Fault Detection in Small Diameter Pipes Using
Fault Detection in Small Diameter Pipes Using
April 2013
This page is intentionally left blank.
Abstract
Ultrasonic guided wave technology is one of the more recent developments in the
field of non-destructive evaluation. It was introduced in 1998, mainly in the oil and
gas industry, to monitor pipelines for corrosion, defects and loss of metal. Ultrasonic
guided wave is an attractive technology as it requires only exposing the area where
the transducers will be placed, hence requiring minimal insulation removal and
excavation for a buried pipe. In this project small diameter pipes (with diameter less
than 5 cm) are considered due to a request by Scotia Gas Network Ltd. The aim
of this project was to investigate propagation of torsional mode, T(0,1), in various
different scenarios.
Experimental study was conducted on a bare steel pipe with a diameter of 3.4 cm
and a wall thickness of 0.5 cm. A defect was machined onto the pipe which occupied
8.3% cross sectional area of the pipe wall. The aim was to find the frequency at
which the wave must be transmitted, in order to detect the defect located in the
pipe. The results obtained show that if the torsional mode, T(0,1), was transmitted
at a frequency of 50 kHz the defect can be detected. Whereas, when the guided
waves were transmitted at higher frequencies there was high reverberation level that
prevented the defect echo to be clearly distinguished. This high reverberation was
due to the limited number of transducers being used in the experiment. The next
experiment involved rotating the receiving transducer around the circumference of
the pipe so that multiple readings could be taken; it was then averaged to give a
resulting waveform. This experiment demonstrated taking multiple readings around
the circumference of the pipe reduces reverberation level at high frequencies, by
eliminating non-axisymmetric modes as it has phase difference at different positions.
Next, guided waves were propagated through a bitumen coated pipe, to investig-
ate the response of guided waves. The results obtained were poor as guided waves
travelled very limited distance, due to high attenuation caused by the bitumen. Lastly,
guided waves were propagated through a buried bare pipe in order to investigate
whether guided waves can detect the defect. The results were positive as the defect
echo was clearly identifiable. Further work using more transducers, using a chirp
instead of tone burst input signal and testing on a corroded pipe is recommended.
iv
Abstract iii
List of Figures ix
List of Abbreviations xv
Acknowledgements xvii
1 Introduction 1
1.1 Motivation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
1.2 Aims and Objectives . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
1.3 Brief outline of dissertation . . . . . . . . . . . . . . . . . . . . . . . 3
2 Pipeline Inspection 5
2.1 Ultrasonic Guided Wave Technology . . . . . . . . . . . . . . . . . . 5
2.2 Background . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Guided waves in pipes . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.3.1 Types of guided waves in pipes . . . . . . . . . . . . . . . . . 7
2.3.2 Naming of guided waves . . . . . . . . . . . . . . . . . . . . . 8
2.3.3 Background . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.3.4 Choosing a wave mode and frequency for inspection . . . . . 9
2.3.5 Torsional vs. Longitudinal waves . . . . . . . . . . . . . . . . 10
2.3.6 Dispersion curves . . . . . . . . . . . . . . . . . . . . . . . . 11
2.4 Ultrasonic Guided Wave Technology vs. Conventional Ultrasonics . 12
2.5 Generation of guided waves . . . . . . . . . . . . . . . . . . . . . . . 14
3 Experiments 17
3.1 Single position measurement . . . . . . . . . . . . . . . . . . . . . . 17
3.1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.1.2 Experimental setup . . . . . . . . . . . . . . . . . . . . . . . 17
3.1.3 Results and analysis . . . . . . . . . . . . . . . . . . . . . . . 21
3.1.4 Discussion and experimental conclusion . . . . . . . . . . . . 26
vi CONTENTS
4 Conclusion 59
4.1 Main findings of this dissertation . . . . . . . . . . . . . . . . . . . . 59
4.2 Future work . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
References 61
Appendices
C CD V
3.24 Setup for the experiment showing two transducers placed on the
bitumen coated pipe. The transmitting and receiving transducers here
are placed 6.2 cm apart from each other and the bitumen coating is
removed so that the transducers can couple on the pipe. . . . . . . . 42
3.25 Comparing how the amplitude of received signal varies in a bitumen
coated pipe, for guided waves transmitted at different frequencies,
when a single reading at zero degree configuration was taken. . . . 44
3.26 Variation in amplitude of direct echo in non-coated and bitumen
coated pipe. The graphs shows results from normal pipe where the
input burst was transmitted at Vpk−pk of 19 V and from bitumen coated
pipe where the input tone burst was transmitted at Vpk−pk of 19 V and
4.75 V. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
3.27 Variation in normalised reverberation RMS in a bitumen coated pipe,
for guided waves transmitted at different frequencies, when the tone
burst was transmitted at Vpk−pk of 19 V and 4.75 V. . . . . . . . . . . 46
3.28 Dimensions of the box through which the pipe goes through. This
was then filled with 0.0324 m3 of sand. . . . . . . . . . . . . . . . . . 48
3.29 Part of pipe along with the defect buried in the sand box. The box
contained 50 kg of sand. . . . . . . . . . . . . . . . . . . . . . . . . . 49
3.30 Transmitting and receiving transducers placed on the pipe, at 0 ◦
configuration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
3.31 Setup for the experiment showing two transducers placed on the pipe.
The transmitting and receiving transducers here are placed 6.2 cm
apart from each other and it has a defect buried in sand. . . . . . . . 50
3.32 Amplitude of average received signal, for guided waves transmitted
at different frequencies through the buried pipe, when eight evenly
spaced readings are taken circumferentially around the pipe. . . . . 52
3.33 Comparing how the average normalised reverberation RMS varies,
for guided waves transmitted at different frequencies, when the pipe
is buried and not buried. . . . . . . . . . . . . . . . . . . . . . . . . . 53
3.34 Comparing how SRR varies, for guided waves transmitted at different
frequencies, when the pipe is buried and not buried. . . . . . . . . . 53
3.35 Variation in ratio between peak amplitude for a buried and non-buried
pipe, when guided waves are transmitted at different frequencies. . . 54
This page is intentionally left blank.
List of Tables
a m Radius of pipe
c ms−1 Wave velocity
E Nm−2 Young’s modulus
f Hz Frequency
h m Thickness
k m−1 Wave-number (spatial frequency)
L m Length of pipe
m Dimensionless Counter variable / Mode
n Dimensionless Order of circumferential variation
N Dimensionless Number of . . .
r m Radial direction of pipe
SNR dB Signal to Noise Ratio
SRR dB Signal to Reverberation Ratio
U m Displacement vector
x m Distance
z m Axial direction of pipe
Subscript Description
bitumen bitumen
deg degrees
ext external
g group
i initial
int internal
mean arithmetic mean
p phase
pipe pipe
pk-pk peak to peak
r readings
rev reverberation
rms root mean square
s signal
t transducers
List of Abbreviations
Abbreviation Description
Firstly I would like to express gratitude to my supervisors for their enthusiasm, en-
couragement, dedication, support and helpful guidance throughout this dissertation.
I am greatly honoured to have supervisors who added great value to this dissertation
with his vast knowledge and experience of working with ultrasonic technology. His
knowledge and advice have helped me to stay on track and work at a smooth pace.
I am greatly indebted and thankful to my family and would like to dedicate this
dissertation to them. I would like to thank my parents for striving hard to provide
a good education for me. Thank you for all your understanding, love, support and
motivation that you have provided me with at every stage of my life. I would also like
to thank my twin brother, Rohit, for his support and regular humour that has helped
in relieving stress throughout the course.
Introduction
1.1 Motivation
Pipelines play an important role in our everyday lives. With energy companies,
spending billions in exploring and extracting fossil fuels; people tend to neglect the
role of pipelines in delivering these fuels to the final consumer.
