Operation, Fundamentals, and Analysis: Bruker Multi-Mode 8 Atomic Force Microscope

This document provides instructions for operating a Bruker Multi-mode 8 Atomic Force Microscope. It describes the key components of the AFM, guidelines for sample preparation including solid state and solution state imaging, and standard operating procedures for setup and imaging. Sample preparation methods like dropcasting, spincoating and direct deposition are outlined. Substrate preparation for different materials is also covered. Detailed steps are provided for powering on the system, selecting imaging modes, and loading samples and probes.

Uploaded by

syam00
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
167 views12 pages

Operation, Fundamentals, and Analysis: Bruker Multi-Mode 8 Atomic Force Microscope

This document provides instructions for operating a Bruker Multi-mode 8 Atomic Force Microscope. It describes the key components of the AFM, guidelines for sample preparation including solid state and solution state imaging, and standard operating procedures for setup and imaging. Sample preparation methods like dropcasting, spincoating and direct deposition are outlined. Substrate preparation for different materials is also covered. Detailed steps are provided for powering on the system, selecting imaging modes, and loading samples and probes.

Uploaded by

syam00
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 12

Bruker Multi-mode 8 Atomic Force Microscope

Operation, Fundamentals, and Analysis


Rachel A. Letteri, April 2018

Bruker Multi-mode 8 AFM

Figure 1. (a) Bruker Multimode 8 AFM: Zoomed in images of the (b) front and (c) back of the AFM head,
(d) z-scanner motor switch (up and down directions indicate tip movements relative to the sample); (e)
labeled transparent image of AFM head components; (f) names of AFM components labeled in green.
Images (c) and (e) were taken and/or adapted from the Bruker AFM Multimode 8 user guide. 1

Sample preparation
• Note that sample preparation will vary based on sample and substrate composition, and
many other factors. General guidelines are provided below, but for best results, optimize
concentrations, substrates, and sample preparation procedures.
• For solid state measurements: Substrates must be first adhered to magnetic pucks, using
sticky tabs (similar to double-sided tape) or carbon tape (for a stronger bond). Next, either
(1) dropcast, (2) spincoat, or (3) dispense liquid solution on a substrate for 2-3 minutes to
enable adsorption and wick away excess solvent.
o Dropcast: While dropcasting and allowing samples to dry completely can produce
undesired drying patterns, it is often the preferred method for materials that do not
adhere easily to substrates (e.g., by electrostatic interactions). For neutral
materials, e.g., polyethylene glycol, this is an especially valuable sample
preparation method.
▪ Prepare a 0.5-1 mg/mL solution of your polymer/nanostructure. Dropcast
2-4 μL of this solution and allow to dry.
o Spincoat: This method can also be useful for coating substrates with materials that
don’t easily adhere. Prepare a 0.1-0.5 mg/mL solution of your
polymer/nanostructure.

