ASCL LAB Ejemplos en Pureza
ASCL LAB Ejemplos en Pureza
NOTE ASCLD/LAB customers may find value in using this guidance document to better understand
and apply AL-PD-3060 ASCLD/LAB Policy on Measurement Uncertainty.
Document History / AL-PD-3063
Description of Activity or
Date Version Approved By Effective Date
Revision
ASCLD/LAB Executive
May 22, 2013 1.0 Reviewed and approved for distribution May 22, 2013
Director
All intellectual property rights in this publication are the property of the
American Society of Crime Laboratory Directors / Laboratory Accreditation Board ASCLD/LAB
ASCLD/LAB
139 J Technology Drive
Garner, North Carolina 27529
USA
919-773-2600
Purpose........................................................................................................................................................................... 5
Scope .............................................................................................................................................................................. 5
The NIST 8-Step Process for Estimating and Reporting Measurement Uncertainty ....................................................... 6
EXAMPLE 1, Step 6: Expand the combined standard uncertainty by coverage factor (k) .................................. 18
EXAMPLE 2, Step 6: Expand the combined standard uncertainty by coverage factor (k) .................................. 30
EXAMPLE 3, Step 6: Expand the combined standard uncertainty by coverage factor (k) .................................. 51
To provide guidance to laboratories that must achieve compliance with the ASCLD/LAB Policy on Measurement
Uncertainty 1 as they prepare for and maintain accreditation under the ASCLD/LAB-International Testing Program.
Scope
• Use of the NIST 8-Step Process for Estimating and Reporting Measurement Uncertainty 2 for:
o a measurement of the weight of controlled substance evidence
o a measurement of the volume of controlled substance evidence
o a measurement of the quantity of a controlled substance when reported as a weight or volume fraction
(purity) of the whole
• This general framework conforms to the principles in the Guide to the Expression of Uncertainty in
Measurement (GUM)3
• Laboratories that are accredited or that are seeking accreditation under the ASCLD/LAB-International Testing
Program in the discipline of Drug Chemistry,
• Technical assessors for the ASCLD/LAB-International program, and
• Users of ASCLD/LAB-International accredited laboratory services.
Denotes a definition. The definition will be from the VIM 4 unless otherwise noted
Denotes a requirement from the cited accreditation requirement document. Accreditation documents include, but
are not limited to: ISO/IEC 17025:2005, 5 ASCLD/LAB-International Supplemental Requirements for the Testing 6
program, ASCLD/LAB Policy on Measurement Uncertainty and ASCLD/LAB Board Interpretations 7. (A complete
listing of accreditation requirements is available at www.ascld-lab.org.)
Portions of the ASCLD/LAB Guidance on the Estimation of Measurement Uncertainty - Overview8 and the
ASCLD/LAB Guidance on the Estimation of Measurement Uncertainty – ANNEX A Details on the NIST 8-Step
Process 9 have been included in this document but the reader is encouraged to refer to both of these documents for
additional information.
Per the ASCLD/LAB Policy on Measurement Uncertainty, applicant and accredited laboratories in the Drug Chemistry
Discipline shall have and shall apply a procedure to estimate the uncertainty of measurement when values are
reported for the quantity (weight or volume) of controlled substance evidence or the quantity of a controlled substance
when reported as a weight or volume fraction (purity) of the whole. This guidance document will use the NIST 8-Step
Process to work through an example estimation of measurement uncertainty for each type of measurement. These
examples are meant to be illustrative only. The NIST 8-Step Process is a tool that can be used by a laboratory to
meet accreditation requirements but a laboratory is not required to use this process. ASCLD/LAB-International
The NIST 8-Step Process for Estimating and Reporting Measurement Uncertainty
The steps in the process are:
The SWGDRUG Supplemental Document SD-3 for Part IVC – Quality Assurance/Uncertainty Measurement
Uncertainty for Weight Determinations in Seized Drug Analysis 10 has been referenced for this example. Readers
are encouraged to review the SWGDRUG document in its entirety.
Measurement Traceability:
The traceability for this measurement process is established through the calibration of the measuring equipment,
the balance, and through the mass reference standards used to confirm the continued calibration status of the
balance.
• The calibration of the balance was determined to have a significant effect on the accuracy or validity of the
test result.
o The external calibration of all balances is performed by a calibration laboratory that meets Section 3.2.3
of the ASCLD/LAB Policy on Measurement Traceability 11
o The mass reference standards are calibrated by an external calibration laboratory that meets Section
3.2.3 of the ASCLD/LAB Policy on Measurement Traceability
Measurement Assurance:
• The laboratory determined that intermediate checks are needed to maintain confidence in the calibration
status of the balances during the interval between external calibrations. These checks are carried out
according to a defined schedule and procedure (ISO/IEC 17025:2005, Clause 5.5.10).
o The laboratory procedure includes the use of calibrated mass reference standards
• Data is available from both an initial study and ongoing quality control that evaluates the variation in the
measurement process using a check standard 12 surrogate. All analysts participate. All balances are used as
are all methods of weighing allowed by the test method procedure. The check standard used is a powder
material, packaged for stability that is sensitive to environmental changes in the laboratory.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(a) Statement defining the measurand
Quantity (VIM 1.1): property of a phenomenon, body, or substance, where the property has a magnitude that can
be expressed as a number and a reference
The statement defining the measurand can be a written statement, a visual diagram and/or a mathematical
expression. To be clear about what measurement process the estimation is being calculated for, it is important to be
as specific as possible when defining the measurand. To distinguish one measurement process from another in a
laboratory it may be necessary to include a reference to a specific type of equipment used, a specific procedure, etc.
in the statement defining the measurand.
Drug Chemistry - Weight Determination Using Balance Model 123; Procedure No. XXXX, Revision ZZZZ
Weight Determination
• Weight determination of a solid (non-liquid) material using a balance with a readability of 0.01 gram.
• Range of measurement: laboratory determined minimum balance load to maximum balance load
The measurement of weight is not made directly. Weight is determined through a functional relationship based on
the amount of force on the balance. The functional relationship can be expressed by the mathematical equation:
y = mx + b +/- U
Where,
y = the measurement result
m = slope or sensitivity of the measurement instrument linearity
x = the indication
b = bias; in this example, the scale error determined during the external calibration process or during
the intermediate checks, is a systematic bias in the measurement
U = expanded uncertainty
y can further be expressed mathematically as a function of the influences on the measurement process
identified by a laboratory.
ISO/IEC 17025:2005, Clause 5.4.6.2 states “… the laboratory shall at least attempt to identify all the components
of uncertainty…”
ISO/IEC 17025:2005, Clause 5.4.6.3 states “when estimating the uncertainty of measurement, all uncertainty
components which are of importance in the given situation shall be taken into account …”
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(b) Statement of how traceability is established for the measurement
(c) The equipment (e.g., measuring device[s] or instrument[s]) used
(d) All uncertainty components considered
ASCLD/LAB Policy on Measurement Uncertainty, Clause 5.3.2 requires that the specific measuring device or
instrument used for a reported test result to have been evaluated in the estimation of measurement uncertainty for
that test method.
Traceability is established for this measurement through the calibration of the balance(s), the measuring
equipment, used to perform the measurement and through the mass reference standards used to confirm the
continued calibration status of the balance(s).
The equipment (e.g., measuring device[s] or instrument[s]) used in this test method is:
Top-loading balances Model 123; readability of 0.01 g
NOTE: Residual sample not transferred to the weighing vessel during the performance of the test method to
obtain a weight is not included as a component of uncertainty in the measurement.
A laboratory may identify different uncertainty components when an evaluation of their specific measurement
process is performed.
This list of potential uncertainty components could also be compiled into a fishbone diagram or into any other format of
the laboratory’s choosing. Two fishbone diagram examples depicting possible uncertainty components and one blank
fishbone diagram are available on the ASCLD/LAB website at www.ascld-lab.org. Appendix A to this document
provides an example of using an Excel spreadsheet (available at www.ascld-lab.org) to display the information from
the NIST 8-Step Process.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(e) All uncertainty components of significance and how they were evaluated
(f) Data used to estimate repeatability and/or reproducibility
(g) All calculations performed
Review the list of uncertainty components created in Step 2. Identify which of the components listed can be quantified
by existing data from the measurement process.
The GUM (2.3.2) refers to this method of analysis of one or more uncertainty components as:
Type A evaluation (of uncertainty): method of evaluation of uncertainty by the statistical analysis of series
of observations
The quantification of the remaining uncertainty components is performed by a different method of evaluation referred
to by the GUM (2.3.3) as:
Type B evaluation (of uncertainty): method of evaluation of uncertainty by means other than the statistical
analysis of series of observations
Data is available from both an initial study and ongoing quality control that evaluates the variation in the
measurement process using a check standard surrogate that reasonably mimics the weight determination
process. All balances are used as are all methods of weighing allowed by the test method procedure. All analysts
participate. The laboratory has data spanning the work day, all days of the work week, and normal laboratory
environmental conditions over greater than a one year time period.
