Upload 00093075 1525633476936
Upload 00093075 1525633476936
Electrolyte Analyzer
Instructions for use
Copyright © 2007 Alfa Wassermann, Diagnostic Technologies, LLC, all rights reserved
Copyright, 2007, Alfa Wassermann, Diagnostic Technologies, LLC. All rights reserved. Unless otherwise noted, no part of this
publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any language in any
form without the written permission of Alfa Wassermann, Diagnostic Technologies, LLC.
ISETROL is a trademarks of Roche Diagnostics GmbH.
STARLYTE is a trademark of Alfa Wassermann.
Rev. 1.0, Oct 2007
– Important Information! – Follow Instructions Carefully! –
These Instructions for Use contain vital warning and safety information.
This instrument is intended to be used only for the specialized purpose described in the instructions. The
most important prerequisites for use, operation, and safety are explained to ensure smooth operation. No
warranty or liability claims will be covered if the instrument is used in ways other than those described or if
the necessary prerequisites and safety measures are not observed.
The instrument may be operated only by persons whose qualifications enable them to comply with the safety
measures that are necessary during operation of the instrument.
Adjustments and maintenance performed with removed covers and connected power may be attempted only
by a qualified technician who is aware of the associated dangers.
Instrument repairs are to be performed only by the manufacturer or qualified service personnel.
Only accessories and supplies either delivered by or approved by Alfa Wassermann are to be used with
the instrument. These items are manufactured especially for use with this instrument and meet the highest
quality requirements.
Operation of the instrument with solutions whose composition is not consistent with that of the original
solutions can negatively affect, above all, the long term measurement accuracy. Deviations in the composition
of the solutions can also decrease the service life of the electrodes.
The quality control requirements must be completed at least once daily for safety reasons. Because accurate
measurement results depend not only on the proper functioning of the instrument, but also on a number of
other factors (such as preanalytics), the results produced by the instrument should be examined by a trained
expert before subsequent decisions are reached that are based on the measurement values.
Explanation:
The instrument has been constructed and tested according to the following European Stan-
dards:
• IEC/EN 61010-1:2001
• IEC/EN 61010-2-101:2002
• IEC/EN 61010-2-081:2002 + A1:2003
It was delivered from the factory in flawless condition with regards to safety features. In order
to preserve this condition and ensure safe operation, the user must observe the notices and
warnings that are contained in these Instructions for Use.
• This instrument is classified under the protection class I according to IEC 1010-1 /
EN 61010-1.
• If an object or liquid enters the internal areas of the instrument, remove the instrument
from its power supply and allow an expert to check it thoroughly before using it again.
CAUTION:
• The power cord may be plugged into a grounded socket only. When using an extension
cord, make sure it is properly grounded.
• Any rupture of the ground lead inside or outside the instrument or a loose ground connec-
tion may result in hazardous operating conditions. Intentional disconnection of the
grounding is not permitted.
1 Introduction
1.1 General notes ................................................................................................................... 1-2
1.2 Measurement and calibration procedures ................................................................ 1-5
1.3 Measurement evaluation .............................................................................................. 1-5
1.4 Important safety instructions ....................................................................................... 1-6
1.5 System description ......................................................................................................... 1-7
1.6 Installation ...................................................................................................................... 1-12
1.7 Shutdown ........................................................................................................................ 1-25
1.8 Standby mode ................................................................................................................ 1-29
2 Specifications
2.1 Performance parameters .............................................................................................. 2-1
2.2 Linearity ............................................................................................................................ 2-5
2.3 Interferences .................................................................................................................... 2-6
2.4 Limitations ........................................................................................................................ 2-7
2.5 Relationship of ionized calcium to total calcium ..................................................... 2-7
2.6 Bibliography ..................................................................................................................... 2-8
2.7 Sample throughput ......................................................................................................... 2-9
2.8 Sample volumes .............................................................................................................. 2-9
2.9 Sample types .................................................................................................................... 2-9
2.10 Calibrations ...................................................................................................................... 2-9
2.11 Environmental parameters .......................................................................................... 2-10
2.12 Product data ................................................................................................................... 2-13
2.13 Printer .............................................................................................................................. 2-14
2.14 Display ............................................................................................................................. 2-14
3 Measurement
3.1 Preanalytics ...................................................................................................................... 3-1
3.2 Limitations of clinical analysis ..................................................................................... 3-3
3.3 Measuring procedure .................................................................................................... 3-3
3.4 Normal ranges ................................................................................................................. 3-7
3.5 Correlation factors .......................................................................................................... 3-8
3.6 Additional settings .......................................................................................................... 3-9
4 Quality control
4.1 General QC concept ........................................................................................................4-1
4.2 Material setup ..................................................................................................................4-2
4.3 Performing a QC measurement ....................................................................................4-3
4.4 Printing a QC report ........................................................................................................4-5
5 Maintenance
5.1 Decontamination .............................................................................................................5-1
5.2 Daily maintenance ...........................................................................................................5-3
5.3 Weekly maintenance .......................................................................................................5-5
5.4 Monthly maintenance .....................................................................................................5-6
5.5 Semi annual maintenance .............................................................................................5-9
5.6 Annual maintenance .................................................................................................... 5-10
5.7 Unscheduled maintenance ......................................................................................... 5-12
6 Troubleshooting
6.1 Error messages .................................................................................................................6-1
6.2 Service functions .............................................................................................................6-7
6.3 Service codes ................................................................................................................ 6-11
6.4 Deleting data ................................................................................................................. 6-13
7 Theoretical foundations
7.1 Clinical significance ........................................................................................................7-1
7.2 Principles of Operation ...................................................................................................7-7
7.3 Electrode specifications .............................................................................................. 7-10
8 Appendix
8.1 Description of reports .....................................................................................................8-1
8.2 Specifications of the solutions .....................................................................................8-3
8.3 Maintenance schedule ...................................................................................................8-4
9 Index
1 Introduction
The STARLYTE V analyzer is a powerful tool designed to quickly, accurately and efficiently
conduct basic electrolyte testing in the convenience of the laboratory.
This manual will help guide through setting up the analyzer and analyzing samples. As the
user become familiar with the operation of the unit, the manual may be used as a reference
for day-to-day routines and as a guide for maintenance and troubleshooting.
Fig. 1-1
1.1.1 Symbols
Lot number
Electrodes: This date indicates the limit of the maximum storage time of an
electrode. The electrode must be installed in the instrument no later than
the imprinted date.
If the installation takes place on the imprinted date, it still falls within the
specifications. The calculation of the “Install before” date is based on the
production date of the elctrode.
Storage note
The conditions necessary to preserve the product's shelf life before opening.
Store upright
Sections marked with this symbol (see Instructions for Use) contain information
that must be observed to avoid potential injuries (to patients, users and third par-
ties).
Risk of infection!
All sections / passages that are marked with this symbol describe procedures and/or
indicate conditions or dangers that could damage or lead to a malfunction in the
STARLYTE V analyzer, and therefore should never be attempted.
TIP: All sections / text locations marked with "TIP" describe safe procedures that are intend-
ed to provide the user with additional help.
• Always handle blood samples and collection devices with appropriate care.
• Use approved protective gloves to avoid direct contact with sample.
• Aseptic procedures are required when cleaning the sampling probe to avoid contamina-
tion.
• Dispose of Fluid pack according to local regulations.
Dispose of used Fluid pack, electrodes and the instrument according to applicable
laboratory regulations (biologically contaminated–hazardous waste!)
Display Keypad
Front cover
Sample door
Fluid pack
Fig. 1-2
Valves
Fig. 1-3
Display
The analyzer communicates to the user through a dot matrix display. This two line, alpha-
numeric read-out allows up to 16 characters per line, displaying the activities of
the analyzer, sample results and other programmed information.
Na K Cl READY
Fig. 1-4
Keypad
The communication with the analyzer is performed through a keypad with YES and NO keys.
With these keys, all analyzer functions, including: sample measurement, data input, pro-
gramming and quality control testing can be performed.
Fig. 1-5
Printer
The thermal printer uses heat-sensitive paper to output information in 16 columns. The
analyzer will print measured values, calibration values, electrode voltages, and amount of
liquid remaining in the Fluid pack as well as cleaning and maintenance information. The
unit is configured to allow convenient storage of a second roll of paper in the paper tray.
Fig. 1-6
Measuring chamber
The measuring chamber consists of the movable left locking device that holds the electrodes
in place, the electrodes, the right electrode holder with sample sensor connector, and the
measuring chamber base.
Fig. 1-7
Peristaltic pump
A peristaltic pump transports all sample and operating fluids inside the instrument.
Fig. 1-8
The sample probe mechanism is located behind the sample door at the front of the unit.
Fig. 1-9
Valves
Fig. 1-10
Fluid pack
The self-contained Fluid pack is constructed to ensure that waste cannot spill out of the
package.
Fig. 1-11
Rear panel
The rear panel of the unit contains a serial number plate, as well as the power switch/power
receptacle module and an RS232 interface port.
Fig. 1-12
Serial port
The analyzer is equipped with a serial RS232 interface and allows data to be exchanged with
commercially available computer systems. Measurement data is transferred after each mea-
surement and calibration data is transferred after each calibration. The interface is always
active, and data is sent independently of the printer setup.
A higher leakage current can be expected when using the serial interface. This must
be checked by suitably qualified personnel, depending on the local regulations.
1.6 Installation
1.6.1 Location
Location is important for optimal operation of your analyzer. Before you begin setup,
choose a site that is convenient for your sampling needs and meets the following physical
requirements of the unit:
• Ambient temperature + 15°C to + 32°C
• Avoid exposure to direct sunlight, vibration and strong electromagnetic fields (electric
motors, transformers, X-ray equipment, cellular phones, etc.).
• Use a stable and level work surface.
• Maximum relative humidity of 85 % (15 % - 85 %)
• Ample room to allow air to circulate freely around the unit, see Fig. 1-13.
• Avoid exposure to explosive gases or vapors.
• Check for correct voltage: 100 to 240 VAC, 50/60 Hz.
340 mm
350 mm
330 mm
Fig. 1-13
After setting up the STARLYTE V analyzer in a location that meets the above requirements,
perform the following steps to ensure the instrument is ready for operation:
• Refer to the packing slip to check for the completeness of the shipment.
• If the shipment is incomplete, please inform Alfa Wassermann immediately.
If the instrument was damaged during shipment, immediately inform the company that
made the delivery. Retain all packaging materials and products as this may be needed as evi-
dence in the event of a damage claim.
Do not remove the analyzer from the shipping carton by pulling upward on the poly-
styrene (styrofoam) packing materials. These packaging materials do not provide
strength to support the analyzer.
Accessories
The following parts are delivered as standard equipment with the STARLYTE V analyzer:
• Printer paper
• Quad Ring, 1.78x1.02 mm
• Fuse set
• Sample probe stylet
• Syringe, 12 mL, with tapered tip
• Quick Reference Guide
You also need a supply of lint-free tissues and disposable sample cups, which should be kept
in a location convenient to the analyzer.
1.6.2 Setting up
Prior to beginning the actual installation, it is recommended to completely read through
this chapter to develop an understanding of the procedures that are required.
Begin by placing the analyzer on a secure table top that allows plenty of working space and
is convenient to a power connection.
Open the analyzer main door. Locate and carefully remove the five red relief clamps. Save
the clamps for reuse to prevent damaging the tubes, if the analyzer is later shut down for
any reason.
TIP: Save the relief clamps for reuse to prevent damaging the tubes, if the analyzer is later
shut down (see section "Shutdown").
• Slip the two pump windings around the analyzer pump rollers, making sure not to over-
stretch the tubing.
Fig. 1-16
• Unscrew the red transport housing from the reference electrode and check that the
o-ring on the electrode is properly seated. Rinse, dry and save the transport housing for
storage of the reference electrode in the event the analyzer is turned off or taken out of
service for any reason.
Fig. 1-17
• Carefully screw the reference electrode into the reference electrode housing and place it
with the other electrodes.
reference electrode
housing reference electrode
Fig. 1-18
• Slide the measurement chamber forward until it locks in the front position.
• Unclamp the left electrode holder by moving the clamp forward.
• Locate the sample sensor cable and ensure that it is securely inserted in the receptacle
above the measurement chamber. Check that an o-ring is present in the right electrode
holder
Fig. 1-21
Now install the electrodes in the measurement chamber, beginning on the right and working
to the left (the reference electrode will be installed last).
TIP: The electrode on the right may be one of the following: chloride, ionized calcium, lith-
ium or a dummy electrode, if no third test is needed. If sodium and lithium are selected,
the potassium electrode is replaced by a dummy electrode.
• Check to make sure that the lettering on the electrode corresponds with the proper elec-
trode station number on the tray. Also, note that all electrodes have a lip on the bottom
that rests on the flat edge of the measurement chamber to aid in proper positioning.
Fig. 1-22
• After all electrodes have been installed, close the clamp on the left electrode holder by
lifting it upward until it locks in the back position, and ensure that the electrodes are
properly seated.
Fig. 1-23
Fig. 1-24
• Plug the tubing connector of the reference housing assembly in the receptacle below the
left side of the measurement chamber.
Fig. 1-25
Na + Na + Na + Na + Na + Na +
K+ K+ K+ Dummy K+ Dummy
Cl - / Ca ++ / Li + Dummy Cl - Cl - Ca 2+ Ca 2+
Na + Na + Na + (Na + ) Na +
K+ K+ Dummy Dummy Ca 2+
Cl - / Ca ++ / Li + Li + Li + Li + Li +
Fig. 1-26 Parameter configurations
1. Selecting language
Before the analyzer is powered on, the language setting must be selected. The factory-set
language is English.
• Remove the paper tray and use a pen to set the switch to the desired language position
as indicated on the label.
TIP: If you decide to change the language after the unit is powered on, you must cycle the
power to activate the new language.
2. Start-up
• Locate the power switch on the back of the unit and make sure that it is in the OFF (O)
position.
• Plug the power cord into the power receptacle module on the back of the unit, then plug
the cord into a grounded electrical outlet (100 - 240 V, 50/60 Hz).
Fig. 1-27
• Push the power switch to the ON (I) position. The unit will automatically begin to
operate.
Fig. 1-28
Now that the STARLYTE V analyzer is functioning, begin using the keypad interface to com-
municate with the instrument. Use the NO key to make changes, the YES key to accept the
displayed values or information.
TIP: After start-up, date and time should be set and the printer paper inserted.
The analyzer will display a default date and time and will allow the operator to input the
correct date and time:
Date: 01-JAN-02
Time: 00:00
• Record the installation date of the Fluid pack on the label on the Fluid pack.
• Remove the protective strip and slide the Fluid pack into position on the left side of the
analyzer. Save the protective strip for the case of a shutdown or disposal.
Once the protective strip is removed, be sure to keep the Fluid pack upright to avoid
spillage. Save the protective strip to use to close the nipples on the Fluid pack prior
to disposing.
TIP: The printer paper is heat sensitive on one side only. Please make sure that you insert
the paper roll correctly.
• Place a the paper roll in the paper tray and thread it into the feeder slot.
Fig. 1-31
• Press the paper feed button to completely feed the paper through the printer.
Fig. 1-32
TIP: By pressing the paper advance button once, the paper will automatically advance 10
lines.
Do not pull the paper out of the printer to avoid damage to the printer.
Prior to performing the first calibration or running the first sample, the STARLYTE V ana-
lyzer needs to undergo a simple cleaning and conditioning procedure that helps ensure that
the analyzer will perform properly. This procedure is called daily maintenance, because it
must be performed each day the analyzer is used to conduct sampling.
The process involves cleaning and conditioning the sample path and electrodes, which pre-
pares the STARLYTE V analyzer for calibration. The bottles containing Cleaning Solution
and Electrolyte Conditioning Solution should be ready, along with a package of lint-free tis-
sues that will be used to dry the probe.
Check expiration date on the bottles of the Cleaning Solution and the Electrolyte
Conditioning Solution.
In some cases, when the analyzer prompts you for an action and you do not respond
within a set period of time, an alarm will sound and the unit will discontinue its cur-
rent operation.
Fig. 1-33
• Hold the Cleaning Solution under the probe, until [Wipe Probe/Close Sample
Door] is displayed, and use a lint-free tissue to remove the cleaning solution from the
probe. Close the sample door.
Fig. 1-34
• While the analyzer displays [Thank You!] and a brief countdown, pour a small
amount of conditioning solution into a clean container.
• Answer the prompt [Perform Daily Conditioning?] by pressing YES .
• At the prompt [Open Sample Door Introduce Sample] , lift the sample door and
the pump will begin to aspirate.
Fig. 1-35
• Hold the Electrolyte Conditioning Solution under the probe, until [Wipe Probe/
Close Sample Door] , use a lint-free tissue to remove the conditioning solution from
the probe. Close the sample door.
Fig. 1-36
• After the analyzer has displayed [Thank You!] and a brief countdown, the prompt
[Remain in Daily Maintenance?] can be answered with NO, and an automatic cal-
ibration will be started.
It is very important that the main door is closed during calibration, since it provides
shielding from sources of electromagnetic interference.
After finishing the calibration, the analyzer returns to [READY] and is now ready for QC.
After completion of the daily maintenance procedure, the analyzer starts an automatic cal-
ibration for sodium and potassium.
To select a different parameter configuration, interrupt the calibration by pressing NO .
• Continue pressing NO, until [OPERATOR FUNCTIONS?] is displayed. Press YES .
• Press NO until the prompt [Select Parameter Configuration?] is displayed, see
Fig. 1-26. Press YES and the current configuration will be displayed. The default config-
uration is the following:
Sel. Parameter:
[Na][K][ ] ok?
• Press NO until the desired configuration is displayed. Press YES to accept this selection.
TIP: If [(Na)][ ][Li] is selected, only lithium is reported. Install the sodium electrode
and calibrate for lithium.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each replace-
ment of the Fluid pack, after startup of the instrument as well as after monthly, semi
annual and annual maintenance steps.
Additionally, at least once daily one QC measurement has to be performed in alter-
nating levels (low, normal, high) (e.g., day 1 - level 1, day 2 - level 2, day 3 - level 3,
day 4 - level 1, etc.). When required by local regulations, QC measurements must be
performed more often.
A quality control program for electrolytes includes the analysis of sample materials
with known ranges of expected values and the comparison of these values with ana-
lyzer results. For further information, see chapter 4: "Quality control".
1.7 Shutdown
A complete shutdown of the analyzer may be indicated to prepare the analyzer for shipping
or in case the analyzer is not being used for an extended period of time. For this procedure,
you will need a special shutdown kit (not supplied with the analyzer).
Never attempt to turn the power off for an extended period of time without perform-
ing a complete shutdown of the analyzer.
In case the analyzer is not being used for several days only, it is not recommended to
perform a complete shutdown, but to put the analyzer in Standby Mode.
To perform the shutdown, you will need the following items:
• One shutdown plug (supplied with shutdown kit).
