Upgrade Instructions System: Upgrade VB30/39/40 To VB41

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
180 views30 pages

Upgrade Instructions System: Upgrade VB30/39/40 To VB41

Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 30

Sensation

Upgrade Instructions
System
Upgrade VB30/39/40 to VB41

Answers for life.


2 Document Version / Disclaimer / Copyright

Document Version
Siemens reserves the right to change its products and services at any time.
In addition, manuals are subject to change without notice. The hardcopy documents cor‐
respond to the version at the time of system delivery and/or printout. Versions to hard‐
copy documentation are not automatically distributed.
Please contact your local Siemens office to order a current version or refer to our website
http://www.healthcare.siemens.com.

Disclaimer
Siemens provides this documentation “as is” without the assumption of any liability under
any theory of law.
The content described herein requires superior understanding of our equipment and may
only be performed by qualified personnel who are specially trained for such installation
and/or service.

Copyright
“© Siemens, 2008” refers to the copyright of a Siemens entity such as:
Siemens Healthcare GmbH - Germany
Siemens Aktiengesellschaft - Germany
Siemens Shenzhen Magnetic Resonance Ltd. - China
Siemens Shanghai Medical Equipment Ltd. - China
Siemens Medical Solutions USA Inc. - USA
Siemens Healthcare Diagnostics Inc. - USA and/or
Siemens Healthcare Diagnostics Products GmbH - Germany

Sensation C2-023.892.02.01.02 Page 2 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Table of Contents 3

1 General information 4

1.1 General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4


1.1.1 Reason for the Update. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.2 Time for Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.3 Prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1.4 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4

2 Preparations 5

2.1 Preparations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1.1 General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2.1.2 Required data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5

3 Installation 9

3.1 HW and SW Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9


3.1.1 Status of SRS installation (check only!) . . . . . . . . . . . . . . . . . . . . . 9
3.1.2 Software installation CTAWP (ICS) and IRSmx2s . . . . . . . . . . . . . . 10
3.1.3 Software installation CTWP (IES) . . . . . . . . . . . . . . . . . . . . . . . . . . 18

4 Upgrade Completion 21

4.1 Additional work steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21


4.1.1 Completion of configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.1.2 BHC configuration ICS (CTAWP) . . . . . . . . . . . . . . . . . . . . . . . . . . 21
4.1.3 Tune-up (only for Upgrade VB30/VB39 to VB41) . . . . . . . . . . . . . 21
4.1.4 Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.1.5 Quality Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.1.6 Completion Test ICS (CTAWP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
4.2 Upgrade Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.1 Creating backup files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
4.2.2 Recovery DVD/CD creation for ICS and IES (CTAWP and CTWP). . . 24
4.2.3 Checking Recovery DVD's/CD's for CTAWP and CTWP . . . . . . . . . . 26
4.2.4 Informing the customer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
4.2.5 Update Installed Volume Database . . . . . . . . . . . . . . . . . . . . . . . . 27

5 Changes to previous version 28

6 List of Hazard IDs 29

© Siemens, 2008 C2-023.892.02.01.02 Page 3 of 30 Sensation


07.11 H IM CR CT SCM EN
4 1 General information

1.1 General information

 To prevent missing data, the customer needs to manually store reports that have been
created as well as all other data on the H-drive prior to installing the update.

 For safety reasons, the customer needs to archive all images and raw data prior to instal‐
ling the upgrade.

1.1.1 Reason for the Update


n Release of new software version VB41
- For more information, refer to the Software Release Notes

1.1.2 Time for Upgrade


The upgrade will take approx 7 hours.

1.1.3 Prerequisites
n SOMARIS/5 VB30/VB39/VB40 is running.
n ICS & IES new service key and licenses for VB41.
n Customer has archived all images and raw data.
n For the configuration of Virtual Simulation software (VSim) you need support
from Siemens OCS CSE or the Applications Specialist and the Physicist.

1.1.4 Ordering Information


There are 2 Upgrade Kits available:
1. Software Upgrade Kit: includes the VB41 software
2. Soft- and Hardware Upgrade Kit: includes the VB41 software and an IRSmx2s

Sensation C2-023.892.02.01.02 Page 4 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Preparations 2 5

2.1 Preparations

2.1.1 General Information


This document describes the upgrade of Somaris/5 VB30/39/40 to VB41.

