LC Ms Training Protocol 2
LC Ms Training Protocol 2
Wendy Gavin
Version 2
July 2017
1
Table of Contents
1. References…………………………………………………………………………………………………………………. page 3
2. LC/MS Instrument……………………………………………………………………………………………………… page 4
3. Basics of LC/MS………………………………………………………………………………………………………….. page 6
4. Atmospheric Pressure Ionization……………………………………………………………………………….. page 7
a. Electrospray Ionization………………………………………………………………………………….. page 7
b. Atmospheric Chemical Pressure Ionization……………………………………………………. page 9
c. DUIS……………………………………………………………………………………………………………… page 9
5. Triple Quadrupole……………………………………………………………………………………………………… page 11
a. Tandem MS/MS……………………………………………………………………………………………. page 14
b. Isotopes……………………………………………………………………………………………………….. page 16
6. Common ESI adducts…………………………………………………………………………………………………. page 17
7. Liquid Chromatography…………………………………………………………………………………………….. page 19
8. Analysis Modes…………………………………………………………………………………………………………. page 21
9. Sample Prep/ Important Facts ………………………………………………………………………………….. page 26
2
1. References
LC-MS Training Manual pdf.
Steps for LC-MS pdf.
LCMSMS_OperatorsGuide.pdf.
EEM-07-002-A Level 1 Safety Training.pdf
RES-07-001-ETIC Emergency evacuation.pdf
3
2. LC/MS Instrument-Shimadzu LC/MS 8040
A. LC Unit
B. Pumps –LC-20AD
Pumps
4 LC-20AD
C. Degasser Unit
Degasser
D. Auto-Sampler
SIL-20AC
Cools samples
5
3. Sample Preparation
Dissolve sample in acetonitrile, methanol or water. One can use a mixture of these
solvents. Do NOT use THF, DMF, DMSO.
Filter through 0.22 micron filter
0.1% v/v, thus 1ml of acid in 1 liter of solvent
4. Basics of LC/MS
Liquid Chromatography Mass Spectrometry (LC/MS) is an analytical technique that combines the
separation capabilities of liquid chromatography with the mass analysis capabilities of mass
spectrometry. LC/MS is oriented towards specific detection and potential identification of chemicals in
the presence of other chemicals (mixture).
Liquid Chromatography is a fundamental technique used to separate a sample into its individual
components. The separation occurs based on the interactions of the sample with the mobile phase and
stationary phase. Liquid chromatography separates the components of the sample based on differences
in their affinity for the stationary phase or mobile phase. LC can separate of wide range of organic
compounds from small-molecule metabolites to peptides and proteins.
Traditional detectors for LC include: refractive index, electrochemical, fluorescence, and ultraviolet-
visible (UV-Vis). Such detectors quantify substances based on retention time and quantitate substances
based on peak intensity and peak area. Chromatography offers great resolution, but quantifying and
quantitating substances accurately can be difficult, especially when components elute at the same time.
Mass Spectrometry (MS) offers highly sensitive detection technique that ionizes the sample components
using various methods. The resulting ions are separated in vacuum based on their mass-to-charge ratio
and the intensity of each ion is measured. MS detectors generate valuable information about molecular
weight, structure, identity, quantity, and purity of the sample. MS detectors increase confidence of
qualitative and quantitative analyses.
MS is a more sensitive and more specific detector than all other LC detectors. It can analyze compounds
without a suitable chromophore. It can identify compounds in unresolved chromatography peaks,
reducing the need for perfect chromatography.
LCMS unit consists of a unit for separating components (HPLC), an ion source that ionizes the sample, an
electrostatic lens that introduces the generated ions, a mass analyzer that separates ions based on their
mass-to-charge ratio, and a detector unit that detects the separated ions.
6
5. API-Atmospheric Pressure Ionization
The atmospheric pressure ionization unit ionizes the sample sent from the LC under atmospheric
conditions.
Ion sources ionize the analyte molecules and separate the resulting ions from the mobile phase.
Atmospheric pressure ionization (API) ionizes analyte molecules first, at atmospheric pressure. These
analyte ions are mechanically and electrostatically separated from neutral molecules. Ions generated
are stripped of solvent, focused into a beam, then delivered to the quadrupole. Electrospray ionization
(ESI) and atmospheric pressure chemical ionization (APCI) are two common API techniques. ESI is best
suited for ionic compounds of high polarity, where APCI is good for low or medium polarity compounds.
A. ESI
The ESI source operates at atmospheric pressure. The sample solution is sprayed from a small tube into
a strong electric field in the presence of nitrogen to assist desolvation. The liquid aerosolizes as it exits
the capillary at atmospheric pressure, the desolvating droplets shedding ions that flow into the mass
spectrometer, induced by the effects of electrostatic attraction and vacuum. ESI can be used for a wide
range of samples such as natural products, proteins, peptides, biological macromolecules and
pharmaceuticals.
The principle of ionization for an ESI probe is that the sample solution is drawn into a capillary tube with
a high voltage of around 3 to 5 Kv applied to it. Nebulizer gas is blown out around the outside of the
capillary tube, spraying the solution and generating fine droplets that are electrostatically charged with
the same sign as the applied voltage. During the course of movement, the charged droplets are subject
to vaporization of the solvent and they disintegrate when the repulsive force among the charges
exceeds the surface tension of the liquid. Through repetition of vaporization and disintegration, very
fine droplets are achieved, and ultimately it is thought that sample ions are released in the vapor phase.
7
The ionization process does not fragment the sample into smaller charged particles. Instead, it turns the
sample into smaller droplets which in turn will be further desolvated into even smaller droplets. This
creates molecules with attached protons. These protonated and desolvated molecular ions will pass
through the mass analyzer to the detector where the mass of the sample can be determined.
ESI allows for the production of multiply charged species. This results in the ability to analyze very high
molecular weight species. One sees a distribution of ions for peptides and proteins.
ESI
Proteins have many suitable sites for protonation as all of the backbone amide nitrogen atoms could be
protonated theoretically, as well as certain amino acid side chains such as lysine and arginine which
contain primary amine functionalities. ESI can ionize peptides and proteins since they can take on
multiple charges. Since the mass spectrometer measures m/z, this brings the apparent mass down to
500 to 2000m/z.
8
B. APCI
APCI ionizes sample and solvent molecules by spraying the sample solution into a heater using nitrogen.
Solvent molecules are ionized by a corona discharge to generate stable reaction ions. Ions are produced
in the discharge and extracted to the mass spectrometer. Protons are transferred between these
reaction ions and sample molecules by either adding or removing a proton. APCI is used for analyzing
highly fat-soluble compounds or compounds that do not ionize in solution.
The solvent gas heated (from 300 to 500°C) inside the APCI probe is ionized on the occurrence of corona
discharge due to the application of a high voltage (±3 to 5kV) to the needle. The sample molecules are
ionized as a result of ion-molecular reactions (CI reactions) with the solvent ions. Nebulizer gas is used
to spray the liquid in the same way as with ESI.
APCI
9
C. DUIS
Dual Ion Source (DUIS) is an ionization method in which data can be obtained in both ESI and APCI
modes. High voltages are applied simultaneously to the nebulizer for ESI and the corona needle for APCI
and heated gas is used to assist ionization in the APCI mode. This probe is used as the nebulizer for ESI.
The DUIS ion source adds a corona discharge needle with a voltage of few kV. This latter induces ion
formation from low polarity compounds that may not ionize with ESI. This is not a true APCI source.
In LCMS, the ionization mode is selected depending on the polarity of the sample. ESI is usually chosen
for high-polarity compounds such as those found in drugs and pesticides. APCI is chosen for compounds
having lower polarity.
ESI Ionization is achieved by applying a high voltage to the capillary through which a stream of ionized
analyte molecules exits in nebulized drop form. In APCI, reaction ions are formed by the electrical
discharge around a high voltage corona needle. These ions react in the gas phase with neutral analyte
molecules to produce ionization.
For unknown compounds, you can select both APCI and ESI mode using the dual ion source or DUIS.
Both ESI and APCI ionization are executed concurrently and continuously without relying on switching
between modes. A standard ESI probe is used in the DUIS interface, meaning that uncompromised ESI
performance is readily achieved.
10
The ionization flow using a DUIS is as follows:
6. Triple Quadrupole
The mass analyzer used for electrospray ionization is a quadrupole mass spectrometer. A
quadrupole mass spectrometer uses four charged rods, two negatively charged and two positively
charged, that have alternating AC and DC currents. The rods are connected to both the positive
terminal of the DC voltage and the negative terminal. Each pair of rods contains a negatively
charged rod and a positively charged rod. The molecular ions are then sped through
the chamber between these pairs of oppositely charged rods making use of a potential difference
to do so. To maintain charge, and ultimately be readable by the detector, the molecular ions must
travel through the quadrupole chamber without touching any of the four charged rods. If a
molecular ion does run into one of the rods it will deem it neutral and undetectable.
Quadrupole MS contains four parallel cylindrical metal rods inside a vacuum chamber,
positioned equidistant from the center axis, arranged in a square. The analyte ions are directed
down the center of the square. Voltages applied to the rods generate electromagnetic fields.
The electromagnetic fields determine which mass-to charge ratio of ions can past through the
filter.
11
Quadrupole MS systems can separate ions at lower vacuum levels (10-2 to 10-3 Pa) than other mass
separation methods. Thus, when interfaced with LC unit, the drop in vacuum level caused by the
interface has minimum effect on mass separation performance. Quadrupole MS has a mass
measurement range of up to 2000m/z, which enables qualitative analysis in a practical range of
molecular masses. High-speed switching enables simultaneously monitoring multiple selected ions (SIM)
for higher-sensitivity simultaneous quantitative analysis of multiple components.
Quadrupole MS can operate in two modes called scanning (scan) mode or selected ion monitoring (SIM)
mode. In scan mode, the mass analyzer will monitor for a range of mass-to-charge ratios. Scan mode is
good for qualitative and quantitative analysis. In SIM mode, it will monitor for only a few mass-to
charge ratios. SIM mode allows quantitative analysis.
12
The mechanism of analysis for the mass spectrometer is that samples are introduced from the liquid
chromatography. ❶ The sample is sprayed and ionized under atmospheric pressure by the atmospheric
pressure ionization probe. ❷The ionized sample is introduced through the sample introduction unit or
desolvation line (DL), oriented 90° to the spray, into the first stage primary vacuum chamber, where it is
efficiently focused at the tip pf the skimmer by the Qarray where multi-stage high frequency ion guides
are arranged. ❸The rear section of the skimmer [asses through the high frequency ion guides (multi-
poles) arranged in the second and third vacuum chambers. ❹The ions are separated according to their
mass-to-charge ratio (m/z) by the quadrupole mass filter with pre-rod and the collision cell and are
detected by the detector. ❺The detected ions signals are first amplified by the amplifier and then
processed by the LabSolutions data processing software.
The charged droplets that have been spray ionized are heated by the desolvation line (DL). This heating
serves to remove the solvent and introduce ions into the vacuum. The lens system is made up of the
Qarray, skimmer, multipole and entrance lens. The ions generated under atmospheric pressure are
efficiently focused, and then introduced into the quadrupole rods by the lens system.
The vacuum housing comprises the atmospheric pressure chamber and the first, second, third, and
fourth vacuum chambers. The rotary pump evacuates the first vacuum chamber and the rear part of the
triple inlet turbo molecular pump, while a single inlet turbo molecular pump evacuates the second,
third, and fourth vacuum chambers. The pressure is measured by the Pirani gauge fitted to the first
vacuum chamber and the ion gauge fitted to the fourth vacuum chamber.
13
The nitrogen gas supplies 3 types of gas: nebulizer gas, drying gas, and standard sample delivery gas.
A mass spectrum tells us the molecular weight; however, we also want to see the fragment ions.
Fragments ions provide structural information for the compound of interest. The first mass
spectrometer is used to select a single (precursor) mass that is characteristic of a given analyte in a
mixture. The mass-selected ions pass through a CID where they are activated to cause them to fall apart
to produce fragment (product) ions. The second mass spectrometer is used to separate the fragment
ions according to mass. The resulting MS/MS spectrum consists only of product ions from the selected
precursor.
ɱ1+= precursor ion
ɱ1 + ɱ2 + +ɴ
ɱ2+= product ion
ɴ = neutral loss
If we perform an MS/MS experiment to find out all of the product ions, ɱ2+ that result from the
decomposition of a specified parent ion ɱ1+, and then this is a product-ion scan. It is used to determine
structurally significant fragment ions for a selected precursor ion.
A precursor-ion scan experiment tells us all of the possible precursor ions, ɱ1+ that decompose to
produce a specified product ion ɱ2+. This is useful when you know that a particular product (fragment)
14
ion mass is characteristic of a class of compounds. You can identify the mixture components that belong
to that compound class.
This is an MS/MS experiment that looks for all pairs of precursor ions and product ions that differ by a
constant neutral loss, N. This is useful when you know that a particular neutral loss mass is characteristic
of a class of compounds, and you would like to identify the mixture components that belong to that
class of compound.
Multiple reaction monitoring is the method used by the majority of scientists performing mass
spectrometric quantitation. MRM delivers a unique fragment ion that can be monitored and
quantified in the midst of a very complicated matrix. MRM plots are very simple, usually
containing only a single peak. This characteristic makes the MRM plot ideal for sensitive and
specific quantitation.
The MRM experiment is accomplished by specifying the parent mass of the compound for
MS/MS fragmentation and then specifically monitoring for a single fragment ion. One could
think of this operation as the SIM of a fragment ion. The specific experiment in known as a
"transition" and can be written (parent mass -> fragment mass) For example 534->375.
To obtain structural information, analyte ions are fragmented by colliding them with neutral gas in CID
15
B. Isotopes
A mass spectrometer has the ability to distinguish different isotopes. The presence of isotopes gives
each fragment a characteristic series of peaks with different intensities. These intensities can be
predicted based on the natural abundance of each isotope. An isotope has the same number to protons
but a different number of neutrons which results in a different overall mass. A lot of elements have a
variety of naturally occurring isotopes, like carbon, bromine and chlorine.
Each isotope will show up as a separate line in a mass spectrum. The Y-axis is relative intensity, so the
height of each peak will correspond to the relative abundance of each isotope in the sample. In the
example below, the compound contains a Bromine group. It’s isotope of 81Br is two atomic mass units
more of the actual mass of the compound. Notice that the peak heights are nearly identical since each
isotope is almost 50% naturally occurring.
79
Br and 81Br
Another sample is a compound containing a 35Cl. It has a naturally occurring isotope of 37Cl which is two
atomic mass units more at 251amu.
16
35
Cl and 37Cl
Another compound is 12C which has a naturally occurring isotope of 13C. The sample below show the
exact mass of 325 m/z with a 13C isotope at 326m/z. The Base Peak of this spectrum is 325 and would be
labeled M+. and the 13C with an atomic mass of 1 greater would be labeled (M+1)+. The intensity of each
isotope peak is proportional to the abundance of the isotope.
13C
Adduct ions are prevalent in LC-MS analyses and can come from any number of
sources. An adduct ion is any ion formed by adduction of an ionic species to a molecule,
ionization polarities (positive or negative ionization).
17
Example of glucose in positive Scan mode [M + Na]+ = 204. Can’t detect [M + H]+ = 181
Molecular weight of glucose=180.26
[M + Na]+ = 204
18
8. Liquid Chromatography
LCMS API (atmospheric pressure ionization) is used to analyze organic compounds with low to medium
polarity or medium to high polarity. Reversed-phase LC mode is best for separating and ionizing these
types of compounds. API is used for ionizing compounds with relatively high polarity by spraying the
mobile phase into a strong electric field to form a fine aerosol of charged droplets. Therefore, the
mobile phase must be able to dissolve polar compounds (water and polar organic solvents) and readily
form charged droplets (solvents with low viscosity and volatile salts).
Must be volatile
Must be aprotic
Common choices
Water
Acetonitrile
Methanol
Ethanol
Propanol
Isopropanol
Butanol
Hexane
Cyclohexane
Toluene
Ethyl Acetate
Mobile Phase
Must be MS grade
Be aware of source limitations on flow rates
On our instrument never go above 1ml/min
19
Mobile Phases Suitable for API
Solvents not to be used for LCMS are phosphate buffer solutions and non-polar solvents such as hexane.
Phosphate buffer solution can precipitate its non-volatile salts at the interface and potentially cause
mechanical damage. Non-polar solvents like hexane contribute very little to ionizing sample molecules
using APCI. Analytical conditions using this solvent would need modifications.
Since there are a limited number of mobile phases that are suitable for LCMS, one can change the
column as another option. C18(octadecyl) is the standard reverse-phase column, but one could try a
C30 to increase retention or a C8 to reduce retention. A phenyl column could increase separation
selectivity. Other options for better LC methods are to try isocratic or a different gradient elution.
TIC -Total Ion current chromatogram is a chromatogram this is displayed using all of the detected ions.
MC- Mass chromatogram shows the change in the intensity of m/z over time for a selected ion with a
specific mass.
20
Caffeine in MeOH
Caffeine in CH3CN
9. Analysis Modes
a. MS Analysis
1) MS Analysis is possible using either Q1 or Q3 in the MS analysis mode. Scan analysis
and SIM analysis are also possible. High speed scanning at 15,000µ/sec can be
performed in Q3 MS mode.
b. MS Scan Mode
1) Scan measurement is performed while scanning the mass ranges once every fixed
interval. This method is mainly used for qualitative analysis.
Q1 Scan Reserpine
Q3 Scan Reserpine
21
c. SIM Mode
1) In scan mode, a mass spectrum is obtained continually. In SIM mode, only ions with
the target mass are selectively detected. This enables high-sensitivity analysis with
no detection time wasted on the detection of ions with masses that are not
required. Since the peak height and area are stable, SIM is normally used for
quantitative analysis.
Q1 SIM Reserpine
Q3 Sim Reserpine
22
e. Precursor Ion Scan Mode
1) This mode involves performing scanning in Q1, fixing Q3 to a specific m/z, and
selectively analyzing the ions generated in the CID. This allows examination of
precursor ions with common product ions, and it is suitable for screening ions with
common substructures.
23
a.
24
Neutral loss scan of reserpine
25
10. Sample Preparation-Very Important
a) Never use 100% water for C18 column – bad for column
b) Solvent A is always Water (Acid or base)
c) Solvent B is always Organic ( MeOH, Acetonitrile)
d) Water has 0.1% v/v solution of acid or base
a. Example: 1ml formic acid in 1 liter of H2O.
e) Use H2O from wall unit or LC/MS grade
f) Store column in 50/50 mixture of MeOH/H2O (no acid or base)
g) Read about your particular column.
h) Never use a flow rate above 1mL/min.
i) Samples ionize better in acetonitrile.
26