Pipelines are similar to a human body, as it ages, it deteriorates and regular tests
must be performed to prevent further deterioration. They are similar to other modes
of transportation available in a country, except that they are more capital-intensive
and are long-term assets to a country. Regular inspection of pipelines is a must, in
order to avoid complete replacement of pipelines, which can be expensive. Testing
of pipelines is a very important issue especially in the petrochemical industry to
avoid leakages, as they transport inflammable liquids, which can cause harm to both
the environment and living organisms. These pipelines can be as long as millions
of kilometres, therefore it can be quite difficult to regularly monitor these lines. As
regular maintenance of these pipelines is required it is important to find techniques
that have high sensitivity, reliability, can perform inspections over long distances, are
quick and at the same time are not very expensive to operate.
There are many different techniques available that can be used to detect defects
and corrosion in pipes. Some of the more common techniques include radiography,
eddy-current, magnetic flux, electrochemical and the use of ultrasonic guided waves.
Ultrasonic Guided Wave Technology (UGWT) is one of the more attractive methods
of inspection as this technique allows quick, accurate and long range inspection of
pipes. It only requires excavation during initial instalment of the transducer arrays
that excite the ultrasonic waves and during the replacement of the damaged areas.
This technique was developed in NDT Lab at Imperial College [1] and most of the
research has been performed on large diameter pipes using torsional mode, T(0,1).
In this project, tests were performed using the torsional mode, T(0,1), to detect a
2 Introduction
defect located in the pipe and further tests were performed to understand how this
technique responds when the pipe is coated with bitumen and when the pipe is
buried in sand.
Chapter 3, describes the experiments performed along with the results and
conclusion for each experiment. First experiment focuses on transmitting guided
waves at different frequencies and taking a single measurement to determine the
frequency at which the defect on the pipe is clearly detected. Next experiment
involved taking multiple readings around the circumference of the pipe to see how
it affects the results. Tests were even performed on a bitumen coated pipe and a
buried pipe to determine how the guided waves respond in this scenario.
Pipeline Inspection
This chapter introduces the basic concepts and theories of ultrasonic guided wave
propagation in hollow cylindrical pipes; it explains how this technology operates and
detects defects in pipes. UGWT is regarded as an attractive non-destructive testing
(NDT) method available for inspection of pipelines. There are two known methods
to inspect defects in pipes, destructive and non-destructive. Destructive testing is
where the sample pipe is taken to a laboratory for the mechanical properties to be
examined. It uses hydrostatic testing to ensure that the pipe is within safe operating
conditions. Whereas, NDT is generally used to detect flaws, cracks and corrosion; it
is a preferred method as the pipeline operation is not usually disrupted [2].
Guided wave inspection, also known as long range inspection, exploits ultrasonic
waves that propagate along a medium and are guided by the geometric boundaries
of the medium [3]. These waves are generated by exciting a transducer with high
voltage electric pulses; these are converted to a mechanical vibration by a trans-
ducer. The waves then propagate through the material and are reflected back if
there are any defects, welds or pipe ends along the path of the wave (see Figure
2.1). The reflected wave is then received by the transducer and converted back to
an electrical signal to be displayed on the oscilloscope. Received signal usually
have higher amplitude than the background noise level [4, 5].
The difficulty with guided waves is that the velocity is highly dependent on
wavelength and frequency, known as dispersion. Hence, dispersion curves are used
as it shows relation between the wave mode, velocities and frequency.
If the wave velocity and the travel time are known they can be used to calculate
the location of the reflection i.e. discontinuity. There are two known approaches
to ultrasonic testing - through transmission and pulse echo. Through transmission
6 Pipeline Inspection
method is used when there is access to two opposing sides of the material being
tested, where the transmitting and receiving transducers are placed far away from
each other. Pulse echo is when only one side of the material being tested is
accessible and therefore both transmitting and receiving transducers or sometimes
one transmitter-receiver transducer is located here. Pulse echo transmission method
is adopted for this project.
Figure 2.1: Transmission of guided waves along the pipe, with a piezoelectric
transducer (green circle) placed in each module.
2.2 Background
Guided waves were first studied by Rayleigh and Lamb, early in the 19th century;
they studied elastic wave propagation in traction-free, isotropic plates; this was later
acknowledged by naming guided waves after them1 [6]. Pochamer, in 1876, and
Chree, in 1889, investigated the propagation of guided waves in a free bar [1]. In
1959, Gazis released a paper where he discussed the propagation of waves in hollow
cylinder. He found an exact solution for propagation of free harmonic waves along
the hollow cylinder, using linear theory of elasticity; he also spoke about evolution of
wave structures of different modes at different frequency [7]. This was followed by
Fitch performing an experiment using axisymmetric and non-axisymmetric modes
to confirm the result suggested by Gazis [8]. Research was also performed by Pao
and Mindilin (1960), Onoe et al, Meeker (1962) and Meitzler (1972) which developed
three dimensional problem of a circular cylinder in vacuum [1]. In 1965, Viktorov
et al. analysed the method of generating guided wave in isotropic layers using a
wedge, both theoretically and analytically [9].
[10]. In 1979, Silk and Bainton researched the use of guided waves for inspection
of small heat exchange tubing; they found similarities between ultrasonic waves in
hollow cylinder and lamb waves in flat plates. A new notation for the wave modes
was also provided by them [11], discussed ahead in this report. In recent times
Alleyne and Cawley (1995) have reported on how it is possible to excite lamb waves
using dry coupled piezoelectric transducers [12]; Lowe et al. (1998) performed
corrosion screening in pipes using long range guided waves [13] and Alleyne et
al. (2001) used guided waves to detect cracks and corrosion in chemical plant
pipes [14].
Guided waves
in pipes
Longitudinal
Torsional mode Flexural mode
mode
Figure 2.2: Flow chart showing the types of guided waves in pipes.
fields [16].
2.3.3 Background
When detecting faults in pipes the fundamental torsional mode, T(0,1), and lon-
gitudinal mode, L(0,2), are the most commonly used modes. These are preferred,
because of their non-dispersive nature over a wide frequency band. Non-dispersive
motions are those in which the wave motion does not change as it propagates; their
phase velocity is independent of wave-number.
Traditionally L(0,2) has been preferred mode by many researchers, it was used
by Silk and Bainton when they were experimenting on heat exchange tubing [11], by
Alleyne et al. while testing on chemical pipes [14] and by Lowe et al. when detecting
defects in pipes [13]. Though, recently there has been a lot of work performed
using the torsional mode T(0,1). In 2001, Alleyne and Pavalakovic used T(0,1)
along with L(0,2) to detect defects in a pipe [14]. Demma in 2003, experimented
with T(0,1) to analyse and study reflections from defects, cracks and notches in
pipes [1, 18]. Davies and Crawley in 2009, performed synthetic focusing using T(0,1)
for imaging crack-like defects [19]. In 2010, Cardente et al. assessed the scattering
of T(0,1) due to tapered notches in pipes [20] and recently in 2012, Lovstad and
Crawley studied the reflection of T(0,1) due to pitting in pipes [21]. For this project
the fundamental torsion mode, T(0,1), has been used; the reasons for choosing this
mode are discussed in the sections ahead.
As mentioned in the section above, there are three different modes to choose
from after which the frequency at which the wave is transmitted at must be decided.
There are certain modes and frequencies that are sensitive to particular defects and
therefore in order to obtain reliable results it is necessary to choose appropriate
mode for the wave. Wilcox et al. defined a list of factors which influence the mode
and frequency selection and these include dispersion, attenuation, sensitivity, excit-
ability, detectability and mode selectivity and in practical implementation speed of
single test, testing tool design and level of difficulty in analysing data must also be
taken into account [1].
The frequency at which guided waves are transmitted determines the smallest
defect that can be detected. As the frequency increases the waves become more
sensitive to small defects, but the frequency is limited by vibration levels. Whereas
when testing a long pipe, low frequency guided waves are selected due to their
ability to travel further [5]. In this experiment the torsional mode T(0,1) was used as
it is non-dispersive and it is transmitted at frequency of 50 kHz, though the aim is to
try to be able to transmit the waves at a frequency of 100 kHz, using two transducers.