 For safety reasons, the customer needs to archive all images prior to installing the up‐
date.

 To prevent missing data, the CSE needs to manually store the postprocessing applica‐
tions data on the H-drive for the customer prior to installing the update.

The customer needs to archive all raw data prior to installing the update, because the
 data will definitely be deleted during the software installation of the IRS. The IRS soft‐
ware has to be installed completely (boot from CD/DVD).

To ensure successful installation of the software, ensure the following:


1. There must be no restart pending due to changes made in the configuration. Open
LocalService and go to Configuration. Check whether a "restart pending" is shown in the
message bar. If this is the case, click HOME to start the missing automatic restart.

 2. All applications (e.g. Explorer, Patient Browser) must be closed and stay closed during
the entire update unless requested in the instructions.
3. Walk through the instructions step by step and do not try to save time by working out
your own "update strategy". This is especially important when installing the IES. WAIT
when it asks you to wait until the ICS is ready.

It may be helpful and save time if the license file and all DICOM nodes are noted prior to
the update.


n Open Windows Explorer
n Search for C:\somaris\service\html\report\report.htm, right-click on the file and use
Send to... to save to a floppy
n Copy the file license.dat in folder C:\Somaris\Config\Licensing also to a floppy.

2.1.2 Required data


The Navigator and Wizard have to be switched on.
Read out the following configuration data and enter them in the fields provided. These
data will not be restored. Additionally, you can store the file report.htm.

© Siemens, 2008 C2-023.892.02.01.02 Page 5 of 30 Sensation


07.11 H IM CR CT SCM EN
6 2 Preparations

2.1.2.1 ICS / CTAWP


Date and time settings
n From the syngo user interface, select Options -> Configuration.
n In the configuration panel, double-click the "Regional and Language Options" icon.
n Write down the following data:

Regional Options Language


Regional Options -> Cus‐ Time / Time Separa‐
tomize tor
Regional Options -> Cus‐ Date / Date Separa‐
tomize tor
Languages ->Details Input language

Service key
n Write down the service key.

Service key
Service key = ...................14 characters................... ....last 6 characters....
20 characters

Perform the following steps to ensure that all archiving, filming, copying, and networking
jobs on the Navigator are finished, and that the respective queues are empty:
n Open the Patient Browser.
n Click Transfer -> Select "Local Job Status" and, if necessary, delete all entries.
n Click Transfer -> Select "Network Job Status" and, if necessary, delete all entries.
n Click Patient -> Select "Film Task Status" and, if necessary, delete all entries.

BHC adaptation
n From the service software home menu, select Utilities.
n Under source, select Escape to OS and at the command select NT command Inter‐
preter.
n Under Parameters, type "Regedit" and press Enter.
n Under HKEY_LOCAL_MACHINE\SOFTWARE\Siemens\SOMARIS\Config\Site\Masme,
select BHC_XXKV,
for example, BHC_120KV, write down the number, e.g.: the V1 in registry is
KV120_STA

2.1.2.2 IES / CTWP (optional)


Date and time settings
n From the syngo user interface, select Options -> Configuration.
n In the configuration panel, double-click the "Regional and Language Options" icon.

Sensation C2-023.892.02.01.02 Page 6 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Preparations 2 7

n Write down the following data:

Regional Options Language


Regional Options -> Cus‐ Time/ Time Separa‐
tomize tor
Regional Options -> Cus‐ Date / Date Separa‐
tomize tor
Languages ->Details Input language

Service key
n Write down the service key.

Service key
Service key = ...................14 characters................... ....last 6 characters....
20 characters

Perform the following steps to ensure that all archiving, filming, copying, and networking
jobs on the Wizard are finished, and that the respective queues are empty:
n Open the Patient Browser.
n Click Transfer -> Select "Local Job Status" and, if necessary, delete all entries.
n Click Transfer -> Select "Network Job Status" and, if necessary, delete all entries.
n Click Patient -> Select "Film Task Status" and, if necessary, delete all entries.

