Elicrom Peru Calibration Scope

This document provides the scope of accreditation for ELICROM PERU S.A.C. to ISO/IEC 17025 for calibration in various fields including acoustics, chemistry, dimensional measurements, electrical measurements, and insulation resistance. The laboratory is accredited to calibrate equipment such as sound level meters, conductivity meters, pH meters, calipers, gauges, microscopes, and electrical equipment for measurements of voltage, current, resistance, and more. Calibration methods, measurement ranges, and uncertainties are defined for numerous parameters and types of equipment.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
147 views18 pages

Elicrom Peru Calibration Scope

This document provides the scope of accreditation for ELICROM PERU S.A.C. to ISO/IEC 17025 for calibration in various fields including acoustics, chemistry, dimensional measurements, electrical measurements, and insulation resistance. The laboratory is accredited to calibrate equipment such as sound level meters, conductivity meters, pH meters, calipers, gauges, microscopes, and electrical equipment for measurements of voltage, current, resistance, and more. Calibration methods, measurement ranges, and uncertainties are defined for numerous parameters and types of equipment.
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 18

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2017

ELICROM PERU S.A.C.


Av. Faustino Sánchez Carrión N° 615, Off # 502
Lima, Perú 15076
José Aparcana Phone: 051 017669297

CALIBRATION

Valid To: August 31, 2021 Certificate Number: 4286.04

In recognition of the successful completion of the A2LA evaluation process, accreditation is granted to this
laboratory to perform the following calibrations1, 6:

I. Acoustics

Parameter/Range Frequency CMC2, 5 () Comments

Sound Level Meters and


Dosimeters –

Acoustic Calibration
(94, 104, 114) dB 31.5 Hz 0.20 dB Brüel & Kjær 4226
re: 20 μPa 63 Hz 0.20 dB
125 Hz 0.20 dB
250 Hz 0.15 dB
500 Hz 0.15 dB
1000 Hz 0.13 dB
2000 Hz 0.20 dB
4000 Hz 0.20 dB
8000 Hz 0.28 dB
12 500 Hz 0.51 dB
16 000 Hz 0.51 dB

Electric Calibration
(30 to 150) dB (20 to 20 000) Hz 0.078 dB Transmille 3041A

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 1 of 17


Parameter/Equipment Range CMC2 () Comments

Acoustic Calibrator 94 dB 0.20 dB Sound Level Meter


114 dB 0.20 dB Transmille 3041A

94 dB to 1 kHz 0.24 Hz
114 dB to 1 kHz 0.24 Hz

II. Chemical

Parameter/Equipment Range CMC2, 5 () Comments

Conductivity Meters3 1 µS/cm 0.56 µS/cm Conductivity standard


5 µS/cm 0.56 µS/cm solution
10 µS/cm 0.56 µS/cm
100 µS/cm 2.2 µS/cm
1000 µS/cm 4.7 µS/cm
1413 µS/cm 4.7 µS/cm
10 000 µS/cm 41 µS/cm
100 000 µS/cm 370 µS/cm

