Syringe Pumps Ap 12 / 22 Operating Manual
Syringe Pumps Ap 12 / 22 Operating Manual
OPERATING MANUAL
0197
Dear Customer,
thank you very much for purchasing medical equipment from Ascor S.A.
We can assure you that you have made a good choice. Our Company
devotes much effort and time to improve its products. It will be a
pleasure and satisfaction for us if our product fulfils your requirements
and expectations. If during operation of our equipment you have any
doubts or remarks regarding its operation or functionality, please do not
hesitate to inform us about them. We will do our best to remedy all your
problems and your remarks will be taken into consideration in designing
our new products.
Table of Contents
1. Be sure to read this section! 4
3. Pump construction 8
3.1 Pump AP 12 8
3.2 Pump AP 22 9
4. Pump functioning 10
7. Manufacturer’s Responsibilities 30
8. User Tests 30
Before starting the pump operation, read carefully this instruction manual.
• The pump may only be operated by qualified medical personnel, familiar with this
instruction manual and/or trained by authorized personnel of Ascor S.A.
CODAN 10 ml,
10 CODAN/ONCE CODAN Medical System
ONCE 10 ml
3. Pump construction
3.1 Pump AP 12
3.2 Pump AP 22
4. Pump functioning
In order to limit the risk of mistake, pumps AP are equipped with a sensor detecting the
syringe size (10, 20-30, 50ml) and comparing them with preset parameters.
Pumps AP are also equipped with a sensor for detecting the correct fixing of syringe
plunger in pump arm, which makes it impossible to start the pump in case of incorrect fixing
of the syringe.
Operation of the pump consists in moving the syringe plunger with user-preset speed,
which corresponds to a defined infusion flow rate. The pump arm is driven by a step motor,
whose speed of rotation is controlled by a microprocessor. The step motor drives the guide
screw, which causes the arm movement.
The internal microprocessor calculates with high accuracy the speed of rotation of the
motor in accordance with properly preset infusion parameters. Moreover, the microprocessor
controls infusion timing and volume, displays information and messages, checks for
occlusion and monitors the battery status.
The operation of the microprocessor is supervised by additional safety circuit, so called
“watchdog”, which stops the pump operation and activates an alarm in case of detecting any
irregularities.
Preparing the pump for operation needs performing a few simple steps. The pump can
be fixed to a pole (diameter between 20 and 40 mm) or placed on a horizontal surface near
the patient’s bed. The pump can be positioned above, below or at patient level.
It is possible for the pump to be mounted on a horizontal tube (e.g. by patient’s bed).
This can be done by removing two screws that hold the clamp (Fig. 1, pos. 5) turn the clamp
o
by 90 and replace the two fixing screws. Make sure that the clamp is correctly fixed to the
pump body and begin the installation on a horizontal tube.
Attention! With this positioning there is a risk that the pump turns downwards, e.g. during
setting up, if the clamp is not tightened correctly.
The pump is ready for inserting the syringe and setting up infusion parameters.
General notes
• Syringe, as a disposable component, should not be used longer than 24 hours.
• Attention! Removing the full syringe, which is connected with patient’s body
through the extension line, is a danger for the patient. Lifting such a syringe more
than 30 cm over the level of needle position (in patient’s body) may result in intrinsic
infusion caused by the under-pressure in the syringe.
• Attention! The above described situation of free flow of the drug from the syringe to
patient’s vein may also appear when the pump is positioned more than 30 cm above
the level of needle in patient’s vein, and when the syringe plunger is incorrectly fixed
on the pump arm.
• The syringe should be fixed in such way that its stem end is inside of pump arm end
(Fig.1, pos. 1), and the syringe wings are in the gap between the casing and the
locking plate (Fig. 1, pos. 10). Then the syringe must be locked with clamp (Fig.1,
pos. 2) positioning it perpendicularly to syringe axis. Make sure the syringe plunger
is positioned along syringe axis
• After inserting the syringe with drug, pressing „<” button will cause slow movement
of the plunger, which fills the drain with the fluid. Care should be taken to remove all
air from the drain – infusion fluid must flow out of the needle – before the needle is
inserted into patient’s vein. Manufacturer recommends filling the extension line
with fluid using the pump, as described above.
• To achieve proper accuracy for activation of „empty syringe” alarm, the pump, from
time to time, carries out a self-test in order to make corrections to the arm setting.
