ANALIZADOR EDAN H50 User Manual PDF
ANALIZADOR EDAN H50 User Manual PDF
I
A Table of Contents
Chapter 8 QC ................................................................................................................................ - 26 -
Chapter 9 Calibration ................................................................................................................. - 29 -
-3-
H50/H51 Hematology Analyzer User Manual Safety Guidance
The reliability of the analyzer and the safety of operators are considered during product
design and production. The following safety and preventive measures should be carried
out:
WARNING
Safety Warnings
-1-
H50/H51 Hematology Analyzer User Manual Safety Guidance
WARNING
9. Connect the analyzer to a grounded socket and make certain that the power supply
meets the requirements specified in the user manual.
10. Do not exceed the maximum permitted load when using multiple portable socket-
outlets to supply the system.
11. SHOCK HAZARD - Do not attempt to connect or disconnect a power cord with wet
hands. Make certain that your hands are clean and dry before touching a power cord.
12. If you have any questions about the power adaptor or the power cord prior to using,
consult EDAN or its authorized distributors for service.
13. The analyzer is not waterproof. Do not use it in locations where water or any liquid
leakage may occur.
14. To avoid the possibility of electrostatic shock and damage to the system, avoid using
aerosol spray cleansers on the analyzer screen.
15. EXPLOSION HAZARD – The analyzer is not suitable for use in the presence of a
flammable anesthetic mixture with oxygen or other flammable compounds.
16. To avoid electrical shock, never use the system in altitude exceeding 3,000 meters
above sea level.
17. LASER RADIATION – The bar code scanner is a laser product. Do not view beam
generated by the scanner with eyes directly.
18. Periodically have the safety of the analyzer checked by a qualified service engineer.
19. Only accessories supplied or recommended by the manufacturer should be used.
Otherwise, performance and electric shock protection can‘t be guaranteed.
20. Blood samples should be collected according to proper medical guidelines which
contains collection details, such as selection of blood collection site, collection
procedures, sample handling, etc. Sterile techniques should be followed to prevent
the blood collection site from being contaminated.
21. Handle blood samples and collection devices with care, and wear approved
protective gloves to avoid direct contact with samples.
22. To avoid electrical shock and damage to the system, turn off the analyzer and
disconnect the analyzer from the AC power source before cleaning and disinfecting.
23. To avoid being injured, never touch the sampling needle directly with hands.
24. The system is for in vitro diagnostic use only.
25. Perform quality control (QC) tests to make sure that the system works smoothly.
26. The sample collection tube should only be used a single time.
27. The sample is contained in the sample tubes, so sample tubes need to be
disposed of as bio-hazardous waste, complying with local regulatory guidelines.
-2-
H50/H51 Hematology Analyzer User Manual Safety Guidance
WARNING
28. The analyzer needs to be disposed of according to local regulations after their useful
lives. Alternatively, they can be returned to the dealer or the manufacturer for
recycling or proper disposal. Please delete or back up safety data contained in the
analyzer before disposing of it.
29. Misuse of electrical equipment may cause electrocution, burns, fire and other
hazards.
30. Due to the use and the environment of this equipment, all parts and surfaces of
H50/H51 Analyzer are potentially infective. Wearing gloves when operating or
maintaining the analyzer and hands washing after work are essential.
31. Only reagents including Lyse, Cleaner, Diluent, Bleach, Hematology Controls and
Hematology Calibrator provided by EDAN or its authorized distributors should be
used. If other reagents are used, Edan will not assure the accuracy of test results
and will not be responsible for the analyzer’s maintenance and repair caused by the
reagents.
32. Don't position the analyzer in some places where it's difficult to disconnect the AC
plug.
CAUTION
1. Do not use the analyzer in a dusty environment with bad ventilation or in the
presence of corrosives.
2. Make sure that the time and date of the system are correct.
3. The system is intended to analyze whole blood samples. Never use it to analyze
serum or plasma.
4. If there are clots or bubbles in the blood sample, discard it and collect samples again.
5. Transport, store and use the analyzer, diluent, lyse, cleaner, Hematology Controls
and Hematology Calibrator according to the corresponding user manual.
6. Connect the analyzer with the peripherals recommended by EDAN.
7. Maintain the system as described in the user manual to avoid potential damage.
8. Do not clean the analyzer and accessories with abrasive fabric.
9. Do not immerse the analyzer into liquid under any circumstances.
10. Extreme humidity may affect test results. A relative humidity greater than 80% may
cause inaccurate results.
