BioFire COVID 19 Instructions For Use DFA2 PRT 0080 04
BioFire COVID 19 Instructions For Use DFA2 PRT 0080 04
BioFire ®
COVID-19 Test
Instructions for Use
Manufactured by
BioFire Defense, LLC
79 West 4500 South, Suite 14
Salt Lake City, UT 84107 USA
TABLE OF CONTENTS
INTENDED USE ............................................................................................................. 1
SUMMARY AND EXPLANATION OF THE TEST ................................................................. 1
PRINCIPLE OF THE PROCEDURE ..................................................................................... 2
MATERIALS PROVIDED .................................................................................................. 3
MATERIALS REQUIRED BUT NOT PROVIDED .................................................................. 3
WARNINGS AND PRECAUTIONS .................................................................................... 4
General Precautions ........................................................................................................ 4
Safety Precautions........................................................................................................... 4
Laboratory Precautions ................................................................................................... 5
Precaution Related to Public Health in the United States ............................................... 6
REAGENT STORAGE, HANDLING, AND STABILITY ........................................................... 6
SAMPLE REQUIREMENTS............................................................................................... 7
BIOFIRE® COVID-19 TEST PROCEDURE............................................................................ 7
Preparing Specimens for Pooling .................................................................................... 7
Step 1: Prepare Pouch ..................................................................................................... 8
Step 2: Hydrate Pouch ..................................................................................................... 8
Step 3: Prepare Sample Mix ............................................................................................ 9
Step 4: Load Sample Mix ................................................................................................. 9
Step 5: Run Pouch ......................................................................................................... 11
FilmArray 2.0 .............................................................................................................. 11
FilmArray Torch .......................................................................................................... 12
QUALITY CONTROL ...................................................................................................... 13
Process Controls ............................................................................................................ 13
Monitoring Test System Performance .......................................................................... 13
BIOFIRE® COVID-19 TEST EXTERNAL CONTROLS ........................................................... 14
External Controls ........................................................................................................... 14
BioFire® COVID-19 Test External Control Procedure (+) ................................................ 15
BioFire® COVID-19 Test External Control Procedure (-) ................................................. 16
INTERPRETATION OF RESULTS ..................................................................................... 17
Assay Interpretation ...................................................................................................... 17
Organism Interpretation ............................................................................................... 18
SARS-CoV-2 ................................................................................................................. 18
The BioFire COVID-19 Test is also for the qualitative detection of nucleic acid from SARS-CoV-2 in pooled
samples containing up to eight nasopharyngeal swabs collected individually in transport media from individuals
suspected of COVID-19 by their healthcare provider. Testing of pooled specimens is limited to DoD laboratories
that meet the requirements to perform high complexity tests.
Specimens should only be pooled in areas with low SARS-CoV-2 prevalence, and when testing demand exceeds
laboratory capacity or reagent availability. For pooled specimen testing, authorized laboratories will adhere to a
protocol for ongoing monitoring of the pooling strategy or limit testing to individuals who are subjected to a
detailed infection prevention and control plan.
Results are for the identification of SARS-CoV-2 RNA. The SARS-CoV-2 RNA is generally detectable in
nasopharyngeal swabs in transport media during the acute phase of infection. Positive results are indicative of
the presence of SARS-CoV-2 RNA; clinical correlation with patient history and other diagnostic information is
necessary to determine patient infection status. Positive results do not rule out bacterial infection or co-infection
with other viruses. Pooled samples with positive or equivocal results must be tested individually prior to reporting
results. The agent detected may not be the definite cause of disease. Laboratories within the United States and
its territories are required to report all results to the appropriate public health authorities.
Negative results do not preclude SARS-CoV-2 infection and should not be used as the sole basis for patient
management decisions. Negative results must be combined with clinical observations, patient history, and
epidemiological information. Negative results from pooled samples should be reported as presumptive.
Specimens with low viral genetic material may not be detected in pooled samples due to decreased sensitivity. If
clinical signs and symptoms are inconsistent with a negative result or results are necessary for patient
management, the patient should be considered for individual testing.
The BioFire COVID-19 Test is intended for use by laboratory personnel who have received specific training on
the use of the FilmArray® 2.0 and/or the FilmArray® Torch Instrument Systems. The BioFire COVID-19 Test is
only for use under the Food and Drug Administration’s Emergency Use Authorization.
General Precautions
1. For in vitro diagnostic (IVD) use under Emergency Use Authorization only.
2. Positive results are indicative of the presence of SARS-CoV-2 RNA.
3. Testing of pooled specimens should only be performed in high complexity laboratories.
4. Laboratories within the Unites States and its territories are required to report all results to the appropriate
public health authorities.
5. The BioFire COVID-19 Test has not been FDA cleared or approved but has been authorized by FDA
under an Emergency Use Authorization (EUA) for use by laboratories which are certified under the
Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. §263a, and meet the
requirements to perform moderate and high complexity tests.
