This document provides information about oil-free air compressors, including: 1) It describes the components and features of DH-B series air compressors, including that they are oil-free, have an internal air dryer and filter, over-voltage protection, and electronic drain valve. 2) It provides technical specifications for three models (DH-3000B, DH-6000B, DH-9000B) and lists included components and spare parts. 3) It outlines installation, operation, and safety instructions, noting the equipment should be in a clean, well-ventilated environment and not operated before cooling water is connected. Proper maintenance is also discussed.

Uploaded by

Yusuf Hungund
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
222 views15 pages

DH-B Serials Air Compressor PDF

This document provides information about oil-free air compressors, including: 1) It describes the components and features of DH-B series air compressors, including that they are oil-free, have an internal air dryer and filter, over-voltage protection, and electronic drain valve. 2) It provides technical specifications for three models (DH-3000B, DH-6000B, DH-9000B) and lists included components and spare parts. 3) It outlines installation, operation, and safety instructions, noting the equipment should be in a clean, well-ventilated environment and not operated before cooling water is connected. Proper maintenance is also discussed.

Uploaded by

Yusuf Hungund
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 15

AIR COMPRESSOR

DH-B serials Air compressor


User Manual
□3000B □6000B □9000B
Cabinet air-cooled Air compressor
1.Oil-free
2. Internal Air Dryer
3. Over-Voltage Protection
4. Internal Delicate Filter
5. Internal Electronic Drain Valve

Note:Equipment should be installed in clean environment with good ventilation!!


To prevent backflow water, gas-liquid mixing connection should be installed strictly according to this guide
The interval between shutdown and startup should be more than 5seconds!
keep equipments in clean and dry environment for a longer lifespan
It is prohibited to operate the equipment before connecting with cooling water
VersionNo.:20150707

Before installation, be sure to read the instructions


-1-
AIR COMPRESSOR
Catalogue
Preface………………………………………………………………………………………………………...………………….........3
Safety Notice…………………………………………………..………………………………………………………..……………4
Chapter 1 Outline …………………………….…………………………………………………………………...………………….5
1.1 Brief Introduction……………………………………….………………………………………………..…………………….5
1.2 Product Feature……..………………………………………………………………………………………...……………….5
Chapter 2 Product mode & Technical parameter………………………………………………………………..…………....6
Chapter 3 Control panel and interface function Explanation…….……………………………………………...…….……7
3.1 Control panel component function Explanation……………………………………...………………………..……………7
3.2 Side interface Explanation……………………………………………………………………………………..……………..7
Chapter 4 Installation and operation Notice...................................................................................................................9
4.1 Installation field requirements……………..………………………………….…………………………………..………….9
4.2 Piping installation ………………………………………………………………..…………………………………..………..9
4.3 Electrical installation…………………………………………………………………………………………………………..9
4.4 Power switch on off operation..…………………………………………………………………………………..…………..9
4.5 Note for Use…………………………………………………………………..……….…………………………...…………..9
Chapter 5 Maintenance&trouble shooting...................................................................................................................10
5.1 Warranty terms…………………………………………………………..…………….…..……………………………..….10
5.2 Maintenance Items…………………………………………………………………….………………………….………….11
5.3 Maintenance Items Period……………………..……………………………………..…………………………..…………11
5.4 Failure Analysis and Trouble Shooting………………………………………………………………………..……...……12
Installation Connection Schematic Diagram……………………….………………………..……….……………………..…13
Electric Circuit Diagram………………………………………………….……………………..….…................................……14

-2-
AIR COMPRESSOR
Preface
Thank you for using DH-B series oil-free air compressor. The products have been strictly tested before delivery from our
factory. To ensure the best use of the equipment, please read through the user manual before operation.

The user manual is applicable as follows:


The user manual is for our DH-B series products. Please keep it properly and make it accessible

at all time。

The User Manual Usage:


It is used as the instruction for operation personnel
Knowledge required for understanding the user manual:
You will have a better understanding of the user manual if you have mechanical and electrical
background, and will qualify for the equipment maintenance. It can be used for the project designer to select air
compressor and design .