Bruker Multimode 8 Atomic Force Microscope 1


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
▪ On our spin-coater, this program is already in place, but if not, write one
with a similar procedure: step 1 – 500 RPM for 10 s; step 2 – 3000 RPM
for 2 min.
▪ Place substrate-containing magnetic puck on the spincoater (vacuum will
hold it in place).
▪ Draw 20 μL of your solution in a pipette
▪ Start the program, and during the first slow spinning step, add your solution
onto the spinning substrate. This will encourage even distribution of the
materials.
▪ Dry substrates under a stream of N2(g) to remove any residual liquid.
▪ Adjust sample concentration, or repeat depositions as needed.
o Dispense solution, then wick away excess solution
▪ For nanostructures that exhibit attractive interactions with the surface (e.g.,
electrostatic), this method is particularly useful, as it prevents any adverse
drying features
▪ Dropcast a 0.1-0.5 mg/mL solution such that it covers the entire substrate.
▪ Allow to stand 1-3 min
▪ Tilt substrate onto filter paper to wick away excess solution, then dry
substrate under a stream of N2(g)
• Solution-state imaging
o Nanostructures must adhere to substrate, either via covalent interactions or
attractive non-covalent interactions (e.g., electrostatics). As such, often sample
preparation means tailoring the substrates to facilitate adsorption to
nanostructures of interest.
o Static solution imaging: prepare a 0.1-0.5 mg/mL solution of
polymer/nanostructures. Place the substrate onto the magnetic sample holder,
align tip, remove the tip assembly, add a drop or two onto the substrate, then
replace tip holder assembly
o Dynamic imaging under flow: Prepare a very dilute solution of nanostructures
(since we’re adding more volume – maybe 10-100x more dilute relative to static
imaging), and pull into a syringe
• Substrate preparation
o Mica: This is an anionic, layered substrate. After adhering to magnetic pucks, mica
can be reused by cleaving off the top layer. Apply Scotch tape, then pull away. A
mica disc should be visible on the piece of tape. Repeat until 2 almost whole discs
are cleaved, and appear on the tape. This is a great substrate for solution-state
imaging of cationic nanostructures.
o Glass, silicon wafers: Anionic, non-layered/non-reusable substrates. Clean with
soap and water, then wash with acetone and water to remove oily and water-
soluble contaminents, respectively. Dry under N2(g).
o Coating substrates to changes surface chemistry: Spincoat a 0.1 mg/mL solution
of polylysine, or coat substrates with a silane (e.g., 3-aminopropyltriethoxysilane)
to render the surface cationic. Consult the literature for best procedures, since
both methods can generate aggregates (polylysine can aggregate upon drying to
produce 2-4 nm aggregates, and water-induced cross-linking of silanes also
produces aggregates). As such, be sure to image coated substrates to determine
surface roughness prior to imaging nanostructures.

Bruker Multimode 8 Atomic Force Microscope 2


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
SOP
The tip holder should always be secured in the
AFM head when not in use. Please do not
store on the counter.
• Power on controller (back left of box,
as shown in Figure 2)
• Power on TV and light, turning the dial
up almost all the way to the right to get
sufficient light intensity to allow
visualization of the surface and tip
• Power on computer and monitor
o Note that both the computer
and monitor should be powered off when not in use
o Username: nanoscope, Figure 2. Power-on procedures: Photos of the (a) front
password: none (just press and (b) back of the AFM controller, with the power
enter) button indicated in green; (c) photo of the TV showing
• Click on Nanoscope 8.15 an optical microscopy image of a cantilever; (d) light
o Note that the Nanoscope 8.15, intensity adjustment on light source for the optical
Nanoscope Analysis, and microscope. In the manual,1 the optical microscope
Multimode manual shortcuts video set-up is abbreviated OMV.
are located at the center of the
desktop
• Select experiment category, e.g. ‘ScanAsyst’ (or Tapping Mode, Contact Mode,
Quantitative Analysis, etc.), as shown in Figure 3
o Select experiment group, e.g., Scanasyst in air (or in fluid)
o Select experiment, e.g., Scanasyst in air (or in fluid)

Figure 3. Experiment selection: Screenshots of experiment selection for (a) Scanasyst, (b)
tapping, and (c) contact modes performed in air. For solution-state imaging, select the
appropriate experiment group and experiment.

Bruker Multimode 8 Atomic Force Microscope 3


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
• Load sample and probe:
o Remove tip/probe holder assembly (part 8) from instrument by unscrewing knob
(part 15) on back-center of AFM head, and place on a table with the tip facing up
(Figure 4).
o Place sample (adhered to magnetic puck) in the center the magnetic platform, and
push the lever at the base of the instrument (Figure 1d) in the up direction to move
the sample sufficiently far down (tip up) so that the probe will not crash into the
sample when the probe holder is placed in
the AFM head
▪ The ‘up’ and ‘down’ labels on the
switch in Figure 1d indicate the
direction of tip motion relative to the
sample, but the sample moves (i.e.,
the up direction moves the stage
down such that it appears that the tip
is moving upwards away from the Figure 4. Tip holder assemblies for (a)
stage, and vice versa) solid-state and (b) solution-state imaging.
o For loading probes onto either solid or
solution-state tip holder assemblies, push down gently on probe holder to open
clip. Use forceps to carefully place tip on platform and then close clip
▪ Note: it is strongly advised to practice placing broken tips into the probe
holder ca. 2-3 times before handling good tips – even for experienced users
(especially if you’ve had a lot of coffee!!)
▪ For solution-state imaging, it’s important that the metal clip secures the tip
– it falls out easily if the clip is not perpendicular to the long axis of the
cantilever. Test the security of the tip by inverting the tip holder after
installing a broken tip.
o Insert probe holder with tip facing down into scope such that pins fit on the
protruding pins of the AFM head (part 2), as shown in Figure 1b. Secure probe
holder by tightening knob in the back-center of the AFM head (part 15) until the
clip contacts the tip holder assembly
▪ Do not overtighten, as long as the pins touch the probe holder, it will be
secure
o Solution-state imaging note: It’s easiest to align the laser in air before adding
solution.
• Align the laser:
o Focus the camera on the tip by using the large black knobs (part 1), as shown in
Figure 2c
▪ If the cantilever is not visible, move the AFM relative to the camera into the
illuminated light circle using the micrometer knobs at the bottom of the
instrument (part 6) or move the AFM head assembly (and tip) relative to
the sample and camera using the knobs on the AFM head (part 11)
o Look for the red laser spot, and if needed, move the laser close to the tip in the x
and y directions using the knobs on the top of the AFM head (part 12)
o Alignment method 1: Raise camera using the focus knobs (part 1), and holding
the AFM head carefully, unhook the springs (part 3), and hold head such that the
laser spot is aimed on a sheet of paper (or your hand)
▪ As shown in Figure 5, align laser such that the laser is on the tip using the
two laser dials (one on right of head to move laser in x direction and one
on the back to move in y direction)

Bruker Multimode 8 Atomic Force Microscope 4


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
▪ Replace head in microscope. Look at the sum reading (want this to be
about 6)

Figure 5. Tip alignment procedure, using method 1. In method 2, the same


procedure can be used, except that the laser spot will be visible on the paper
covering the detector when it’s hitting the cantilever (whereas in method 1, the
laser is not visible when it’s hitting the
cantilever).

o Alignment method 2: Slide a piece of paper in


front of the detector such that when the laser
bounces off the cantilever, it will be visible on
the sheet of paper – as shown in Figure 6.
▪ Move the laser in x and y to center it on
the back of the cantilever closest to the
tip following a similar procedure as
shown in Figure 5, however in this
method, you will see a laser spot when Figure 6. Photo of laser alignment
using method 2
it’s hitting the cantilever, unlike in
method 1 when the laser is hidden from
the paper directly below it when it’s reflected off the cantilever
▪ Ensure that once the spot is centered and hitting the piece of paper, that it
is a round/Gaussian spot rather than elliptical or distorted (otherwise make
small x-y adjustments using the knobs, part 12).
• This last step allowing us to ensure good laser spot shape
represents an advantage over method 1 (in addition to not having
to remove the AFM head).
o Note that, if the laser is aligned and the sum (looks like a speedometer on the
digital display, Figure 7) is still below
4-5, try adjusting the lever on back of
the AFM head (part 14) to tilt the
mirror, and thereby redirect laser
toward the detector, to maximize the
sum
• Align the detector: Move the detector such
that the laser spot hits the center of the
detector by adjusting the knobs on the left/top
of the AFM head (part 13) one at a time. This
will change the numbers in the displays in
Figure 7. Adjust the position of the detector
until one of the display values approaches 0,
then adjust the other knob so that both
Figure 7. Photo of the Multimode 8 displays, with
numbers go to 0 (+/-0.1-0.3 is preferable the sum indicated by an arrow.
since these numbers will change during
imaging, but +/- 1 is fine)

Bruker Multimode 8 Atomic Force Microscope 5


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
o If the numbers don’t change immediately, it’s totally fine – it just means the detector
center is still far away from the laser spot. Turn one of the knobs all the way in
one direction, then all the way in the other. At some point, a change should occur.
If the detector is at the edge of its movement range and still needs to move more
to bring the display numbers (indicative of the displacement of the laser spot from
the center of the detector) to 0, adjust the mirror using the lever on the back of the
AFM head (part 14) to direct the laser to a different spot.
• Focus tip just above surface:
o First, focus far above the surface turning knobs (part 1), then slowly focus
downwards. You will first pass through the tip, then the surface, and finally the tip
reflection (which looks just like the actual tip, so you’ll want to be sure you know
you’re focused above the surface on the tip before moving the tip).
o Focus on the surface, then move the focus just above the surface (such that the
surface looks blurry)
o Use the ‘tip down’ switch (Figure 1d) to move the tip slowly towards the surface
until it’s in focus. Since the focus is just above the surface, when the tip comes
into focus, it too will be just above the surface.
o Once the tip is in place just above surface, turn off light (Figure 2d) to avoid
interference with the laser light traveling to the detector
o Solution-state imaging note: As the tip is immersed in solution, the alignment/sum
will decrease due to the different bending of light/refraction across the air/water
interface. Move only the mirror (part 14) to direct the laser back toward the
detector.
• You’re ready to image – see the sections below for procedures for imaging in Scanasyst,
contact, and tapping modes.

Imaging in PeakForce Tapping/ScanAsyst mode (most common)


PeakForce Tapping mode combines the benefits of contact and tapping mode, and adds
tremendously beneficial automation, allowing users to obtain nice images irrespective of
experience level. At each point along a line, the cantilever is moved relative to the sample until
the ‘peak force’ is reached, and then withdrawn. As such, force-deflection curves are obtained at
each point, enabling mechanical mapping in a less destructive way than contact mode.

• Specify imaging parameters in the


software, as shown in Figure 8. Good
initial parameters are:
o Scan size = 1-5 μm is
o Scan speed = 1 Hz
o Aspect ratio = 1
o Samples/per line = 512
▪ While 512 points/line is
best for publication, 256 is
a great way to survey a
larger area in half the time
o The red button expands
parameters, but this shouldn’t be
necessary in routine imaging Figure 8. ScanAsyst parameters. Adjust those in the
• Engage (this moves the tip into contact green box if desired and make sure ScanAsyst Auto
with the surface) Control is on; the others are adjusted automatically.

Bruker Multimode 8 Atomic Force Microscope 6


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
• The height and peak force error data are collected automatically
o If you’d like both the trace and retrace data from each line, you’ll have to collect
this in separate channels (just open and select new channels)
• Withdraw the tip when imaging is complete, click several times

Contact Mode Imaging


In contact mode, the cantilever is pressed into the sample until it deflects to a specified set point.
While this is the most traditional imaging mode and allows acquisition of mechanical information
(i.e., force-distance curves), the constant contact with the sample can be destructive for
soft/polymeric samples.

• Set initial scan parameters


o Scan rate = 1 Hz
o Scan size = 100 nm
o Scan angle = 0 º
o Slow scan axis = enabled
o Integral gain = 2; proportional gain = 5
o Deflection setpoint = 0 V
o Units = metric
o Z limit, deflection limit = max (enter really large number and the software will
convert it back to an acceptable value
• Engage
• Increase deflection setpoint by 0.5 V at a time until you see an image, and the trace and
retrace line/height profiles match well
o This is telling the tip to press harder into the sample such that it deflects more
o Ideally, you have the lowest deflection setpoint possible while still following the
sample to avoid wear on the tip.
• Height and deflection error will be collected automatically in channels 1, 2. If not, select
these. Note that it doesn’t matter whether you collect trace or retrace data (i.e., that
collected as each line of the sample is scanned and retraced)
• Withdraw when imaging is complete

Tapping Mode Imaging


Tapping mode minimizes sample damage by contacting the sample only periodically, while
providing additional chemical information about the surface. The tip is oscillated near its
resonance frequency, and then brought near the surface. As the tip approaches the surface, the
amplitude/oscillation is dampened (slowed) by enhanced sample-tip interactions. In tapping
mode, an amplitude setpoint is specified, and the sample moved up and down to maintain a
constant amplitude, enabling height and ‘phase’ imaging. Phase imaging provides insight into
tip-surface interactions, and is especially useful for block polymer films. Even if the films have a
constant height, the morphology can be imaged by contrast between tip-sample interactions in
different polymer domains.

• For tapping mode, it’s suggested to use a silicon probe


• Set scan parameters – the following are good starting points:
o Scan size =1 μm
o X, Y offsets = 0 nm
o Scan angle = 0º
o Scan rate = 1 Hz
o Samples/line = 256-512
Bruker Multimode 8 Atomic Force Microscope 7
Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
o Slow scan axis = enabled
o Integral gain = 4
o Proportional gain = 8 (note that but gains for tapping mode will be ca. 10% of those
used in contact mode imaging)
o Amplitude set point = leave as is (this will be set by the software as the tip engages
the surface)
o Drive frequency, drive amplitude = leave as is (this will be set during cantilever
tuning)
o Limits: set to max by entering large numbers and allowing software to make the
correction to the maximum possible value
• Tune the cantilever
This step measures the resonance frequency of the cantilever and adjusts the piezo
voltage to oscillate the cantilever such that it vibrates at a frequency near resonance to
ensure good amplitude. The tuning process involves application of a range of frequencies
to cantilever, and the amplitude measured to determine the resonance frequency. The
amplitude vs. frequency curve should be symmetrical and Gaussian, though some
deviation is okay. If the peak shape looks bad/multimodal, change the probe.
o Click on the tuning fork icon or select microscope -> cantilever tune
o In the tune window (Figure 9), set mode = main
o Start and end frequencies should be set so they are +/-20-50 Hz of the listed
resonance frequency f0 (found on the probe box)
o Click auto tune
o After tuning is complete, click center peak and zero phase.

Figure 9. Cantilever tune window.

• Reset gains: integral gain = 0.5; proportional gain = 5


• Height, amplitude error, and phase should be automatically collected, but if not, add these
channels. The line direction can either be trace or retrace.

Bruker Multimode 8 Atomic Force Microscope 8


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
• Engage
• Adjust the scan rate, gains, and amplitude set point to bring the trace and retrace lines as
close together as possible.
o Sometimes increasing drive amplitude can be helpful

Solution-state imaging (under flow)


• Align probe in air
• Wet a silicone thin ring and place in ring around tip platform
(widest end should fit into the ring, and the narrow end of the
ring extend upward. The water helps to hold the ring in place,
even when the tip holder assembly is inverted.
• Bring the tip into focus just above the surface (as described
above) – this helps to create a good seal between the
substrate and the silicone ring prior to introduction of fluids.
• Prepare a very dilute solution of nanostructures (since we’re
adding more volume – maybe 10-100x more dilute relative to
static imaging).
• Pull sample into a syringe and remove as much air as
possible, then introduce using the inlets and outlets shown in
Figure 10. Figure 10. Solution-state
• Add a vial to collect solution flowing out of the cell. imaging under flow
• Very slowly introduce solution into cell – rapid flow may apply
shear sufficient to break the tip.
• If there are no leaks, solution should flow out of the outlet as you introduce solution.
o If leaks occur, remove and re-install silicone ring.

Image capture
• Select capture directory by clicking the file button next to the text box for file name entry
(top right of software window) and selecting the folder in which you’d like your data to be
saved
• Name file, and press enter (to save the name)
• Click capture (or continuous capture, which will keep capturing images as the sample is
scanned)
o The image will be captured once the scan (from top to bottom, or bottom to top) is
complete – though you can always force capture by clicking capture again
o If you wish to start a scan from the bottom click the button containing an up arrow
(middle top of software window)
o To start a scan from the top, click the button containing a down arrow
- The zoom button at the bottom of the height channel image allows you to draw a box, and
then upon clicking execute, an image will be acquired in the place specified by the
rectangle (or close to it – the zooming is not always exact)
- Click offset, then place the cursor at a point on the image. Upon clicking execute, a new
image with the same dimensions will be acquired, this time centered on the point indicated.
- When done imaging, click withdraw several times to lift tip off of surface in a controlled
manner to avoid breaking the tip, then use switch on the instrument to move tip faster
away from surface

Image analysis
• Open Nanoscope analysis, and then open the image file you wish to analyze

Bruker Multimode 8 Atomic Force Microscope 9


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
• Flatten to correct systematic errors/features, such as substrate tilt, which can often be
parabolic. Click the flatten icon, boxed in Figure 11. 2nd order flattening is a good starting
point. Select no thresholding, the whole image. Click execute.

Figure 11. Flatten image to remove systematic variation in substrate curvature, etc. Click the
boxed flatten icon to access the window shown above, then once inputs are as desired, click
execute.

• Set data scale by right clicking on the image color bar, and click color scale
o In the window that appears, click the modify data scale tab and set the max and
min of each image. When comparing multiple samples, it’s best that this the data
scale be kept constant
• Save image as a jpg, by right clicking on the image, select export, then screen display
• Section analysis: Use this feature to obtain line profiles (e.g., height vs. x-y dimension).
o As shown in Figure 12, click the section analysis icon and draw a line to generate
a profile, which can be exported as txt files and plotted in origin. It’s a good idea
to also save an image with the section lines drawn, so you know at which point on
the image you took the line profile.

Figure 12. Section analysis, including instructions on how to bring up the window, draw a line to
generate a line profile, and export x-z data for plotting in Origin or excel.

Bruker Multimode 8 Atomic Force Microscope 10


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
Nanomechanical measurements
(based on training by Cody Chalker, April 2018)

Since contact and PeakForce tapping modes generate


force vs. displacement curves, these can be used to
extract mechanical information about surfaces. In
PeakForce mode, where force-deflection curves are
acquired at each point, mechanical mapping is
possible. However, since determination of mechanical
properties (e.g., modulus) is highly dependent on probe
geometry, probe spring constant, and sample/probe
composition (as well as contamination), it’s crucial that
users exercise diligence in ensuring these
measurements are conducted properly and reported
data are representative. As such, mechanical
properties should always be listed as a range – never
as a particular value.

• Adjust the load on the table by moving the load


adjust such that the vertical position indicator
spot is at the line (not too high or low), as shown
in Figure 13a, to minimize noise. If further
noise reduction is desired, the AFM can be
placed on a tripod, as shown in Figure 13b.
• Calibrate deflection sensitivity.
o Use a hard sample (e.g., sapphire or
silicon) Figure 13. Minimize noise by leveling
o Check noise: In the quantitative the vibration-reducing table (a) or
nanomechanical analysis mode, set scan placing the AFM on a tripod (b).
size = 0 and click engage.
o Activate ramp mode, and click start single ramp (this will acquire a force-deflection
curve in a single point).
o On the resulting force distance curve, place the cursors at 2 points in the linear
range closest to the surface (highest z value), and the software will calculate the
slope.
o This will bring up update deflection sensitivity window and it will ask you if it’s OK.
If you click OK, this will update the deflection sensitivity in the calibration window.
o Repeat this test to make sure you obtain similar values.
o Note that it’s actually best to put silicon and sample on same tip, so as not to move
tip after this deflection sensitivity measurement.
• Calibrate the probe spring constant
o Spring constant values listed on the probe boxes are nominal, and vary from batch
to batch, so calibration is necessary. To learn more, see the Veeco application
note on spring constant calibration2 and the papers by Sader and coworkers.3-4
o Perform a thermal tune
▪ Note that this measures the vibration as a function of frequency without
excitation (i.e., without shaking the cantilever with a piezo).
▪ Fit the power spectra density curve, or power as a function of frequency to
the simple harmonic oscillator equation. – damping is used for tapping
mode since the oscillations are near the surface. Fit by adding the 2

