Perkin Elmer FT-NIR: Palm Oil Analyzer

The document discusses FT-NIR technology for analyzing palm oil and its derivatives. FT-NIR allows fast, simultaneous and non-destructive analysis of multiple parameters without sample preparation or chemicals. It has advantages over dispersive NIR including higher speed, sensitivity and accuracy due to its simultaneous collection of spectral data.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
166 views

Perkin Elmer FT-NIR: Palm Oil Analyzer

The document discusses FT-NIR technology for analyzing palm oil and its derivatives. FT-NIR allows fast, simultaneous and non-destructive analysis of multiple parameters without sample preparation or chemicals. It has advantages over dispersive NIR including higher speed, sensitivity and accuracy due to its simultaneous collection of spectral data.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 33

Perkin Elmer FT-NIR

Palm Oil Analyzer

Farisa Rizki

1 © 2009 PerkinElmer
Introduction: The Electromagnetic Spectrum

Increasing Wavelength

50,000 12,820 4000 400 cm-1

Gamma X-ray UV NIR IR FIR Micro Radio

200 380 780 2500 25 000 nm

Increasing Energy

K-shell electrons Outer-shell electrons Molecular vibrations Molecular Rotation


What is NIR?

NIR is fast prediction method, which able to


Fast do simultaneous parameter analysis in less
than two minutes

NIR method doesn’t need any sample


Free preparation and not uses any chemical
Chemical

NIR method is the easy and simple to


Easy to use for operator
Use
PerkinElmer Palm Oil Masterpiece
Portable IR early
MIR/NIR – Fertilizer
disease in detection

Oil palm
plantation
MIR/NIR – NIR-PCF, EFB,
margarine, soap, Cake, CPO
shortening, feed
HATR-Sludge

Oleochem, Perkin Elmer


specialty Palm Oil Palm oil
fat, biofuel mill
Masterpiece
SMP, SFC, TFM,
Glycerin, Moisture, VM, OC, FFA, MI
NaCl etc

Palm oil
bulking &
refinery

MIR/NIR – CPO, FFA, IV, MI, AV, PV,


RBD, SBE, OC

4
Spectrum Two N, FT-
NIR
 Plug-play solid and liquid accessory with
in seconds
 Compact, Rugged and Robust
 Full Software Suite : spectrum quant,
spectrum 10, and adulteration (option)
 Easy to use operational and calibration
software
 Heating liquid module
 High Humidity resistance (up to 75%)
 Wavelength 680 – 2500 nm
 No Consumable part

5
5
Spectrum Two Heatable Transmission Module

Transmission

Disposable cuvettes Analysis whole range palm oil Three heatable vial positions

No sample cross contamination Analysis of whole range palm oil On-board heatable sample
and increased throughput with liquid Decrease capital output holder
disposable vials  CPO, RBDPO, Olein, Stearin,  Heatable liquid sampling to
 Vials 8 mm for palm oil analysis PFAD*, PKO for parameter FFA, prevent solidify oil
IV, Moisture, AV*, PC*, DOBI**
 Reuse by solvent cleaning  Two preheating, while one
 Affordable analyzing
 Non contact accesory with main
instrument – safety

6
2. Sampling:
Near Infrared Reflection Module (NIRM)

Key features Analysis many types solid sample Removable spinner

 Highest sensitivity and  Press cake fiber : Oil & moisture  Heatable liquid sampling to
reproducibility - The premier  Sludge : Oil prevent solidify oil
choice for NIR reflectance  Two preheating, while one
sampling  Palm Kernel : Oil and moisture
analyzing
 Zero alignment  Bleaching earth*
 Non contact accesory with main
 Empty Fruit Bunch* instrument – safety

7
Types of Instrument Configurations

Frontier
with
NIRA

Perkin Elmer
Frontier
with NIR Palm Oil Spectrum
2N with Oil
accy Analyzer accy
Configuration

Spectrum
2N with
NIRA

8
The Instrument – Robust

No Consumable Part
Same Industry leading Spectrum 2
Dynascan Interferometer

Tungsten
halogen source

Detector
options:
Standard +
Enhanced
InGaAs

Gold Coated
Optics

Filter wheel as
standard
9 Confidence with over 5000 instruments shipped
AVI, Patented Technology of PerkinElmer
For Transferable Instrument callibration data

• HCl at 4cm-1 resolution


Without AVI After AVI

2912.0 2908 2904 2900.0 2912.0 2908 2904 2900.0


cm-1 cm-1

Fig 12: AVI correcting differences between instruments


10
Spectrum Touch software provides a simpler interface

• Spectrum 10 is the easiest to use


IR/NIR software today
• The Touch interface exploits latest
Windows touchscreen technology for
simple pushbutton solutions
Is FT-NIR Expensive?
ROI : 1 Year

Chemical and lab consumable Saving :


Saving 10 jt/ month ---- 120 jt / tahun

Increase Oil Productivity :