6. The BioFire COVID-19 Test has been authorized only for the detection of nucleic acid from SARS-CoV-
2, not for any other viruses or pathogens.
7. The BioFire COVID-19 Test is only authorized for the duration of the declaration that circumstances exist
justifying the authorization of emergency use of in vitro diagnostics for detection and/or diagnosis of
COVID-19 under Section 564(b)(1) of the Federal Food, Drug and Cosmetic Act, 21 U.S.C. § 360bbb-
3(b)(1), unless the declaration is terminated or authorization is revoked sooner..
8. BioFire COVID-19 Test pouches are only for use with FilmArray 2.0 and FilmArray Torch systems.
9. BioFire COVID-19 External Control Kit (+) is only for use with FilmArray 2.0 and FilmArray Torch
systems.
10. Always check the expiration date on the pouch. Do not use a pouch after its expiration date.
11. FilmArray pouches are stored under vacuum in individually-wrapped canisters. To preserve the integrity
of the pouch vacuum for proper operation, be sure that a FilmArray instrument/module will be available
and operational before unwrapping any pouches for loading.
12. Bleach introduced in a sample may damage nucleic acids in the sample, which may lead to a false
negative result.
13. If infection with SARS-CoV-2 is suspected based on current clinical and epidemiological screening
criteria recommended by public health authorities, specimens should be collected with appropriate
infection control precautions.
Safety Precautions
1. Wear appropriate Personal Protective Equipment (PPE), including (but not limited to) disposable clean
powder-free gloves. Protect skin, eyes, and mucus membranes. Change gloves often when handling
reagents or samples.
2. Handle all samples and waste materials as if they were capable of transmitting infectious agents.
Observe safety guidelines such as those outlined in:
• CDC/NIH Biosafety in Microbiological and Biomedical Laboratories1
• CLSI Document M29 Protection of Laboratory Workers from Occupationally Acquired Infections.2
• Refer to Interim Laboratory Safety Guidelines for Handling and Processing Specimens Associated
with SARS-CoV-2 www.cdc.gov/coronavirus/2019-nCoV/lab-biosafety-guidelines.html.
3. Follow your institution's safety procedures for handling biological samples.
4. Dispose of materials used in this assay, including reagents, samples, and used buffer tubes, according
to federal, state, and local regulations.
5. Sample Buffer is assigned the following classifications:
• Acute toxicity (Category 4),
Laboratory Precautions
1. Preventing Organism Contamination
Due to the sensitive nature of the BioFire COVID-19 Test, it is important to guard against contamination of
the sample and work area by carefully following the testing process outlined in this instruction document,
including these guidelines:
• Do not handle samples or pouches in a biosafety cabinet which is used for SARS-CoV-2 culture or
immunofluorescence testing.
• Prior to processing samples, thoroughly clean both the work area and the FilmArray® Pouch Loading
Station using a suitable cleaner such as freshly prepared 10% bleach or a similar disinfectant. To avoid
residue buildup and potential damage to the sample or interference from disinfectants, wipe disinfected
surfaces with water.
• Samples and pouches should be handled and/or tested one-at-a-time. Always change gloves and
clean the work area between each pouch and sample.
• Use clean gloves to remove materials from bulk packaging bags and reseal bulk-packaging bags when
not in use.
1. Store the test and control kit, including reagent pouches and provided buffers, at room temperature (15-
30°C). DO NOT REFRIGERATE.
2. Avoid storage of any materials near heating or cooling vents, or in direct sunlight.
3. All kit components should be stored and used together. Do not use components from one kit with those of
another kit. Discard any extra components from the kit after all pouches have been consumed.
4. Do not remove pouches from their packaging until a sample is ready to be tested. Once the pouch
packaging has been opened, the pouch should be loaded as soon as possible (within approximately 30
minutes).
5. Once a pouch has been loaded, the test run should be started as soon as possible (within approximately 60
minutes). Do not expose a loaded pouch to temperatures above 40°C (104°F) prior to testing.
NOTE: Bleach can damage organisms/nucleic acids within the specimen, potentially causing false negative
results. Contact between bleach and specimens during collection, disinfection, and testing procedures should
be avoided.
Refer to the BioFire COVID-19 Test Quick Guide or the appropriate FilmArray operator’s manual for more
details.
3. Check the expiration date on the pouch. Do not use expired products.
4. Insert the pouch into the FilmArray Pouch Loading Station, aligning
the red and blue labels on the pouch with the red and blue arrows on
the FilmArray Pouch Loading Station.
5. Place a Sample Injection Vial (with red cover) into the red well of the
FilmArray Pouch Loading Station.
6. Place a Hydration Injection Vial (with blue cover) into the blue well of
the FilmArray Pouch Loading Station.
