, go to [Parameters],
[H - Inter-Related Tests], and [B - Sample Blank]. Enter the test number for the color reaction test. See
Section 5.5 Sample Blank for more information.
Rate Assay
This assay determines the rate of absorbance variation per minute or change OD/min by
calculating the slope of the change by least squares in absorbance variations for all OD values
within the OD limits and included in the defined photometric points. Note at least three
consecutive OD values must be within the OD range limit. If no-lag-time is defined as yes, then
values before the defined first read point can be used for fast reactions.
Instruction 4093
The Y axis is the reaction OD and the X axis is the photometric read point or time in minutes.
OD
∆OD/min.
OD limit
R1 S R2 27
OD
DOD(1)
DOD(2)
R1 S Pz Pw R2 Py Px 27
OD
R1 S R2 A B 27
OD
DOD(1)
DOD(2)
R1 S Pz Pw R2 Py Px 27
Sample Blank
The sample blank method uses the specified Color Test and Blank Test as defined in [Parameters],
[Inter-related Tests], [Sample Blank].
OD calculated = color blank OD - [Reagent blank OD]
If a Color Test is selected, the Blank Test must be specified in the blank test field on the [Sample
Blank] screen. OLP10sb
STANDBY
?
1998/11/03 10:10
Sample Blank
ALARMCLEAR
If a Color Test is selected on the screen, a corresponding Blank Test must be specified.
Once these tests are inter-related in the parameters shown above, and the Color Test is ordered, the
test specified as the Blank Test will automatically be ordered.
See Also
For LIH procedures, refer to [Parameters], [Specific Test Parameters] located in the Software Chapter.
Program three tests with the following wavelength pairs: 410 nm and 480 nm (P1-P0) LIH1,
570 nm and 600 nm (P2-P0) LIH2, 660 nm and 800 nm (P3-P0) LIH3 to allow the determination
of the ODs at the six wavelengths used in the LIH OD calculations. These three tests may have
sample volume equal to 18.0 ul, sample diluent volume equal to 0.0 and reagent 1 volume plus
reagent 1 diluent volume equal to 282 ul. (It is highly recommended that these volumes be equal
to the numbers programmed in LIH.) Test No. 96 LIH MUST BE selected and use normal saline
as Reagent 1. Set the calibration for these three temporary tests to Cal type 1 MB with factor
equal to 1000 and decimal to 1. Set the Method to End, Reaction to + and reagent OD limits to
-2.0 to 2.5.
Run replicates of the cutoff levels of each type of interferant. Print the OD data for each sample
and each temporary LIH test. See [Routine], [Reaction Monitor] to select "List Print of Display
Data."
• To calculate the lipemia reaction OD, the equation is (OD 660-OD 800) both at P3 minus
reagent blank OD (660-800).
• To calculate the icterus reaction OD, the equation is (OD 480 at P1 - OD 570 at P2)
minus (OD 600 at P2 - OD 800 at P3) minus reagent blank.
• To calculate the hemolysis reaction OD, the equation is (OD 410 - OD 480) both at P1
minus (OD 600 at P2 - OD 800 at P3) minus reagent blank.
The final values for lipemia, icterus and hemolysis (LIH) are compared to the OD levels set in
[Parameters], [Specific Test Parameters] after the following adjustment is made:
LIH check data = Reaction OD x (R1 Volume + R1 Dilution Volume + Sample Volume +
Sample Dilution Volume)/(300) x 18/ (Sample Volume + Sample Dilution Volume)
OR
Reaction OD X R1 vol. + R1 dil. vol. + sample vol. + sample dil. vol. 18
LIH check data = 300
X Sample vol. + sample dil. vol
For example: If the lipemia, icterus and hemolysis reaction ODs are less than the OD limit set
for Level + for each of these, then the LIH flags are printed as LIP N, ICT N and HEM N. If a
reaction OD is equal to or greater than the Level OD limit, but less than the next higher Level
OD limit, then the appropriate Level flag of + to +++++ is printed. If the reaction OD limit at
any of the 6 wavelengths is greater than 2.5 or if OD 410 exceeds 2.0 then ABN for abnormal is
printed for Software versions up to 8.3. Software version 9.0 differentiates between ABN L for
abnormal low and ABN H for abnormal high depending on how high or low the OD wavelength
reaction limit is exceeded. If the sample was not analyzed for LIH it is left blank.
