CLX - XT (Model 28030 Manual (Rev 1.5)
CLX - XT (Model 28030 Manual (Rev 1.5)
HF scientific
3170 Metro Parkway
Ft. Myers, FL 33916
Phone: 239-337-2116
Toll Free: 888-203-7248
Fax: 239-332-7643
EMail:[email protected]
Website: www.hfscientific.com
Importer’s Name:
Importer’s Address:
I, the undersigned, hereby declare that the equipment specified above conforms to the
above Directive and Standard
Specifications....................................................................................................1
1.0 Overview...........................................................................................................2
1.1 Unpacking and Inspection of the Instrument and Accessories ...................2
1.2 The Display .................................................................................................3
1.3 The Touch Pad ............................................................................................3
1.4 Orientation ..................................................................................................4
2.0 Safety.................................................................................................................4
2.1 Symbols Used In CLX-XT....................................................................5
Section Page
9.0 Troubleshooting.............................................................................................24
9.1 CLX-XT Fault Detection.....................................................................24
9.2 Setting Flow Rate.................................................................................25
9.3 Clearing Faults.....................................................................................25
9.4 Reagent Clogs .....................................................................................25
9.5 Diagnostic Chart .................................................................................26
9.6 Technical and Customer Assistance ...................................................26
Instruction Manual 1
Mounting Kit 1
Remove the instrument from the packing carton. Carefully inspect all items to ensure that
no visible damage has occurred during shipment. If the items received do not match the
order, please immediately contact the local distributor or the HF scientific Customer
Service Department.
Front View
Bottom View
2.0 Safety
This manual contains basic instructions that must be followed during the commissioning,
operation, care and maintenance of the instrument. The safety protection provided by this
equipment may be impaired if it is commissioned and/or used in a manner not described in
this manual. Consequently, all responsible personnel must read this manual prior to
working with this instrument.
In certain instances “Notes”, or helpful hints, have been highlighted to give further
clarification to the instructions. Refer to the Table of Contents to easily find specific
topics and to learn about unfamiliar terms.
The measurement chamber consists of a sample inlet, a purge drain, and an overflow. The
reagent is added from the check valves integrated into the lower portion. A green LED
provides the 515 nm source lamp, a red LED is used for sample level and flow
measurement. A single detector is located 180 º from the green LED. A replaceable glass
cuvette separates the LEDs from the detector and maintains the measurement path length.
Sample water flows in the inlet at the bottom, through the measurement cuvette and out
through an overflow drain. This flow is used to both fill the cuvette and flush the system.
The reagents are dispensed from two replaceable bottles. One bottle has a buffer to control
the pH; the second has an indicator that contains the DPD, which produces color when
chlorine is present in the sample. The degree of color is dependant on the amount of
chlorine in the sample water.
The measurement chamber is open to view operations. A white LED backlights the
chamber for a clearer view. The white LED will flash to attract attention in the case of a
warning or failure. Most warnings and failures are also displayed on the screen. To
prevent interference, the white LED is turned off during measurements.
During normal operation the CLX-XT will run through a timed cycle. A simplified cycle
will consist of the following sequences:
Flushing – continuous sample flow
Purging – PURGE valve opens
Zeroing – no flow with cuvette full
Adding Reagents – one pulse of the reagent pump
Mixing with sample – sample flow pulses in
Reading resulting sample – no flow with cuvette full
Purging – PURGE valve opens to remove reacted sample
The cycle above is simplified and does not describe all the actions and testing that occurs.
The CPU continuously diagnoses the entire system for correct operation and sample water
flow. If an error occurs, a message is posted to the message queue on the LCD screen.
The reagent is added by a single pulse of the reagent solenoid. When the reagents require
replacement, the PRIME button is pushed to bring new reagents into the system. During
PRIME the reagent solenoid is pulsed several times to draw fluid from the two reagent
bottles and fill the tubes with new reagent. A complete PRIME takes less than a minute.