pH Meters3 4 pH Unit 0.012 pH Unit pH buffer solution


7 pH Unit 0.012 pH Unit
10 pH Unit 0.012 pH Unit

Turbidity Meters3 (0 to 4000) NTU 1.5 % rdg Stabilized Formazin


turbidity standard

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 2 of 17


III. Dimensional

Parameter/Equipment Range CMC2, 5 () Comments

Calipers3 –

Digital Up to 450 mm 8.2 µm Gauge blocks

Analog (Vernier) Up to 450 mm 16 µm

Analog (Dial) Up to 200 mm 8.2 µm


(200 to 300) mm 16 µm

Depth Gages3 –

Digital Up to 450 mm 5.8 µm Gauge blocks

Analog (Vernier) Up to 450 mm 12 µm

Analog (Dial) Up to 450 mm 5.8 µm

Height Gauges3 –

Digital Up to 450 mm 5.8 µm Gauge blocks

Analog (Vernier) Up to 450 mm 12 µm

Analog (Dial) Up to 450 mm 5.8 µm

Optical and Stereo Up to 400X 4.7 µm Stage micrometer


Microscopes, with Magnification
Reticle3

Ultrasonic Thickness Up to 100 mm 0.58 µm Gauge blocks


Gauges3

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 3 of 17


Parameter/Equipment Range CMC2, 5 () Comments

Outside, Inside and


Depth Micrometers,
Indicators, Thickness
Gauges3 –

Analog Up to 25 mm 1.2 µm Gauge blocks


(25 to 50) mm 1.2 µm
(50 to 75) mm 1.2 µm
(75 to 100) mm 1.2 µm
(100 to 125) mm 1.7 µm
(125 to 150) mm 1.7 µm
(150 to 175) mm 1.7 µm
(175 to 200) mm 1.7 µm
(200 to 225) mm 1.7 µm
(225 to 250) mm 1.7 µm
(250 to 275) mm 1.7 µm
(275 to 300) mm 1.7 µm
(300 to 325) mm 1.7 µm
(325 to 350) mm 1.7 µm
(350 to 375) mm 1.8 µm
(375 to 400) mm 1.8 µm
(400 to 425) mm 1.8 µm
(425 to 450) mm 1.8 µm

Digital Up to 25 mm 0.58 µm
(25 to 50) mm 0.60 µm
(50 to 75) mm 0.60 µm
(75 to 100) mm 0.61 µm
(100 to 125) mm 0.61 µm
(125 to 150) mm 0.62 µm
(150 to 175) mm 0.67 µm
(175 to 200) mm 0.70 µm
(200 to 225) mm 0.70 µm
(225 to 250) mm 0.70 µm
(250 to 275) mm 0.72 µm
(275 to 300) mm 0.76 µm
(300 to 325) mm 0.76 µm
(325 to 350) mm 0.80 µm
(350 to 375) mm 0.83 µm
(375 to 400) mm 0.83 µm
(400 to 425) mm 0.87 µm
(425 to 450) mm 0.87 µm

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 4 of 17


IV. Electrical – DC/Low Frequency

Parameter/Equipment Range CMC2, 5, 7 () Comments

DC Voltage – Generate (0.1 to 120) mV 0.0050 mV Transmille 3041A


(0.12 to 1.2) V 0.000 013 V
(1.2 to 12) V 0.000 12 V
(12 to 120) V 0.000 76 V
(120 to 1000) V 0.0083 V

DC Current – Generate (0.1 to 200) µA 0.0058 µA Transmille 3041A


(0.2 to 1.2) mA 0.000 17 mA
(1.2 to 12) mA 0.000 21 mA
(12 to 120) mA 0.0015 mA
(0.12 to 1.2) mA 0.098 mA
(1.2 to 12) A 0.000 22 A

Parameter/Range Frequency CMC2, 5, 7 () Comments

AC Voltage – Generate
(20 to 120) mV 56 Hz 0.050 mV Transmille 3041A
(20 to 120) mV 1 kHz 0.076 mV
(0.12 to 1.2) V 56 Hz 0.000 22 V
(0.12 to 1.2) V 1 kHz 0.000 34 V
(1.2 to 12) V 56 Hz 0.000 82 V
(1.2 to 12) V 1 kHz 0.0014 V
(12 to 120) V 56 Hz 0.017 V
(12 to 120) V 1 kHz 0.068 V
(120 to 1000) V 56 Hz 0.080 V
(120 to 1000) V 1 kHz 0.20 V

AC Current – Generate
(25 to 200) µA 40 Hz 0.43 µA Transmille 3041A
(25 to 200) µA 56 Hz 0.37 µA
(0.2 to 1.2) mA 40 Hz 0.96 µA
(0.2 to 1.2) mA 56 Hz 0.57 µA
(1.2 to 12) mA 40 Hz 0.0042 mA
(1.2 to 12) mA 56 Hz 0.0015 mA
(12 to 120) mA 40 Hz 0.036 mA
(12 to 120) mA 56 Hz 0.014 mA
(0.12 to 1.2) A 40 Hz 0.34 mA
(0.12 to 1.2) A 56 Hz 0.79 mA
(1.2 to 12) A 40 Hz 0.0038 A
(1.2 to 12) A 56 Hz 0.0017 A