NOTE: DO NOT use the Transfer Pipette to mix the sample once it is loaded into the
Sample Injection Vial.
NOTE: Waiting 5 seconds decreases the risk of dripping and contamination from
the sample.
2. Lift the Sample Injection Vial, leaving the red cover in the well of the FilmArray
Pouch Loading Station, and insert the Sample Injection Vial cannula tip into the
pouch sample port located directly below the red arrow of the FilmArray Pouch
Loading Station.
3. Forcefully push down in a firm and quick motion to puncture seal (a faint “pop” is heard) and sample is
pulled into the pouch by vacuum.
FilmArray 2.0
1. Ensure that the FilmArray 2.0 system (instrument and computer) is powered on and the software is launched.
2. Follow on-screen instructions and procedures described in the FilmArray 2.0 operator’s manual to place
the pouch in an instrument, enter pouch, sample, and operator information.
3. Pouch identification (Lot Number and Serial Number), Pouch Type and Protocol, will be automatically
entered when the barcode is scanned. If it is not possible to scan the barcode, the pouch Lot Number,
Serial Number, and Pouch Type can be manually entered from the information provided on the pouch label
into the appropriate fields. To reduce data entry errors, it is strongly recommended that the pouch
information be entered by scanning the barcode.
NOTE: When selecting a Pouch Type manually, ensure that the Pouch Type matches the label on the
BioFire COVID-19 Test pouch.
4. Enter the Sample ID. The Sample ID can be entered manually or scanned in by using the barcode scanner
when a barcoded Sample ID is used.
5. Select and/or confirm the appropriate protocol for your sample type from the Protocol drop down list. The
BioFire COVID-19 Test has a single NPS2 protocol available from the drop down list.
6. Enter a user name and password in the Name and Password fields.
NOTE: The font color of the user name is red until the user name is recognized by the software.
7. Review the entered run information on the screen. If correct, select Start Run.
Once the run has started, the screen displays a list of the steps being performed by the instrument and the
number of minutes remaining in the run.
NOTE: The bead-beater apparatus can be heard as a high-pitched noise during the first minute of
operation.
8. When the run is finished, follow the on-screen instructions to remove the pouch, then immediately discard it
in a biohazard waste container.
9. The run file is automatically saved in the FilmArray database, and the test report can be printed, viewed,
and/or saved as a PDF file.
10. To view run data, double click on a run file, select the interpretation tab and click on a specific assay result.
NOTE: The font color of the user name is red until the user name is recognized by the software.
8. Review the entered run information on the screen. If correct, select Start Run.
Once the run has started, the screen displays a list of the steps being performed by the Module
(instrument) and the number of minutes remaining in the run.
NOTE: The bead-beater apparatus can be heard as a high-pitched noise during the first minute of
operation.
9. At the end of the run, remove the partially ejected pouch, then immediately discard it in a biohazard waste
container.
10. The run file is automatically saved in the FilmArray database, and the test report can be viewed, printed,
and/or saved as a PDF file.
Process Controls
Two process controls are included in each pouch:
2. PCR2 Control
The PCR2 Control assay detects a DNA target that is dried into wells of the array along with the
corresponding primers. A positive result indicates that PCR2 was successful.
Both control assays must be positive for the test run to pass. If controls fail, the sample should be retested
using a new pouch.
External Controls
For quality control and laboratory test verification, BioFire Defense provides an optional external positive assayed
control kit to monitor the performance of in vitro laboratory nucleic acid testing procedures for the qualitative
detection of the BioFire COVID-19 Test performed on FilmArray 2.0 and FilmArray Torch systems. Offered
separately, the BioFire COVID-19 Test External Control (+) kit is a surrogate control material comprised of dried
synthetic RNA in buffer and stabilizer, supplied in an External Control Vial that is used directly with the BioFire
COVID-19 Test. The BioFire COVID-19 Test External Control (+) Kit contains six (6) BioFire COVID-19 Test
External Control (+) Vials. The RNA in the external control includes RNA segments to monitor whether the PCR
primers for each SARS-CoV-2 assay are present for both stages of the nested PCR.
The BioFire COVID-19 Test External Control (+) Kit contains no biological hazards and is 100% non-infectious.
This control is stored at 15-30°C. To run a positive external control, reference BioFire COVID-19 Test External
Control Procedure (+) below or BioFire COVID-19 External Control Kit (+) Quick Guide. To run a negative
external control, use the BioFire COVID-19 Test and reference BioFire COVID-19 External Control Procedure (-)
below.
An optional verification protocol may be obtained from the BioFire Defense Product Support webpage
www.biofiredefense.com/covid-19test. Customers may also use these BioFire COVID-19 Test External Control
Kit (+) or an alternate Quality Control material for verification testing.
Good laboratory practice recommends running positive and negative external controls regularly. Evaluation of
external controls is recommended prior to using a new shipment or new lot of BioFire COVID-19 Test Kits.