OLP4086E
In quality control, usually a normal control and an abnormal control are used. The following
figure shows a two-dimensional plot with the normal control on the x-axis and the abnormal
control on the y-axis.
Abnormal
region sample
+2SD
(1) is an example of normal
control results ~ - 1SD below
the mean and abnormal
MEAN control > -2SD below the mean.
-2SD
(1)
When the normal and abnormal controls are within control limits, but both controls are
repeatedly biased high or low, investigate system errors by performing a reagent blank or
calibration. If the problem persists reagent deterioration may be the cause; replace reagent
calibrate and repeat controls.
The twin plot offers the advantage of classifying problems as either a system error or a random
error. To be effective it should be combined with multi-rule control procedures to determine if
drift is occurring or if the reagent is unstable.
Multi-Rule Control
In the -R control, a control error is checked by examining multiple control rules, but it is
difficult if not impossible to confirm numerous tests on a real-time basis. The multi-rule control
technique makes it possible to speedily cope with an error real-time, as this control method
notifies the worker of just which rule an error (when generated) violates based on an error flag.
When using this control technique, test samples of both the normal range and the abnormal
OLP4087E
range. Refer to the following figure:
Control OUT-OF-CONTROL REJECT RUN
Data
Yes Yes
No No No
12S Trend IN-CONTROL Trend
Yes No
No No No No
13S 22S R4S 4 1S 10x
Tip
The term “consecutive” applies to either one identical control substance or a high-concentration and low-
concentration control substance.
4 Error Flag
R4S is a judgment level for determining if two control results, with high and low
concentrations, exceed the control limit “mean + 2SD” and if the other exceeds the control
limit of “mean – 2SD.” If the data is out of the control limit, then quality control has not
been attained. If the data is within the control limit, see the next judgment level, 41S.
5 Error Flag
41S is a judgment level for determining if four consecutive control results exceed the control
limit of either ‘mean +1 SD’ or ‘mean -1 SD.’ If they have exceeded the limit, then quality
control has not been attained. If they have not exceeded either control limit, an inquiry is
made to the next judgment level 10 .
6 Error Flag
10 is a judgment level for determining if 7 - 10 consecutive control results fall on one side
of the control mean (above or below). If they have exceeded the control limit, then quality
control has not been attained. If they have not exceeded the control limit, then quality control
has been attained.
7 Error Flag
Trend: Indicates that five through ten QC data points (user defined) exhibit steadily
increasing or decreasing values. (A number 7 error flag is generated.)
See Also
For more information on Error Flags, refer to the Error Flags Chapter.
For more information on Quality Control, many sources are available such as the following:
Tietz, N. W., ed. "Westgard Multi-Rule Control Chart," Fundamentals of Clinical Chemistry,
Third Edition. Philadelphia: Harcourt Brace Jovanovich, Inc., 1987. 246-247.
Example of Control Errors According to the Multi-Rule Control are shown below
High value sample 2 2S : Systematic error extending over the 2 type concentration range 1 3S
R 4S : Random error
3SD
2SD
1SD
Mean
Ð1SD
Ð2SD
Ð3SD
3SD
2SD
1SD
Mean
Ð1SD
Ð2SD
Ð3SD
The following describes the possible causes and corrective actions for random errors and
systematic errors shown in the figure above.
Random Errors
• Dispensing Problem (sample, reagent)
For troubleshooting information refer to the Troubleshooting Chapter.
• Reagent Deterioration
For troubleshooting information refer to "Troubleshooting Data Flags" in the Error Flags
Chapter.
See Also
For troubleshooting information refer to the Troubleshooting Chapter.
• Periodic Maintenance Not Performed Properly
For information on scheduled maintenance refer to the Maintenance Chapter.
• Poor Mixing
For troubleshooting information refer to the Troubleshooting Chapter.