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 5 of 17


Parameter/Range Frequency CMC2, 5, 7 () Comments

AC Current – Clamp On
Meters
Up to 100 A 56 Hz 0.41 A Transmille 3041A with
(>100 to 300) A 1.1 A EA015
(>300 to 500) A 1.7 A
(>500 to 700) A 2.3 A
(>700 to 1000) A 3.0 A

Parameter/Equipment Range CMC2, 5, 7 () Comments

Resistance – Generate 1Ω 0.0057 Ω Transmille 3041A


10 Ω 0.0064 Ω
100 Ω 0.0011 Ω
1 kΩ 0.000 11 kΩ
10 kΩ 0.0010 kΩ
100 kΩ 0.0094 kΩ
1 MΩ 0.000 20 MΩ
10 MΩ 0.0055 MΩ
100 MΩ 0.59 MΩ

Insulation Resistance Up to 1 MΩ 0.0024 MΩ Transmille 3041A


Meters –

Fixed Points 5 MΩ 0.013 MΩ Standard Resistance


10 MΩ 0.12 MΩ Box
50 MΩ 0.58 MΩ
100 MΩ 1.2 MΩ
500 MΩ 5.8 MΩ
1 GΩ 0.012 GΩ
5 GΩ 0.058 GΩ
10 GΩ 0.12 GΩ
15 GΩ 0.17 GΩ
20 GΩ 0.24 GΩ

Thermocouple Simulators –

Type J (-20 to 760) °C 0.76 °C Transmille 8080

Type K (-20 to 1000) °C 1.0 °C

Type R (0 to 1000) °C 0.41 °C

Type S (0 to 1000) °C 0.37 °C

Type T (-20 to 400) °C 0.82 °C

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 6 of 17


Parameter/Equipment Range CMC2, 5 () Comments

Electrical Simulation of (-20 to 1000) °C 0.37 °C Transmille 3041A


Temperature Indicators

V. Fluid Quantities

Parameter/Equipment Range CMC2, 5 () Comments

Volume – Laboratory
Glassware
Single Volume Pipettes 0.5 ml 0.000 34 ml Gravimetric method
1 ml 0.000 59 ml
2 ml 0.000 34 ml
5 ml 0.000 94 ml
10 ml 0.0011 ml
20 ml 0.0014 ml
25 ml 0.0014 ml
50 ml 0.0016 ml
100 ml 0.0026 ml

One-mark Volumetric 1 ml 0.0022 ml


Flasks 2 ml 0.0022 ml
5 ml 0.0022 ml
10 ml 0.0022 ml
20 ml 0.0037 ml
25 ml 0.0037 ml
50 ml 0.0055 ml
100 ml 0.0077 ml
200 ml 0.012 ml
250 ml 0.012 ml
500 ml 0.017 ml
1000 ml 0.098 ml

Graduated Pipettes Up to 1 ml 0.000 19 ml


(> 1 to 25) ml 0.0082 ml

Graduated Measuring Up to 5 ml 0.000 20 ml


Cylinders (> 5 to 10) ml 0.000 21 ml
(> 10 to 25) ml 0.000 28 ml
(> 25 to 50) ml 0.000 45 ml
(> 50 to 100) ml 0.000 45 ml
(> 100 to 250) ml 0.0047 ml
(> 250 to 500) ml 0.0051 ml
(> 500 to 1000) ml 0.095 ml

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 7 of 17


Parameter/Equipment Range CMC2, 5 () Comments

Volume – Laboratory
Glassware (cont)

Burettes Up to 10 ml 0.0082 ml Gravimetric method


(10 to 25) ml 0.020 ml
(25 to 50) ml 0.041 ml
(50 to 100) ml 0.082 ml

Pyknometers: Up to 10 ml 0.000 22 ml
Type 3 (Gay-Lussac), (> 10 to 25) ml 0.000 37 ml
Type 5 (Hubbard), (> 25 to 50) ml 0.000 55 ml
Type 6 (With (> 50 to 100) ml 0.000 77 ml
Thermometer Coupled)