10 Pipeline Inspection
One problem with L(0,2) is that in the operating frequencies above 35 kHz
there is another mode, L(0,1), which interferes as it has lower group velocity than
L(0,2). Whereas T(0,1) is the only torsional mode in the frequency range of interest,
as T(0,2) has higher cut-off frequency [3]. Table 2.1 shows advantages of using
torsional mode.
• Even though torsional modes (3.2 kms−1 ) travel along the pipe slower com-
pared to longitudinal modes (5.4 kms−1 ), they are more sensitive to discon-
tinuities [14, 22].
• Torsional modes have less mode conversion, refraction and reflection com-
pared to longitudinal modes; this is because they propagate slower along
the pipe [4, 23].
• Torsional mode has the ability to detect longitudinal defects, whereas the
longitudinal mode is less able detect defects which are narrow and parallel
to pipe axis. This is because the torsional mode has dominant shear strain
which is perpendicular to axis [14]. In one of the experiments carried out
by Zenghua Liu et al. in 2006, he shows that T(0,1) mode can detect both
longitudinal and circumferential defects [24].
and then in that case the extent of mode conversion produced by a reflector can
be measured. It also reduces the test range, then in that case L(0,2) would be
preferable [14].
Reflectors
in pipe
Symmetrical Asymmetrical
reflectors reflectors
Dispersion curves can be plotted for either phase velocity or group velocity, they
are usually calculated for pipe in vacuum as there is not much difference between
solutions of air or vacuum, due to acoustic impedance of air being negligible com-
pared to a pipe.
Phase velocity dispersion curves show the velocity of the wave cycles within a
signal; it depends on the thickness of the pipe. Group velocity dispersion curves
show velocity of finite-time wave packets, hence they are used to calculate the travel
time of wave signal and it depends on stress magnitude [3, 26].
Gazis, in 1959 derived the dispersion curve for a hollow cylinder [7]. In 1967,
Vikotrov calculated the dispersion curve for an isotropic cylinder and compared the
results with the experimental results obtained from piezoelectric transducers. Viko-
trov used a form where the wave velocity was on y-axis and thickness/wavelength
ratio on x-axis. Though, the form that is currently preferred is wave velocity on y-axis
12 Pipeline Inspection
Table 2.2: Typical range the guided waves propagate with a standard transducer [29].
Guided waves have the advantage of being able to inspect coated pipes. It only
requires exposing a small area of circumference to place the transducers, therefore
saving time and money. It also has the ability to inspect buried pipes, therefore
making it cheaper to detect faults if the pipelines are buried below a road. Con-
ventional method can result in road being excavated costing up to US$50000 [30],
whereas when using guided waves excavation is only required during installation
and replacement [30] (see Table 2.3 for more advantages).
2.4 Ultrasonic Guided Wave Technology vs. Conventional Ultrasonics 13
• It is highly safe and doesn’t emit harmful radiation like radiography inspection
methods [3].
• It can detect changes in wall thickness, which might be due to corrosion [4].
• It has the ability to detect defects in buried, insulated or elevated pipes and
also detect in areas that would be inaccessible otherwise [30].
• Propagates wave in both direction of long pipe from a single test point [31].
• Has a higher sensitivity than conventional ultrasonic and other NDT methods
[31].
• Saves time and money as insulation needs to be removed only where the
transducers will be placed [31].
Table 2.3: Advantages of Ultrasonic Guided Wave Technology for pipeline inspection.
Though UGWT has quite a few benefits, one of the drawbacks is the limitation in
distance of wave propagation for heavily corroded and bitumen coated pipes [3] (see
Table 2.2). UGWT also has limitation in ability to detect small volume defects like
14 Pipeline Inspection
localized pitting [3] and it does not have the ability to distinguish between internal
and external corrosion, which means visual inspection must be performed leading
to additional costs as insulation needs to be removed at these locations [32].
The other problem is that the output waveform depends on the type of material
being inspected e.g. cast iron and other grain structures give poor results due to
low sound transmission and high signal noise [4]. Furthermore, UGWT being a
sophisticated technology requires very experienced technicians to operate it [4].
Despite these drawbacks, there are still positives about this technology and it
is one of the technologies constantly improving. One of the leading oil companies,
BAPCO, had started using this technology in 2000 [32].
The paper also states that the number of piezoelectric transducers required is
2.5 Generation of guided waves 15
Figure 2.8: Group velocity dispersion curve for a steel pipe with an external diameter
of 75 mm [13].
This page is intentionally left blank.
Chapter 3
Experiments
3.1.1 Introduction
As mentioned earlier, UGWT allows long range inspection of pipes in order to
detect corrosion and defects. This experiment was performed to gain a deeper
understanding of how the technology operates, what each echo represents and to
find out at what frequency the T(0,1) mode must be transmitted so that the defect
and pipe end echoes are clearly distinguishable. This is an important experiment,
called frequency tuning, and has been performed by past researchers also [23, 24].
The selection of frequency at which guided wave is transmitted is important as it
determines the ability to send the ultrasonic energy and it must be carefully selected
so that this energy can travel from position A to B [23].
As known from theory, high frequency waves are more sensitive to small defects
and the lowest frequency range used in practical testing with torsional mode T(0,1)
is 45-65 kHz for a 3 inch pipe and 10-20 kHz for 24 inch pipe [33]. It is also stated
that frequency range of 20-100 kHz can be used to detect defect as small as 5% of
the pipe’s CSA [33]. Therefore the hypothesis was that as the frequency increases
it can become difficult to detect the defect echo, due to high reverberation.
3.2). The tests were performed with the pipe horizontally placed and the ends
resting on supports. Torsional fundamental mode, T(0,1), was transmitted along the
pipe for all the experiments. It was transmitted over a range of frequencies varying
from 30 kHz to 200 kHz, in order to determine the frequencies at which the defect
echo was identifiable.
Figure 3.1: Properties of the bare steel pipe used in the experiment.
For all the experiments two dry coupled piezoelectric transducers were used,
where one of them transmitted the guided waves along the pipe and other received
it (pulse echo transmission method was adopted) (see Figure 3.3). Piezoelectric
transducers were used because of their high sensitivity, low price and convenient
usage [34]. It produced tangential displacement, thereby generating the torsional
mode. The transducers were placed 6.2 cm apart from each other, this was the
minimum possible distance. The receiving transducer was placed 186.8 cm to
the left of the defect, while the transmitting transducer was located 54 cm to the
right of the left pipe end wall; they were placed at zero degrees configuration (see
Figure 3.4). Function generator2 was connected to the transmitting transducer that
2
Function generator HP 33120A was used.
3.1 Single position measurement 19
provided an input tone burst of three cycles at a burst rate of 10 kHz and Vpk−pk
of 19 V. This was the maximum voltage that could be applied; applying a lower
voltage would result in smaller amplitude as the transducers would get excited slowly.
Function generator was connected to the channel 1 of the oscilloscope3 to display
the input waveform. The receiver transducer was connected to a signal amplifier4
that increased the amplitude of the output. This then led to a bandpass filter that
removed the high frequency signals and low frequency noise, thereby making the
output waveforms smoother before being displayed by the oscilloscope (see Figure
3.5). A General Purpose Interface Bus (GPIB) cable was used, that enabled the
data to be transferred to a computer so that the results can be recorded by a Matlab
data capture script that also allowed controlling of the oscilloscope. The script was
edited by Zaghari (2012).
Figure 3.3: A module with a piezoelectric transducer in place. The module was
placed on the pipe so that the transducer can produce tangential displacement in
order to generate torsional mode. Two modules were used in the experiment, each
with a transducer where one transmitted the guided wave and the other received it.
Figure 3.4: Transmitting and receiving transducers placed at zero degree configura-
tion.
3
Oscilloscope LeCroy 9304C was used.
4
ORTEC Precision AC Amplifier 9452 was used.
20
Figure 3.5: Setup for the experiment showing two transducers placed on the pipe. The transmitting and receiving transducers here are
placed 6.2 cm apart from each other.
Experiments
3.1 Single position measurement 21
In this section, the results obtained from the experiment are looked at. The first
part of the experiment involved analysing the received signals to distinguish the
defect and the pipe end echoes.
