General Specifications: Model SC4A Conductivity Sensors and Fittings For 2-Electrode Systems
General Specifications: Model SC4A Conductivity Sensors and Fittings For 2-Electrode Systems
FEATURES
• Compact style sensor
• Built-in temperature resistor: Pt1000
• Fast temperature response
• Plug and cable form a water tight connection to IP67
• Selection of two cell constant 0.02 cm-1 or 0.10 cm-1
• Sensor with 8 pin Variopin with ID chip for SENCOM SA11-C1 use
• Certified for hazardous area
• Direct process insertion
• Wide range of mountings
• Compatible with PR10 retractable assembly
• The sensor is polished to meet pharmaceutical requirements
• Standard quality inspection certificate with delivery of sensor
• Peek and O-rings according to FDA standard
System Configuration
GS 12D07J04-01EN-P
2nd Edition November, 2021
2
1. SPECIFICATIONS
General Specifications
Wetted parts adapters/fitting : /PS, /FF for SC4A-*-AD : Stainless Steel AISI 316L
/PF for SC4A-*-AD PVDF
/SA1(2), /SB1(2), /SC1 for
SC4A-E-SA (SB, SC) Stainless Steel AISI 316L
* for colors see Section 5.1, TPE-O for outer jacket, color black
Note 1 : Cable is not considered as wetted part
Note 2: The temperature sensor included in the sensor is designed for process compensation and for indication.
It is NOT designed for process temperature control.
GS 12D07J04-01EN-P
3
Temperature @ 100 KPa : 0 °C ... 125 °C (32 °F to 257 °F) - For VarioPin type sensors
(1 Bar, 14.5 PSI) 0 °C ... 100 °C (32 °F to 212 °F) - For fixed cable type sensors
Remark:
For all types 135°C (275 °F) max. for a short period of time during sterilization.
In combination with a fitting or option, the specification of the most critical part is leading.
GS 12D07J04-01EN-P
4
1.7 Shipping details
Package size (LxWxH) : 220 x 220 x 90 mm (8.7 x 8.7 x 3.6 inch) - Fixed cable version
300 x 100 x 75 mm (11.8 x 3.9 x 3.0 inch) - VarioPin type version
Package weight (max.) SC4A-*** connection type: -03/VS -05 -10 -15 -20
0.63 0.78 1.13 1.48 1.83 kg
1.4 1.7 2.5 3.0 4.0 lbs
GS 12D07J04-01EN-P
5
1.10 Regulatory compliances:
GS 12D07J04-01EN-P
6
2. SPECIFICATIONS SC4A-AD Dimensions SC4A-AD
General Specifications
90 (3.55")
177 .2415 FDA migration tested 21
CFR 177.2415 +, PIM and
152 (5.98")
152 (5.98")
USP class VI
Surface Roughness : 0.8µm (R432)
Ø 19,1
1)
2)
Suffix -AD-09 not available with -VS.
With option /FF: option /PS is mandatory.
Figure 4. - Mounted sensor with the option /
PF and /PS
GS 12D07J04-01EN-P
7
3/4 “NPT
1/4 “NPT
100 (3.94”)
1/4 “NPT
30(1.18”)
Figure 5.1. Dimensional drawing /FF with the Figure 5. Dimensional drawing /FF with the
option /PS option /PS
GS 12D07J04-01EN-P
8
3. SPECIFICATIONS SC4A-SA/SB/SC Dimensions SC4A-SA
General Specifications
95 (3.74")
Ø 25
95 (3.74")
(1.00")
Ø 24.9
3.1 Materials Wetted Parts
Body & electrodes : Stainless steel AISI 316L (1.4435)
O-ring : EPDM, FDA migration tested Ø 19.1
21 CFR 177.2600
Mounting adapter : Stainless steel AISI 316L Tri-clamp * units in mm (inch)
(1.00")
70 (2.75")
Ø 50.5
Ø 25
80
70
Ø 19.1 Ø 19.1
* units in mm (inch)
GS 12D07J04-01EN-P
9
3.4 Installation SC4A-SA/SB/SC
GS 12D07J04-01EN-P
10
4. SPECIFICATIONS SC4A-PR
General Specifications
Dimensions SC4A-PR
4.1 Materials Wetted Parts
Body & electrodes : Stainless steel AISI 316L
or Titanium grade 2 or 3
O-ring : Viton
Mounting adapter : Retractable fitting
Insulation :PEEK (Polyether Ether Ketone)
M19x1.5
4.2 Operating Specifications (Sensor)
See paragraph 1.6
48 (1.89")
52 (2.04)
(1.08)
41 (1.6)
the sensor is explained in the Instruction Manual of this fitting.