The SERVICE button empties the cuvette, stops the flow of sample water, and clears any
errors. This provides a convenient way to replace reagents and the measurement cuvette. If
more extensive servicing is performed, all power to the CLX-XT should be removed. If
the CLX-XT is to be turned off, it is recommended that the instrument be placed in
SERVICE mode before removing power. This ensures that the cuvette is emptied and the
flow is off.
To control this condition the CLX-XT has a built-in cooling chamber for the indicator
reagent. The temperature of this chamber is kept below 24˚C. This cooling chamber will
cycle similar to a normal refrigerator to keep the indicator reagent at the desired
temperature. This temperature is factory set and should not be adjusted.
To allow for this feature to operate correctly, this cooling chamber must be constantly
powered.
On the right side of the enclosure is a fan. This is required for proper operation of the
cooling chamber and must never be blocked.
The buffer reagent is required to control the sample pH to bring the sample to neutral.
The reagents are provided already hydrated and require little operator preparation.
The indicator reagents are packaged a plastic box. Each 125 mL indicator will require the
addition and mixing of the DPD powder contained in the small brown bottle. This
activates the indicator reagent. The indicator reagents should be changed out on a 90 day
basis even if they are not fully used.
The 500 ml buffer does not require any preparation and is good for one year.
After one year all the reagents will have been consumed and replacement reagent set will
be required. This can be ordered through agent Catalog No.09991. This set will include a
new plastic box with four sets of Indicator reagents plus one 500 mL bottle of Buffer.
POWER IN
INSTRUMENT
17.140 [435.36]
SIGNAL OUT
4-20MA 15.130 [384.30]
RS-485/Modbus
SIGNAL OUT
ALARMS
INSTRUMENT DRAIN
INSTRUMENT WATER INLET EXHAUST AIR
CHILLER
INTAKE AIR
CHILLER
ENCLOSURE VENT
POWER IN
CHILLER
14.305 [363.34]
It is critical that the instrument be mounted as close as possible to the sampling point to
ensure a quick response time (within 2-3 meters (6-10 ft) of the sampling point).
The provided mounting feet will need to be installed with the provided screws. These can
be rotated as shown above. Suggested mounting screws are up to M6 (¼”).
The instrument is equipped with an internal cabinet drain (vent) to prevent damage in the
event of a tubing failure.
The main sample drain tubing connects to a hose barb. The rated tubing size is ½” ID
tubing. It is recommended that opaque tubing be used to prevent algae growth. Keep this
tubing as short as possible. This drain must be kept open to the atmosphere.
The fluid waste from drain connection of this instrument contains reagents
diluted with large quantities of sample water. HF scientific recommends that
operators check with local authorities concerning proper disposal of waste
fluids. This waste fluid must NEVER be reintroduced into the incoming water
stream.
A ½ “ ID tubing can be placed over the cabinet vent to redirect accidental
spills of reagent to a suitable container.
The CLX-XT is intended for cord connection with a three wire non-locking
grounded power cord, however rigid or flexible conduit connections can be used.
A set of one 120 VAC and one 240 VAC (EURO) power cords can be purchased
separately from the factory (Catalog No. 20779S). The CLX-XT is not supplied
with a power cord. If the CLX-XT is to be used in the U.S. or Canada the power
cord must be UL Listed & CSA Certified. Please consult all local electrical codes
for proper connection. All wiring to the instrument must be rated for 70 ˚C or
higher.
The connection block is marked N for Neutral and L for line the third symbol
indicates a secure earth ground. The green removable terminal block is suitable for
wire gauges 18 to 12 AWG. In addition a chassis common earth bond point is
provided on the chassis below the instrument safety access cover.
4.3.2 Cooling Chamber Power
Power must be connected to the cooling control box mounted on the top right side
of the cabinet. The requirements are 100-240VA, 47-63 Hz and 100VA. A circuit
breaker must be placed prior to the power connection in close proximity and within
easy reach to allow for service. This circuit breaker must be marked to indicate that
it is a disconnecting means for the cooling circuit for the instrument. All wiring to
the cooling chamber must be rated for 90 ˚C or
higher.