-3-
AIR COMPRESSOR
Safety notes
Qualified operation personnel
Operation personnel must have systematic training.
1. The cable of air compressor power supply should match the power of compressor motor, and the motor protection
circuit breaker should be installed, and make sure there is reliable grounding device to protect the equipment
2. Before startup, make sure nobody else is around the machine, and no tools or other article
left in it.
3. Compressed air and the compressor can be dangerous. When repairing it or making
maintenance, make sure it is powered off and place a warning sign of ‘ under maintenance’
or ‘ no switching on’
4. For the first time startup or cable connection has been changed, make sure the motor rotary
direction is the same as the arrow direction.
5. The output pressure must not be higher than the specified pressure on the label. Otherwise,
the air compressor will be overloaded and damaged.
6. If the machine is damaged or has potential danger, the machine must not be turned on.
Make sure the power is cut off, and put a distinct marking.
7. Finger and clothing should be away from the rotating fan, drive coupling and the hot
components, for example, oil-gas separator, piping, etc. When the machine is running, it is
prohibited to dismantle screws or components. The machine’ s high temperature liquid or
compressed air will cause serious injury, or even death.
8 .Before making maintenance, make sure the machine is shut down; compressed air has been
released; the compressor has been disconnected with other gas source; and compressor unit
has been cooled down.
9. When cleaning the compressor components, only use non-corrosive solvent. It is prohibited
to use flammable, explosive and volatile solvent.
10. Compressor unit assembly is only for highly qualified professional personnel with
professional tools. Customer must not dismantle or repair compressor unit. If the
compressor unit is damaged, please contact our after sale customer service.AIR COMPRESSOR
11. The components must be from original manufacturer.
13. The machine must not be used in enclosed room or underground tunnel. The compressed
air must not be used for breathing purpose.

-4-
AIR COMPRESSOR
Chapter I Product Outline
1.1、Product outline:

The product utilize high quality, high flow swinging piston compressor as power source, providing stable oil-free air
source which avoids contaminated oil getting into human body, or damaging the machine. The components are all of high
quality, and the complete air compressor is designed to match the oxygen generator: high air flow, low noise level, dry and
clean gas source, stable operation and automatic control. When the internal pressure of the air cylinder reach the lower
limit or upper limit, the air compressor will start or shutdown automatically. The product is suitable for the air source of
oxygen generator or air feed ozone generator.

1.2、DH-B serials Product Features:

⑴、Internal air dryer, delicate filter, electronic drain valve for dry and clean air source

⑵、Output gas is oil-free, dry and clean, so oil removal processing is not required. The output

gas can be used in food, pharmaceutical, medical, etc.。

⑶、Low noise level, the noise level is about half of the one from traditional piston compressor.。

⑷、Each compressor is equipped with cooling fan, ensuring the continuous operation of the

equipment.

⑸、Output pressure automatic control. The compressor stops working when the output pressure reaches its upper

limit 4kg; and will resume to work when the pressure is lower than 0.2kg.

⑹、Internal air dryer and delicate filter have electronic drain valve it can automatically drain water every 20 minutes.

1.3、Components and spare parts list

Item name Material quantity Usage area Usage


User Manual 1
4“ male coupling Cooper 1
4“ female ball valve Cooper 1 Air outlet valve
Air outlet pipeline
12-04 joint Cooper 1
12*8 PU tube PU 2 Meter Air output pipe
8-04 joint Cooper 1 Drainage outlet
8*6 PU Tube PU 2 Meter pipeline Drainage pipe

-5-
AIR COMPRESSOR
Chapter 2 Product Model and Technical Specification
2.1、Product Model and Technical Specification:

Model
DH-3000B DH-6000B DH-9000B
Item
3 3 3
Exhaust gas capability 0.2Nm /min 0.4Nm /min 0.6Nm /min
Exhaust gas pressure <0.4Mpa

Exhaust gas temperature <35℃

Power supply 220V/50Hz 220V/50Hz 3*380V/50Hz


Power 2.2KW 3.5KW 4.5KW
Outlet size 1/2“inner thread
Auto-drainage outlet size 1/2“inner thread

Ambient temperature <37℃

Ambient humidity <70%

Connection cable size


2 3*1.0 3*2.5 5*1.5
(mm )

Dimension(mm) 580*500*1300 680*680*1300 600*900*1300

Net weight(kg) 65 108 122

-6-
AIR COMPRESSOR
Chapter 3 Control Panel and Interface Function Explanation

3.1、Control panel components function explanation:

⑴、②:gas source temperature gauge:display gas source temperature, normal temperature rang15~25℃

⑵、③:Power switch: turn on/off the machine.

⑶、④:gas source over- temperature alarm light:when the gas source temperature is over 30 ℃, the alarm light

is on

⑷、⑤:The over pressure indicator light: when the output pressure is over 4kg/cm2, the light is on, and Main compressor

head is stopped working; When this failure happens, the output pipe need check; if it is blocked, and the valve of gas
pipe is closed; and if the oxygen generator or ozone generator are switched on to work.
⑸、⑥:Power indicator: the indicator light is on after powered.

⑹、⑦:Output pressure meter: internal air receiver pressure display, the normal output pressure range is 1.5-3.0 kg/cm2.

3.2、 Machine side interface Explanation:

⑴、①:Automatic water drainage outlet: internal air dryer and delicate filter water drainage outlet for draining water every

20 minutes.
⑵、⑧:Power interface: specified electrical power supply

⑶、⑨:Air outlet: output dry and clean and compressed air.

(4)、⑩:Over-load switch: when the internal compressor is over loaded, the over-load switch can provide electricity-cutting

protection; When it is under this protection, please check if the voltage of power supply is too low, and the startup
pressure is set up too high( the default pressure is lower than 0.2kg to start)

DH-3000Bcontrol panel and joint schematic Diagram

-7-
AIR COMPRESSOR

DH-6000B Control Panel and interface schematic diagram

DH-6000B air compressor operation notice:

1、There are 2 heads inside of DH-6000B, there are 2 switch for heads on the control panel, the two heads can

work at the same time and also can work independently(DH-6000B is usually working with two sets of ozone generator, if
there is only one set ozone generator working, then only one head need powered on inside of DH-6000. If the two heads
are working together under this situation, there will be over-pressure alarm.
2、Before start the DH-6000B, please select the working compressor head(start the compressor switch 1 or 2 or both),
make sure the pressure meter value is lower than 0.2kg/cm2, then press the start-up switch of the compressor. It is
forbidden to start up the other head in the running of one head, or the head cannot be started normally, then caused the
over load protection or damage

DH-9000Bcontrol panel and interface schematic diagram

-8-
AIR COMPRESSOR
Chapter 4 Installation and operation notice
4.1 Installation Field requirement:
(1)Equipment should be installed indoor, with clean, adequate light and good air circulation
environment. It is prohibited to be installed in the environment of exposure to sunlight,
rain, water vapor, moisture, corrosive gas, explosive gas and dust
(2)Environment temperature should be lower than 40 ℃ , humidity lower than 70 ℃
(preferably, install air conditioner for cooling and drying). Environment temperature too
high may lead to equipment cease and lower ozone generator production. High
humidity may lead to controlling circuit malfunction or damaged.
(3) The distance between equipment and wall, ceiling should be more than 600mm, which is
convenient for future maintenance.
(4) Equipment should be installed on solid, horizontal ground. Installing on uneven ground
may lead to a high level of equipment noise and vibration
4.2 Piping installation
⑴ Output piping should have 1°-2° angle of inclination which will be good for draining out
of the water in the piping.
⑵ Output piping should use less elbows or valves, in order to avoid pressure loss.