• Five solenoid relief clamps (supplied with analyzer and shutdown kit).
• Two reference electrode transport housings (supplied with shutdown kit).
• Two containers, one filled with at least 100 mL of water, the other one empty.
• Protective strip for Fluid pack (supplied with shutdown kit).
Fig. 1-37
• Press YES .
The analyzer will take approximately one minute to flush all lines with water. Upon com-
pletion, you will receive the prompt [Remove Water and Press YES] .
• Remove the line with the blue mark from the distilled water and place it on a cloth.
• Press YES . During this cycle, all lines are purged of distilled water.
• Upon completion, [All Electrodes & Plug Removed?] will be displayed.
Fig. 1-40
Fig. 1-41
Fig. 1-42
• Press YES .
Never insert the solenoid relief clamps with the Fluid pack in place.
Fig. 1-43
• Press YES . The prompt [Do You Wish to DELETE All Data?] will appear.
• By pressing YES , all QC values and statistics will be deleted and the sample number is
reset to 0. QC and normal ranges as well as correlation factors, printer settings, date/
time and Fluid pack volume are reset to default. Also, all service codes are deactivated.
A deletion of all data is recommended in case the analyzer is used by different labora-
tory later on.
• Press NO to retain all data.
See chapter 6.4.2: "Clearing all data".
Depending on the duration of the standby mode, an automatic calibration can be per-
formed. After that, the analyzer returns to [READY] .
2.1.2 Reproducibility................................................................................................................................................2-1
Material: ISETROL Protein Based Aqueous Control Material (Level 1), n=80 ........................................2-2
Material: ISETROL Protein Based Aqueous Control Material (Level 2), n=80 ........................................2-2
Material: ISETROL Protein Based Aqueous Control Material (Level 3), n=80 ........................................2-2
Material: RNA EQUIL Reduced Bovine Hemoglobin Solution (Level 2), n=80 .......................................2-2
Material: Aqueous Standard Solution (Level 1), n=80 ....................................................................................2-3
Material: Aqueous Standard Solution (Level 2), n=80 ....................................................................................2-3
Material: Pooled Human Serum, n=80 ..................................................................................................................2-3
Material: Acetate Dialysate Solution, n=80 ..........................................................................................................2-3
Material: Bicarbonate Dialysate Solution, n=80 .................................................................................................2-4
Material: Urine, n=80 ....................................................................................................................................................2-4
2 Specifications
2.1 Performance parameters
2.1.2 Reproducibility
Typical Within-Run (S wr ), Between-Day (S dd ) and Total (S T ) Precision is determined from
2 runs per day with 2 replicates per run for 20 days on two STARLYTE V analyzers. Values
for sodium and potassium are average of all instruments, while values of chloride, ionized
calcium and lithium are determined from the measurement of two of each respective unit
configuration. All values are reported in mmol/L.
Material: I S ETROL Protein Based Aqueous Control Material (Level 1), n=80
Material: I S ETROL Protein Based Aqueous Control Material (Level 2), n=80
Material: I S ETROL Protein Based Aqueous Control Material (Level 3), n=80
2.2 Linearity
Correlation
Parameter Slope Intercept Coefficient Sy·x Range [mmol/L] n
Sodium 0.99993 0.0128 0.99995 0.666 51 − 196 300
Potassium 0.99838 0.0119 0.99919 0.194 2.0 − 12.6 300
Chloride 0.97556 -0.1775 0.99994 0.674 56 − 194 100
ionized Calcium 1.01552 -0.0078 0.99980 0.037 0.4 − 3.3 100
Lithium 0.99850 0.0087 0.99985 0.038 0.3 − 5.3 100
Correlation to flame
Correlation to chloridometry
2.3 Interferences
Salicylate, in extremely high levels, is known to interfere with the chloride electrode and
results in a positive bias of the chloride result. At therapeutic levels of salicylate concentra-
tion, the influence on chloride is clinically insignificant.
The lithium electrode shows a slight sensitivity to the ionized calcium present in the sample
and results in a negative bias of the lithium result. At normal physiological ionized calcium
concentrations, the influence of ionized calcium is clinically insignificant.
Use only lithium-free sampling containers for the determination of lithium mea-
surement values! If sample containers are used which contain lithium as an antico-
agulant, this may lead to incorrect patient measurements, which may result in
incorrect clinical decisions, possibly endangering the patient's health.
2.4 Limitations
A number of substances have been reported to cause physiological changes in blood, serum
and plasma analyte concentrations. A comprehensive discussion concerning these and other
interfering substances, their blood, serum or plasma concentrations, and their possible
physiological involvement is beyond the scope of this Instructions for Use. No significant
effect on serum has been demonstrated from bromide, ammonium and iodide.
As with any clinical reaction, users must be alert to the possible effect on results due to
unknown interference from medications or endogenous substances. All patient results must
be evaluated by the laboratory and the physician in light of the total clinical status of the
patient.
Opening and closing the fist with a tourniquet in place results in an increase in potassium
levels by as much as 10 to 20 %. It is recommended that the blood sample is obtained with-
out a tourniquet, or that the tourniquet be released after the needle has entered the vein and
2 minutes elapsed before the sample is withdrawn.
Since the concentration of potassium inside erythrocytes is much greater than that in extra
cellular fluid, hemolysis should be avoided, and the serum should be separated from the
cells as soon as possible after collection.
The lithium electrode response is dependent on the actual sodium concentration of the sam-
ple.
The STARLYTE V analyzer reports lithium in the range of 95 - 180 mmol/L
(89.6 - 169.5 mmol/L for blood/serum) Na + .
1. Tietz, Norbert W., Ed., Clinical Guide to Laboratory Tests, 2nd Ed., (Philadelphia: W.B.Saunders, Co., 1990) p.98-99, 118-119,
456-459, 510-511, 720-721.
2.6 Bibliography
Bishop ML, Duben-Engelkirk JL, Fody EP. Clinical Chemistry Principles Procedures Corre-
lations, 2nd Ed., (Philadelphia: J.B.Lippincott Co.),1992,p.281.
Burritt MF, Pierides AM, Offord KP: Comparative studies of total and ionized serum calci-
um values in normal subjects and in patients with renal disorders. Mayo Clinic Proc. 55:606,
1980.
Burtis C, Ashwood E (Eds.), Tietz Textbook of Clinical Chemistry, 2nd Ed., (Philadelphia:
W.B. Saunders, Co.,1994) pp.1354-1360,2180-2206.
Calbreath, Donald F., Clinical Chemistry A Fundamental Textbook, (Philadelphia: W.B.
Saunders Co., 1992) pp.371, 376, 390-395.
Henry, R.J., Clinical Chemistry - Principles and Technics, (New York, Harper and Row,
1974)
Kost GJ.Arch.Path.Lab.Med., Vol.117, Sep.1993, p.890-95
National Committee for Clinical Laboratory Standards. Protection of Laboratory Workers
from Occupationally Acquired Infections, Second Edition; Approved Guidelines; NCCLS
Document M29-A2, (2001).
National Committee for Clinical Laboratory Standards. Additives for Blood Collection
Devices: Heparin; Tentative Standard; NCCLS Document H24-T, (1988).
National Committee for Clinical Laboratory Standards. Evaluation of Precision Perfor-
mance of Clinical Chemistry Devices, Second Edition; Tentative Guideline. NCCLS Docu-
ment EP5-T2, (1992).
Rose, Burton David, Clinical Physiology of Acid-Base and Electrolyte Disorders, 4th Ed.,
(New York: McGraw-Hill, Inc., 1993) pp. 346-348, 432, 797-798.
Schoeff, Larry E & Williams, Robert H. (Eds.) Principles of Laboratory Instruments, (St.
Louis: Mosby Year Book Inc., 1993) pp. 150-157, 161-164.
Snyder John R., Senhauser Donald A, (Eds.), Administration and Supervision in Laboratory
Medicine, 2nd Ed, (Philadelphia: J.B.Lippincott Co., 1989) pp.262-284.
Tietz, Norbert W.,Ed.,Clinical Guide to Laboratory Tests, 2nd Ed., (Philadelphia: W.B.
Saunders Co., 1990), pp.98-99, 118-119, 456-459, 510-511, 720-721
Tietz, Norbert W.,Ed., Textbook of Clinical Chemistry, 2nd Ed., (Philadelphia: W.B. Saun-
ders, Co.,1986), pp.1816, 1837, 1840-1842, 1845.
Toffaletti J, Gitelman JH, Savory J: Separation and quantification of serum constituents
associated with calcium by gel filtration. Clin Chem 22: 1968-72, 1976.
2.10 Calibrations
1-point calibration after each sample
2-point or 3-point calibration every 4 hours
Instrument
Operating conditions:
Ambient temperature ........................................... 15 to 32 °C
Relative humidity ................................................. < 85 % (not condensing)
Electrodes
Operating conditions:
Operating temperature .................................................. 15 to 33 °C
Relative humidity ............................................................ 20 - 95 %, if T ≥ 15 to ≤ 31 °C
20 - 90 %, if T > 31 to ≤ 33 °C
Solutions
Fluid pack
Standard A
For calibration of sodium, potassium, ionized calcium and lithium in the STARLYTE V ana-
lyzer.
Contents: 350 mL
Active ingredients: Na + 150 mmol/L
K+ 5.0 mmol/L
Cl - 115 mmol/L
Ca 2+ 0.9 mmol/L
Li + 0.3 mmol/L
Additives: Germicides
Storage temperature: 15 - 30 °C
Stability: Expiration date and lot number are printed on each container label
On-board stability: 14 weeks
Standard B
For calibration of sodium, potassium, ionized calcium and lithium in the STARLYTE V ana-
lyzer.
