Analytical Last
Analytical Last
143
Absorption of IR radiation by molecule
Molecules are made up of atoms linked by chemical bonds.
∆E = hv
146
There infrared spectrum have two fundamental region.
(a) Functional group region (4000 cm-1 - 1250 cm-1)
dependent only on the functional group that absorbs
(b) Fingerprint region (1250 cm-1 - 600 cm-1).
dependent on the complete molecular structure
147
The IR spectrum is plot is % transmittance against wavenumber
148
Instrument
Design
149
Types of Molecular Vibrations
The positions of atoms in a molecules are not fixed; they are subject to a
number of different vibrations.
Vibrations fall into the two main categories: stretching and bending.
Symmetrical Asymmetrical
stretching stretching
150
Bending: Change in angle between two bonds. There are four types of
bend.
151
Theoretical number of fundamental modes of vibration
It is possible to deduce the number & kinds of vibrations in simple
diatomic and triatomic molecules & whether vibrations will lead to
absorption.
152
In a non-linear molecule, 3 of these degrees of freedom are rotational, 3
are translational and the remainder is fundamental vibrations.
153
Vibrational Frequency
In diatomic molecule with two masses m1 and m2 connected by a
massless spring the stretching frequency of a bond can be approximated
by Hooke’s Law.
155
Quantum Treatment of Vibrations
Using the harmonic oscillator and wave equations of quantum
mechanics, the energy can be written as
157
The frequency of radiation v that will bring about this change is
identical to the classical vibrational frequency of the bond.
Or
158
k is the force constant (N/m), c (cm/s), and µ is the reduced mass (kg).
Generally, k for most single bonds ranges average value of 5 x 102 N/m
and double - 1 X 103 and triple bonds-1.5 X 103 N/m.
159
Example:
Calculate the approximate wavenumber and wavelength of the
fundamental absorption due to the stretching vibration of a carbonyl
group C=O.
160
and the reduced mass µ is given by
As noted earlier, the force constant for the typical double bond is about
1 x 103 N/m. Substituting this value and µ into the equation gives
162
(a) Dispersive IR spectrophotometers
It is a double-beam like spectrophotometer, recording instruments,
which use reflection gratings for dispersing radiation.
163
The grating separates the wavelengths of light in the spectral range and
directs each wavelength individually through a slit to the detector.
Each wavelength is measured one at a time, with the slit monitoring the
spectral bandwidth and the grating moving to select the wavelength being
measured.
In the 1870’s Michelson was measuring light and its speed with great
precision and reported the speed of light with the greatest precision to
be 299,940 km/s and for this he was awarded the Nobel Prize in 1907.
Then the MI recombines the two beams and conducts them into the
detector where the difference of the intensity of these two beams are
measured as a function of the difference of the paths.
The detector now reports variation in energy vs time for all wavelengths
simultaneously.
167
A plot of the signal versus mirror displacement is the interferogram.
168
This is the "raw data" which can be Fourier transformed to get the
actual spectrum.
169
170
Procedures of data acquisition using FTIR instrument
1. Reference scan
First reference interferogram is obtained by scanning a reference
empty sample compartment (usually air) .
Then, the reference spectrum is calculated by co-adding the data,
and storing the results in the memory of the instrument computer
(usually after transforming it to the spectrum).
2. Sample scans
A sample is then inserted in the radiation path (in the sample
compartment) and the sample spectrum is obtained.
The ratio of sample and reference spectral data is then computed to
give the transmittance at various frequencies.
From this ratio, the absorbance is calculated as a function of
wavenumber. 171
Advantages of FTIR instrument
Have higher signal-to-noise ratio for a given scan-time → because
information from all wavelengths is collected simultaneously.
172
IR SOURCES AND TRANSDUCERS
(1) Light source
IR sources consist of an inert solid that is heated electrically to a
temperature between 1500 and 2200 K.
This heated material will then emit IR radiation. The following are
some of the source.
Instrument
Design
177
(c) Solid samples
Solid dissolved in solvent
a convenient way of obtaining IR spectra is on solutions
prepared to contain a known concentration of sample.
178
Most organic compounds exhibit numerous absorption bands throughout
the mid-IR region, and hence finding a solvent that does not have
overlapping bands is often impossible.
On the other hand the solid sample must be ground until its particle size
is much less than the wavelength of the radiation to avoid the effects of
scattered radiation.
There are some techniques for handling solid samples for handling solid
samples
Pellet techniques
Mull techniques
Thin film
179
(a) Pellet Technique (Disk Method)
• The small amount of finally ground solid sample is mixed with 100 times
its weight of powdered KBr.
• Although KBr is the most frequently used pelleting salt, materials such
as CsI and CsBr are sometimes used.
• This paste is then sandwiched between two salt plates and used for
spectral measurement.
181
Mull is formed grinding a 2.5 mg of the sample & 2-3 drops of nujol oil.
182
(3) Detectors
There are three kinds of IR transducers
Thermal detector
Pyroelectric detector (specialized “thermal detector”).
Photoconducting detector (transducer).
The two fine wires are welded in blackened gold foil (to improve its
heat-absorbing capacity) and sealed in an evacuated chamber with a
window that is transparent to IR radiation (not glass o quartz).
184
Instrument
Design
hn vacuum
semiconductor
Transparent
glass to IR
186
(3) Pyroelectric Detectors
constructed from single crystalline wafers of pyroelectric materials,
which are insulators with very special thermal and electrical properties.
Pyroelectric substances, in contrast, retain a strong temperature-
dependent polarization after removal of the field.
Thus, by sandwiching the pyroelectric crystal between two electrodes,
one of which is IR transparent, a temperature-dependent capacitor is
produced.
Pyroelectric transducers exhibit response times that are fast enough to
allow them to track the changes in the time domain signal from an
interferometer.
For this reason, many FTIR spectrometers for the mid-IR region use this
type of transducer.
187
3.4.3 Applications of Infrared Spectrometry
Infrared spectroscopy is widely used in industry as well as in research.
(classification of unknowns)
188
Different types of groups of atoms (C-H, O-H, N-H, etc…) absorb
infrared radiation at different characteristic wavenumbers.
189
(a) Group Frequencies
Group frequencies for several functional groups fall in the range of 3600-
1250 cm-1 & a few fall in the fingerprint region, such as C-O-C & C-Cl
stretching vibration at 1200 cm-1 & 700 to 800 cm-1 respectively.
In general, the IR spectrum can be split into four regions for interpretation
190
Figure 3D-7 The group frequency region of the mid-IR (3600–1250 cm-1) is used to
identify common functional groups. The fingerprint region (~1200–600 cm-1) is used
to identify compounds.
191
Factors Influencing Group Frequencies
There are various factors that contribute to the position, intensity and
appearance of IR bands
a) Symmetry
b) Coupling
c) Resonance
d) Hydrogen Bonding
e) Ring Strain
f) Electronic Effects
192
(a) Symmetry
For a particular vibration to be IR active there must be a change in dipole
moment during the course of the particular vibration.
+ - vibration + -
C O C O
193
(b) Coupling
In a multi-atomic molecule, coupling vibration for functional groups occur
for adjoining bonds and absent in a molecule with more than two bonds
separated atoms.
For example, the calculated and observed ̅ for most C=C bonds is
around 1650 cm-1
In allene, however, coupling of the two C=C systems gives two IR bands
at 1960 and 1070 cm-1 due to coupling.
H
H
C C C
H
H
194
(c) Hydrogen Bonding
Intramolecular H-bonding with carbonyl compounds can serve to lower
the absorption frequency.
CH3
CH3
O
1680 cm-1 O
1724 cm-1
O
O
H
O
195
Gas phase Neat liquid
The exact position of this band depends to a large extent upon the
degree of hydrogen bonding or association to which the hydroxyl group
is subject.
Upon association, the energy and force constant of the O-H bond
decreases, and the absorption band is therefore shifted to a lower
frequency by about 200 cm-1.
196
120
Vapor phase spectrum(135 °C)
100
80
% Transmittance
liquid film
60
40
20
0
4000 3500 3000 2500 2000 1500 1000
Wavenumbers, cm-1
C
Resonance
X
199
O
C X X= NH2 CH3 Cl NO2
H3C
1677 1687 1692 1700 cm-1
O O
O
N C
H2N C CH3 vs.
O CH3
Sterics
In this case the presence of the methyl group “misaligns” the conjugated
system, and resonance cannot occur as efficiently.
O O
CH3
C=O: 1686 cm-1 C=O: 1693 cm-1
200
Frequency decreases with increasing atomic mass.
201
Classification of IR bands
Infrared spectrum intensity of signals are classified into four groups
Strong signals (s): when signal intensity is approximately 5% T.
Medium signals (m): when signal intensity is between 30 and 60% T.
Weak signals (w): when signal intensity does not exceed 70% T.
202
IR spectra interpretation
There are a few general rules for the determination of a molecular structure.
a. Look first at the high-wavenumber end of the spectrum (>1500 cm−1) and
concentrate initially on the major bands.
b. Do not expect to assign every band in the spectrum.
c. Exploit negative evidence as well as positive evidence. Eg. if there is no
band in the 1850–1600 cm−1 region, it is most unlikely that a carbonyl group
is present.
d. Keep ‘cross-checking’ wherever possible. Eg. an aldehyde should absorb
near 1730 cm−1 and in the region 2900–2700 cm−1.
e. Use the lower-wavenumber end of the spectrum for the confirmation or
elaboration of possible structural elements.
203
The C-H stretching Region
Due to the small electronegativity difference between C and H,
hydrocarbon bands are of medium intensity at best and give simple
spectra.
204
C-H sp stretch ~ 3300 cm-1 UNSATURATED
C-H sp2 stretch > 3000 cm-1
C-H sp3 stretch < 3000 cm-1
C-H aldehyde two ~ 2850 and 2750 cm-1 SATURATED
peaks (both weak)
205
205
The C-H stretching
Methylene group
stretching vibrations
206
Alkanes
There are two types of bonds (C-H and C-C) and the spectra can be
interpreted in terms of four vibrations.
(a) C-H stretching from 3000–2850 cm-1
(b) C-H bendingC–H bend or scissoring from 1470-1450 cm-1
Instrument
(c) C–H rock, methyl from 1370-1350 cm-1
Design
(d) C–H rock, methyl, seen only in long chain alkanes, from 725-720 cm-1
(c) C-C stretching: are weak, & appear in broad region(1200-800 cm-1)
C-C stretching is of little value for identification.
(d) C-C bending: (below 500 cm-1, do not appear in normal spectra).
207
Hexane
208
Alkenes
The spectra of alkenes are usually more complex (more peaks) than the
spectra of alkanes.
Only alkenes and aromatics show a C-H stretch slightly >3000 cm-1.
210
210
Overlay a spectrum of 1-octene with octane
211
Alkynes
The –C≡C– stretch appears as a weak band from 2260-2100 cm-1
A terminal alkyne (but not an internal alkyne) will show a C–H stretch as
a strong, narrow band in the range 3330-3270 cm-1.
A terminal alkyne will show a C–H bending vibration in the region 700-
610 cm-1.
212
In a manner very similar to alkynes, nitriles show a prominent band
around 2250 due to C≡N bond.
This band has a sharp, pointed shape just like the alkyne C≡C bond, but
because the C≡N bond is more polar, this band is stronger than in
alkynes.
The cyano group often gives a strong, sharp peak due to its large dipole
moment.
The C≡C bond gives a sharp peak, but it is often weak due to a lack of a
dipole.
213
1-Hexyne
214
1-Hexyne
215
Aromatic
The aromatic sp2 =C-H stretch occurs at frequencies beyond 3000 cm-1.
Only alkenes and aromatics show a C–H stretch slightly higher than 3000
cm-1. Compounds that do not have a C=C bond show C–H stretches only
below 3000 cm-1
Summary
C–H stretching → 3000-3100 cm-1
C=C (ring) stretching → 1600-1585 cm-1 and 1500-1400 cm-1
C–H in-plane bending → 1250-1000 cm-1
Out of plane =C-H (C-H) stretching (intense) → 900-690 cm-1
216
Aromatic
Toluene
217
O-H stretching
218
Alcohols have characteristic IR absorptions associated with both the O-H
and the C-O stretching vibrations.
The free ‒OH band appears as a sharp signal of medium to low intensity
near 3620 cm-1.
Summary
O–H stretch, hydrogen bonded→ 3500-3200 cm-1
C–O stretch 1260-1050 cm-1 (strong)
219
1-Butanol
220
Cyclohexanol
221
O-H bending
Alcohols (1260-1000 cm–1) Phenols (1800-1260 cm–1)
primary alcohol: 1050-1085 cm-1
secondary alcohol: 1085-1125-1
tertiary alcohol: 1125-1200 cm-1
222
Ethers
The main difference between the IR spectra of ethers and alcohols is
that ethers lack O-H signals.
The C-O stretching allows for the differentiation of ethers from alkanes.
225
IR spectrum of aniline
226
IR spectrum of diethylamine
227
IR spectrum of triethylamine
228
Carbonyl Stretching Region
The C=O absorption is one of the most characteristic in the entire
spectrum, and it is also most likely to be the most intense spectral.
230
Acetophenone
231
Aldehyde
C-H (H–C=O) stretch → 2850-2700 cm-1 (often two bands are observed)
C=O stretch:
aliphatic aldehydes → 1740-1720 cm-1
alpha, beta-unsaturated aldehydes (aromatic) → 1710-1685 cm-1
232
233
Carboxylic acids
O-H stretch → often centered around 3000 cm-1
C=O stretch:
Saturated acids → 1730-1700 cm-1
Aromatic carboxylic acids → 1710-1680 cm-1
234
Benzoic acids
235
Acid anhydrides
C=O stretch: 2 bands separated by 60 -30 cm-1)
Saturated anhydrides, 1750 and 1820 cm-1;
unsaturated anhydrides, 1775 and 1720 cm-1
1750
1820
236
Esters
C=O stretch
aliphatic →1760-1735 cm-1
α, β-unsaturated, →1730-1715 cm-1
phenyl or vinyl esters → 1770-1780 cm–1
237
238
Amides
C=O stretch between 1680-1630 cm-1
N-H stretch
1° amines show two bands (3370-3170 cm-1)
2° amines have one band (3370-3170 cm-1)
3° amines have no absorbance in this region
N-H bend vibrations between 1640-1550 cm-1
239
240
has C=O band (1650-1800 cm-1) does not have
(very strong
IR Spectrum
C=O band
C-H, O-H, and N-H
3700-2500 cm-1
Aldehyde C≡C , C≡N
C=O 2300 - 2000 cm-1 Alkanes
1740-1720 (saturated) sp3 C-H stretch 2850-3000
1710-1685 (unsaturated) sp3 C-H bend 1460 & 1380
C-H Nitriles Alkenes
C≡N ≈2250
28500-2700 sp2 C-H stretch 3000-3100
Instrument
Ketone sp2 C-H bend 650-1000
Alkynes
C=ODesign C≡C
C=C 1600-1660
≈2150
1715 ± 10 (saturated) sp C-H stretch 3300 Aromatics
1700-1640(unsaturated) sp C-H bend 620 sp2 C-H stretch 3050-3150
sp2 C-H bend 690-900
Esters C=C 1600 & 1480
C=O
1760-1735 (saturated) Alcohols
1730-1715 (unsaturated) O-H stretch 3600-3500
Ethers
C-O stretch 1120 (alphatic)
241
1040 & 1250 (aromatic)
The IR interpretation is the qualitative tool widely useful in
pharmaceutical ,chemical and fertilize industry’s to identify the functional
groups.
Identification of functional group and structure elucidation.
243
Figure 3D.11 The baseline method for determining the
absorbance of an absorption maximum. 244
IR spectrum provides uniqueness to a degree of specificity that is
matched or exceeded by relatively few other analytical methods.
245
Advantages of Infrared Spectroscopy
It is a universal technique. Solids, liquids, gases, semi-solids, powders,
and polymers are all routinely analyzed.
246
Disadvantages of Infrared Spectroscopy
It is impossible to determine
atoms or monatomic ions because they don’t have chemical bonds.
Aqueous solutions
247
248
Reflection
Absorption
Scattering Transimission
249
History of Raman spectroscopy
C.V. Raman and the Raman Effect
In 1928, the Indian physicist C. V. Raman
discovered that the visible wavelength of a small
fraction of the radiation scattered by certain
molecules differs from that of the incident beam.
Violet Scattering
Liquid
Light
Rayleigh Raman
Violet filter scattered scattered
light light
According Raman effect visible wavelength
Green filter
of a small fraction of the radiation scattered
by certain molecules differs from that of the Green
incident beam and furthermore that the
shifts in wavelength depend upon the
chemical structure of the molecules
responsible for the scattering.
251
Principle to Raman spectroscopy
Raman spectroscopy is vibrational spectroscopic technique, which
provide a unique molecular fingerprint for any molecule.
It is a type of spectroscopy which deals with the scattering of light by
the molecules.
When light is scattered by matter, almost all of the light scattered have
the same wavelength with that of the incident light and called elastic
(Rayleigh scattering).
However, a very small percentage of light will be scattered with different
wavelength from incident light and called inelastic scattering (Raman
scattering).
This inelastic scattering of light was observed experimentally by Indian
physicist C. V. Raman and his student in 1928 and then called Raman
scattering (Raman effect).
252
Light scattering
hv0
Rayleigh
hv0
scattering
hvex- ΔE hvex+ΔE
hvex hvv
Energy
V3
V2
V1
ΔE ΔE
V0
Ground state
Rayleigh Scattering Stokes Raman scattering Anti-Stokes Raman scattering
253
Light scattering
hv0
Rayleigh
hv0
scattering
hvex- ΔE hvex+ΔE
hvex hvv
Energy
V3
V2
V1
ΔE ΔE
V0
Ground state
Rayleigh Scattering Stokes Raman scattering Anti-Stokes Raman scattering
254
Stokes shifted Raman bands involve the transitions from lower to higher
energy vibrational levels.
Therefore, stokes bands are more intense than anti-Stokes bands and
hence are measured in conventional Raman spectroscopy.
255
Selection rules
Raman effect arises from polarization (distortion of the electron cloud) of
the scattering molecules that is caused by the electric vector of EMR.
256
Polarizability
257
Homonuclear diatomic molecules such as Cl2 do not absorb IR
radiation, because they have no dipole moment.
Example
1. Cl2 stretching vibration is IR-inactive but Raman active.
1337 cm-1
Raman active
258
Raman scattering information is presented as an intensity-versus-
wavelength shift in cm-1 and can be conveniently compared to its IR
counterparts.
259
Raman instrumentation
Raman spectrophotometers can be dispersive or non-dispersive.
264
Short wavelength sources such as argon ion and krypton ion lasers can
produce significant fluorescence and cause photodecomposition of the
sample.
265
(2) Sample Illumination System
Liquid Samples:
A major advantage of sample handling in Raman spectroscopy
compared with infrared arises because water is a weak Raman
scattered but a strong absorber of infrared radiation. Thus, aqueous
solutions can be studied by Raman spectroscopy but not by infrared.
This advantage is particularly important for biological and inorganic
systems and in studies dealing with water pollution problems.
Solid Samples:
Raman spectra of solid samples are often acquired by filling a small
cavity with the sample after it has been ground to a fine powder.
Polymers can usually be examined directly with no sample
pretreatment.
Gas samples:
Gas are normally contain in glass tubes, 1-2 cm in diameter and
about 1 mm thick. Gases can also be sealed in small capillary tubes.
266
(3) Raman Spectrometers
Raman spectrometers were similar in design and used the same type of
components as the classical ultraviolet/visible dispersing instruments.
Most spectrometers used double-grating systems to minimize the
amount of stray and Rayleigh-scattered radiation reaching the
transducer. Photomultipliers served as transducers.
Now Raman spectrometers being marketed are either Fourier transform
instruments equipped with cooled germanium transducers or
multichannel instruments based upon charge- coupled devices.
Wavelength-Selection Devices and Transducers
A high-quality wavelength-selection device is required in Raman
spectroscopy to separate the relatively weak Raman lines from the
intense Rayleigh-scattered radiation.
Traditional dispersive Raman spectrometers used double- or even triple-
grating monochromators for this purpose.
267
(3) Raman Spectrometers
Raman spectrometers were similar in design and used the same type of
components as the classical ultraviolet/visible dispersing instruments.
Most spectrometers used double-grating systems to minimize the
amount of stray and Rayleigh-scattered radiation reaching the
transducer. Photomultipliers served as transducers.
Now Raman spectrometers being marketed are either Fourier transform
instruments equipped with cooled germanium transducers or
multichannel instruments based upon charge- coupled devices.
Wavelength-Selection Devices and Transducers
A high-quality wavelength-selection device is required in Raman
spectroscopy to separate the relatively weak Raman lines from the
intense Rayleigh-scattered radiation.
Traditional dispersive Raman spectrometers used double- or even triple-
grating monochromators for this purpose.
268
In recent years, holographic interference filters, called notch filters, and
holographic grating have eliminated the need for multiple-grating
monochromators.
The photodiode array was the first array detector to be used. It allows the
simultaneous collection of entire Raman spectra.
269
Figure 3D-3 fiber-optic Raman spectrometer with spectrograph and CCD detector.
The bandpass filter (Bp) is used to isolate a single laser line. The band-rejection
filter (BR) minimizes the Rayleigh-scattered radiation.
270
Fourier Transform Raman Spectrometers
The Fourier transform Raman (FT-Raman) instrument uses a Michelson
interferometer, similar to that used in FTIR spectrometers, and a
continuous-wave (CW).
The vibration should have a dipole The molecule need not possess a
moment change due to that vibration. permanent dipole moment
Water generally cannot be used due Water can be used as a solvent.
to its intense absorption.
Indicates the ionic character in the Indicates the covalent character in the
molecule. molecule.
273
Other types of Raman spectroscopy
Advancements in tunable lasers led to several new Raman
spectroscopic methods in the early 1970s.
1. Resonance Raman spectroscopy
Resonance Raman scattering refers to a phenomenon in which Raman
line intensities are greatly enhanced by excitation with wavelengths that
closely approach that of an electronic absorption band of an analyte.
The frequency coincidence (or resonance) can lead to greatly enhanced
intensity of the Raman scattering, which facilitates the study of
compounds present at low concentrations.
Under this circumstance, the magnitudes of Raman peaks associated
with the most symmetric vibrations are enhanced by a factor of 102 to
106.
As a consequence, resonance Raman spectra have been obtained at
analyte concentrations as low as 10-8 M.
274
True energy state
Virtual state
V3
Vibrational V2
energy states V1
V0
Ground state
Stokes Anti-stokes Resonance Raman Fluorescence
scattering
Raman scattering
Figure 3D-5. (A) Raman spectra of β-carotene solution excited at 488, 514,
532 and 633 nm, respectively. The orange dash curves are the baselines
for Raman spectra. (B) The absorbance curve of β-carotene solution in a 2
mm thick cuvette. The numbers with letter ‘C’ represent Raman peaks from
β-carotene, and the numbers with ‘A’ are peaks from acetone molecules.
Journal of Photochemistry & Photobiology, B: Biology 179 (2018) 18–22
276
The most important application of resonance Raman spectroscopy is to
study biological molecules under physiologically significant conditions;
that is , in the presence of water and at low to moderate concentration.
As an example, the technique has been used to determine the oxidation
state and spin of iron atoms in hemoglobin and cytochrome-c.
Non-resonance Raman scattering occurs when the radiation interacts with
a molecule resulting in polarization of the molecule’s electrons. The
increase in energy from the radiation excites the electrons to an unstable
virtual state; and the radiation is emitted (scattered) at a slightly different
energy than the incident radiation.
Resonance Raman scattering occurs is when the incident radiation is at a
frequency near the frequency of an electronic transition of the molecule of
interest which provides enough energy to excite the electrons to a higher
electronic state.
277
2. Surface-Enhanced Raman Spectroscopy
In surface-enhanced Raman spectroscopy (SERS), Raman spectra are
acquired in the usual way on samples that are adsorbed on the
roughened surface of or colloidal metal nanoparticles (usually silver, gold,
or copper) or on surfaces of pieces of these metals.
As a rule of thumb, to achieve maximum Raman signal enhancement, the
target analyte needs to either directly adsorbed on or placed very close to
the metal nanostructure surface.
In SERS, the Raman signal of the probe molecule is amplified through
excitation of Localized surface plasmon resonance (LSPR) of the
substrate which enhances signal of proximate analyte molecules.
LSPRs are excited when electromagnetic radiation interacts with a
nanoparticle to create coherent oscillations (excitations) of the conduction
electrons.
278
-
-- - - +
+ ++ + +
++ +++
- - - --
279
Au,
Metal nanostructures with dimension
Ag of sub-wavelength, typically <100 nm
The molecule be attached (or at least brought close) to a metallic substrate 281
Application of Raman spectroscopy(RS)
Raman spectroscopy finds applications, in variety of fields and industrial
disciplines.
282
Table 3D-1 Selected Group Frequencies for Common Inorganic Compounds
283
Raman spectroscopy for Organic Species
Raman spectra are similar to infrared spectra in that they have regions
that are useful for functional group detection and fingerprint regions that
permit the identification of specific compounds.
For example, the Raman band of C=C stretch is intense than IR bands,
and its position is sensitive to the nature of substituents as well as to
their geometry.
However, in the IR the intensity is highly variable and may range from
very strong to completely missing.
Thus, Raman studies are likely to yield useful information about the
olefinic functional group that may not be revealed by IR spectra.
284
Figure 3D-6 Raman and IR spectra for 1-Hexene
285
Figure 3D-7 Raman and IR spectra for cis-3-Heptene
286
Figure 3D-8 Raman and IR spectra for tran-3-Heptene
287
2. Material science
Raman spectroscopy is suitable technique for
characterization of nanomaterials as nanocomposites, nano-sized
crystals, polymers and semiconductors.
real-time monitoring of polymerization reactions.
determination of semiconductor impurities in silicone substrates and
diamond-like carbon coatings.
α-Fe2O3 nanoflakes
Carbon-coated/Fe3O4 nanoflakes
Figure 3D-10. Raman spectra of cocaine samples in different forms: (A) yellow freebase paste, (B)
white freebase paste, (C) crack rock, (D) hydrochloride cocaine powder, and (E) freebase cocaine
powder. J. Raman Spectrosc. 2016, 47, 28–38 289
RS also used in security forces and fire brigades for identification of
unknown or hazardous substances, by instance detection of explosives.
Figure 3D-12: Standard Raman (A) and SERS (B) spectra of two different (i, ii) double-
stranded thiolated DNA oligomers J. Am. Chem. Soc. 2008, 130, 5523–5529 291
Figure 3D-13 SERS spectra of three bacterial meningitis pathogens (a) SERS spectra of
N. meningitides, (b) H. influenzae and (c) S. pneumoniae using PCR product of different
concentrations.
292
Advantages of Raman spectroscopy
Quantitative analysis using Raman spectroscopy is more simpler than IR
spectroscopy of due to the following reason.
Raman spectra tend to be less cluttered with bands than IR spectra
and there is less peak overlap in mixtures
suitable for analysis many organic and inorganic materials (solids,
liquids, polymers or vapors).
Less or no- sample preparation needed.
not interference from water.
non-destructive.
highly specific
Raman spectra are acquired quickly within seconds.
samples can be analyzed using glass or a polymer cell.
A very small volume a (< 1 μm in diameter) is required
inorganic materials are easily analyzable with Raman spectroscopy.
293
Disadvantages of Raman spectroscopy
Very low Raman scattering effect
high levels of fluorescence (intrinsic or caused by impurities) overlaying
the Raman bands.
intense laser radiation leads to sample photodecomposition
Expensive instrument and maintenance fee
294
295
Introduction of NMR spectroscopy
Nuclear magnetic resonance, NMR, is a physical phenomenon of
resonance transition between magnetic energy levels, happening when
atomic nuclei are exposed to external magnetic field and applied an EMR
with specific frequency (radio- frequency region of roughly 4 to 900 MHz).
Therefore, a spinning nucleus acts like magnet oriented along the spin
rotation axis. This tiny magnet is often called a nuclear spin.
297
Nuclear Resonance
When a spinning nucleus is placed in a strong external magnetic field, B0,
the spinning nucleus will have a small magnetic moment, µ.
The magnetic fields of the spinning nuclei will tends to align itself either in
the same direction (+½ or α) of the external magnetic field (B0) or
opposite direction (-½ or β) to it.
298
NMR spectroscopy employs an magnetic energy absorption process,
which orients spinning nuclei in a strong external magnetic field.
alignment opposes
the applied magnetic
field and is higher in
energy.
∆ = hv
The precessing nuclei will absorb energy from the radio frequency
source, and if the precessing frequency is the same as the frequency of
the radio frequency beam.
When this occurs, the nucleus and radio frequency beam are said to be
resonance, hence the term “ nuclear magnetic resonance”:
300
NMR spectrometer
305
(a) Number of Signals
Shows number of different chemical environments in which the nuclei in
the sample reside.
The protons that reside in the same magnetic environment are termed
chemically equivalent protons.
306
(b) Splitting of Signals (Spin-Spin Splitting)
The splitting of chemical-shift resonances occurs as the magnetic
moment of a nucleus interacts with the magnetic moments of immediately
adjacent nuclei.
The pattern into which some signals split indicates the number of protons
on adjacent atoms.
307
Table 3E-1: Relative Intensities of First-order Multiplets (I = 1/2)
The spacing between the peaks in hertz is called the coupling constant,
J.
Two signals that arise from nuclei that interact with each other will have
the same value of J.
308
309
(c) Position of Signals (Chemical Shift)
The chemical shift is telling us the strength of the magnetic field that the
nucleus feels.
The shift in the position of the NMR region resulting from the shielding
and deshielding by electrons is called chemical shift.
Chemical shift =
δ= × 10 ppm
In TMS all 12 hydrogen nuclei are chemically equivalent; that is, they are
all exposed to the same shielding and give a single absorption peak.
311
Nucleus feel different magnetic fields depending upon their location and
the neighbour atom they directly attached to.
Since the field experienced by the nucleus defines the energy difference
between the two spin states, the frequency and hence, the chemical shift,
δ /ppm, will change depending on the electron density around the proton.
312
Shielding
Under an applied magnetic field, circulating electrons in the electron
cloud produce a small opposing magnetic field, ultimately decreasing the
effective magnetic field, shifting the signal to the right (or up field).
If there is higher electron density around the nucleus, the electron cloud
“shields” the nucleus from the applied magnetic field, resulting
diamagnetic shielding.
Because the nucleus experiences lower external magnetic field, it needs
a lower frequency to achieve resonance, and it shifts chemical
shift upfield (lower ppm).
B0 B0
e- e- e-
Magnet
Magnet
e- - e- e-
- e
e e-
313
Deshielding
If the electron density around a nucleus decreases, the opposing
magnetic field becomes small and therefore, the nucleus feels more the
external magnetic field B0, and therefore it is said to be deshielded.
B0
Magnet
Magnet
314
Magnet Magnet
RF
RF
e- e-
e-
shielded Deshielded
Low E High E
resonance
315
Figure 3E-7 Abscissa scales for NMR spectra. 316
Effective Magnetic Field, Beff is what the nuclei sense through its
electron filled environment.
Beff= Bapplied – Blocal → for shielding
Bapplied is the magnetic field supplied by the applied external magnetic
force (NMR spectrometer). This number is the same for all the nuclei in
the molecule.
According to this formula, the greater the electron density is around the
nuclei, the larger the Blocal will be, and the more the proton is shielded
from the applied magnetic field.
317
Factors affecting Chemical shift
2. Magnetic anisotropy
3. Hydrogen bonding
318
1. Electronegativity
Electronegative atoms pull electron density from the proton toward itself .
This causes the proton to have less electron density, and the proton is
said to be less shielded (deshielding).
d- d+
Cl C H
d-
The less shielded proton feels more applied magnetic field, and this leads
to a higher E and a higher chemical shift.
Protons that are closer to the electronegative atom are in a less electron
dense environment, which means that their chemical shifts will be larger.
319
Electronegativity dependence of Chemical Shift
Dependence of the Chemical Shift of CH3X on the Element X
320
Substitution Effects on Chemical Shift
321
2. Magnetic Anisotropy Effect
The presence of a nearby π-bond or π-system greatly affects the
chemical shift.
This can either oppose or reinforce the location of proton or the space
occupied by the proton causing either shielding or deshielding effect.
322
Aromatic Protons d = 7-8 ppm Vinyl Protons d = 5-6 ppm
Electronegative
oxygen atom
323
3. Hydrogen bonding
The chemical shift depends on how much hydrogen bonding is taking
place.
Hydrogen bonding lengthens the O-H bond and reduces
the valence electron density around the proton -it is
R deshielded and shifted downfield in the NMR spectrum.
O H H
O H O R
R
Alcohols vary in chemical shift from 0.5 ppm (free OH) to
about 5.0 ppm (lots of H bonding).
With carboxylic acids the O-H absorptions are found between 10 and 12
ppm very far downfield.
325
Proton NMR Shifts
326
Interpretation of 1H NMR spectra
327
Proton NMR spectrum
How many types of H ?
The number of signals shows how many different kinds of protons are
present?
Where on spectrum?
The location of the signals shows how shielded or deshielded the
proton is-Chemical shift
How big?
Indicated by the integration (relative area) of the signal for each
group. The intensity of the signal shows the number of protons of that
type.
What is the connectivity ?
Look at the coupling patterns. This tells you what is next to each
group
328
Spin-spin splitting
Signal splitting: Splitting of an NMR signal into a set of peaks by the
influence of neighboring nonequivalent hydrogens.
Peak: The units into which an NMR signal is split is doublet, triplet,
quartet, multiplet, etc.
329
Example-1: For each of the following compounds, calculate the number of
multiplets for each band and their relative areas:
(a) Cl(CH2)3Cl; (b) CH3CHBrCH3; (c) CH3CH2OCH3.
2 +1 = 3 1 +1 = 2
(a) CH3CH2OCH3
0 +1 = 1
3 +1 = 4
2 +1 = 3
330
=>
331
13C NMR Spectroscopy
332
Sensitivity of Carbon NMR
About 99% of the carbon atoms in a natural sample are the isotope 12C.
The less abundant isotope 13C has an odd number of neutrons, giving it
a magnetic spin ½ of just like a proton.
333
Still 13C NMR has several advantages over proton NMR in terms of its
power to elucidate organic and biochemical structures.
334
Carbon Chemical Shifts
Like 1H NMR, many 13C signals are deshielded by electron-withdrawing
substituents → has a substantial effect for a carbon atom.
Carbon chemical shifts are ~15 to 20 times larger than proton
chemical shifts, because the carbon atom is one atom closer to a
shielding or deshielding group than its attached hydrogen.
For example, consider 1H NMR and 13C NMR spectra of 1,2,2-
trichloropropane.
335
Similarly, an aldehyde proton absorbs at around δ = 9.4 1H NMR in the
spectrum, and the carbonyl carbon atom absorbs around 180 ppm
downfield.
Since only 1% of the carbon atoms in the 13H NMR sample are
magnetic, there is only a small probability of natural two 13C atoms
occurring adjacent to each other is small.
Furthermore, coupling between 13C and 12C does not occur because the
spin quantum number of 12C is zero.
Therefore, carbon–carbon splitting can be ignored. However, carbon–
hydrogen coupling is common.
Most carbon atoms are bonded directly to hydrogen atoms or are
sufficiently close to hydrogen atoms for carbon–hydrogen spin-spin
coupling to be observed. 337
The two most common types of proton decoupling in 13C NMR are
broadband decoupling;
off-resonance decoupling; and
For off-resonance decoupling, the 13C nuclei are split only by the protons
directly bonded to them and n + 1 rule applies.
339
The CCl2 group appears as a singlet,
the CH2Cl group as a triplet, and the
CH3 group as a quartet.
341
Depth 13C NMR
DEPT (Distortionless Enhanced Polarization Transfer) is a more recent
technique.
DEPT gives better sensitivity, and it avoids overlapping multiplets
because all the peaks remain decoupled singlets.
Each 13C nucleus is magnetically coupled to the protons bonded to it.
A DEPT experiment usually includes three spectral scans:
The normal decoupled scan, in which each type of nucleus appears as a
singlet.
The DEPT-90 scan, in which only the CH (methine) carbons bonded to
exactly one proton appear.
The DEPT-135 scan, in which the (methyl) groups and CH (methine)
groups appear normally, and the CH2 groups give negative peaks.
However, carbons that are bonded to no protons do not appear.
342
In general
Carbons with no H’s appear only in the
normal spectrum, but not in either
DEPT spectrum.
343
Table 13-4 Summary of DEPT Spectra
344
Interpreting Carbon NMR Spectra
345
Interpreting 13C NMR spectra uses the same principles as interpreting
1H NMR spectra.
346
Figure 3E-14 Off-resonance-decoupled and broadband-decoupled
spectra of δ-valerolactone of molecular formula C5H8O2. 347
The chemical shift of 70 ppm is about 20 times the chemical shift
of a proton on a carbon bonded to an electronegative element.
The two signals at 19 and 22 ppm are from carbon atoms that are
not directly bonded to any deshielding group, although the carbon
at 22 ppm is probably closer to one of the oxygen atoms.
348
349
Introduction
Mass spectrometry is a technique used for measuring the molecular
weight and determining the molecular formula of an organic compound.
M + e- → [M] +• + 2e-
The radical cation M+• is called the molecular ion or parent ion and the
mass of M+• represents the molecular weight of M.
The ions are sorted according to their mass to charge ratios (m/z), and
and relative abundances
350
Since M is unstable, some ions decompose to form fragments of radicals
and cations that have a lower molecular weight than M+•.
But only the positively charged fragments are detected by the mass
spectrometer.
351
Molecular ion (M): A radical cation formed by removal of a single electron
from a parent molecule in a mass spectrometer.
For our purposes, it does not matter which electron is lost; radical cation
character is delocalized throughout the molecule.
352
Mass Spectrometer
353
Mass spectrometers
Instrument that produces ions and separate them according to their mass
to charge ratios, m/z.
2) Ion sources
3) Mass analyzer
4) detectors
5) Readout
354
The MS instrument consists of three major components:
Ion Source: For producing gaseous ions from the substance being
studied.
Mass Analyzer: For resolving the ions into their characteristics mass
components according to their mass-to-charge ratio.
Detector System: For detecting the ions and recording the relative
abundance of each of the resolved ionic species.
355
A mass spectrometer should always perform the following processes:
Produce ions from the sample in the ionization source.
Detect the ions emerging from the last analyzer and measure their
abundance with the detector that converts the ions into electrical
signals.
356
Separation of Ions:
A beam of electrons causes molecules to ionize and fragment.
Since z is almost always +1, m/z actually measures the mass (m) of the
individual ions.
The tallest peak in the mass spectrum is called the base peak.
The base peak may also be the M peak, but not always.
359
Isotopes
The presence of significant amounts of heavier isotopes leads to small
peaks that have masses that are higher than the parent ion peak.
360
M+1 Peaks
Though most C atoms have atomic mass of 12, ≈1% have a mass of 13.
365
Examples: C6H5CH2NH2 MW =107
H2NCH2CH2NH2 MW = 60
366
Fragmentation patterns of alkanes
The mass spectra of simple hydrocarbons have peaks at m/z values
corresponding to the ions produced by breaking C-C bonds.
369
Fragmentation patterns of alkenes
The molecular ion of alkenes, is usually distinct especially in compounds
containing multiple double bonds.
Shows fragment ions of CnH2n+ and CnH2n-1+ than the CnH2n+1 peak of
alkanes.
370
Figure 3E-7 The radical cation of trans-hex-2-ene cleaves at an
allylic bond to give a resonance-stabilized methallyl cation, m/z = 55.
371
Fragmentation patterns of aromatic
Compounds containing aromatic rings tend to fragment at the carbon
(called a benzylic carbon) next to the aromatic ring. Such a cleavage
forms a resonance-stabilized benzylic cation.
372
Fragmentation patterns of alcohol
One of the most common fragmentation patterns of alcohols is loss of
H2O to give a peak which corresponds to M-18.
373
Figure 3E-8: The mass spectrum of 3-methylbutan-1-ol. The strong peak at m/z 70 is actually the
M–18 peak, corresponding to loss of water. The molecular ion is not visible because it loses water
easily. 374
Evaluation of Unknown Compounds by Mass Spectroscopy
1. Get an overview of the spectrum. Is it simple? Complex? Are there
groups of peaks?
Are there significant peaks due to isotopes (e.g. M+1, M+2, etc.)?
375
3. Evaluate the major fragments
What mass is lost from M+ to give these peaks?
376
377
Introduction to Atomic Spectrometry
Atomic spectrometry is the most widely used method for elemental
analysis.
380
Excitation of electrons in ground state atoms requires an input of
sufficient energy to transfer the electron to one of the excited state.
Excited electrons will only spend a short time in the excited state where
upon relaxation an excited electron will emit a photon and return to the
ground state.
Figure 4A-2 (a) Partial absorption spectrum for sodium vapor. (b)
Electronic transitions responsible for the absorption lines in (a) 383
The techniques of atomic spectroscopy can be divided into three
categories.
Atomic absorption spectrometry
Atomic emission spectrometry
Atomic fluorescence spectrometry
384
Atomization methods
The atomization device must normally perform the complex task of
converting analyte species in solution into gas-phase free atoms and/or
elementary ions
To obtain atomic spectra, the constituents of a sample must be converted
to gaseous atoms.
Flame Furnace
plasma 385
Atomization – the process by which a sample is converted to an atomic
vapor.
Flame atomization
387
Several processes occur during atomization including:
a) nebulization – solution sample, get into fine droplets by spraying via thin
nozzle or passing over vibrating crystal.
389
Introduction of solution samples
There are several sample introduction system in atomic spectroscopy.
390
Pneumatic Nebulizers
Samples in solution are usually easily introduced into the atomizer by a
simple nebulization, aspiration, process.
Nebulization converts the solution into an aerosol of very fine droplets
using a jet of compressed gas.
The flow of gas carries the aerosol droplets to the atomization chamber
or region.
Figure 4A-4 Types of pneumatic nebulizers: (a) concentric tube, (b) cross-flow, (c) fritted disk, (d)
Babington. 391
Introduction of Solid Samples
Several techniques have been used for the direct introduction of solids into
atomizers.
(a) Direct sample Insertion
direct manual insertion of the solid into the atomization device
(b) Electrothermal vaporization
vaporization of the sample and transfer of the vapor into the atomization
region,
(c) Arc or spark ablation
arc, spark, or laser ablation of the solid to produce a vapor that is then
swept into the atomizer,
(d) Glow-Discharge Technique
slurry nebulization in which the finely divided solid sample is carried
into the atomizer as an aerosol consisting of a suspension of the solid
in a liquid medium, and sputtering in a glow-discharge device 392
Atomization techniques
1. Flame atomization
In a flame atomizer, a solution of the sample is nebulized by a flow
of gaseous oxidant, mixed with a gaseous fuel, and carried into a
flame where atomization occurs.
394
Flame Structure
Interzonal region is the hottest part of the flame and best for atomic
absorption.
Oxidation of the atoms occurs in the secondary combustion zone where
the atoms will form molecular oxides and are dispersed into the
surroundings.
399
It is used for atomic absorption and atomic fluorescence measurements
but have not been applied for direct production of emission spectra.
They are also used for vaporizing samples in inductively coupled plasma
emission spectroscopy.
401
3. Inductively Coupled Plasmas
involves use of high temperature plasma for sample
atomization/excitation.
403
Effects of temperature on atomic spectra
Temperature determines the
degree to which a sample breaks down into atoms and
the extent to which a given atom is found in its ground, excited, or
ionized states.
405
Example: The lowest excited state of a sodium atom lies 3.371 x 10-19
J/atom above the ground state. The degeneracy of the excited state is 2,
whereas that of the ground state is 1. What is the fraction of Na in the
excited state in an acetylene-air flame at 2600 K is
That is, less than 0.02% of the atoms are in the excited state.
If the temperature were 2610 K, fraction of atoms in the excited state is
The fraction of atoms in the excited state is still less than 0.02%, but that
fraction has increased by 100(1.74 - 1.67)/1.67 = 4%
406
Absorption arises from ground-state atoms, but emission arises from
excited-state atoms.
408
AAS is a spectroanalytical procedure for qualitative & quantitative
analysis of chemical elements based on absorption of optical radiation
(light) by free atoms in the gaseous state.
409
AAS makes use of absorption spectrometry to assess the conc. of an
analyte sample in solution.
412
Figure 4A-10: Block Diagram of an Atomic-Absorption Spectrometer 413
Flame Atomic Absorption spectrometry
atomizer
414
A. Light source
The light source, in FAAS can be either
(a) The Hollow Cathode Lamp (HCL) or
(b) Electrodeless Discharge Lamp (EDL).
(a) The Hollow Cathode Lamp (HCL)
It consists of a
Cathode made of analyte metal or serves as a support for a coating
of analyte metal of interest
The generated electrons collide with the metal atoms, raising them to the
ES.
420
D. Detectors
The detector is normally a photomultiplier tube, which produces an
electrical current dependent on the light intensity.
421
Flame atomic absorption is very convenient & widespread, it has an
acceptable level of accuracy for most analytes.
423
A) The cold vapor mercury technique
The cold vapor method is only used for
the determination of mercury.
10 20 30 40 50 60 70 80 90 100
Concentration ( mg/L )
428
Interferences in AAS
Interference is any effect that changes the signal while analyte
concentration remains unchanged.
429
A. Spectral interferences
It refers to the overlap of analyte signal with signals due to other
elements or molecules in the sample or with signals due to the flame or
furnace that the monochromator cannot resolve them.
Solution
The best solution is choosing altenative wavelength for analysis.
Example: V line is 308.2 nm and Al line is at 308.2 nm.
Choose a different Al line at 309.2 nm.
Solution
Use a higher flame temperature (nitrous oxide/acetylene)
Use a releasing agent
Use protective chelation
431
Determination of calcium in the presence of sulfate or phosphate (e.g. in
natural waters)
3Ca2+ + 2PO43- → Ca3(PO4)2 (stable compound)
Releasing agent
Add 1000 ppm of LaCl3
2LaCl3 + Ca3(PO4)2 → 3CaCl2 + 2La(PO4)
CaCl2 readily dissociates
Protective chelation
Ca3(PO4)2 + 3EDTA → 3Ca(EDTA) + 2PO43-
Ca(EDTA) dissociates readily.
432
Ionization interferences
Ionization interference can be also a problem in the analysis of alkali
metals at relatively low temperature because they have low ionization
potentials.
M(g) M+(g) + e-
Ions have energy levels different from those of neutral atoms, so the
desired signal is decreased.
433
Solution
Add an ionisation suppressor which are easily ionised element such as
Cs.
436
Atomic emission spectroscopy measures radiation emitted when valence
electron in a higher energy atomic orbital returns to a lower energy.
437
Instrumentation
The instrumentation of atomic emission spectroscopy is the same as that
of atomic absorption, but without the presence of a radiation source.
In atomic emission the sample is atomized and the analyte atoms are
excited to higher energy levels all in the atomizer.
The flame (1700 – 3150 oC) is most useful for elements with relatively
low excitation energies like sodium, potassium and calcium .
The plasma (6000 – 8000 oC) has a very high temperature and is useful
for elements of high excitation energies.
439
Emission spectrometry atomization and sources
Plasma sources
A plasma is a hot, partially ionized gas that contains an abundant
concentration of cations and electrons.
441
Inductively Coupled Plasma Source
The ICP torch consists of three concentric
quartz tubes, surrounded at the top by a
radio-frequency induction coil.
The resulting ions and the electrons are subjected to the fluctuating
magnetic field produced by the induction coil and flow in the closed
annular paths within the coils.
These electrons rapidly acquire enough energy from the oscillatory field
generated by the induction coil and sustain a high degree of ionisation.
The spacing between coil and torch, the distance between individual
copper coils and the concentricity of both coil and torch are important
parameters in the formation of stable plasma.
443
Figure 3B-2 Schematic diagram of an inductively coupled plasma torch.
444
Instrumentation for ICP-AES
445
Quantitative Analysis using ICP-AES
Example: 10-g portion of the salt substitute is dissolved in 10 mL of 3 M HCl and
100 mL of distilled water and diluted to a 250-mL. A series of standard additions is
prepared by placing 25-mL portions of the diluted sample into separate 50-mL
volumetric flasks, spiking each with a known amount of an approximately 10 mg/L
standard solution of Na+, and diluting to volume. The emission intensity is
measured for each of the standard addition. The results of a flame atomic emission
analysis of the standards is
447
Advantages of ICP-AES
The wide applicability of ICP-AES in the quantitative analysis of various
materials can be attributed to the following advantages
Rapid and simultaneous multi-element analysis.
Lack of chemical interferences.
High temperature of the excitation source : 5000 to 10000K.
Low sample requirements.
Low detection limits: 1 to 100 ng/g or µg/l (part per billion).
Good accuracy and precision: relative standard deviation of about
1 per cent.
Applicability to elements that are difficult to be determined by AAS:
B, C, Ce, La, Nb, Pr, S, P, Ti, Ta, V and Zr can also be measured. .
448
Applications of ICP-AES
The ICP-AES technique is versatile tool in the hands of analytical chemists.
Agricultural science:
Analysis of agricultural products and foods besides soil analysis.
Health sciences
Determination of Al in blood, Cu in brain tissue, Se in liver, Na in
breast milk.
Direct determination of Ca, Fe, Cu, Mg, Na and K in serum samples.
Geological sciences
Presence of lanthanides and other elements in rock samples.
Forensic Sciences
Crime scene soil analysis.
449
Forensic Sciences
Crime scene soil analysis.
Metallurgy
Analysis of trace elements in stainless steel.
Environmental science
Waste water analysis, determination of pollutant metals in variety of
matrices.
Industry
Presence of metals like Cu, Fe, Ni, and Si in lubricating oils or gasoline
at tracer concentration.
Traces of metals like Ca, Cu, Fe, Mn, Mg, P, K and Zn in beer or wine;
determination of trace elements in polymers, evaluation of catalysts,
and so on.
450