Pyknometer Type 4 Up to 25 ml 0.000 45 ml


(Reischauer) (> 25 to 50) ml 0.0055 ml
(> 50 to 100) ml 0.0077 ml

Centrifuge Tube (6 in) Up to 25 ml 0.0081 ml


(> 25 to 50) ml 0.0084 ml
(> 50 to 100) ml 0.013 ml

Centrifuge Tube (8 in) Up to 10 ml 0.0081 ml


(> 10 to 25) ml 0.0084 ml
(> 25 to 100) ml 0.013 ml

Imhoff Cone Up to 100 ml 0.095 ml


(> 100 to 1000) ml 0.10 ml

Beaker Up to 10 ml 0.000 45 ml
(> 10 to 50) ml 0.0021 ml
(> 50 to 100) ml 0.0041 ml
(> 100 to 150) ml 0.0096 ml
(> 150 to 200) ml 0.012 ml
(> 200 to 250) ml 0.013 ml
(> 250 to 500) ml 0.042 ml
(> 500 to 1000) ml 0.10 ml

Water Trap Up to 25 ml 0.0081 ml

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 8 of 17


Parameter/Equipment Range CMC2, 5 () Comments

Piston Operated
Volumetric Apparatus–

Piston Pipettes (10 to 200) µl 0.052 µl Gravimetric method


(> 200 to 500) µl 0.19 µl
(> 500 to 2000) µl 0.19 µl
(> 2000 to 5000) µl 0.20 µl
(> 5000 to 10 000) µl 0.23 µl

Piston Burettes Up to 2 ml 0.000 19 ml


(> 2 to 5) ml 0.000 20 ml
(> 5 to 10) ml 0.000 23 ml
(> 10 to 25) ml 0.000 38 ml
(> 25 to 50) ml 0.000 68 ml

Dispensers Up to 0.5 ml 0.000 19 ml


(> 0.5 to 2) ml 0.000 19 ml
(> 2 to 5) ml 0.000 20 ml
(> 5 to 10) ml 0.000 23 ml
(> 10 to 25) ml 0.000 38 ml
(> 25 to 50) ml 0.000 68 ml
(> 50 to 100) ml 0.0013 ml

Viscometer3 –

Rotational Devices (5.711 to 621.1) cP 0.29 % of reading Viscosity standard


(621.1 to 2859) cP 0.38 % of reading solution
(2859 to 6149) cP 0.38 % of reading
(6149 to 26 600) cP 0.44 % of reading

Viscosity Cups3 –

Ford (2, 3, 4, 5) Cups 0.16 % Viscosity standard


solution
Zahn (1, 2, 3, 4, 5) Cups 0.16 %

Shell (1, 2, 3, 4, 5, 6) Cups 0.16 %

Stein Hall Stein Hall Cup 0.0058 s

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 9 of 17


Parameter/Equipment Range CMC2, 5 () Comments

Hydrometry:
Specific Gravity (Relative
Density) Hydrometers and
Equivalent Values in Other
Hydrometer Scales:
Density (g/ml, g/cm³,
kg/m³) (cont)

Baumé Heavy (0.0 to 30.0) °Bé Hy 0.021 °Bé Hy Cuckow method


Hydrometers (30.0 to 50.0) °Bé Hy 0.043 °Bé Hy
(50.0 to 70.0) °Bé Hy 0.025 °Bé Hy

API Hydrometers and (-1.0 to 11.0) °API 0.011 °API


Thermo-hydrometers (9.0 to 21.0) °API 0.012 °API
(API Scale for (19.0 to 31.0) °API 0.016 °API
Petroleum Products) (29.0 to 41.0) °API 0.015 °API
(39.0 to 51.0) °API 0.018 °API
(49.0 to 61.0) °API 0.021 °API
(59.0 to 71.0) °API 0.024 °API
(69.0 to 81.0) °API 0.029 °API
(79.0 to 91.0) °API 0.034 °API
(89.0 to 101.0) °API 0.041 °API