11. Use this system at a temperature between 18 °C and 32 °C. Outside this range, the
system may produce inaccurate results.
12. The cable tie must be removed before the analyzer is firstly turned on after
completion of its first installation. If not, the analyzer’s operation might be affected or
even worse the analyzer might be damaged.
-3-
H50/H51 Hematology Analyzer User Manual Safety Guidance
CAUTION
13. Make sure that there is no intense electromagnetic interference source around the
analyzer, such as radio transmitters, mobile phones etc. Attention: large medical
electrical equipment such as electrosurgical equipment, radiological equipment and
magnetic resonance imaging equipment is likely to bring electromagnetic
interference.
3. Recycle
4. On/Off button
5. Network port
6. Serial port
9. Disposal Method
10. Manufacturer
-4-
H50/H51 Hematology Analyzer User Manual Safety Guidance
18. Warning
(Background: yellow Symbol and outline: black)
21. Caution
-5-
H50/H51 Hematology Analyzer User Manual Measurement Principle
-6-
H50/H51 Hematology Analyzer User Manual Measurement Principle
-7-
H50/H51 Hematology Analyzer User Manual Measurement Principle
1
3
-8-
H50/H51 Hematology Analyzer User Manual Measurement Principle
4 USB Ports
TCP/IP Port
Not Used
P/S 24V
-9-
H50/H51 Hematology Analyzer User Manual Measurement Principle
NOTE:
If the power cord provided with the instrument needs to be replaced, the new cord must
be in compliance with the local regulation and the instrument specification in matter of
consumption.
2.2.6 Printer
H50/H51 Analyzer is not delivered with the printer. The instrument is equipped already with
PCL3 & PCL6 standard printer drivers which cover a large selection of printer models, such as
HP Laserjet Professional P1106.
- 10 -
H50/H51 Hematology Analyzer User Manual Measurement Principle
2.3 Configuration
2.3.2 Options
x 1 USB Printer
x 1 USB Keyboard
NOTE:
The following table lists the differences between H50 Hematology Analyzer and H51
Hematology Analyzer.
- 11 -
H50/H51 Hematology Analyzer User Manual Measurement Principle
- 12 -
H50/H51 Hematology Analyzer User Manual Measurement Principle
WBC 5 differential percentages are obtained by optic measurements. The measured pulses on the
two optical channels are displayed on DIF Plot ALL (Y axis) and FSC (X axis). Each dot on the
DIF Plot represents the height in ALL and FSC of at least one pulse.
- 13 -
H50/H51 Hematology Analyzer User Manual Measurement Principle
- 14 -
H50/H51 Hematology Analyzer User Manual Measurement Principle
7) Abnormal lymphocyte absolute value (ALY#) is made from ALY% and lymphocyte absolute
value by formula below:
ALY# = ALY% x LYM#
8) IMM% is the number of elements in the IMM population area divided by sum of elements
in DIF scattergram including IMM
9) Immature monocyte absolute value (IMM#) is made from IMM% and WBC total counts by
formula below:
IMM# = IMM% x WBC (total counts)
Where HCT is expressed in %, RBC is expressed in 1012/L and MCV is expressed in fL.
ʹሻ Mean Corpuscular Hemoglobin (MCH) calculation is obtained from HGB and RBC by the
following formula
ୌୋ
MCH =
ୖେ
Where MCH is expressed in pg, HGB is expressed in g/L and RBC is expressed in 10 12/L.
3) Mean Corpuscular Hemoglobin Concentration (MCHC) is made from HGB and HCT by the
formula below:
ୌୋ୶ଵ
MCHC =
ୌେ
Where MCHC is expressed in g/L, HGB is expressed in g/L and HCT is expressed in %.
4) RDW CV parameter is derived from the RBC histogram.
5) RDW SD parameter is calculated on curve distribution (defined as curve width at 20% of
peak)
Where PCT is expressed in %, PLT is expressed in 10 9/L and MPV is expressed in fL.
- 15 -
H50/H51 Hematology Analyzer User Manual Measurement Principle
2) PLCR (Platelet Large Cell Ratio) indicates the percentage of large platelets with a
volume >12 fL. Aside from the two flexible discriminators which delimit the volume
distribution curve, there is additionally a fixed discriminator at 12 fL (marked in red on
below picture). The PLCR is the percentage of cells higher than 12fL regarding the whole
platelets count.
