General

3 Series
Specifications Coriolis Mass Flow and Density Meter
with Marine Approval
GS 01R04B10-00E-E

RCCF31 + RCCR31
RCCS34 - 39/IR +
RCCY038/039

Features
Contents
Page 1
FEATURES
• ROTAMASS is a flowmeter for nearly all fluids, including
Principle of measurement Page 1
Performance specifications Page 2
high viscosity liquids, slurries and multi-phase media
Normal operating conditions Page 3
with a certain gas content
Mechanical specifications Page 4
• Field-mount and rack-mount remote converter available
Electrical specifications Page 4 • Refined digital signal processing enables accurate and
Remote cable specification Page 5 stable measurement
CE Declaration Page 5 • A special detector decoupling system makes the device
Hazardous area specifications Page 6 highly independent from external loads or vibrations.
Planning and installation hints Page 9 • Simple flow path means self-draining, simple cleaning.
Dimensions Page 11 • High accuracy and high stability over a wide flow range
Model-, suffix- and option- codes Page 14 • Accurate density measurement, up to +/- 0.0005 g/cm³
• Concentration measurement for solutions, suspensions
and emulsions (e.g. water cut)
ROTAMASS is a Coriolis mass flowmeter with highly
• Volume flow with reference density
refined digital signal processing electronics, so that accurate
• Temperature measurement
and stable mass flow measurement is achieved.
• One analog output, two pulse outputs or status-out and
one status-in as standard I/O
ROTAMASS employs a flame-proof type converter case
• Available in explosion proof versions (ATEX, IECEx,
suitable for use in the hazardous area together with its
INMETRO, NEPSI, KOSHA)
intrinsically safety type detector.
• American NTEP approval for Custody Transfer Measurement
(see GS 01R04B07-00E-E)
ROTAMASS´s signal processing, housing protection and its
• Wide process temperature range –200°C to 230°C
detector´s special decoupling system against external loads and
• Microprocessor-based multifunction capability
vibrations, realize high performance in real applications.
• EEPROM protects parameter settings and totalized
values during power failure.
• High visibly LCD display
PRINCIPLE OF MEASUREMENT • HART® communication function is available.
Mass flow measurement according to the Coriolis principle. • MODBUS® communication is available
For abrasive or highly corrosive fluids please contact your • EN, ASME or JIS flanges as standard, others on request
Yokogawa representative. • Type Approval acc. DNV Rules for classification of ships

Rota Yokogawa GmbH & Co. KG GS 01R04B10-00E-E


Rheinstr. 8 ©Copyright January 2015 (RYG)
D-79664 Wehr 3rd edition, July 2017 (RYG)
Germany
2

PERFORMANCE SPECIFICATIONS Density Measurement


Adjustment with water and air at calibration temperature.
Model
- Remote detector RCCS34 to 39/IR : 2 tube design Measuring range:
- Remote field-mount converter RCCF31 RCCS34 - 38 : 0.3 kg/l to 5 kg/l
- Remote rack-mount converter RCCR31 RCCS39 -39/IR : 0.3 kg/l to 2 kg/l
No density measurement for gas applications.
Fluid to be measured : Liquid, gas or slurry With option /K4 thermal stabilization is acquired.
For further details about the option /K6 please refer to “special
Measurement Items : Mass flow, density, temperature calibrations” on page 3.
and derived from these values: Calibration condition standard
concentration, volume flow and Density : 0.9 kg/l ≤ r ≤ 1.1 kg/l
net flow Temp. Fluid : 22.5°C ± 12.5°C
Flow Rate : about 0.2 * Qnom as defined for each
Mass Flow Measurement
model
Table 1: measuring range
Calibration condition for /K6:
RCCS39
Model RCCS34 RCCS36 RCCS38 RCCS39
/IR Density : 0.7 kg/l ≤ r ≤ 1.65 kg/l
t/h 5 17 50 170 300 Temp. Fluid : 20°C ≤ T ≤ 80°C
Qmax
lb/h 11023 37478 110231 374785 661386 Temp. Ambient : 20°C ± 3K
t/h 3 10 32 100 250 Flow Rate : about 0.2 * Qnom as defined for each
Qnom model
lb/h 6613 22046 70547 220462 551155

Qnom is the water flow rate at about 1 bar pressure drop. Table 4: Accuracy (at calibration conditions):
The flowmeter has a default low cut of 0.05% of Qnom. Type Standard Option /K4 Option /K6

Liquid RCCS34 - 39/IR: RCCS34 0.003 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
± 0.1% of flow rate ± zero stability / flow rate *100% RCCS36 0.0022 g/cm³ 0.001 g/cm³ 0.0005 g/cm³

Accuracy of volume flow : RCCS38 0.0015 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
SQRT ( (mass flow error in %)² + (density error in %)²) RCCS39 0.0015 g/cm³ 0.001 g/cm³ 0.0005 g/cm³
RCCS9/IR 0.0015 g/cm³ ------ ------
Accuracy based on the frequency output includes the combined
effects of repeatability, linearity and hysteresis. Repeatability:
RCCS34-39/IR : ± 0.0005 g/cm³ (Std, /K4)
Repeatability of liquids: Static pressure effect:
± 0.05% ± (zero stability/2) / flow rate *100% Compensated if static pressure is set in the menu.
1.0
0.9
0.8 Specification of high performance density measurement
0.7
0.6 option /K6: Density calibration
0.5
0.4 Density range : 0.3 to 2.5 kg/l
0.3
Ambient temp. range : -10°C to 50°C (14°F to 122°F)
Error in %

0.2
0.1
0 Process temp. range Standard: -50°C to 150°C (58°F to 302°F)
-0.1
-0.2
-0.3
Minimum flow rate for specified accuracy:
-0.4
-0.5
- RCCS36 to RCCS39 : 700 kg/h (1543 lb/h)
-0.6
-0.7 - RCCS34 : 140 kg/h (308 lb/h)
-0.8
-0.9
-1.0
0 50 100 150 Maximum flow rate : Qnom
Flow in % of Qnom F10.EPS Repeatability : ±0.0002 g/cm³
Temperature measurement:
±(0.5°C+0.002*abs(Tmedium-20°C))
Table 2 : Zero stability Density accuracy : only for liquids, one phase
Model RCCS34 RCCS36 RCCS38 RCCS39
RCCS39 Process temperature influence :
/IR ±0.000015 g/cm³ * abs(Tmedium-20°C)
kg/h 0.15 0.5 1.6 5 13
lb/h 0.33 1.1 3.5 11 28.6 Temperature Measurement
Temperature measuring range of converter :
Pressure Dependency Standard, /LT, /MT : -200°C to 230°C (-328°F to 448°F)
The stiffness of the ROTAMASS tubes is slightly line pressure Accuracy:
dependent. The static pressure effect of mass flow and density Standard (-70°C to 150°C / -94°F to 302°F)
can be corrected by setting the static pressure manually via menu. : ±(0.5°C+0.005*abs(Tmedium-20°C))
Option /LT (-200°C to 150°C / -328°F to 302°F)
Table 3 : Static pressure effect on mass flow (if not corrected)
: ±(1.0°C+0.008*abs(Tmedium-20°C))
Model RCCS34 RCCS36 RCCS38 RCCS39
RCCS39 Option /MT (-70°C to 230°C / -94°F to 446°F)
/IR : ±(0.5°C+0.005*abs(Tmedium-20°C))
% of For process temperatures more than 80°C (176°F) higher
rate 0.00081 / 0.00346 / 0.00950 / 0.01058 / 0.0047 /
per bar (0.01175) (0.05018) (0.1378) (0.15345) (0.0682)
/ (psi)

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
3

than ambient temperature the detector should be insulated to


maintain optimum accuracy.
NORMAL OPERATING CONDITIONS
Ambient Temperature Ranges
Calibration for Liquids and Gases : - Remote detector RCCS3:
The ROTAMASS flowmeters are always factory calibrated with Standard : -30°C to +70°C (-22°F to 158°F)
water. Option /LT : -30°C to +70°C (-22°F to 158°F)
Calibration Conditions: Option /MT : -30°C to +70°C (-22°F to 158°F)
- Water : 22.5°C ± 12.5°C (72.5°F ± 22.5°F) Terminal box temperature should not exceed 90°C
- Ambient temperature : 22.5°C ± 12.5°C (72.5°F ± 22.5°F) - Remote converter RCCF31, RCCR31:
- Process pressure : 1 to 2 bar abs Display operating range
- Installation: RCCS34 to RCCS38 vertical : -20°C to +60°C (-4°F to 140°F)
RCCS39 to RCCSS39/IR horizontal Electronic operating range
All specifications are based on above mentioned calibration : -40°C to +60°C (-40°F to 140°F)
reference conditions, a flow calibration protocol is attached to Cold start : above -30°C (-22°F)
each instrument. Where meters are mounted in direct sunlight, it is recom-
mended to install a sunshade. This is particularly important in
Special Calibrations countries with high ambient temperatures.
- Mass-/Volume flow calibration with factory certificate (option /K2):
Calibration with water at customer specified flow values Ambient Humidity Range : 0 to 95% RH
according calibration order sheet.
- Mass-/Volume flow calibration with DAkkS certificate acc. Process Temperature Ranges
EN17025 (option /K5): Detector :
Calibration with water at customer specified flow values - RCCS34 to 39/IR : -70°C to 150°C (-94°F to 302°F)
according calibration order sheet. - RCCS34 to 39/IR /LT : -200°C to 150°C (-328°F to 302°F)
- Density calibration with factory certificate (option /K6): - RCCS34 to 39/IR /MT : -70°C to 230°C (-94°F to 446°F)
Adjustment and check with 3 different fluids, fluid temperature
influence adjustment for low ambient temperature influence For use in hazardous area see “Hazardous Area Specifications”
and thermal treatment for long term density measurement
stability, improved temperature measurement accuracy (see
Process Pressure Limit
also page 10).
The process pressure limit depends on the option /MC2, /MC3
as shown in the table below.

Classes of piping systems:
Marine option /MC2 /MC3
Piping system for Class II Class III
p (bar) t (°C) p (bar) t (°C)
Steam ≤ 16 ≤ 300 1)
≤7 ≤ 170
Fuel oil, ≤ 16 ≤ 150 ≤7 ≤ 60
lubricating oil,
flammable
hydraulic oil
Other media ≤ 40 ≤ 300 1)
≤ 16 ≤ 200

1)
The specified temperatures are defined by DNV classes for
piping systems. The de facto process temperature ranges of
ROTAMASS depend on the instrument configuration.

The pressures mentioned in the table above are valid for


DNV Type approval. Table 9 shows the maximum possible
design pressures. The respectivly lowest pressure must be
used.
The maximum temperature and process pressure limits of an
instrument are marked on the name plate as TS and PS.
The given temperature/pressure ranges are calculated and
approved without corrosion or erosion effects. In case of
heavy corrosion and/or erosion the instrument may not
withstand the pressure and an incident may happen with
human and/or environmental harm. Yokogawa will not take any
liability regarding damage caused by corrosion / erosion.
If corrosion / erosion may happen, the user has to check
periodically if the necessary wall thickness is still in place.

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
4

Gas Content Limits for Liquid / Gas Mixtures Wetted Parts


Gas content limit is defined as the amount of gas in a liquid / - RCCS34 to 39/IR :
gas mixture which generates an error in the converter. The Measuring tubes and process connection
gas content limit is dependent on viscosity, surface tension : 316L/1.4404/14435
and bubble size of the liquid / gas mixture.
Furthermore it is highly flow rate dependent (the higher the Table 5 : Diameter of measuring tubes
flow rate, the lower the gas content limits). The stated values RCCS39
are for a flow of 50% of Qnom and water / air without /HP: Type RCCS34 RCCS36 RCCS38 RCCS39
/IR
Model Gas content limit Inner mm 7.7 13.4 22.1 37.2 54.5
diameter inches 0.303 0.528 0.870 1.485 2.146
RCCS34 no limitation
Wall mm 0.89 1.24 1.65 2.6 2.9
RCCS36 approx. 50% thickness inches 0.035 0.049 0.065 0.102 0.114
RCCS38 approx. 30%
Chemical composition of wetted materials 316L/1.4404
RCCS39 approx. 7% conform with
RCCS39/IR approx. 3% ANSI / NACE-MR0175 / ISO15156-2
ANSI / NACE-MR0175 / ISO15156-3
With option /HP the gas content limits are improved. NACE MR0103
With liquid/gas mixtures the specified mass flow accuracy will For details please see RYG´s declaration about NACE
not be achieved. conformity 8660001
For short time aeration a function can be activated to keep
the current outputs constant during the aeration time. Pressure Equipment Directive 2014/68/EU
Detectors comply with Directive 2014/68/EU on Pressure
Other 2 Phase Flow, liquid/solid and liquid/liquid Equipment for fluid group 1 and 2.
Two phase flow can generate minus span errors. The errors Note: PED-compliance is not applicable for Rotamass
are proportional to the difference in density between the 2 assembled in the USA that are identifiable by nameplate
phases and the amount of the second phase. If the particles containing “Assembled in USA with Foreign and Domestic Parts”.
or droplets are very small no errors will be generated.
CRN : CRN 0F12074.5
Secondary Containment Approved process connections
Model Typical rupture Option /J1 see table 12
pressure pressure test *)
Vibration Test :
RCCS34-36 120 bar / (1740 psi) 60 bar / (870 psi) - Acc. IEC 60068-2-64
RCCS38 120 bar / (1740 psi) 40 bar / (580 psi) - Acc. DNV standard for Certification No. 2.4
RCCS39 80 bar / (1160 psi) 10 bar / (145 psi) Location class:
RCCS39/IR 50 bar / (725 psi) ---
Detector: B, accuracy for some frequencies within
±1% of span
*) Pressure test with safety factor S=1.1 Converter: A, accuracy no impact
If the detector housing is exposed to a pressure close to
the rupture pressure it will deform and measurement will be
strongly influenced. Therefore the pressure test of the housing ELECTRICAL SPECIFICATIONS
(option /J1) can only be done as shown in above table. Power Supply
- DC- type : 20.5 V to 28.8 V DC
Power consumption : max. 25 VA / 10 W
MECHANICAL SPECIFICATIONS External circuit breaker rating : 5 A, 250 V (The converter
Protection Class doesn’t feature an installed power switch).
- RCCF31 : IP66/67
- RCCR31 : IP20
- RCCS3 : IP66/67 Fuse on Base Board
In case of RCCR31 for marine application the customer has - DC- type : 2 A, T, breaking capacity 1500A
to ensure the minimum required IP grade in accordance with
the installation area on board. I/O Signals, including /KF1, /EF1, /UF1, /NF1
- One active current output: Iout1
Materials 4 to 20 mA DC, galvanic separated from other signals,
- Detector housing : Stainless steel 304/1.4301 Load resistance : 20 Ω to 600 Ω
- Detector terminal box : 316L/1.4404 Ambient temperature effect : < 0.05% of span/10°C
- Detector gas filling plug: 1.4305
Linearity : 0.008 mA = 0.05% of span
- Detector insulation housing
: Stainless steel 304/1.4301 Setting range URV for liquids:
- Field- mount converter housing -Qmax to + Qmax , Abs(URV-LRV) ≥ 5% of Qnom
: Aluminium alloy with Polyurethane Setting range URV for gases:
corrosion-resistant coating or -Qmax to + Qmax , Abs(URV-LRV) ≥ 1% of Qnom
epoxy coating (option /X1) - Two pulse outputs / status outputs : Pout, Sout
- Field- mount converter mounting bracket: Passive transistor contact output, 30 V DC, 200 mA
: Stainless steel 304/1.4301 Output rate
- Rack- mount converter housing Output 1 : 0 to 10000 pulses/s
: Aluminium
Output 2 : 0 to 2000 pulses/s
- Name plates : Stainless steel 304/1.4301
Option /NM : passive, according EN 60947-5-6
Coating Color Option /AP : active output, 12 V, 6 mA, RL> 10 kΩ
- Field-mount converter case As frequency output
: Mint green Output 1 : 20 Hz to 10000 Hz
- Status input : Voltage-free contact
Closed : < 200 Ω
Open : > 100 kΩ

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
5

Digital Communication Safety Requirement Standards


- HART® communication protocol rev. 5, superimposed on Acc. IEC 61010-1
4 -20 mA DC signal (Iout1) IEC 61010-2-030
- Load resistance : 230 Ω to 600 Ω (including cable) Overvoltage category II
- Power line spacing : >15 cm, avoid parallel wiring Pollution degree 2
- Cable length : ≤ 2 km if „CEV” cables are used
- MODBUS® communication (/MB2 + /MB3)
- Physical interface RS485 as two wire data bus according CE DECLARATION
EIA/TIA-485 Pressure Equipment Directive 2014/68/EU
- Maximal bus length is depending from bus topology and Detectors comply with Directive 2014/68/EU on Pressure
communication speed. Equipment for fluid group 1 and 2.
- Addresses: 1 to 247 Note: PED- compliance is not applicable for Rotamass
- Baud rates: 1200, 2400, 4800, 9600, 19200, 38400, assembled in the USA that are identifiable by nameplate
57600, 76800, 115200 containing “Assembled in USA with Foreign and Domestic Parts”.
- Data formats: E/1, O/1, N/1, N/2 (parity/stop bits)
- Modes: RTU, ASCII Electromagnetic Compatibility
Acc. IEC 61326-1: Class A, Table 2
Setting Functions IEC 61326-2-3
Parameter setting is possible by using the infrared switches IEC 61000-3-2
on the display or with HART® communication. For comfortable IEC 61000-3-3
setting we recommend to use Yokogawas Field Mate setting
tool (DTM based). Safety Requirement Standards
The option /MB2 can be set via Modbus and display. Acc. IEC 61010-1
The option /MB3 can be set via Modbus, HART® and display. IEC 61010-2-030
Overvoltage category II
Display Function Pollution degree 2
- Up to 4 lines.
- Language English Intended use
Rotamass 3 series Coriolis mass flow and density meter is
intended to measure mass flow of fluids and gases while
Damping Functions simultaneously also capturing their density and temperature.
The damping function is adjustable from 0 to 200 seconds Yokogawa is not liable for damages or penalties caused by
and affects the display and outputs. use deemed contrary to the Intended Use of the product.

Isolation Resistance of Converter Rotamass 3 series Coriolis mass flow and density meter is
When surge arrestors are removed a tailor-made product that is specifically designed to both be
- between power and ground terminal: >100 MΩ / 500 V DC installed and function within:
- between power and I/O terminals : >20 MΩ / 100 V DC - Large-scale fixed installations
- between I/O terminals and ground : >20 MΩ / 100 V DC - Means of transport for persons or goods, excluding electric
two-wheel vehicles which are not type-approved
The EU Directive 2011/65/EU (RoHS II) does not apply to this
Dielectric Strength product and is not part of its CE-Marking. This product must
When surge arrestors are removed not be used beyond its Intended Use. Please consider the
- between power and ground terminal : 1,500 V AC for 1 minute above if the intended installation location of your device is
inside European Economic Area.
Lightning Protection
Arresters 2000 A are inside of the converter for power
supply lines.

