Non Conformance Form
Non Conformance Form
CP N-05 : Building and Civil Engineering Work for Depot Building Including
Related Infrastructure and Track Work Ballast
(Contract No : NSCREX-
03-N05)
Drawing/Specification No.
It has been observed that the Universal Testing Machine (SERIAL NO.16801) used in the testing of rebars was giving a discrepancy/ inaccurate reading during the
recently conducted retest/validation of the 36-mm rebar yield strength of the rebar from STEEL ASIA .After the completion of the test it was found that the final reading
reflected on the measuring device screen is showing a reading that is lower than the actual reading being shown on the stress/strain graph of the machine during the
test. (See attached photos). UTL laboratory is instructed to investigate, identify and correct as necessary the cause of this uncertainties in their UTM.
a) Calibrated or verified, or both, at specified intervals, or prior to use, against measurement standards traceable to international or national measurement standards;
when no such standards exist, the basis used for calibration or verification shall be retained as documented information;
b) Identified in order to determine their status;
c) safeguarded from adjustments, damage or deterioration that would invalidate the calibration status and subsequent measurement results.
The organization shall determine if the validity of previous measurement results has been adversely affected when measuring equipment is found to be unfit for its
intended purpose, and shall take appropriate action as necessary.
INITIATOR: REVIEW/VALIDATION
IMMEDIATE CORRECTION:
SIGNATURE
NAME
DATE
SIGNATURE
NAME
POSITION
DATE
Name:
Signature: Date: