Certificate of Calibration: Customer Information

This certificate summarizes the calibration of a pressure sensor used by Goodyear Dunlop Dudelange Tires Operations. The sensor was calibrated on February 16, 2021 using a multifunctional calibrator and found to be within its specified accuracy of 2.5 bar across its 0 to 1000 bar range. Test results showed differences between the reference values and instrument values between -1.7 and -1.3 bar, well within the sensor's accuracy tolerance.

Uploaded by

Sazzath Hossain
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
100% found this document useful (1 vote)
464 views2 pages

Certificate of Calibration: Customer Information

This certificate summarizes the calibration of a pressure sensor used by Goodyear Dunlop Dudelange Tires Operations. The sensor was calibrated on February 16, 2021 using a multifunctional calibrator and found to be within its specified accuracy of 2.5 bar across its 0 to 1000 bar range. Test results showed differences between the reference values and instrument values between -1.7 and -1.3 bar, well within the sensor's accuracy tolerance.

Uploaded by

Sazzath Hossain
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 2

CERTIFICATE OF CALIBRATION

2102-16060
Customer information
Client : Goodyear Dunlop Dudelange Tires Operations S.A.
Contact : Ahmet KOSE
Address : Zone Industrielle Riedgen
3451, Dudelange
Luxembourg

Reference Trescal : TP-37823

Instrument information
Make : GOODYEAR
Type : HMI
Description : Pressure Before GP
Serial number : #13L APEX IB Pressure Sensor
Identification number : 13LP10029
Range : 0 .. 1000 bar
Accuracy : 2.5bar.

Date of calibration : 16 February 2021

Used method
P1-02-E.001 Calibration of electrical measurement instruments
Electrical instruments are calibrated with a multifunctional calibrator and/or a digital multimeter.

Environmental conditions
During this service the following conditions were not exceeded :
Ambient temperature : 23°C ± 5°C
Relative humidity : 45%rh ± 20%rh

Used reference
The equipment used is traceable to National and/or International standards.
R1965 Hygro- / Thermometer cert n° 200119422
R4452 Multifunctional calibrator cert n° 200413969
Note
This service was performed at the clients location.
The instrument is measured but not adjusted, so the results are both 'as found' as 'as left'.

Technician Issue date 26 February 2021 Head of the laboratory


Gregory Gaspart Luc Van Pelt

o.t. Wim Van Camp


This document may not be reproduced other than in full, except with the prior written approval of the head of the issuing laboratory.
Unless otherwise stated, the calibration was performed at the address mentioned in the footnote.

Trescal nv | Vosstraat 200 | Berchem (Antwerpen) | Belgium | Telephone + 32 3 542 62 90 | Email: [email protected]

Page 1 to 2
CERTIFICATE OF CALIBRATION
2102-16060
Initial results:
- Range Reference Units Calculated Instrument Difference Uncertainty Units
value value value ±
1 0 tot 1000 bar 4,000 mA 0,0 0,0 0,0 0,4 bar
2 0 tot 1000 bar 8,000 mA 250,0 248,3 -1,7 0,4 bar
3 0 tot 1000 bar 12,000 mA 500,0 498,6 -1,4 0,4 bar
4 0 tot 1000 bar 16,000 mA 750,0 748,3 -1,7 0,4 bar
5 0 tot 1000 bar 20,000 mA 1000,0 998,7 -1,3 0,4 bar

The stated uncertainty is that of the entire set-up including the object under test.
The reported uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing a
level of confidence of approximately 95% .
The uncertainty is calculated following EA-4/02 in accordance with the requirements of the ISO/IEC 17025.

Page 2 to 2