Lactometers (Quevenne (15 to 40) °Q 0.090 °Q


Scale)

Flow Rate – Liquids

Volumetric Up to 70 l/min 0.58 % Direct comparison


with Coriolis
reference standard

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 10 of 17


VI. Mechanical

Parameter/Equipment Range CMC2, 5 () Comments

Mass –

OIML Classes F1, F2 1 mg 0.020 mg Class E2 weight set


& M1 2 mg 0.020 mg
5 mg 0.020 mg
10 mg 0.025 mg
20 mg 0.03 mg
50 mg 0.04 mg
100 mg 0.05 mg
200 mg 0.06 mg
500 mg 0.08 mg
1g 0.030 mg
2g 0.040 mg
5g 0.050 mg
10 g 0.060 mg
20 g 0.080 mg
50 g 0.10 mg
100 g 0.16 mg
200 g 0.30 mg
500 g 2.5 mg
1 kg 5 mg
2 kg 10 mg

OIML Classes M2 & 5 kg 250 mg Class M1 weight set


M3 10 kg 500 mg
20 kg 1000 mg

Scales and Balances3 (0 to 220) g 0.020 mg Standard weights


(>220 to 1200) g 0.056 mg
(>1200 to 2500) g 0.068 mg
(>2500 to 6200) g 0.068 mg
(>6.2 to 30.1) kg 0.50 g
(>30.1 to 1000) kg 0.35 kg

Torque Wrenches and (0.013 to 813) Nꞏm 1.8 % of reading Torque transducers
Drivers

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 11 of 17


Parameter/Equipment Range CMC2, 5 () Comments

Pressure –

Pneumatic (Including (-10 to 0) psi 00.0068 psi + 0.062 % of Additel ADT681,


Pressure Modules and (-68.95 to 0) kPa reading ADT914 digital
Pressure Switch)3 pressure calibrator
(0 to 270) psi 0.0068 psi + 0.016 % of
(0 to 1861.58) kPa reading

(0 to 27) psi 0.058 psi + 0.22 % of


(0 to 186.16) kPa reading

(0 to 97) psi 0.29 psi + 0.30 % of


(0 to 668.79) kPa reading

(0 to 270) psi 0.46 psi + 0.30 % of


(0 to 1861.58) kPa reading

(0 to 1000) psi 0.061 psi + 0.0071 % of


(0 to 6894.76) kPa reading

(0 to 10000) psi 0.046 psi + 0.0058 % of


(0 to 68947.59) kPa reading

(-21 to 0) inHg 0.58 inHg + 0.22 % of


(-71.11 to 0) kPa reading
Additel ADT681
Pneumatic3 (-10 to 0) psi 0.0068 % series digital pressure
(-68.95 to 0) kPa & vacuum gages with
pumps
(0 to 270) psi 0.016 %
(0 to 1861.58) kPa

(0 to 27) psi 0.21 %


(0 to 186.16) kPa

(0 to 97) psi 0.30 %


(0 to 668.79) kPa

(0 to 270) psi 0.30 %


(0 to 1861.58) kPa

(-21 to 0) inHg 0.39 %


(-71.11 to 0) kPa

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 12 of 17


Parameter/Equipment Range CMC2, 5 () Comments

Centrifugues3 –

Rotational Speed (240 to 599.9) rpm 1.5 rpm Digital tachometer


(>600 to 3700) rpm 1.0 rpm
(>3700 to 7320) rpm 1.1 rpm
(>7320 to 9959) rpm 1.2 rpm
(>9959 to 9999) rpm 1.3 rpm
(>9999 to 11 321) rpm 1.5 rpm
(>11 321 to 13 809) rpm 1.6 rpm
(>13 809 to 16 055) rpm 1.7 rpm
(>16 055 to 18 146) rpm 1.8 rpm
(>18 146 to 20 000) rpm 1.9 rpm