Evaluation of external controls is also recommended when there is a new operator, and following
replacement/repair of a FilmArray® 2.0 or FilmArray® Torch system.
External controls may also be used in initial laboratory validations of the FilmArray 2.0 or FilmArray Torch system
used with the BioFire COVID-19 Test in accordance with appropriate federal, state, and local guidelines or
accreditation requirements, as applicable.
It is ultimately the responsibility of each laboratory to determine the frequency and type of material used for
external control testing as part of the laboratory’s Quality Control program.
Information on how to obtain optional external control material is posted on the BioFire Defense webpage.
BioFire COVID-19 Test External Control Kit (+)
Part Number: 423748
6. Recap the External Control Vial (+) and mix by gently inverting three (3) times.
7. Pour the rehydrated External Control (+) into the Sample Injection Vial and
immediately dispose of the External Control Vial (+).
• Change gloves.
8. Tightly close lid of Sample Injection Vial and mix by gently inverting at least three (3)
times.
9. Return Sample Injection Vial to red well of Pouch Loading Station.
10. Continue at Step 4 of the BioFire COVID-19 Test Panel Procedure to load the
pouch and run it on the FilmArray.
6. Remove the Sample Injection Vial from the FilmArray Pouch Loading Station
and invert the vial at least three (3) times to mix.
7. Return the Sample Injection Vial to the red well of the FilmArray Pouch
Loading Station.
8. Continue at Step 4 of the BioFire COVID-19 Test Procedure to load the
pouch and run the FilmArray.
SARS-CoV-2a ORF1ab
SARS-CoV-2d ORF1ab
SARS-CoV-2e ORF8
Assay Interpretation
When PCR2 is complete, the FilmArray instrument performs a DNA melting analysis on the PCR products and
measures the fluorescence signal generated in each well (for more information see appropriate FilmArray
operator’s manual). The FilmArray Software then performs several analyses and assigns a final assay result
for every well. The steps in the analyses are described below.
Analysis of Melt Curves. The FilmArray Software evaluates the DNA melt curve for each well of the PCR2 array
to determine if a PCR product was present in that well. If the melt profile indicates the presence of a PCR
product, then the analysis software calculates the melting temperature (Tm) of the curve and compares it against
the expected Tm range for the assay. If the software determines that the Tm of the curve is within the assay
specification Tm range, the melt curve is called positive. If the software determines that the Tm of the curve is not
in the appropriate Tm range, the melt curve is called negative.
Analysis of Replicates. Once positive melt curves have been identified, the software evaluates the three
replicates for each assay to determine the assay result. For an assay to be called positive, at least two of the
three associated melt curves must be called positive, and the Tm for at least two of the three positive melt curves
must be similar (within 1.0°C). Assays that do not meet these criteria are called negative.
Table 4. Interpretation of Internal Controls Field on the BioFire COVID-19 Test Report
Internal
Controls Explanation Action
Result
The run was successfully
completed
None
Passed AND Report the results provided on the test
report.
Both pouch controls were
successful.
The run was successfully
completed
Result Summary
The Result Summary section of the test report lists the result for the overall target and each individualized
assay result. Possible results for each assay are ‘Detected’, ‘Equivocal’, ‘Not Detected’, or ‘Invalid’. Table 5
provides an explanation for each interpretation and any follow-up necessary to obtain a final result. The SARS-
CoV-2 target and the three associated assays are listed in the Results Summary section. According to the
result for the target, ‘Detected’, ‘Not Detected’, ‘Equivocal’, or ‘Invalid’ will be indicated to the left of the target
name. According to the result for each associated assay, ‘Detected’, ‘Not Detected’, or ‘Invalid’ will be
indicated to the left of each assay name.
SARS-CoV-2
Explanation Action
Results
The run was successfully completed
Change Summary
It is possible to edit the Sample ID once a run has completed. If this information has been changed, an
additional section called Change Summary will be added to the test report. This Change Summary section
lists the field that was changed, the original entry, the revised entry, the operator that made the change, and
the date that the change was made. Sample ID is the only field of the report that can be changed.
To assist clinical laboratories running the BioFire COVID-19 Test, the relevant Conditions of Authorization are
listed below, and are required to be met by laboratories performing the EUA test.
A. Authorized laboratories1 using the BioFire COVID-19 Test will include with test result reports, all
authorized Fact Sheets. Under exigent circumstances, other appropriate methods for disseminating these
Fact Sheets may be used, which may include mass media.
B. Authorized laboratories using the BioFire COVID-19 Test will use the BioFire COVID-19 Test as outlined
in the Instructions for Use. Deviations from the authorized procedures, including the authorized
instruments, authorized extraction methods, authorized clinical specimen types, authorized control
materials, authorized other ancillary reagents and authorized materials required to use the BioFire
COVID-19 Test are not permitted.
