LPS Series: Operation Manual
LPS Series: Operation Manual
OPERATION MANUAL
1.1 Introduction
1.2 Safety considerations
1.3 Accessories
1.4 Output Isolation
1.5 Specifications
T w o: GETTING STARTED
3.1 Introduction
3.2 Operating Ranges
3.3 Load Consideration Connecting Multiple Loads
3.4 Parallel Connection Operation
3.5 Series Connection Operation
3.6 Tracking Operation (LPS-304)
3.7 Range Operation (LPS-30 l/302)
1.1 Introduction
This section contains a general description of your power supply as well as its performance specifications.
Information about options and accessories are also provided.
SAFETY SYMBOLS
WARNING The WARNING sign denotes a hazard. It calls attention to a procedure, practice,
or the like, which, if not correctly performed or adhered to, could result in
personal injury. Do not proceed beyond a WARNING sign until the indicated
conditions are fully understood and met
CAUTION The CAUTION sign denotes a hazard. It calls attention to a procedure, practice,
or the like, which, if not correctly performed or adhered to, could result in
damage to or destruction of part or all of the product.. Do not proceed beyond a
CAUTION sign until the indicated conditions are fully understood and met
1.3 Accessories
Power cable
Operation manual
Fuse
The output of the power supply is isolated from earth ground. Either output terminal may be grounded or
the output may floated up to +/- 240 Vdc (including output voltage) from chassis ground
1.5 Specifications
bit 7: 0=nothing
1 =digital output overload
Sept 4: Measure the DC current from positive channel output terminals with
the DMM and enter the measured current value as +I Lo.
Sept 5: Repeat step 4. Enter the measured value as +I Hi.
(For LPS-304 Only)
Sept 6: Measure the DC voltage from negative channel output terminals (- a;
COMl) with the DMM, and enter the measured value as -V Lo.
Sept 7: Repeat step 6. Enter the measured value- as -V Hi.------- ---
Sept 8: Measure the DC current from negative channel output terminals with
the DMM and enter the measured current value as -1 Lo.
Sept 9: Repeat step. 8. Enter the measured value as -1 Hi. Calibration is
completed.'
NOTE: If Error #0002 appear on the LCD after inputting ca1ibration parameter
please make sure all the calibration parameters are correct-and the
re-enter them again.