Humacount 5D: - Quick Guide

The document provides a quick guide for using the HumaCount 5D device to perform measurements in capillary blood mode. It outlines the following steps: 1) 480μl of diluent is automatically dispensed into a bullet tube by the HC5D. 2) Select the "Capillary Whole Blood (CTA)" mode. 3) Collect a blood sample via finger prick using a HUMAN capillary tube containing EDTA. Wipe away excess blood. 4) Insert the capillary tube containing blood into the bullet tube of diluent to suspend the sample homogeneously. 5) Insert the bullet tube into the HC5D and press the button to start

Uploaded by

Mohamed Zina
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
91 views2 pages

Humacount 5D: - Quick Guide

The document provides a quick guide for using the HumaCount 5D device to perform measurements in capillary blood mode. It outlines the following steps: 1) 480μl of diluent is automatically dispensed into a bullet tube by the HC5D. 2) Select the "Capillary Whole Blood (CTA)" mode. 3) Collect a blood sample via finger prick using a HUMAN capillary tube containing EDTA. Wipe away excess blood. 4) Insert the capillary tube containing blood into the bullet tube of diluent to suspend the sample homogeneously. 5) Insert the bullet tube into the HC5D and press the button to start

Uploaded by

Mohamed Zina
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

HumaCount 5D

| Quick Guide

Cat No. 16450/5-1


1 2
480µl of HC5D- Click “Mode&ID” in the measuring
Diluent are screen and select “Capillary Whole
automatically Blood (CTA)” mode
dispensed into
a bullet tube by
the HC5D.

3
Blood collection by finger prick, and 20µl
HUMAN capillary tube (EDTA coded). After
blood collection is completed, wipe away
excess blood on wet lint free tissue. Avoid
any contact of tissue and blood inside the
capillary tube

For measurements in the

Quick Guide capillary blood mode you


need
for measurements in - HumaCount 5D device
Capillary Blood Mode - Capillary tube (REF 16070/30)
- Bullet tube
- Moist tissue

4
Put the capillary into the bullet tube
filled with Diluent, close the lid and
invert the sample carefully, until the
suspension is homogenous.

6
Results of 29 parameters will
be shown in the measuring

< <
screen

5• Open the lid, insert the bullet tube to the


collection needle and press the big red button
to start the recording. Make sure the needle
always touches the bottom of the bullet tube.
• HC5D automatically draws demanded volume
from the bullet tube