This supply is separated from the instrument power
connection because the cooling chamber must be
operated full time to prevent premature degradation
of the indicator reagent. Power cable enters from the
bottom of the cabinet. The connection block is
marked N for Neutral and L for line the third symbol
indicates a secure earth ground. The removable
terminal block is suitable for wire gauges 18 to 12
AWG. In addition a chassis common earth bond
point is provided on the chassis below the instrument safety access cover.
4.3.4 Relays
The Alarm 1 and Alarm 2 relays are mechanical relays rated at 240 VAC 2A.
Please note that the relays are labeled NO (Normally Open), NC (Normally
Closed) and C (Common). As these alarms are configured fail-safe, the normal
condition is with power applied to the CLX-XT and in a non-alarm condition.
Operation of these alarms is covered in section 7.4 Configuring the Alarms.
The lever operated terminal blocks are rated for wire gauges 28-14.
4.3.5 4-20 mA
The 4-20 mA output is driven by a 15 VDC power source and can drive recorder
loads up to 600 ohms. Transformer isolation is provided on the CLX-XT. Do not
run 4-20 mA cables in the same conduit as power. Operation of this output is
covered in section 7.1 Setting the 4-20 mA.
To prevent damage to the instrument, and for general safety ensure that power is
disconnected to the CLX-XT prior to making any connections. Polarities of the
connections are labeled beside this termination on the PC board.
The recommended cable is 22 AWG shielded twisted pair. To prevent ground
loops, connect the shield at either the CLX-XT or at its destination, but not both.
The grey terminal block is removable to assist in making connections.
The 4-20mA is factory calibrated. An adjustment is available on the 4-20mA in
sections 7.13 and 7.14. In addition to making adjustments, these menus output
continuous 4 mA or 20 mA and can be used as a signal test. Remember that the
configuration mode will timeout after 15 minutes.
The Buffer is read to use and is provided in a 500 ml bottle. The indicator requires some
preparation and is supplied in four 125 ml bottles. Each indicator reagent will last a
maximum of 3 months based on reagent aging when kept in the operating cooling
chamber.
Please note that calibrations will not be allowed until a reading is posted.
The security code (333) must be entered to gain access to CAL or CONFIG menus.
Notice that the first number in the code is flashing. The flashing indicates that this is the
number to be changed. Use the or arrows to select
the first of the three numbers in the code and then press
the 8 button to accept the first number of the code.
Now enter the second number in the code. Proceed as
with the first number followed by 8. Then repeat the
process for the third number in the access code, and
finish with the 8 button.
If the valid access code has been selected, the instrument will be directed to the calibration
mode. If the wrong access code is selected, the instrument will return to the AUTO mode.
Refer to section 7.5 Enabling the Security Access for more information.
The white LED is also used to draw attention to a problem as described in section 9.1
CLX-XT Fault Detection. In these instances the white LED blinks at a constant rate
dependant on the severity of the problem, but is still turned off as described above. Please
note that any fault is always posted to message queue on the lower portion of the LCD.
5. The screen will show the current reading on the CLX-XT. Using the & buttons
adjust the reading to agree with the laboratory method or portable photometer.
6. Press 8 to accept the calibration adjustment and return to AUTO measurement
mode.
There is a limit to the size of the change that can be made to a current
reading. The upper limit is the current reading times 1.5. The lower limit is
the current reading divided by 1.5.
Ensure a reading is posted to the display before calibrating to avoid a nOnE
error.
4. The screen will show the current reading on the CLX-XT. Since there is no
chlorine, the only reading may be a slight offset due to the absorbance of the
reagents. There should be no pink color developed.
5. Press 8 to perform a Zero calibration. When the calibration has completed the
instrument will return to AUTO measurement mode automatically.
There is a limit of ±0.20 mg/L total adjustment available. A ZERO Cal.
greater than this will cause a CAL warning and no calibration will have
occurred. Enter SERVICE mode to clear this error.
