0% found this document useful (0 votes)
4K views1 page

GSE-Series Calibracion

The document provides calibration procedures for several Industrial Weighing Systems scale models: The GSE-350 calibration procedure involves pressing keys in a specific sequence to zero the scale and then tare a calibration weight. The GSE 450/460 procedure has users press keys to zero the scale, then enter and tare the calibration weight value. The GSE 455/465/550 procedure has users enter a code, zero the scale empty, then enter and tare the calibration weight. The GSE 625/622 procedure involves using mode and zero keys to enter auto calibration, entering a bypass code, zeroing the empty scale, then entering and taring the calibration weight amount.

Uploaded by

Luis Huerta
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
4K views1 page

GSE-Series Calibracion

The document provides calibration procedures for several Industrial Weighing Systems scale models: The GSE-350 calibration procedure involves pressing keys in a specific sequence to zero the scale and then tare a calibration weight. The GSE 450/460 procedure has users press keys to zero the scale, then enter and tare the calibration weight value. The GSE 455/465/550 procedure has users enter a code, zero the scale empty, then enter and tare the calibration weight. The GSE 625/622 procedure involves using mode and zero keys to enter auto calibration, entering a bypass code, zeroing the empty scale, then entering and taring the calibration weight amount.

Uploaded by

Luis Huerta
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Industrial Weighing Systems

9 Richmond St. Picton, ON Canada K0K 2T0


Ph: 613-786-0016 Cell: 613-921-0397 Fax: 613-476-5293
E-mail [email protected] Website: www.iwsystems.ca

Procedure for GSE-350


Press Zero and Select simultaneously
Press Zero, Print, Units Tare one after the other scale should say zero.
With nothing on the scale press Tare
Use Print and Units keys
Enter the load calibration weight then press Tare
Add calibration weight press Tare
Keep pressing Tare until out

Procedure for GSE 450 460 calibration weight Scale


1: Press [Zero]+[Select] Simultaneously
2: Press [Zero] [PRINT] [UNITS] [ENTER] In that order
3: Remove any Load, keep pressing [Select] until "New Zero" is displayed
4: Press [Enter] to set zero
5: Place calibration weight on scale
6: Key in the calibration weight value
7: Press [Enter] "Cal OK"
8: Press [Enter] 3 times

Procedure for GSE 455 465 550


1) From the calibration weight screen key in 100 [Select]
Displays Key in Code: The code must be keyed in within 5 seconds
2) Press 54321 [ID] [Enter]
3) Prompt with quick cal, empty scale and Press [Enter]
4) Apply cal calibration weight, Key in calibration weight and Press [Enter]
5) Cal OK Press [Enter], Save press [Enter], Exit press [Enter]

Procedure for GSE 625 622


Use the scale select key to choose the scale you wish to calibrate
Press the [GSE] [MODE] to put the instrument into offset display mode
Press the [GSE] [ZERO]. The scale is now in the auto cal mode
Enter the program jumper bypass code [23640] [GSE] [.]
The lower display will prompt you through the set up of Auto Cal.
First remove all weight and press [Zero].
During zero lower display counts to 10, count will reset if motion is sensed
With count at 10 place cal weights on scale, key in amount and press [ENTER]
When the procedure is complete press [MODE] to return to offset displayed
If you want to cal another scale use scale select and repeat or:
Press [ZERO] to re-enter the Weigh Mode