0% found this document useful (0 votes)
86 views7 pages

113.scope Location 1 CC-2807

The document is a calibration certificate from Wika Instruments India Pvt. Ltd. that lists the measurement capabilities and calibration methods for various electrical quantities such as voltage, current, resistance, power, and temperature. It provides the measurement ranges, frequencies, and calibration uncertainties for quantities like DC voltage, AC voltage, DC current, AC current, resistance, frequency, capacitance, power, and temperature simulations using thermocouples and RTDs. The calibration methods include using traceable equipment like Fluke calibrators and digital multimeters through direct measurement techniques.

Uploaded by

Ravichandran D
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
86 views7 pages

113.scope Location 1 CC-2807

The document is a calibration certificate from Wika Instruments India Pvt. Ltd. that lists the measurement capabilities and calibration methods for various electrical quantities such as voltage, current, resistance, power, and temperature. It provides the measurement ranges, frequencies, and calibration uncertainties for quantities like DC voltage, AC voltage, DC current, AC current, resistance, frequency, capacitance, power, and temperature simulations using thermocouples and RTDs. The calibration methods include using traceable equipment like Fluke calibrators and digital multimeters through direct measurement techniques.

Uploaded by

Ravichandran D
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as DOCX, PDF, TXT or read online on Scribd
You are on page 1/ 7

Laboratory Wika Instruments India Pvt. Ltd.

, 128, SIDCO North Phase, Ambattur


Industrial Estate, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2807 Page 1 of 7
Validity 16.08.2018 to 15.08.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

ELECTRO-TECHNICAL CALIBRATION

I. SOURCE

1. DC Voltage# 100 µV to 1 mV 1.2 % to 0.12 % Using Fluke calibrator


1 mV to 329 mV 0.12 % to 0.003 % 5522A by Direct Method
329 mV to 10 V 0.003 % to 0.002 %
10 V to 1000 V 0.002 %

2. AC Voltage# 45 Hz to 1 kHz Using Fluke calibrator


1 mV to 3 mV 0.72 % to 0.31 % 5522A by Direct Method
3 mV to 30 mV 0.31 % to 0.044 %
30 mV to 300 mV 0.044 % to 0.022 %
300 mV to 300 V 0.022 % to 0.025 %
300 V to 1000 V 0.025 % to 0.07 %

3. DC Current# 10 µA to 329 µA 0.025 % to 0.024 % Using Fluke calibrator


329 µA to 1.9 mA 0.024 % to 0.015 % 5522A by Direct Method
1.9 mA to 329 mA 0.015 % to 0.013 %
329 mA to 1.09 A 0.014 % to 0.027 %
1.09 A to 2.99 A 0.027 % to 0.046 %
2.99 A to 10.9 A 0.046 % to 0.53 %
10.9 A to 20 A 0.53 % to 0.42 %
20 A to 1000 A 0.5 % to 0.3 %

4. AC Current# 45 Hz to 1 kHz Using Fluke calibrator


33 µA to 329 µA 0.52 % to 0.22 % 5522A by Direct Method
329 µA to 3.29 mA 0.22 % to 0.13 %
3.29 mA to 2.99A 0.13 % to 0.08 %
2.99 A to 20 A 0.08 % to 0.23 %

50 Hz Using Fluke 5522A &


20 A to 1000 A 0.9 % to 0.38 % Current Coil by Direct
Method

Sangeeta Kunwar Avijit Das


Convenor Program Manager
Laboratory Wika Instruments India Pvt. Ltd., 128, SIDCO North Phase, Ambattur
Industrial Estate, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2807 Page 2 of 7
Validity 16.08.2018 to 15.08.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

6. DC Resistance# 0.1 Ω to 2 Ω 1.16 % to 0.062 % Using Fluke calibrator


2 Ω to 109 Ω 0.062 % to 0.005 % 5522A & Precision
109 Ω to 1.09 MΩ 0.005 % to 0.004 % Resistance Box by
1.09 MΩ to 10.9 MΩ 0.004 % to 0.016 % Direct Method
10.9 MΩ to 109 MΩ 0.016 % to 0.062 %
109 MΩ to 119 MΩ 0.062 % to 0.44 %
119 MΩ to 1000 MΩ 0.44 % to 1.81 %