The display shows the following message:
The pump arm should then be moved completely back using „>” button.
It should be kept in mind that the pump detects occlusion only as soon as the pump arm is
stopped by excessive pressure in the extension line and syringe. As the pressure increases
gradually, due to expansion of the extension line, occlusion indication is delayed, and the
delay depends on the infusion flow rate and length and flexibility of the extension line. In
order to shorten the time of occlusion detection as well as reduce volume of the drug
collected in the expanding extension line, usage of special high-pressure, short, low internal
diameter and thick-wall extension lines is recommended, especially with low infusion flow
rates.
- Class BF device
- Separate collection
Infusion parameters are set up the same way for both types of pumps. For pump AP 22
setup is carried out separately for syringe A and syringe B.
Infusion parameters can be divided into three groups:
1. Basic parameters:
• syringe type,
• flow rate,
• volume or time of infusion,
• drug name
or:
• syringe type,
• volume,
• time,
• drug name
Additionally, the following parameters may be used:
• infusion time,
• time of pause,
These parameters are activated during setup of parameters included in group of additional
parameters (see below).
2. Auxiliary parameters:
• BOLUS rate,
• BOLUS volume,
• occlusion pressure.
3. Additional parameters:
• password,
• infusion rate in K.V.O. mode,
• volume and type of sound alarm,
• drug library modification,
• STAND-BY function on/off,
• ANTI–BOLUS function on/off,
• name of hospital ward.
Parameters of the first group must be set or confirmed before each infusion.
The second group includes auxiliary parameters, which can (but need not to) be set for each
infusion. They can be changed during infusion, if necessary.
The third group includes parameters that are not set very often. They can be changed only
before starting the infusion.
After starting the pump with ON/OFF button, the following message is displayed:
informing the user of the self-checking tests being carried out by the pump.
If any irregularities are detected, the pump is automatically locked and a sound alarm is
activated. If such situation appears repeatedly, please contact our maintenance personnel,
in order they carry out inspection or repair.
As soon as the self-testing is completed, the pump starts its work preparation cycle.
The following situations are possible:
lamp is flashing red. If the pump should be powered from an internal battery, press YES.
Otherwise he pump should be connected to the mains.
Confirming with YES button starts procedure of changing syringe type. Selection can be
made by pressing NO repeatedly until the right type of syringe appears on the display.
The choice should be confirmed by pressing YES button.
The next parameter to be set is:
Flow rates below 100 ml/h can be set with accuracy of 0,1 ml. Higher ones – with accuracy
of 1 ml.
The pump enables volume over time operation. This can be done by pressing NO button
(at the moment when rate is displayed) thus deleting the infusion flow rate, and pressing NO
button once again for abt. 1 second. The display will show:
When setting up the infusion flow rate, it should be kept in mind that, depending on
previously set type of syringe, maximum infusion flow rates are: 500 ml/h (for syringes
50 ml), 300 ml/h (for syringes 30 ml), 250 ml/h (for syringes 20 ml) and 150 ml/h (for
syringes 10 ml). If the flow rate setting is too high, the system informs of this fact by
displaying the maximum acceptable value for the chosen syringe type:
This message can be deleted by pressing any key (it disappears automatically after approx.
5 seconds), and then correct value can be entered and confirmed by pressing YES button.
If the previous parameter was flow rate, and not volume and time, the next parameter would
be:
volume XX.X ml
The field XX.X shows the volume of the previously performed infusion. Pressing NO will
delete this value and allow for entering a new one, which should be confirmed by pressing
YES button. The maximum value of infusion volume is 1000 ml. Leaving this parameter
out (by pressing YES button without entering any number) or setting it to zero is
acceptable, but in such case “END OF INFUSION” alarm will not be activated (see
chapter 5.5), and the pump will not stop working until the syringe is emptied.
Instead of infusion volume, it is possible to set up the time of infusion. Obviously, only one of
these two parameters can be set. Change from volume to time setting (or vice versa) is
achieved by pressing NO button, in order to delete the existing value, and then
pressing it again and holding for more than approx. 1 second. As a result, the following
will be displayed:
The final parameter to be set is drug name. The following question is displayed:
choose a drug ?
Pressing NO will automatically finish setting up the primary parameters, which means that a
correct syringe is to be inserted and a programmed infusion started.
