0% found this document useful (0 votes)
455 views1 page

N F: NFT S C: EAR Ield Tandard Alibration Tube

The document discusses calibration tubes for near field testing. It recommends the CT45 calibration tube for its quick and simple calibration. The tube contains artificial internal defects of varying depths for calibration, including GR40 at 40% wall thickness, GR60 at 60% wall thickness, and H100 being a through-wall defect. It provides a chart of defect parameters based on tube wall thickness and outer diameter. It also includes a variable chart with dimensional values for the calibration tube in both inches and millimeters.

Uploaded by

Mahmood Khan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
455 views1 page

N F: NFT S C: EAR Ield Tandard Alibration Tube

The document discusses calibration tubes for near field testing. It recommends the CT45 calibration tube for its quick and simple calibration. The tube contains artificial internal defects of varying depths for calibration, including GR40 at 40% wall thickness, GR60 at 60% wall thickness, and H100 being a through-wall defect. It provides a chart of defect parameters based on tube wall thickness and outer diameter. It also includes a variable chart with dimensional values for the calibration tube in both inches and millimeters.

Uploaded by

Mahmood Khan
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

Olympus NDT Canada

505, boul. du Parc-Technologique Tel. : (418) 872-1155


Québec (Québec) G1P 4S9 Fax. : (418) 872-5431
Canada Web site : www.olympusndt.com

NEAR FIELD: NFT STANDARD CALIBRATION TUBE

RECOMMENDED CALIBRATION TUBES


The CT45 calibration tube model has been designed for a quick and simple NFT calibration. As near field only
detects internal defects and wall thinning, it is totally unnecessary to have fins on the calibration tube.

GR60
(internal) GR40
H100
(internal)

GR60 H100 GR40

ARTIFICIAL DEFECT PARAMETERS IN FUNCTION OF TUBE WT AND OD


Defect Depth (%WT) Diameter

GR40 40 % (internal)

GR60 60 % (internal)
If OD>1.00”, 0.250”
H100 Thru
Else: 0.188”

VARIABLE CHART IN FUNCTION OF TUBE OD


Value Value
Variable Variable
in mm in mm

A 4.75” 121mm D 5.00” 127mm

B 3.50” 89mm GL 0.75" 19mm

C 2.75” 70mm TL 17.50” 444mm