PaX-i3D Smart Troubleshooting Manual
Version 3.0
General Information
The document is created by Vatech Co., Ltd. aiming to give overview of troubleshooting guide for
PaX-i3D Smart on this manual is the confidential property of Vatech Co., Ltd. and any unauthorized
distribution or copying of this document is strictly prohibited. The terms and conditions are subject
to change at any time without notice by Vatech Co., Ltd. This manual will be continuously revised
and distributed with system upgrade or design modification.
Product Information
PaX-i3D Smart Platform 1 ~ 3
( W-Pano + AnyCeph & AnyCeph-X )
Manual Revision
PaX-i3D Smart Platform 1~3 Troubleshooting Manual Ver. 3.0 - ENG
Revision Date
2017. 06. 02.
Safety Information
To perform troubleshooting of PaX-i3D Smart, operators should perform the
Image procedure according to the following instruction. This Manual is intended
only for Vatech Co., Ltd. authorized engineers to avoid personal injury and
damage.
To effectively describe safety information, the following signs are used in this
manual.
Useful information or possible harmful situations are
indicated.
Necessary instructions or application information are
indicated.
Important application instructions that could result serious
personal injury or equipment damage, are described.
Critical application information that could result critically
severe injury or damage on both operator and equipment,
are described.
Radioactivity sign to signify the radiation hazard.
Contents
I. Tube Alarm H003, H012 ............................................................................5
1. FLOWCHART ......................................................................................................... 5
2. ABOUT SYMPTOM................................................................................................. 6
3. REASON OF DEFECT ............................................................................................ 6
4. TROUBLESHOOTING ............................................................................................ 6
4.1 Is power board output voltage within the normal range? ......................................................... 6
4.2 Check each spare part for serviceability .................................................................................. 7
5. SPARE PARTS ....................................................................................................... 9
I. Tube Alarm H003, H012
1. FLOWCHART
Tube Alarm H003
Tube Alarm H012
4.1
1 Is power board output voltage No
within normal the range?
Replace the power board
Yes
4.2
Check each defective part
Select your case
[Case 3] [Case 4]
[Case 2]
[Case 1] Replace Replace Generator,
Replace Generator
Replace generator Generator and CI collimator and
and CI collimator
W-PANO sensor W-PANO sensor
2. ABOUT SYMPTOM
2.1 Image Capturing is not available due to the occurrence of [TUBE ALARM : H003]
or [TUBE ALARM : H012].
3. REASON OF DEFECT
3.1 [TUBE ALARM : H003] Current is more than maximum permissible current while
Inverter Board is working (X-Ray exposure).
Error “2” is displayed on Inverter Board.
3.2 [TUBE ALARM : H012] During X-Ray exposure, kVp Feedback is less than -20 kVp off
from the setting value.
Error “3” is displayed on Inverter Board.
4. TROUBLESHOOTING
4.1 Is power board output voltage within the normal range?
(1) Check the voltage supply status by measuring the wall voltage. If the power supply is unstable, install
the AVR.
(2) Turn off the power of unit and disassemble Vertical Top case.
(3) Turn on the power of the unit.
(4) By using the DMM (Digital Multimeter), measure the output voltage of power board.
[Figure 1]
(5) Check whether the output voltage is within the normal range.
(6) If the output voltage of power board is not normal, replace the power board.
[Figure 1]
SAP Code Item Name Detail
B0000392 BOARD POWER/(VT-PWR-007)(SI-665N)/MASTER N/A
4.2 Check each spare part for serviceability
(1) Turn on the unit.
(2) Execute the Terminal 1.9b.
(3) After entering the [SPM_CVER] command, click ‘Send’.
And, Check the feedback result whether communication is normal or not.
Normal
Abnormal
(4) After entering the [SPM_FISS_0001] command, click ‘Send’.
Check the feedback result to make sure if the communication is normal or not.
Normal
Abnormal
(5) Case 1: is normal for all responses.
(6) Case 2: CI Collimator does not response.
(7) Case 3: W-PANO does not response.
(8) Case 4: CI Collimator and W-PANO do not response.
4.2.1 [Case 1] Replace generator
(1) If all parts response normally, replace the generator. As the generator is composed with of the tube
tank and the inverter board, replace them together when replacing the generator.
(2) After replacing the generator, it is mandatory to perform the X-Ray beam alignment mandatorily.
(3) After replacing generator, it is recommended to proceed with the Sensor calibration and Geometry
settings.
SAP Code Item Name Detail
G0000031 DG-07C11T2 (H-TYPE)(SMART VISTAPANO GREEN 15X15) /07CT2 N/A
H-TYPE TANK AND INV-11 INVERTER/MASTER
4.2.2 [Case 2] Replace generator and CI collimator
(1) If the CI collimator does not response, replace CI collimator and generator by referring to the [Case 1].
SAP Code Item Name Detail
A0300240 L-P-3 ASSY CI (PHT-30LFO)/CS PART)(141010)/MASTER N/A
4.2.3 [Case 3] Replace generator and CT/PANO sensor
(1) Check sensor type.
(2) If the CT/PANO sensor does not response, replace CT/PANO sensor and generator by referring to the
[Case 1].
SAP Code Item Name Detail Cut-in Serial
S0000258 Xmaru1404CF/W-PANO, PPE/B1A-148Z03 Single Mode ~ 061-1829
S0000341 Xmaru1404CF/W-pano_R1/MASTER Multi Mode 61-1830
4.2.4 [Case 4] Replace generator, CI collimator and CT/PANO sensor
(1) If CI collimator, CT/PANO sensor do not response, replace them with the above mentioned spare
parts.
5. SPARE PARTS
5.1 Essential Part
SAP Code Item Name Detail
DG-07C11T2 (H-TYPE)(SMART VISTAPANO GREEN 15X15) /07CT2 H-
G0000031 N/A
TYPE TANK AND INV-11 INVERTER/MASTER
5.2 Recommended Part
Cut-in
SAP Code Item Name Detail
Serial
B0000392 BOARD POWER/(VT-PWR-007)(SI-665N)/MASTER N/A N/A
A0300240 L-P-3 ASSY CI (PHT-30LFO)/CS PART)(141010)/MASTER N/A N/A
S0000258 Xmaru1404CF/W-PANO, PPE/B1A-148Z03 Single mode ~ 061-1829
S0000341 Xmaru1404CF/W-pano_R1/MASTER Multi Mode 061-1830 ~
After replacing the Generator, please make the sensor calibration / X-ray beam alignment /
Geometry calibration / CT number calibration.
Refer the “PI3DS_110T_45A_en]Technical_Manual.pdf "
Download: http://vcsm.vatech.co.kr > Download > Manual > No.9161