PLCR is significantly decreased in patients with thrombocytosis while it is increased in
thrombocytopenia. In patients with high counts, PLCR is significantly decreased in reactive
thrombocytosis than neoplastic thrombocytosis. PLCR is more likely higher in destructive
thrombocytopenia than hypo-proliferative thrombocytopenia. PLCR is inversely related to
platelet count and directly related to PDW and MPV.
3) Large platelets cell count (PLCC) is made from PLCR and PLT by formula below:
PLCC = PLCR x PLT
Where PLCC is expressed in 109/L, PLCR is expressed in % and PLT is expressed in 10 9/L.
4) PDW means the size distribution spread of the platelets population derived from the PLT
histogram.
- 16 -
H50/H51 Hematology Analyzer User Manual Installation Guidance
4.2 Unpacking
NOTE:
If any part is missing, contact your distributor immediately.
- 17 -
H50/H51 Hematology Analyzer User Manual Installation Guidance
4.3 Installation
Installation of the device must be done according to the procedures described below.
1. Remove the instrument from the cardboard and place it on a stable table.
2. Remove the accessory boxes from the reagent compartment and unpack them.
- 18 -
H50/H51 Hematology Analyzer User Manual Installation Guidance
3. Connect one end of the diluent pickup tubing to the white diluent connector located at the
back of the instrument and connect another end to the diluent container.
NOTE:
Diluent container must be placed on the same level as the instrument.
4. Connect one end of the waste pickup tubing to the red waste connector located at the back of
the instrument and put another end into a waste container.
NOTE:
Waste container must be placed on the floor.
5. Tighten the cap of the Lyse pickup tubing (Yellow color sleeve) and the cap of the Cleaner
pickup tubing (Blue sleeve) to the corresponding bottles. Install lyse and cleaner bottles in the
reagent compartment of the instrument.
- 19 -
H50/H51 Hematology Analyzer User Manual Installation Guidance
6. Connect one end of the power supply cable to the power supply adapter and connect the other
end of the power supply cable to the instrument. The connection way is shown below.
7. Plug the main cord to the main power supply. 100-240 VAC; 50/60 Hz; Single phase with
ground
NOTE:
In order to assure data safety contained in the analyzer, the user can change the
password (123456) provided by the manufacture after first login in accordance with the
procedures described in Section 13.5.6 User Management Setting.
- 20 -
H50/H51 Hematology Analyzer User Manual Reagents Management
PREVIOUS: Allow to return to the previous screen except when the system in under the
Main Screen.
TOOLS: Depending on the screen, you can access to some of the following options.
MENU: Allow to return to Main Screen except when the system in under the Main Screen.
ABOUT: Allow accessing to the following screen to know about the analyzer’s information.
Move to the top of the table Move to the left of the table
Move up in the table step by step Move to the left step by step
Move down in the table step by step Move to the right of the table
- 21 -
H50/H51 Hematology Analyzer User Manual Reagents Management
z RESULTS z QC SCREEN
z REAGENTS z REPEATABILITY
z SERVICE z CALIBRATION
Users can access into different sub-screens from the main screen by pressing the corresponding
buttons. Each sub-screen has the detailed introduction in its corresponding chapter.
Sub-screens Chapter
RUN SAMPLE Chapter 11
RESULTS Chapter 11
REAGENTS Chapter 6
SERVICE Chapter 13
STARTUP Chapter 7
QC SCREEN Chapter 8
REPEATABILITY Chapter 10
CALIBRATION Chapter 9
LOG OUT Chapter 14
SHUT DOWN Chapter 14
- 22 -
H50/H51 Hematology Analyzer User Manual Reagents Management
Screen 5-1 Screen 5-2
- 23 -
H50/H51 Hematology Analyzer User Manual Reagents Management
Press Cycles Counter on Screen 5-1 to Press Reagents Log on Screen 5-1 to enter
enter into the Screen. Operator and into the Screen. Users can view reagents
Administrator can view information on this information on this screen.
screen but only Technician is allowed to
perform Reset operation.
.
2) Waste Drainage: If the customer uses the
waste drainage instead waste container to
treat the waste, directly select Waste Open
Drain.
Screen 5-3
NOTE:
If select Waste Open Drain, Waste Capacity and Reset will be removed from the
screen.