2.1.2.3 2nd IES / CTWP (optional)


Date and time settings
n From the syngo user interface, select Options -> Configuration.
n In the configuration panel, double-click the "Regional and Language Options" icon.
n Write down the following data:

Regional Options Language


Regional Options -> Cus‐ Time / Time Separa‐
tomize tor
Regional Options -> Cus‐ Date / Date Separa‐
tomize tor
Languages ->Details Input language

Service key
n Write down the service key.

© Siemens, 2008 C2-023.892.02.01.02 Page 7 of 30 Sensation


07.11 H IM CR CT SCM EN
8 2 Preparations

Service key
Service key = ...................14 characters................... ....last 6 characters....
20 characters

Perform the following steps to ensure that all archiving, filming, copying, and networking
jobs on the Wizard are finished, and that the respective queues are empty:
n Open the Patient Browser.
n Click Transfer -> Select "Local Job Status" and, if necessary, delete all entries.
n Click Transfer -> Select "Network Job Status" and, if necessary, delete all entries.
n Click Patient -> Select "Film Task Status" and, if necessary, delete all entries.

Do not perform a backup now. The backup will be done at the beginning of the software

 installation.
You will be prompted to insert a blank CD for the backup.

Sensation C2-023.892.02.01.02 Page 8 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Installation 3 9

3.1 HW and SW Installation

3.1.1 Status of SRS installation (check only!)

CTAWP (ICS) and CTWP (IES)


n In the Local Service Home Menu, select Configuration.
n Go to Service -> SRS
n Click > to start the function check.
Fig. 1: SRS Function Check

n Write the result of the SRS function check into the table below.
n Click Finish and Home.
n Cancel the requested restart.

CTAWP/ICS CTAWP/ICS CTWP/IES CTWP/IES


Desired Actual Desired Actual
SRS Basic Function
Somatom Life
Utilization Management
System Management
Virus protection

© Siemens, 2008 C2-023.892.02.01.02 Page 9 of 30 Sensation


07.11 H IM CR CT SCM EN
10 3 Installation

SRS is correctly installed if the status of the Desired and Actual checkboxes are the same


--> SRS should also work after the upgrade due to backup/restore!
If the checkboxes differ, an error message is output --> SRS will not work, SRS trouble‐
shooting is required after the upgrade.

3.1.2 Software installation CTAWP (ICS) and IRSmx2s


Software installation for CTAWP (ICS)
n Shut down and switch off the CTWP (IES)
n Insert the DVD_Som/5 _ICS/IES_VB41A_1 of 2
n Select Options -> Service -> Local Service, and enter the ServiceKey.
n In the service software menu, select Utilities.
n Under Source, select Escape to OS and at "Command" select NT Command Interpret‐
er.
n Under Parameters, enter R:\sinstall.bat
n Activate with Return or GO.
 The "Silent Installation" starts automatically when the message appears:
"Please insert CD for Backup and press OK.", take out the DVD_Som/5 _ICS/
IES_VB41A_1 of 2 and insert a blank CD for the backup and click OK." Backup
is running. "Please wait ..." will appear.

 If the backup was not successful (e.g. a bad CD), the message "Error occurred with Silent
Installation" will come up. You have to begin the installation again with R:\sinstall.bat

n After backup is finished, the message "Please insert the OEMSetup CD..." is dis‐
played.
n Insert the DVD “Som/5_ICS/IES_VB41A_1 of 2”, and press the OK button.
 System reboots from DVD.
n Select “Restore the systemdrive” using the drop-down menu in the window, and
click OK
n In response to the message "The 1st Harddisk will be completely erased!!! Do you
want to continue?", click Yes. In response to the message "Please insert the DVD for
Restore and click OK" pop-up, click OK.
 GHOST is running.
n When the message “Symantec Ghost needs to open the next part of the image.
Please insert the next disk and click ok,...” pop up, please insert DVD “Som/5_ICS/
IES_VB41A_2 of 2”, and then click “OK” button.
n Wait until the DVD rejects automatically. At the same time, a message "Restore of the
system is finished successfully ... ..." pops up. Take out the DVD and click OK. The
installation will start automatically.
n If the window “Startup Error -> Failed to start process: pdfcamp-registry” appears,
press OK

Sensation C2-023.892.02.01.02 Page 10 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Installation 3 11

n After a successful software installation the message "Please insert the Backup CD for
Restore and press OK" displays.