7.4.1 Alarm 1
Alarm 1 Function: The ALM1 is displayed and the display indicates the current
function of alarm 1 (HI, LO, OFF, or Error). Use the or buttons to cycle
through and select the desired function. Press the 8 button to accept the selection.
If the alarm was turned OFF, a prompt will appear to set up alarm 2 (go to section
Alarm 1 Set Point: This prompt is used to select the set point for this alarm; this
is indicated by “S/P” shown on the lower row of the display. Select the desired
alarm level by using the and buttons. Once the desired set point has been set,
press the 8 button to accept it.
7.4.2 Alarm 2
Repeat the procedure listed in section 7.4.1 to set up the parameters for alarm 2. If
a selection was made to turn the alarm OFF, the next selection for the speed of
response RESP is shown.
Due to the cyclic nature of the CLX-XT, relay chatter is not an issue. There is
no need for alarm delays or hysteresis.
If extended settings are set to OFF, pressing the button will save all settings and the
CLX-XT will automatically return to the normal AUTO mode of the instrument.
The factory setting is averaging of 2. Select the desired AVG using the and buttons and
press the 8 button to accept it.
In addition to averaging, the CLX-XT has a software filter that limits the change between
consecutive readings to 20%. For example, the reading following a reading of 1.00 ppm
could not be displayed as higher than 1.20 ppm or lower than 0.80 ppm. This filter also
helps smooth out large changes and eliminate reading
spikes. After 3 consecutive readings, it is assumed that
the large change in the readings is "real", and the filter
will be disabled. For example, the change between
reading 1 & 2, and readings 2 & 3 will be limited to
20%, but the change between readings 3 & 4 will not be
limited.
Note: This software filter is completely disabled when the averaging (AVG) is set to 1.
Make selections using the and buttons then press the 8 button to move to the next
menu.
The indicator reagent has a defined life after being mixed regardless of the cycle time
setting. If the reagent is kept in the powered cooling chamber the life is 90 days.
This feature is generally not used in ballast water applications and should be set to
OFF.
The use of this option may result in some slight loss of accuracy. The factory setting for
this option is OFF.
Make selections using the and buttons then press the button to exit to AUTO mode
and save all configuration settings. If the 4-20 mA in section 7.1 is turned ON, there are
two additional menus that will appear before returning to AUTO mode.
This setting will be slightly different on each instrument as each CLX-XT will be factory
set to 4.00mA. Press the button when adjustments are complete to save this setting and
move on to the 20mA adjustment.
When complete with the 20mA adjustment, press the button to exit to AUTO mode and save all
configuration settings.
A level 4 fault is simply a screen indication that one of the alarm levels has been
activated. This fault level will not affect the 4-20 mA and will only affect the alarm
activated. The sample back light blinks at a rate of once every 4 seconds.
A level 3 fault indicates a failure or a problem that usually can be corrected by the
operator. Refer to the chart below. If any of these errors occur, the instrument will still
display readings and probably will operate correctly. These faults will self-clear when the
problem is corrected. If any of these faults occur, they will affect the 4-20mA and any
alarm set for fault detection (Error). The sample back light blinks at a rate of once every
2.5 seconds.
Level 3 (Self-Clearing) fault conditions
Message Description of Fault Corrective Action
MA 4-20 mA enabled & loop open Check 4-20 mA wiring or turn off 4-
20mA if not used
CAL Calibration invalid –not accepted Recalibrate if needed
WATR No water flowing Check water flow
FAST Intake water flow too fast Set flow rate (see section 9.2)
SLOW Sample cuvette filling too slowly Set flow rate (see section 9.2)
PURG Sample cuvette has slow purge Check drain lines
NPRG Sample cuvette not purging Check drain lines
ISOL Problem with intake solenoid Check wiring, check for clogged solenoid
PSOL Problem with purge solenoid Check wiring, check for clogged solenoid
RGNT Problem with reagent Check reagent and lines
GLAS Dirty cuvette Replace or clean cuvette
WCAL Water Level Calibration Invalid Clear fault (see section 9.3)
A level 2 fault indicates a severe problem that will usually require technical assistance
from HF scientific customer service (see section 9.6). The queued display will show
POST. If this fault occurs it will affect the 4-20mA and any alarm set for fault detection
(Error). The sample back light blinks at rate of once every 1 second.