If, however, YES button is pressed, the display will show the name of the drug infused
previously. Drug library can be scrolled by pressing NO repeatedly. When a desired drug
name is found, press YES to confirm it.
In case a desired drug name is not found, pressing PROG button will take the operator back
to the previous question: “choose a drug?”. Pressing NO will finish setting up the primary
parameters.
The drug library is described in more details in Chapter 5.2.3. (page 21,22).
If up to that moment the syringe has not been fixed on the pump assembly, correct settings
of the pump are indicated alternately with the following message:
and automatic adjustment (movement right) of the pump arm. It should be remembered
that the syringe should be filled with abt. 2-3 ml of drug more to fill the extension line
at the moment when FILL THE LINE message is displayed. In some cases, instead of
the above message, the following can be displayed:
wait !!!
This means, that the pump is carrying out a self-test in order to correct the stetting of the
pump arm. Wait a moment and then continue, following the messages displayed.
Pumps AP have a sensor for detecting syringe sizes (10, 20-30, 50ml) and comparing them
with the preset types. If a syringe of volume different than the preset one is inserted, a
message will be displayed:
WRONG SYRINGE !
When the syringe is inserted and clamped on the pump, the following message is displayed:
READY to run .. ??
means that the pump is ready for operation.
To start the infusion, button YES or START should be pressed.
Pressing the PROG button, it is possible to check (read) the preset infusion parameters. If
any of them is incorrect, it can be deleted with NO button and correct value can be entered.
After any change it should be confirmed with YES button.
All primary parameters, except for syringe type, can be changed at any time, also
during the infusion process.
Its value must not exceed the volume of the syringe chosen.
If the BOLUS volume value is not preset, infusion in BOLUS mode is continued as
long as BOLUS button is held. The volume of the delivered dose in ml is displayed. The
volume of fluid administered during activated BOLUS mode adds up to total volume of drug
administered to the patient.
Starting the BOLUS function is described with more details in Chapter 5.4.2. BOLUS
function can be stopped by pressing START/STOP button.
Setting the limit level of infusion pressure.
The next parameter that will be shown after pressing the YES button is pressure level, at
which ”OCCLUSION” alarm should be activated. The reason for activating this alarm may
be, e.g. crushed extension tube (drain) or obstructed syringe needle. Four pressure levels
can be preset:
• minimum 0.04 MPa (300 mm Hg),
• low 0.06 MPa (450 mm Hg),
• medium 0.09 MPa (675 mm Hg),
PRESS minimum
PRESS low
PRESS medium
PRESS high
The selection is made by pressing NO button repeatedly until the right pressure level is
displayed. Pressing YES button will result in setting up and memorizing the selected value.
The accuracy of measuring the occlusion pressure is proportional to the quality and volume
of the syringes used. The lower the volume (diameter) of the syringe, the more difficult an
accurate measurement is. In case occlusion pressure is a critical parameter, usage of large
syringes, at volume of 50÷60 ml, is recommended.
Attention!
• Syringes at volume of 10÷20 ml should only be used for very low infusion flow
rates, i.e. below 1 ÷ 2 ml/hour.
• Low quality syringes (not designed for usage with infusion pumps – with a cone
instead of thread) may cause accidental, false occlusion alarms for the lowest
infusion pressure levels.
This group of parameters enables the user to customize the pump features. Setting up
parameters in this group should begin with switching the pump off using ON/OFF switch.
Then press PROG button and, holding it, switch the pump on with ON/OFF. The following
message will be displayed:
password 0
Entering a number in the range between 1 and 999999 enables protecting the infusion
parameters against any modification by unauthorized persons. If a password has been
entered, then, before any modification of any infusion parameter, a message is displayed:
PASSWORD
Correct password must be entered from the keyboard and confirmed by pressing YES
button. If password protection is not to be used, [0] should be entered as the
parameter and confirmed with YES button.
ATTENTION!
If you forgot the password code 555.555 will unblock the system.
KVO (Keep Vein Open) function consists in the fact that instead of stopping infusion, a
minimum flow rate is maintained in order to keep needle patiency (avoiding clot creation).
Typical flow rate of such infusion is 0.5 ml/h, but in certain cases it should be different. Its
value can therefore be set at 0 - 5.0 ml/h. Leaving this parameter without any value is equal
to setting it to 0,5 ml/h, while entering 0 ml/h will switch this function off. Volume of fluid
received by the patient during activated K.V.O. function adds up to total quantity of drug
received by the patient.