 The message window may be behind the Internet Explorer.

n For systems with IRS3 continue with ( IRSmx2 Installation (only for systems with
IRS3) / Page 11).
For systems with IRSmx2s continue with ( Software Update IRSmx2s / Page 18).

3.1.2.1 IRSmx2 Installation (only for systems with IRS3)

 The IRS installation is only necessary for systems with IRS3. The IRS3 must be exchanged
for an IRSmx2s.

 Switch off the system before exchanging the IRS.

Removing the old IRS3

1. Shut down and switch off the system.

© Siemens, 2008 C2-023.892.02.01.02 Page 11 of 30 Sensation


07.11 H IM CR CT SCM EN
12 3 Installation

2. If necessary remove the IRS3 mountings.


Fig. 2: IRS mounting to prevent tipping

(A) Front view


(B) Top view
(C) Side view
(1) Holding bracket (L-bracket) for securing to the wall. Mount it down or up alterna tively.
(2) Spacer for mounting front side
(3) Spacer for holding bracket(Z-bracket) rear side
(4) Drilling for pos.2)
(5) Z-bracket
(6) Wall spacer (mounted on the cabinet)
(7) Wall
3. Remove the cables from the IRS3.

Cable connections at Cable ID


IRS power supply W8
IRS: X101 (PE) W11
G Rec D24 W55 (fiber-optic cable)
IRS: IRSTCR D32 / X1 W75
IRS switch (Ethernet cable W62
at Port 1)

4. Remove the IRS3 from the installation location.

Sensation C2-023.892.02.01.02 Page 12 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Installation 3 13

Installing the new IRSmx2s

Transporting
Move the IRS on its pallet near its final installation location.

 Do not move the IRS over uneven sections of the floor. Vibrations can damage the IRS.

Unpacking the IRS

1. Remove all nails and fasteners with the pry bar if necessary.
2. Remove the top and sides of the wooden crate if necessary.

Checking for transport damage


After removing the transport crate, examine the IRS for transport damage, e.g., due to
moisture or external forces.

 In case of transport damage, inform the local project manager immediately.

Removing the IRS from the pallet


n Adhere to the instructions hanging from the side of the IRS.
Fig. 3: IRSmx2/3 delivery state

© Siemens, 2008 C2-023.892.02.01.02 Page 13 of 30 Sensation


07.11 H IM CR CT SCM EN
14 3 Installation

Assembling the IRS at its installation location

The IRS is air-cooled by internal ventilators. Ensure unimpaired circulation from the in‐


take and outlet. For correct cooling, a minimum distance of 100 mm at the back and 150
mm at the front between the IRS and other objects is required!
Never put the IRS into a closed cabinet; otherwise, it will overheat!

 For the exact installation location, refer to the project plan or ask the on-site project
manager. The IRS can also be installed in a room adjacent to the examination room.

n For the installation, fold down the spacers at the back of the IRS. This is necessary to
maintain the minimum distance to the wall.
Fig. 4: IRSmx-towers: Spacers and rollers

1 2
(1) Spacers with rollers UP; transporting position
(2) Spacers with rollers DOWN; installation position

IRS cabling
Power cable preparation:
1. Connect the adapter cable (delivered with the upgrade kit) to the IRS cable W8.
Fig. 5: Connecting the IRS adapter cable

PE cable preparation:
1. Remove the cable shoe from the IRS PE cable W11.

Sensation C2-023.892.02.01.02 Page 14 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Installation 3 15

2. Use a crimping tool to attach the cable shoe (delivered with the upgrade kit) to the IRS
PE cable W11.

IRSmx2s cable connections:


Connect the following cables to the IRSmx2s.

Cable connection at IRS Cable ID Cable routes to


AC Power In (power supply) W8 PDC: X254
X100 (PE) W11 PDC: X100 (PE)
D24 - FO from Gantry W55 (fiber-optic cable) Gantry left (direct connec‐
(GREC, fiber-optic port) tion), CT Gantry D23 / X1
D32.X1 (CAN bus) W75 D301 X27
LAN#1 - LAN to ICS (Ether‐ W62 ICS (twisted pair crossover)
net port)

© Siemens, 2008 C2-023.892.02.01.02 Page 15 of 30 Sensation


07.11 H IM CR CT SCM EN
16 3 Installation

Fig. 6: IRSmx2s (8871910)

(1) D24
(2) D32.X1
(3) LAN to ICS
(4) Power
(5) PE


After connecting the IRS power cable, attach the electric shock protection and switch on
the IRS main power switch at the back ( Fig. 7 Page 17).

Sensation C2-023.892.02.01.02 Page 16 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Installation 3 17

Fig. 7: IRSmx2/3 electric shock protection and power switch

1 2

(1) Electric shock protection


(2) Main power switch

Measuring the protective conductor

 Measure the protective conductor meter as described in the following product-specific


contents.

n Make sure that the measurement instrument is calibrated according to manufacturer's


specifications.
n Use a protective conductor meter to perform the test.
n To ensure correct results, unroll the test cable completely from the test device prior to
performing the test. Check to make sure the zero setting is correct.
n Measure the protective conductor and resistance.

Checking the protective conductor


IRSmx2s: protective conductor resistance ≤ 300 mΩ
n From: Protective ground terminal in on-site power distribution.
To: Uncoated metallic part.

Switching on the System


Switch on the system on the control box.

 The CTWP (IES) must be switched off!

© Siemens, 2008 C2-023.892.02.01.02 Page 17 of 30 Sensation


07.11 H IM CR CT SCM EN
18 3 Installation

3.1.2.2 Software Update IRSmx2s

Check (via console or Options->Version) if VB41 is already installed at the delivered


IRSmx2s. If necessary update the IRS Software to VB41.
For more information about the IRS console refer to the document “Troubleshooting
Guide CT02-023.840.04.xx.02”

Software Update for IRSmx2s


n Insert the DVD SOM/5_IRS_VB41A_V1.0 into the CD drive and close it.
n The software update starts automatically.
n After about 15 minutes, the IRS update is finished.
 Check the update status via IRS console.
n After the IRS software update perform a system restart. Select System -> End -> Re‐
start System.

3.1.2.3 Restore CTAWP (ICS) and IRSmx2s


n When the message "Please insert the Backup CD for Restore and press OK" displays
at the ICS , insert the backup CD, and click OK.
n After the restore is finished, the ICS starts to reboot automatically.

 The message window may be behind the Internet Explorer.

 If a restore step fails, perform a manual restore of this step at the local service.

Settings for CTAWP (ICS)


1. Insert the new license CD.
2. Open the Local Service Home menu. Input the new Service Key.
3. Go to Configuration -> Service -> Licensing -> Select New License Files.
4. Select New License File under Browse to select the license file.
5. Click Save and Finish.

3.1.3 Software installation CTWP (IES)

If the system is equipped with IES Tower 8D or 8E, DVD_Som/5_IES_VB41_1of2_64bit

 and DVD_Som/5_IES_VB41_2of2_64bit are needed.


For all other IES Towers, DVD Som/5_ICS/IES_VB41_1 of 1_32bit is needed.

Sensation C2-023.892.02.01.02 Page 18 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Installation 3 19

Software installation for CTWP (IES)


n Switch on the CTWP (IES).
n Insert the DVD_Som/5 _IES_VB41A_1 of 2.
n Select Options -> Service -> Local Service, and enter the ServiceKey.
n In the service software menu, select Utilities.
n Under Source, select Escape to OS and at "Command" select NT Command Interpret‐
er.
n Under Parameters, enter R:\sinstall.bat
n Activate with Return or GO.
 The "Silent Installation" starts automatically when the message appears:
"Please insert CD for Backup and press OK.", take out the DVD_Som/5
_IES_VB41A_1 of 2and insert a blank CD for the backup and click OK." Backup
is running. "Please wait ..." will appear.

 If the backup was not successful (e.g. a bad CD), the message "Error occurred with Silent
Installation" will come up. You have to begin the installation again with R:\sinstall.bat

If you miss inserting the requested CD on time in the following work steps, a warning is
displayed:

 No package has been found. Please insert the CD "........" in the CD drive to continue the
installation.
Insert the requested CD and press any key to continue.

n After backup is finished, the message "Please insert the OEMSetup CD..." is dis‐
played.
n Insert the DVD “Som/5_IES_VB41A_1 of 2”, and press the OK button.
 System reboots from DVD.
n Select “Restore the systemdrive” using the drop-down menu in the window, and
click OK
n In response to the message "The 1st Harddisk will be completely erased!!! Do you
want to continue?", click Yes. In response to the message "Please insert the DVD for
Restore and click OK" pop-up, click OK.
 GHOST is running.
n When the message “Symantec Ghost needs to open the next part of the image.
Please insert the next disk and click ok,...” pop up, please insert DVD “Som/
5_IES_VB41A_2 of 2”, and then click “OK” button.
n Wait until the DVD rejects automatically. At the same time, a message "Restore of the
system is finished successfully ... ..." pops up. Take out the DVD and click OK. The
installation will start automatically.
n If the window “Startup Error -> Failed to start process: pdfcamp-registry” appears,
press OK
n When the message "Please insert the Backup CD for Restore and press OK" displays,
insert the backup CD, and click OK.

© Siemens, 2008 C2-023.892.02.01.02 Page 19 of 30 Sensation


07.11 H IM CR CT SCM EN
20 3 Installation

n After the restore is finished, the IES starts to reboot automatically.

 The message window may be behind the Internet Explorer.

Settings for CTWP (IES)


n Insert the new license CD.
n Open the Local Service Home menu. Input the new Service Key.
n Go to Configuration -> Service -> Licensing -> Select New License Files.
n Select New License File under Browse to select the license file.
n Click Save and Finish.

Restart the CTWP (IES).


n Select System -> End -> Shutdown the system.
n Switch off the system and wait 2 min.
n Switch on the system.

Sensation C2-023.892.02.01.02 Page 20 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Upgrade Completion 4 21

4.1 Additional work steps

4.1.1 Completion of configuration


Complete the configuration described in the Document “Configuration of SOMARIS/5
VB30/38/39/40/VB41 CT00-000.843.07.xx.02”.

4.1.2 BHC configuration ICS (CTAWP)


n From the service software home menu, select Utilities at the ICS.
n Under Source, select Escape to OS and at "Command" select NT Command Interpret‐
er.
n Under Parameters, type “Regedit”, and press Enter.
n Under HKEY_LOCAL_MACHINE\SOFTWARE\Siemens\SOMARIS\Config\ Site\Masme
\BHC_XXXKV, right click V1, and select Modify. In the windows that are displayed,
change the related number along with the record.
Fig. 8: BHC parameter

n Under C:\Somaris\Tools\Bin, double-click the BHC_configure.exe file.


n Click Start.

4.1.3 Tune-up (only for Upgrade VB30/VB39 to VB41)


Perform the following tune-up steps:
n Open the Local Service Home menu and go to TuneUp.
n Select Expert Mode.
n Select and execute the following steps:
- Channel correction, select all.
- Air calibration, select all (Basecal + Diffcal).
- Water scaling, select all.

© Siemens, 2008 C2-023.892.02.01.02 Page 21 of 30 Sensation


07.11 H IM CR CT SCM EN
22 4 Upgrade Completion

n Click Go and follow the software instructions.

4.1.4 Tests

4.1.4.1 Tune CD burning


n In the service software menu, select Utilities.
n Under Source, select Tune CD Burning.
n Activate checkbox Tune CD Burning for current hardware.
n Activate with Go.
n Wait until message “Driver has been successfully installed.” appears and click OK.

4.1.4.2 Software self-test ICS and IES (CTAWP and CTWP)

 Before performing the software self test at the CTWP, “Tune CD Burning” has to be per‐
formed.

Perform the software self-test at the CTAWP and CTWP.


n Open the Local Service Home menu and go to Test Tools.
n Click the Software SelfTest and execute all subtests.
 In case of errors, refer to the hints for troubleshooting.

4.1.5 Quality Measurement

4.1.5.1 Constancy reference measurement


n Open the Local Service Home menu and go to Quality.
n Select Constancy Reference.
n Select Reference.
n In Constancy Reference you have to perform all steps.

4.1.6 Completion Test ICS (CTAWP)


Perform the Completion Test at the CTAWP.
n Open the Local Service Home menu and go to Test Tools.
n Click the Completion Test and click Go.

Sensation C2-023.892.02.01.02 Page 22 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Upgrade Completion 4 23

4.2 Upgrade Completion

4.2.1 Creating backup files

4.2.1.1 Backup at ICS / CTAWP


n Start the service software and select the Backup & Restore function in the main
menu.
n Insert the backup CD-R for the CTAWP.
n From Command, select Backup.
n From Drives, select the CD drive, e.g., [-R-] CD-R.
n From Packages, select CT-Settings-PXXX and click Go.
 BackupRestore Running ...
 Wait until Ready is displayed.
n From Packages, select SW-Settings02-PXXX and click Go.
 BackupRestore Running ...
 Wait until Ready is displayed.
n From Packages, select Security Settings PXXX and click Go.
 BackupRestore Running ...
 Wait until Ready is displayed.
n From Packages, select VSim (optional) and click Go.
 BackupRestore Running ...
 Wait until Ready is displayed.
n When "Ready," remove the backup medium.
n Click Home and terminate the service software by closing the browser window.

The backup files have been saved correctly when ok is appended to the end of the back‐

 up filename,
e.g. <hostname>-SW-Settings02-PXXX-date_time_ok.ar

4.2.1.2 Backup at IES / CTWP (optional)


n Start the service software and select the Backup & Restore function in the main
menu.
n Insert the backup CD-R for the CTWP.
n From Command, select Backup.
n From Drives, select the CD drive, e.g., [-R-] CD-R.

© Siemens, 2008 C2-023.892.02.01.02 Page 23 of 30 Sensation


07.11 H IM CR CT SCM EN
24 4 Upgrade Completion

n From Packages, select CT-Settings-CTWP and click Go.


 BackupRestore Running ...
 Wait until Ready is displayed.
n From Packages, select SW-Settings02 and click Go.
 BackupRestore Running ...
 Wait until Ready is displayed.
n From Packages, select Security Settings CTWP and click Go.
 BackupRestore Running ...
 Wait until Ready is displayed.
n From Packages, select VSim (optional) and click Go.
 BackupRestore Running ...
 Wait until Ready is displayed.
n From Packages, select InSpace (optional) and click Go.
 BackupRestore Running ...
 Wait until Ready is displayed.
n When "Ready," remove the backup medium.
n Click Home and terminate the service software by closing the browser window.

The backup files have been saved correctly when ok is appended to the end of the back‐

 up filename,
e.g. <hostname>-SW-Settings02-PXXX-date_time_ok.ar

4.2.2 Recovery DVD/CD creation for ICS and IES (CTAWP and CTWP)

According to regulations, a recovery DVD/CD has to be created after the installation is


 finished. Also, please take the original VB40 software DVD´s/CD´s back and do not leave
them at the customer's site.

CD_SOM/5_ICS/IES_VB41A_32bit_BARE &CD_SOM/5_ICS/IES_VB41A_64bit_BARE
are the startup CDs which are used for recovery DVD/CD installation, creation, and check‐
ing.

 If the system is equipped with CTWP Tower 8D/8E, 10D/10E, and IES 11D/11E, CD_SOM/
5_IES_VB41A_64bit_BARE and Somaris_Recovery DVD/CD (for CTWP) are needed.
For all other CTWP Towers:
CD_SOM/5_ICS/IES_VB41A_32bit_BARE & Somaris_Recovery_DVD/CD (for CTWP) are nee‐
ded.