Increase 0.1% Oil productivity --- 350 jt/tahun

Less chemical waste $$$


Time Saving $$$
Less Labor dependent $$$
Summary Benefit of FT-NIR

 Fast (1 minute)
 Simultaneous multiparameter analysis
 No sample preparation
 Non-destructive
 Chemical / reagents free
 Safe and very simple to use
 Precise

13
Current IR User (Palm Oil)
Austral Edible Oil
Felda Kernel Oil
Yanmar
Sime Darby R&D centre
MPOB-AOTD
MPOB-Chemistry
Sarawak Oil Palm
Sarawak Oil Palm
Mewah group - MOI Food Sdn Bhd
Bintulu Biport Bulkers
Tanjung Manis Bulkers
Oleon klang
FGV Oleochemicals
Pacific Oleo Sdn Bhd
Henkel Oleochemicals Banting
Cognis Oleo
Nat Oleo
FPG Oleochemicals Sdn Bhd
Vance Bioenergy
FPG Kuantan
IOI Plantations Sdn Bhd
IOI Plantations Sdn Bhd
Mahamurni Plantations Sdn Bhd
Mahamurni Plantations Sdn Bhd
Mahamurni Plantations Sdn Bhd
Mahamurni Plantations Sdn Bhd
Kirana Refinery Sdn Bhd
1
4
Thank you
Farisa Rizki, MSc
Product Specialist, Indonesia
Comparisons of PerkinElmer FTNIR and
Dispersive Monochromator NIR
• Fourier Transform FTNIR Spectroscopy Technology
• All frequencies are measured simultaneously, single scan more rapid.
• No monochromator and slit system, higher energy throughput and higher S/N ratio.
• Laser wavelength is used, accurate wavenumber and excellent repeatability.
• No stray light issue.
• Resolution is constant across spectrum.
• No grating or filter changes, no discontinuities in the FTIR spectra.
• Re-calibration models are not necessary.
• Easy in software operation and calibration.
• Low maintenance and consumable costs, laser is changed every 5-10 years, source is changed
every 2-3 years.

• Dispersive Monocrhomator NIR Spectroscopy Technology


• Scan continuously and sequentially due to grating system, slower in single scan.
• Monochromator and slit system, low energy throughput and lower S/N ratio.
• Stray light issue due to monochromator is used.
• Slit system resulted variable resolution in spectrum.
• Discontinuities in spectra due to grating or filter changes.
• Complicated in software operation and calibration.
• Re-calibration models are needed at 3-6 month interval
• High maintainace and consumable costs, Tungsten lamp is replaced every 6 months
Press Cake Fiber – (Oil Content & Volatile Matter)

17
Sludge – NIRA (Oil Content & Volatile Matter)

18
CPO – NIR Glass Transmission (FFA & Moisture)

19
Free Fatty Acids Calibration for Crude Palm Oil

Fig 1 : Free Fatty Acids calibration curve for Crude Palm Oil

20
Iodine Value Calibration for Crude Palm Oil

Fig 2: Iodine Value calibration curve for Crude Palm Oil

21
Moisture Calibration for Crude Palm Oil

Fig 3: Moisture calibration curve for Crude Palm Oil

22
Anisidine Calibration for Crude Palm Oil

Outliers removed

Fig : Anisidine calibration curve for Crude Palm Oil

23
Dobi value of CPO by FTNIR

24
Oil Content Calibration Curve for Palm Kernel Expellar

Fig 6: Oil Content calibration curve for Palm Kernel Expellar

25
Moisture and Volatile Matter Calibration Curve for Palm Kernel Expellar

Fig 7: Moisture and Volatile Matter calibration curve for Palm Kernel Expellar

26
Additional Parameters
Debris, C12 and FAC Analysis with NIR

27
Debris by FTNIR

28
FAC Analysis with NIR

29
C12 calibration of CPO, Olein, Stearin, PKO and CPO-PKO blend by NIR

30
Full model calibration of CPO, Olein, Stearin, PKO and CPO-PKO blend by NIR

31
Full model calibration of CPO, Olein, Stearin, PKO and CPO-PKO blend by NIR

FAC Calibration of C12 FAC Calibration of C18

FAC Calibration of C14 FAC Calibration of C18:1

FAC Calibration of C16 FAC Calibration of C18:2


32
Sample prediction of Stearin, PKO and CPO-PKO blend model

Sample Name C12 C14 C18:0 C18.1 C18.2 C16


FAC-Stearin1 0.2 1.1 4.8 36.5 7.4 49.3
PKO 1-1 47.5 15.6 2.2 15.8 2.5 8.4
CPO10-PKO90-131-1 42.78 14.1 2.5 18.4 3.3 11.7

Fig : Model input (from GC data) for stearin, PKO and CPO10-PKO

Fig : Prediction for same samples of stearin, PKO and CPO10-PKO


33