D. Authorized laboratories using the BioFire COVID-19 Test will have a process in place for reporting test
results to healthcare providers and relevant public health authorities, as appropriate.
E. Authorized laboratories will collect information on the performance of the BioFire COVID-19 Test and
report to DMD/OHT7-OIR/OPEQ/CDRH (via email: [email protected]) and BioFire
Defense Product Support website https://www.biofiredefense.com/product-support/filmarray-
support/adverse-reporting-biofire-covid19-test/ any suspected occurrence of false positive or false
negative results and significant deviations from the established performance characteristics of the BioFire
COVID-19 Test of which they become aware.
F. All laboratory personnel using the BioFire COVID-19 Test must be appropriately trained in RT-PCR
techniques and use appropriate personal protective equipment when handling this kit, and use the BioFire
COVID-19 Test in accordance with the authorized labeling.
G. For pooled specimen testing, authorized laboratories will adhere to a protocol for ongoing monitoring of
the pooling strategy or limit testing to individuals who are subjected to a detailed infection prevention and
control plan.
H. Authorized laboratories using specimen pooling strategies when testing patient specimens with the
BioFire COVID-19 Test will include with test result reports for specific patients whose specimen(s) were
the subject of pooling, a notice that pooling was used during testing and that “Patient specimens with low
viral loads may not be detected in sample pools due to the decreased sensitivity of pooled testing.”
I. Authorized laboratories implementing pooling strategies for testing patient specimens must use the
“Specimen Pooling Implementation and Monitoring” recommendations available in the authorized labeling
to evaluate the appropriateness of continuing to use such strategies based on the recommendations in the
protocol.
J. Authorized laboratories will keep records of specimen pooling strategies implemented including type of
strategy, date implemented, and quantities tested, and test result data generated as part of the “Specimen
Pooling Implementation and Monitoring”. For the first 12 months from the date of their creation, such
records will be made available to FDA within 48 business hours for inspection upon request, and will be
made available within a reasonable time after 12 months from the date of their creation.
K. BioFire Defense, LLC, authorized distributors, and authorized laboratories using the BioFire COVID-19
Test will ensure that any records associated with this EUA are maintained until otherwise notified by FDA.
Such records will be made available to FDA for inspection upon request.
1
For ease of reference, the letter of authorization refers to “authorized laboratories” as follows: Testing of non-pooled specimens is limited to
laboratories certified under the Clinical Laboratory Improvement Amendments of 1988 (CLIA), 42 U.S.C. §263a, that meet the requirements to
perform high or moderate complexity tests, and similarly qualified U.S. Department of Defense (DoD) and non-U.S. laboratories. Testing of
pooled specimens is limited to DoD laboratories that meet the requirements to perform high complexity tests.
Clinical Summary
The clinical performance was evaluated by testing individual archived nasopharyngeal swab specimens collected
in transport media. The overall clinical performance including only individual clinical specimens is summarized in
Table 7. The studies contributing to this summary are detailed below.
Table 7. Summary of Individual Clinical Specimens Evaluated with the BioFire COVID-19 Test
PPA % NPA %
Agreement 29/30 96.7% 71/71 100%
95% CI [83.3-99.4%] [94.9-100%]
Positive Percent Agreement (PPA) was calculated as 100% x (TP / (TP +FN)). Negative Percent Agreement
(NPA) was calculated as 100% x (TN / (TN+FP)). Nine of out ten positive samples were Detected by the BioFire
COVID-19 Test and five out of five negative samples were Not Detected, resulting in 90% PPA and 100% NPA
(Table 8).
Positive specimens were re-tested individually on the BioFire COVID-19 Test. Single individual positive
specimens were combined with the negative specimens in pools of 5 and 8 specimens. Twenty pools of each
size were tested. Pooled test results were compared to individual test results to evaluate the effect of pooling on
SARS-CoV-2 detection. Results are shown in Table 10.
SARS-CoV-2 was detected by the BioFire COVID-19 Test in 20/20 (100% PPA) of the pools of 5 specimens and
in 19/20 (95% PPA) of the pools of 8 specimens. For the single 8-pooled sample run in which SARS-CoV-2 was
not detected, the positive specimen included in this pool had late amplification when tested individually and when
included in a pool of 5 specimens, indicating analyte levels near the Limit of Detection (LoD).