- 24 -
H50/H51 Hematology Analyzer User Manual Startup
Chapter 7 Startup
Press to turn on the analyzer and the
system will automatically do STARTUP. After
Startup, the system will enter into Login
Screen. After login, the system directly enters
into Run Sample Screen to display the
STARTUP results.
- 25 -
H50/H51 Hematology Analyzer User Manual QC
Chapter 8 QC
To control the stability of the instrument before running patient samples, it is mandatory to run
QC test at the beginning of the working day. If the QC test results are out of control, it is
recommended to perform a calibration, then run QC test again.
NOTE:
Take safety measures when working with biological samples, such as wearing
approved gloves, etc.
Only controls provided by EDAN or its authorized distributors can be used. If not, Edan
is not responsible for any adverse consequences such as test results’ accuracy,
analyzer’s troubleshooting, maintenance and repair, etc.
Press QC to enter into Screen 7-1. Press Changes to enter into Screen 7-2. Press Results to enter
into Screen 7-3.
“CHANGE” button allows creating a new QC
lot or re-editing the existent QC information.
But all its associated results will be deleted if
you re-edit the existent QC information.
Screen 7-2
- 26 -
H50/H51 Hematology Analyzer User Manual QC
Screen 7-3
QC results can be viewed in Table mode as displayed in Screen 7-3. But the results can also be
viewed in Details mode by pressing “DETAILS” to enter into Screen 7-4 or in LJ Graphs mode
by pressing “LJ GRAPHS” to enter into Screen 7-5.
Screen 7-4
High Limit
Target Target
Low Limit
Test result Low Limit Yellow Line
NOTE:
The position of yellow line represents a QC test.
Screen 7-5 Therefore the test result varies as the yellow
line’s position changes.
- 27 -
H50/H51 Hematology Analyzer User Manual QC
The system will indicate whether the result of each parameter is outside the limits with H/L. If
the results of some parameters are outside the acceptable ranges, check the following items:
z Refer to the user manual to confirm that the test procedures are correct.
z Controls are stored properly and don’t expire.
z The system passes the STARTUP.
z If the QC test fails, conduct sample test with the sample which must have been tested
yesterday after the analyzer’s QC test succeeds and then observe whether there are some
differences between the test results obtained today and the results obtained yesterday. If
there is no difference, please contact Edan’s service engineer. If there is difference, it
means the analyzer might need to be calibrated. Please do the calibration and then re-
conduct QC test. If the QC test still fails, please contact Edan’s service engineer.
z If the QC test fails, re-conduct QC test with the previous lot of controls. Contact Edan’s
service engineer if the QC test with the previous lot of controls succeeds. Calibrate the
analyzer if the QC test with the previous lot of controls still fails. Re-conduct QC test
after calibration. If the QC test still fails after the analyzer’s calibration, please contact
Edan’s service engineer.
z It is recommended to perform instrument calibration if at least one of parameters is out
of limits from a significant degree.
NOTE:
1) Concerning Controls’ usage, storage, handling and other details, please strictly
respect its user manual.
2) Perform QC tests every day before sample analysis. If not, all results will be
displayed and printed with the indication “QC not done”.
3) If QC test fails, all results will be displayed the indication “QC FAILED” unless a new
QC test succeeds.
4) The analyzer can keep 100 results (results and histograms) per lot for 12 different
lots.
5) The column allows selecting or unselecting a result in the table. Mean, SD and
CV are only decided by the selected results ( ) in the table. Automatic calculations
will be implemented accordingly.
- 28 -
H50/H51 Hematology Analyzer User Manual Calibration
Chapter 9 Calibration
The H50/H51 Hematology Analyzer has been calibrated prior to delivery. However, errors may
still be introduced to the system during transport or installation. To maintain the accuracy and
precision of this analyzer, Edan recommends that QC test needs to be conducted prior to being
used for sample analysis for the first time and after every repair and maintenance of the analyzer.
If the QC test fails, which is not caused by the improper operation or the controls, the analyzer
needs to be calibrated.
NOTE:
Only the calibration materials provided by EDAN or its authorized distributors should be
used.
Press Calibration to enter into Screen 8-1. Press Change to enter into Screen 8-2. Press Results
to enter into Screen 8-3.
“CHANGE” button allows editing new calibration
information or re-editing the existent calibration
information. But all its associated results will be
deleted if you re-edit the existent calibration
information.