∑ (x )
n
2
i −x
s= i =1
n −1
The standard deviation of the reproducibility data in this example = 0.031 g
The purpose of this uncertainty component is to account for the rounding that is automatically performed by the
balance. Because rounding automatically occurs at both zero (taring) and at load, two components will be
included.
The measurement process reproducibility data may double-count variation separately quantified for the display
resolution at zero and at load. Any double-counting will result in an over estimation of the measurement
uncertainty. (Note: a laboratory may be able to support a position that these components can be covered by
sufficient data from a measurement assurance program).
Balance Linearity:
The laboratory procedure to confirm the continued calibration status of the balance (ISO/IEC 17025:2005, Clause
5.5.10) and procedures to ensure proper functioning of the balance (ISO/IEC 17025:2005, Clause 5.5.6) both
have set pre-defined performance criteria of +/- 0.01 g across the entire useable range of the balance(s) used for
this measurement.
The value for the temperature coefficient is provided by the balance manufacturer based on the material(s) used
in the construction of the balance. It may be expressed as a nominal value or a percentage change per degree of
temperature (0C) in relation to the amount of load on the balance.
The laboratory in this example, uses this function of the balance and per a defined schedule and procedure
performs a check of the calibration status of the balance to provide objective evidence that this adjustment has
had minimal impact on the calibration performed by the external calibration laboratory. The calibration status
check procedure, using calibrated mass reference standards, covers the weighing range of the balance.
The laboratory’s pre-defined criteria for the calibration status checks preclude use of the balance outside of the
pre-defined performance criteria. This criterion will control variation due to many aspects including temperature of
the balance. Within this operating criterion, variations in environmental conditions in the laboratory, including
temperature, are captured through the use of the measurement assurance check standard (Type A Evaluation).
NOTE: if a laboratory is not evaluating the impact of balance temperature through other laboratory processes and
therefore through other uncertainty components, then the manufacturer’s temperature coefficient of sensitivity
would be used to quantify this component. A laboratory can determine if the component will be quantified as a
worst case scenario for the entire range of the balance as below, quantified for specific weight ranges or as a
relative uncertainty percentage.
Evaluating this at the upper limit of the balance will provide the largest possible impact
Upper limit of balance = 6000 g
The buoyant effect of weighing the measurand in air: From the SWGDRUG document10
“Buoyancy is difficult to account for in seized drug cases because the density of the material being weighed must
be known. However, for material that has a lower density than the steel calibration weights (8.0 g/cm3) the bias
imparted is always negative and the weight displayed by the balance will be less than the true weight of the
material. Ignoring buoyancy contributes a small systematic error that represents no more than 0.1 % bias to the
weight. Therefore, buoyancy corrections are not made in any uncertainty computations shown in this document
(SD-3).”
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
All uncertainty components must be expressed as a standard uncertainty and in the same measurement unit.
Upper limit = +a
Lower limit = -a
Possible range of values = (+a) - (-a) = 2a
standard uncertainty = a / √3
The certificate indicates this expanded uncertainty assumes a normal distribution, a coverage factor of k = 2 and
a coverage probability of approximately 95 %.
The uncertainty on the calibration certificate will be divided by the coverage factor, 2, to arrive at a standard
uncertainty.
standard uncertainty = 0.013 g / 2
= 0.007 g
Balance Linearity:
This component is evaluated as a rectangular distribution:
Upper limit = +a
Lower limit = -a
Possible range of values = (+a) - (-a) = 2a
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(g) All calculations performed
(h) The combined standard uncertainty,…
Individual standard uncertainties quantified by Type A and Type B methods of evaluation are now combined to
calculate the combined standard uncertainty for the measurement process. The combined standard uncertainty is an
estimated standard deviation and characterizes the dispersion of the values that could reasonably be attributed to the
measurement result at the equivalent of one standard deviation, a coverage probability of 68.26 %.
∑(c u )
2
uc ( y ) = i i
This is the RSS formula when the measurement result is the sum of a series of components. The formula also
assumes that the uncertainty components are independent or uncorrelated (a change in one component does not
cause a change in another component).
The estimation will assume that the uncertainty components are independent or uncorrelated and that the
measurement result is the sum of a series of components.
Care must be taken if the measurement results lie over a range of values. If the laboratory has information that
an individual component does not vary consistently across the measurement range, then the laboratory must
decide how this will be handled.
The measurement process reproducibility may double-count variation quantified by the scale readability at zero
and scale readability at load uncertainty components. This double-counting cannot be quantified. Any double-
counting will result in an over estimation of the measurement uncertainty.
uc ( y ) =s 2process + ureadabiltyzero
2
+ ureadabiltyload
2
+ ucalibrationunc
2
+ ulinearity
2
uc ( y ) =0.0312process + 0.003readabilityzero
2
+ 0.003readabilityload
2
+ 0.007calibrationunc
2
+ 0.006linearity
2
uc ( y ) = 0.001
uc ( y ) = 0.032 g
Evaluation of bias
Measurement accuracy 15 encompasses both precision 16 and bias 17. A measurement is more accurate when it has
less bias and greater precision. An assessment of bias requires the comparison of a measurement result to a
reference value.
The GUM3 in section 3.2.4 states “it is assumed that the result of a measurement has been corrected for all
recognized significant systematic effects and that every effort has been made to identify such effects.”
The laboratory in this example views the monitoring of bias as a component of ensuring the validity of the test
method (ISO/IEC 17025:2005, Clause 5.9.1). The laboratory routinely evaluates bias during the procedure for
confirmation of the calibration status of the balance through the use of multiple calibrated mass reference
standards.
The laboratory procedure requires that all calibration status checks be within 0.01 g of the reference value and
specifies what an analyst is to do if this does not occur (ISO/IEC 17025:2005, Clause 5.9.2). A review of the
data, including data points outside of this administrative limit with no known cause, yields a mean value of each
reference standard that is within the range of the reference value plus or minus the combined standard
uncertainty. The bias is viewed as insignificant by the laboratory.
Continuing to follow a conservative approach in this example, an additional component will be added to the
uncertainty estimation that will address the uncertainty in the reference value of the reference standards used for
the evaluation of bias.
uc ( y ) = s 2process + ureadabiltyzero
2
+ ureadabiltyload
2
+ ucalibrationunc
2
+ ulinearity
2
+ ucalibrefstd
2
uc ( y ) = 0.001
uc ( y ) = 0.033 g
The impact to the calculated combined standard uncertainty and can only be seen when additional digits are
shown in calculations.
The combined standard uncertainty, calculated in Step 5, is an estimated standard deviation and characterizes the
dispersion of the values that could reasonably be attributed to the measurement result at the equivalent of one
standard deviation, a coverage probability of 68.26 %.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 3.1.2 requires the coverage probability to be a minimum
of 95.45 % (often referred to as approximately 95 %).
The coverage probability of the combined standard uncertainty must be expanded to comply with the ASCLD/LAB
Policy.
EXAMPLE 1, Step 6: Expand the combined standard uncertainty by coverage factor (k)
The data from the measurement process is assumed to follow a normal distribution.
The laboratory has objective evidence supporting this assumption. The data was graphed in a histogram and
was found to be symmetrical about the mean with the number of data points 1, 2 and 3 standard deviations
from the mean following the expected percentages.
The laboratory has greater than 200 measurements of the check standard.
Refer to the Student’s t-distribution, Table 1 in ASCLD/LAB Guidance on the Estimation of Measurement
Uncertainty – ANNEX A9 for the specific value of k. NOTE: The assumptions for a normal distribution are
modified when a limited number of measurements are used for repeatability data and/or reproducibility data.
To expand the uncertainty to a 95.45 % coverage probability the coverage factor k = 2.0003 will be used. A
laboratory can choose to increase the coverage probability. To increase the coverage probability to 99.73 %, a
coverage factor k=3.0003 will be used. NOTE: A laboratory could choose to use the more conservative values for
k for 100 degrees of freedom from the Student’s t-distribution table or could interpolate the values of the t-
distribution table for the number of measurements available.
Step 7 is the time to critically evaluate the estimation of uncertainty and to determine if it “makes sense” and is
“reasonable.”
If the laboratory determines that the expanded uncertainty is not acceptable, areas of method improvement (e.g.,
measurement method standardization, improved calibration of equipment, increased readability of the measuring
instrument, etc.) can be identified and evaluated for the impact that a change would have on the estimation of
uncertainty using the information available from Steps 3 and 4.
Once changes to a test method have been validated, the appropriate edits to Steps 1-6 can be made and the
estimation of uncertainty reevaluated.
Is there a requirement regarding the relationship between the calculated expanded uncertainty and the
measurement result or the minimum sample load of a test method?
ISO/IEC 17025:2005, Clause 5.4.6.1 for calibration laboratories and Clause 5.4.6.2 for testing laboratories
requires laboratories to have and apply a procedure for estimating uncertainty of measurement.
Neither ISO/IEC 17025:2005 nor ASCLD/LAB-International Policy on Measurement Uncertainty specifies how large
the expanded uncertainty can be in relationship to the measurement result. It will be up to the laboratory to determine
what is acceptable as part of the laboratory procedure for estimating uncertainty of measurement.
The Eurachem/CITAC Guide on Quantifying Uncertainty in Analytical Measurement19 provides the following
information on this topic. Although written for analytical measurements, the points are applicable.
• At low concentration, an increasing number of effects become important
• The relative uncertainty associated with the result tends to increase, first to a substantial fraction of the result
and finally to the point where the (symmetric) uncertainty interval includes zero. This region is typically
associated with the practical limit of detection or limit of quantitation for a given method.
• The GUM does not give explicit instructions when the results are small and the uncertainties large compared
to the results. Indeed, the basic form of the ‘law of propagation of uncertainties’ may cease to apply
accurately in this region; one assumption on which the calculation is based is that the uncertainty is small
relative to the value of the measurand.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 3.1.1 requires the laboratory to have a procedure for the
process of rounding the uncertainty.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 3.1.2 requires the coverage probability of the expanded
uncertainty to be a minimum of 95.45 % (often referred to as approximately 95 %).
Coverage probability (VIM 2.37): probability that the set of true quantity values of a measurand is contained
within a specified coverage interval
Note 1 This definition pertains to the Uncertainty Approach as presented in the GUM.
Note 2 The coverage probability is also termed “level of confidence” in the GUM.
ASCLD/LAB requirements for the reporting of measurement uncertainty can be found in the Policy on
Measurement Uncertainty in Clauses 5.1 - 5.3
Clauses 5.1 and 5.3 apply to testing laboratories
ASCLD/LAB Policy on Measurement Uncertainty, Clause 5.3.1 states that the rounded expanded uncertainty
shall be reported to at most two significant digits, unless the laboratory has a documented rationale for reporting
additional significant figures. The need for reporting additional significant figures in a rounded expanded
uncertainty is rare.
Refer to the Policy for all reporting requirements as only selected reporting requirements are being covered in this
Guidance Document.
5.1. 2 When measurement uncertainty is reported, the value shall be reported in the test report or in an
attachment to the report, and shall be expressed as an expanded uncertainty and include the coverage
probability.
This measurement result shall include the measured quantity value, y, along with the associated
expanded uncertainty, U, and this measurement result shall be reported as y ± U where U is consistent
with the units of y.
NOTE 1 For asymmetrical uncertainties, it may be inappropriate to quote a single result for the
uncertainty and presentations other than y ± U may be needed.
As required in Section 3.1.1 of the ASCLD/LAB Policy on Measurement Uncertainty, the laboratory has
established a procedure for the process of rounding the expanded uncertainty. Following that procedure and
considering Section 5.3.3 of the Policy that requires the measurement result and the rounded expanded
uncertainty to be reported to the same level of significance, the expanded uncertainty will be reported as:
k = 2.000 U= 0.07 g
k = 3.000 U= 0.10 g
The net weight of Item 1 was found to be 30.03 g +/- 0.07 g at a coverage probability of 95.45 %.
The net weight of Item 1 was found to be 30.03 g +/- 0.10 g at a coverage probability of 99.73 %.
A laboratory can include additional information regarding the test result and the relationship to a specification in the
test report but it must be additional information and not in place of that which is required by the ASCLD/LAB Policy on
Measurement Uncertainty. If a laboratory wishes to compare a test result to a specification (e.g., statute limit), the
expanded uncertainty is taken into account in evaluating compliance to the specification. The report must accurately,
clearly, unambiguously and objectively communicate this information.
The net weight of Item 1 was found to be 30.03 g +/- 0.07 g at a coverage probability of 95.45 %. Taking the
estimated measurement uncertainty into consideration, the weight of Item 1 is greater than one ounce (28.35 g).
The net weight of Item 1 was found to be 30.03 g +/- 0.10 g at a coverage probability of 99.73 %. Taking the
estimated measurement uncertainty into consideration, the weight of Item 1 is greater than one ounce (28.35 g)
If the laboratory, desires to include additional information in a test report regarding a specification but the
measurement result and the expanded uncertainty are not clearly outside (above or below) a specification,
can the laboratory still include this additional information?
A laboratory can still include additional information regarding a specification, a statute limit. The report must
accurately, clearly, unambiguously and objectively communicate this information.
The measurement
result +/- the
expanded
uncertainty is above
The measurement the limit.
result +/- the
expanded
uncertainty is not
above the limit.
LIMIT
The measurement
result +/- the
expanded
uncertainty is not
The measurement below the limit.
result +/- the
expanded
uncertainty is below
the limit.
The net weight of Item 1 was found to be 28.43 g +/- 0.07 g at a coverage probability of 95.45 %. Taking the
estimated measurement uncertainty into consideration, the weight of Item 1 is greater than one ounce (28.35 g).
The net weight of Item 1 was found to be 28.43 g +/- 0.10 g at a coverage probability of 99.73 %. Taking the
estimated measurement uncertainty into consideration, there is a possibility that the weight of Item 1 is less than
one ounce (28.35 g).
If an item contains multiple sub-items and the laboratory desires to report a total weight, how is the
measurement uncertainty handled?
If a laboratory desires to issue an overall weight for an item containing two or more sub-items, then the uncertainty in
each measurement must be accounted for in the reported overall uncertainty. If the measurements are made using
the same equipment, the calibration of the equipment is also the same, and there is correlation between some
uncertainty components. The difficulty comes in quantifying the amount of correlation.
It is the laboratory’s responsibility to consult references or seek the advice of a statistician as necessary when
evaluating correlation of uncertainty components.
If the measurements use different equipment, two different types of balances, these are viewed as independent
measurements. This position is supported based on the major components to uncertainty, measurement process
reproducibility and readability, being different. The weights and their respective uncertainties would be added
together.
Example: Laboratory obtains a gross weight on a balance that reads to 0.1 gram, obtains the weight of the
packaging on a balance that reads to 0.001 gram. Per laboratory procedure the net weight is truncated to the
level of significance of the least sensitive balance. The combined uncertainty is rounded per laboratory
procedure.
Report statement:
The net weight of Item 1 was found to be 174.4 g +/- 1.0 g at a coverage probability of 95.45 %.
Measurement Traceability:
The traceability for this measurement process is established through the calibration of the measuring equipment,
a graduated cylinder, used to perform the measurement.
• The calibration of the graduated cylinder was determined to have a significant effect on the accuracy or
validity of the test result.
o The calibration of the graduated cylinder is performed by an external calibration laboratory that meets
Section 3.2.3 of the ASCLD/LAB Policy on Measurement Traceability.11
Measurement Assurance:
• The laboratory determined that no intermediate checks are needed to maintain confidence in the calibration
status of the graduated cylinder during the interval between external calibrations of the equipment. (ISO/IEC
17025:2005, Clause 5.5.10). The graduated cylinder is inspected visually to ensure proper functioning
(ISO/IEC 17025:2005, Clause 5.5.6)
• Training and competency testing of each analyst as well as an intra-laboratory comparison where multiple
analysts read the volume of different types of liquid routinely encountered in casework provides data on the
variation of multiple analysts reading a volume of liquid using the graduated cylinder.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(a) Statement defining the measurand
Quantity (VIM 1.1): property of a phenomenon, body, or substance, where the property has a magnitude that can
be expressed as a number and a reference
The statement defining the measurand can be a written statement, a visual diagram and/or a mathematical
expression. To be clear about what measurement process the estimation is being calculated for, it is important to be
as specific as possible when defining the measurand. To distinguish one measurement process from another in a
laboratory it may be necessary to include a reference to a specific type of equipment used, a specific procedure, etc.
in the statement defining the measurand.
Drug Chemistry Volume Determination - Test Method No. XXXX, Revision ZZZZ
Where:
y is the best estimate of the measurand Y from the test method and is expressed by
y = f (x1, x2, ..., xN )
y is a function of the components of uncertainty
b = bias; in this example, the calibrated scale error is a systematic bias in the measurement
U = expanded uncertainty
ISO/IEC 17025:2005, Clause 5.4.6.2 states “… the laboratory shall at least attempt to identify all the components
of uncertainty…”
ISO/IEC 17025:2005, Clause 5.4.6.3 states “when estimating the uncertainty of measurement, all uncertainty
components which are of importance in the given situation shall be taken into account …”
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(b) Statement of how traceability is established for the measurement
(c) The equipment (e.g., measuring device[s] or instrument[s]) used
(d) All uncertainty components considered
ASCLD/LAB Policy on Measurement Uncertainty, Clause 5.3.2 requires that the specific measuring device or
instrument used for a reported test result to have been evaluated in the estimation of measurement uncertainty for
that test method.
Traceability is established for this measurement through the calibration of the graduated cylinder used for the
measurement.
The equipment (e.g., measuring device[s] or instrument[s]) used in this test method is:
All measurements are made using one calibrated 100 mL graduated cylinder.
A laboratory may identify different uncertainty components when an evaluation of their specific measurement
process is performed.
This list of potential uncertainty components could also be compiled into a fishbone diagram or into any other format of
the laboratory’s choosing. Two fishbone diagram examples depicting possible uncertainty components and one blank
fishbone diagram are available on the ASCLD/LAB website at www.ascld-lab.org. Appendix B to this document
provides an example of using an Excel spreadsheet (available at www.ascld-lab.org) to display the information from
the NIST 8-Step Process.
ISO/IEC 17025:2005, Clause 5.4.6.3 states “when estimating the uncertainty of measurement, all uncertainty
components which are of importance in the given situation shall be taken into account using appropriate methods
of analysis.”
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(e) All uncertainty components of significance and how they were evaluated
(f) Data used to estimate repeatability and/or reproducibility
(g) All calculations performed
Review the list of uncertainty components created in Step 2. Identify which of the components listed can be quantified
by existing data from the measurement process.
The GUM (2.3.2) refers to this method of evaluation of one or more uncertainty components as:
Type A evaluation (of uncertainty): method of evaluation of uncertainty by the statistical analysis of series
of observations
Type B evaluation (of uncertainty): method of evaluation of uncertainty by means other than the statistical
analysis of series of observations
An intra-laboratory comparison requiring all analysts to measure the volume of typical liquids was conducted to
gather initial data. Prior to this initial study, the laboratory had standardized the procedure for reading the
meniscus. The procedure includes a meniscus reader that must be used. The procedure follows NIST GMP 3 -
Good Measurement Practice for Method of Reading A Meniscus Using Water or Other Wetting Liquid 20 (Type A
Evaluation).
The laboratory has determined that the largest difference between analysts is less than one increment on the
measurement scale of the graduated cylinder, 1 mL (Type B Evaluation).
Each new analyst during training practical exercises and competency testing must provide objective evidence that
their performance meets the laboratory pre-defined criteria for measurement competency.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(g) All calculations performed
All uncertainty components must be expressed as a standard uncertainty and in the same measurement unit.
Upper limit = +a
Lower limit = -a
standard uncertainty = a / √3
A review of the calibration certificate from the accredited external calibration laboratory identified that the greatest
calibration expanded uncertainty over the usable range is 50 µL.
The certificate indicates this expanded uncertainty assumes a normal distribution, a coverage factor of k = 2 and a
coverage probability of approximately 95 %.
The uncertainty on the calibration certificate will be divided by the coverage factor, 2, to arrive at a standard
uncertainty.
standard uncertainty = 50 µL / 2
= 25 µL
= 0.025 mL
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(g) All calculations performed
(h) The combined standard uncertainty,…
Individual standard uncertainties quantified by Type A and Type B methods of evaluation are now combined to
calculate the combined standard uncertainty for the measurement process. The combined standard uncertainty is an
estimated standard deviation and characterizes the dispersion of the values that could reasonably be attributed to the
measurement result at the equivalent of one standard deviation, a coverage probability of 68.26 %.
The combined standard uncertainty, denoted by uc(y), is the positive square root of the variance of all components
combined. Variance is equal to the Standard Deviation squared. This formula is commonly called the Root Sum of
the Squares or RSS.
∑(c u )
2
uc ( y ) = i i
This is the RSS formula when the measurement result is the sum of a series of components. The formula also
assumes that the uncertainty components are independent or uncorrelated (a change in one component does not
cause a change in another component).
The estimation will assume that the uncertainty components are independent or uncorrelated and that the
measurement result is the sum of a series of components.
=uc ( y ) 2
uscalereadability + ucalibunc
2
uc ( y ) = 0.334
uc ( y ) = 0.578 mL
The combined standard uncertainty, calculated in Step 5, is an estimated standard deviation and characterizes the
dispersion of the values that could reasonably be attributed to the measurement result at the equivalent of one
standard deviation, a coverage probability of 68.26 %.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 3.1.2 requires the coverage probability to be a minimum
of 95.45 % (often referred to as approximately 95 %).
The coverage probability of the combined standard uncertainty must be expanded to comply with the ASCLD/LAB
Policy.
EXAMPLE 2, Step 6: Expand the combined standard uncertainty by coverage factor (k)
Both components in this example are based on an infinite number of observations. To expand the uncertainty to
a 95.45 % coverage probability the coverage factor k = 2.0003 will be used.
A laboratory can choose to increase the coverage probability to a greater percentage than that required in the
Policy. If, for example, a laboratory chooses to increase the coverage probability to 99.73 %, then the coverage
factor k = 3.0003 will be used.
Step 7 is the time to critically evaluate the estimation of uncertainty and to determine if it “makes sense” and is
“reasonable.”
If the laboratory determines that the expanded uncertainty is not acceptable, areas of method improvement (e.g.,
measurement method standardization, improved calibration of equipment, increased readability of the measuring
instrument, etc.) can be identified and evaluated for the impact that a change would have on the estimation of
uncertainty using the information available from Steps 3 and 4.
Once changes to a test method have been validated, the appropriate edits to Steps 1-6 can be made and the
estimation of uncertainty reevaluated.
Is there a requirement regarding the relationship between the calculated expanded uncertainty and the
measurement result or the minimum sample load of a test method?
ISO/IEC 17025:2005, Clause 5.4.6.1 for calibration laboratories and Clause 5.4.6.2 for testing laboratories
requires laboratories to have and apply a procedure for estimating uncertainty of measurement.
Neither ISO/IEC 17025:2005 nor ASCLD/LAB-International Policy on Measurement Uncertainty specifies how large
the expanded uncertainty can be in relationship to the measurement result. It will be up to the laboratory to determine
what is acceptable as part of the laboratory procedure for estimating uncertainty of measurement.
The Eurachem/CITAC Guide on Quantifying Uncertainty in Analytical Measurement19 provides the following
information on this topic. Although written for analytical measurements, the points are applicable.
• At low concentration, an increasing number of effects become important
• The relative uncertainty associated with the result tends to increase, first to a substantial fraction of the result
and finally to the point where the (symmetric) uncertainty interval includes zero. This region is typically
associated with the practical limit of detection or limit of quantitation for a given method.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 3.1.1 requires the laboratory to have a procedure for the
process of rounding the uncertainty.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 3.1.2 requires the coverage probability of the expanded
uncertainty to be a minimum of 95.45 % (often referred to as approximately 95 %).
Coverage probability (VIM 2.37): probability that the set of true quantity values of a measurand is contained
within a specified coverage interval
Note 1 This definition pertains to the Uncertainty Approach as presented in the GUM.
Note 2 The coverage probability is also termed “level of confidence” in the GUM.
ASCLD/LAB requirements for the reporting of measurement uncertainty can be found in the Section 5 of the
ASCLD/LAB Policy on Measurement Uncertainty.
Clauses 5.1 and 5.3 apply to testing laboratories
ASCLD/LAB Policy on Measurement Uncertainty, Clause 5.3.1 states that the rounded expanded uncertainty
shall be reported to at most two significant digits, unless the laboratory has a documented rationale for reporting
additional significant figures. The need for reporting additional significant figures in a rounded expanded
uncertainty is rare.
Refer to the Policy for all reporting requirements as only selected reporting requirements are being covered in this
Guidance Document.
5.1. 2 When measurement uncertainty is reported, the value shall be reported in the test report or in an
attachment to the report, and shall be expressed as an expanded uncertainty and include the coverage
probability.
This measurement result shall include the measured quantity value, y, along with the associated
expanded uncertainty, U, and this measurement result shall be reported as y ± U where U is consistent
with the units of y.
5.3.3 The measurement result and the rounded expanded uncertainty shall be reported to the same level of
significance.
As required in Clause 3.1.1 of the ASCLD/LAB Policy on Measurement Uncertainty, the laboratory has
established a procedure for the process of rounding the expanded uncertainty. Following that procedure and
considering Clause 5.3.3 of the Policy that requires the measurement result and the rounded expanded
uncertainty to be reported to the same level of significance, the expanded uncertainty will be:
k = 2.000 U = 1 mL
k = 3.000 U = 2 mL
Scale Error
The known scale error, a correction, must be taken into account prior to reporting the measurement result. A
review of the calibration certificate for the graduated cylinder indicates that the scale error is +0.080 mL. The
correction to the measurement result is -0.080 mL.
If the volume of Item 1 is measured to be 23 mL, then the corrected measurement result is:
23 mL – 0.080 mL = 22.92 mL
The laboratory procedure will need to address if the measurement result is rounded or truncated and if rounded
what the procedure for rounding will be. For this example the result will be rounded down.
The volume of liquid to be measured is greater than the volume of the calibrated graduated cylinder so more
than one measurement is necessary. How is this handled?
This example does make only one measurement of volume. If the volume of liquid is greater than 100 mL requiring
two or more measurements, then the uncertainty in each measurement must be accounted for in the overall
uncertainty.
If the measurements are made using the same equipment, the calibration of the equipment is the same there is
correlation or dependence between the uncertainty components. The difficulty comes in quantifying the amount of
The SWGDRUG has published Supplemental Document SD-4 for Part IVC – Quality Assurance/Uncertainty
Measurement Uncertainty for Purity Determinations in Seized Drug Analysis 21. Readers are encouraged to
review the SWGDRUG document in its entirety.
Measurement Traceability:
The traceability for this measurement process is established through the calibrators used to establish the
calibration curve on the measuring system 23 as well as through the calibration of other equipment used in the
measurement process where the calibration of the equipment is viewed to have a significant effect on sampling or
the accuracy of the test result.
• All calibrations of measuring equipment are performed by external calibration laboratories that meet Section
3.2.3 of the ASCLD/LAB Policy on Measurement Traceability11
o The following equipment is calibrated:
Balance
Volumetric flasks
Pipette used to aliquot the internal solution
Measurement Assurance
• This laboratory has prepared a measurement assurance sample to use as quality control to ensure the
ongoing performance of the analytical test method (ISO/IEC 17025:2005, Clause 5.9.1). The measurement
assurance check standard12 contains the analyte of interest and a common diluent. The reference material
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(a) Statement defining the measurand
Quantity (VIM 1.1): property of a phenomenon, body, or substance, where the property has a magnitude that can
be expressed as a number and a reference
The statement defining the measurand can be a written statement, a visual diagram and/or a mathematical
expression. To be clear about what measurement process the estimation is being calculated for, it is important to be
as specific as possible when defining the measurand. To distinguish one measurement process from another in a
laboratory it may be necessary to include a reference to a specific type of equipment used, a specific procedure, etc.
in the statement defining the measurand.
Drug Chemistry – Purity Determination of Drug R using Procedure No. XXXX, Revision ZZZZ
Imeasurand
Cmeasurand = Ccalibratorsx + b +/- U
Icalibrators
Where,
I is the instrument response
C is the concentration
b is a bias
U = expanded uncertainty
Each of these influences (inputs) on the measuring process will have uncertainty components that will be
considered.
ISO/IEC 17025:2005, Clause 5.4.6.2 states “… the laboratory shall at least attempt to identify all the components
of uncertainty…”
ISO/IEC 17025:2005, Clause 5.4.6.3 states “when estimating the uncertainty of measurement, all uncertainty
components which are of importance in the given situation shall be taken into account …”
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(b) Statement of how traceability is established for the measurement
(c) The equipment (e.g., measuring device[s] or instrument[s]) used
(d) All uncertainty components considered
ASCLD/LAB Policy on Measurement Uncertainty, Clause 5.3.2 requires that the specific measuring device or
instrument used for a reported test result to have been evaluated in the estimation of measurement uncertainty for
that test method.
The Eurachem/CITAC Guide on Quantifying Uncertainty in Analytical Measurement19 suggests that for analytical
methods it is helpful to break the analytical procedure into a set of generic steps when identifying uncertainty
components. This reference includes the following steps that are then further broken down to individual components:
Sampling
Sample Preparation
Presentation of Certified Reference Materials to the measuring instrument
Calibration of the Instrument
Analysis (data acquisition)
Sample preparation may include, but may not be limited to extraction, concentration, and derivatization. Each of these
steps will have uncertainty components to consider.
Traceability is established for this measurement through the preparation of the calibrators used to establish
the calibration curve on the measuring system and through the calibration of other equipment used in the
measurement process where the calibration of the equipment is viewed to have a significant effect on sampling or
the accuracy of the test result.
• Calibrators
o Prepared using a reference material of known purity, a calibrated balance and calibrated volumetric
flasks.
Both the balance and the volumetric flasks are considered to be equipment that has a significant
effect on the accuracy or validity of the test result (ISO/IEC 17025:2005, Clause 5.6.1).
o The multi-point calibration curve covering the reportable concentration range is established each
analytical run.
• Internal Standard
o The pipette used to add the specified volume of internal standard is considered to be a piece of
equipment that has a significant effect on sampling (ISO/IEC 17025:2005, Clause 5.6.1).
The laboratory has the pipette calibrated by an appropriately accredited external calibration
laboratory.
The procedure used to both check the calibration status (ISO/IEC 17025:2005, Clause 5.5.10) and
to ensure proper functioning of the pipette (ISO/IEC 17025:2005, Clause 5.5.6) requires the use of a
calibrated balance.
• Balance
o The laboratory has determined that the measurements made in the procedure to ensure proper
functioning of the pipette are to have measurement traceability. This is a second reason, to that above,
that the laboratory has the balance calibrated by an appropriately accredited calibration laboratory.
Intermediate checks performed to ensure continued confidence in the calibration status of the
balance, are performed using calibrated mass reference standards.
The equipment (e.g., measuring device[s] or instrument[s]) used in this test method is:
• Chromatographic measuring system including sampling system and data processing software
• Analytical balance used in the preparation of calibrators, quality control samples and measurand samples as
well as the procedure to ensure the proper functioning of the pipette
• Volumetric flask(s) used to prepare calibrators
• Pipette used to measure the specified volume of internal standard
A laboratory may identify different uncertainty components when an evaluation of their specific measurement
process is performed.
This list of potential uncertainty components could also be compiled into a fishbone diagram or into any other format of
the laboratory’s choosing. Two fishbone diagram examples depicting possible uncertainty components and one blank
fishbone diagram are available on the ASCLD/LAB website at www.ascld-lab.org. Appendix C to this document
provides an example of using an Excel spreadsheet (available at www.ascld-lab.org) to display the information from
the NIST 8-Step Process.
ISO/IEC 17025:2005, Clause 5.4.6.3 states “when estimating the uncertainty of measurement, all uncertainty
components which are of importance in the given situation shall be taken into account using appropriate methods
of analysis.”
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(e) All uncertainty components of significance and how they were evaluated
(f) Data used to estimate repeatability and/or reproducibility
(g) All calculations performed
Review the list of uncertainty components created in Step 2. Identify which of the components listed can be quantified
by existing data from the measurement process.
The GUM (2.3.2) refers to this method of evaluation of one or more uncertainty components as:
Type A evaluation (of uncertainty): method of evaluation of uncertainty by the statistical analysis of series
of observations
The quantification of the remaining uncertainty components is performed by a different method of evaluation referred
to by the GUM (2.3.3) as:
Type B evaluation (of uncertainty): method of evaluation of uncertainty by means other than the statistical
analysis of series of observations
The calibration model was determined during method validation and was shown through the use of residual plots
to have equal variation across the linear range. Therefore, the laboratory is going to calculate an estimation of
measurement uncertainty for the entire reportable concentration range.
(Note: if variation is not constant across the reportable range, a laboratory can still choose to estimate a
single measurement uncertainty but will have to base calculations on the widest variation seen in the
measurement process or a laboratory can choose to break the reportable concentration range into two or
more sections and estimate the measurement uncertainty for each concentration section.)
Data is available from ongoing quality control using the same check standard over a long period of time. This
check standard contains the analyte at approximately 50 % purity mixed with a common diluent. The laboratory
The analytical process involves the weighing of similar weight samples for all calibrators, quality control samples
and replicate samples of the measurand. This weighing process uses the same balance, the same operator, and
occurs at approximately the same time. Since the purity concentration of the measurand is relational to the purity
concentration of the calibrators, any impact due to uncertainty in the balance calibration, drift in linearity,
readability of the balance and impact of facility temperature (see Example #1) are viewed to be captured by the
quality control check standard that goes through the same process.
Two uncertainty components related to the internal standard were listed for consideration. Using professional
judgment and assuming that the same batch of internal standard is used for all samples in an analytical batch, the
only aspect of the internal standard that has an influence on the measurement result is variation in the amount of
internal standard added to each calibrator, quality control sample or measurand sample by the pipette.