Table 11. SARS-CoV-2 Virus Test Results at 1× and 0.1× LoD for the BioFire COVID-19 Test
Concentration Tested Test Result (% Detection)
×LoD
Genomic Copies/mL TCID50/mL Detected Equivocal Not Detected
1× 3.3E+02 2.2E-02 20/20 (100%) 0/20 (0%) 0/20 (0%)
Table 12: Summary of LoD Confirmation Result using the FDA SARS-CoV-2 Reference Panel
Reference Materials Cross-
Specimen Type Product LoD
Provided by FDA Reactivity
SARS-CoV-2 5.4E+03 NDU/mL1 N/A2
NPS in transport medium
MERS-CoV N/A2 ND3
1
NDU: Nucleic acid amplification test (NAAT) Detectable Units
2
N/A: Not applicable
3
ND: Not detected
As of March 16, 2020, the majority of assay primers showed 100% homology to all available SARS-CoV-2
genome sequences. Four primers, across 2 of the 3 BioFire COVID-19 Test assays, showed single mismatches
to sequences of SARS-CoV-2. In the majority of these cases these were single isolate sequences. In both
assays the mismatches were single base pair differences. Such mismatches are well tolerated on the BioFire
platform and the assays are predicted to be reactive in the BioFire COVID-19 test despite such mismatches.
Table 13. Organisms Tested for Evaluation of BioFire COVID-19 Test in silico Cross-Reactivity
Recommended Organisms Additional Organisms
Human coronavirus 229E Parechovirus
Human coronavirus OC43 Corynebacterium diphtheria
Human coronavirus HKU1 Bacillus anthracis
Human coronavirus NL63 Moraxella catarrhalis
SARS-coronavirus Neisseria elongata
MERS-coronavirus Neisseria meningitidis
Adenovirus Pseudomonas aeruginosa
Human Metapneumonovirus (hMPV) Leptospira
Parainfluenza virus 1-4 Chlamydia psittaci
Influenza A & B Coxiella burnetii
Enterovirus Staphylococcus aureus
Respiratory syncytial virus Homo sapiens
Rhinovirus SARS-coronavirus
Chlamydia pneumoniae Coronavirus
Haemophilus influenza Recombinant SARSr-CoV
Legionella pneumophila SARS2
Mycobacterium tuberculosis SARS coronavirus ExoN1
Streptococcus pneumonia SARS coronavirus wtic-MB
Streptococcus pyogenes SARS coronavirus MA15
Bordetella pertussis SARS coronavirus MA15 ExoN1
Mycoplasma pneumoniae Bat Betacoronavirus SARS related virus
Pneumocystis jirovecii Coronaviridae
Candia albicans Coronavirinae
Pseudomonas aeruginosa
Staphylococcus epidermidis
Staphylococcus salivarius
All assays show 80% or greater homology to Bat coronavirus RaTG13 (accession: MN996532). In addition, the
SARS-CoV-2e assay shows greater than 80% homology to Pangolin coronavirus isolate MP789 (accession:
MT084071). It is unlikely that these isolates would be found in our sample matrix of nasopharyngeal swabs;
however, little is known about their potential to infect a human host, or their evolutionary relationship to SARS-
CoV-2. No other significant amplification of non-target sequences is predicted.
Table 14. Organisms Tested for Evaluation of BioFire COVID-19 Test Analytical Specificity
Assay Detections
SARS-CoV-2d
SARS-CoV-2a
SARS-CoV-2e
SARS-CoV-2
Organism ID Test Concentration
Test Result
SARS-CoV-2d
SARS-CoV-2a
SARS-CoV-2e
SARS-CoV-2
Organism ID Test Concentration
Test Result
Zeptometrix 5.10E+07
Adenovirus 7 (species B) 0/3 0/3 0/3 Not Detected
0810021CF TCID50/mL
Human
Zeptometrix
Metapneumovirus 1.78E+05 TCID50/mL 0/3 0/3 0/3 Not Detected
0810161CF
(hMPV)
Parainfluenza virus 1 BEI NR-48681 8.0E+05 TCID50/mL 0/3 0/3 0/3 Not Detected
Zeptometrix
Parainfluenza virus 2 1.10E+06 TCID50/mL 0/3 0/3 0/3 Not Detected
0810504CF
5.10E+07 TCID50/mL
Parainfluenza virus 3 BEI NR-3233 0/3 0/3 0/3 Not Detected
(7.0E+08 copies/mL)
Zeptometrix
Parainfluenza virus 4 1.70E+07 TCID50/mL 0/3 0/3 0/3 Not Detected
08010060BCF
Zeptometrix
Influenza A subtype H1 7.05E+04 TCID50/mL 0/3 0/3 0/3 Not Detected
0810036CFN
Zeptometrix 7.05E+04 TCID50/mL
Influenza A subtype H3 0/3 0/3 0/3 Not Detected
0810252CF (1.92E+08 copies/mL)
Zeptometrix
Influenza B 4.78E+06 TCID50/mL 0/3 0/3 0/3 Not Detected
0810239CF
Enterovirus species A NCPV 5.0E+08 TCID50/mL
0/3 0/3 0/3 Not Detected
(EV71) 0812215v (3.8E+08 copies/mL)
Enterovirus species B Zeptometrix 5.10E+07 TCID50/mL
0/3 0/3 0/3 Not Detected
(Echovirus 6) 0810076CF (1.10E+08 copies/mL)
Enterovirus species C 7.90E+05 TCID50/mL
ATCC VR-1023 0/3 0/3 0/3 Not Detected
(Coxsackievirus A17) (3.17E+06 copies/mL)
Enterovirus species D Zeptometrix
1.58E+06 TCID50/mL 0/3 0/3 0/3 Not Detected
(68) 0810237CF
Respiratory syncytial Zeptometrix
1.05E+06 TCID50/mL 0/3 0/3 0/3 Not Detected
virus 0810040ACF
Zeptometrix
Rhinovirus 1.26E+06 TCID50/mL 0/3 0/3 0/3 Not Detected
0810012CFN
Pneumocystis jirovecii
ATCC PRA-159 1E+07 CFU/mL 0/3 0/3 0/3 Not Detected
(PJP)
ATCC MYA-
Candida albicans 7.88E+08 CFU/mL 0/3 0/3 0/3 Not Detected
2876
Pooled human nasal
- - - - - -
wash b
a
The human coronavirus HKU1 used in this study was a previously collected clinical specimen. The concentration of virus in the sample was
estimated based on the results of a previously performed real-time PCR test.