K.V.O mode infusion is activated after stopping the pump with START/STOP button or after
displaying the alarm message ”empty syringe” or ”end of infusion”.
ATTENTION! If the value of flow rate set for K.V.O. function is higher than the basic
infusion flow rate (delivery), then infusion in K.V.O. mode will be effected with flow
rate equal to basic infusion (e.g. if basic infusion flow rate = 0.5 ml/h, and K.V.O. flow
rate = 1.5 ml/h, then infusion in K.V.O. mode = 0.5 ml/h, i.e. infusion is not stopped).
The next parameter defines type of signal used as acoustic alarm. The required option can
be chosen with NO button.
or
As soon as the selected option is confirmed with YES button, parameter defining the volume
level of the acoustic signal.
ALARM (1-3,*)
”1” means the lowest level. * – alarm volume increasing over time (press button '.' )
The next two parameters refer to the drug library.
The drug library consists of two lists: “manufacturer’s basic list” and “operator’s extended
list”.
“Manufacturer’s basic list” consists of the drug names entered by the manufacturer and
cannot be modified by the operator. The drug names are listed in alphabetical order and
displayed in small letters.
“Operator’s extended list” can be created by the user by adding new drug names (up to
fourteen). The following question is displayed:
Pressing YES button enables entering a new drug name, which will be displayed in capital
letters. Each letter to be entered can be chosen with < or > buttons and must be
confirmed by pressing YES. The letter order is alphabetical. When a new drug name is
entered, pressing YES will include it in the drug library and then once again message “add
a drug ???” is displayed. It is possible to enter several drug names in a row, as long as the
above question is answered by pressing YES.
This function enables to remove an unnecessary drug name from the “operator’s extended
list”. After pressing YES choose a drug name to be erased by pressing NO repeatedly and
when it is displayed, press YES to erase it.
STAND_BY yes
or
STAND_BY no
enabling to switch on/off the STAND_BY function. With this function infusion is periodically
held up. Choosing ”yes” results in adding two parameters to primary group:
and
XX.XX:XX defines infusion time and infusion breaks in hours. min. sec. The
cycle begins with infusion whose duration is equal to the preset ”infusion”
parameter. After this time an alarm is activated and the infusion is stopped.
Break in infusion begins, its duration being defined by „pause” parameter.
Completion of this period is signaled with another alarm and demand of
continuing the infusion. The cycle is repeated until the patient receives the
set volume of drug or the pump is switched off.
Another additional parameter is ANTI- BOLUS. This function ensures reduction of negative
consequences of occlusion occurring in the infusion fluid transportation system. Activating
this function (see Chapter ALARMS) will cause return of the pump arm, reduction of
pressure in extension line and withdrawal of drug excess from the flexible extension line to
the syringe. Thanks to this, when the line is patent again, there is no uncontrolled flow of
drug under pressure.
ANTI-BOLUS. Yes
Reduce BOLUS ?
If the answer is YES, then the over pressure caused by occlusion will be
automatically reduced. This function reduces the risk of injecting a dose of
drug under too high pressure when the fluid transportation system is again
patent, i.e. so called „occlusion bolus” volume is reduced.
Attention! During infusion break, caused by occlusion, the K.V.O.
function is inactive.
The next parameter, which can be entered, is name of the hospital unit, where the pump is
used.
If NO button is pressed, the name can be modified choosing the required letters with < or
> keys and confirming them with YES button. The characters are arranged in alphabetical
mode. As soon as the name is entered and confirmed with YES button, the following
message will be displayed:
Password
This function enables creating a separate password for the entered hospital unit name.
Entering a number in the range between 1 and 999999 enables protecting the hospital unit
name parameter against changing by unauthorized persons. If a password has been
entered, then, before any modification of the hospital unit name, a message is displayed:
PASSWORD
Correct password must be entered from the keyboard and confirmed by pressing YES
button. If password protection is not to be used, 0 should be entered as the parameter and
confirmed with YES button.
ATTENTION!
If you forgot the password, code 555.555 will unblock the system.
As soon as the presetting procedure is completed, the following messages are displayed
alternately:
They inform of the necessity of fixing a syringe, previously filled with drug, on the pump
assembly.
The pump arm will be adjusted automatically (will move right) to enable inserting the syringe
filled with a drug (the syringe volume + about 2,5 – 3 ml for priming the extension line). So
the syringe should be filled with an according volume of a drug.
with the clamp (Fig. 1, pos. 2) positioning it perpendicularly to syringe axis. Make sure
the syringe plunger is positioned along syringe axis.
Pumps AP have a sensor for detecting syringe sizes (10, 20-30, 50 ml) and comparing them
with preset types. If a syringe of volume different than the preset one is inserted, a message
will be displayed:
The extension line should be filled with infusion fluid by pressing < button in such a way that
all air bubbles are removed from it.
Attention! Before starting the infusion process make sure there are no air bubbles in
the syringe and in the extension tube (drain). Otherwise, they should be removed using
the above procedure.
READY to run....??
5.4 Infusion
5.4.1 Starting and stopping infusion
Having set pump parameters and fixed the syringe, infusion can be started. It can be done
by pressing START/STOP button, which starts or stops the pump operation. Infusion is
indicated by blinking of a yellow lamp described with a word START as well as by the
following alternate messages:
Infusion can also be stopped using ON/OFF button, which switches the pump off. When the
pump is switched on again, it is possible to continue the infusion by pressing YES in reply to
CONTINUE ? message, (the set parameters are memorized by the system). Also the status
of the counter of drug volume administered to the patient is memorized. This method of
stopping infusion is recommended in case infusion has to be stopped for more than
2 minutes.
As soon as the pump delivers the preset volume of the fluid, the following message will be
displayed:
END of infusion
Setting up the next infusion is possible as soon as the empty syringe is removed from the
pump.
Pumps AP 12/22 enable administering a bolus dose to the patient during infusion or when
the infusion has been stopped with START/STOP button. Bolus function will be activated
after 3 seconds from pressing and holding the < button. This will be announced with the
message:
Pressing NO button again will result in return to displaying infusion flow rate. Cyclical
scrolling of information on the infusion status can be executed as often as required, without
affecting the process of infusion. The same method can be used for viewing information
when the pump is stopped as well as when „empty syringe” or „end of infusion” or
„occlusion” alarms are activated.
It is possible to change almost all infusion parameters in running course without the
necessity of the infusion interruption. It is especially easy to change the flow rate. Just enter
a new value from the numeric keypad and confirm it with YES button. Lack of
confirmation/entering with YES button, for safety reasons, will result in keeping the previous
infusion flow rate, which will appear on the display.
Other basic parameters (like: volume or time of infusion) can be changed by pressing PROG
button shortly, thus entering the possibility of scrolling pre-set parameters of infusion. When
the parameter to be modified is displayed, the previous value should be deleted by pressing
NO, and the new value introduced with number keys and confirmed with YES.
The auxiliary parameters can be changed by pressing PROG button (for at least 1 second)
and following the steps used for setting them up.
This does not refer to group of additional parameters (e.g. K.V.O, STAND-BY etc.), whose
change requires switching off the pump. It is also impossible to change syringe type during
the infusion process. This parameter can be modified after stopping infusion and removing
the syringe from the pump.
This procedure is carried out as follows: first, information on the preset volume should be
displayed, and then NO button should be pressed and held for 3 seconds. After this time a
short acoustic signal will sound and the system will start counting volume from zero. Volume
counter will also be reset when NO button is pressed to answer the message
“CONTINUE ?”, appearing during pump operation, or on completion of the infusion.
Attention! Resetting the volume counter during infusion will cause loss of information
on the quantity of drug administered to the patient till that moment.
Situations requiring intervention of the personnel are signaled with acoustic and visual
alarms (display flashing). Acoustic alarm can be switched of using
button. Type of the alarm is described on the display with adequate message.
Five minutes before complete emptying of the syringe the system switches on an acoustic
alarm informing the personnel of the approaching end of infusion. When
5 min. PREALARM
If the preset or resulting infusion time is shorter than 5 minutes – the alarm will not be
activated.
When the drug in the syringe runs low – approx. 0.5 ml before its complete emptying – an
acoustic alarm sounds and the following message is displayed:
EMPTY SYRINGE !!
The pump is automatically switched over to K.V.O. mode. Infusion can be restarted as soon
as a new syringe, filled with the drug, is inserted into the pump. Answering YES to the
message „CONTINUE ?” will enable adding up the fluid volumes of the subsequent
syringes.
It is also possible to empty the syringe completely (including 0.5 ml remnant). To do this,
having reset EMPTY SYRINGE alarm, button START should be pressed. As soon as the
syringe is emptied completely, „empty syringe” alarm will reappear.
If infusion volume or time was set, then as soon as the set value is exceeded, infusion will
be stopped and the pump will be switched over to K.V.O mode. End of infusion will be
signaled with acoustic alarm and messages displayed alternately:
Stoppage of the pump arm movement caused by occlusion is indicated by a sound and the
following message:
OCCLUSION !!!
drug under pressure. The risk of infusing Bolus after occlusion release is thus significantly
reduced.
ANTI-BOLUS is started by pressing [>] button after muting OCCLUSION alarm. Then the
following question appears on the display:
Reduce BOLUS ?
If the answer is YES, the pump will automatically reduce the overpressure caused by
occlusion.
Attention! During the infusion break, caused by occlusion, K.V.O. function is inactive.
Occlusion alarm may also appear as a result of increased infusion resistance caused by
high density of the fluid delivered at high flow rates. In this case, the preset occlusion
pressure should be increased or the syringe used should be changed. Also flexible tube
(drain) of greater internal diameter is recommended.
The accuracy of measuring the occlusion pressure is proportional to the quality and volume
of the syringes used. The lower the volume (diameter) of the syringe, the more difficult an
accurate measurement is. In case occlusion pressure is a critical parameter, usage of large
syringes, at volume of 50÷60 ml, is recommended.
Attention! Syringes at volume of 10÷20 ml should only be used for very low infusion
flow rates, i.e. below 1 ÷ 2 ml/hour.
Having reset “OCCLUSION” alarm and pressing NO button it is possible, just as during
infusion, view all information on the current infusion, i.e. entered volume, time till emptying
the syringe, flow rate, occlusion pressure, battery status. As soon as the cause of occlusion
is removed, pressing the START/STOP button allows for continuing the process.
If during infusion the syringe is removed by the personnel or (accidentally) by the patient, the
following message will be displayed:
NO SYRINGE !!
and alarm will be activated. After resetting the alarm and inserting the syringe, the question
will appear:
Continue..... ?
Pressing YES button means completing the stopped infusion. Pressing NO button means
preparing the pump for a new infusion.
Alarm:
NO MAINS !!!
and flashing of red lamp
informs the operator of the fact that the pump has automatically switched over to power
supply from internal battery. This could happen as a result of mains failure, fuse blow-out,
disconnecting the supply cable or switching off the main switch (Fig.1 pos. 9). Pressing YES
button means confirming this situation.
Alarm:
LOW BATTERY !!
means that the battery will be completely discharged within 30 minutes (at medium infusion
flow rates, i.e. approx. 5 ml/h). The pump should be connected to the mains as soon as
possible. If the infusion is continued without connection to the mains, the infusion message
“inf. XX.X ml/h” will be displayed alternately with the following message:
• Infusion must be carried out only with the syringe whose name has been set. Using
another syringe, even if its appearance and volume are identical, does not
guarantee achieving the set infusion parameters with accuracy indicated in the
pump technical data. This may have an impact on patient’s health.
• Usage of syringes with thread for fixing the extension line – Luer Lock – is
recommended. The thread secures the extension line against slipping off, e.g. when
the pressure increases because of occlusion.
• Attention ! The pump does not have its own system for detecting air in the line. The
pump user must check if there are any bubbles of air in the extension or in the
syringe. To fill the extension line with fluid, „<” button should be used to shift the
pump arm.
• During operation the pump should be positioned below or at patient level. This
secures the patients against free flow of drug from the syringe to his vein in case the
syringe is removed from the pump → see General Notes on Page 13.
• Connection of the pump to the mains is indicated by a green lamp
It is recommended that the pump be connected to the mains also when it is not
operated, in order to continue recharging of the pump battery (the pump can stay
connected to the mains for any time). This guarantees a complete recharging of the
battery.
• It is also recommended that the pump be supplied from the battery only in situations
where mains supply is impossible e.g. power failure, transporting the patient. Time
necessary for full recharging the completely discharged battery (message: VERY
LOW BATT !!!) is approx. 24 hours. The pump can be normally operated during
recharging of the battery.
