Temperature Measurement Theory
Temperature Measurement Theory
TEMPERATURE MEASUREMENT
THEORY
All methods of temperature measurement can be divided into direct ( contact methods affected by
means of thermometers, thermocouples and resistance thermometers ) and indirect ( contactless,
for instance, optical methods ). In the former, the sensor of measuring instrument is placed in
close vicinity of a heated object whose temperature is to be measured. Optical methods measure
either the intensity of radiation, or else a change in the spectral distribution of radiation as a
function of temperature.
Thermocouples
A thermocouple is a system composed of two homogeneous metal wires dissimilar in chemical
composition that are soldered, fused or welded together at one end, forming a closed circuit when
the free-ends are connected to an indicating instrument which is mostly a potentiometer. ( See
Figure 1 )
Types of Thermocouples
Any dissimilar metal junction could generate a thermo-emf when it is heated, but those pairs used
for temperature measurements are limited. The thermocouples and the symbols used for them are
listed in Table 1.
Several types of thermocouples may have similar accuracy and initial cost. The objective is to
select the type with greatest stability and longest life. Selection may depend on whether the
atmosphere will corrode or poison the thermocouple.
Type J : Not used below 0oC because condensation may cause rusting of iron. They are best used
in reducing atmospheres.
Type T: May be used in slightly oxidizing or reducing atmospheres and are frequently used for
low temperature work.
2
Type K: Used only in an oxidizing atmosphere. A reducing atmosphere may cause green rot,
which shifts calibration.
OBJECTIVE
To conduct temperature measurement tests in order to demonstrate the use of thermocouples for
high temperature measurements.
APPARATUS
1. Vertical tube furnace
2. Thermocouple calibration cell
3. Thermocouple
4. Potentiometer
5. Mercury in glass thermometer
PROCEDURE
PART A :
1. Assemble the thermocouple calibration cell ( See Figure 2 ).
2. Power up the vertical tube furnace, by setting up the controller to a desired temperature
( a temperature above the melting point of calibration material ).
3. Measure the temperature of the furnace intermittently until it reaches the desired temperature
and stabilizes.
4. Connect thermocouple from the calibration cell to a potentiometer and turn the furnace off.
5. Measure the thermo-emf from the calibration cell every 30 seconds and record them until the
temperature given by the instructor.
6. Apply corrections to the measured thermo-emf values if the cold junction temperature is not
0oC. Use the following procedure for correction :
i. Measure the reference ( cold ) junction temperature of your thermocouple in oC using a
mercury in glass thermometer.
ii. Find out the corresponding voltage to the temperature measured in step (i) from the calibration
tables.
iii. Add voltage measured from the thermocouple and the voltage found in step (ii) to obtain the
corrected voltage.
iv. Find out the corresponding temperature to the corrected voltage in step (iii) from the
calibration table.
3
Figure 2. Schematic Cross-section of Thermocouple Calibration Cell.
PART B: ( OPTIONAL )
Use ice-water mixture to construct a cold junction and assemble the thermocouple calibration cell
as given in Figure 3. Then, repeat the procedure given in Part A, except the correction part.
SUGGESTED READINGS