See Also
For troubleshooting information refer to "Troubleshooting Data Flags" in the Error Flags Chapter.
• Deteriorated Reagent
For troubleshooting information refer to "Troubleshooting Data Flags" in the Error Flags
Chapter.
• Temperature Problems
For temperature specifications refer to the Specifications Chapter.
W1
Abbreviation for automatic wash of cuvettes. If analysis was stopped, a W1 is performed to
remove the sample remaining in the cuvette and the cuvette is washed.
W2
Abbreviation for automatic wash of cuvettes and tubing. Perform a W2 weekly. After performing
W2 be sure to perform the photocal measurement.
See Also
For information on performing a W2, refer to the Maintenance Chapter.
Photocal Measurement
To obtain appropriate analysis results, cuvettes are checked for stains or scratches. Confirm
the photocal data obtained from a photocal measurement using the analysis status keys and the
[Photocal Monitor] screen.
See Also
• For detailed information on performing a photocal, refer to the Maintenance Chapter.
• For detailed information about analysis status keys, refer to the Software Chapter.
• For detailed information about the [Photocal Monitor] screen, refer to the Software
Chapter.
Profile
The profile screen sets up test panels. The individual tests are assigned for each profile. Profile
No. 99 is assigned to the One Touch Mode. Therefore, this profile must be used for the most
frequently used analysis tests. Set the profiles by choosing [Profile] from the [Parameters]
screen.
Round
A round is a category in which a set of analysis tests used for similar analysis are grouped
together so that specific analysis tests can be accessed quickly. Set the desired round by
choosing [Round] from the [Parameters] screen. Example: Designate the analysis tests used
for the routine analysis to Round 1, and the analysis tests used for drug analysis to Round 2.
Perform routine analysis under Round 1 and switch to Round 2 for drug analysis as required.
ACAL
Abbreviation for auto-calibration. It represents the automatic creation of calibration curves.
A calibration curve is automatically created using the yellow rack. It is mainly used for the
analysis tests in the end point assay method.
MCAL
RB
Abbreviation for reagent blank. In routine analysis the reagent blank serves as the reference
value for the reagents at each photometric point of individual analysis tests. It also becomes the
Y-segment first data point of calibration curves created by ACAL.
LIH Testing
Performs a test for lipemia, icteric, and/or hemolysis in serum. LIH is the symbol used for
testing lipemia (L), icterus (I), and hemolysis (H). Checks the level of interference present that
may make the sample analysis result abnormal.
END
This is one of the methods available for individual test parameters of the system. In the end
point assay method, it defines the analysis method that uses the water blank absorbance as
the reference for measurement data at each photometric point. The water blank absorbance is
obtained from the photocal measurement. “END” is a method test for individual test parameters.
This is another end point assay method identical to “END 1.” It uses reagent blank absorbance
as the reference for measurement data at each photometric point. END 1 does not use the
reagent blank.
RATE
Set-up test for individual test parameters of the system. In the rate assay method, it defines
the analysis method that uses water blank absorbance as the reference for measurement data at
each photometric point. Water blank absorbance is obtained from the photocal measurement.
“RATE” is a set-up test for individual test parameters. RATE 1 is another rate assay method
identical to “RATE.” RATE uses reagent blank absorbance as the reference for measurement
data at each photometric point. RATE 1 does not use the reagent blank.
FIXED
Set-up test for individual test parameters of the system. In the fixed point assay method, it
defines the analysis method that uses water blank absorbance as the reference for measurement
data at each photometric point. Water blank absorbance is obtained from photocal measurement.
“FIXED” is a set-up test for individual test parameters. This is another fixed point assay method
identical to “FIXED 1.” It uses reagent blank absorbance as the reference for measurement data
at each photometric point. FIXED 1 does not use the reagent blank.
NO-LAG_TIME
For example, if analysis for a pathologic human sample using this system ends too quickly due
to rapid reaction, two or more photometric points of effective analysis data may not be obtained.
If this is the case, the system can be set-up to calculate the analysis result using the data in the
lag phase. Used for one of the analysis tests in the rate assay method.