b
Pooled nasal wash was not evaluated in this study; however, approximately 50 negative residual NPS samples were evaluated during the
clinical evaluation of the test, and no cross-reactivity of test assays to flora present in NPS samples was observed.
Substances listed below include endogenous substances that may be found in specimens at normal or
elevated levels (e.g., blood, mucus/mucin, human genomic DNA), medications, washes or topical applications
for the nasal passage, various swabs and transport media for specimen collection, and substances used to
clean, decontaminate, or disinfect work areas. The concentration of substance added to the samples was
equal to or greater than the highest level expected to be in NPS specimens.
None of the substances were shown to interfere with the RP2 Panel function and are not expected to interfere
with the BioFire COVID-19 Test. However, it was observed that exposure of samples to bleach prior to testing
could damage the organisms/nucleic acids in the sample, leading to inaccurate test results (lack of analyte
detection). The effect of bleach was dependent on the concentration and/or length of time the bleach was
allowed to interact with the sample.
Various commensal or infectious microorganisms typically found in NPS specimens were tested and did not
interfere with the performance of the RP2 Panel Internal Controls. These organisms have not been tested on
the BioFire COVID-19 Test but due to similarities in the internal control, they are not expected to interfere with
the BioFire COVID-19 Test Internal controls. See Table 15b for a list of competitive organisms tested.
Pooling Implementation
Pooling must only be performed by U.S. Department of Defense on individuals who are subjected to a detailed
infection prevention and control plan prior to and during operations, or by laboratories that can adhere to a full
protocol for ongoing monitoring of the pooling strategy per these Instructions for Use. Pooling of specimens
allows for testing of more individuals with fewer reagents. When resource availability is sufficient to meet testing
demand, laboratories should reconsider whether the risks of reduced test sensitivity with pooling continue to
outweigh the benefits of resource conservation. Pooling of specimens should also be considered in context of the
SARS-CoV-2 positivity rate within the test population. Higher positivity rates generally decrease the efficiency of
pooling samples because specimens in positive pools must be retested individually. The BioFire COVID-19 Test
has been validated for pooling up to eight samples.
Before implementing a pooling strategy, laboratories should determine the percent positivity rate of the testing
population and choose an appropriate pooling sample size that is within the maximum validated pool size of eight
samples.
Using historical data for individual specimens from the previous 7-10 days, the percent positivity rate (Pindividual)
can be determined by dividing the number of positive specimens by the total number of specimens tested during
that date range.
Refer to Table 16 to identify which pooling sample size provides the greatest testing efficiency for the determined
Pindividual within the validated pool sizes for the assay. If Pindividual is 2% or less, then the largest validated pool size
(n=8) should be used to maximize efficiency. If the Pindividual is greater than 25%, then pooling is not efficient and
should not be implemented. An example of the efficiency calculation for 5-sample pooling uses the formula F=1/
(1+ 1/5-(1-P)5), when P individual is 1%, the efficiency F is 3.46 for n=5. It means that 1,000 tests can cover
testing of 3,460 patients on average.
If historical data for individual specimens from the previous 7-10 days are not available for a laboratory as
described above, pooling may be implemented with the maximum pool size of (n=8). However, efficiency may not
be maximized if Pindividual has not been determined.
(Ppools) = (Number of positive specimens in pools / Total number of specimens tested in pools) * 100
For DoD Laboratories that Can Adhere to a Full Protocol for Ongoing Monitoring of the
Pooling Strategy
Continue to monitor n-sample pooling strategy by calculating the positivity rate among patient samples during n-
sample pooling (Ppools-x) for subsequent 7-10 day period based on n-sample pool testing. (Ppools-x) should be
updated daily using a moving average.