Figure 3.6: Group velocity dispersion curve, for torsional mode T(0,m), for a steel
pipe with an external diameter of 35 mm. This graph was produced by running a
MATLAB script which is written by Zaghari (2012).
Using the group velocity, 3309 ms−1 , for torsional mode T(0,1) and the distances
of defect and pipe end walls it was possible to estimate the arrival of each echo. A
schematic representation of the expected path line of wave propagation is shown in
figure 3.7. The diagram shows a burst signal that is transmitted in both directions by
the transmitting transducer. The signal transmitted to the right is first received by
the receiving transducer known as direct signal, whereas the signal transmitted to
the left hits the pipe end and is reflected back to be received by the transducer. The
burst signals propagate along the pipe till the energy dissipates.
22 Experiments
Figure 3.7: Schematic representation of the expected path line for wave propagation
along the bare pipe containing a defect. The burst signal is transmitted in both
directions of the pipe.
Figure 3.8, shows the signal obtained when the guided waves were transmitted
at 50 kHz. This was analysed to understand what feature each echo represents.
Table 3.1 shows the expected arrival time of each echo calculated using speed,
3309 ms−1 (0.3309 cm(µsec)−1 ) and the actual arrival time, the instance at which
the echo arrives. As shown in the table, the arrival of direct echo has the largest
difference with theoretical estimate, because the distance was short and the echo
arrived quickly making it difficult to measure the arrival time, whereas the other
echoes arrive close to the expected time. The other reasons for the difference
are discussed in the next section. The second echo due to the defect was not
distinguishable, due to it being too close to the previous echo.
The hypothesis for this experiment was that as the frequency at which guided
waves are transmitted increases, it is less likely the defect echo will be detected due
to high reverberation. As the time at which the defect echo arrives is known, it is
possible to compare how this echo varies when guided waves are transmitted at
different frequencies. Guided waves were transmitted, for a range of frequencies
varying from 30 kHz to 200 kHz, to assess how the defect echo varies (see Figure
3.9).
As shown in figure 3.9, it was possible to detect the defect echo when guided
waves were transmitted at 50 kHz. When transmitted at 100 kHz the defect echo
is present but it is difficult to distinguish and identify, whereas at frequency of 50
kHz the pipe end and defect echoes are clearly visible and distinguishable. The
3.1 Single position measurement 23
reason that the defect echo was not visible when waves were transmitted at higher
frequencies is because high frequencies produce very strong reverberation which
causes the defect echo to be lost, while at low frequencies (30 kHz) there is high
noise due to which the defect echo cannot be distinguished.
Figure 3.8: Received signal, showing different echoes, for guided waves transmitted
at 50 kHz and the transducers placed at a zero degree configuration.
Table 3.1: Comparison between theoretical and experimental arrival times of the
echoes, in a received signal.
24
Figure 3.9: Amplitude of received signal, for guided waves transmitted at different frequencies, when a single reading at zero degree
configuration was taken. The time interval between the black dotted lines is analysed ahead for reverberation root mean square and the time
interval between the red dotted lines is used to calculate signal root mean square, which is used to calculate the signal to reverberation ratio.
Experiments
3.1 Single position measurement 25
Reverberation RMS (VRMS ) was calculated using the formula 3.1, where Nr is
the total number of readings taken, Vmean is the arithmetic mean of all the readings
and Vi is the original reading.
v
u
u 1 X Nr
VRMS = t (Vi − Vmean )2 (3.1)
Nr i=1
Figure 3.10: Variation in normalised reverberation RMS, for guided waves transmit-
ted at different frequencies, when a single reading at zero degree configuration was
taken.
As shown in figure 3.10, the reverberation RMS increases, as the guided waves
were transmitted at higher frequencies. It also shows that the reverberation RMS
was lowest for guided waves were transmitted at 50 kHz; hence the defect echo was
easily distinguishable.
26 Experiments
The hypothesis set before the experiment was proven correct and also the two
findings are related to each other. The reason it is difficult to detect the defect echo
is due to the fact that when the piezoelectric transducers are coupled onto the pipe
though the aim is to excite only torsional wave, through tangential displacement, it
also excites some longitudinal and flexural modes at the same time. This is more
clearly seen using the group velocity dispersion curve, figure 3.11, where it is visible
that other modes are also excited at the same time. For example it can be seen
that when T(0,1) mode is excited at 50 kHz, it also excites other modes like L(0,1),
F(1,1), F(1,2), F(2,1) and F(3,1). Similarly, as the transmission frequency increases
there are more non-axially symmetric modes that get excited as a result of which the
received waveform displays signals from other modes that interfere with the defect
signal, thereby making the defect echo less easily identifiable and distinguishable.
Figure 3.9 shows, that the arrival of defect and pipe end echoes is not the same for
all frequencies. This is due to there being a mode change when the guided waves
hits the pipe end or defect, thereby resulting in variation in arrival speeds.
Reverberation occurs due to reflection of part of the transmitted wave that is then
received by the transducer, thereby making it difficult to detect the defect. It occurs
due to both axisymmetric and non-axisymmetric modes. Reverberation is different
from noise. If noise is a limiting factor then the signal can be transmitted at higher
level, thereby increasing transmission echo and not noise; but if reverberation is a
limiting factor then increasing the signal will increase reverberation. Reverberation
increases with frequency and it determines the frequency the test can be performed
at, as shown in figure 3.9, whereas frequency determines the smallest defect that
can be detected by the guided wave.
3.1 Single position measurement 27
Figure 3.11: Group velocity dispersion for a steel pipe with an external diameter of
34 mm and wall thickness of 5 mm. This graph is produced from PCDISP MATLAB
scripts written by Seco and Jimenez (2012).
It is known that guided waves transmitted at high frequencies are more sensitive
to smaller defects than waves transmitted at low frequencies. The problem occurs
that the defect echo gets lost in reverberation, when waves are transmitted at high
frequencies using one transmitting and receiving transducer. In order to be able to
clearly distinguish the defect echo this reverberation must be reduced.
It was mentioned earlier, in section 2.5, that increasing the number of transducers
enable to suppress the additional unwanted modes that get excited alongside the
torsional mode; thereby reducing the reverberation level when guided waves are
transmitted at high frequencies. Therefore the second experiment focuses on rotat-
ing the receiving transducer around the circumference of the pipe and averaging the
signal to verify if reverberation reduces at high frequencies, hence making it easier
to detect the defect echo.
28 Experiments
3.2.1 Introduction
The aim of performing this experiment was to try to reduce the reverberation
level that occurs when guided waves are transmitted at high frequencies. Alleyne
and Cawley stated that it is possible to suppress the non-axisymmetric modes if
array of transducers are attached around the circumference of the pipe [12]. This
experiment was performed to test whether the theory is valid in practical application,
and if it reduces reverberation, thereby enabling ultrasonic signals to be transmitted
at higher frequencies. The aim was to be able to get a defect echo that can be
clearly distinguished and identified, when transmitting waves at 100 kHz.
The spacing between the transducers must be equal to or less than half the
wavelength of mode that is supposed to be suppressed. In this case the aim was
to try to eliminate all the modes that have a lower group velocity than T(0,1) when
transmitted at 100 kHz. According to suggestions by Alleyne and Cawley, a minimum
of 7 transducers were required around the circumference of the pipe in order to be
able to detect the defect echo when guided waves are transmitted at 100 kHz [12]
(refer to Appendix B). One of the major problems is coupling between the receiving
transducer and the pipe, as it is difficult to produce same coupling each time the
receiving transducer is rotated. There is also possibility of errors in measurement of
angles, when rotating the transducer, as this was measured by finding the equivalent
distance around the circumference using tape measure. Hence the readings for
all frequencies were taken at the position of transducer, before rotating it, to keep
results consistent. The other approach to performing this experiment is to have a
collar of transmitting and receiving transducer around the circumference of the pipe,
3.2 Multiple circumferential position measurements 29
see figure 3.13, this method was not adopted due to budget constraints.
