LABORATORY

PROCESS

SOFT WARE

AUTOMATION

UniPol L-Series
Po l a r i m e t e r
Wi t h Au t o m a t i c
Wa ve l e n g t h Co n t r o l
(AW C )
SPECIFICATIONS UniPol L UniPol L1000 UniPol L2000

°Optical rotation, °Specific rotation, °Z International Sugar Scale, % Concentration (g/mL, g/100mL, g/L)
Measurement scales
7 scales freely definable

Measuring range ± 360° / ± 259°Z


SPECIFICATIONS UniPol L UniPol L1000 UniPol L2000
Resolution 0.01° / 0.02°Z 0.001° / 0.01°Z 0.001° / 0.01°Z
°Optical rotation, °Specific rotation, °Z International Sugar Scale, % Concentration (g/mL, g/100mL, g/L)
Measurement
Precision scales ± 0.01° /freely
± 0.03°Z * ± 0.005° / ± 0.02°Z * ± 0.005° / ± 0.02°Z *
7 scales definable

Reproducibility
Measuring range ± 0.01° / ± 0.03°Z ±±360°
0.005°
/ ±/259°Z
± 0.02°Z ± 0.005° / ± 0.02°Z

Sensitivity
Resolution Up to OD
0.01° 3
/ 0.02°Z 0.001° / 0.01°Z 0.001° / 0.01°Z

Wavelength
Precision 589 nm/(other
± 0.01° on *request)
± 0.03°Z ± 0.005° / ± 0.02°Z * ±405 and/ 589
0.005° nm *
± 0.02°Z

Response time
Reproducibility 6
± to 8 sec.
0.01° / ± over the entire measuring range
0.03°Z ± 0.005° / ± 0.02°Z ± 0.005° / ± 0.02°Z

Sensitivity Different
Up to ODmodels,
2 10 to 200 mm length; Material: glass, stainless steel, acid-proof stainless steel; stainless steel tubes
Measuring tubes
with integrated temperature sensor***; T-Cell Polarimeter tubes**** (Temperature control via Peltier-elements)
Wavelength 589 nm (other on request) 405 and 589 nm
Temperature measurement NTC Sensor
Response time 6 - 8 sec. over the entire measuring range
Temperature range 0°C to 99°C
Different models, 10 to 200 mm length; Material: glass, stainless steel, acid-proof stainless steel;
Measuring tubes
Resolution 0.01°C
stainless steel tubes with integrated temperature sensor***;T-Cell Polarimeter tubes (Temperature control via Peltier-elements)

Precision
Temperature measurement ± 0.03°C
NTC Sensor

Temperature regulation - T-cell tubes 18°C to 40°C

Resolution 0.01°C

Precision ± 0.5°C

Light source LED, interference filter

Display LCD-Display, monochrom

Operation Alpha numerical keybord, 20 characters inclusive function keys

Interfaces / Communication RS232 (2x), Parallel (1x), USB**, Ethernet**

Standard models UniPol L ; UniPol L P**** UniPol L1000 ; UniPol L1000 P**** UniPol L2000 ; UniPol L2000 P****

Conformity International Pharmacopoeia, OIML, ASTM, ICUMSA, Australian Standard K157

Standard circle polarimeter for common applications; GLP/GMP conform; 21 CFR part 11 ready; energy saving LED light source;
Highlights
Modell L 1000 / L 2000 with autom. wavelength control (AWC); capable for micro cuvettes

* Standard conditions
** Optional
Polarimeter applications *** Certificate on request
Polarimetry is an instrumental analytical method using the optical activity of inorganic and organic compounds **** P-version for T-cell tubes

as a non-destructive measure of their concentration in a solution.

Applications often used Typical applications of the models


• Determination of concentration • Raw, intermediate and final products of sugar cane and beet processing
© Schmidt + Haensch reserved all rights over texts and images

• Purity analysis • Food (sugar, starch, milk and food additives, sugar-free sweeteners like isomalt)
• Quality control • Dairy products (lactose, sucrose, lactoglobulin, lactic acids)
• Scientific analysis • Pharmacy (reception and product control)
Subject to modification without notice 10/15

For more information contact:


analytikLtd (UK and Ireland Distributor)
Barn B, 2 Cygnus Business Park, Middle Watch, Swavesey, Cambridge, CB24 4AA
T: +44 (0)870 991 4044 E: [email protected] www.analytik.co.uk

Schmidt + Haensch GmbH & Co.


Waldstraße 80-81, 13403 Berlin, Germany
Phone: + 49 (0 30) 417072-0, Fax: + 49 (0 30) 417072-99
[email protected], www.schmidt-haensch.com ISO 9001: 2008