Bruker Multimode 8 Atomic Force Microscope 11


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018
cursors on either side of the peak (just like a GPC peak), entering the
temperature, and clicking fit. This will calculate the spring constant
• Calibrate the tip radius
o Values listed on the probe boxes are nominal, and vary from batch to batch, so
calibration is necessary. There are several methods for measuring probe radius.
o TEM is a great way to calibrate tip radius, especially for small tips, however
requires special probe holders.
o Tip reconstruction
▪ Input nominal tip radius
▪ Engage nioprobe sample/standard using a scan area of 200 nm. Acquire
several images, then calculate probe radius using algorithms in SPIP or
Gwiddion to do ‘blind tip reconstruction’.
o Consider larger (30-50 nm) or beaded tips for force-distance curve acquisition in a
more reliable way, and be sure to check how the tip radius and mechanical
properties change over time.
• Notes
o References and advice on calibration for mechanical measurements.2-5
o For best results, clean AFM tips with UV ozone and perform mechanical
measurements in argon or nitrogen atmospheres to minimize capillary forces
▪ Consider storing tips and standard samples in a controlled humidity
chamber or in a glove bag
o Purchase new PS/LDPE QNM standards and use for ensuring measurements are
conducted properly
o JKR modulus models are best to use for mechanical measurements, rather than
DMT modulus – all improvements upon Hertzian mechanical models

Recent review on AFM of polymer samples6

References
1. Veeco Instruments, I., Multimode8 with ScanAsyst - User Guide and Instruction Manual. 2010.
2. Ohler, B., Practical Advice on the Determination of Cantilever Spring Constants. Veeco
Instruments, Inc. Application Note 2007.
3. Sader, J. E.; Chon, J. W. M.; Mulvaney, P., Calibration of rectangular atomic force microscope
cantilevers. Review of Scientific Instruments 1999, 70 (10), 3967-3969.
4. Sader, J. E.; Sanelli, J. A.; Adamson, B. D.; Monty, J. P.; Wei, X.; Crawford, S. A.; Friend, J.
R.; Marusic, I.; Mulvaney, P.; Bieske, E. J., Spring constant calibration of atomic force
microscope cantilevers of arbitrary shape. Rev Sci Instrum 2012, 83 (10), 103705.
5. te Riet, J.; Katan, A. J.; Rankl, C.; Stahl, S. W.; van Buul, A. M.; Phang, I. Y.; Gomez-Casado,
A.; Schon, P.; Gerritsen, J. W.; Cambi, A.; Rowan, A. E.; Vancso, G. J.; Jonkheijm, P.;
Huskens, J.; Oosterkamp, T. H.; Gaub, H.; Hinterdorfer, P.; Figdor, C. G.; Speller, S.,
Interlaboratory round robin on cantilever calibration for AFM force spectroscopy.
Ultramicroscopy 2011, 111 (12), 1659-69.
6. Wang, D.; Russell, T. P., Advances in Atomic Force Microscopy for Probing Polymer Structure
and Properties. Macromolecules 2017, 51 (1), 3-24.

Bruker Multimode 8 Atomic Force Microscope 12


Operation, Fundamentals, and Analysis
Rachel A. Letteri, April 2018