Analysis
Instrument parameter settings (e.g. oven
temperature(s), gas flow, split ratios, integration Covered in Type A Evaluation of process
parameters, aging of chromatographic column, reproducibility data
autosampler syringe, autosampler precision, etc)
Duplicate listing of component – see
Interference from the matrix
Measurand section above
This component is not an uncertainty
component but is a quality control concern
The laboratory analyzes a matrix blank that
Interference from reagents contains no analyte but does evaluate all
reagents used in the analytical method. The
laboratory procedure specifies acceptable
criteria for this quality control sample
This component is not an uncertainty
component but is a quality control concern
The laboratory, as part of the procedure to
Interference from other compounds identify the analyte prior to quantitation, will
have ensured that specificity of the
chromatographic quantitation test method is
acceptable.
Quantified in Type A Evaluation of process
reproducibility data QC sample and through the
Stability of sample(s) from preparation through analysis
procedure administrative requirement for
agreement of replicates (Type B Evaluation)
Data Processing
Covered in Type A Evaluation of process
Calibration model
reproducibility data
Covered in Type A Evaluation of process
Integration parameters
reproducibility data
Covered in Type A Evaluation of process
Processing algorithms
reproducibility data
The number of measurements from ongoing quality control sample data in this example is 26.
The statistic that will be calculated is the percent relative standard deviation.
To begin the mean (average) and standard deviation of the data will be calculated.
Mean (Sample):
1 n
x = ∑ xi
n i =1
( x1 + x2 + x3 + ...xn )
x=
n
∑ (x )
n
2
i −x
s= i =1
n −1
The standard deviation of the reproducibility data in this example = 2.98 %
s
RSD =
x
2.98
RSD =
49.58
RSD = 0.060
The laboratory procedural requirement for replicate agreement is an example of an administrative requirement
that restricts variation in the measurement method.
• It is up to a laboratory to determine if an administrative control such as this example will be used.
o The decision will be based on, but may not be limited to, knowledge of the measurement process, the
impact of repeat analysis on cost and process efficiency, and the required expanded uncertainty.
• Measurement data may at times exceed the set administrative limit and not be a statistical outlier.
The two components, process reproducibility and interference from matrix, quantify a number of the same
uncertainty components. The matrix control, over a longer period of time, holds the impact from matrix constant
while the effects from equipment, calibration, operators, and the laboratory environmental conditions vary. The
replicate samples of the measurand provide information on the measurand matrix and a short–term evaluation of
the effect from equipment, calibration, operators, and the laboratory environment. Any double-counting of a
component will result in an over estimation of the measurement uncertainty.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(g) All calculations performed
All uncertainty components must be expressed as a standard uncertainty and in the same measurement unit.
In this example, the measurand is sampled in duplicate and the laboratory procedure for the reported purity is to
average the two measurement results. Repeat measurements of the measurand provide more information and
more confidence that the reported result is the best estimate of the true value. When multiple measurements are
made of the measurand and the average is reported and the measurement process reproducibility data is based
on single measurements of quality control samples, the standard deviation of the measurement process can be
divided by the square root of the number of measurements. This statistic is the standard deviation of the mean.
If a single measurement result for the measurand is selected to be reported (e.g., the lowest value), then the
standard deviation of the mean calculation is not applicable. If the laboratory makes an equal number of multiple
measurements of the quality control sample as it does of the measurand and averages the results to evaluate the
acceptability of the quality control sample, then the standard deviation of the mean calculation is not applicable.
The % relative standard deviation of the reproducibility data in this example = 6.0 %
The mathematical expression for relative standard deviation of the mean:
RSD
RSDmean =
n
RSDmean = 4.24%
Upper limit = +a
Lower limit = -a
Possible range of values = (+a) - (-a) = 2a
standard uncertainty = a / √3
The reference material was purchased from an authoritative source. Stated purity is >99 %.
99 % purity is the minimum and 100 % pure the maximum. Therefore, the range can be expressed as 99.5 % +/-
0.5 %.
standard uncertainty = a / √3
The certificate indicates this expanded uncertainty assumes a normal distribution, a coverage factor of k = 2 and
a coverage probability of approximately 95 %.
The uncertainty on the calibration certificate will be divided by the coverage factor, 2, to arrive at a standard
uncertainty.
The certificate indicates this expanded uncertainty assumes a normal distribution, a coverage factor of k = 2 and
a coverage probability of approximately 95 %.
Upper limit = +a
Lower limit = -a
Possible range of values = (+a) - (-a) = 2a
standard uncertainty = a / √3
ASCLD/LAB Policy on Measurement Uncertainty, Clause 4.1 requires ASCLD/LAB-International applicant and
accredited testing laboratories to record for each estimation of measurement uncertainty:
(g) All calculations performed
(h) The combined standard uncertainty,…
Individual standard uncertainties quantified by Type A and Type B methods of evaluation are now combined to
calculate the combined standard uncertainty for the measurement process. The combined standard uncertainty is an
The combined standard uncertainty, denoted by uc(y), is the positive square root of the variance of all components
combined. Variance is equal to the Standard Deviation squared. This formula is commonly called the Root Sum of
the Squares or RSS.
∑(c u )
2
uc ( y ) = i i
This is the RSS formula when the measurement result is the sum of a series of components. The formula also
assumes that the uncertainty components are independent or uncorrelated (a change in one component does not
cause a change in another component).
The estimation will assume that the uncertainty components are independent or uncorrelated and that the
measurement result is the sum of a series of components.
Care must be taken if the measurement results lie over a range of values. If the laboratory has information that
an individual component does not vary consistently across the measurement range, then the laboratory must
decide how this will be handled. In this scenario, the calibration model was determined during method validation
and was shown through the use of residual plots to have equal variation across the linear range so a single
estimation of measurement uncertainty can be calculated for the entire concentration range. Another option is to
calculate a measurement uncertainty for different ranges of measurements.
As noted in Step 3, the two components, process reproducibility and interference from matrix, quantify a number
of the same uncertainty components. The matrix control, over a longer period of time, holds the impact from
matrix constant while the effects from equipment, calibration, operators, and the laboratory environmental
conditions vary. The replicate samples of the measurand provide information on the measurand matrix and a
short–term evaluation of the effect from equipment, calibration, operators, and the laboratory environment. Any
double-counting of a component will result in an over estimation of the measurement uncertainty.
u=
c ( y)
2
sreproducibility + uMatrix
2
+ uRMpurity
2
+ u 2pipettecal + u 2flaskcal + u 2flaskvolerr
u=
c ( y) 4.242reproducibility + 5.77 2Matrix + 0.292RMpurity + 0.092pipettecal + 0.022flaskcal + 0.082flaskvolerr
uc ( y ) = 51.44
uc ( y ) = 7.17 %
ASCLD/LAB Policy on Measurement Uncertainty, Clause 3.1.2 requires the coverage probability to be a minimum
of 95.45 % (often referred to as approximately 95 %).
The coverage probability of the combined standard uncertainty must be expanded to comply with the ASCLD/LAB
Policy.
EXAMPLE 3, Step 6: Expand the combined standard uncertainty by coverage factor (k)
The data from the measurement process is assumed to follow a normal distribution.
The laboratory has objective evidence supporting this assumption. The data was graphed in a histogram and
was found to be symmetrical about the mean with the number of data points 1, 2 and 3 standard deviations
from the mean following the expected percentages.
The assumptions for a normal distribution are modified when a limited number of measurements are used for
repeatability data and/or reproducibility data. Refer to the Student’s t-distribution table, Table 1 in ASCLD/LAB
Guidance on the Estimation of Measurement Uncertainty – ANNEX A9 for the specific values of k.
To expand the uncertainty to a 95.45 % coverage probability the coverage factor k = 2.113 will be used.
A laboratory can choose to increase the coverage probability. If a laboratory chose to increase the coverage
probability to 99.73 %, then the coverage factor k = 3.333 will be used.
Step 7 is the time to critically evaluate the estimation of uncertainty and to determine if it “makes sense” and is
“reasonable.”
If the laboratory determines that the expanded uncertainty is not acceptable, areas of method improvement (e.g.,
measurement method standardization, improved calibration of equipment, etc.) can be identified and evaluated for the
impact that a change would have on the estimation of uncertainty using the information available from Steps 3 and 4.
Once changes to a test method have been validated, the appropriate edits to Steps 1-6 can be made and the
estimation of uncertainty reevaluated.
Is there a requirement regarding the relationship between the calculated expanded uncertainty and the
measurement result or the limit of quantification of a test method?
