Chemical Engineering Laboratory 1: Experiment 3: Discharge Coefficient (Orifice Meter)
Chemical Engineering Laboratory 1: Experiment 3: Discharge Coefficient (Orifice Meter)
CHE150-1L/B41
ABSTRACT
Two types of flow meters have been observed in this experiment; Area meter and head meter. Mass flow rate, rotameter
reading, and manometer reading are the axes used in generating a calibration curve for rotameter and orifice meter. It was
found that the curve of rotameter was more linear than orifice meter that has a square root relationship. Moreover, pressure
drop and discharge coefficient of the orifice was also calculated. Areas and velocities for both pipe and orifice were
computed for the needed values. Pressure drop was observed to be increasing with the mass flow rate. For pipe taps, the
pressure difference has been found to be really permanent loss and not orifice differential. However, discharge coefficient
had no trend with values 0.859, 0.863, 0.905, 0.817, 0.683 since the Reynolds number was large enough for it to be
independent of mass flow rate. The possible sources of error are the square root relationship itself and straight pipe that has
not been provided.
1. Introduction
should be in 30 seconds. The excess in time, whether it rotameter. The curve represents the relationship of the mass
seem to be small, should be added into account and cannot flow rate and rotameter reading which is directly
consider to be negligible. It is because this time was used in proportional. As shown, it has a positive slope but is not
calculating values that is essential to the experiment. quite linear which is unusual to the rotameter. If data from
table 3.2 was checked again, it can be observed that the
Mass of water, kg Time, s difference in the mass flow rate of trial 1 and 2 which is
3.5 30.16 0.027 was way greater that the difference of the rest which
4.3 30.03 just fall on the range of from 0.004-0.007.
4.5 30
4.7 30.04
4.85 30.24 Calibration Curve for Rotameter
Table 3.1 Noted and Calculated Data of Mass and 3.5
Time 3
reading were so close with an increment of just 0.5. Then, values of mass flow rate and manometer reading which will
with these values, a calibration curve was able to generate serve as the x- and y-axis. Manometer readings were the
with mass flow rate as the x-axis and rotameter reading as ones from the manometer recorded every trial.
0.2 Table 3.5 shows the values of the velocity of the pipe,
velocity of orifice, pressure drop, and the discharge
0
0.1 0.15 0.2 coefficient (left to right column). These are computed using
Mass Flow Rate (kg/s)
equations 3.2 – 3.5.
m1
U 1= (Equation 3.3)
Figure 3.2 Calibration Curve for Orifice
S1 ρ
√ 2 gc
1−β
( −∆ρ P )
4
(Equation 3.2)
References