7. Frequency# 1 Hz to 120 Hz 0.058 % to 0.0048 % Using Fluke calibrator


120 Hz to 100 kHz 0.0048 % to 0.00067 % 5522A by Direct Method
100 kHz to 2 MHz 0.00067 % to 0.029 %

8. Capacitance# 100 Hz Using Fluke calibrator


0.33 µF to 7 µF 0.66 % to 0.46 % 5522A by Direct Method
7 µF to 300 µF 0.46 % to 0.64 %

1 kHz Using Fluke calibrator


0.22 nF to 2 nF 5.83 % to 1.19 % 5522A by Direct Method
2 nF to 330 nF 1.19 % to 1.47 %

10. DC Power# 1 V to 1000 V Using Fluke calibrator


1 A to 10 A 5522A by Direct Method
1 W to 10 kW 0.11 % to 0.13 %

11. AC Power# 50 Hz Using Fluke calibrator


60 V to 240 V 2.8 % to 0.13 % 5522A by Direct Method
0.4 A to 5 A
± 0.2 to UPF
(Lead/ Lag)
0.48 W to 1200 W

12. Power Factor 1Ø# ± 0.2 to UPF 0.0034 PF to 0.0021 PF Using Fluke calibrator
(Lead/Lag) 5522A by Direct Method

Sangeeta Kunwar Avijit Das


Convenor Program Manager
Laboratory Wika Instruments India Pvt. Ltd., 128, SIDCO North Phase, Ambattur
Industrial Estate, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2807 Page 3 of 7
Validity 16.08.2018 to 15.08.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

13. Temperature Simulations#


Pt 100 (-) 200 °C to 800 °C 0.12 C Using Fluke Calibrator
Pt 1000 (-) 200 °C to 630 °C 0.12 C 5522A by Direct Method
K Type T/C (-) 200 °C to 1370 °C 0.20 C
J Type T/C (-) 210 °C to 1200 °C 0.16 C
E Type T/C (-) 250 °C to 1000 °C 0.32 C
B Type T/C 600 °C to 1800 °C 0.61 C
S Type T/C 0 to 1760 °C
0.69 C
R Type T/C 0 to 1760 °C
0.65 C
N Type T/C (-) 200 °C to 1300 °C
T Type T/C (-) 250 °C to 400 °C 0.32 C
C Type T/C 0 to 2300 °C 0.48 C
L Type T/C (-) 200 °C to 900 °C 0.52 C
U Type T/C (-) 200 °C to 600 °C 0.29 C
0.47 C

II. MEASURE

1. DC Voltage# 0.1 mV to 1 mV 1.2 % to 0.12 % Using 8½ Digit


1 mV to 1000 V 0.12 % to 0.003 % Multimeter by Direct
Method

2. AC Voltage# 100 Hz to 1 kHz Using 8½ Digit


1 mV to 190 mV 3.53 % to 0.060 % Multimeter by Direct
Method
50 Hz to 1 kHz Using 8½ Digit
190 mV to 1.9 V 0.088 % to 0.065 % Multimeter
1.9 V to 190 V 0.065 % to 0.081 %
190 V to 750 V 0.081 % to 0.12 %

4. DC Current# 1 µA to 10 µA 0.62 % to 0.098 % Using 8½ Digit


10 µA to 190 mA 0.098 % to 0.046 % Multimeter by Direct
190 mA to 1.9 A 0.046 % to 0.09 % Method

Sangeeta Kunwar Avijit Das


Convenor Program Manager
Laboratory Wika Instruments India Pvt. Ltd., 128, SIDCO North Phase, Ambattur
Industrial Estate, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2807 Page 4 of 7
Validity 16.08.2018 to 15.08.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

5. AC Current# 50 Hz to 1 kHz Using 8½ Digit


30 µA to 190 µA 0.52 % to 0.46 % Multimeter by Direct
190 µA to 19 mA 0.46 % to 0.37 % Method
19 mA to 1.9 A 0.37 % to 0.42 %

6. DC Resistance# 100 mΩ to 1 Ω 0.21 % to 0.016 % Using 8½ Digit


1 Ω to 100 Ω 0.016 % to 0.003 % Multimeter by Direct
100 Ω to 100 KΩ 0.003 % to 0.004 % Method
100 KΩ to 1 MΩ 0.004 % to 0.0082 %
1 MΩ to 100 MΩ 0.0082 % to 0.065 %

7. Frequency# 1 Hz to 100 Hz 0.036 % to 0.035 % Using 8½ Digit


100 Hz to 100 kHz 0.035 % Multimeter by Direct
100 kHz to 1 MHz 0.035 % Method
1 MHz to 5.5 MHz 0.035 %

8. Temperature Measurement# Using Fluke calibrator


Pt 100 (-) 200 °C to 800 °C 0.18 C 5522A by Direct Method
Pt 1000 (-) 200 °C to 630 °C 0.19 C
K Type T/C (-) 200 °C to 1370 °C 0.21 C
J Type T/C (-) 210 °C to 1200 °C 0.16 C
E Type T/C (-) 250 °C to 1000 °C 0.32 C
B Type T/C 600 °C to 1800 °C 0.61 C
S Type T/C 0 to 1750 °C
0.69 C
R Type T/C 0 to 1760°C
0.65 C
N Type T/C (-) 200 °C to 1300 °C
T Type T/C (-) 200 °C to 400 °C 0.32 C
C Type T/C 0 to 2300 °C 0.48 C
L Type T/C (-) 200 °C to 900 °C 0.59 C
U Type T/C (-) 200 °C to 600 °C 0.29 C
0.47 C

Sangeeta Kunwar Avijit Das


Convenor Program Manager
Laboratory Wika Instruments India Pvt. Ltd., 128, SIDCO North Phase, Ambattur
Industrial Estate, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2807 Page 5 of 7
Validity 16.08.2018 to 15.08.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

MEC
HANICAL CALIBRATION

I. PRESSURE INDICATING DEVICES

1. Pressure (Hydraulic) 6 bar to 60 bar 0.009 % of Rdg Using DH Budenberg


Digital Pressure Hydraulic Dead Wt.
Gauges/ Analog >60 bar to 2600 bar 0.009 % of Rdg Tester By Direct Method
Gauges, Pressure Based on
Transmitters, DKD-R-6-1
Pressure Switches$
>2600 bar to 4000 bar 1.72 bar Using High Pressure
Calibrator By
Comparison Method
Based on DKD-R-6-1

2. Pressure (Pneumatic) 0.1 to 1 bar 0.011 % of Rdg Using DH Budenberg


Digital Pressure >1 bar to 25 bar 0.012 % of Rdg Hydraulic Dead Wt.
Gauges/ Analog >25 bar to 100 bar 0.012 % of Rdg Tester By Direct method
Gauges, Pressure Based on DKD-R-6-1
Transmitters,
Pressure Switches$

3. Low Pressure 0 to 25 mbar 0.016 mbar Using Low Pressure


(Pneumatic) >25 mbar to 100 mbar 0.018 mbar Digital Calibrator By
Digital Pressure Comparison Method
Gauges/ Analog Based on DKD-R-6-1
Gauges, Pressure
Transmitters,
Pressure Switches$

4. Absolute Pressure 0.015 bar to 6 bar 0.0002 bar Using Absolute


(Pneumatic) Pressure Controller
Digital Pressure By Comparison method
Gauges/ Analog Based on DKD-R-6-1
Gauges, Pressure
Transmitters,

Sangeeta Kunwar Avijit Das


Convenor Program Manager
Laboratory Wika Instruments India Pvt. Ltd., 128, SIDCO North Phase, Ambattur
Industrial Estate, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2807 Page 6 of 7
Validity 16.08.2018 to 15.08.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

Pressure Switches$
5. Vacuum (-) 0.03 bar to (-)1 bar 0.012 % of Rdg Using Vacuum Dead
Digital Vacuum Weight Tester By Direct
Gauges/ Analog Method Based on
Gauges, Vacuum DKD-R-6-1
Transmitters$

6. Low Vacuum 0 to (-) 25 mbar 0.018 mbar Using Low Vacuum


Digital Vacuum >(-) 25 mbar to 0.014 mbar Digital Calibrator By
Gauges/ Analog (-) 100 mbar Comparison Method
Gauges, Vacuum Based on DKD-R-6-1
Transmitters$

7. Pressure (Hydraulic) 0 to 50 bar 0.007 bar Using Digital Pressure


Digital Pressure >50 bar to 100 bar 0.016 bar Standard By
Gauges/ Analog >100 bar to 400 bar 0.084 bar Comparison Method
Gauges, Pressure >400 bar to 700 bar 0.12 bar Based on DKD-R-6-1
Transmitters, >700 bar to 1200 bar 0.4 bar
Pressure Switches♣

8. Pressure (Pneumatic) 0 to 0.5 bar 0.00006 bar Using Digital


Digital Pressure >0.5 bar to 1.5 bar 0.0002 bar Pressure Standard
Gauges/ Analog >1.5 bar to 7 bar 0.001 bar By Comparison Method
Gauges, Pressure >7 bar to 21 bar 0.0025 bar Based on DKD-R-6-1
Transmitters,
Pressure Switches♣

9. Vacuum 0 to (-) 0.5 bar 0.00007 bar Using Digital Vacuum


Digital Vacuum > (-) 0.5 bar to (-) 1 bar 0.00015 bar Standard By
Gauges/Analog Comparison method
Gauges, Vacuum Based on DKD-R-6-1
Transmitters♣

Sangeeta Kunwar Avijit Das


Convenor Program Manager
Laboratory Wika Instruments India Pvt. Ltd., 128, SIDCO North Phase, Ambattur
Industrial Estate, Chennai, Tamil Nadu
Accreditation Standard ISO/IEC 17025: 2005
Certificate Number CC-2807 Page 7 of 7
Validity 16.08.2018 to 15.08.2020 Last Amended on -

Sl. Quantity Measured / Range/Frequency *Calibration Measurement Remarks


Instrument Capability (±)

THERMAL CALIBRATION

I. TEMPERATURE
1. RTD/ Thermocouple (-) 39 ⁰C to 400 ⁰C 0.1 ⁰C Using Dry Well Bath and
Sensors with or PRT Sensor and
Without Indicators/ Precision Thermometer
Controllers/ By Comparison Method
Recorders, Dial /
Digital Temperature
Gauges, Digital
Thermometers#

2. RTD/ Thermocouple 400 ⁰C to 650 ⁰C 0.12 ⁰C Using Dry Well Bath and
Sensors with or PRT Sensor and
without Indicators/ Precision Thermometer
Controllers/ By Comparison Method
Recorders, Dial/
Digital Temperature
Gauges, Digital
Thermometers#

3. Thermocouple 400 ⁰C to 800 ⁰C 1.1 ⁰C Using Dry Well Bath


Sensors with or and S type T/C Sensor
without Indicators/ 800 ⁰C to 1200 ⁰C 1.6 ⁰C with Process Calibrator
Controllers/ By Comparison Method
Recorders, Dial/
Digital Temperature
Gauges, Digital
Thermometers#
* Measurement Capability is expressed as an uncertainty () at a confidence probability of 95%
$
Only in Permanent Laboratory

Only for Site Calibration
#
The laboratory is also capable for site calibration however, the uncertainty at site depends on the
prevailing actual environmental conditions and master equipment used.

Sangeeta Kunwar Avijit Das


Convenor Program Manager

You might also like