D-p5-Bv-pd-019 - DRT, Issue 01, Rev 00 - Digital Radiography Testing
D-p5-Bv-pd-019 - DRT, Issue 01, Rev 00 - Digital Radiography Testing
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
W/ Order No. :
Package No. : 5
THIS DOCUMENT IS CONTROLLED BY THE BUREAU VERITAS - VIETNAM AND MAY NOT BE AMENDED, REVISED OR ALTERED IN
ANY OTHER WAY WITHOUT THE CONSENT OF THAT COMPANY.
THE SIGNATURES BELOW AUTHORISE ALL PAGES OF THIS PROCEDURE FOR USE FROM THE DATE OF APPROVAL AS SHOWN.
Page 1 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
REVISION HISTORY
Page 2 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
INDEX
1.0 Scope 4
15.0 Reporting 20
Page 3 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
1.0 SCOPE:
1.1 This procedure has been established for Radiographic Examination on Full Penetration
Butt Welded Joints, Profile radiography to access the corrosion of the material carried
out by BVV.
1.2 This procedure is applicable for Radiographic Examination of Welds in accordance to the
following referenced codes, standards and documents
Page 4 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
AS 2177.1-
Non Radiography of welded butt joints in metal - Methods of test.
1994
AS2885.2-
Pipelines-Gas and liquid petroleum Part 2 Welding
2007.
Page 5 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
4.1 The weld ripples or the weld surface irregularities on both the inside (where
accessible) and outside shall be removed by any suitable process to such a degree
that the resulting radiographic image due to any surface irregularities cannot mask
or be confused with the image of any discontinuity.
4.2 The finished surface of all butt welded joints may be flushed with the base material
or may have reasonably uniform crowns with reinforcement not exceeding that
specified in the referencing code section.
5.1 Radiographic examination shall be performed with the INC 100 Gamma Ray
Projector using IR –192 source operated by cranking unit (wind out), or with the
Andrex CMA 306 (300 KV) X-Ray machine.
5.2 The Radiation source size for Radiography with Gamma rays using Iridium 192
Isotope shall be determined with SE-1114, Standard Test Method for Determining
the Focal Size of Iridium-192 Industrial Radiographic Source.
5.3 The Radiation source size for Radiography with X-rays using machine operated at
500kV and less, the spot size may be determined by pinhole method or in
accordance with SE-1165, Standard Test Method for measurement of Focal.
6.1 DXR 250P Direct Radiography Panel (Detector) and its characteristics are as follows:
Page 6 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
Detector Specifications
Power Supply
Weight 3 Kg (7 lb)
7.1 Digital radiography is also known as direct radiography, for short DR, with DR-
technology; there is an immediate conversion of radiation intensity into digital image
Page 7 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
information.
7.2 Exposure and image formation happen simultaneously, allowing near real-time image
capture, with the image / radiograph available for review only seconds after the
exposure.
7.3 This almost instant image formation is the reason that DR is considered the only genuine
method of digital radiography.
7.4 Some of the Devices even provide a true real-time mode with display rates upto 30
images per seconds.
7.5 For DR, Flat panel detectors in a variety of sizes are used that convert incident radiation
intensity into proportional and digitized electronic signals.
Page 8 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
7.6 These digital signals can, by means of a computer and screen (workstation), without
intermediate steps, be presented as a coherent radiographic image.
7.7 A cable typically links the detector to this workstation from which the panel is controlled
as well.
7.8 There are different types and suppliers of DR-flat panel systems.
7.9 A variety of flat panel systems exist with a wide range of pixel sizes and resolutions.
7.10 The more and smaller the pixels the higher the potential resolution of the system.
7.11 A sensor materials amorphous silicon and amorphous selenium are in use.
7.12 As sensors CCD’s (charge couples devices) and CMOS (Complementary Metal Oxide
semiconductor) are applied.
7.13 The most common high resolution flat panels use amorphous silicon technology.
7.16 This light is turn is converted from into a proportional electric signal by a scintillator made
of eg. Structured Cesium Iodide (CsI) photodiodes and integrated thin film transistors
(TFT’s).
7.17 Each picture element (pixel) contributes to the radiographic image formed on the screen
of the workstation.
7.18 Each element is square in effective area, with pixel typically ranging from 50 to 400
micron.
Page 9 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
8.1 IQIs shall be selected from either the same alloy material group or grade as identified in
SE-747or from the alloy material group or grade with less radiation absorption than the
material being radiographed.
8.2.1 Source Side IQI(s). The IQI(s) shall be placed on the source side of the part
being examined.
8.2.2 There is no restriction on the placing the IQI(s) on a separate blocks when, due
to part or weld configuration or size, is not practical to be placed, provided the
IQI(s) /area-of interest density tolerance requirement are met.
Page 10 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
8.2.3 The IQI(s) on the source side of the separate block shall be placed no closer to
the film than the source side of the part being radiographed.
8.2.4 The block dimensions shall exceed the IQI(s) dimensions such that the outline of
at least three sides of the IQI(s) image shall be visible on the radiograph.
8.2.5 Film side IQI could be placed when inaccessibility prevents hand placing the
IQI9s) on the source side. A lead letter F shall be placed on the IQI(s) or
adjacent but shall not mask the essential wire or hole.
8.3.1 Wire IQI: The IQI shall be placed on the weld so that the length of the wires is
perpendicular to the length of the weld.
8.3.2 The identification numbers and when in used the lead letter ‘F’ shall not be in the
area of interest, except when geometric configuration makes it impractical.
8.4.1 The IQI with IQI identification number(s), and, when used, the lead letter “F’ may
be placed in the area of interest.
8.5 If using more than one IQI, one shall be representative of the lightest area of interest: the
intervening densities on the radiograph shall be considered as having acceptable density
8.6 The IQI shall be selected from Table T-276 of ASME Section V Article 2.
8.7 Other equivalent IQI (DIN, EN or BS) may be used documented comparison is made for
the respective ASTM and approved by NDT Level III.
Page 11 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
Page 12 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
Page 13 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
Table – 1
Material Thickness Range plus estimated reinforcement (not exceeding maximum permitted by referencing code
section) actual measurement not required. Back rings or strips shall not be considered as part of the thickness in IQI
selection.
9.1 The image brightness through the body of the hole-type IQI or adjacent to the
designated wire of the wire-type IQI, shall be judged to be equal to or greater than the
image brightness in the area of interest for a negative image format.
9.2 This image brightness requirement is reversed for a positive image format.
9.3 The contrast and brightness range that demonstrates the required sensitivity shall be
considered valid contrast and brightness values for interpretation.
Page 14 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
9.4 When multiple IQIs are utilized to cover different thickness ranges:
9.4.1 The contrast and brightness range that demonstrates the required IQI image of
each IQI shall be determined.
9.4.2 Intervening thicknesses may be interpreted using the overlapping portions of the
determined contrast and brightness ranges.
10.1.1.1 In the single wall technique, the radiation passes through only one
wall of the weld material.
10.1.1.2 This technique is used for plates and for the pipe it shall be used as
panoramic technique with diameter greater than or equal to 20”.
10.1.1.3 In this technique, the source is kept at the center of the pipe and the
films are kept on the external side covering the circumference of the
weld (Refer to Appendix A).
10.1.2.1 In this technique the radiation passes through two walls and only the
weld (material) on the film side-wall is viewed for acceptance on the
radiograph.
Page 15 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
10.1.2.3 This technique should be used for pipe diameter greater than or equal
to 3” (Refer to Appendix B).
10.1.3.2 In this technique, the radiation passes through two walls and both
walls are viewed for acceptance on the radiograph.
10.1.3.3 This technique should be used for pipe diameter less than or equal to
3”.
10.1.3.4 Only a source side IQI shall be used. (See IQI selection in this
procedure.)
10.1.3.5 Double Wall Double Image may be performed where the source is
offset to the weld (Elliptical technique), or where the source is at right
angles to the weld (material) and the image of both walls are
superimposed (Superimposed technique).
Page 16 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
10.2.1 The radiation energy used for any radiographic technique shall achieve the
density and IQI image requirements stated section 8.0 above.
10.3.1 The direction of central beam of radiation shall be centered on the area of
interest whenever practical.
10.4.1 The source to object distance, object to film distance shall be selected in such a
manner that for given source size, the radiograph shall achieve geometric
unsharpness not exceed the values given below for the specified thickness
ranges
Page 17 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
Page 18 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
10.5.2 Film side markers shall be placed on the film when radiographing curved or
spherical components whose concave side is toward the source and when the
source to material distance is greater than the inside radius.
10.5.3 Either side (source or film side) marker may be placed when radiographing either
curved or spherical components whose concave side is toward the source and
source to material distance equals the inside radius of component.
10.5.4 The location markers used shall be made of lead and shall be placed on part (not
on exposure holder/cassette) and their locations marked on the part surface
(permanently, when permitted) being radiographed to help in mapping of
defective area, traceability and identification of part with respect to radiograph.
11.2.2 The digital image shall be free of system-induced artifacts, such as nonfunctional
pixels in the detector in the area of interest that could mask or be confused with
the image of any discontinuity.
Page 19 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
12.1 Viewing facilities shall provide subdued background lighting of an intensity that will not
interfere with the Image viewing and interpretation.
12.2 Masks shall be provided for the viewing equipment such that light from around the outer
edge of the film being viewed or from low-density areas of the film will not interfere with
interpretation.
13.1 Acceptance criteria shall be done in accordance with the criteria given in the referencing
code specified on the contractual documents or engineering drawings or customer
issued work order/purchase order.
14.1 This procedure shall be reviewed annually from the date of revision. As a minimum, the
following shall be considered while reviewing:
14.1.1 Any revision to the reference documents and its impact on this procedure.
14.2 Any other additional information that is deemed necessary for the intent of this
procedure.
15.0 REPORTING:
15.1 All the radiographic test results shall be recorded in the TEST REPORT (Please see the
Appendix E).
Page 20 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
APPENDIX A
EXPOSURE TECHNIQUE 1
FILM
SOURCE
Page 21 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
APPENDIX B
EXPOSURE TECHNIQUE 2
SOURCE
FILM
Page 22 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
APPENDIX C
EXPOSURE TECHNIQUE 3
SOURCE
FILM
Page 23 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
APPENDIX D
EXPOSURE TECHNIQUE
SOURCE
FILM
Page 24 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
APPENDIX E
Page 25 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
APPENDIX F
ACCEPTANCE CRITERIA FOR ASME SEC VIII DIV 1 UW 51(b), API 650, ASME SEC I,
ASME B 31.1 AND ASME SEC IX
Rounded Indications Max. Size of Rounded Indication unacceptable Max. Size of Rounded Indication
if unacceptable if
a) ¼ t or 4mm (whichever is smaller) if distance a) > 0.2 t or 3 mm (whichever is
between isolated indication separated by < 25 smaller) [for all weld thickness]
mm. b) > 12 no. of rounded indication {for
b) (1/3) t or 6 mm (whichever is smaller) if those acceptable per a)} per 150 mm
distance between isolated indication separated weld length. For weld length less
by > 25 mm than 150 mm proportionately fewer
c) 10 mm for t > 50 mm are allowed. [weld thickness < 3
mm]
c) Greater than shown in Appendix I
ASME SEC IX [for weld thickness
> 3 mm]
Aligned Rounded Indications Unacceptable if
a) Summation of diameters > t in a length of 12 t
b) Maximum group length of aligned indications
L= 6 mm for t <19 mm
L= 1/3 t for 19 mm < t < 57 mm
L = 19 mm for t > 57 mm
c) Maximum allowable spacing between groups of aligned indication = 3L (where L is the
longest indication)
Rounded Indication / Cluster of rounded Indications
a) Refer to Appendix 4, ASME SEC VIII, Div I for pressure vessels, tanks for oil
storage.
b) Refer to appendix A-250, ASME SEC I for power boilers and power piping.
Page 26 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
APPENDIX G
Concave root 1. For Weld Type 1 & 2: Unacceptable if Depth of Root surface concavity is < Total joint thickness
Surface (Suck inclusive of weld reinforcement
Up) 2. N/A for Fillet Welds.
Page 27 of 28
BUREAU VERITAS
VIETNAM
NSRP DOCUMENT No.:
DCSM PROJECT 2019 DIGITAL RADIOGRAPHIC D-P5-BV-PD-019
Note:
1) Weld Type 1: Girth & Miter Groove
2) Weld Type 2: Longitudinal Groove (includes Straight and Spiral Seam)
3) Weld Type 3: Fillet Welds (includes socket and seal welds, and attachment welds for slip-on flanges, branch
reinforcement, and supports)
4) Weld Type 4: Branch Connection (Includes Pressure containing welds in branches and fabricated laps)
5) “N/A” indicates Code does not establish acceptance criteria or does not require evaluation for this kind of
imperfection for this type of weld
Page 28 of 28