Ø 19.1
Part no. Description Quantity (0.75)
* units in mm (inch)
K1500BE O-ring set, Viton, 15.6 x 1.78 10
K1500ED O-ring set, Kalrez, 15.6 x 1.78 1 Figure 11. SC4A -PR
GS 12D07J04-01EN-P
11
5. ADDENDUM 1: Typical Installations
For optimum measurement results, the SC4A sensor should be installed in a location that offers an acceptable representation of the
process composition and DOES NOT exceed the specifications of the sensor.
Figure 12. Mounted sensor with option /PF and /PS Figure 13. Dimensional drawing /FF with
option /PS
5.2 Typical installation SC4A-E-SA (SB, SC) with standard options / spare parts
The typical installation of the SC4A-E-SA sensor is done by using a standardized 25 mm port. Stainless Steel welding sockets are
available as straight version (spare part /SA1) or 15° angled version (spare part /SA2). Both versions are delivered with mounting nut.
In Figure 14 an example is shown how to install the sensor using spare part /SA1. Sensors SC4A-E-SB and SC4A-E-SC are installed
by tri-clamp method, see Figure 15. For the SB version two Stainless Steel tri-clamps are available, in a 1” size (spare part /SB1) or in a
1½“ size (spare part /SB2). The SC version has just one Stainless Steel tri-clamp in a 2” size (spare part /SC1).
/SA1
51 (2.00")
55 (2.16")
Figure 14. Mounted SC4A-E-SA sensor with Figure 15. Mounted SC4A-E-SB(-SC) sensor
spare part/ SA1
GS 12D07J04-01EN-P
12
5.3 Typical installation SC4A-*-PR
The installation of the SC4A-*-PR sensor is done using the Model PR10 retractable fitting.
Figure 16. Dimensional drawing PR10...-D32 Figure 17. Dimensional drawing PR10...-D50
with mounted SC4A sensor with mounted SC4A sensor
GS 12D07J04-01EN-P
13
6. ADDENDUM 2: Available model codes
GS 12D07J04-01EN-P
14
7. ADDENDUM 3: Control drawings
7.10 FM-United States
Applying standards : FM Class 3600
FM Class 3610
FM Class 3810
ANSI/ISA 60079-0
ANSI/ISA 60079-11
Certificate no.* : FM20US0123X
IS CL I, DIV 1, GP ABCD, T4…T6 CL I, ZN 0, AEx ia IIC, T4…T6 Ga
Control Drawing: D&E 2020-024-A50
Electrical data : (See Note)
Specific conditions of use : See Control Drawing D&E 2020-024-A50.
Temperature classes for SC4A models are defined T4…T6 (See Note).
Note: Intrinsically safe, entity, for Class I, Division 1, Groups A, B, C and D; Class I, Zone 0, AEx ia IIC, Ga (entity) for
hazardous (classified) locations when installed per control drawing D&E 2020-024-A50.
Maximum sensor input parameters:
Ui= 14.4 V; Ii= 116.5 mA; Pi= 0.3424 W
Li= 0.1 mH (models with fixed cable) or Li= 0 mH (VS type)
Ci= 150 nF (models with fixed cable) or Ci= 0.4 nF (VS type)
Ambient temperature:
-30°C to +40°C for temperature class T6,
-30°C to +95°C for temperature class T5,
-30°C to +125°C for temperature class T4 (VS type),
-30°C to +130°C for temperature class T4 (models with fixed cable).
* Certification is subject to change, due to new regulations or changes in the product itself.When a certificate is updated the
certificate is released under the same number with a new date.(FM-United States: FM20US0123X, effective from 03-2021)
GS 12D07J04-01EN-P
15
7.2 Control drawing: D&E 2020-024-A50 (part 1)
GS 12D07J04-01EN-P
16
7.3 Remarks:
1. No revision to this drawing without prior approval of FM.
2. Installation must be in accordance with the National Electrical Code (ANSI/NFPA 70), ANSI/ISA-RP12.06.01,
and relevant local codes.
3. The sensor shall be installed to a certified intrinsically safe HOST with the following maximum values: Uo= 14.4
V, Io = 116.5 mA, Po = 342.4 mW.
4. The sensor does not provide isolation from earth. Installers shall take necessary measures to prevent the
possibility of sparking resulting from differing earth potentials between the sensors and interconnecting
equipment. This can be realized for example by selecting interconnecting equipment which provides input-to-
output and input-to-earth isolation up to 500 V rms.