Figure 3.12: Rotation of receiver transducer around the circumference of the pipe.
Figure 3.15 shows the results obtained, and it is visible that the pipe end and de-
fect echoes are more clearly identifiable when multiple readings were taken around
the pipe rather than a single reading. The reason for this improvement is by taking
readings around the circumference of the pipe and averaging the results it reduces
the reverberation caused by non-axially symmetric modes, flexural modes, and it
also averages out noise.
Figure 3.14: The first graph shows the received signals for 8 evenly spaced positions of the receiving transducer around the circumference of
the pipe, for guided waves transmitted at 100 kHz, whereas the second graph shows the average signal of these 8 waveforms.
31
32
Figure 3.15: Amplitude of average received signal, for guided waves transmitted at different frequencies, when 8 evenly spaced readings are
taken circumferentially around the pipe. The time interval between the black dotted lines is then analysed ahead for reverberation RMS and
the time interval between the red dotted lines is used to calculate signal RMS, which is used to calculate the signal to reverberation ratio.
Experiments
3.2 Multiple circumferential position measurements 33
Figure 3.16: Variation in ratio between the relative reverberation RMS, for an average
of eight evenly spaced readings around the circumference of the pipe and a single
reading at zero degree configuration, when guided waves were transmitted at
different frequencies.
Figure 3.16, shows how the ratio of relative reverberation RMS, between average
of eight evenly spaced readings around the circumference of the pipe and a single
reading at zero degree configuration, varies for guided waves transmitted at differ-
ent frequencies. As shown from the graph, reverberation RMS reduces for all the
frequencies when more readings are taken around the circumference of the pipe, as
the ratio is less than 1. The most improvement occurs for waves transmitted at 100
kHz, as expected. The improvement at 100 kHz occurs not only against noise, but
also against reverberation by elimination of the non-axisymmetric modes. If there
was improvement only against noise then the ratio is expected to follow formula
3.2, where Vrev,RMS is the reverberation RMS for waveform produced from single
reading or ’k’ that is the average waveform of the multiple readings taken, and Nr is
the number of readings.
p
Vrev,RMS,k ∗ Nr = Nr ∗ Vrev,RMS,1 (3.2)
After this the signal RMS was calculated for the time interval of the defect echo.
These two RMS were used to calculate the Signal to Reverberation Ratio (SRR).
SRR is the ratio between signal RMS (Vs,RMS ) and the reverberation RMS (Vrev,RMS )
and is calculated in dB using the formula below -
Vs,RMS
SRR = 20 log10 ( ) (3.3)
Vrev,RMS
The aim is to have a resulting waveform with high SRR, as it would mean that
the received waveform has more useful information rather than reflected signal. A
waveform with high SRR would enable the user to distinguish and clearly identify
the defect, as it would not be lost in reverberation.
Figure 3.17: Comparing how the signal to reverberation ratio varies, for guided
waves transmitted at different frequencies, when a single reading at zero degree
configuration and when eight evenly spaced measurements are taken around the
circumference of the pipe.
As shown in figure 3.17, SRR improved at all frequencies when multiple readings
were performed, but the most significant improvement was at 50 kHz and 100 kHz.
This improvement at 100 kHz was due to reduction in reverberation that enabled the
defect echo to be distinguished. The other interesting point is that at frequencies
150 and 200 kHz, for single measurement, despite the SRR being better than at 50
kHz the defect was not visible. This could be because of scattering of signals due to
high energy and also because of high reverberation due to which the defect echo
gets lost and is not clearly distinguishable.
3.2 Multiple circumferential position measurements 35
In the second part of this experiment it was looked at how the output waveforms
vary with the position of receiver transducer, for the guided waves transmitted at 100
kHz. The problem with this part of the experiment was that all the readings were
performed at different times; hence there might be variation in coupling between the
transducer and the pipe. As can be seen from figure 3.18, the defect echo was not
visible when the receiving transducer was placed at zero degree configuration or
below the transmitting transducer due to high reverberation whereas it was clearly
distinguishable when 8 or 16 readings were taken around the circumference of the
pipe. It is also noticeable that the waveform produced when 16 readings were taken
around the pipe has echoes that are clearer and have larger amplitude making them
easily distinguishable compared to the waveform produced for 8 readings.
Normalised reverberation RMS was calculated for these waveforms using formula
3.1, so that it can be compared how reverberation varies with the transducer position
and the number of readings taken. It was calculated for the same time interval, as in
previous cases, between the pipe end echo and the defect echo. The waveforms
were normalised, where the peak to peak amplitude was made to unity so that the
comparison is consistent. Figure 3.19, shows the variation in reverberation RMS
with the number of readings taken around the circumference of the pipe. The figure
shows that when more readings were taken around the circumference of the pipe,
the reverberation RMS was lower. It is also noticeable that the reverberation RMS
does not improve vastly when the number of readings around the circumference of
the pipe increases from 8 to 16. This is because there is no reduction in reverbera-
tion due to non-axisymmetric modes and the slight improvement is against noise.
The reverberation RMS was higher when the receiving transducer was placed below
the transmitting transducer than at zero degree configuration, this is even visible
when looked at figure 3.18, due to which the defect echo was not identifiable.
Lastly, SRR was calculated for all waveforms following the same procedure as
in previous experiment and using formula 3.3. The higher the SRR the better the
chances of detecting the defect echo. As shown in figure 3.20, taking multiple
readings around the circumference of the pipe improves the SRR. Thereby demon-
strating that the probability of detecting defect echo is higher when more readings
are taken around the circumference of the pipe. It can be seen that the SRR doubles
from approximately 3.5 dB to 7 dB, when the number of readings increases from 8
to 16. The reason why the SRR is better when 16 readings were taken is, because
there was an improvement in signal RMS as the amplitude of the defect echo is
greater than when 8 readings were taken. The SRR was very low and almost similar
when only one reading was taken, either at zero degree configuration or when the
receiving transducer is placed below the transmitting transducer.
36
Figure 3.18: Comparing how resulting waveforms vary, for guided waves transmitted at 100 kHz, with variation in transducer positions. It
shows, the output waveforms obtained when the transmitting and receiving transducer are placed at a zero degree configuration, when
the receiving transducer is placed below the transmitting transducer, and average of 8 (at 45 ◦ each) and 16 (at 22.5 ◦ each) readings taken
around the circumference of the pipe.
Experiments
3.2 Multiple circumferential position measurements 37
Figure 3.19: Comparing how normalised reverberation RMS varies, for guided waves
transmitted at 100 kHz, when a single reading at zero degree configuration, when
the receiver transducer is placed below the transmitting transducer and when eight
and sixteen evenly spaced readings are taken around the circumference of the pipe.
Figure 3.20: Comparing how signal to reverberation ratio varies, for guided waves
transmitted at 100 kHz, when a single reading at zero degree configuration, when
the receiver transducer is placed below the transmitting transducer and when eight
and sixteen evenly spaced readings are taken around the circumference of the pipe.
38 Experiments
The hypothesis set before the experiment has been proven correct by the results
obtained. The results show that the reverberation decreases if more readings are
taken around the circumference of the pipe.
The other part of the experiment involved analysing the waveforms obtained, for
waves transmitted at frequency of 100 kHz, at four different transducer positions.
3.2 Multiple circumferential position measurements 39
The results prove that it is possible to detect defect echo more clearly if there are
more transducers around the circumference of the pipe or if a transducer is reposi-
tioned physically over the circumference of the pipe and multiple readings taken.
It can be seen that when there was only one reading taken for waves transmitted
at 100 kHz, be it the case with zero degree configuration or receiver placed below
transmitter, the reverberation RMS is very high resulting in low SRR and thereby
resulting in defect echo getting lost in reverberation. Whereas, when there were
8 or 16 readings taken around the pipe then the reverberation RMS was low, but
still not zero because axisymmetric modes cannot be eliminated (see Table 3.2).