VII. Optical Quantities

Parameter/Equipment Range CMC2, 5 () Comments

Spectrophotometers3 –

Absorbance (0.29 to 1.53) A 0.0022 A Neutral density


glass filters
Transmittance (2.979 to 12.407) % 0.046 % of Transmittance
(20.1 to 26.93) % 0.18 % of Transmittance
(42.7 to 51.3) % 0.19 % of Transmittance

Spectrophotometers
Filters –

Absorbance (0.29 to 1.53) A 0.0023 A Standard


spectrophotometer
Transmittance (3.0 to 12.4) % 0.089 % of Transmittance
(20.0 to 26.8) % 0.37 % of Transmittance
(42.6 to 51.1) % 0.40 % of Transmittance
Wavelength
(278 to 880) nm 0.1 nm

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 13 of 17


Parameter/Equipment Range CMC2, 5 () Comments

Spectrophotometer
Wavelength3, 4 278.77 nm 0.09 nm Holmium glass filter
287.35 nm 0.11 nm
333.75 nm 0.10 nm
347.55 nm 0.21 nm
360.58 nm 0.10 nm
381.53 nm 0.09 nm
385.83 nm 0.11 nm
418.62 nm 0.12 nm
445.70 nm 0.10 nm
453.48 nm 0.10 nm
460.03 nm 0.12 nm
473.70 nm 0.19 nm
484.17 nm 0.17 nm
536.48 nm 0.11 nm
637.53 nm 0.21 nm

431.40 nm 0.09 nm Didymium glass filter


440.18 nm 0.14 nm
472.40 nm 0.09 nm
513.27 nm 0.08 nm
529.03 nm 0.14 nm
573.22 nm 0.22 nm
585.30 nm 0.59 nm
629.17 nm 0.29 nm
684.47 nm 0.10 nm
740.47 nm 0.32 nm
748.50 nm 0.10 nm
807.20 nm 0.12 nm
879.50 nm 0.29 nm

VIII. Thermodynamics

Parameter/Equipment Range CMC2, 5 () Comments

Liquid and Dry Baths (Dry (-25 to 50) °C 0.029 °C Reference RTD
Wells) 3 (50 to 100) °C 0.033 °C thermometers,
(100 to 200) °C 0.053 °C thermistors,
(200 to 300) °C 0.11 °C thermocouples
(300 to 400) °C 0.16 °C
(400 to 600) °C 0.40 °C
(600 to 1100) °C 1.5 °C

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 14 of 17


Parameter/Equipment Range CMC2, 5 () Comments

Bi-metal Thermometers3 (-20 to 150) °C 0.59 °C Reference RTD


(150 to 250) °C 1.17°C thermometers and
(250 to 400) °C 2.9 °C temperature blocks
(400 to 550) °C 2.9 °C

Digital Thermometers3 (-20 to -5) °C 0.064 °C Reference RTD


(-5 to 25) °C 0.064 °C thermometers and
(25 to 50) °C 0.064 °C temperature blocks
(50 to 75) °C 0.064 °C
(75 to 100) °C 0.064 °C
(100 to 150) °C 0.097 °C
(150 to 200) °C 0.15 °C
(200 to 400) °C 0.29 °C
(400 to 600) °C 1.1 °C
(600 to 1100) °C 2.3 °C

Partial Immersion Liquid- (-20 to 0) °C 0.086 °C Reference RTD


In-Glass Thermometers (0 to 25) °C 0.086 °C thermometers and
(25 to 50) °C 0.086 °C temperature blocks
(50 to 70) °C 0.086 °C
(70 to 90) °C 0.086 °C
(90 to 110) °C 0.086 °C
(110 to 130) °C 0.10 °C
(130 to 150) °C 0.11 °C
(150 to 200) °C 0.16 °C
(200 to 250) °C 0.19 °C
(250 to 300) °C 0.21 °C