Sensation C2-023.892.02.01.02 Page 24 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Upgrade Completion 4 25

Create a Recovery DVD/CD for CTAWP and CTWP.


n Insert the correct CD:
- For CTWP Tower 8D/8E, select CD_SOM/5_IES_VB41A_64bit_BARE.
- For the other CTAWP/CTWP Tower, select CD_SOM/5_ICS/
IES_VB41A_32bit_BARE.
n Select System -> End -> Restart system and boot from CD.
n The following pop-up window appears.
Fig. 9: Backing up the system

n Select Backup system, and press OK.


n The following pop-up window appears.
Fig. 10: Insert a blank DVD/CD

 You can create recovery DVDs or recovery CDs.

n Remove the BARE CD and insert an empty DVD/CD. Wait for 15 seconds, and press OK.
 An image of the C: partition is created.
n Insert the next DVD/CD when you are asked for it. Wait 15 seconds and click OK.
n When the recovery process is finished, please remove the DVD/CD and press OK to re‐
boot the system.

© Siemens, 2008 C2-023.892.02.01.02 Page 25 of 30 Sensation


07.11 H IM CR CT SCM EN
26 4 Upgrade Completion

n Label the recovery DVDs/CDs with the following information: Tower part no., revision
no., software version, system serial no. and so on.

4.2.3 Checking Recovery DVD's/CD's for CTAWP and CTWP


n Insert the correct CD:
- For CTWP Tower 8D/8E, select CD_SOM/5_IES_VB41A_64bit_BARE.
- For the other CTAWP/CTWP Tower, select CD_SOM/5_ICS/
IES_VB41A_32bit_BARE.
n Select System -> End -> Restart system and boot from CD.
n The following pop-up window appears.
Fig. 11: Check recovery DVD/CD

n Select Check recovery DVD/CD, and click OK.


n Insert the first recovery DVD/CD. Wait for 15 seconds, and click OK.
n The Ghost window appears. System is checking the first recovery DVD/CD. Insert the
next recovery DVD/CD when you are asked for it.
- If the message “The recovery DVD is OK! Please ...” appears, it means that the re‐
covery DVDs/CDs are OK.
- If the message “Cannot read from file & The recovery DVD has some errors!...”
appears, you need to create new recovery DVDs/CDs.
n Remove the recovery DVD/CD and click OK. The system will reboot automatically.
n Leave the Recovery DVDs/CDs at the customer site and take back the original
VB40 software DVDs.

4.2.4 Informing the customer


n Inform the customer that the update is complete.
n Update the Customer Documentation.

Sensation C2-023.892.02.01.02 Page 26 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
Upgrade Completion 4 27

n Take the new/additional customer documentation and file it with the customer’s docu‐
mentation.

4.2.5 Update Installed Volume Database


n IVK data must always be current.
n Update the installed volume database with the new items of this upgrade.
Relevant items are:
- SW License
- IRS

© Siemens, 2008 C2-023.892.02.01.02 Page 27 of 30 Sensation


07.11 H IM CR CT SCM EN
28 5 Changes to previous version

Page Change Reason


All Initial version Change No.: 556527

Sensation C2-023.892.02.01.02 Page 28 of 30 © Siemens, 2008


07.11 H IM CR CT SCM EN
List of Hazard IDs 6 29

There are no Hazard IDs in this document.

© Siemens, 2008 C2-023.892.02.01.02 Page 29 of 30 Sensation


07.11 H IM CR CT SCM EN
All documents may only be used for
rendering services on Siemens
Healthcare Products. Any document
in electronic form may be printed
once. Copy and distribution of elec‐
tronic documents and hardcopies is
prohibited. Offenders will be liable
for damages. All other rights are re‐
served.

healthcare.siemens.com/services

Siemens Healthcare Headquarters


Siemens Healthcare GmbH
Henkestr. 127
91052 Erlangen
Germany
Telephone: +49 9131 84-0
siemens.com/healthcare

Print No.: C2-023.892.02.01.02 | Replaces: n.a.


Doc. Gen. Date: 07.11 | Language: English
Part No.: 07543015
© Siemens Healthcare GmbH, 2008

siemens.com/healthcare

You might also like