Due to this the SRR is high and the defect echo visible, when multiple readings are
taken around the circumference of the pipe. There is a possibility there might be
imperfections in coupling, when rotating the transducer, thereby resulting in some
non-axisymmetric modes still being present.
Table 3.2: Analysis of signals obtained when 8 and 16 readings are taken around
the circumference of the pipe, for guided waves transmitted at 100 kHz.
In the next section another important experiment was performed to see how
guided wave propagate through bitumen coated pipes. This is an important ex-
periment as most pipes are coated externally and sometimes internally in order to
prevent corrosion, increase flow of gas and reduce energy costs in pumping and
compressor stations [36].
40 Experiments
3.3.1 Introduction
The actual aim of this experiment was to perform tests on a corroded pipe to
determine if UGWT can be used to detect the presence of corrosion on pipes. The
company that was supplying the pipe, sent a pipe that was coated with bitumen
instead of corroded pipe.
Due to time constraint, the experiment was performed on the bitumen coated
pipe. The aim was to understand the propagation of waves through it and to see
if the waves attenuate as strongly, as mentioned in theory. This is also a useful
experiment as bitumen has been used to protect pipes from corrosion for over 40
years [37]. Therefore most old pipes are coated with bitumen and in order to inspect
these pipes for defects, it is important to know how guided waves perform when the
pipe is coated with bitumen.
As shown by table 2.2, guided waves propagate very limited distance in a bitumen
coated pipe. The hypothesis set for this experiment was that the bitumen coating
will strongly attenuate the waves and hence the pipe end echoes5 will be less visible
compared to a bare pipe and the amplitude of echoes is also expected to be smaller
in this case.
5
This pipe had no defect so there was no defect echo, as with previous cases.
6
Bitumen coating was only removed for the area where the transducers were placed and not all
the way around the pipe.
3.3 Bitumen coated pipe 41
Figure 3.21: Bitumen coated pipe along with the area for which the coating was
removed (the black region) to place the transducers.
Figure 3.22: Transducers (receiving at right and transmitting at left) placed on the
region of pipe where bitumen coating has been removed. It was only removed for
the top section of the pipe and not all the way around the pipe.
Figure 3.24: Setup for the experiment showing two transducers placed on the bitumen coated pipe. The transmitting and receiving transducers
here are placed 6.2 cm apart from each other and the bitumen coating is removed so that the transducers can couple on the pipe.
Experiments
3.3 Bitumen coated pipe 43
As shown in figure 3.25, guided waves respond very poorly when transmitted
through bitumen coated pipe, they attenuate very highly and as a result dissipate
quickly. It is visible that at lower frequencies, i.e. 30 kHz and to an extent 50 kHz,
there is some aspects of pipe end echo visible, whereas at higher frequencies the
only visible echo is due to direct transmission. This could be because as known low
frequency waves travel further than high frequency waves. Another experiment was
performed with an input burst applied at Vpk−pk of 4.75 V to investigate if transmitting
the input signal at a lower strength has an affect on wave propagation through the
pipe. The results obtained were very similar to figure 3.25, but the only difference
was that the echo had a lower amplitude. The results show that propagation of
waves is not affected by the strength at which input burst is transmitted at.
The amplitude of direct echo produced for guided waves transmitted at different
frequencies was compared for the input tone burst transmitted at Vpk−pk of 4.75 V
and 19 V through a bitumen coated pipe and when the input burst was transmitted
at Vpk−pk of 19 V through a bare pipe. The distance between the transmitter and
receiving transducer was 6.2 cm in all cases, to ensure results are consistent.
Figure 3.26 shows how the amplitude of direct echo varies in each case. The graph
shows that there was a similar trend for all the experiments and that the amplitude
at 50 kHz is lower than all other frequencies. This is because the reverberation
is low for a single measurement, when the guided waves are transmitted at a
frequency of 50 kHz and hence there is less disturbance. It also shows that the
received signal for bitumen coated pipe has a lower amplitude compared to results
from the bare pipe, because of high attenuation caused by bitumen coating. The
results display that when the tone burst was transmitted at a Vpk−pk of 4.75 V it
produced amplitude almost 4 times smaller than the amplitude produced for tone
burst transmitted at Vpk−pk of 19 V, as it has 4 times more energy than at 4.75 V.
Therefore, the amplitude of the direct echo is proportional to the voltage the tone
burst is transmitted at through a bitumen coated pipe.
7
In the previous experiments also, the tone burst was applied at Vpk−pk of 19 V.
44
Figure 3.25: Comparing how the amplitude of received signal varies in a bitumen coated pipe, for guided waves transmitted at different
frequencies (burst applied at Vpk−pk of 19 V), when a single reading at zero degree configuration was taken. The time interval between the
black dotted lines is then analysed ahead for reverberation RMS.
Experiments
3.3 Bitumen coated pipe 45
Figure 3.26: Variation in amplitude of direct echo in non-coated and bitumen coated
pipe. The graphs shows results from normal pipe where the input burst was trans-
mitted at Vpk−pk of 19 V and from bitumen coated pipe where the input tone burst
was transmitted at Vpk−pk of 19 V and 4.75 V.
To analyse how the reverberation varies with frequency in a bitumen coated pipe,
reverberation RMS was calculated. The signals were normalised first to ensure
comparison can be made and then reverberation RMS was calculated for the time
interval between the direct echo and the pipe end echo, using formula 3.1. Figure
3.27, displays how the reverberation RMS varies with frequency when the tone burst
was transmitted at Vpk−pk of 19 V and 4.75 V. To ensure reliable results the readings
for both scenarios were taken for the same position of transducers, so that coupling
between the pipe and transducer stays consistent, when transmitting tone burst at
both voltages.
The results demonstrate that reverberation RMS is almost same for the tone
bursts transmitted at Vpk−pk 19 V and 4.75 V. This implies that the reverberation
does not depend on the energy of tone bursts, but rather on the frequency at which
waves are transmitted at. It is visible that the results are consistent with the previous
experiments, as the reverberation RMS was lowest at 50 kHz for a single point
measurement. This suggests that the trend in which reverberation varies across
frequencies does not depend on whether the pipe is coated with bitumen or not,
though the relative values could be different.
46 Experiments
• Reverberation RMS does not depend on the voltage of the tone burst and only
depends on the transmission frequency of the guided waves.
The first finding is one of the reasons why UGWT is considered poor for detecting
defects in bitumen coated pipes. This is a disadvantage because high wave attenu-
ation limits the defect detection sensitivity and monitoring range of the technology.
When pipes are coated with a viscoelastic material, it causes the energy of guided
waves to leak into the coated material from the pipe wall. The rate of leakage
is determined by the strength of the coating. Bitumen being a strongly adhesive
material binds firmly to the pipe, thereby resulting in leakage into the coating and
3.3 Bitumen coated pipe 47
causing significantly high attenuation of guided wave [38]. This high attenuation,
thereby results in limitation to the distance the guided waves can propagate. Bitu-
men coated pipes attenuate guided waves as the pipe is a bi-layer system and any
energy carried in bitumen layer is attenuated quickly, whereas in a corroded pipe the
test ranges is less as energy is scattered by the corroded surface [30]. It is known
from previous research that insulation coatings that have strong adhesive strength,
such as bitumen, cause high attenuation of guided waves whereas mineral wool
coating almost has no effect as it does not bond strongly with the pipe [39]. It must
also be remembered that the thickness and age of coating also have an effect. The
other factor that influences attenuation is the frequency at which guided waves are
transmitted and as the frequency increases the attenuation increases (see Figure
3.25). This is because the wavelength of the wave is shorter, thereby dissipating
easily into the bitumen coating [38].
The next findings are related to the amplitude and reverberation of waveforms in
a bitumen coated pipe. This experiment showed that the amplitude of the direct echo
varies in a similar manner across the frequencies and it is lower for bitumen coated
pipe, due to high reverberation and absorption by the viscous coating when com-
pared with the bare pipe for the input burst transmitted at Vpk−pk of 19 V. Whereas,
when input burst was transmitted at Vpk−pk of 4.75 V the amplitude is lower, because
the input burst was transmitted at lower energy, but the trend in which the amplitude
varies across frequencies is similar.
Most pipelines are coated with bitumen or with other viscoelastic material for
protection purposes, and in practice most pipelines are buried underground. Similar
to insulation coating on a pipe, soil also has an influence on the propagation of
waves along the pipe. In the next experiment performed propagation of guided
waves through a buried pipe was looked at.
48 Experiments
3.4.1 Introduction
The aim of performing this experiment was to assess how guided waves perform,
in a practical scenario, where pipes are buried in soil. UGWT is preferred over
convectional ultrasonic when inspecting buried pipes, because it allows for screen-
ing of pipe by placing transducers at a single convenient location thereby requiring
excavation in only one area.
Figure 3.28: Dimensions of the box through which the pipe goes through. This was
then filled with 0.0324 m3 of sand.
3.4 Buried Pipe 49
Figure 3.29: Part of pipe along with the defect buried in the sand box. The box
contained 50 kg of sand.
Figure 3.31: Setup for the experiment showing two transducers placed on the pipe. The transmitting and receiving transducers here are
placed 6.2 cm apart from each other and it has a defect buried in sand.
Experiments
3.4 Buried Pipe 51
Figure 3.32, shows the waveforms obtained when the guided wave were trans-
mitted through a buried pipe. As shown in the figure, guided waves propagate better
through a buried pipe than through a bitumen coated pipe. It also shows that the
defect echo is visible at high frequencies as it can be identified clearly at 50 kHz
and 100 kHz, whereas at 150 and 200 kHz it is present but not clearly identifiable
due to reverberation. The waveforms produced for the buried pipe were very similar
to the waveforms obtained for the non-buried pipe (see Figure 3.15). This was a
surprising result as it was expected the buried pipe to have a higher reverberation
than non-buried pipe.
After calculating reverberation RMS, signal RMS was calculated for the time
interval of the defect echo. The two RMS was then used to calculate SRR, using
formula 3.3. Figure 3.34 shows how the SRR varies with transmission frequency
of guided waves for buried and non-buried pipe. As shown in figure, when guided
waves were transmitted at low frequencies the SRR was high for non-buried pipe;
this is linked to low reverberation RMS for non-buried pipe. When transmitted at 100
kHz, the results were similar for both scenarios and for high frequency waves the
buried pipe has a better SRR than non-buried pipe, because at high frequencies the
reverberation is lower for the buried pipe.
52
Figure 3.32: Amplitude of average received signal, for guided waves transmitted at different frequencies through the buried pipe, when eight
evenly spaced readings are taken circumferentially around the pipe. The area between the black dotted lines is then analysed ahead for
reverberation root mean square and the area between the red dotted line is used to calculate signal root mean square, which is used to
calculate the signal to reverberation ratio.
Experiments
3.4 Buried Pipe 53
Figure 3.33: Comparing how the average normalised reverberation RMS varies, for
guided waves transmitted at different frequencies, when the pipe is buried and not
buried.
Figure 3.34: Comparing how SRR varies, for guided waves transmitted at different
frequencies, when the pipe is buried and not buried.
54 Experiments
Figure 3.35: Variation in ratio between peak amplitude for a buried and non-buried
pipe, when guided waves are transmitted at different frequencies.
In order to compare how the peak amplitude varies in the two scenarios, ratio
between the peak amplitude of buried pipe and non-buried pipe was calculated for
the average waveforms of eight readings taken around the circumference of the pipe.
Figure 3.35, shows how the ratio of peak amplitude varies with frequency. As can be
seen from the figure, the amplitude ratio is almost one when the guided waves were
transmitted at 100 kHz. This result was expected as the reverberation RMS and
SRR were almost the same for both cases at that frequency. As 8 readings were
taken around the pipe it mitigates all the non-axisymmetric modes that get trans-
mitted at 100 kHz. At lower and higher frequencies the amplitude ratio is less than
one, as the peak amplitude for a buried pipe is lower than when the pipe is not buried.
3.4 Buried Pipe 55
• Transmission of guided waves at 100 kHz produces similar results for both bur-
ied and non-buried pipes, when 8 readings are taken around the circumference
of the pipe.
The first finding is a plus point of UGWT and one of the reasons it is used by
petrochemical industries, to detect defects, as most pipes are buried underground.
The results produced show that sand does not bind with the pipe as strongly as thick
bitumen coating does; therefore there is less leakage of signal. Though it must be
noted attenuation of guided wave depends on soil type - clay, silt, sharp sand, soft
sand etc. and the depth of soil. Previous researchers have shown that some soil
types cause very high attenuation [40–42].
This experiment demonstrated that reverberation is less for a buried pipe than
a non-buried pipe, when guided waves are transmitted at high frequencies. This
results in high SRR at higher frequencies, as more useful signal is obtained. Re-
verberation for buried pipe is low at high frequencies, because the extra modes are
suppressed by the sand. The effect of low reverberation at high frequencies is the
reasons why at high frequencies the peak amplitude is lower, than the non-buried
pipe case, as there is less disturbance in the received signal. It can also be seen
that when the signal is transmitted at 100 kHz, for 8 readings taken around the cir-
cumference of the pipe, the reverberation RMS and SRR is same for both scenarios
and the amplitude of peak waveform is almost equal in both cases. The reason
for similar result at 100 kHz is because when eight readings are taken around the
circumference of the pipe it eliminates the reverberation, due to non-axisymmetric
modes and thereby gives a similar waveform in both cases where the defect echo
can be identified. It shows that taking multiple readings around the circumference of
56 Experiments
From this experiment it can be concluded that UGWT performs well when testing
for a defect in buried pipes and it does not produce a high attenuation. Though
there are other factors also that determine how well the guided waves perform when
testing through a buried pipe and these must be taken into account, such as the
type of soil the pipe is buried in and how adhesive the soil is to the pipe. If the soil
is strongly adhesive then guided waves would perform poorly, because the waves
would get leaked to the soil.
3.5 Finding attenuation of the bare pipe 57
Conclusion
In this dissertation, ultrasonic guided wave technology was studied and the
propagation of these waves was assessed in different scenarios. Experiments were
performed to investigate at what frequency the guided waves must be transmitted
to detect the defect located in the pipe, how rotating the receiver around the pipe
affects the received signal and defect detection. Tests were even performed to
understand how the technology responds, in industrial scenarios, when the pipe is
coated with bitumen and when the pipe is buried in sand.
The experimental tests were conducted on a steel pipe with a diameter of 3.4 cm
a wall thickness of 5 mm. A defect of 8.3% CSA in size was machined onto the pipe
to investigate at what frequency the guided waves must be transmitted to detect
the defect, it was detected at 50 kHz. At higher frequencies the defect echo was
not visible, as the echo was lost in the reverberation when two transducers were
used. In order to decrease the reverberation level at higher frequencies, it required
increasing the number of transmitting and receiving transducers to suppress the
non-axisymmetric modes. Due to budget limitation it was not feasible to purchase
many transducers. Therefore receiver transducer was physically rotated around
the circumference of the pipe and readings taken, that allowed the defect echo to
be detected at higher frequencies. In averaging the signal, the non-axisymmetric
modes were eliminated, due to phase difference when the receiving transducer is
rotated; hence leaving a resultant signal with low reverberation level.
The next experiment involved testing the technology on a bitumen coated pipe
where results were very poor as the guided waves attenuated highly when transmit-
ted through a bitumen coated pipe; hence the distance the guided waves propagated
was very limited. The other experiment was performed on a buried pipe where
results were encouraging as the defect that was buried in sand was detected by the
60 Conclusion
technology. This was possible because the sand on the outside does not bind to the
pipe as strongly as bitumen coating, thereby there is less leakage of energy.
If there is a higher budget available then an experiment should be carried out with
a transducer that is capable of transmitting and receiving signals, rather than using
two transducers. Also, instead of the transmitting guided waves in both directions of
the pipe it must be looked if it is possible to transmit guided wave in one direction so
that the transducer can be placed close to pipe end. This would reduce the number
of echoes received, making it easier to identify the defect echo. The other change is
to increase the number of transducers used around the circumference of the pipe.