⑶ Output piping size should not be less than outlet size (0.5 inch), to avoid pressure loss。

4.2、Pipeline installation:

⑴.Air outlet pipeline: included 12*8 PU tubes and joints, when install it, please make reference the attached pipeline
installation schematic diagram; if need the high level installation, user can use DN20 PVC pipe (or metal pipe) for
connection.
⑵.Drainage outlet pipeline: included 8*6 PU tube and joint,when install it, please make reference the attached
pipeline installation schematic diagram; if need the high level installation, user can use DN20 PVC pipe (or metal
pipe) for connection.
4.3 Electrical connection
..(1) The connection cable size should not be less than specified in the specification, and when
…..cable length is more than 10 meters, a bigger size should be used.
.(2) The power source should meet the rated voltage and frequency requirement.
.(3) The voltage fluctuation cannot be too high, and it should be within the range of
…-10%~+10% of the nominal voltage. It cannot be used if voltage is out of range,
….otherwise it will cause equipment damage, and voltage stabilizer or regulator should be
….used to meet the voltage range
⑷ Voltage unbalance rate of the three phases cannot be too high, and should be within 5%.
….Stabilizer should be used if there is high fluctuation.
4.4、Switch on/off procedure:

⑴ Check before startup: check whether the piping is blocked, whether the component is
…..abnormal, whether the voltage is matched.
⑵ Startup Operation: open the air outlet valve, firstly power on the oxygen generator, ozone generator, then
press .the .panel power switch air compressor, air compressor starts to work, gradually increasing output pressure,
working pressure should be in the range of 1.5~2.5kg/cm2.
⑶.Shut down: Press the power switch and the air compressor will stop working。
4.5 Operation Notice
⑴ The machine is oil-free air compressor. Do not use lubricating oil!

⑵ Please check one time every day if the drainage outlet is draining water, the drainage outlet to drain water every 20
minutes.
⑶ Before start the DH-6000B, please select the working compressor head(start the compressor switch 1 or 2 or both),
make sure the pressure meter value is lower than 0.2kg/cm2, then press the start-up switch of the compressor. It is
forbidden to start up the other head in the running of one head, or the head cannot be started normally, then caused the
over load protection or damage

-9-
AIR COMPRESSOR
Chapter 5 Maintenance and trouble shooting
5.1、Warranty terms:

Dear customer:
Thank you for choosing BNP products. To ensure you enjoy the perfect after sale service from
BNP, please read through this manual carefully after receiving the product, and keep it properly.

BNP after sale service commitment:

1. All products (not including consumable material, accessories and the L series products) from
BNP are warranted for one year and lifetime maintenance from the date of purchasing.
2. The following occasions are excluded from the warranty, and BNP may charge for the
repairing materials.
A. Product damages due to high humidity working environment or Oxygen concentrator flooded by water;
B. Product damages due to unauthorized operation unless instructed by BNP or BNP authorized distributor;
C. Product damages due to natural disaster, or any other force majeure;
D. Fail to provide proof of purchase or warranty card.
3. If on-site service is required, customer will be required to pay for the cost of travelling, and
BNP or distributor can decide to come or not according to the distance.
4. In the case of sterilization standard is not met due to improper operation or
malfunction of the product, BNP does not take any responsibility.

----------------------------------------------------------------------------------------------------------

Warranty Card

User name Purchase date

Contact person Contact phone

Address Postcode

Product name Serial No.

Customer Distributor

(Signature) (Signature)

Please fill the blank then send the card to our company.

After receiving this form, we will confirm to you and the warranty is valid

- 10 -
AIR COMPRESSOR
5.2 Maintenance items
⑴、Fan Dust Filter screen cleaning /replacement: As there is dust in the air, the intake filter screen will needs regular
cleaning/replacement (refer to below maintenance cycle). Before cleaning, make sure the machine is power off. Dismantle
the filter screen, then dismantle the cover and
take out the screen. Clean the screen with clean water, and install it after it is dry., the irregular cleaning will cause the
problem of bad radiant.
⑵、The intake air filter element cleaning/Replacement: in the proceeding of use, there is dust in the air, the intake air
filter element need regular cleaning /replacement(refer to below maintenance cycle). Before cleaning, make sure the
machine is powered off and Unscrew the air filter element , then remove the cover of air filter element, and take out the
filter element, then clean it with clean water, then dry it and install it (replace new filter element is acceptable too)
⑶、Electronic drainage valve cleaning/replacement: in the bottom of delicate filter, the internal filter screen of the
electronic drainage valve need clean or replace regularly, (refer to below maintenance cycle) Before cleaning, make sure
the machine is powered off and release the air in the air receiver (pressure is 0). Unscrew the screw cap on the side of the
solenoid valve, take out the filtering net, and clean it with clean water, then install the filtering screen again。

⑷、Delicate filter element replacement: in the bottom of delicate filter, the internal filter screen of the electronic drainage
valve need clean or replace regularly, (refer to below maintenance cycle) Before cleaning, make sure the machine is
powered off and release the air in the air receiver (pressure is 0). Unscrew the screw cap on the side of the solenoid valve,
take out the filtering net, and clean it with clean water, then install the filtering screen again。


5.3、Maintenance cycle table(Tab 5-1)
Maintenance cycle table
Item content Half Two Remark
daily weekly Monthly Quarterly yearly
yearly yearly
Fan filter Clean Regular Clean
○ ●
screen replacement、 replacement
Head inlet Clean Regular Clean
○ ●
filter element replacement replacement
Regular clean the
Drainage Clean
○ ● drainage valve
solenoid valve replacement
interior filter screen
Check the air
compressor
Check the
draingae outlet
auto-drainage Check ○
working well or not,
working status
every 20 minutes
once
Delicate filter
replacement ● At most 2years
element
Note:
(1)、○:means inspection, adjust or clean;●:means replacement

(2)、The above table is the suggested maintenance/ replacement cycle from manufacturer and
customer can also conduct the maintenance/replacement according the actual operation
environment and intensity of work

- 11 -
AIR COMPRESSOR
5.4、Trouble analysis and inspection (table 5-2):
Serial
Failure Analysis Solution
no.
⑴ Power plug/ power cabinet (1) Check power source.
1 No power (2) Switch on the internal
no power
breaker
(2) Internal breaker switch off
⑴ Air outlet ball valve is not open。 ⑴ open the ball valve。

⑵ Ozone generator and oxygen generator ⑵ Start the ozone and oxygen

2 Over-pressure alarm are not started generator, then start the air
⑶ lower pressure alarm。 compressor.
⑶ readjust the pressure switch or
replaceit
If the problem still exists after the above inspection, please contact our company or local
distributor.

- 12 -
AIR COMPRESSOR

Equipment Installation Connection Schematic Diagram(DH-B Serials)

- 13 -
AIR COMPRESSOR

Electric Circuit Diagram( DH-3000/6000B)


1 2 3 4

D
DH-6000B空压机电气原理图(内含冷干机+过滤器)-160530 D

4.0 1
L
4.0
N 0.5
4.0 2 电子排水阀2/L 8
PE
电子排水阀1/L 7 45 0 抽风机/L
QF1
4.0 8# 网络 6 空压机风扇*4/L
22 0V 电源 40 A
QF2 5
3*2.5 平方
5A 电子排水阀1/N 4 11 00F温度控制器/N
L1

L2

运行灯/ 压力高/ 温度高灯/N 3 45 0 抽风机/N


3 KM1-A2 2 空压机风扇*4/N
KM1 2# 网络 1 电子排水阀2/N

L3
32 A

4
T1

T2

C 11 00F控制点 KM1 C
电源开关
自锁

T3
3
4

1
压力开关
5.0-4.0

3
5 6 7 8

过载开关*2
10 A

A1
1.5 1.5 1.0
N L

N L

N L

N L

N L
KM1
B B
L

L
N

A2

空压机1 空压机2 冷干机 11 00F温度控制器 运行灯/ 绿 压力高/ 警报灯 温度高/ 警报灯 电子排水阀1 电子排水阀2 15 0 风扇*4 45 0 抽风机
1.5KW 1.5KW 60 0W 20 分钟排2 秒 20 分钟排2 秒
20 分钟排2 秒

1 9
24 V- mA+
2 10
24 V+ V+
11
R TD+ PT10 0 探头/+ 棕色/N
3 至主图45 0 抽风机
NO 黑色/L
至主图11 00F控制点 4
COM
5
NC
蓝色
7 12
N TC + PT10 0 探头/- 黄/ 绿
至主图11 00F温度控制器 8 AL1 设置为37 13
L TC -
AH1 设置为0 Title
A 4E-45 0 风机 A
11 00F温度控制器
LOC 设置为0
AL1 设置为35 Size Number R evisio n
AH1 设置为0
A4
SDIS设置为1
Date: 19 -Ju l-20 16 Sh eet of
File: E:\ZY\protel 文件\ 电气原理图\pro 版电气原理图
Drawn\ 在用文件
B y: \ 空压机\DH-60 00B 空压机电气原理图
1 2 3 4

- 14 -
AIR COMPRESSOR

Electric Circuit Diagram( DH-9000B)


1 2 3 4

D
DH-9000B空压机电气原理图(内含冷干机+过滤器)-160530 D

2.5
L1
2.5
L2
2.5
L3
2.5
N
2.5 1
PE
电子排水阀2/L 9 45 0 抽风机2/L
0.5
2 电子排水阀1/L 8 45 0 抽风机1/L
3*38 0V 电源
1.5 1.5 1.5 8# 网络 7 空压机风扇*6/L
5*2.5 平方 QF1 25 A
6
L1

L2

L3
QF2 电子排水阀2/N 5
5A 电子排水阀1/N 4 11 00F温度控制器/N
KM1 运行灯/ 压力高/ 温度高灯/N 3 45 0 抽风机2/N
25 A 3 KM1-A2 2 45 0 抽风机1/N
2# 网络 1 空压机风扇*6/N

L3
T1

T2

T3

4
C C

11 00F控制点 KM1
电源开关
自锁

T3
3
4
1.5+1.0

1
1.5

1.5

1.5

压力开关
4.0-3.5

3
过载开关*3 5 6 7 8
10 A
1.0

B A1 B
N L

N L

N L

N L

N L

N L

N L

N L

N L
KM1
A2

15 00 空压机*3 冷干机 11 00F温度控制器 运行灯/ 绿 压力高/ 警报灯 温度高/ 警报灯 电子排水阀1 电子排水阀2 15 0 风扇*6 45 0 抽风机*2
20 分钟排1 秒 20 分钟排2 秒 20 分钟排2 秒

1 9
24 V- mA+
2 10
24 V+ V+
11
R TD+ PT10 0 探头/+ 棕色/N
3 至主图45 0 抽风机
NO 黑色/L
至主图11 00F控制点 4
C OM
5
NC
蓝色
7 12
N TC + PT10 0 探头/- 黄/ 绿
至主图11 00F温度控制器 8 AL1 设置为37 13
L TC -
AH1 设置为0 Title
A 45 0 风机 A
11 00F温度控制器
LOC 设置为0
AL1 设置为35 Size Number R evisio n
AH1 设置为0
A4
SDIS设置为1
Date: 19 -Ju l-20 16 Sh eet of
File: E:\ZY\protel 文件\ 电气原理图\pro 版电气原理图
Drawn\ 在用文件
B y: \ 空压机\DH-90 00B 空压机电气原理图
1 2 3 4

- 15 -