FLUORIMETRY
FLUORIMETRY
PRINCIPLE: - Molecule contains electrons, electrons and nonbonding (n) electron. The
molecular orbital (HOMO).It has lest energy and more stable. When the molecules absorbs
radiant energy from a light source, the bonding electrons may be promoted to anti bonding
molecular orbital (LUMO). It has more energy and hence less stable.
The process of promotion of electrons from HOMO to LUMO with absorption of energy is called
as excitation. Singlet state:-a state in which all the electrons in a molecule are paired
Doublet state:- a state in which un paired electrons is present or Triplet state:- a state in
which unpaired electrons of same spin present Singlet excited state:- a state in which
electrons are unpaired but of opposite spin like (un paired and opposite spin)
When light of appropriate wavelength is absorbed by a molecule the electrons are promoted
from singlet ground state to singlet excited state. Once the molecule is in this excited state
relaxation can occur via several process. The process can be the following 1) Collisional
Collisional de activation: - In which entire energy lost due to collision de activation and no
electrons from excited singlet to singlet ground state with emission of light.
Phosphorescence:-At favorable condition like low temperature and absence of oxygen there is
transition from excited singlet state to triplet state which is called as inner system crossing. The
emission of radiation when electrons undergo transition from triplet state to singlet ground
is less than 0.02 A=log IoIt or A= abc Io=intensity of incident light a= absorptivity of constant b=
1.0 Since Some Energy Is Lost By Radiation less Pathways (Collisional, Intersystem Crossing,
Vibrational Relaxation)
Quenchers: - Quenching is the reduction of fluorescence intensity by the presence of
substance in the sample other than the fluorescent analyte. Quenching is following types:-
INSTRUMENTATION
Typically fluorometers utilize a double beam. These two beams work in tandem to
decrease the noise created from radiant power fluctuations. The upper beam is passed
through a filter or monochromator and passes through the sample. The lower beam is
passed through an attenuator and adjusted to try and match the fluorescent power
given off from the sample. Light from the fluorescence of the sample and the lower,
Within the machine the transducer that detects fluorescence created from the upper
beam is located a distance away from the sample and at a 90-degree angle from the
incident, upper beam. The machine is constructed like this to decrease the stray
light from the upper beam that may strike the detector. The optimal angle is 90
degrees. There are two different approaches to handling the selection of incident light
that gives way to different type’s fluorometers. If filters are used to select wavelengths
use two monochromators, some spectrofluorometers may use one filter and one
monochromator. Where, in this case, the broad band filter acts to reduce stray light,
monochromator.
Light sources for fluorimeters are often dependent on the type of sample being
tested. Among the most common light source for fluorometers is the low-
most versatile. However, this lamp is not a continuous source of radiation. The xenon
arc lamp is used when a continuous source of radiation is needed. Both of these
induces chemiluminescence.
Glass and silica cuvettes are often the vessels in which the sample is placed. Care
must be taken to not leave fingerprints or any other sort of mark on the outside of the
cuvette, because this can produce unwanted fluorescence. "Spectro grade" solvents
such as methanol are sometimes used to clean the vessel surfaces to minimize these
problems
APPLICATIONS
Detemination of phenytoin.