HF scientific
3170 Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Toll Free (888) 203-7248
Fax: (239) 332-7643
Email: [email protected]
www.hfscientific.com
The CLX-XT is shipped with one CLX-XT Tubing/Cuvette kit, HF part #. The kit consists
of the following:
Qty Part
1 Spare Buffer Cap Assembly
1 Spare Indicator Cap Assembly
8 Pump Tubes
1 Spare Cuvette
Two replacement sets of reagent cap/tubing assemblies are supplied with the CLX-XT;
one for the buffer and one for the indicator.
Additional kits can be ordered from your local HF scientific distributor or representative.
It is recommended to keep one kit on hand at all times.
Generally, both pump tubes should be replaced annually. Spare cap assemblies for both
the buffer and the indicator are supplied and can be replaced when needed.
Every 90 days
1. The indicator reagent required for operating this instrument must be changed at least
every 90 days. The buffer reagent can be used up to one year.
2. The internal strainer should be checked and cleaned if necessary.
3. The glass cuvette should be inspected. Check for excessive debris on the inside
surface of the glass. It is suggested to keep a spare cuvette to replace when required.
The old cuvette may be cleaned, if possible, for future replacement.
*The 1 year buffer and 90 day indicator reagent life is primarily used on ballast water
applications. The assumption is based on the factory default of a 2 ½ minute cycle and up
to 60 hours total operation per month.
Steps:
INLET OUTLET
CHECK VALVE DETAIL
1 Flush the system as described above to reduce personal contact with the reagents.
2 Press SERVICE to stop the flow of sample water and drain the cuvette.
3 Remove and retain the thumb screw on top of the pump; pull the pump hammer and
spring up and out of the way. There is no need to completely remove the hammer and
spring.
4 Working on one reagent side at a time. Replace the black tubing between optics inlet
and the check valve OUTLET. Discard the old tubing.
5 Ensure the check valve is placed into its seat.
6 Repeat steps 4 & 5 for the other reagent.
7 Replace the hammer and spring back into place and secure with the thumb screw.
8 Check the drawing on the following page to ensure correct installation.
9 Return to operation as described.
Steps:
1 Flush the system as described above to reduce personal contact with the reagents.
2 Press SERVICE to stop the flow of sample water and drain the cuvette.
3 Remove the buffer cap assembly from the cabinet vent and the connection at the check
valve.
4 Replace the buffer cap (blue cap) assembly. Note that the vent tube is not centered and
does not connect to the suction tube.
5 Repeat steps 3 & 4 for the indicator side (white cap).
6 Replace reagents, prime and return to service.
Cuvette Replacement
CLX-XT (28652) (5/11) Page 29
REV 1.5
Check the condition of cuvette and change if it appears badly soiled or discolored. Follow
the steps below.
Steps:
1. Replace the two Cap Assemblies.
2. Turn the knurled top on the optics system counterclockwise (as viewed for the top)
until the cuvette just “pops” out, but do not remove the top.
3. When the cuvette “pops” out, move the retaining o-ring & remove the cuvette. You
may need a stiff wire such as a bent paper clip to grasp the cuvette. Retain this cuvette
for future use if it can be cleaned.
4. Install the new cuvette by pushing it firmly in place and turning the knurled top
clockwise until the cuvette is held securely.
5. Check the drawing on the following page to ensure correct installation.
6. Return to operation as described.
T-Strainer Cleaning
A T- strainer is integral to the instrument and must be checked occasionally. When
necessary it must be removed and cleaned.
Steps:
1. Press SERVICE to stop the flow of sample water and
drain the cuvette.