5. Sensor Model code:
GS 12D07J04-01EN-P
17
7.6 Control drawing: D&E 2020-024-A50 (part 2)
GS 12D07J04-01EN-P
18
7.7 Remarks:
1. No revision to this drawing without prior approval of FM.
2. Installation must be in accordance with the National Electrical Code (ANSI/NFPA 70), ANSI/ISA-RP12.06.01,
and relevant local codes.
3. The sensor shall be installed to a certified intrinsically safe Smart Adapter, model SA11-C1, with the following
maximum values: Uo= 6.6 V, Io = 100 mA, Po= 165 mW.
4. The Installers shall take necessary measures to prevent the possibility of sparking resulting from differing earth
potentials between the sensors and interconnecting equipment. The sensor itself does not provide 500 V rms
isolation from e arth, the interconnecting equipment Model SA11-C1 Smart Adapter however provides this
required isolation.
5. Sensor Model code:
GS 12D07J04-01EN-P
19
7.10 FM-Canada
Applying standards : CAN/CSA-C22.2 No. 60079-0
CAN/CSA-C22.2 No. 60079-11
Certificate no.* : FM20CA0062X
S CL I, DIV 1, GP ABCD, T4…T6 CL I, ZN 0, Ex ia IIC, T4…T6 Ga
Control Drawing: D&E 2020-024-A51
Electrical data : (See Note 1)
Specific conditions of use : See Control Drawing D&E 2020-024-A51.
Temperature classes for SC4A models are defined T4…T6 (See Note 1).
Note 1 : Intrinsically safe, entity, for Class I, Division 1, Groups A, B, C and D; Class I, Zone 0, Ex ia IIC, Ga (entity) for
hazardous (classified) locations When installed per control drawing D&E 2020-024-A51.
Maximum sensor input parameters:
Ui= 14.4 V; Ii= 116.5 mA; Pi= 0.3424 W;
Li= 0.1 mH (models with fixed cable) or Li= 0 mH (VS type);
Ci= 150 nF (models with fixed cable) or Ci= 0.4 nF (VS type
Ambient temperature:
- 30°C to +40°C for temperature class T6,
- 30°C to +95°C for temperature class T5,
- 30°C to +125°C for temperature class T4 (VS type),
- 30°C to +130°C for temperature class T4 (models with fixed cable).
* Certification is subject to change, due to new regulations or changes in the product itself. When a certificate is updated the
certificate is released under the same number with a new date. - FM-Canada: FM20CA0062X (effective from 03-2021)
GS 12D07J04-01EN-P
20
7.11 Control drawing: D&E 2020-024-A51 (part 1)
GS 12D07J04-01EN-P
21
7.12 Remarks:
1. No revision to this drawing without prior approval of FM.
2. Installation must be in accordance with the National Electrical Code (CEC) CSA22.1 and relevant local codes.
3. The sensor shall be installed to a certified intrinsically safe HOST with the following maximum values: Uo= 14.4
V, Io = 116.5 mA, Po = 342.4 mW.
4. The sensor does not provide isolation from earth. Installers shall take necessary measures to prevent the
possibility of sparking resulting from differing earth potentials between the sensors and interconnecting
equipment. This can be realized for example by selecting interconnecting equipment which provides input-to-
output and input-to-earth isolation up to 500 V rms.
5. Sensor Model code:
GS 12D07J04-01EN-P
22
7.15 Control drawing: D&E 2020-024-A51 (part 2)
GS 12D07J04-01EN-P
23
7.16 Remarks:
1. No revision to this drawing without prior approval of FM.
2. Installation must be in accordance with the National Electrical Code (CEC) CSA22.1 and relevant local codes.
3. The sensor shall be installed to a certified intrinsically safe Smart Adapter, model SA11-C1, with the following
maximum values: Uo= 6.6 V, Io = 100 mA, Po = 165mW.
4. The Installers shall take necessary measures to prevent the possibility of sparking resulting from differing
earth potentials between the sensors and interconnecting equipment. The sensor itself does not provide 500 V
rms isolation from earth, the interconnecting equipment Model SA11-C1 Smart Adapter however provides this
required isolation.
5. Sensor Model code:
GS 12D07J04-01EN-P
Yokogawa has an extensive sales
and distribution network.
Please refer to the European website
(www.yokogawa.com/eu) to contact
your nearest representative.
All Rights Reserved. Copyright © 2021, Yokogawa Process Analyzers Europe BV GS 12D07J04-01EN-P Nov.11, 2021-02