High Flow System Setup Guide

A B

Setup Start

Humidification Press 5 sec.


chamber
preparation

Green
light: Correct
disinfection has
Chamber been completed
fitting since last
use

Last
Click Disinfection:
#16
Orange:
Disinfection
H2O NOT completed

Water bag
connection

Rolling circle:
Warm-up time
32
o
c 15 21
L/min
%O 2

Breathing tube
connection

New target
settings

Access / leave
Junior Mode:
Press 5 sec.
34
c
o
15 21
L/min
%O 2

Target settings
FLOW and
% O2 : See
Scan the relevant QR code above
to download the AIRVO 2 simulator
C
app to your smartphone.

This content does not replace the User Instructions. For a full product setup description, cautions, hazards and warnings, please refer to the User Instructions provided with the product.

www.fphcare.com
High Flow System Setup Guide
C D

Settings Patient interfaces

1 Target Flow Setting 2 % O2 Setting

Access the target Adjust the O2 flow meter


flow setting. to change % O2 value.

37 30 Nasal Tracheostomy
c L/min 35
o

OPT942 OJR416 OPT970


%O 2
OPT944 OJR418
OPT946

Available target settings


Move past the Note: This will NOT affect Temp °C Flow L/min
temperature setting menu. the delivered patient flow. 31 34 37 2 5 10 15 20 25 – 55 60

OJR416 2 20

OJR418 2 25
37
c o OPT942 S 10 50
37 60
c L/min 90
o
OPT944 M 10 60
%O 2

OPT946 L 10 60
Adjust target flow:
OPT970 10 60

If the setting menu is Turn Off


locked ( ), unlock by Read the
pressing 3 sec: measured
% O2 value. Press 5 sec.

30L/min
Cleaning and disinfection

3When the A B

Parameter setting target flow setting Clean the Surface


adjustment outlet elbow cleaning
is adjusted: 1

New target flow setting


adjustment (item 1) will
change the delivered % O2.

60L/min
If required, adjust the
% O2 (item 2).
3

C
Exit the menu settings. 1 2 3 4
Automatic
disinfection
cycle 55:00 00:00