Electromagnetic Compatibility
Acc. IEC 61326-1: Class A, Table 2
IEC 61326-2-3
IEC 61000-3-2
IEC 61000-3-3
DNV standard for Certification No. 2.4, Class A

REMOTE CABLE RCCY03 SPECIFICATION


DNV certified cable
Max. 300m / 900ft may be installed
Table 6 : Cable specifications
Model code Temperature Wire gauge Resistance of Capacitance Capacitance Inductance of
range loop wire/wire wire/shield loop
-30°C to +90°C
RCCY037 7x2x0.5mm² 80.8 Ω/km 81 nF/km 63 nF/km 0.63 mH/km
-22°F to +176°F

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
6

HAZARDOUS AREA SPECIFICATIONS Electrical data remote converter RCCF31, RCCR31:


- Driving circuit : terminals D+ / D-
Ex [ib] IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W
ATEX Lo = 15 mH; Co = 0.65 µF
Ex [ib] IIB : Uo = 11.7 V; Io = 124 mA; Po = 0.363 W
Remote detector RCCS34 ... 39/IR (Option /KS1): Lo = 8 mH; Co = 10.3 µF
- KEMA 01ATEX 1075 X - Sensor circuits: terminals S1+/ S1- or S2+ / S2-
- Intrinsically safe Ex [ib] IIB/IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W
- II 2G Ex ib IIB/IIC T1 ... T6 Gb Ex [ib] IIC : Lo = 15 mH; Co = 0.65 µF
- II 2D Ex ib IIIC Txxx Db Ex [ib] IIB : Lo = 60 mH; Co = 4.07 µF
(xxx = max. surface temperature see below) - Temperature sensor circuit : terminals TP1, TP2, TP3
- Max. surface temperature : Ex [ib] IIB/IIC : Uo = 13.3 V; Io = 40 mA; Po = 0.133 W
Standard + /LT : 150°C (302°F) Ex [ib] IIC : Lo = 20 mH; Co = 0.91 µF
/MT : 220°C (500°F) Ex [ib] IIB : Lo = 80 mH; Co = 5.6 µF
- Degree of protection : IP66/67
- Ambient humidity : 0 to 95% RH
- Ambient temperature range For temperature classification see table 7.
Standard , option /LT and option /MT
: -30°C to +70°C (-22°F to 158°F)
- Process temperature limits :
Standard : -50°C to 150°C (-58°F to 302°F)
Option /LT : -200°C to 150°C (-328°F to 302°F)
Option /MT : -50°C to 220°C (-58°F to 428°F)

Remote converter RCCF31 (Option /KF1) :


- KEMA 02ATEX 2183 X
- Flame proof with Intrinsically safe connection to detector (ib)
- II 2G Ex d [ib] IIC T6 Gb or Ex d e [ib] IIC T6 Gb
- II 2G Ex d [ib] IIB T6 Gb or Ex d e [ib] IIB T6 Gb
with option /HP
- II 2D Ex tb [ib] IIIC T75°C Db
- Max. surface temperature : 75°C (167°F)
- Degree of protection : IP66/67
- Power supply : 20.5 to 28.8 V DC
- Power consumption : max. 25 VA / 10 W
- Ambient humidity : 0 to 95% RH
- Ambient temperature range: -30°C to +55°C (-22°F to 131°F)

Remote converter RCCR31 (Option /KS1) :


- KEMA 02ATEX 2183 X
- Associated apparatus with Intrinsically safe connection
to detector (ib)
- II (2) G [Ex ib Gb] IIC
- II (2) G [Ex ib Gb] IIB with option /HP
- II (2) D [Ex ib Db] IIIC
- Power supply : 20.5 to 28.8 V DC
- Power consumption : max. 25 VA / 10 W
- Ambient humidity : 0 to 95% RH
- Ambient temperature range: -30°C to +55°C (-22°F to 131°F)

WARNING
Remote rack-mount converter RCCR31 must be installed in
safe area !

Electrical data remote detector RCCS34 ... 39/IR :


- Driving circuit : terminals D+ and D-
Ex ib IIC : Ui = 16 V; Ii = 53 mA; Pi = 0.212 W
Li = 3.2 mH; Ci = negligible small
Ex ib IIB : Ui = 16 V; Ii = 153 mA; Pi = 0.612 W
Li = 3.2 mH; Ci = negligible small
- Sensor circuits: terminals S1+ and S1- or S2+ and S2-
Ex ib IIC : Ui = 16 V; Ii = 80 mA; Pi = 0.32 W
Li = 2.1 mH; Ci = negligible small
- Temperature sensor circuit : terminals TP1, TP2, TP3
Ex ib IIC : Ui = 16 V; Ii = 50 mA; Pi = 0.2 W
Li = negligible small; Ci = negligible small

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
7

IECEx APPROVAL Electrical data remote converter RCCF31, RCCR31:


Certificate: IECEx KEM 06.0031X - Driving circuit : terminals D+ / D-
Ex [ib] IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W
Remote detector RCCS34 ... 39/IR (Option /ES1): Lo = 15 mH; Co = 0.65 µF
- Intrinsically safe Ex [ib] IIB : Uo = 11.7 V; Io = 124 mA; Po = 0.363 W
- Ex ib IIB/IIC T1 ... T6 Gb Lo = 8 mH; Co = 10.3 µF
- Ex ib IIIC Txxx Db - Sensor circuits: terminals S1+/ S1- or S2+ / S2-
(xxx = max. surface temperature see below) Ex [ib] IIB/IIC : Uo = 14.5 V; Io = 47 mA; Po = 0.171 W
-Max. surface temperature : Ex [ib] IIC : Lo = 15 mH; Co = 0.65 µF
Standard + /LT : 150°C (302°F) Ex [ib] IIB : Lo = 60 mH; Co = 4.07 µF
/MT : 220°C (500°F) - Temperature sensor circuit : terminals TP1, TP2, TP3
- Degree of protection : IP66/67 Ex [ib] IIB/IIC : Uo = 13.3 V; Io = 40 mA; Po = 0.133 W
- Ambient humidity : 0 to 95% RH Ex [ib] IIC : Lo = 20 mH; Co = 0.91 µF
- Ambient temperature range Ex [ib] IIB : Lo = 80 mH; Co = 5.6 µF
Standard , option /LT and option /MT
: -30°C to +70°C (-22°F to 158°F) For temperature classification see table 7.
- Process temperature range :
Standard : -50°C to 150°C (-58°F to 302°F)
Option /LT : -200°C to 150°C (-328°F to 302°F)
Option /MT : -50°C to 220°C (-58°F to 428°F)

Remote converter RCCF31 (Option /EF1) :


- Flame proof with Intrinsically safe connection to detector (ib)
- Ex d [ib] IIC T6 Gb or Ex d e [ib] IIC T6 Gb
- Ex d [ib] IIB T6 Gb or Ex d e [ib] IIB T6 Gb
with option /HP
- Ex tb [ib] IIIC T75°C Db
- Max. surface temperature : 75°C (167°F)
- Degree of protection : IP66/67
- Power supply : 20.5 to 28.8 V DC
- Power consumption : max. 25 VA / 10 W
- Ambient humidity : 0 to 95% RH
- Ambient temperature range: -30°C to +55°C (-22°F to 131°F)

Remote converter RCCR31 (Option /ES1) :


- Associated apparatus with Intrinsically safe connection to
detector (ib)
- [Ex ib Gb] IIC
- [Ex ib Gb] IIB with option /HP
- [Ex ib Db] IIIC
- Power supply : 20.5 to 28.8 V DC
- Power consumption : max. 25 VA / 10 W
- Ambient humidity : 0 to 95% RH
- Ambient temperature range: -30°C to +55°C (-22°F to 131°F)

WARNING
Remote rack-mount converter RCCR31 must be installed in
safe area !

Electrical data remote detector RCCS34 ... 39/IR :


- Driving circuit : terminals D+ and D-
Ex ib IIC : Ui = 16 V; Ii = 53 mA; Pi = 0.212 W
Li = 3.2 mH; Ci = negligible small
Ex ib IIB : Ui = 16 V; Ii = 153 mA; Pi = 0.612 W
Li = 3.2 mH; Ci = negligible small
- Sensor circuits: terminals S1+ and S1- or S2+ and S2-
Ex ib IIC : Ui = 16 V; Ii = 80 mA; Pi = 0.32 W
Li = 2.1 mH; Ci = negligible small
- Temperature sensor circuit : terminals TP1, TP2, TP3
Ex ib IIC : Ui = 16 V; Ii = 50 mA; Pi = 0.2 W
Li = negligible small; Ci = negligible small

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
8

Table 7 : Temperature classification for ATEX, IECEx, INMETRO, NEPSI and KOSHA certified flowmeter
RCCS34 to RCCS39/IR
without insulation
Temp. class Max. ambient Max. process
temperature temperature
T6 40°C / 104°F 40°C / 104°F
T5 55°C / 131°F 55°C / 131°F
70°C / 158°F 100°C / 212°F
T4
40°C / 104°F 120°C / 248°F
70°C / 158°F 160°C / 320°F
T3
40°C / 104°F 180°C / 356°F
T2 70°C / 158°F 220°C / 428°F
T1 70°C / 158°F 220°C / 428°F

INMETRO APPROVAL (For Brazil)


Certificate TÜV 11.0419 X
RCCS3 with option /US1 same as IECEx /ES1
RCCF31 with options /UF1
RCCR31 with option /US1
Same parameters and specifications as IECEx approval.

NEPSI APPROVAL (For China)


Certificate GYJ12.1381X
RCCS3 with option /NS1
RCCF31 with options /NF1
RCCR31 with option /NS1
Same parameters and specifications as IECEx approval
except NEPSI has no dust proof certification.

KOSHA APPROVAL (For Korea)


Same parameters and specifications as IECEx approval.
Meter with IECEx option must be ordered.

TS APPROVAL (For Taiwan)


Certificate ML0412007041H6
Same parameters and specifications as IECEx approval.
Meter with IECEx option must be ordered.
For export to Taiwan please contact your Yokogawa
representative regarding Taiwan Safety Mark.

PESO APPROVAL (For India)


Same parameters and specifications as ATEX approval.
Meter with ATEX option must be ordered.

EAC APPROVAL (For Russia, Kazakhstan, Belorussia)


Certificate RU C-DE.ГБ08.B.00208
RCCS3 with option /GS1
RCCT3 with options /GF1 or /GF5
RCCF31 with options /GF1 or /GF5
RCCR31 with option /GS1
Same parameters and specifications as IECEx approval.

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
9

Zero Adjustment Function


PLANNING AND INSTALLATION HINTS Zero point can be adjusted either by setting the switches on
display, with the HART® communication, with Modbus
Design Limits communication or with status input when the fluid flow is
It is the responsibility of the user to use the instrument within stopped and the detector filled.
the given design limits. Erosion and corrosion influence the To ensure no flow conditions isolation valves should be
accuracy and may restrict the temperature / pressure limits. installed. To achieve the specified accuracy a zero should be
Therefore corrosion and erosion should be avoided.
performed at process conditions (temperature, pressure).
Installation
Pressure / Temperature Dependencies of Process Connections
The flowmeter can be installed vertically, horizontally or in any
See also process pressure limits in chapter ”Normal operation
other position, as long as the measuring tubes are completely
conditions”.
filled with the measured liquid during measurement.
Explosion Proof Concept
Redundant Installation
The detector is intrinsically safe Ex ib, the converter RCCF31
If two flowmeters of the same size are installed in series
are flame (explosion) proof. The converter RCCR31 is an
mutual interference called cross talk may take place. Cross
talk occurs due to the fact that both meters have the same intrinsically safe associated apparatus. The driving power from
resonance frequency. If serial installation is planned please converter to detector is limited and protected by an intrinsically
contact your Yokogawa representative who can ensure that safe barrier, which is part of the converter. The barrier is
a frequency adjustment is made to one of the meters at the protecting the detector either for gas group IIC or IIB (option /HP).
factory.
Batch Process
Sizing The specified mass flow accuracy applies if the batch process
The measuring range and accuracy are virtually independent is >1 minute. For shorter batch time (Dt in s) the accuracy
of fluid conditions and size of the connecting pipe. Select a decreases with the square root of 60/Dt.
suitable nominal size from pressure loss calculation. Check For short batches the opening and closing times of the valves
whether the measuring range and accuracy at minimal flow have to be longer than 2 seconds.
fit the application. The calculations of the pressure loss are
based on newtonian fluids. For correct calculation please con- Concentration Measurement for Liquids
tact your local Yokogawa representative. The Standard Concentration Measurement (option /CST)
is suitable for concentration measurement of emulsions or
Cavitation suspensions, where the density of the solid is assumed to
To avoid cavitation keep the back pressure of the fluid suf- be fix. It can also be used for (mainly low concentration)
ficiently above the vapor pressure of the fluid. For low viscous solutions if the two fluids are not strongly interacting. The
fluids following condition should be fulfilled at the given density change of the liquid components due to temperature
temperature: can normally be described with a linear or quadratic function
pback > pvapor + 0.7*∆p with very high accuracy within the desired measurement
With ∆p = pressure loss range. The coefficients of these function (linear and quadratic
thermal expansion coefficients) must be either known or have
Long Term Stability to be determined prior to using this function.
To get stable deflection of the tubes by the Coriolis forces the For interacting liquids the Advanced Concentration
stiffness and therefore the wall thickness has to kept constant Measurement options should be used, these options can
during measuring. With corrosion or erosion the meter factor be ordered using the appropriate /C concentration
is drifting with time and recalibration is necessary. measurement option. For more information please see
TI 01R04B04-04E-E “Concentration Measurement with
Recalibration Service ROTAMASS”.
Yokogawa offers full recalibration service, if necessary with a
certificate traceable to German national standards. Please
contact your Yokogawa affiliate or directly Rota Yokogawa,
Germany.

Installation above 100°C (212°F) Process Temperature


To provide enough cooling the instrument should be installed
vertically or horizontally with the converter down.

Installation below 0°C (32°F) Process Temperature


The detector can be insulated to prevent ice capping either
by the customer or by the manufacturer. Ask your Yokogawa
representative for special insulation. If the customer wants to
insulate by themselves a closed cell foam as insulation mate-
rial is recommended to avoid water siphon. In this case option
/S2 should be selected. For temperatures below -70°C (-94°F)
option /LT is recommended (on request).

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
10

Density Measurement
There are 3 levels of density measurement. The standard adjustment and /K4 delivers an accuracy up to 0.001 g/cm³, if the fluid den-
sity is around 1 kg/l. However, at elevated temperatures the density error may increase. For option /K4 the instrument is
preheated ensuring long term stability. Option /K6 includes preheating, a full calibration at 3 different densities, increased temperature
measurement specification and individual adjustment of the fluid temperature dependency. Multi-phase flow can generate higher de-
viations. The higher the density differences of the single components are the more likely it is that a negative density error is generated.
Aeration has to be avoided fully to receive good density measurement.
For more information please see TI 01R04B04-05E ”Density Measurement with ROTAMASS”.
Note: Density specification under calibration condition only with flow direction “forward” according the arrow on the meter.
Table 8: Overview density-/volume- flow measurement:
Option Accuracy Certificate Description Typical Application
Standard ± 0.0015 g/cm³ to Standard (mass flow) - Standard adjustment with water and air - Process medium and environment are
± 0.02 g/cm³ factory calibration - Density constants given in mass flow approximately at room temperature, the density
certificate certificate range is 0.9 kg/l to 1.1 kg/l
Option /K4 ± 0.001 g/cm³ Standard (mass flow) - Thermal treatment of the sensor and - Improved volume flow accuracy
factory calibration special hardware design - Process medium up to 150°C, for higher
certificate - Standard adjustment with water and air temperature select option /MT
- Density constants given in mass flow - Density range is 0.9 kg/l to 1.1 kg/l
certificate
Option /K6 ± 0.0005 g/cm³ Separate factory - Thermal treatment of the sensor and - Density and concentration measurement in
density calibration special hardware design addition to the mass flow:
certificate - Density calibration with 3 different liquids - Process medium up to 150°C
- Individual adjustment of the fluid - Density range 0.3 kg/l to 2.5 kg/l
temperature dependency - Best volume flow accuracy

Table 9 : Pressure rating


Process Temperature
Type of process connection
RT 1) 50°C 100°C 150°C 200°C 250°C
A1 Flange acc. ASME B16.5 Class 150 19 bar 18.4 bar 16.2 bar 14.8 bar 13.7 bar 12.1 bar
A2 Flange acc. ASME B16.5 Class 300 49.6 bar 48.1 bar 42.2 bar 38.5 bar 35.7 bar 33.4 bar
A3 Flange acc. ASME B16.5 Class 600 excl. RCCS39/IR 99.3 bar 96.2 bar 84.4 bar 77 bar 71.3 bar 66.8 bar
A3 Flange acc. ASME B16.5 Class 600 for RCCS39/IR 92 bar 89 bar 80 bar 73 bar 67 bar 62 bar
A4 Flange acc. ASME B16.5 Class 900 127 bar 127 bar 110.5 bar 100.8 bar 93.4 bar 87.4 bar
D2 Flange acc. EN 1092-1 PN 16 16 bar 15.6 bar 15.1 bar 13.7 bar 12.7 bar 11.9 bar
D4 Flange acc. EN 1092-1 PN 40 40 bar 39.0 bar 37.9 bar 34.4 bar 31.8 bar 29.9 bar
D5 Flange acc. EN 1092-1 PN 63 63 bar 61.4 bar 59.7 bar 54.3 bar 50.1 bar 47.1 bar
D6 Flange acc. EN 1092-1 PN 100 100 bar 97.4 bar 94.7 bar 86.1 bar 79.5 bar 74.7 bar
D7 Flange acc. EN 1092-1 PN 160 160 bar 156.9 bar 151.6 bar 137.9 bar 127.2 bar 119.6 bar
G9 Internal thread RCCS34 181 bar 181 bar 164 bar 149 bar 136 bar 126 bar
T9 Internal thread NPT RCCS34 181 bar 181 bar 164 bar 149 bar 136 bar 126 bar
Process Temperature
up to 120°C 220°C
J1 Flange acc. JIS B 2220 10K 14 bar 12 bar
J2 Flange acc. JIS B 2220 20K 34 bar 31 bar
Process Temperature
Type of process connection
RT 1) 120°F 210°F 300°F 390°F 480°F
A1 Flange acc. ASME B16.5 Class 150 276 psi 267 psi 235 psi 215 psi 199 psi 175 psi
A2 Flange acc. ASME B16.5 Class 300 719 psi 698 psi 612 psi 558 psi 518 psi 484 psi
A3 Flange acc. ASME B16.5 Class 600 excl. RCCS39/IR 1440 psi 1395 psi 1224 psi 1117 psi 1034 psi 969 psi
A3 Flange acc. ASME B16.5 Class 600 for RCCS39/IR 1331 psi 1291 psi 1160 psi 1059 psi 972 psi 899 psi
A4 Flange acc. ASME B16.5 Class 900 1838 psi 1838 psi 1667 psi 1461 psi 1384 psi 1283 psi
D2 Flange acc. EN 1092-1 PN 16 232 psi 226 psi 219 psi 198 psi 184 psi 172 psi
D4 Flange acc. EN 1092-1 PN 40 580 psi 565 psi 549 psi 499 psi 458 psi 433 psi
D5 Flange acc. EN 1092-1 PN 63 914 psi 890 psi 865 psi 787 psi 726 psi 683 psi
D6 Flange acc. EN 1092-1 PN 100 1450 psi 1412 psi 1373 psi 1248 psi 1152 psi 1083 psi
D7 Flange acc. EN 1092-1 PN 160 2320 psi 2275 psi 2198 psi 1999 psi 1844 psi 1734 psi
G9 Internal thread RCCS34 2620 psi 2620 psi 2375 psi 2159 psi 1972 psi 1828 psi
T9 Internal thread NPT RCCS34 2620 psi 2620 psi 2375 psi 2159 psi 1972 psi 1828 psi
Process Temperature
up to 248°F 428°F
J1 Flange acc. JIS B 2220 10K 1203 psi 174 psi
J2 Flange acc. JIS B 2220 20K 493 psi 449 psi
1)
RT = Room Temperature; EN1092: -10°C to 50°C; ASME B16.5: -29°C to 38°C