- 29 -
H50/H51 Hematology Analyzer User Manual Calibration
Screen 8-4
Press on to restart the calibration
NOTE:
1. 3 calibration tests at least and 12 calibration tests at most are required to be
conducted successively.
2. The column allows selecting or unselecting a result in the table. Mean, SD and
CV are only decided by the selected results ( ) in the table. Automatic calculations
will be implemented accordingly.
3. Concerning calibration materials’ usage, storage, handling and other details, please
strictly respect its user manual.
4. If the room temperature changes more than 10°C during the working day, the
instrument must be calibrated more frequently.
- 30 -
H50/H51 Hematology Analyzer User Manual Repeatability
Chapter 10 Repeatability
Press Repeatability to enter into Screen 9-1. Press Details to enter into Screen 9-2.
Under the Screen 9-1, the sampling needle will
move down to aspirate the prepared materials
by pushing the start key and then the system
will automatically finish the whole analysis
and display the test results on the screen.
Screen 9-2
NOTE:
1) 3 repeatability tests at least and 50 repeatability tests at most are required to be
conducted successively.
2) The column allows selecting or unselecting a result in the table. Mean, SD and
CV are only decided by the selected results ( ) in the table. Automatic calculations
will be implemented accordingly.
- 31 -
H50/H51 Hematology Analyzer User Manual Sample Test
it needs to be stored at 2-8ć for no more than 24 hours. If the sample has been stored at 2-8ȭ, it
Screen 10-1
Under the Screen 10-1, the sampling needle will move down for aspirating the prepared sample.
The system will finish the whole sample test automatically and finally display test results on the
screen.
A mode button is located on the Screen 10-1.
Press the mode button, the system will pop
out Screen 10-2.
Screen 10-3
The operator can edit the following information in Screen 10-3.
1. NAME: Patient name (1-20 characters)
2. PID: Patient identification (1-16 characters)
3. SID: Sample identification(1-16 characters)
4. Birth Date: patient’s birth date
5. Doctor: Doctor ID or doctor’s name (1-16 characters)
6. Comments: Add comments if needed
7. Gender: F(female)/M(male)/U(unknown)
8. Group: Reference Range
9. Mode: DIF WB, DIF PD, CBC WB, CBC PD, DIF CAP, CBC CAP
NOTE:
1) WB stands for whole blood and PD stands for pre-dilution. If PD mode is selected,
the button “DIL DISPENSE” will appear on the Screen 10-1. Press the button and
the system will dispense 300μL dilution via the sampling needle in to the prepared
tube.
2) If the sampling needle is not down in sampling position, press on start key and
needle will go down in sampling position
Start key
- 33 -
H50/H51 Hematology Analyzer User Manual Sample Test
3) Two thirds of the sampling needle shall be within the blood sample to ensure the
successful sample aspiration and then press on start key.
two-thirds of
the needle Blood volume
inside
4) During sample aspiration, ON/Off button will blink. The system will beep after
completion of sample aspiration, then the sampling needle will move up and the
sample tube can be removed.
5) As soon as the cycle is launched, the SID is incremented automatically.
3) Press TABLE VIEW in Screen 10-6, you will be viewed the selected test results in Table
mode.
to save changes.
- 35 -
H50/H51 Hematology Analyzer User Manual Test Results
Chapter 12 Flags
Results flags are
displayed behind the
test values.
++++ The measured result is above the measurement range. Result is replaced by ++++
h or l: result is colored in
The test result is above or below the corresponding
light orange
H/h/L/l “Reference Ranges” as the Reference Ranges Screen
H or L: result is colored in
in Section 12.5.1.5 Reference Ranges shows.
red
- 36 -
H50/H51 Hematology Analyzer User Manual Test Results
Table 11-1
Detailed information in
System Alerts Triggers
pop-out window
Background test results don’t meet
S-UP Failed Startup Failed
background requirements.
Means that the Instrument temperature is
INS-T Instrument temperature out
lower than 18°C or higher than 32°C and the
of range
test results may be affected.
QC Not Done During the
QC NOT DONE QC tests are not done during the day.
Day
- 37 -
H50/H51 Hematology Analyzer User Manual Test Results
Press SERVICE/SETTINGS/LAB
PARAMETERS/CBC THRESHOLDS
AND FLAGS and the system will enter
into Screen11-1 as the left screen shows.