Infrared Temperature – (>-15 to 0) °C 0.26 °C Fluke 4180 black body


Measuring Equipment (>0 to 50) °C 0.63 °C Fluke 4181 black body
(>50 to 120) °C 1.2 °C
(>120 to 200) °C 1.3 °C
(>200 to 300) °C 1.4 °C
(>300 to 450) °C 2.4 °C
(>450 to 500) °C 3.4 °C

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 15 of 17


Parameter/Equipment Range CMC2, 5 () Comments

Thermo-Hygrometers,
Dataloggers and
Heat Stress WBGT
Meters*–

Humidity4 (10 to 25) % RH 1.8 % Reference thermo-


(25 to 45) % RH 2.2 % hygrometer and
(45 to 75) % RH 2.2 % climatic chamber
(75 to 90) % RH 2.7 %
*Including Wet bulb,
Temperature (0 to 10) °C 0.35 °C dry bulb, globe
(10 to 20) °C 0.30 °C temperature, and dew
(20 to 30) °C 0.67 °C point
(30 to 40) °C 0.67 °C
(40 to 50) °C 1.0 °C

IX. Time & Frequency

Parameter/Equipment Range CMC2, 5 () Comments

Stopwatches3 Up to 24 h 0.037 s/24h Reference stopwatch

Non-Contact (240 to 9000) RPM 1.0 RPM Transmille 3041A and


Tachometers (9001 to 9500) RPM 1.1 RPM work station
(9501 to 10 000) RPM 1.2 RPM
(10 001 to 20 000) RPM 1.3 RPM
_____________________________________________
1
This laboratory offers commercial calibration and field calibration service.
2
Calibration and Measurement Capability Uncertainty (CMC) is the smallest uncertainty of measurement
that a laboratory can achieve within its scope of accreditation when performing more or less routine
calibrations of nearly ideal measurement standards or nearly ideal measuring equipment. CMCs represent
expanded uncertainties expressed at approximately the 95 % level of confidence, usually using a coverage
factor of k = 2. The actual measurement uncertainty of a specific calibration performed by the laboratory
may be greater than the CMC due to the behavior of the customer’s device and to influences from the
circumstances of the specific calibration.

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 16 of 17


3
Field calibration service is available for this calibration and this laboratory meets A2LA R104 – General
Requirements: Accreditation of Field Testing and Field Calibration Laboratories for these calibrations.
Please note the actual measurement uncertainties achievable on a customer's site can normally be
expected to be larger than the CMC found on the A2LA Scope. Allowance must be made for aspects such
as the environment at the place of calibration and for other possible adverse effects such as those caused
by transportation of the calibration equipment. The usual allowance for the actual uncertainty introduced
by the item being calibrated, (e.g. resolution) must also be considered and this, on its own, could result
in the actual measurement uncertainty achievable on a customer’s site being larger than the CMC.
4
CMC’s for intermediate values are calculated using linear interpolation.
5 The contributions from the “best existing device” are not included in the CMC claim.
6
This scope meets A2LA’s P112 Flexible Scope Policy.
7
The type of instrument or material being calibrated is defined by the parameter. This indicates the
laboratory is capable of calibrating instruments that measure or generate the values in the ranges indicated
for the listed measurement parameter

(A2LA Cert. No. 4286.04) Revised 10/16/2019 Page 17 of 17


Accredited Laboratory
A2LA has accredited

ELICROM PERU S.A.C.


Lima, PERU
for technical competence in the field of

Calibration
This laboratory is accredited in accordance with the recognized International Standard ISO/IEC 17025:2017
General requirements for the competence of testing and calibration laboratories. This laboratory also meets R205 – Specific
Requirements: Calibration Laboratory Accreditation Program. This accreditation demonstrates technical competence for a
defined scope and the operation of a laboratory quality management system
(refer to joint ISO-ILAC-IAF Communiqué dated April 2017).

Presented this 10th day of October 2018.

_______________________
Vice President, Accreditation Services
For the Accreditation Council
Certificate Number 4286.04
Valid to August 31, 2021
Revised October 16, 2019

For the calibrations to which this accreditation applies, please refer to the laboratory’s Calibration Scope of Accreditation.