For marine application acc. DNV Type approval the process pressure is limited according section „Process pressure limit“ (page 3).

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
11

DIMENSIONS
Remote field-mount Converter RCCF31
Without Display

17 (0.66)
Ø102 (4.01) 75 (2.95)

201(7.91)
(4.72)
91.6 (3.6)

120
Ø123(4.84)

(2.51)

(12)
305
122 (4.80)
64
29 32 29
(1.14) (1.25) (1.14)

85 (3.34)
124 122
(4.88) (4.80) 148 (5.82)

With Display 17 (0.66)


Ø102 (4.01) 75 (2.95)

201(7.91)
(4.72)
120
91.6 (3.6)

Ø123 (4.84)

(12)
305
(2.51)
122 (4.80)
64

49 32 29
(1.92) (1.25) (1.14)

85 (3.34)
144 (5.66) 122 (4.80) 148 (5.82)

Weight with bracket: 5.5 kg (12.1 lbs) (depends on type)

Dimensions in mm (inches)

Remote rack-mount Converter RCCR31


Terminal Board, back view Cassette, side view Cassette front view
128 (3HE)

22 228 107 (21TE)

Dimensios in mm
19-inch rack, acc. IEC / DIN EN 60297-3-101
Weight: 1.5 kg

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
12

Remote Detector RCCS34, RCCS36, RCCS38, RCCS39


Ø102 (4.01)

L1 ± 5 (0.2) Ø102 (4.01)

(3.14)
80
98 (3.85)

H6
H5
H4

H1
L3 W1 Option /S2, /MT,
Option /LT with
L2
Hazardous Area Approval
W2

Note: The flange dimensions depend on size and pressure rating of the flange.
Model L1 L2 L3 H1 W1 W2 H4 H5 H6 Weight
mm 10-21 kg
RCCS34 see table 11 272 (10.7) 212 (8.35) 177 (6.97) 60 (2.36) 80 (3.15) 80 (3.15) 138 (5.43) 218 (8.58)
(inches) (22-46 lbs)
mm 15-35 kg
RCCS36 see table 11 400 (15.7) 266 (10.5) 230 (9.06) 76 (2.99) 90 (3.54) 80 (3.15) 138 (5.43) 218 (8.58)
(inches) (33-77 lbs)
mm 25-50 kg
RCCS38 see table 11 490 (19.3) 267 (10.5) 269 (10.6) 89 (3.5) 110 (4.33) 100 (3.94) 148 (5.82) 228 (8.97)
(inches) (55-110 lbs)
mm 60-103 kg
RCCS39 see table 11 850 (33.5) 379 (14.9) 370 (14.6) 129 (5.08) 160 (6.3) 135 (5.31) 166 (6.53) 245 (9.65)
(inches) (132-227 lbs)
Dimensions in mm (inches). Weights with smallest and biggest flanges.

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
13

Remote Detector RCCS39/IR- /V2


(4.02)
(0.2) (see table 11)

(2.2.8)
(3.85)

(10.47)
(7.32)
(6.93)

(21.9)

(21.9)
(27.19)
(6.63)
(35.12)

(4.02)
(6.93)

Weight 95...149 kg (209...328 lbs)

Dimensions in mm (inches). Weights with smallest and biggest flanges

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
14

MODEL, SUFFIX AND OPTION CODES


Remote Detector RCCS3, Model and Suffix Code
Model Suffix Code Description Restrictions

RCCS34 Nominal Value : 3 t/h = 50 kg/min (110.2 lbs/min)


RCCS36 Nominal Value : 10 t/h = 170 kg/min (374.78 lbs/min)
RCCS38 Nominal Value : 32 t/h = 533 kg/min (1175 lbs/min)
RCCS39 Nominal Value : 100 t/h = 1670 kg/min (3681.72 lbs/min)
RCCS39/IR Nominal Value : 250 t/h = 4170 kg/min (9193.27 lbs/min) only with /V2
Cable conduit connection -M M20 x 1.5, female thread with cable glands
-A ANSI ½” NPT, female thread only with cable gland for detector connection
Process connection size 1)
41 ¼” see table 11
01 DN 15, ½” see table 11
23 ¾” see table 11
02 DN 25, 1” see table 11
04 DN 40, 1½” see table 11
05 DN 50, 2” see table 11
06 DN 65, 2½” see table 11
08 DN 80, 3” see table 11
10 DN 100, 4” see table 11
12 DN 125, 5” see table 11
15 DN 150, 6” see table 11
20 DN 200, 8” see table 11
Process connection rating and style 1) A1 ASME flange class 150, process connection dim. + facing acc. ASME B16.5 see table 11
A2 ASME flange class 300, process connection dim. + facing acc. ASME B16.5 see table 11
A3 ASME flange class 600, process connection dim. + facing acc. ASME B16.5 see table 11
A4 ASME flange class 900, process connection dim. + facing acc. ASME B16.5 see table 11
D2 EN flange PN 16, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D4 EN flange PN 40, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D5 EN flange PN 63, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D6 EN flange PN 100, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
D7 EN flange PN 160, process connection dim. + facing acc. EN 1092-1 Form B1 see table 11
J1 JIS flange 10K, JIS B 2220 see table 11
J2 JIS flange 20K, JIS B 2220 see table 11
G9 G female thread see table 11
T9 NPT female thread see table 11
Material of wetted parts 1)
SL Stainless steel 316L (1.4404)

1)
see selection table „Process connection and materials“ (table 11)

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
15

Remote Detector RCCS3, Option Code


Options Option Description Restrictions
code
Hazardous Area Approvals 1)
/KS1 ATEX intrinsically safe approval
/ES1 IECEx intrinsically safe approval
/US1 INMETRO intrinsically safe approval for Brazil
/NS1 NEPSI intrinsically safe approval for China
Marine Approval 2) /MC2 Marine approval acc. DNV piping class 2 (see section “Process Pressure limits”) not for thermal oil application
/MC3 Marine approval acc. DNV piping class 3 not for thermal oil application
Custody Transfer Measurement /Q20 NTEP approval 12-080A2, Accuracy class 0.3 acc. NIST Handbook 44 Only RCCS34 to 39 (Refer to GS 01R04B07-00E)
/Q21 NTEP approval 12-080A2, Accuracy class 0,3 acc. NIST Handbook 44, Heated Only RCCS39/IR ; not with /Q01, /Q20 (Refer to GS
Products 01R04B07-00E)
Tag Number /BG With customer specified tag number on name plate max. 16 digits
Flange Facing /DN Flange with safety grooves acc. to EN 1092-1 form D only for D2 to D6
/EN Flange with spigot acc. to EN 1092-1 form E only for D2 to D6
/FN Flange with recess acc. to EN 1092-1 form F only for D2 to D6
/RJ Ring Type Joint Flanges only for A3, A4
Low temperature version /LT -200°C ≤ Tmedium ≤ 150°C (-328°F to 302°F) not with /MT, in combination with Hazardous Area Approval
only with /S2
Extended temperature range /MT -70°C ≤ Tmedium ≤ 230°C (-94°F to 446°F) for RCCS34 to 39/XR; always with /S2;

Special Calibration /K2 4)


Custom 5 pts mass-/volume-flow calibration using water with factory certificate
(traceable to German national standards)
/K4 Density adjustment + thermal treatment; (accuracy: 0.001 g/cm³) only RCCS34 to 39
/K5 4) Custom 10 pts mass-/volume-flow calibration using water with DAkkS certificate
(according EN-17025:2005)
/K6 Density calibration with 3 different fluids incl. individual temperature compensation only RCCS34 to 39; not with /LT; not with /MT
with certificate (accuracy: 0.0005 g/cm³)
Certificates /P2 Certificate of compliance with the order acc. to EN 10204:2004 -2.1
/P3 Test Report acc. EN10204:2004-2.2(QIC) including the content of option /P2
/P6 Material certificate acc to EN 10204: 2004 -3.1
/P8 Pressure test report measuring system
/H1 Oil and fat free for wetted surface acc. ASTM G93-03 level C
/WP WPS acc. DIN EN ISO 15609-1 (Welding Procedure Specification) only for butt weld between process connection and
WPQR acc. DIN EN ISO 15614-1 (Welding Procedure Qualification Record) flow divider
WQC acc. DIN EN 287-1 (Welder Qualification Certificate)
/L2 Calibration certificate level 2: Declaration and the Calibration Equipment List are
issued
/L3 Calibration certificate level 3: Declaration and the Primary Standard List are issued
/L4 Calibration certificate level 4: Declaration and the Yokogawa Measuring
Instruments Control System are issued
Housing Pressure Test /J1 Rupture pressure proof test and certificate (see page 4)
Customer insulation / Heating /S2 Terminal box on extension for high or low process temperature
X-Ray Examination /RT X-ray examination of flange welding only RCCS34 with /K4, /K6 or /LT only one-sided
PMI Certificate /PM6 PMI Test on wetted parts RCCS34 to 39/IR
Dye Penetrant Test /PT Dye Penetrant Test acc. DIN EN ISO 3452-1 at the weldings of the process
connection, with certificate
Version IR /V2 Version 2 always with RCCS39/IR
Delivery to Korea /KC With KC-mark for Korea
Delivery to China /CN For meters which are delivered to China combination with Hazardous Area Approval only with /NS1
1)
Select affiliated converter RCCF31/RCCR31 with the same approval type (e.g. ATEX).
2)
Select affiliated converter RCCF31/RCCR31 with the same approval type (/MC2 or /MC3).
3)
Select affiliated converter RCCF31 with the same approval type (/Q20 or /Q21).
4)
Calibration order sheet must be delivered with the order. This is available on the Flow Center Page at Coriolis/RCCS3/Technical Information.
Volume calibration: Mass flow calibration converted by density to volume flow.

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
16

Remote field-mount Converter RCCF31, Model, Suffix and Option Code


Model Suffix Code Option Description Restrictions
Code
RCCF31 Remote field-mount converter to be connected to RCCS3;
when ordered without detector combination option /NC must be selected
Power supply -D 24 V DC
Indicator direction H2 With indicator
N0 Without indicator
Cable conduit connection M M20 x 1.5, female thread with cable glands
A ANSI ½” NPT, female thread, only cable gland for detector connection

Hazardous Area Approvals 1) /KF1 ATEX Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC
/EF1 IECEx Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC
/UF1 INMETRO Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC
/NF1 NEPSI Flame proof converter + Intrinsically safe detector output with /HP not for gas group IIC
Marine Approval 2) /MC2 Marine approval acc. DNV piping class 2 not for thermal oil application
/MC3 Marine approval acc. DNV piping class 3 not for thermal oil application
Custody Transfer Measurement /Q20 NTEP approval 12-080A2, Accuracy class 0.3 acc. NIST Handbook 44 Only in combination with RCCS34 to 39
(Refer to GS 01R04B07-00E)
/Q21 NTEP approval 12-080A2, Accuracy class 0,3 acc. NIST Handbook 44, Heated Only in combination with RCCS39/IR; not with /Q01,
Products /Q20 (Refer to GS 01R04B07-00E)
High Driving Power /HP High Driving Power, recommended for combination with RCCS36 to 39, strongly
recommended for combination with RCCS39/IR, Please see „ Hazardous Area
Specifications“
Communication /MB2 Modbus communication protocol for all parameters and values not with /Q20, /Q21, /AP, /NM, /MB3
/MB3 Modbus communication protocol for all parameters and values incl. HART not with /MB2
Communication, analogue output, pulse/status output, status input
NAMUR Switch /NM One pulse output acc. EN 60947-5-6 (NAMUR) not with /AP
Active Pulse Output /AP One active pulse output not with /NM
Tag Number /BG With customer specified tag number on name plate max. 16 characters; the last 8 significant characters are
used for HART® Tag;
characters acc. Codepage 850 (Multi-lingual)
HART® Tag Number (Software Tag) /BT1 With customer specified tag number for HART® communication in converter 22 characters for long tag;
characters acc. Codepage 850 (Multi-lingual)
No Combination /NC No combination with detector
Epoxy Coating /X1 Epoxy coating of converter housing
Delivery to Korea /KC With KC-mark for Korea
Delivery to China /CN For delivery to China combination with Hazardous Area Approval only with /NF1
Concentration Measurement 4) /CST Standard concentration measurement not with /C
/C Advanced concentration measurement, details see table „Advanced Concentration not with /CST
Measurement Options“
Certificates /P2 Certificate of compliance with the order acc. EN 10204:2004 -2.1
/P3 Test Report acc. EN10204:2004-2.2(QIC) including the content of option /P2
Instruction Manuals /IE Quantity of instruction manuals in English  = 1 to 3 selectable 5)

/ID Quantity of instruction manuals in German  = 1 to 3 selectable 5)

/IF Quantity of instruction manuals in French  = 1 to 3 selectable 5)

1)
Select affiliated RCCS3 with the same approval type (e.g. /KF with /KS1).
2)
Select affiliated RCCS3 with the same approval type (/MC2 or /MC3).
3)
Select affiliated converter RCCF31 with the same approval type (/Q20 or /Q21).
4)
For detailed information please see TI 01R04B04-04E-E. Option /K6 of RCCS3 is recommended with concentration measurement.
5)
If no instruction manual is selected, only a DVD with instruction manuals is shipped with the instrument. More than 3 manuals of one language on request available.

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
17

Remote rack-mount Converter RCCR31, Model, Suffix and Option Code


Model Suffix Code Option Description Restrictions
Code
RCCR31 Remote converter for 19” rack mounting to be connected to RCCS3
when ordered without detector combination option /NC must be selected
Power supply -D 24 V DC

Hazardous Area Approvals 1) /KS1 ATEX associated apparatus for intrinsically safe detector connection for gas group IIC with /HP not for gas group IIC
/ES1 IECEx associated apparatus for intrinsically safe detector output for gas group IIC with /HP not for gas group IIC
/US1 INMETRO associated apparatus for intrinsically safe detector output for gas group IIC with /HP not for gas group IIC
/NS1 NEPSI associated apparatus for intrinsically safe detector output for gas group IIC with /HP not for gas group IIC
Marine Approval 2) /MC2 Marine approval acc. DNV piping class 2 not for thermal oil application
/MC3 Marine approval acc. DNV piping class 3 not for thermal oil application
High Driving Power /HP High Driving Power, recommended for combination with RCCS36 to 39, strongly
recommended for combination with RCCS39/IR, Please see „ Hazardous Area
Specifications“
Active Pulse Output /AP One active pulse output not with /NM
NAMUR Switch /NM One pulse output acc. EN 60947-5-6 (NAMUR) not with /AP
Tag Number /BG With customer specified tag number on name plate max. 16 characters; the last 8 significant characters are
used for HART® Tag;
characters acc. Codepage 850 (Multi-lingual)
HART® Tag Number (Software Tag) /BT1 With customer specified tag number for HART® communication in converter 22 characters for long tag;
characters acc. Codepage 850 (Multi-lingual)
Communication /MB2 Modbus communication protocol for all parameters and values not with /AP, /NM, /MB3
/MB3 Modbus communication protocol for all parameters and values incl. HART not with /MB2
Communication, analogue output, pulse/status output, status input
No Combination /NC No combination with detector
Concentration Measurement 3) /CST Standard concentration measurement not with /C
/C Advanced concentration measurement, details see table „Advanced Concentration not with /CST
Measurement Options“
Subrack /SR2 Subrack for 2 converter RCCR31 with mounting
/SR4 Subrack for 4 converter RCCR31 with mounting
Delivery to Korea /KC With KC-mark for Korea
Delivery to China /CN For meters which are delivered to China combination with Hazardous Area Approval only with /NS1
Certificates /P2 Certificate of compliance with the order acc. EN 10204:2004 -2.1
/P3 Test Report acc. EN10204:2004-2.2(QIC) including the content of option /P2
Instruction Manuals /IE Quantity of instruction manuals in English  = 1 to 3 selectable 4)

/ID Quantity of instruction manuals in German  = 1 to 3 selectable 4)

/IF Quantity of instruction manuals in French  = 1 to 3 selectable 4)

1)
Select affiliated RCCS3 with the same approval type (e.g. /KS1 with /KS1).
2)
Select affiliated RCCS3 with the same approval type (/MC2 or /MC3).
3)
For detailed information please see TI 01R04B04-04E-E. Option /K6 of RCCS3 is recommended with concentration measurement.
4)
If no instruction manual is selected, only a DVD with instruction manuals is shipped with the instrument. More than 3 manuals of one language on request available.

Remote Cable RCCY03, Model, Suffix and Option Code

Model Suffix Code Option Description Restrictions


Code
RCCY037 Length in “meter” max. ambient temperature 90°C (176°F)
Cable ends -0 No termination, with one termination kit
Cable length L0005 Available standard length the following lengths can be ordered (e.g. 15m = L0015):
L0010 RCCY037-0: 5m; 15m; 40m; 100m; 1000m
L0040
L0100
L1000
Options:
Termination kits /TK001 Quantity of additional termination kits
/TK002
/TK003
/TK004
/TK005
/TK006
/TK007
/TK008
/TK009
/TK010
/TK015
/TK020
/TK025
/TK030
/TK035
/TK040
/TK050
/TK060
/TK070
/TK080
/TK090
/TK099
/TK150
/TK200
Delivery to China /CN For meters which are delivered to China

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
18

Advanced Concentration Measurement Options (others on request), recommended with Option /K6
Option Display Components Concentration Temp. range Source of concentration- / density table
range
/C00 --- --- --- --- Advanced concentration measurement function.
There are no pre-defined concentration setups (coefficients are set to zero).
The concentration coefficients will be set by customer making use of either
FieldMate or the stand- alone concentration tool and manual entry by HHT.
For more information, please contact your regional Yokogawa office.
/C01 °Brix Sugar / Water 0 - 85 °Brix 0 - 80°C PTB- Messages 100 5/90:
(32 - 176°F) „The density of watery Saccarose solutions after the introduction of the
international temperature scale of 1990 (ITS1990)“ Table 5
/C02 WT% NaOH / Water 2 - 50 WT% 0 - 100°C D´Ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
(32 - 212°F)
/C03 WT% KOH / Water 0 - 60 WT% 54 - 100°C D´Ans-Lax, Handbook for chemists and physicists Vol.1, 3rd edition, 1967
(129 - 212°F)
/C04 WT% NH4NO3 / Water 1 - 50 WT% 0 - 80°C Data table on request
(32 - 176°F)
/C05 WT% NH4NO3 / Water 20 - 70 WT% 20 - 100°C Data table on request
(68 - 212°F)
/C07 WT% HNO3 / Water 50 - 67 WT% 10 - 60°C Data table on request
(50 - 140°F)
/C09 WT% H2O2 / Water 30 - 75 WT% 4 - 44°C Data table on request
(39.2 - 111.2°F)
/C10 WT% Ethylene Glycol / 10 - 50 WT% -20 - 40°C Data table on request
Water (-4 - 104°F)
/C11 WT% Amylum = starch / Water 33 - 43 WT% 35 - 45°C Data table on request
(95 - 113°F)
/C12 WT% Methanol / Water 35 - 60 WT% 0 - 40°C Data table on request
(32 - 104°F)
/C20 VOL% Alcohol / Water 55 - 100 VOL% 10 - 40°C Data table on request
(50 - 104°F)
/C21 °Brix Sugar / Water 40 - 80 °Brix 75 - 100°C Data table on request
(167 - 212°F)
/C30 WT% Alcohol / Water 66 - 100 WT% 15 - 40°C Standard Copersucar 1967
(59 - 104°F)
/C37 WT% Alcohol / Water 66 - 100 WT% 10 - 40°C Brazilian Standard ABNT
(50 - 104°F)

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
19

Table 11 : Selection Table Process Connection and Materials, Installation Length in mm

RCCS39/IR
RCCS34

RCCS36

RCCS38

RCCS39
01A1 ½“-150 370 ---- ---- ---- ----
01A2 ½“-300 370 ---- ---- ---- ----
01A3 ½“-600 380 ---- ---- ---- ----
02A1 1“-150 370 500 ---- ---- ----
02A2 1“-300 370 500 ---- ---- ----
02A3 1“-600 390 520 ---- ---- ----
04A1 1½“-150 380 500 600 ---- ----
04A2 1½“-300 380 510 600 ---- ----
04A3 1½“-600 400 530 620 ---- ----
04A4 1½“-900 470 600 640 ---- ----
05A1 2“-150 ---- 510 600 ---- ----
Flanges according to ASME B16.5

05A2 2“-300 ---- 510 600 ---- ----


05A3 2“-600 ---- 540 630 ---- ----
05A4 2“-900 ---- 660 720 ---- ----
06A1 2½“-150 ---- ---- 610 ---- ----
06A2 2½“-300 ---- ---- 610 ---- ----
06A3 2½“-600 ---- ---- 640 ---- ----
06A4 2½“-900 ---- ---- 760 ---- ----
08A1 3“-150 ---- ---- 610 1000 ----
08A2 3“-300 ---- ---- 620 1000 ----
08A3 3“-600 ---- ---- 640 1000 ----
08A4 3“-900 ---- ---- 760 ---- ----
10A1 4“-150 ---- ---- ---- 1000 1100
10A2 4“-300 ---- ---- ---- 1000 1100
10A3 4“-600 ---- ---- ---- 1030 1100
12A1 5“-150 ---- ---- ---- 1000 1100
12A2 5“-300 ---- ---- ---- 1000 1100
12A3 5“-600 ---- ---- ---- 1040 1160
15A1 6“-150 ---- ---- ---- ---- 1100
15A2 6“-300 ---- ---- ---- ---- 1100
15A3 6“-600 ---- ---- ---- ---- 1200
20A1 8“-150 ---- ---- ---- ---- 1140
20A2 8“-300 ---- ---- ---- ---- 1140
01D4 DN 15 PN 40 370 ---- ---- ---- ----
01D6 DN 15 PN 100 380 ---- ---- ---- ----
02D4 DN 25 PN 40 370 500 ---- ---- ----
02D6 DN 25 PN 100 390 520 ---- ---- ----
04D4 DN 40 PN 40 370 500 600 ---- ----
04D6 DN 40 PN 100 450 560 620 ---- ----
05D4 DN 50 PN 40 ---- 500 600 ---- ----
05D5 DN 50 PN 63 ---- 520 620 ---- ----
Flanges according to EN 1092-1

05D6 DN 50 PN 100 ---- 590 660 ---- ----


05D7 DN 50 PN 160 ---- 590 660 ---- ----
08D4 DN 80 PN 40 ---- ---- 610 1000 ----
08D5 DN 80 PN 63 ---- ---- 620 1000 ----
08D6 DN 80 PN 100 ---- ---- 730 1000 ----
10D2 DN 100 PN 16 ---- ---- ---- 1000 1100
10D4 DN 100 PN 40 ---- ---- ---- 1000 1100
10D5 DN 100 PN 63 ---- ---- ---- 1000 1100
10D6 DN 100 PN 100 ---- ---- ---- 1050 1100
12D2 DN 125 PN 16 ---- ---- ---- 1000 1100
12D4 DN 125 PN 40 ---- ---- ---- 1000 1100
12D5 DN 125 PN 63 ---- ---- ---- 1000 1100
12D6 DN 125 PN 100 ---- ---- ---- 1100 1140
12D7 DN 125 PN 160 ---- ---- ---- ---- 1100
15D2 DN 150 PN 16 ---- ---- ---- ---- 1100
15D4 DN 150 PN 40 ---- ---- ---- ---- 1100
15D5 DN 150 PN 63 ---- ---- ---- ---- 1140
15D6 DN 150 PN 100 ---- ---- ---- ---- 1180

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
20

Table 11 : Selection Table Process Connection and Materials, Installation Length in mm (continued)

RCCS39/IR
RCCS34

RCCS36

RCCS38

RCCS39
01J1 DN 15 10K 370 ---- ---- ---- ----
01J2 DN 15 20K 370 ---- ---- ---- ----
02J1 DN 25 10K 370 500 ---- ---- ----
Flanges according to JIS B 2220

02J2 DN 25 20K 370 500 ---- ---- ----


04J1 DN 40 10K 370 500 600 ---- ----
04J2 DN 40 20K 370 500 600 ---- ----
05J1 DN 50 10K ---- 500 600 ---- ----
05J2 DN 50 20K ---- 500 600 ---- ----
08J1 DN 80 10K ---- ---- 600 1000 ----
08J2 DN 80 20K ---- ---- 610 1000 ----
10J1 DN 100 10K ---- ---- ---- 1000 1100
10J2 DN 100 20K ---- ---- ---- 1000 1100
12J1 DN 125 10K ---- ---- ---- 1000 1100
12J2 DN 125 20K ---- ---- ---- 1000 1100
15J1 DN 150 10K ---- ---- ---- ---- 1100
15J2 DN 150 20K ---- ---- ---- ---- 1100
41G9 G¼“ female ---- ---- ---- ---- ----
01G9 G½“ female 390 ---- ---- ---- ----
Thread

23G9 G¾“ female 390 ---- ---- ---- ----


41T9 NPT¼“ female ---- ---- ---- ---- ----
01T9 NPT½“ female 390 ---- ---- ---- ----
23T9 NPT¾“ female 390 ---- ---- ---- ----

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
21

Table 11 : Selection Table Process Connection and Materials, Installation Length in inch

RCCS39/IR
RCCS34

RCCS36

RCCS38

RCCS39
01A1 ½“-150 14.57 ---- ---- ---- ----
01A2 ½“-300 14.57 ---- ---- ---- ----
01A3 ½“-600 14.96 ---- ---- ---- ----
02A1 1“-150 14.57 19.69 ---- ---- ----
02A2 1“-300 14.57 19.69 ---- ---- ----
02A3 1“-600 15.75 20.47 ---- ---- ----
04A1 1½“-150 14.96 19.69 23.62 ---- ----
04A2 1½“-300 14.96 20.08 23.62 ---- ----
04A3 1½“-600 15.35 20.87 24.21 ---- ----
04A4 1½“-900 18.5 23.62 25.2 ---- ----
05A1 2“-150 ---- 20.08 23.62 ---- ----
Flanges according to ASME B16.5

05A2 2“-300 ---- 20.08 23.62 ---- ----


05A3 2“-600 ---- 21.26 24.8 ---- ----
05A4 2“-900 ---- 25.98 28.35 ---- ----
06A1 2½“-150 ---- ---- 24.02 ---- ----
06A2 2½“-300 ---- ---- 24.02 ---- ----
06A3 2½“-600 ---- ---- 25.2 ---- ----
06A4 2½“-900 ---- ---- 29.92 ---- ----
08A1 3“-150 ---- ---- 24.02 39.37 ----
08A2 3“-300 ---- ---- 24.21 39.37 ----
08A3 3“-600 ---- ---- 25.2 39.37 ----
08A4 3“-900 ---- ---- 29.92 ---- ----
10A1 4“-150 ---- ---- ---- 39.37 43.3
10A2 4“-300 ---- ---- ---- 39.37 43.3
10A3 4“-600 ---- ---- ---- 40.55 43.3
12A1 5“-150 ---- ---- ---- 39.37 43.3
12A2 5“-300 ---- ---- ---- 39.37 43.3
12A3 5“-600 ---- ---- ---- 40.94 45.67
15A1 6“-150 ---- ---- ---- ---- 43.3
15A2 6“-300 ---- ---- ---- ---- 43.3
15A3 6“-600 ---- ---- ---- ---- 47.24
20A1 8“-150 ---- ---- ---- ---- 44.9
20A2 8“-300 ---- ---- ---- ---- 44.9
01D4 DN 15 PN 40 14.57 ---- ---- ---- ----
01D6 DN 15 PN 10014.96 ---- ---- ---- ----
02D4 DN 25 PN 40 14.57 19.69 ---- ---- ----
02D6 DN 25 PN 10015.35 20.47 ---- ---- ----
04D4 DN 40 PN 40 15.35 19.69 23.62 ---- ----
04D6 DN 40 PN 10017.72 22.05 24.41 ---- ----
05D4 DN 50 PN 40 ---- 19.69 23.62 ---- ----
05D5 DN 50 PN 63 ---- 20.47 24.41 ---- ----
Flanges according to EN 1092-1

05D6 DN 50 PN 100 ---- 23.23 25.98 ---- ----


05D7 DN 50 PN 160 ---- 23.23 25.98 ---- ----
08D4 DN 80 PN 40 ---- ---- 24.02 39.37 ----
08D5 DN 80 PN 63 ---- ---- 24.41 39.37 ----
08D6 DN 80 PN 100 ---- ---- 28.74 39.37 ----
10D2 DN 100 PN 16 ---- ---- ---- 39.37 43.31
10D4 DN 100 PN 40 ---- ---- ---- 39.37 43.31
10D5 DN 100 PN 63 ---- ---- ---- 39.37 43.31
10D6 DN 100 PN 100 ---- ---- ---- 41.34 43.31
12D2 DN 125 PN 16 ---- ---- ---- 39.37 43.31
12D4 DN 125 PN 40 ---- ---- ---- 39.37 43.31
12D5 DN 125 PN 63 ---- ---- ---- 39.37 43.31
12D6 DN 125 PN 100 ---- ---- ---- 43.31 44.88
12D7 DN 125 PN 160 ---- ---- ---- ---- 43.31
15D2 DN 150 PN 16 ---- ---- ---- ---- 43.31
15D4 DN 150 PN 40 ---- ---- ---- ---- 43.31
15D5 DN 150 PN 63 ---- ---- ---- ---- 44.88
15D6 DN 150 PN 100 ---- ---- ---- ---- 46.46

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
22

Table 11 : Selection Table Process Connection and Materials, Installation Length in inch (continued)

RCCS39/IR
RCCS34

RCCS36

RCCS38

RCCS39
01J1 DN 15 10K 14.57 ---- ---- ---- ----
01J2 DN 15 20K 14.57 ---- ---- ---- ----
02J1 DN 25 10K 14.57 19.69 ---- ---- ----
Flanges according to JIS B 2220

02J2 DN 25 20K 14.57 19.69 ---- ---- ----


04J1 DN 40 10K 14.57 19.69 23.62 ---- ----
04J2 DN 40 20K 14.57 19.69 23.62 ---- ----
05J1 DN 50 10K ---- 19.69 23.62 ---- ----
05J2 DN 50 20K ---- 19.69 23.62 ---- ----
08J1 DN 80 10K ---- ---- 23.62 39.37 ----
08J2 DN 80 20K ---- ---- 24.02 39.37 ----
10J1 DN 100 10K ---- ---- ---- 39.37 43.31
10J2 DN 100 20K ---- ---- ---- 39.37 43.31
12J1 DN 125 10K ---- ---- ---- 39.37 43.31
12J2 DN 125 20K ---- ---- ---- 39.37 43.31
15J1 DN 150 10K ---- ---- ---- ---- 43.31
15J2 DN 150 20K ---- ---- ---- ---- 43.31
41G9 G¼“ female ---- ---- ---- ---- ----
01G9 G½“ female 15.35 ---- ---- ---- ----
Thread

23G9 G¾“ female 15.35 ---- ---- ---- ----


41T9 NPT¼“ female ---- ---- ---- ---- ----
01T9 NPT½“ female 15.35 ---- ---- ---- ----
23T9 NPT¾“ female 15.35 ---- ---- ---- ----

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa
23

Table 12 : CRN approved process connections

RCCS39/IR
RCCS34

RCCS36

RCCS38

RCCS39
01A1 ½“-150 X ---- ---- ---- ----
01A2 ½“-300 X ---- ---- ---- ----
01A3 ½“-600 X ---- ---- ---- ----
01A5 ½“-900/1500 X ---- ---- ---- ----
02A1 1“-150 X X ---- ---- ----
02A2 1“-300 X X ---- ---- ----
02A3 1“-600 X X ---- ---- ----
02A5 1“-900/1500 X X ---- ---- ----
04A1 1½“-150 X ---- X ---- ----
04A2 1½“-300 X X X ---- ----
04A3 1½“-600 X X X ---- ----
04A4 1½“-900 ---- ---- X ---- ----
04A5 1½“-900/1500 X X ---- ---- ----
Flanges according to ASME B16.5

05A1 2“-150 ---- X X ---- ----


05A2 2“-300 ---- X X ---- ----
05A3 2“-600 ---- X X ---- ----
05A4 2“-900 ---- ---- X ---- ----
05A5 2“-900/1500 ---- X ---- ---- ----
06A1 2½“-150 ---- ---- X ---- ----
06A2 2½“-300 ---- ---- X ---- ----
06A3 2½“-600 ---- ---- X ---- ----
06A4 2½“-900 ---- ---- X ---- ----
08A1 3“-150 ---- ---- X X ----
08A2 3“-300 ---- ---- X X ----
08A3 3“-600 ---- ---- X X ----
08A4 3“-900 ---- ---- X ---- ----
10A1 4“-150 ---- ---- ---- X X
10A2 4“-300 ---- ---- ---- X X
10A3 4“-600 ---- ---- ---- X X
12A1 5“-150 ---- ---- ---- X X
12A2 5“-300 ---- ---- ---- X X
12A3 5“-600 ---- ---- ---- X X
15A1 6“-150 ---- ---- ---- ---- X
15A2 6“-300 ---- ---- ---- ---- X
15A3 6“-600 ---- ---- ---- ---- X
20A1 8“-150 ---- ---- ---- ---- ----
20A2 8“-300 ---- ---- ---- ---- ----

All Rights Reserved. Copyright © 2015, Rota Yokogawa GS 01R04B10-00E-E 3rd edition July 22, 2017-00
24

REGISTERED TRADEMARKS
HART® : Registered trademark of HART Communication Foundation, Austin, TX, USA
MODBUS® : Registered trademark of MODBUS Organization

YOKOGAWA ELECTRIC CORPORATION


Headquarters
2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 JAPAN
Phone : 81-422-52-5555
Branch Sales Offices
Osaka, Nagoya, Hiroshima, Kurashiki, Fukuoka, Kitakyusyu

YOKOGAWA CORPORATION OF AMERICA


Head Office
12530 West Airport Blvd, Sugar Land, Texas 77478, USA
Phone : 1-281-340-3800 Fax : 1-281-340-3838
Georgia Office
2 Dart Road, Newnan, Georgia 30265, USA
Phone : 1-800-888-6400/ 1-770-253-7000 Fax : 1-770-254-0928

YOKOGAWA AMERICA DO SUL LTDA.


Praca Acapulco, 31 - Santo Amaro, Sáo Paulo/SP, BRAZIL, CEP-04675-190
Phone : 55-11-5681-2400 Fax : 55-11-5681-4434

YOKOGAWA EUROPE B. V.
Euroweg 2, 3825 HD Amersfoort, THE NETHERLANDS
Phone : 31-88-4641000 Fax : 31-88-4641111

YOKOGAWA ELECTRIC CIS LTD.


Grokholskiy per 13 Building 2, 4th Floor 129090, Moscow, RUSSIA
Phone : 7-495-737-7868 Fax : 7-495-737-7869

YOKOGAWA CHINA CO., LTD.


3F Tower D Cartelo Crocodile Building, No.568 West Tianshan Road,
Shanghai 200335, CHINA
Phone : 86-21-62396262 Fax : 86-21-62387866

YOKOGAWA ELECTRIC KOREA CO., LTD.


(Yokogawa B/D, Yangpyeong-dong 4-Ga), 21, Seonyu-ro 45-gil, Yeongdeungpo-gu,
Seoul, 150-866, KOREA
Phone : 82-2-2628-6000 Fax : 82-2-2628-6400

YOKOGAWA ENGINEERING ASIA PTE. LTD.


5 Bedok South Road, Singapore 469270, SINGAPORE
Phone : 65-6241-9933 Fax : 65-6241-2606

YOKOGAWA INDIA LTD.


Plot No.96, Electronic City Complex, Hosur Road, Bangalore - 560 100, INDIA
Phone : 91-80-4158-6000 Fax : 91-80-2852-1442

YOKOGAWA AUSTRALIA PTY. LTD.


Tower A, 112-118 Talavera Road, Macquarie Park NSW 2113, AUSTRALIA
Phone : 61-2-8870-1100 Fax : 61-2-8870-1111

YOKOGAWA MIDDLE EAST & AFRICA B.S.C.(C)


P.O. Box 10070, Manama, Building 577, Road 2516, Busaiteen 225, Muharraq,
Kingdom of BAHRAIN
Phone : 973-17358100 Fax : 973-17336100
Apr. '15

GS 01R04B10-00E-E 3rd edition July 22, 2017-00 All Rights Reserved. Copyright © 2015, Rota Yokogawa