Morphologic alarms will be affected if the
thresholds displayed in the left screen are
changed. Therefore, only technical
persons are recommended to edit values
on this screen.
Screen 11-1
- 38 -
H50/H51 Hematology Analyzer User Manual Test Results
Counts
fL
1) The alarm WBC - L1 will be triggered if this population at the left of L1 is higher than an
absolute limit AND a percentage of the LYM.
2) The alarm WBC – L5 will be triggered if this population at the right of L5 is higher than an
absolute limit AND a percentage of the WBC.
NOTE:
9 First peak on the left side of the histogram represents the lymphocytes cells; the
other one located on the right side represents all the others WBC populations.
9 Only technical persons are recommended to edit thresholds (L1 and L5) and flags in
Screen 11-1 because any change might trigger or affect the alarm WBC - L1 and
WBC – L5.
CR1 CR2
fL
1) The alarm RBC - Micro will be triggered if this population at the left of CR1 is higher than
an absolute limit OR a percentage of the RBC.
2) The alarm RBC - Macro will be triggered if this population at the right of CR2 is higher than
- 39 -
H50/H51 Hematology Analyzer User Manual Test Results
Counts
fL
1) The alarm PLT - Debris will be triggered if this population at the left of CP1 is higher than
an absolute limit OR a percentage of the PLT.
2) The alarm PLT - Schizo will be triggered if this population between CP2 and P is higher than
an absolute limit OR a percentage of the PLT.
3) The alarm PLT - Micro will be triggered if this population defined by CP3-2 width at the
right of CP3 is higher than an absolute limit AND a percentage of the PLT.
NOTE:
Only technical persons are recommended to edit thresholds (P, PL, CP1, CP2, CP3 and CP3-2)
and flags in Screen 11-1 because any change might trigger or affect the alarm PLT – Debris,
PLT - Schizo and PLT - Micro.
- 40 -
H50/H51 Hematology Analyzer User Manual Test Results
Press SERVICE/SETTINGS/LAB
PARAMETERS/DIF FLAGS and the
system will enter into Screen11-2 as the left
screen shows. The following morphologic
alarms will be affected if the values
displayed in the left screen are changed.
Therefore, only technical persons are
recommended to edit values on this screen.
Screen11-2
DIF: Lymphocytes
- 42 -
H50/H51 Hematology Analyzer User Manual Test Results
- 43 -
H50/H51 Hematology Analyzer User Manual Service
Chapter 13 Service
Press Service on the main screen to enter into the Screen 12-1.
Screen 12-1
Press System Init ( ) and the system will automatically perform mechanical
initialization.
- 44 -
H50/H51 Hematology Analyzer User Manual Service
Screen 12-2
NOTE:
- 46 -
H50/H51 Hematology Analyzer User Manual Service
13.5.1.2 Units
Press Units to enter into Units Screen as the following two screens show. There are seven units
system. Users can select one unit system from the dropdown menu in accordance with actual
needs.
NOTE:
The difference between SI and SI MODE is that “g” is used in SI unit system while
“mmol” is used in SI MODE unit system.
CBC Thresholds and Flags Screen Please refer to Section 12.3 in this manual for
more details.
- 47 -
H50/H51 Hematology Analyzer User Manual Service
press to save.
- 48 -
H50/H51 Hematology Analyzer User Manual Service
NOTE:
NOTE:
Only technical persons are allowed to change the calibration factors for any changes
can affect the analyzer’s performance.
- 49 -
H50/H51 Hematology Analyzer User Manual Service
Date/Time Screen
Auto Clean: If the operator sets the auto clean frequency (analysis runs number), the system will
launch clean operation at the designated frequency. The auto clean frequency ranges from 10 to
5000 and the default frequency is 100. The operator can change the default value in accordance
with actual needs.
Auto Shutdown: If the operator enables this function, the system will shut down after the
designated idle duration. The idle duration ranges from 30 minutes to 720 minutes and the default
idle duration is 180 minutes. The operator can change the default value in accordance with actual
needs.
NOTE: mn means minute.
- 50 -
H50/H51 Hematology Analyzer User Manual Service
3) Click “Not Installed” and Screen 12-4 will displayed. Press and Screen 12-5 will pop
out.
4) Select the targeted printer’s name and Press to finish the printer’s management.