This would help to reduce reverberation and allow the defect echo to be detected at
high frequencies as high frequency waves are more sensitive to small defects.
The other experiment that can be performed is using a chirp as an input signal,
instead of tone burst. A tone burst is a sinusoidal wave with limited cycles, whereas
chirp is a signal where frequency increases or decreases with time. A chirp signal is
preferred as it might help to reduce reverberation, due to it producing a frequency
sweep. In order to get a good resolution, short acoustic pulses must be transmitted;
the signal energy is proportional to the signal pulse duration and to get a good SNR
the pulses must be longer [43]. The solution to achieving the balance between single
echo resolution and SNR performance is to use a chirp instead of conventional tone
burst [43].
Lastly, guided waves can also be transmitted through a corroded pipe to investig-
ate how guided waves respond and whether they can detect the corrosion on the
pipe. The corrosion can also be introduced electrochemically onto the pipe. Guided
waves can even be propagated through pipes that are coated with other materials
such as epoxy, wax, polyethylene, etc. to see how the technology respond in those
cases.
References
[2] V. Varma, “State of the art natural gas pipe inspection,” u. s. department of
energy report, Oak Ridge National Laboratory, Oak Ridge, Oak Ridge, TN
37831, Sept. 2002.
[5] T.H.Gan, “The application of guided wave ultrasonic technique for the inspection
of corrosion under insulation (cui) of pipes and pipelines.” Petromin Pipeliner,
Sept. 2010.
[6] J.W.S. Rayleigh, “The theory of sound.” Dover publications, Inc., Newyork, N. Y.
10014, 1945.
[11] M. Silk and K. Bainton, “The propagation in metal tubing of ultrasonic wave
modes equivalent to lamb waves,” Ultrasonic International, vol. 17, pp. 11–19,
1979.
[12] D. N. Alleyne and P. Cawley, “The excitation of lamb waves in pipes using
dry-coupled piezoelectric transducers,” Journal of Nondestructive Evaluation,
vol. 15, pp. 11–20, 1996.
[13] M. Lowe, D. Alleyne, and P. Cawley, “Defect detection in pipes using guided
waves,” Ultrasonic International, vol. 36, pp. 147–154, 1998.
[14] D. Alleyne, B. Pavlakovic, M. Lowe, and P. Cawley, “Rapid, long range inspection
of chemical plant pipework using guided waves,” Review of progress in quantitve
nondestructive evaluation, vol. 20, pp. 180–187, 2000.
[15] J.Davies, Inspection Of Pipes Using Low Frequency Focused Guided Waves.
PhD thesis, Imperial College, London, 2008.
[16] D.Sonia and B.Fouad, “Propagation of guided waves in a hollow circular cylinder
application to non destructive testing,” in 14th International Conference on
Experimental Mechanics, (Poitiers, France), June. 2010.
[19] J. Davies and P. Cawley, “The application of synthetic focusing for imaging
crack-like defects in pipelines using guided waves,” IEEE Transactions on
Ultrasonics, Ferroelectrics, and Frequency Control, vol. 56, p. 760, 2009.
[21] A. Lvstad and P. Cawley, “The reflection of the fundamental torsional guided
wave from multiple circular holes in pipes,” NDT&E International, vol. 44,
pp. 553–555, 2011.
[22] B.E. Paton, V.A. Troitskii and A.I. Bondarenko (2010), “Method and technology
of diagonstics of pipeline systems by guided low-frequency ultrasonic waves.”
Guided NDE LLC. http://www.ndt.net/article/ecndt2010/reports/1_03_
38.pdf, [PDF], [Accessed 10 Oct. 2012].
REFERENCES 63
[24] Z. Liu, C. He, B.Wu, X.Wang, and S. Yang, “Circumferential and longitudinal de-
fect detection using t(0,1) mode excited by thickness shear mode piezoelectric
elements,” Ultrasonics, vol. 44, pp. e1135–e1138, 2006.
[27] G. K. Sohi, Study of Rayleigh Lamb Wave Propagation in elastic and micropolar
solids. PhD thesis, Thapar University, Punjab India, 2012.
[28] M. Lowe (2011), “The guided wave testing method for screening of pipelines.”
V Pan American Conference for Non Destructive Testing Cancun. http:
//www.ndt.net/article/panndt2011/presentations/143_Lowe.pdf, [PDF],
[Accessed 18 Nov. 2012].
[29] T.M.Webb (2011), “Benefits and challenges of ultrasonic guided wave inspec-
tion.” 2011 API Inspection Summit and Expo. http://new.api.org/meetings/
proceedings/2011inspectionsummit.cfm?renderforprint=1, [PPT], [Ac-
cessed 18 Nov. 2012].
[30] M. Lowe and P.Crawley, “Long range guided wave inspection usage current
commercial capabilities and research directions,” tech. rep., Department of
Mechanical Engineering Imperial College London, London UK, 2006.
[31] J. L. Rose, “A baseline and vision of ultrasonic guided wave inspection potential,”
Journal of Pressure Vessel Technology, vol. 124, pp. 273–277, 2002.
[36] Pipelines International (2010), “Inside and out: benefits of pipeline coating.”
Pipelines International - December 2010. http://pipelinesinternational.
com/news/inside_and_out_benefits_of_pipeline_coating/053589/, [On-
line], [Accessed 21 March 2013].
[37] F. Simonetti and P. Cawley, “On the nature of shear horizontal wave propagation
in elastic plates coated with viscoelastic materials,” The Royal Society.
[39] J. wen Cheng, The Attenuation of Guided Wave Propagation on the Pipelines.
PhD thesis, National Sun Yat-Sen University, China - Taiwan, 2006.
[41] J. Rose, J. Mu, J. Hua, R. Royer, and S. Kenefick, “Ultrasonic guided wave
phased array inspection of coated and buried pipe (part1: theory),” in JRC-NDE
2009 - Developments in Ultrasonic Guided Wave Inspection, (-), June 2010.
[42] J. Rose, J. Mu, J. Hua, R. Royer, and S. Kenefick, “Ultrasonic guided wave
phased array inspection of coated and buried pipe (part2: experiment),” in
JRC-NDE 2009 - Developments in Ultrasonic Guided Wave Inspection, (-),
June 2010.
[44] D. J. Chinn, M. J. Quarry, and J. L. Rose, “Group velocity and phase velocity,”
tech. rep., Lawrence Livermore National Laboratory, Livermore, CA, 2005.
[45] M. S. Santhanam (2009), “The guided wave testing method for screening
of pipelines.” IISER, Pune/PRL, Ahmedabad. http://www.iiserpune.ac.in/
~santh/gvel.pdf, [PDF], [Accessed 27 Sept. 2012].
Appendices
Appendix A
∂(kCp ) ∂Cp
Cg = = Cp + k (A.3)
∂k ∂k
Group velocity is used in identifying and locating a defect, whereas phase velocity
is used in design of guided wave sensor for mode excitation. More details about the
concept and equations can be found in Ultrasonic waves in solid media by Rose [17].
This page is intentionally left blank.
Appendix B
The calculation performed here is for T(0,1) wave mode that is transmitted at 100
kHz along a pipe with an outer diameter of 34 mm. First the group velocity is found
using the dispersion curve. The group velocity for T(0,1) remains constant through-
out the frequency range at 3309 ms−1 . Using the group velocity and frequency the
wavelength of the wave is calculated using -
Cg = f × λ (B.1)
λ
△≤ (B.2)
2
The minimum number of transducers (Nt ) required is then calculated using
the formula below, where ’aext ’ is the external pipe radius which is 17 mm and the
spacing being 16.5 mm.
2πaext
Nt = (B.3)
△
Therefore the minimum number of piezoelectric transducers required to
propagate T(0,1) at 100 kHz in a 34 mm diameter pipe is 7. Similarly to propagate it
at 50 kHz we would require 4 piezoelectric transducers.
This page is intentionally left blank.
Appendix C
CD