Contents: 85 mL
Active ingredients: Na + 100 mmol/L
K+ 1.8 mmol/L
Cl - 72 mmol/L
Ca 2+ 1.5 mmol/L
Li + 0.3 mmol/L
Additives: Germicides
Storage temperature: 15 - 30 °C
Stability: Expiration date and lot number are printed on each container label
On-board stability: 14 weeks
Standard C
For calibration of sodium, potassium, ionized calcium and lithium in the STARLYTE V ana-
lyzer.
Contents: 85 mL
Active ingredients: Na + 150 mmol/L
K+ 5.0 mmol/L
Cl - 115 mmol/L
Ca 2+ 0.9 mmol/L
Li + 1.4 mmol/L
Additives: Germicides
Storage temperature: 15 - 30 °C
Stability: Expiration date and lot number are printed on each container label
On-board stability: 14 weeks
Reference Solution
A salt bridge for calibration and measurement in the STARLYTE V analyzer.
Contents: 85 mL
Active ingredients: Potassium chloride 1.2 mmol/L
Additives: Germicides
Storage temperature: 15 - 30 °C
Stability: Expiration date and lot number are printed on each container label
On-board stability: 14 weeks
A waste container is provided with the Fluid pack which, when used, holds human
body fluids which may be potentially infectious; handle with appropriate care to
avoid skin contact or ingestion.
Urine diluent
For use as a diluent for the measurement of urine samples in the STARLYTE V analyzer.
QC material
Temperature ............................ 15 to 30 °C
Temperature ............................ 2 to 30 °C
2 to 45 °C < 5 days
2.12.2 Classification
Safety Category ........................... I
Overvoltage category ................... II
Contamination level .................... 2
2.12.3 Dimensions
Height ......................................... 33.5 cm
Width .......................................... 31.5 cm
Depth .......................................... 29.5 cm
Weight ........................................ < 6 kg
2.13 Printer
Type thermal printer
Resolution 16 characters / line
Printing speed 4 mm / sec.
Paper width 38 mm
Paper length about 30 m
2.14 Display
Type monochrome LCD
Resolution 2 lines, 16 characters per line
3.2.2 Electrolytes.......................................................................................................................................................3-3
3 Measurement
3.1 Preanalytics
Safety
Follow the usual applicable safety precautions when drawing blood samples. When
handling blood samples, there always exists the danger of transmission of HIV, hep-
atitis B and C viruses or other pathogens transmissible by blood. Employ suitable
blood sampling techniques in order to reduce risk to personnel.
Suitable protective equipment, like laboratory clothing, protective gloves, protective
goggles and if necessary mouth protectors, must be worn to prevent direct contact
with biological working materials. In addition, a face mask is required if there is a
risk.
Please refer to NCCLS document M29-A2, "Protection of Laboratory Workers from Occupa-
tionally Acquired Infections", Approved Guidelines - Second Edition 2001, for further infor-
mation on safe handling of specimens.
Sample requirements
Only qualified personnel may perform the collection of blood needed for analytical purpos-
es.
The puncture site may never be squeezed! Mixing the blood sample with tissue fluid
may lead to the premature onset of clotting despite sufficient heparinization of the
sample collection containers! Incorrect sample collection or the use of an unsuitable
sample collection container may lead to errors and discrepancies in the measure-
ment values.
See, for example, NCCLS document H11-A3, "Procedures for the collection of arte-
rial blood specimens – Third Edition", Approved Standard, 1999, for detailed infor-
mation about blood sampling, storage, and handling.
Acceptable anticoagulants
The only clot inhibitors that may be used for analyses in the STARLYTE V analyzer are hep-
arin salts. Other clot inhibitors, such as EDTA, citrate, oxalate, fluoride, and ammonium-
based materials have significant influence on parameters and may not be used for this rea-
son.
Use only lithium-free sampling containers for the determination of lithium mea-
surement values! If sample containers are used which contain lithium as an antico-
agulant, this may lead to incorrect patient measurements, which may result in
incorrect clinical decisions, possibly endangering the patient's health.
Use only lithium-free sampling containers for the determination of lithium mea-
surement values! If sample containers are used which contain lithium as an antico-
agulant, this may lead to incorrect patient measurements, which may result in
incorrect clinical decisions, possibly endangering the patient's health.
Whole blood
Withdraw whole blood samples using heparinized syringes or capillaries. Analyze the sam-
ples as soon as possible after sampling. Remove air bubbles from the sample collection con-
tainer immediately after the sampling procedure. Samples analyzed within 15 minutes of
collection can be stored at room temperature. If analysis cannot be performed within
15 minutes, store the sample in an ice bath (Do not freeze!) immediately after collection.
Samples must be analyzed within 60 minutes of collection. Please consult the relevant liter-
ature.
Serum
After the appearance of spontaneous clotting, process the sample in a centrifuge to separate
the cellular, solid components and the fibrin from the watery serum. Transfer the serum to
a suitable sample container and seal.
If it is necessary to store the sample, close the sample container tightly and cool it to
4 - 8 °C. If a sample has been cooled, warm it to room temperature (15 - 33 °C) before anal-
ysis.
Plasma
Plasma samples are obtained by centrifuging heparinized whole blood, during which the cel-
lular components of the blood are removed from plasma.
Complete the analysis as quickly as possible.
If it is necessary to store the sample, close the sample container tightly and cool it to 4 -
8 °C. If a sample has been cooled, warm it to room temperature (15 - 33 °C) before analysis.
Plasma samples older than 1 hour must be re-centrifuged in order to remove fibrin clumps
that may have formed.
Each laboratory should determine the acceptability of its own blood collection syringes,
capillaries and tubes and the serum or plasma separation products. Variations in these prod-
ucts exist between manufacturers, and at times, from lot to lot.
Aqueous solutions
The Standard Mode ALWAYS reports direct ISE values and is not affected by corre-
lation factors or the setting of QCC (see chapter 6.3: "Service codes").
Dialysate
Either acetate or bicarbonate dialysate fluids may be analyzed. Dialysate samples can be
stored longer than whole blood, though preferably capped and placed in the refrigerator.
Prior to measurement, always bring the sample to room temperature.
3.2.1 General
The literature lists various substances which may negatively impact upon the measurement
result of the blood and plasma/serum sample material. A detailed discussion of these phe-
nomena can be found at different places in the technical literature. With respect to the
STARLYTE V analyzer, an attempt was made to identify or evaluate these possible influenc-
es. But since it is not possible to check all medication or substances, the user should be
immediately informed with abnormal deviations of the measurement results—as with every
clinical analysis—and evaluate the complete picture of the patient or perform expanded
measurements in his own laboratory, if necessary.
3.2.2 Electrolytes
It is well-known, for example, that the potassium value of a patient can vary by up to 20 %
from the normal state, simply because of the presence of a pressure bandage. Hence, taking
a blood sample while a pressure bandage is present should be avoided. In general, a local
hemolysis caused by pressure should be avoided prior to taking a blood sample.
fluid pack, after startup of the instrument as well as after monthly, semi annual and annual
maintenance steps.
Additionally, at least once daily one QC measurement has to be performed in alternating
levels (low, normal, high) (e.g., day 1 - level 1, day 2 - level 2, day 3 - level 3, day 4 - level 1,
etc.). When required by local regulations, QC measurements must be performed more often.
A quality control program for electrolytes includes the analysis of sample materials with
known ranges of expected values and the comparison of these values with analyzer results.
It is very important that the main door is closed during sampling to provide shield-
ing from sources of electromagnetic interference.
Urine samples require dilution, and must be analyzed in the urine mode. Instruc-
tions for analyzing urine samples: see chapter 3.3.4: "Urine samples".
• Open the sample door. The prompt [Introduce Sample] will be displayed and the
pump will start to aspirate.
Fig. 3-1
It is very important that the main door be closed during sampling to provide shield-
ing from sources of electromagnetic interference.
• Hold the ampoule under the probe until [Wipe Probe Close Sample Door] is dis-
played.
• Use a lint-free tissue to clean the probe, then close the sample door when prompted.
Fig. 3-2
It is very important that the sample probe is carefully cleaned each time following a
sample introduction.
The analyzer will display [Thank You!] and a brief countdown will begin. At completion
of analysis, the test results will be displayed and printed.
Na K Cl - STARLYTE V -
143.9 4.93 103.5 ELECTROLYTE ISE
- NA-K-CL -
HOSPITAL WEST
Name: ..........
..........
Sample: STANDARD
Sample No.28
Na=143.9 mmol/L
K = 4.93 mmol/L
Cl=103.1 mmol/L
Values that are higher or lower than the programmed normal range will be indicated
by an arrow pointing up or down.
If the Na+ result is above 180 mmol/L (above 169.5 mmol/L for blood/serum) or
below 95 mmol/L (below 89.6 mmol/L for blood/serum), the Li+ value cannot be
derived.
If an additional sample report is desired or the automatic sample report is turned off, the
results may be printed by following these steps:
• Press NO.
• The prompt [PRINT FUNCTIONS?] appears.
• Press YES .
• The prompt [Print last Sample Report?] is displayed.
• Press YES .
• The measurement results will be displayed, and the report will be printed.
• The analyzer returns to [READY] .
The Standard Mode ALWAYS reports direct ISE values and is not affected by corre-
lation factors or the setting of QCC (see chapter 6.3: "Service codes").
Na K Cl - STARLYTE V - - STARLYTE V -
263 18.3 118 ELECTROLYTE ISE ELECTROLYTE ISE
- NA-K-CL - - NA-K-CL -
HOSPITAL WEST HOSPITAL WEST
If the result of the urine sample is given with ↑↑↑↑ the K + value of the sample is
higher than 60 mmol/L, and outside the measurement range of the analyzer. The
measurement must be repeated using the following procedure:
• Record the Na + value (as well as Cl - , if activated) of the first urine measurement.
• Dilute the diluted urine (already diluted 1:2 with urine diluent) with distilled water in
the ratio of 1:1 (e.g., 1 mL of diluted urine and 1 mL distilled water).
• Thoroughly mix the sample.
• Run a second urine measurement with the twice-diluted urine sample.
• Ignore the Na + value (and Cl - , if activated).
• Multiply the K + value by 2 and record.
TIP: If an incorrect number was entered inadvertently, press NO , and the analyzer allows the
operator to retype the normal range.
• Press YES and the display will advance to the next parameter.
Repeat the above procedure for all parameters. After all activated parameters have been pro-
grammed, the display prompts [Program Corr Factors?] . This allows programming the
correlation factors.
In case the QC samples were switched to report direct ISE values (Code QCC, see
'Service Codes'), the values are NOT affected by correlation factors.
The measurement range as well as normal and QC ranges are ALWAYS checked
against the DISPLAYED values. Therefore, you may have to adjust the QC and nor-
mal ranges to your correlation factors.
The correlation factors can be changed through the programming menu. Follow the instruc-
tions for [PROGRAM INSTRUMENT?] in chapter 4.2 and enter the password.
Follow the same procedure for the intercept (b) and slope (m) for K + , Cl - , Ca 2+ and Li + .
The programmed factors are automatically applied to acetate and bicarbonate measure-
ments.
A separate set of correlation factors is available for urine samples.
• At the prompt [Program Urine Corr Factors?] , press YES .
• Program the factors as described above.
Note that there are no urine correlation factors available for Ca 2+ and Li + , since these
parameters are not measured in urine mode.
With factory settings, a measurement report is printed automatically after each measure-
ment, a calibration report is not printed automatically. These settings can be changed in the
programming menu.
• Press NO, until [Program Instrument?] is displayed. Press YES .
For safety and security, the analyzer can only be programmed or have existing
parameters changed by entering the correct password.
The analyzer prompts [Enter Code: AAA] . The code " KEY " has to be entered as follows:
• Press NO, until the character K appears.
• Press YES and the cursor will advance to the second position.
• Press NO, until the character E appears.
• Press YES and the cursor will advance to the last position.
• Press NO, until Y appears.
TIP: If the desired character was missed, keep pressing the NO key until the correct character
appears.
TIP: A copy of the last measurement report and the last calibration report as well as the QC
report can be printed in the menu [PRINT FUNCTIONS?] .
• The analyzer displays [Enter Comment Line?] . This setting allows the operator to
enter a comment (one line up to 16 characters) which will be printed on each report.
• Press YES , [Enter Comment:] is displayed.
• Press NO to select the desired letters A-Z or numbers 0-9 above the cursor.
• Press YES to advance the cursor to the next character.
Repeat this procedure for all characters.
4 Quality control
QC measurements must be performed in their entirety (i.e., all three QC levels must
be measured). Omitting QC measurements or ignoring QC measurement results may
lead to incorrect patient measurements, which may result in incorrect clinical deci-
sions, possibly endangering the patient's health.
The target values listed in the package insert sheet should be taken as 2 SD values
(SD = standard deviation).
The QC measurement results within the target value range ± 2SD are acceptable.
If QC measurement results fall outside the target value range ± 3SD, the parameter must not
be used for further measurements!
QC measurement results that are greater than the target value ± 2SD, but less than the target
value ± 3SD, must be treated accordingly.
For safety and security, the analyzer can only be programmed or have existing
parameters changed by entering the correct password.
The analyzer prompts [Enter Code: AAA] . The code " KEY " has to be entered as follows:
• Press NO, until the character K appears.
• Press YES and the cursor will advance to the second position.
• Press NO, until the character E appears.
• Press YES and the cursor will advance to the last position.
• Press NO, until Y appears.
TIP: If the desired character was missed, keep pressing the NO key until the correct character
appears.
• If the code has been entered correctly, press YES . The analyzer is ready for program-
ming and the display reads [Program QC Level 1 Ranges?] .
TIP: If a wrong code is entered accidentally, the following display appears: [CODE ERROR!
RETRY?] .
Press YES , and the system returns to the prompt [Enter Code:] . To exit this menu,
press NO .
If the pack of ISETROL electrolyte control is new, enter the lot number and the target ranges
into the analyzer. Each level has its own lot number which can be found on the information
sheet enclosed with the ISETROL pack.
The ISETROL electrolyte control is made specifically for use in the STARLYTE V
analyzer. To ensure the accuracy of the analyzer, only ISETROL should be used for
quality control.
• To enter the lot number, go to the display [Program QC Level 1 Ranges?] and
press YES .
• The analyzer display reads [Current Lot:xxxx Change Lot #?]
TIP: When a lot number is entered for the first time, the analyzer shows a preset lot number
which still has to be changed. If the current lot information does not have to be changed
and only the programmed QC ranges need to be verified, press NO .
• Press YES , and the analyzer display reads [Print old Values and Statistics?] .
• After you press YES , a log containing the following information is printed: mean value,
standard deviation (SD) and coefficient of variation (CV) of the saved data.
• Press NO. The display [NEW LOT! Delete old Data?] appears.
If this function is selected, all statistical data saved for this QC lot will be deleted
automatically.
TIP: If an incorrect lot number is entered, press NO to return to the prompt [Enter last
4 Dig of Lot: xxxx] .
The analyzer now shows the upper and lower ranges of the QC level parameters, e.g.:
Na low = 040
Na high = 205
TIP: Up to 35 QC measurements can be stored per level. The values can be stored and the
statistics be printed at any time.
It is very important that the main door be closed during sampling, since it provides
shielding from sources of electromagnetic interference.
• Hold the ampoule under the probe until [Wipe Probe/Close Sample Door]
is displayed.
• Use a lint-free tissue to clean the probe, then close the sample door.
It is very important that the sample probe be carefully cleaned each time following
a sample introduction.
• The instrument will display [QC Level 1 in progress] and a countdown will
begin, during which the QC will be analyzed. Upon completion, the results will be
briefly displayed, such as:
Na K ↓Cl
125.1 3.05 77.8
The STARLYTE V analyzer “flags” values that are above or below the programmed
target ranges by using an “up” or “down” arrow.
In case the value is outside the measurement range (↑↑↑↑, ↓↓↓↓ or ERR ), it will
automatically be rejected.
If the values have been saved, [Values Accepted!] will be displayed shortly, followed by
the prompt [QC Level 2 Sample?] .
For further QC measurements (Level 2 and Level 3), follow the outlined instructions for
Level 1, being sure to use a Level 2 or Level 3 ISETROL.
TIP: Responding YES to the [Remain in QC/Std./Urine Sample?] prompt, the ana-
lyzer returns to [QC Level 1 Sample?] .
If correlation factors are changed or QC values switched to direct ISE, ALL values in
memory are recalculated and printed per the new setting.
5 Maintenance
After use, components of the STARLYTE V analyzer, including tubing, fill port,
fluid pack etc., contain biological fluids and therefore represents a possible infec-
tious risk.
Handle these components with care and according to regulations surrounding
potentially infectious materials.
Suitable protective equipment, like laboratory clothing, protective gloves, protective
goggles and if necessary mouth protectors, must be worn to prevent direct contact
with biological working materials. In addition, a face mask is required if there is a
risk of splashes. Suitable disinfection and sterilization procedures must be applied.
5.1 Decontamination
5.1.1 Decontamination
The purpose of this procedure is to minimize the risk of infections when replacing items that
were in contact with blood.
Perform these decontamination procedures regularly.
Recommended disinfectants
Surfaces
Protein remover (Roche deproteinizer)
• Potential dangers
Due to the alkaline and oxidizing character of this preparation, local irritation to the
skin, eyes, and mucous membranes is possible.
• First Aid measures
After inhalation: breath fresh air, drink large amounts of water
After skin contact: wash with generous amounts of water, remove contaminated cloth-
ing.
After eye contact: rinse eyes with generous amounts of water, contact an eye doctor
After drinking: drink large amounts of water, avoid vomiting, contact a doctor.
In case cleaning and/or conditioning has not been performed within the last 24
hours, the analyzer will automatically print on each sample report [Perform Dai-
ly Maintenance!] .
TIP: If fewer than 5 samples are analyzed each day, cleaning should be performed once a
week instead of daily (see "Weekly maintenance").
• Press the NO key, until [DAILY MAINTENANCE?] is displayed, and YES to select.
Select [Perform Daily Cleaning?] by pressing YES .
• At the prompt [Open Sample Door Introduce Sample] , pour a small amount of
Cleaning Solution into a clean container. Lift the sample door and the pump will begin
to aspirate.
• Hold the Cleaning Solution under the probe until [Wipe Probe/Close Sample
Door] is displayed, and use a lint-free tissue to remove the cleaning solution from the
probe. Close the sample door.
• While the analyzer displays [Thank You!] and a brief countdown, pour a small
amount of conditioning solution into a clean container.
• At the prompt [Perform Daily Conditioning?] , press YES .
• At the prompt [Open Sample Door Introduce Sample] , lift the sample door and
the pump will begin to aspirate.
• Hold the conditioning solution under the probe until [Wipe Probe/Close Sample
Door] is displayed, use a lint-free tissue to remove the conditioning solution from the
probe. Close the sample door.
• After the analyzer displays [Thank You!] and a brief countdown, press NO at the
prompt [Remain in Daily Maintenance?] . An automatic calibration will be
started.
It is very important that the main door be closed during calibration, since it provides
shielding from sources of electromagnetic interference.
Wipe the sample probe from top to bottom in order to avoid injury.
Never use strong or abrasive cleaners and no cleaners containing alcohol on the
STARLYTE V analyzer. Use a slightly damp cloth to avoid getting fluid inside the
analyzer.
Fig. 5-3
Fig. 5-4
• Slide the measuring chamber forward until it locks in the front position.
• Unclamp the left electrode holder by moving the clamp forward.
Fig. 5-7
TIP: The reference solution can be taken from the reference electrode housing.
• Pour deproteinizer solution into a small container and submerge the reference housing
into it, ensuring that no air bubbles remain in the housing.
• After 15 minutes, remove the reference housing from the deproteinizer. Thoroughly
rinse the housing with tap water and dry.
• Unscrew the red transport housing from the reference electrode and check that the
o-ring on the electrode is properly seated. Save the transport housing.
• Carefully screw the reference electrode into the reference electrode housing and place
the assembly into the left side of the measuring chamber.
• Close the clamp on the left electrode holder by lifting it until it locks in the back posi-
tion. Ensure the electrodes are seated properly.
Fig. 5-8
• Plug the tubing of the reference electrode into the receptacle below the left side of the
measuring chamber.
• Slide the measuring chamber back until it snaps into position, and close the front cover
of the analyzer.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each replace-
ment of the Fluid pack, after startup of the instrument as well as after monthly, semi
annual and annual maintenance steps.
Additionally, at least once daily one QC measurement has to be performed in alter-
nating levels (low, normal, high) (e.g., day 1 - level 1, day 2 - level 2, day 3 - level 3,
day 4 - level 1, etc.). When required by local regulations, QC measurements must be
performed more often.
• Install the new pump winding plate and slip the new pump windings over the pump
rollers, being careful not to cross the tubes.
• Press NO until [CALIBRATION?] is displayed. Press YES to initiate a calibration cycle.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each replace-
ment of the Fluid pack, after startup of the instrument as well as after monthly, semi
annual and annual maintenance steps.
Additionally, at least once daily one QC measurement has to be performed in alter-
nating levels (low, normal, high) (e.g., day 1 - level 1, day 2 - level 2, day 3 - level 3,
day 4 - level 1, etc.). When required by local regulations, QC measurements must be
performed more often.
• At the [READY] screen press NO, until [SERVICE FUNCTIONS?] appears, and press
YES .
• Press NO, until [Test Pinch Valves?] appears. Press YES .
• Remove the valve caps, sliding them in the direction of the arrow.
• Then remove the pinch bars.
TIP: Each valve consists of the solenoid coil inside the analyzer, the pinch bar and the valve
cap. For further details concerning the position of the valves, see tubing diagram inside
the front door.
• Disconnect the reference electrode and pull out the white reference electrode
receptacle.
• Disconnect the tubing with the green band from the left side electrode holder.
• Open the sample flap and pull out the fill port.
• Remove the Fluid pack and slide out the white Fluid pack receptacle. Disconnect the
tubes from the pump winding plate and pull out the tubing harness.
• Re-install the new main harness in reverse order, use the tubing diagram in the front
door as a guide for installation. Take special care to reconnect the tubes to the correct
nipples of the pump winding.
• Activate each valve by pressing YES , then slide the pinch bar on and snap on the valve
cap in the direction of the arrow.
• Press NO to activate the next valve. Notice the text on the display.
TIP: It may be necessary to pull on the metal tip of the valve in order to put the valve cap
into position.
TIP: It is recommended to replace the sample probe and the fill port at the same time as the
main tubing harness is replaced (see section 5.7.3: Replacing sample probe and fill
port).
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each replace-
ment of the Fluid pack, after startup of the instrument as well as after monthly, semi
annual and annual maintenance steps.
Additionally, at least once daily one QC measurement has to be performed in alter-
nating levels (low, normal, high) (i.e,. day 1 - level 1, day 2 - level 2, day 3 - level 3,
day 4 - level 1, etc.). When required by local regulations, QC measurements must be
performed more often.
Fig. 5-15
• Remove the new electrode from its protective box and check for the presence of an
o-ring in the left side of the electrode.
• Install electrode in its labeled position in the measuring chamber. Note that the elec-
trode has a lip on the bottom that rests on the flat edge of the measuring chamber to aid
in proper positioning.
• Close the clamp on the left electrode holder by lifting it until it locks in the back posi-
tion. Ensure that the electrodes are seated properly.
• Slide the measuring chamber back until it snaps into position.
After installing a new electrode, the STARLYTE V analyzer needs to undergo Daily Mainte-
nance, Calibration and QC Measurement to verify the performance of the electrode.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each replace-
ment of the Fluid pack, after startup of the instrument as well as after monthly, semi
annual and annual maintenance steps.
Additionally, at least once daily one QC measurement has to be performed in alter-
nating levels (low, normal, high) (e.g., day 1 - level 1, day 2 - level 2, day 3 - level 3,
day 4 - level 1, etc.). When required by local regulations, QC measurements must be
performed more often.
A quality control program for electrolytes includes the analysis of sample materials
with known ranges of expected values and the comparison of these values with ana-
lyzer results. For further information, see chapter 4: "Quality control".
5.7.2 Checking reagent fluid level and changing the Fluid pack
The STARLYTE V analyzer monitors the level of solutions in the Fluid pack and displays the
amount remaining. To check the status of fluid remaining in the Fluid pack:
• In the [READY] screen, press the NO key, until [OPERATOR FUNCTIONS?] is displayed.
Press YES .
• The prompt [Check/Change Fl.Pack?] will be displayed. Press YES .
• The analyzer will display the amount of fluid remaining.
• To change the Fluid pack, just grasp the extended portion and slide it out.
Fig. 5-20
As soon as the analyzer detects that the Fluid pack has been removed, the display prompts
[STATUS: NO Fl.Pack] . In addition, the printer will print the current status of the Fluid
pack.
• Prepare the new Fluid pack by carefully removing the protective strip. Save this cover to
close the connectors prior to disposal of the used Fluid pack.
• Write the installation date on the label of the new Fluid pack.
Once the protective strip is removed, be sure to keep the Fluid pack upright to avoid
spillage.
• Slide the new Fluid pack into position on the left side of the analyzer.
The analyzer will prompt [New Fl.Pack Installed?] .
• Press YES to indicate that a new Fluid pack is installed. Again, at the prompt [Are you
sure?] , press YES . The STARLYTE V analyzer will automatically reset the Fluid pack
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each replace-
ment of the Fluid pack, after startup of the instrument as well as after monthly, semi
annual and annual maintenance steps.
Additionally, at least once daily one QC measurement has to be performed in alter-
nating levels (low, normal, high) (e.g., day 1 - level 1, day 2 - level 2, day 3 - level 3,
day 4 - level 1, etc.). When required by local regulations, QC measurements must be
performed more often.
TIP: It is recommended to replace the sample probe and the fill port at the same time as the
main tubing harness is replaced (see section 5.6.1: Exchanging main tubing harness).
Fig. 5-23
tubing.
• Mount the fill port holder at it’s default position.
• Close the sample door and the front cover.
• Press NO until [CALIBRATION?] will be displayed.
• Press YES to start the calibration.
Dispose of the sample probe and the fill port according to applicable regulations
(hazardous waste!).
TIP: The printer paper is heat sensitive on one side only. Please make sure that you insert
the paper roll correctly.
The thermal printer paper supplied by Alfa Wassermann contains an indicator strip to alert
you when the paper roll should be changed. To change the roll:
• Open the front cover of the analyzer.
• Remove the remaining paper by pressing the paper feed button.
• Place a new paper roll in the paper tray and thread it into the feeder slot.
• Press the paper feed button to completely feed the paper through the printer.
TIP: By pressing the paper advance button once, the paper will automatically advance 10
lines.
• Close the front cover and tear off any excess paper.
• Cut the end of the roll to make a square edge.
Do not pull the paper out of the printer to avoid damage to the printer.
6 Troubleshooting
Your STARLYTE V analyzer is designed to provide long, trouble-free service. However, any
scientific measuring device may occasionally malfunction, requiring the operator to identify
the cause of the problem.
The following information lists displayed messages and recommends steps that should help
you return your STARLYTE V analyzer to operation. If your STARLYTE V analyzer does not
perform correctly after conducting the basic steps outlined in this chapter, you should con-
tact Alfa Wassermann for technical assistance.
K+ -700 to +1000 -2500 to +500 -700 to +100 +470 to +1200 -40 to +40
Ca 2+ -3100 to +1000 -2300 to +2500 -3100 to +1000 -660 to -350 -150 to +150
TIP: If the voltage difference A-B or A-C is outside the allowable difference, perform daily
maintenance or replace the electrode.
• After testing the electrodes against the standard solutions, the STARLYTE V analyzer
will prompt [Test External Sample?] . Press YES .
• The instrument will prompt [Open Sample Door Introduce Sample] .
• After introducing the sample, the analyzer will display [Thank you!] shortly, followed
by the electrode voltages.
• Check for stable readings.
• Press NO to exit and after a short [Please Wait] message, the prompt [Remain in
Test Electrodes?] will appear.
• Press NO to exit or to perform other service functions.
If the potassium electrode cannot be calibrated after an electrode exchange, then a serum or
plasma sample should be introduced as an external sample.
• Call up the procedures for testing electrodes (see "Testing the electrodes" on page 6-7).
• Press NO, until [Test External Sample?] is displayed. Press YES .
• Press YES introduce an external sample through the probe. The display will be like:
External: 236
Pump: NO=OFF
• Press NO to exit and after a short [Please Wait] message, the prompt [Test
Fl.Pack Sensor?] will appear.
If the Fluid pack has been removed during this test, a calibration should be per-
formed to prime all fluid lines.
To activate a new language setting, turn the analyzer off and back on.
English ᣣᧄ⺆
Deutsch Polski
Français Português
Español Spare 3
Italiano Spare 4
Fig. 6-1
---Test Pump!---
• Close the front cover and press NO to exit or to perform other service functions.
Be sure not to short pin 2 or pin 3 to ground, otherwise the interface may get dam-
aged.
• The analyzer will display [Interface Test] and tries to send some characters and
checks if they are received within a set time period.
• Check that there was no interface error.
• Press NO to exit or to perform other service functions.
6.2.10 Status-LEDs
The 5 status LEDs at the bottom of the chassis back panel are lit simultaneously during nor-
mal operation. If this is not the case, contact technical support.
TIP: If the desired character was missed, keep pressing the NO key until the correct character
appears.
If an incorrect code is entered during the process, the analyzer will prompt [Code
Error! Retry?] . Press YES to correct the code or NO to exit.
Code Action
DEC This service code will increase the resolution by one digit for all para-
meters in blood and serum samples. QC and standard samples are
always shown in high resolution, urine samples in low resolution.
ECO This service code will automatically put the instrument into Standby
Mode, if no samples are run between two calibrations (4 hrs.). This fea-
ture mainly helps to conserve reagents. When the unit automatically
enters the Standby Mode, it is always necessary to perform a calibration
to get back to the [READY] prompt.
FIF This service code will automatically print the voltage levels of the elec-
trodes on sample and calibration reports.
This code should only be used by trained service technicians. To disable
this code, either enter IFI or turn the power off and back on.
LEM This service code will print the last 20 error messages with the date and
time of occurrence. MEL will erase all error messages stored.
MGL This service code allows you to report Ca 2+ values in mg/dL instead of
Ca 2+ activated mmol/L. When this code is set, the unit automatically converts the cur-
rent high and low values of the normal ranges to mg/dL. Note that QC
and standard measurements are always reported in mmol/L independent
of the MGL setting. This allows you to compare the Ca 2+ values to the
ISETROL assay ranges.
NOB This service code disables the beep.
SLC When this code is set, the current setting of the leasing counter, which
cannot be reset by the operator, is printed on the calibration report.
SFC This code allows the input of a percentage for the Fluid pack counter.
When this code is activated, the analyzer displays:
Fl.Pack Counter:
xxx% ok?
If this percentage is not correct, press NO and enter the correct percen-
tage between 0 and 100 % using the NO and YES keys. Pressing YES to
[ok?] will return the analyzer to the ready screen.
QCC When this code is enabled (using QCC) QC measurements will be repor-
ted in direct ISE values without correlation factors.
In order to disable this code enter the code CCQ. When the code is
disabled (using CCQ) QC measurements will be reported as flame equi-
valent values with correlation factors applied.
NOTE: The default condition is that this code is disabled.
With the exception of FIF, all service codes are permanently stored in memory and will be
retained during power failures. The service codes currently set can be printed on the cali-
bration report.
All QC values and statistics will be deleted and the sample number reset to 0. QC and normal
values as well as correlation factors, printer settings, date/time and liquid content of the
fluid pack will be reset to default values. All service codes will be deactivated. It is recom-
mended that all data be deleted if the analyzer is to be used by other persons at a later date.
7.1.3 Chloride.............................................................................................................................................................7-3
7.1.5 Lithium...............................................................................................................................................................7-6
7 Theoretical foundations
7.1 Clinical significance
7.1.1 Sodium
The vast majority of sodium in organisms is located in the extracellular area (about 97 %).
Even with greatly varying supply with nourishment, the sodium concentration in serum is
subject to strong regula tion. In the kidneys, sodium is glomerularly filtered and most of this
(about 60 - 70 %) is reabsorbed in the proximal tubule.
The most important function of the sodium is to maintain constant osmolarity in the extra-
cellular fluid. For that reason, the levels of sodium and water are always closely interrelated.
However, in pathological conditions, there may be marked tissue hypo or hyperhydration
with values in the standard range. Conversely, an increased, respec tively a decreased sodium
concentration is found in serum due to a loss or gain of water when there is a normal level
of sodium.
An increased sodium level in serum occurs when there is:
• a decreased supply of liquid
• increased loss of water
• through the kidneys
- central diabetes insipidus
- renal diabetes insipidus
- osmotic diuresis (e.g. mannitin fusions)
• through the intestine
- infection deseases (especially dysentery and cholera)
• excessive supply of hypertonic saline solution (infusion therapy dosed too high)
• increase of aldosterone-induced sodium reabsorption
- primary hyperaldosteronism (CONN syndrome)
- secondary hyperaldosteronism
Normal values:
Adult: 135 - 148 mmol/L
Newborn: 134 - 144 mmol/L
Child: 138 - 144 mmol/L
Critical values a :
7.1.2 Potassium
About 97% of potassium within the organism is intracellular. Transport into the cells is reg-
ulated by the Na/K ATPase localized in the cell membrane. Only about 3% of the potassium
is contained in the extracellular fluid. Potassium is glomerularly filtered and most of it
(about 90%) is reabsorbed in the proximal tubule and in Henle's loop. Reabsorption or
excretion in the distal tubules is influenced especially by aldosterone and the blood pH val-
ue.
Due to the high intracellular concentration of potassium, the serum potassium values do not
always reflect the potassium level of the organism. Therefore, the data obtained from the
serum may be interpreted only with careful consideration of the patient's clinical situation
and acid-base status. Consider the following examples: diabetic coma, during which the flow
of potassium into the cell is reduced due to the lack of insulin, and acute intoxication with
heart glycosides with accompanying inhibition of the Na/K ATPase membrane. In both cases
exists, despite a more or less greatly increased serum potassium level, intracellular potassi-
um deficiency.
Increased potassium concentration in serum occurs during:
• decreased excretion through the kidneys
- acute and chronic kidney insufficiency (especially pronounced with oliguria and
anuria)
- Aldosterone deficiency with suprarenal gland insufficiency (M. ADDISON)
- dosage of potassium-saving diuretic
- oral potassium substitution with (possibly unknown) mild limitation of kidney func-
tions
• displacement between intracellular and extracellular potassium
- severe insulin deficiency
- intoxication with heart glycosides
- severe acidosis
- (each 0.1 reduction of the blood pH results in a rise in potassium of 0.4 to 1.2 mmol/
L serum)
- malignant hyperthermia
• release of potassium on massive cell destruction
- hemolytic crisis
- transfusions with cold or very cold blood
- cytostatic therapy for leukemia and others
- burns
- severe soft tissue injuries
- massive diarrhea
- fistulas in the area of the gastrointestinal tract
- villous papillary adenoma
- VERNER-MORRISON syndrome (pancreatic cholera)
• increased renal excretion
- primary hyperaldosteronism (CONN syndrome)
- secondary hyperaldosteronism
- cirrhosis of the liver (caused by decreased aldosterone breakdown)
- therapy with loop diuretics and thiazides
- CUSHING syndrome
- Aldosterone producing suprarenal gland carcinoma
- overdose of mineral corticoids
- renal tubular acidosis
• displacement between intracellular and extracellular potassium
- severe alkalosis
- insulin therapy for diabetic coma (potassium substitution required!)
Normal values:
Adult: 3.5 – 4.5 mmol/L
Newborn: 3.7 – 5.9 mmol/L
Child: 3.4 – 4.7 mmol/L
Critical values a :
7.1.3 Chloride
Chloride is the most important anion in bodily fluids. Chloride is located – like sodium –
mostly in the extracellular area. Erythrocytes represent the highest intracellular content.
The concentration of chloride in serum, like the level of sodium, is held constant within
tight limits in healthy people. Chloride is glomerulary filtered in the kidneys and is tubu-
larly reabsorbed by passively following the sodium.
Chloride may be exchanged for bicarbonates during disturbances to the acid/base status,
causing chloride to adopt the additional task (in addition to maintaining the isotones in the
extracellular area) of working with sodium to regulate the acid/base status.
Changes to the chloride and sodium concentrations in serum usually occur in parallel.
Exceptions to this occur during disturbances to the acid/base status caused by the previously
mentioned exchange of chloride for bicarbonates as well as during massive chloride loss
with gastric juices during extended periods of vomiting (hypochloremic alkalosis).
Normal values a :
Adult: 98 - 107 mmol/L
Normal values a :
Newborn: 98 - 113 mmol/L
a. Tietz Textbook of Clinical Chemistry, 3rd Edition 1999
Critical values a :
Cl - < 75 mmol/L
Cl - > 126 mmol/L
a. Critical Care Testing: A Quick Reference Guide by Andrew St John, First Edition 2001
Calcitonin:
• reduces the level of calcium
Increased concentrations of calcium in serum occur during:
• disturbances to the hormonal regulation of primary and tertiary hyperparathyreoidism
• increased release from the bones
- osteolysis through bone metastasis
- plasmocytome
- paraneoplastic symptom (through ectopic production of parathormone or similar
substances or prostaglandin E2)
- long-lasting immobilization
• vitamin D intoxication within the scope of therapeutic measures
• sarcoidosis
Reduced calcium levels in serum will be noticed as a result of:
• insufficient calcium reabsorption
- undernourishment
- malabsorption syndrome
- vitamin D3 deficiency
- deficiency of 1.25-dihydroxycholecalciferol
- chronic kidney insufficiency
- hypoparathyreoidism
- hypomagnesium
• greatly decreased concentration of albumin in the serum
(Note: ionized calcium is in the normal range!)
- nephrotic syndrome
- cirrhosis of the liver
• acute pancreatitis
Normal values:
Adult: 1.12 – 1.32 mmol/L
Child: 1.10 – 1.50 mmol/L
Critical values a :
7.1.5 Lithium 1
In clinical practice, lithium is given orally in a carbonate or citrate form in the treatment of
recurrent bipolar affective disorders. Bipolar disorders are psychotic illnesses characterized
by alternating mania and depression (typical manic-depressive symptoms).
The mechanism of action of lithium is thought to be based on its effect on the synthesis,
storage, release and reuptake of monoamine neurotransmitters. Onset of the acute antiman-
ic effect of lithium usually occurs within five to seven days; full therapeutic effect requires
10 - 21 days.
Lithium is absorbed quickly by the gastrointestinal tract; approximately 90 - 95 % of the
lithium dosage is excreted through the kidneys. GFR disorders reduce lithium elimination.
Excretion through perspiration, saliva and feces is negligible in normal circumstances.
The therapeutic range of lithium is very narrow. To avoid intoxication (dizziness, nausea,
trembling, and, ultimately, coma and shock), the lithium level must be determined every
4 - 5 days at the start of therapy. A blood sample is taken 8 - 12 hours after the last lithium
dose is administered. If the patient is clinically stable and the lithium level constant, the
monitoring interval can be extended to 1 - 3 months.
Therapeutic range a :
0.6 - 1.2 mmol/L
Toxic range:
Li + > 2 mmol/L
a. Tietz Textbook of Clinical Chemistry and Molecular Diagnostics. Eds: Burtis C & Ashwood E & Bruns
D. Elsevier Saunders 4th Ed., USA 2006
1
Gelenberg AJ, Kane JM, Kekller MB et al. Comparison of standard and low serum levels of lithium for maintenance
treatment of bipolar disorder. N Engl J Med 1989; 321:1489-93.
voltmeter
reference
electrode
inner solution
ion-selective
electrode
glass capillary
ion-selective
membrane
measured
solution
Fig. 7-1
R⋅T
1. E = E′ ± ----------- ⋅ ln a i
n⋅F
or
R⋅T
2. E = E′ ± ----------- ⋅ ln ( f i ⋅ c i ) (+) for cations
n⋅F (–) for anions
Na + electrode K + electrode
ground amplifier display
reference
electrode
housing
reference
solution (KCl)
reference
electrode Cl - electrode
glass
capillary inner electrode
inner electrolyte
sample
chloride-selective
sodium-selective potassium-selective membrane
membrane membrane
Fig. 7-2
If the ion concentration of one measuring solution is known, the ion concentration of the
sample can be determined on the basis of the difference of two measured potentials.
c
i, sample
6. ΔE = E sample – E s tan dard = S ⋅ log -----------------------
-
c i, standard
ΔE the difference between the measured potentials of the sample and the
standard
S the potential difference of the electrode, determined from the potential
difference of two measured standard solutions
c i,sample concentration of the measured ions in the sample
c i,standard concentration of the measured ions in the standard solutions
The unknown ion concentration in the sample can now be determined by:
ΔE ⁄ S
7. c i, sample = c i, s tan dard ⋅ 10
As demonstrated by these equations, the ion selective electrodes do not measure the ion
concentration but the activity of the ions concerned. This activity is a criterion of the ion’s
ability to interact with other ions, in which each ion binds a proportion of its energy.
The ion concentration is calculated on the basis of the ion activity. The correlation is affect-
ed by the total number of ions in the solution. Since sodium is the predominant ion in whole
blood and serum, the known value of the sodium concentration makes it possible to ascer-
tain and adjust for total ion effect and strength.
Construction
O-ring seal
electrolyte chamber
inner electrode
glass capillary
Fig. 7-3
Construction
ion-selective membrane
O-ring seal
electrolyte chamber
measuring capillary
serial number
connecting pin
inner electrode
Fig. 7-4
Construction
ion-selective membrane
O-ring seal
electrolyte chamber
measuring capillary
serial number
connecting pin
inner electrode
Fig. 7-5
Construction
ion-selective membrane
O-ring seal
electrolyte chamber
measuring capillary
serial number
connecting pin
inner electrode
Fig. 7-6
Construction
ion-selective membrane
O-ring seal
electrolyte chamber
measuring capillary
serial number
connecting pin
inner electrode
Fig. 7-7
Construction
The reference electrode assembly consists of two parts: the reference electrode housing and
the reference electrode.
In the reference electrode housing, reference electrolyte solution establishes the electrical
contact between the reference electrode and the sample. At the beginning of each measure-
ment, reference electrolyte is pumped into the housing. At the same time a glass capillary
allows a small amount of reference electrolyte to pass into the measuring capillary, thus
establishing electrical contact between the sample and the reference electrode (see Fig. 7-8).
entrance of exit of
ref. solution ref. solution
glass capillary
Fig. 7-8
Reference electrode
The reference electrode completes the electrical circuit between the reference electrolyte
and electrical ground. This is accomplished by a cotton wool (saturated with reference elec-
trolyte)-calomel (Hg 2 Cl 2 )- mercury (Hg)-platinum wire-connecting pin junction
(see Fig. 7-9).
cable housing
platinum wire
glass tube
cotton wool
calomel (Hg 2 Cl 2 )
mercury (Hg)
Fig. 7-9
8 Appendix
8.1 Description of reports
- STARLYTE V -
ELECTROLYTE ISE
- NA-K-CL -
26SEP03 08:16 - STARLYTE V - - STARLYTE V -
HOSPITAL WEST ELECTROLYTE ISE ELECTROLYTE ISE
- NA-K-CL - - NA-K-CL -
Name: .......... 26SEP03 08:27 26SEP03 08:31
.......... HOSPITAL WEST HOSPITAL WEST
Sample: STANDARD
Name: .......... Name: ..........
Sample No.35 .......... ..........
Sample: SERUM Sample: URINE
Na=151.7 mmol/L
K = 5.56 mmol/L Sample No.36 Sample No.37
Cl=114.5 mmol/L
Na= 263 mmol/L Na=143.9 mmol/L
*PERFORM DAILY * K = 18.3 mmol/L K = 4.93 ↓ mmol/L
* MAINTENANCE! * Cl= 118 mmol/L Cl=103.1 mmol/L
Fig. 8-1 Fig. 8-2 Fig. 8-3
8.1.2 QC report
- STARLYTE V -
ELECTROLYTE ISE
- NA-K-CL -
26SEP03 09:21
HOSPITAL WEST
QUALITY CONTROL
Fig. 8-4
- STARLYTE V -
ELECTROLYTE ISE
Normal Ranges:
- NA-K-CL -
Units: [mmol/L]
26SEP03 09:01
Na: 136 - 145
HOSPITAL WEST
K : 3.5 - 5.1
*CALIBR. REPORT* Corr. Factors:
Cl: 97 - 111
Na Offs.= 0.0
DailyMaintenance Na Slope= 1.000
Performed Last: K Offs.= 0.0
QC Ranges:
26SEP03 08:16 K Slope= 1.000
Units: [mmol/L]
Cl Offs.= 0.0
Standard A Cl Slope= 1.000
*** Level 1 ***
Na = -369mV (4) Bicarb. Factors:
Lot number:1150
K = -1052mV (4) Na Offs.= 0.0
Na: 40 - 145
Cl = -183mV (4) Na Slope= 1.000
K : 1.5 - 15.0
K Offs.= 0.0
Cl: 50 - 200
Difference A-B: K Slope= 1.000
Na = 584mV ( ) Cl Offs.= 0.0
*** Level 2 ***
K = 1056mV ( ) Cl Slope= 1.000
Lot number:1261
Cl = -689mV ( ) Acetate Factors:
Na: 40 - 145
Na Offs.= 0.0
K : 1.5 - 15.0
Temp. = 26.2°C Na Slope= 1.000
Cl: 50 - 200
Fluid pack: K Offs.= 0.0
83% Remaining K Slope= 1.000
*** Level 3 ***
Cl Offs.= 0.0
Lot number:1373
Cl Slope= 1.000
Na: 40 - 145
Urine Factors:
K : 1.5 - 15.0
Na Offs.= 0.0
Cl: 50 - 200
Na Slope= 1.000
K Offs.= 0.0
Service Codes:
K Slope= 1.000
, , ,
Cl Offs.= 0.0
,NOB, ,
Cl Slope= 1.000
, , ,
Fig. 8-5
Caution: A waste container is provided inside the Fluid pack which, when used,
holds human body fluids which may be potentially infectious; handle with appropri-
ate care to avoid skin contact or ingestion!
STARLYTE V analyzer
Maintenance schedule
Month:...........................
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31
DAILY
Perform cleaning
Maintenance schedule
Perform conditioning
WEEKLY
Clean sample probe/fill port
Clean analyzer surfaces
MONTHLY
Clean reference electrode housing
TIP: This page can be used as template for duplication.
SEMI-ANNUALLY
Change pump tubing
ANNUALLY
Change main tubing harness
UNSCHEDULED MAINTENANCE