2. Ensure the source water is turned off.
3. The T-strainer is clamped to the intake regulator. The
removal requires the use of a flat blade screwdriver as
shown in the photo. Once the clamp is opened the T-
strainer can be removed.
4. Disconnect the T-strainer and regulator from the tubing
and clear of the case.
5. Clean the T-strainer screen and replace.
6. Be certain the clamp is tightened fully and the pumping
connections are complete.
7. Turn source water back on.
8. Return to normal operation.
Tubes may darken due to contact with the reagent. This condition does not
affect the performance of these parts.
Once mixed the indicator has an expected life of up to 90 days if kept in the
powered cooling chamber. Write the mixing date on the reagent bottle labels
in the area provided. Dispose of expired reagents correctly.
To replace the reagents, press the SERVICE button; this will empty the cuvette and stop
any flow of water. Remove the cap on both bottles replace with the cap supplied with the
CLX-XT. The buffer is installed on the left and the indicator is installed in the cooling
chamber. The suction tube for both reagents will reach the bottom of the bottles.
To complete the replacement procedure, press the PRIME button and then the button.
This will draw enough of each reagent to completely prime the tubes and replace any old
solution. The system will automatically return to normal operation after it has primed.
Use caution when changing the reagents as they are corrosive. These reagents
may stain clothing. After changing the reagents, operators should wash their
hands.
As a mater of safety, always disconnect any power source to the CLX-XT prior to
attempting any cleaning. It is recommended that the source water is also shut off.
Isopropyl alcohol (rubbing alcohol) on a soft cotton cloth works very well in removing
reagent stains from plastic parts, the key pad and the display. Use care when cleaning
around electrical components. Do not use any harsh cleaning agents as these may cause
damage to the instrument components.
To order any accessory or replacement part, please contact the HF scientific Customer
Service Department. If for any reason technical assistance is needed regarding this
instrument, please do not hesitate to contact the HF scientific Technical Services
Department.
HF scientific
3170 Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Toll Free: (888) 203-7248
Fax: (239) 332-7643
Email: [email protected]
www.hfscientific.com
HF scientific inc., as vendor, warrants to the original purchaser of this instrument that it
will be free of defects in material and workmanship, in normal use and service, for a
period of one year from date of delivery to the original purchaser. HF scientific inc.’s
obligation under this warranty is limited to replacing, at its factory, the instrument or any
part thereof. Parts, which by their nature are normally required to be replaced
periodically, consistent with normal maintenance, specifically reagent, desiccant, sensors,
electrodes, tubing and fuses are excluded. Also excluded are accessories and supply type
items.
Original purchaser is responsible for return of the instruments, or parts thereof, to HF
scientific inc.’s factory. This includes all freight charges incurred in shipping to and from
HF scientific inc.’s factory.
HF scientific inc. is not responsible for damage to the instrument, or parts thereof,
resulting from misuse, environmental corrosion, negligence or accident, or defects
resulting from repairs, alterations or installation made by any person or company not
authorized by HF scientific inc..
HF scientific inc. assumes no liability for consequential damage of any kind, and the
original purchaser, by placement of any order for the instrument, or parts thereof, shall be
deemed liable for any and all damages incurred by the use or misuse of the instruments, or
parts thereof, by the purchaser, its employees, or others, following receipt thereof.
Carefully inspect this product for shipping damage, if damaged, immediately notify the
shipping company and arrange an on-site inspection. HF scientific inc. cannot be
responsible for damage in shipment and cannot assist with claims without an on-site
inspection of the damage.
This warranty is given expressly and in lieu of all other warranties, expressed or implied.
Purchaser agrees that there is no warranty on merchantability and that there are no other
warranties, expressed or implied. No agent is authorized to assume for HF scientific inc.,
any liability except as set forth above.
HF scientific, inc.
3170 Metro Parkway
Fort Myers, Florida 33916-7597
Phone: (239) 337-2116
Toll Free: (888) 203-7248
Fax: (239) 332-7643
Email: [email protected]
Website:www.hfscientific.com