LPS Series: Operation Manual

The document provides operation and calibration instructions for Linear Power Supply (LPS) Series products, including the LPS-301, LPS-302, LPS-303, and LPS-304 models. It contains information on safety, accessories, specifications, front panel controls, display, operation modes, maintenance, and calibration procedures. Calibration involves measuring the voltage and current outputs with a DMM and entering the values to calibrate the power supply readings.

Uploaded by

petrocelli
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
877 views

LPS Series: Operation Manual

The document provides operation and calibration instructions for Linear Power Supply (LPS) Series products, including the LPS-301, LPS-302, LPS-303, and LPS-304 models. It contains information on safety, accessories, specifications, front panel controls, display, operation modes, maintenance, and calibration procedures. Calibration involves measuring the voltage and current outputs with a DMM and entering the values to calibrate the power supply readings.

Uploaded by

petrocelli
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 17

LPS Series

Linear Power Supplies

OPERATION MANUAL

/METE R INTE R NATIONAL COR P .

6 Floor, No. 59, Tsao-Dih Woei,


Wan Shun Liao, Shen Keng Hsiang,
Taipei Hsien, Taiwan, R.O.C.
Tel: 886-2-662-5093 ~5 886-2-662-5194 ~6
Fax: 886-2-662-5097
CONTENTS

ONE: GENERAL INFORMATION

1.1 Introduction
1.2 Safety considerations
1.3 Accessories
1.4 Output Isolation
1.5 Specifications

T w o: GETTING STARTED

2.1 Fromt Panel Controls And Output Terminals


2.2 LCD Display
2.3 How To Set Voltage/Current

THREE: OPERATING CHARACTERISTICS AND CONFIGUATIONS

3.1 Introduction
3.2 Operating Ranges
3.3 Load Consideration Connecting Multiple Loads
3.4 Parallel Connection Operation
3.5 Series Connection Operation
3.6 Tracking Operation (LPS-304)
3.7 Range Operation (LPS-30 l/302)

FOUR: USER MAINTENANCE/SERVICE

4.1 Fuse Replacement


4.2 In Case Of Difficulties

FIVE: USING THE RS-232-C SERIAL INTERFACE


1. GENERAL INFORMATION

1.1 Introduction

This section contains a general description of your power supply as well as its performance specifications.
Information about options and accessories are also provided.

1.2 Safety Considerations

DO NOT EXCEED INPUT RATINGS


This instrument must be connected to a properly grounded receptacle to minimize electric shock hazard.
Operation at line voltages or frequencies in excess of those stated on the data plate may cause leakage
currents in excess of 5.0mA peak.

SAFETY SYMBOLS

WARNING The WARNING sign denotes a hazard. It calls attention to a procedure, practice,
or the like, which, if not correctly performed or adhered to, could result in
personal injury. Do not proceed beyond a WARNING sign until the indicated
conditions are fully understood and met

CAUTION The CAUTION sign denotes a hazard. It calls attention to a procedure, practice,
or the like, which, if not correctly performed or adhered to, could result in
damage to or destruction of part or all of the product.. Do not proceed beyond a
CAUTION sign until the indicated conditions are fully understood and met

1.3 Accessories
Power cable
Operation manual
Fuse

1.4 Output Isolation

The output of the power supply is isolated from earth ground. Either output terminal may be grounded or
the output may floated up to +/- 240 Vdc (including output voltage) from chassis ground
1.5 Specifications

