Model ZR202G---A

Drawings Integrated type


Zirconia Oxygen / High Temperature
Humidity Analyzer
 Reference gas/Except for Pressure compensation (-C, -E)
Unit : mm
48.5

Terminal box side


Display side
Ø123

170
122 125

Flange

L 338 to 351
t
C
Rc1/4 or 1/4NPT
Reference gas inlet

ØA
Ø 50.8

25

ØB

Rc1/4 or 1/4NPT
Calibration gas inlet
Flange
153 to 164
4-G1/2,2-1/2NPT etc.
252 to 265 49 Cable connection port

Unless otherwise specified,differences in the dimensions are specified as: General tolerance = 6 (Criteria of tolerance class it18 in JIS BO401-1986) - 2.

All Rights Reserved, Copyright © 2000, Yokogawa Electric Corporation.


Subject to change without notice.
SD 11M12B02-01E 1/3
1st Edition : Oct. 2000 (YK)
4th Edition : Feb. 2015 (YK)

P-
2

 Reference gas / with Pressure compensation (-P) Unit : mm

48.5

Terminal box side


Display side
Ø123
Ø50.8

170
Flange 122 125

L 342±4
t
C
Rc1/4or1/4NPT
Reference gas inlet

ØA
25
33

*1 ØB
Reference gas outlet
PIPING
:B Flange
PIPING:A 2-G1/2,2-1/1NPT etc.
156±3 Stop Cable connection port *1
256±4 Valve 49 Ensure that the flange gasket does not block
Rc 1/4or1/4NPT the reference gas outlet on the detector flange.
Calibration gas inlet If the flange gasket blocks the outlet,
the detector cannot conduct pressure compensation.
Where necessary, make a notch on the flange gasket.

Flange Weight
Model, Code L PIPING
Specification A B C t (kg)
ZR202G-040--A 400 Approx. 8
ZR202G-070--A 700 Approx. 9
ZR202G-100--A 1000 Approx. 10
ANSI CLASS 150 152.4 120.6 4-Ø19 19 A
ZR202G-150--A 1500 -2-RF Approx. 12
ZR202G-200--A 2000 Approx. 14
ZR202G-250--A 2500 Approx. 16
ZR202G-300--A 3000 Approx. 17
F3.18E.EPS

 HOOD (Option / H)

274 ±4 150 ±3
150 ±3

Hood Material : Aluminum Hood Weight : Approx. 800g

Unless otherwise specified,differences in the dimensions are specified as: General tolerance = 6 (Criteria of tolerance class it18 in JIS BO401-1986) - 2.

SD 11M12B02-01E 2/3
1st Edition : Oct. 2000 (YK)
P- 4th Edition : Feb. 2015 (YK)
3

• Wiring Diagram
Integrated type Zirconia Oxygen/High Temperature Humidity Analyzer
Contact output 1 Contact output 2
Contact input 1

Contact input 2

1 2 3 4 5 6 7
DI-1 DI-2 DI-C DO-1 DO-1 DO-2 DO-2

8 9 10 11 12 13 14
FG AO AO L N G FG
(+) (-)

Analog output 100 to 240 V AC,


4-20 mA DC 50 or 60 Hz
Digital output
F28.EPS

The protective grounding for the analyzer shall be connected either the protective
ground terminal in the equipment or the ground terminal on the case.
Standard regarding grounding: Ground to earth, ground resistance: 100Ω or less.

Unit : mm
• Specified by Option Code

Detector case Detector case

Approx. 44

Approx. 100

40 Ø48

19
Full open height
Calibration gas inlet 58
Rc1/4 or 1/4NPT Calibration gas inlet Rc1/4 or 1/4NPT

 With stop valve (option : /SV) weight : Approx. 200g  Check valve (option : /CV) weight : Approx. 90g

Unless otherwise specified,differences in the dimensions are specified as: General tolerance = 6 (Criteria of tolerance class it18 in JIS BO401-1986) - 2.

SD 11M12B02-01E 3/3