ISO/IEC 17025:2005, Clause 5.4.6.1 for calibration laboratories and Clause 5.4.6.2 for testing laboratories
requires laboratories to have and apply a procedure for estimating uncertainty of measurement.
Neither ISO/IEC 17025:2005 nor ASCLD/LAB-International Policy on Measurement Uncertainty specifies how large
the expanded uncertainty can be in relationship to the measurement result. It will be up to laboratory to determine
what is acceptable as part of the laboratory procedure for estimating uncertainty of measurement.
The Eurachem/CITAC Guide on Quantifying Uncertainty in Analytical Measurement19 provides the following
information on this topic.
ASCLD/LAB Policy on Measurement Uncertainty, Clause 3.1.1 requires the laboratory to have a procedure for the
process of rounding the uncertainty.
Coverage probability (VIM 2.37): probability that the set of true quantity values of a measurand is contained
within a specified coverage interval
Note 1 This definition pertains to the Uncertainty Approach as presented in the GUM.
Note 2 The coverage probability is also termed “level of confidence” in the GUM.
ASCLD/LAB requirements for the reporting of measurement uncertainty can be found in the Policy on
Measurement Uncertainty in Clauses 5.1 - 5.3
Clauses 5.1 and 5.3 apply to testing laboratories
ASCLD/LAB Policy on Measurement Uncertainty, Clause 5.3.1 states that the rounded expanded uncertainty
shall be reported to at most two significant digits, unless the laboratory has a documented rationale for reporting
additional significant figures. The need for reporting additional significant figures in a rounded expanded
uncertainty is rare.
Refer to the Policy for all reporting requirements as only selected reporting requirements are being covered in this
Guidance Document.
5.1. 2 When measurement uncertainty is reported, the value shall be reported in the test report or in an
attachment to the report, and shall be expressed as an expanded uncertainty and include the coverage
probability.
This measurement result shall include the measured quantity value, y, along with the associated
expanded uncertainty, U, and this measurement result shall be reported as y ± U where U is consistent
with the units of y.
NOTE 1 For asymmetrical uncertainties, it may be inappropriate to quote a single result for the
uncertainty and presentations other than y ± U may be needed.
5.3.3 The measurement result and the rounded expanded uncertainty shall be reported to the same level of
significance.
As required in Section 3.1.1 of the ASCLD/LAB Policy on Measurement Uncertainty, the laboratory has
established a procedure for the process of rounding the expanded uncertainty. Following that procedure and
considering Section 5.3.3 of the Policy that requires the measurement result and the rounded expanded
uncertainty to be reported to the same level of significance, the expanded uncertainty will be reported as:
k = 2.11 U= 15 %
k = 3.33 U= 24 %
The percentage of Drug R in Item 1 was found to be 65 % +/- 15 % at a coverage probability of 95.45 %.
The percentage of Drug R in Item 1 was found to be 65 % +/- 24 % at a coverage probability of 99.73 %.
A laboratory can include additional information regarding the test result’s relationship to a specification in the test
report but it must be additional information and not in place of that which is required by the ASCLD/LAB Policy on
Measurement Uncertainty. If a laboratory wishes to compare a test result to a specification (e.g., statute limit), the
expanded uncertainty is taken into account in evaluating compliance to the specification. The report must accurately,
clearly, unambiguously and objectively communicate this information.
The percentage of Drug R in Item 1 was found to be 97 % +/- 15 % at a coverage probability of 95.45 %.
Taking the estimated measurement uncertainty into consideration the concentration of Drug R in Item 1 is greater
80 %.
The percentage of Drug R in Item 1 was found to be 97 % +/- 24 % at a coverage probability of 99.73 %.
Taking the estimated measurement uncertainty into consideration, there is a possibility that the concentration of
Drug R in Item 1 is less than 80 %.
In the scenario above, where the test result considering the expanded uncertainty is not clearly outside a specification,
the report will need to accurately, clearly, unambiguously and objectively report this information.
LIMIT
The measurement
result +/- the
expanded
uncertainty is not
The measurement
below the limit.
result +/- the
expanded
uncertainty is below
the limit.
1 American Society of Crime Laboratory Directors / Laboratory Accreditation Board (ASCLD/LAB): ASCLD/LAB Policy on
Measurement Uncertainty (AL-PD-3060) (Garner, North Carolina: ASCLD/LAB, 2013). Available at www.ascld-lab.org.
2 Adapted from:
National Institute of Standards and Technology, SOP 29 – Standard Operating Procedure for the Assignment of
Uncertainty, (Gaithersburg, Maryland, February 2012). Available for download at
http://www.nist.gov/pml/wmd/labmetrology/upload/SOP_29_20120229.pdf
3
Joint Committee for Guides in Metrology (JCGM), Evaluation of measurement data – Guide to the expression of uncertainty
in measurement (GUM) (GUM 1995 with minor corrections). (Sèvres, France: International Bureau of Weights and Measures
[BIPM]-JCGM 100], September 2008). Available at http://www.bipm.org/en/publications/guides/gum.html.
Even though the electronic version of the 2008 edition of the GUM is available free of charge on the BIPM’s website,
copyright of that document is shared jointly by the JCGM member organizations (BIPM, IEC, IFCC, ILAC, ISO, IUPAC,
IUPAP and OIML).
4 Joint Committee for Guides in Metrology (JCGM), International vocabulary of metrology – Basic and general concepts and
associated terms (VIM), 3rd ed. (Sèvres, France: International Bureau of Weights and Measures [BIPM]-JCGM 200, 2012)
(2008 with minor corrections). Available for download at http://www.bipm.org/en/publications/guides/vim.html.
Even though the electronic version of the 3rd edition of the VIM is available free of charge on the BIPM’s website, copyright
of that document is shared jointly by the JCGM member organizations (BIPM, IEC, IFCC, ILAC, ISO, IUPAC, and OIML).
5 International Organization for Standardization (ISO), ISO/IEC 17025:2005 General requirements for the competence of
testing and calibration laboratories (Geneva, Switzerland: ISO, 2005). Available for purchase at
http://www.iso.org/iso/home/store/catalogue_ics.htm or from other authorized distributors.
Acquisition and use of that document requires a User License Agreement with ASCLD/LAB and an ISO/IEC 17025
Certification Statement. Acquisition forms are available at www.ascld-lab.org or by contacting ASCLD/LAB. Contact
information for ASCLD/LAB is provided on page 2 of this document.
7 ASCLD/LAB, ASCLD/LAB-International Program Applications, Guidance and Board Interpretations [for] Forensic
Science Testing Laboratories (Web only access, available at www.ascld-lab.org).
8 ASCLD/LAB, ASCLD/LAB Guidance on the Estimation of Measurement Uncertainty - Overview (AL-PD-3061) (Garner,
North Carolina: ASCLD/LAB, 2013). Available at www.ascld-lab.org.
9 ASCLD/LAB, ASCLD/LAB Guidance on the Estimation of Measurement Uncertainty - Annex A Details on the NIST 8-
Step Process (AL-PD-3062) (Garner, North Carolina: ASCLD/LAB, 2013). Available at www.ascld-lab.org.
10 SWGDRUG, SUPPLEMENTAL DOCUMENT SD-3 FOR PART IVC – Quality Assurance/Uncertainty Measurement
Uncertainty for Weight Determinations in Seized Drug Analysis.(SWGDRUG 2011-07-07) Available at
http://www.swgdrug.org/supplemental.htm
13 National Institute of Standards and Technology, NISTIR 6919 - Recommended Guide for Determining and reporting
Uncertainties for Balances and Scales, (Gaithersburg, Maryland, January 2002). Available for download at
http://www.nist.gov/pml/wmd/pubs/nistir.cfm
14 ASCLD/LAB, ASCLD/LAB Guidance on Measurement Traceability (AL-PD-3057) (Garner, North Carolina: ASCLD/LAB,
2013). Available at www.ascld-lab.org.
15 VIM Definition 2.13 – measurement accuracy: closeness of agreement between a measured quantity value and a true
quantity value of a measurand. NOTE 1: The concept ‘measurement accuracy’ is not a quantity and is not given a numerical
quantity value.
Source: Joint Committee for Guides in Metrology (JCGM), International vocabulary of metrology – Basic and general
concepts and associated terms (VIM), 3rd ed. (Sèvres, France: International Bureau of Weights and Measures [BIPM]-JCGM
200, 2012) (2008 with minor corrections).
16 VIM Definition 2.15- measurement precision: closeness of agreement between indications or measured quantity values
obtained by replicate measurements on the same or similar objects under specified conditions. NOTE 1: Measurement
precision is usually expressed numerically by measures of imprecision, such as standard deviation, variance, or coefficient of
variation under the specified conditions of measurement.