Model LPS301 LPS-302 LPS-303 LPS-304


MAX. OUTPUT POWER 30W 60W 90W 70W
OUTPUTS Single Single Single Triple
t OUTPUT VOLTAGE I
Output Voltage 30V/15V 30V/15V 3ov +/-30V 5V
Peak Output Voltage 32V/16V 32V/16V +/-32V
Setting Resolution 10mV 10mV 10mV 10mV
OUTPUT CURRENT
Output Current 2.5A 1A 2A
Peak Output Current 2.4A/l.2A 4.4A/2.2A 3A 1.2A
Setting Resolution 1mA 1mA ImA 1mA
4
CONSTANT VOLTAGE CHARACTERISTICS (at rated output)
Line Regulation ( 0% AC) 1mV 1mV 1mV 1mV 5mV
Load Regulation (100% load change) 2mV 2mV 2mV 2mV 10mV
Ripple/Noise rms (1 0Hz to 20MHz) 1mV rms 1mV rms 1mV rms 1mV rms 2mV rms
Ripple/Noise peak (1 0Hz to 20MHz) 1 0mV p-p 10mV p-p 1 0mV p-p 10mV p-p 1 0mV p-p
Transient Response 30ms I
Temperature Coefficient 100ppm/“c
CONSTANT CURRENT CHARACTERISTICS (at rated output)
Line Regulation ( 0% AC) 2mA 2mA 2mA 2mA
Load Regulation (100% load change) 5mA 5mA 5mA 5mA
Ripple/Noise rms (1 0Hz to 20MHz) 1mA rms 1mA rms 1mA rms
Ripple/Noise peak (1 0Hz to 20MHz) 5mA P-P 5mA P-P 5mA p-p 5mA P-P
I
Temperature Coefficient 200ppm/C
OTHER I
Display I 16 character x 2 line backlit LCD
Voltage Accuracy *
Current Accuracy *
Common Mode Voltage 40Vdc
Temperature Range Operating: to 40°C , less than 80% RH; Storage: -45°C to 75°C , less than
80% RH
Dimensions (WxHxL) 8.7”x 3.4”x 8.7”x 3.4”x 8.7”x 3.4”x 8.7”x 3.4”x 11.8”
11.8” 11.8” 11.8”
Weight 10 lbs 12 lbs 12 lbs 12 lbs
Cooling Forced Air
*For output less than 5% of rated output, add 5 digits to the accuracy specification

RS232 Interface: (optional)


1. RS232C DCE interface: 9 pin D-SUB connector
2. Port configuration: asynchronous 2400 baud, 8 data bits, 1 stop bits, no parity
N OTE:

1. All RS232 command are case-nonsensitivity ASCII codes.


2. Use async framing 8 data bits, no parity bit , l stop bit.
3. Bit rate=2400bps
4. Every command string is terminated by CR or LF or BOTH (carriage return)
5. There is one command allowable in a command string
6. A command string enter before “OK” prompt will be reject & no function
7. STATUS opratin explanation:
After a LPS accept A “STAUS” command, it will display a decimal number in ASCII
Convert this decimal number to binary form. each bit indicate a action/status:

bit 0: channel 1 o =c v l=CC

bit 1: channel 2 o =c v l=CC

bit 3,2: 0O: independent


10: tracking to channel 1
11 : tracking to channel 2

bit 4: 0 : digital output off


1: digital output on

bit 5: 0: digtal output 5V


l:digital output 3.3V

bit 6: O: output off


1: output on

bit 7: 0=nothing
1 =digital output overload

bit 8: O=fanoff 1=fan on

bit 9: O=beeper function disbale


l=beeper function enable

bit 10: 0=CC output compensated off


l=CC output compensated on
Calibration For LPS 301-304
Equipment needed for calibration: DMM, such as Fluke model 46 or HP 3479A

Step 1: Simultaneously press and keys to enter


. - - - -
calibration mode.
Setp 2: Measure the DC voltage from the positive channel output terminals
(+ and COMl) with DMM, and then use the arrow keys to enter the
measured value as the +V Lo,
Sept 3 : Repeat step 2. Enter the measured value as +V Hi.

Sept 4: Measure the DC current from positive channel output terminals with
the DMM and enter the measured current value as +I Lo.
Sept 5: Repeat step 4. Enter the measured value as +I Hi.
(For LPS-304 Only)
Sept 6: Measure the DC voltage from negative channel output terminals (- a;
COMl) with the DMM, and enter the measured value as -V Lo.
Sept 7: Repeat step 6. Enter the measured value- as -V Hi.------- ---

Sept 8: Measure the DC current from negative channel output terminals with
the DMM and enter the measured current value as -1 Lo.
Sept 9: Repeat step. 8. Enter the measured value as -1 Hi. Calibration is
completed.'
NOTE: If Error #0002 appear on the LCD after inputting ca1ibration parameter
please make sure all the calibration parameters are correct-and the
re-enter them again.