Compare Ppools-initial to Ppools-x. If Ppools-x is less than 90% of Ppools-initial. (Ppools-x < 0.90 × Ppools-initial), it is
recommended that:
Note: Individual testing as part of either re-assessment study option may be performed using a
different and higher throughput EUA COVID-19 test.
Option 1 Stop n-sample pooling and return to individual testing. Patient samples should be prospectively
individually tested until 10 consecutive positive samples have been collected. These individually tested
samples should then be re-tested in a pool with one positive and n-1 negative samples.
Option 2 Continue n-sample pooling. Individual testing should be performed in parallel to the pooled testing
until 10 consecutive positive samples are obtained. These positive samples should include both positive
individual results generated from individual testing of samples from the non-negative sample pools following
the n-sample pooling and deconvoluting workflow, and positive individual results obtained from individual
testing of samples from the negative sample pools for the time period. Because non-negative pools require
individual testing of samples included in the pool (samples in the positive pools will be tested as a part of
normal n-sample pooling workflow), the study essentially consists of additionally testing individual samples
from the pools with negative results.
If the PPA between pooled-testing results and individual-testing results is less than 90% then:
• If PPA ≤70% (7 out of 10), reduce the pool size (consider a new n as n-1)
• If PPA is 80% (8 out of 10), to compensate for lost sensitivity, reduce the pool size (consider a new n
as n-1) and continue with the reassessment testing until PPA of pooled compared to individual testing
is not less than 90%. OR collect an additional 10 consecutive individually positive samples. Then,
calculate the PPA from the combined data of 20 samples, between pooled-testing results and
individual-testing results. If the PPA is ≥ 85%, then implementation of testing using n-sample pooling is
acceptable.
• If PPA of at least 85% cannot be reached, cease pooling patient specimens.
This updated new positivity rate should be used as Pindividual in the future laboratory monitoring.
For DoD Operations Unable to Adhere to a Full Protocol for Ongoing Monitoring of the
Pooling Strategy
Individuals should be subjected to a detailed infection prevention and control plan prior to and during operations.
This may include for example: restriction of movement, quarantine, isolation, continuous health monitoring
programs and regular molecular SARS-CoV-2 surveillance testing by pooled or individual sample testing with the
BioFire COVID-19 or other authorized molecular SARS-CoV-2 testing.
Continue to monitor n-sample pooling strategy by calculating the positivity rate among patient samples during n-
sample pooling (Ppools-x) for subsequent 7-10 day period based on n-sample pool testing. (Ppools-x) should be
updated daily using a moving average.
Compare Ppools-initial to Ppools-x. If Ppools-x is less than 90% of Ppools-initial. (Ppools-x < 0.90 × Ppools-initial), pooling may
continue, but a new n-sample pooling size may need to be considered. If Ppools-x is greater than 25%, pooling of
patient samples is not efficient and should be discontinued until the percent positivity rate drops below.
Symbols Glossary
The following symbols can be found on labeling for the FilmArray 2.0, FilmArray Torch, and BioFire COVID-19
Test Kits, kit components, BioFire COVID-19 Test External Control Kits (+), and throughout accompanying
packaging.
ISO 15223-1
Graphical symbols for use on equipment – Registered Symbols
5.1.1 5.1.4 5.1.5
Use-By date Batch Code
Manufacturer (YYYY-MM-DD) (Lot Number)
United Nations Globally Harmonized System of Classification and Labeling of chemicals (GHS) (ST/SG/AC.10/30)
Exclamation Mark
Corrosive (Skin Hazardous to the
(Irritant, Acute Toxicity,
Corrosion/Burns, Eye aquatic environment,
Narcotic Effects,
Damage, Corrosive to long-term hazard
Metals) Respiratory Tract
Irritant)
81 FR 38911
Caution: Federal law restricts this device to sale by or on the order of a licensed healthcare
practitioner.
Contact Us by Fax
1-801-447-6907
© Copyright 2020, BioFire Defense, LLC All rights reserved. DFA2-PRT-0080-04, October 2020
The information contained in this document is subject to change without notice. No part of this document may be
reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the express
written permission of BioFire Defense, LLC.
FilmArray Software, Detector, and Metacall software modules © 2002–2020 BioFire Defense, LLC.
BioFire Defense, BioFire®, the BioFire logo, FilmArray® and LCGreen® are trademarks of BioFire Diagnostics, LLC
and/or BioFire Defense, LLC and are registered trademarks in the United States. All other names of products and
brands appearing in this manual are trademarks or registered trademarks of their respective owners.
The purchase of this product includes a limited, non-transferable license under specific claims of one or more U.S.
patents as listed on BioFire Defense’s website (http://www.biofiredefense.com/LegalNotices/) and owned by the
University of Utah Research Foundation and/or BioFire.
References