Source: Joint Committee for Guides in Metrology (JCGM), International vocabulary of metrology – Basic and general
concepts and associated terms (VIM), 3rd ed. (Sèvres, France: International Bureau of Weights and Measures [BIPM]-JCGM
200, 2012) (2008 with minor corrections).
Source: Joint Committee for Guides in Metrology (JCGM), International vocabulary of metrology – Basic and general
concepts and associated terms (VIM), 3rd ed. (Sèvres, France: International Bureau of Weights and Measures [BIPM]-JCGM
200, 2012) (2008 with minor corrections).
18 National Institute of Standards and Technology, SOP 29 – Standard Operating Procedure for the Assignment of
Uncertainty, (Gaithersburg, Maryland, February 2012). Available for download at
http://www.nist.gov/pml/wmd/labmetrology/upload/SOP_29_20120229.pdf
19 S L R Ellison and A Williams (Eds). Eurachem/CITAC Guide: Quantifying Uncertainty in Analytical Measurement, Third
edition, (QUAM: 2012 P1) Available for download at http://www.eurachem.org/index.php/publications/guides
20 National Institute of Standards and Technology, GMP 3 - Good Measurement Practice for Method of Reading A
Meniscus Using Water or Other Wetting Liquid, (Gaithersburg, Maryland, March 2012). Available for download at:
http://www.nist.gov/pml/wmd/labmetrology/upload/GMP_3_20120314.pdf
22 VIM Definition – 5.12 calibrator: measurement standard used in calibration. NOTE: The term “calibrator” is only used in
certain fields.
Source: Joint Committee for Guides in Metrology (JCGM), International vocabulary of metrology – Basic and general
concepts and associated terms (VIM), 3rd ed. (Sèvres, France: International Bureau of Weights and Measures [BIPM]-JCGM
200, 2012) (2008 with minor corrections).
23 VIM Definition – 3.2 measuring system: set of one or more measuring instruments and often other devices, including any
reagent and supply, assembled and adapted to give information used to generate measured quantity values within specified
intervals for quantities of specified kinds
Source: Joint Committee for Guides in Metrology (JCGM), International vocabulary of metrology – Basic and general
concepts and associated terms (VIM), 3rd ed. (Sèvres, France: International Bureau of Weights and Measures [BIPM]-JCGM
200, 2012) (2008 with minor corrections).
ASCLD/LAB NOTE: For interpreting and applying 3.3.2 of this policy, a gas chromatograph is an example of a measuring
system.
NOTE: Regardless of the number of digits that are showing in a cell, Excel carries the maximum number of significant figures in the background and will use the entire number for further calculations
Revision: 1
Step 1
Procedure name and revision:
Step 2
Drug Chemistry No. XXXX, Revision ZZZZ
Estimation prepared by: StepASCLD/LAB
3 Guidance Document Annex B Date Prepared: April 1, 2013
Degrees Component
Line Uncertainty Freedom Standard Contribution 70
Item Component Value Units Distribution Type Divisor (n-1) Uncertainty % 60
1 Measurement Process Reproducibility 0.031 gram A normal 1.00 >200 0.031 63 50
2 Display Resolution - rounding at zero 0.005 gram B rectangular 1.73 ∞ 0.003 6 40
3 Display Resolution - rounding at load 0.005 gram B rectangular 1.73 ∞ 0.003 6 30
4 Balance calibration uncertainty 0.013 gram B normal 2.00 ∞ 0.007 13
20
5 Linearity - allowed tolerance of accuracy checks 0.01 gram B rectangular 1.73 ∞ 0.006 12
Step 5
10
Combined Standard Unc u 0.032 100
0
Expanded Unc in grams U (k = 2.000) 0.065
Step 6
1 2 3 4 5
Expanded Unc in grams U (k = 3.000) Step 4 0.097
Reported Uncertainty (gram): 0.07 gram k = 2.000
Step 8
Reported Uncertainty (gram): 0.10 gram k = 3.000
Step 7
NOTE: Regardless of the number of digits that are showing in a cell, Excel carries the maximum number of significant figures in the background and will use the entire number for further calculations
Revision: 1
NOTE: Regardless of the number of digits that are showing in a cell, Excel carries the maximum number of significant figures in the background and will use the entire number for further calculations
Revision: 1
Step 1
Procedure name and revision: Drug Chemistry No. XXXX, Revision ZZZZ
Estimation prepared by: Step 2 Step 3ASCLD/LAB Guidance Document Annex B Date Prepared: April 1, 2013
Degrees Component
Line Uncertainty Freedom Standard Contribution 70
Item Component Value Units Distribution Type Divisor (n-1) Uncertainty % 60
1 Measurement Process Reproducibility 0.031 gram A normal 1.00 >200 0.031 60
50
2 Display Resolution - rounding at zero 0.005 gram B rectangular 1.73 ∞ 0.003 6
3 Display Resolution - rounding at load 0.005 gram B rectangular 1.73 ∞ 0.003 6 40
4 Balance calibration uncertainty 0.013 gram B normal 2.00 ∞ 0.007 13 30
5 Linearity - allowed tolerance of accuracy checks 0.01 gram B rectangular 1.73 ∞ 0.006 11 20
6 Mass reference standard calibration uncertainty 0.0055 gram B normal 2.00 ∞ 0.003 5 10
Step 5
1 2 3 4 5 6
U (k = 3.000)
Step 4
Expanded Unc in grams 0.098
Reported Uncertainty (gram): 0.07 gram k = 2.000
Step 8
Reported Uncertainty (gram): 0.10 gram k = 3.000 Step 7
NOTE: Regardless of the number of digits that are showing in a cell, Excel carries the maximum number of significant figures in the background and will use the entire number for further calculations
Revision: 1
Revision: 1
Step 1
Procedure name and revision:
Step 2 Test Method No. XXXX, Revision ZZZZ
Step 3
Estimation prepared by: ASCLD/LAB Guidance Document Annex B Date Prepared: April 1, 2013
Degrees Component
Line Uncertainty Freedom Standard Contribution 150
Item Component Value Units Distribution Type Divisor (n-1) Uncertainty %
100
1 Graduated Cylinder Scale Readability 1.0 mL Rectangular B 1.73 ∞ 0.577 96
2 Graduated Cylinder Calibration Uncertainty 0.050 mL Normal B 2.00 ∞ 0.025 4 50
Step 5
1 2
Expanded Unc U (k = 3.000) Step 4 1.734
Reported Uncertainty: 1 mL k = 2.000
Step 8
Reported Uncertainty: 2 mL k = 3.000
Revision: 1
Revision: 1
Step 1
Procedure name and revision: Procedure No. XXXX, Revision ZZZZ
Step 2
Estimation prepared by: Step 3 ASCLD/LAB Guidance Document Annex B Date Prepared: April 1, 2013
Degrees Component
Line Uncertainty Freedom Standard Contribution 60
Item Component Value Units Distribution Type Divisor (n-1) Uncertainty %
50
1 Measurement Process Reproducibility - %RSD 6.0 % normal A 1.41 25 4.24 40
2 Homogeneity 10.0 % rectangular B 1.73 ∞ 5.77 55 40
3 Purity of reference material 0.5 % rectangular B 1.73 ∞ 0.29 3 30
4 Uncertainty of pipette calibration 0.18 % normal B 2 ∞ 0.09 1
20
5 Uncertainty of volumetric flask calibration 0.0344 % normal B 2 ∞ 0.02 0
6 Error in volume of volumetric flask 0.14 % rectangular B 1.73 ∞ 0.08 1 10
Step 5
Step 4 1 2 3 4 5 6
Expanded Unc U (k = 3.33 for n=26; degrees of freedom = 25) 23.88
Reported Uncertainty: 15 % k = 2.11
Step 8
Reported Uncertainty: 24 % k = 3.33
Step 7
The basis for the data above:
1 Measurement Process Reproducibility - this entry is a combined estimation for uncertainty components listed in Step 3 as Type A method of evaluation
This calculation is also based on two replicate samples and the average of the two being reported
Assumes a normal distribution
2 Homogeneity - laboratory control of 10.0 %
3 Purity of reference material - stated purity is greater than or equal to 99%.
99% purity is the minimum and 100% pure the maximum. Therefore, the range can be expressed as 99.5% +/- 0.5%.
4 Uncertainty of pipette calibration = 1.8 µL; at useable volume of 1000 µL this equals a relative percentage uncertainty of 0.18%; certificate indicates k = 2; approximate
95% confidence level
5 Uncertainty of volumetric flask calibration = 0.0086 mL; at volume of 25 mL this equals a relative percentage uncertainty of 0.0344%; certificate indicates k = 2;
approximate 95% confidence level
6 Error in volume of volumetric flask = 0.0350 mL; Equals a relative percentage error for a 25 mL flask of 0.14%
Revision: 1