• If electric voltage is present in the socket, and the message NO MAINS !!! is
displayed, it may mean that:
1. the main switch is in position “0” – pump is switched off;
2. the fuse is blown;
7. Manufacturer’s Responsibilities
Manufacturer is responsible for safe, both for patient and user, and correct operation of the
equipment, according to the technical data detailed in this instruction manual, on condition
that:
• the equipment is operated in accordance with its use and in environment conditions
proper for this type of equipment.
• installation of the equipment is performed in accordance with requirements included
in this instruction manual,
• the equipment is operated in accordance with this instruction manual by trained
medical personnel,
• inspections, repairs and modifications were only made by authorized servicing
company.
Manufacturer recommends carrying out a technical inspection before the end of guarantee
period, and later on every two years.
8. User Tests
Pumps AP 12/22 are equipped with a set of tests with which it is possible to check correct
operation of its main subassemblies. These tests can be helpful for evaluating the condition
of the pump, but they do not guarantee detection of all existing faults. In case of any doubts
regarding the condition of the equipment or accuracy of its operation, the pump should be
immediately delivered to the hospital maintenance department in order to check it and/or
take decisions as regards further steps to be taken, e.g. repair by the authorized servicing
company.
To start User Tests the pump should be switched on with ON/OFF button and then
START/STOP button should be pressed immediately. The display will show a broken line:
-------------------------------------
Enter 0 and confirm with YES button. This will activate the user test module. The first test
will be displayed. To carry out the test, select it pressing YES button. Otherwise press NO
button repeatedly until the right test is displayed.
ALARM test ?
Alarm on
Alarm off
Test can be stopped with YES button. During this test it is possible to check if the pump is
correctly connected to the hospital personnel calling system.
2. Display test.
DISPLAY test.?
It is carried out automatically until the following is displayed:
END OF TEST
Check if the lamps on the control panel are flashing and if the characters are properly
displayed.
3. Keyboard Test.
KEYBOARD test ?
Press buttons indicated on the display. If the keyboard operation is correct, the following
message appears on the display:
END OF TEST
END OF TEST
The following message:
DAMAGE !!!
means that the pump should be sent to an authorized servicing company. Test can be
stopped with YES button.
In order to stop the testing procedures the pump should be switched off using the ON/OFF
button and started again with this button in order to continue the operation.
In order to finish the user tests procedure, the pump should be switched off with ON/OFF
button.
Any repairs of the pumps can be performed only by a trained servicing company indicated
by the manufacturer or distributor of the equipment. Manufacturer will secure after-
guarantee servicing of syringe pumps AP 12 and AP 22.
start-up-graph
trumpet curve
start-up-graph
trumpet curve
Test of dosing accuracy for the syringe
30 ml B-D PLASTIPAK
Performed in accordance with EN 60601-2-24 test 50.102
Flow rate 1ml/h and 5 ml/h. Liquid density 0,99973 [g/ml]
start-up-graph
trumpet curve
start-up-graph
trumpet curve
start-up-graph
trumpet curve
start-up-graph
trumpet curve
The devices meet the requirements of the Council Directive 93/42/ EEC
- Medical Devices and are CE-marked.
.
Guidance and manufacturer’s declaration – electromagnetic
emissions
The AP 12 AND 22 is intended for use in the electromagnetic environment specified below.
The customer or the user of the AP 12 AND 22 should assure that it is used in such an
environment.
Emissions test Compliance Electromagnetic
environment - guidance
RF emissions Group 1 The AP 12 AND 22 uses
energy only for its internal
CISPR 11
function. Therefore, its RF
emissions are very low and
are not likely to cause any
interference in nearby
electronic equipment.
RF emissions Class B
CISPR 11
Harmonic emissions Class A
IEC 61000-3-2
Voltage fluctuations / flicker Complies
emissions
IEC 61000-3-3
d = 1,2 P
NOTE 1: AT 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 : These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects, and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To asses the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the AP 12 AND 22 is used
exceeds the applicable RF compliance is observed, additional measures may be necessary,
such as reorienting or relocating the AP 12 AND 22
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 2,9 V/m.
Manufacturer:
ASCOR S.A.
8, Mory Street
01-330 Warsaw, Poland
tel.: +48-22-836-83-74
fax.: +48-22-836-14-96
e-mail: [email protected]
web side: www.ascor.com.pl
Model:
Serial no: