LCR Meters 4310 4320 4350 43100
• Measurements from 20Hz to 1MHz (43100)
• 0.1% basic accuracy
• <20ms measurement time
• Parameters Z θ Rac C D L Q B G X Y Rdc
• Display four parameters and two tests at once
• Drive level from 10mV to 2Vrms
• RS232, GPIB, USB & LAN Interfaces
• +2V internal / ±40Vdc external bias
• Small size & light weight
The 4300 series is the entry level range of Wayne Kerr LCR Meters. They provide a wide range of features and offer high
performance and very competitive prices.
Frequency Basic Accuracy
Model Frequency Range AC: 0.1%
4310 20Hz to 100kHz DC: 0.2%
Varies with speed, frequency, drive level and impedance of
4320 20Hz to 200kHz
the device under test.
4350 20Hz to 500kHz
Measurement Parameters
43100 20Hz to 1MHz Impedance (Z) Phase Angle (θ)
Frequency Step Size Capacitance (C) Dissipation Factor / tan δ (D)
Frequency Step Size Inductance (L) Quality Factor (Q)
20Hz to 1kHz 5Hz AC Resistance (R) Reactance (X)
1kHz to 10kHz 50Hz DC Resistance (Rdc) Admittance (Y)
10kHz to 100kHz 500Hz Susceptance (B) Conductance (G)
100kHz to 1MHz 5kHz
Series/Parallel Equivalent Circuit
AC Drive Level Any combination of AC parameters
Voltage (into open circuit): 10mV to 2Vrms Range of Readings
Voltage step size: 10mV (200 steps) Parameter Display Measurement Range
Signal source impedance: 100 Ω
Z, R, X 10.0000µΩ to >100.000GΩ
DC Drive Level
Y, B, G 1.00000pS to >10.0000kS
1V or 2V
L 100.000pH to >100.000MH
DC Bias Voltage
C 10.0000fF to >1.00000F
Internal: 2V
D, Q 0.00001 to 99999.9
External: ±40V
o o
A -180.000 to 180.000
Rdc 0.1000mΩ to >10.000GΩ
Measurement Speeds Safety
Four selectable measurement speeds for all functions. Over Complies with the requirements of EN61010-1.
50 measurements per second, depending on frequency and
set up conditions. EMC
Complies with EN61326 for emissions and immunity.
Test Mode
One test or two tests can be triggered automatically. AC Power Input
Measurement parameters, frequency, and drive level can 90VAC to 264VAC (Autoranging); 47Hz to 63Hz
be changed between Test 1 and Test 2.
Display
Test Limits 3.8” ¼ VGA (320 x 240) Black & White
Limits can be set for all measurements in one or two test
mode. Results of limit checks are displayed on screen. Remote Control
PASS/FAIL signal is available on Scaleizer port on rear RS232, GPIB, USB and LAN
panel.
Mechanical
Save and Recall Height 104mm (4.1″) Width 322mm (12.7″)
Up to 20 measurement set ups can be saved with user- Depth 285mm (11.1″) Weight 3kg (6.6lb)
defined names and later recalled.
Temperature Range
External Control Operating: 0°C to 40°C
GPIB, USB, LAN and RS232. Full Accuracy: 23°C ±5°C
Binning (Option) Storage: -40°C to +70°C
User programmable PASS and FAIL bins indicated by
signals available on the rear panel. /B1 provides non-
isolated signals. /B2 provides isolated signals.
Scaleizer (Option)
Provides relay controlled signals available on the rear panel
indicating PASS/FAIL decisions based on user-defined
limits. 2 versions available (/S1 and /S2).
Order Codes Options Recommended Accessories
Model Part No. Description Code Description Part No.
4310 (100kHz) 1J4310R Scaleizer (S1 type) /S1 4-terminal component
1EV1006
4320 (200kHz) 1J4320R Scaleizer (S2 type) /S2 fixture
4350 (500kHz) 1J4350R Bin handler (non-isolated) /B1
SMD tweezers 1EVA40120
43100 (1MHz) 1J43100R Bin handler (isolated 24V) /B2
All units are supplied with User Manual, Kelvin clips and AC power cable as standard.
Only one option may be fitted. To specify an option add the option code to the instrument part number.
Accessories for the 4300 series are being introduced continuously. Consult our website or your local distributor for all accessories available.
CalPlus GmbH CalPlus GmbH
Heerstr. 32 • 14052 Berlin Niederlassung ScopeShop Hamburg
Tel.: 030/ 214 98 2-0 • Fax: 030/ 214 98 2-50 Wendenstr. 379 • 20537 Hamburg
[email protected] • www.calplus.de Tel.: 040/ 30 39 595-0 • Fax: 040/ 30 39 595-50
[email protected] • www.tektronix.calplus.de
October 2011 Issue F
Wayne Kerr’s policy is one of continuous development and consequently the product may vary in detail from the description and specification in this publication.