- 51 -
H50/H51 Hematology Analyzer User Manual Service
Printer Settings
Screen12-8 Screen12-9
With the two screens, the operator can do the communication setting to transmit analysis results,
QC results and calibration test results to some data management systems such as hospital
LIS/HIS system. Please operate in accordance with the following procedures:
1) Do host settings in Screen 12-9:
Enter Address and Port of host which is intended to connect with the H50 analyzer.
2) Do analyzer settings in Screen 12-9:
Dis-select DHCP, manually enter address, mask and gateway. Alternatively, select DHCP and the
analyzer will automatically obtain address, mask and gateway and display them in the screen.
NOTE:
If None is selected, the communication between H50 analyzer and data
management system will fail. Therefore only Serial or Ethernet can be selected if
you want the communication between H50 analyzer and data management system
succeeds.
5) Select format (CSV Format, HL7 Format) and the default communication one is CSV Format.
6) Select which kind of test results (analysis result, QC result or Calibration report) that you
want to automatically transmit from H50 analyzer to the connected data management system.
NOTE:
Both analysis result and QC result can be transmitted from H50 analyzer to the
connected data management system via CSV format or HL7 format, while
Calibration report can only be transmitted from H50 analyzer to the connected data
management system via CSV format
7) Press in Screen 12-8 to save the settings made by the operator and the system will
automatically transmit the selected test results (analysis result, QC result or Calibration
report) to the connected data management system after test.
- 53 -
H50/H51 Hematology Analyzer User Manual Service
Add Login ID
Screen 12-10
Change Password
Screen 12-11
Remove Login ID
- 54 -
H50/H51 Hematology Analyzer User Manual Service
13.6 Troubleshooting
Troubleshooting Screen
- 56 -
H50/H51 Hematology Analyzer User Manual Service
3) When the following screen pops out, open the fluidic door with screw driver and add 4mL of
bleach (Sodium Hypo chloride solution) into RBC bath and WBC bath respectively.
NOTE͵
The system will beep when it requires the user to pour the bleach into the baths
4) Close fluidic door and Press on the pop-out. The system will automatically perform
bleach cleaning.
13.6.1.2 Instructions for Performing Syringe Greasing
Wear rubber glove on one Spread a thin film of With a tool key type T20,
hand and place a bit of grease on each piston. turn the two waste pistons in
silicon grease at the tip of order to spread the grease all
the index. around the pistons.
- 57 -
H50/H51 Hematology Analyzer User Manual Service
- 58 -
H50/H51 Hematology Analyzer User Manual Service
Problems Solutions
Diluent leaks around the 1) Check the rinsing needle block (presence of clots) and clean
it if necessary
needle during sample tests
2) Check the rinsing head.
1) Check the power supply connection wires.
Fail to turn on the analyzer
2) Check the Power Supply block
1) Check whether diluent container is placed at the same level
with the analyzer and whether the diluent tubing is not
All test results are bad.
pinched.
2) Check the diluent’s remaining volume.
No test results are displayed Check the screen wires which are connected to CPU board.
1) Check the paper.
No printing
2) Check the electrical connection.
- 59 -
H50/H51 Hematology Analyzer User Manual Service
13.7 Maintenance
Screen 12-13
Screen 12-14
- 60 -
H50/H51 Hematology Analyzer User Manual Service
Quarterly
None None
(3900 cycles)
Biannually
None None
(7800 cycles)
Annually Needle O-ring
None
(15 600 cycles) Piston greasing
Three years
None Syringes block and pistons exchange
(46800 cycles)
Follow proper safety procedures and take protective measures such as wearing approved
gloves when cleaning and disinfecting the system.
Clean and disinfect the exterior surfaces to remove dust, splatters, etc. Policies regarding the
cleaning and disinfecting intervals are at the discretion of your individual institution.
3) Disconnect the connecting cables if the system is connected to other pieces of equipment.
NOTE:
9 Ethanol (75%)
9 Isopropanol (70%)
- 61 -
H50/H51 Hematology Analyzer User Manual Service
6) Clean the surfaces with the wet cloth to disinfect them after the surfaces are completely dry.
NOTE:
9 Make sure the surfaces are thoroughly dry before disinfecting them.
7) When the surfaces are thoroughly dry, reconnect the power cord, power adaptor and other
connecting cables.
- 62 -
H50/H51 Hematology Analyzer User Manual Logout and Shutdown
14.2 Shutdown
- 63 -
H50/H51 Hematology Analyzer User Manual Warranty and Service
Alternatively, you can send an email to EDAN service department at: [email protected].
- 64 -
H50/H51 Hematology Analyzer User Manual Specifications
Appendix 1 Specifications
A1.1 Environment Requirements
Temperature 18 – 32 °C
Usage
H50/H51 Humidity ʇ80%
Hematology
Analyzer Transport and Temperature -10°C to 50°C
Storage Humidity ʇ80%
- 65 -
H50/H51 Hematology Analyzer User Manual Analytical Performance
A2.2 Background
A2.3 Precision
Parameter Range Precision
WBC 3.5×109 / L ~6.0u 109 / L ≤3.0%
6.1u 109 / L ~15.0×109 / L ≤2.5%
RBC 3.5×1012/ L -6.5×1012/ L ≤1.5%
HGB 110 g/L -180 g/L ≤1.5%
PLT 100×109 / L -500×109 / L ≤4.0%
MCV 70 fL -120 fL ≤1.0%
HCT 35%-55% ≤ 2.0%
A2.4 Carry-over
Parameter Carry-over
WBC ʇ0.5%
RBC ʇ0.5%
HGB ʇ0.6%
PLT ʇ1%
- 66 -
H50/H51 Hematology Analyzer User Manual Analytical Performance
A2.5 Accuracy
Parameter Detection Range Limit (%)
WBC 3.5×109 / L ~9.5× 109 / L ±6
RBC 3.8× 1012/ L ~5.8× 1012/ L ±4.0
HGB 110 g/L ~180 g/L ±3.0
PLT 125×109 / L ~350×109 / L ±13.0
MCV 82 fL ~100 fL ±6.0
HCT 35% ~55% ±6.0
- 67 -
H50/H51 Hematology Analyzer User Manual EMC Information
The system is intended for use in the electromagnetic environment specified below. The user of
the system should ensure that it is used in such an environment.
RF emissions
Class A
CISPR 11
Harmonic
emissions Not Applicable The system is suitable for use in all establishments
IEC 61000-3-2 other than domestic and those directly connected to the
public low-voltage power supply network that supplies
buildings used for domestic purposes.
Voltage
fluctuations/ flicker
Not Applicable
emissions
IEC 61000-3-3
- 68 -
H50/H51 Hematology Analyzer User Manual EMC Information
69
H50/H51 Hematology Analyzer User Manual EMC Information
NOTE 1 UT is the a.c. mains voltage prior to application of the test level.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
NOTE 3: Over the frequency range 150kHz to 80 MHz, field strengths should be less than 3 V/m.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
can not be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the system is used exceeds the applicable RF compliance level
above, the system should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the system.
70
H50/H51 Hematology Analyzer User Manual Safety Specifications
Altitude 3000m
- 71 -
H50/H51 Hematology Analyzer User Manual Glossary
Appendix 5 Glossary
Abbreviation Full Name
WBC Leucocytes Total Count
LYM%/# Lymphocyte percent and absolute value
MON%/# Monocyte percent and absolute value
NEU%/# Neutrophil percent and absolute value
EOS%/# Eosinophil percent and absolute value
BAS%/# Basophils percent and absolute value
ALY%/#* Abnormal Lymphocyte percent and absolute value
IMM%/#* Immature Monocyte percent and absolute value
RBC Red Blood Cell Total count
HGB Hemoglobin
HCT Hematocrit
MCV Mean Cell Volume
MCH Mean Cell Hemoglobin
MCHC Mean Cell Hemoglobin Concentration
RDW-CV Red Blood Cells Distribution Width-CV
RDW-SD Red Blood Cell Distribution Width-SD
PLT Platelets Total count
MPV Mean Platelet Volume
PCT Plateletcrit
PDW Platelets Distribution Width
PLCR Large Platelets Cell Ratio
PLCC Large Platelets Cell Count
RUO Research Use Only
SID Sample Identification
PID Patient Identification
PD Pre-dilution
WB Whole Blood
CAP Capillary
DIFF Differentiation
- 72 -
H50/H51 Hematology Analyzer User Manual User Access
NOTE:
1) ĸ means you have the right to enter into the screen to read its information and
perform operations.
2) X means you can only enter into the screen to read its information.
3) NO means you have no rights to enter into the screen.
- 74 -
- 75 -