Certificate of Accreditation: Perry Johnson Laboratory Accreditation, Inc

This certificate from Perry Johnson Laboratory Accreditation, Inc. accredits Calidad MX, S.A. de C.V. to ISO/IEC 17025:2017 for calibration services. Calidad MX is accredited to perform calibrations of dimensional, optical, chemical, thermodynamic, mechanical, time & frequency, mass, force and weighing devices and electrical calibration. The scope of accreditation details specific calibration equipment, ranges, and measurement capabilities. The certificate is valid until February 28, 2024.

Uploaded by

Eduardo Cortez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
31 views19 pages

Certificate of Accreditation: Perry Johnson Laboratory Accreditation, Inc

This certificate from Perry Johnson Laboratory Accreditation, Inc. accredits Calidad MX, S.A. de C.V. to ISO/IEC 17025:2017 for calibration services. Calidad MX is accredited to perform calibrations of dimensional, optical, chemical, thermodynamic, mechanical, time & frequency, mass, force and weighing devices and electrical calibration. The scope of accreditation details specific calibration equipment, ranges, and measurement capabilities. The certificate is valid until February 28, 2024.

Uploaded by

Eduardo Cortez
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 19

PERRY JOHNSON LABORATORY

ACCREDITATION, INC.

Certificate of Accreditation
Perry Johnson Laboratory Accreditation, Inc. has assessed the Laboratory of:

Calidad MX, S.A. de C.V.


Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160

(Hereinafter called the Organization) and hereby declares that Organization is accredited
in accordance with the recognized International Standard:

ISO/IEC 17025:2017

This accreditation demonstrates technical competence for a defined scope and the
operation of a laboratory quality management system
(as outlined by the joint ISO-ILAC-IAF Communiqué dated April 2017):

Dimensional, Optical, Chemical, Thermodynamic, Mechanical, Time &


Frequency, Mass, Force and Weighing Devices and Electrical Calibration
(As detailed in the supplement)

Accreditation claims for such testing and/or calibration services shall only be made from addresses referenced within this
certificate. This Accreditation is granted subject to the system rules governing the Accreditation referred to above, and the
Organization hereby covenants with the Accreditation body’s duty to observe and comply with the said rules.
For PJLA:

Initial Accreditation Date: Issue Date: Expiration Date:


July 14, 2011 December 21, 2021 February 28, 2024

Tracy Szerszen Accreditation No.: Certificate No.:


President/Operations Manager
70242 L21-791

Perry Johnson Laboratory


Accreditation, Inc. (PJLA)
The validity of this certificate is maintained through ongoing assessments based
755 W. Big Beaver, Suite 1325 on a continuous accreditation cycle. The validity of this certificate should be
Troy, Michigan 48084 confirmed through the PJLA website: www.pjlabs.com

Page 1 of 19
Certificate of Accreditation: Supplement
Calidad MX, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Dimensional
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND REFERENCE
APPROPRIATE CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
CaliperFO 0.105 in to 24 in (7.7 + 1.1 x 10-2L) µm Mitutoyo Gage Block,
MicrometerFO 0.105 in to 24 in (5.82 x 10-1 + 6 x 10-6L) µm Grade 0
CENAM Technical Guide
Micrometer HeadsF 2.6 mm to 25.4 mm (5.76 x 10-4 + 6.1 x 10-6L) mm Gage Block Grade 0
DIN-030 CEM
Coating Thickness 20 µm to 2 600 µm (5.78 x 10-2 + 1.84 x 10-1 L) µm Defelsco Certified
GaugeFO ASTM-D7091-13
Thickness Standards
TapeFO 50 m maximum 0.1 cm Standard Tape
NOM-046-SCFI-1999
RuleFO 100 cm maximum 0.006 cm Standard Rule
Glass Microrule
NOM-040-SCFI-1994
SievesF 45 µm to 13 200 µm 0.38 µm Microscope
Glass Microrule
ASTM E11-17
16 mm to 125 mm 0.012 mm Interior Caliper
ASTM E11-17
CMMO Up to 1 000 mm (1.16 + 1 x 10-6L) µm Gage Blocks
ISO-10360-2
Height GaugesFO Up to 609.6 mm (5.51 x 10-3 + 3 x 10-6L) mm Mitutoyo Gage
Height MasterFO Up to 304.8 mm (6.21 x 10-4 + 1 x 10-6L) mm Block Grade 0
DI-008 CEM
Angle MeterFO Up to 90° 0.007° Angle Block
DI-003 CEM
Granite Surface Plates 0.05 mm 1.2 μm Repeat-O-Meter with Micro
Repeat Measurement Indicator
DI-015 CEM
Roughness TesterFO Roughness Specimen
Ra (Fixed point) 2.94 µm 0.08 µm DI-025 CEM
Ry (Fixed point) 9.3 µm 0.12 µm
Dial IndicatorFO Up to 25.4 mm (4.7 + 4 x 10-3L) μm Head Micrometer
JIS B 7503
Optical Comparator Glass Reticules
min div. 0.001 mn DI-001 CEM
LengthO
X Axis Linerity 1 mm to 200 mm (9.87 x 10-4 + 1 x 10-6L) mm
Y Axis linearity 1 mm to 200 mm (9.87 x 10-4 + 1 x 10-6L) mm
Optical Comparator Up to 90° 0.05º Angular Block
AngularityO DI-001 CEM

Issue XX/20XX This supplement is in conjunction with certificate #L21-XX Page 2 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Dimensional
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND REFERENCE
APPROPRIATE CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
Microscopes FO Up to 100 mm (1.51 x 10-3 + 4.9 x 10-5L) mm Glass Scale, Reticule
DI-006 CEM
Vision SystemO Up to 200 mm (1.51 x 10-3 + 4.9 x 10-5L) mm Glass Reticules
DI-006 CEM
Pin GaugeF 0.152 mm to 25.4 mm 1.2 µm Master Micrometer
DI-016 CEM
Tread Plugs Pitch 0-40 to 4-12 1.4 µm Wire with Micrometer
DiameterF ASME B1.2
Feeler GaugeF 0.03 mm to 0.9 mm 1.3 µm Micrometer Digital
JIS B 7524
Bore GaugesF 10 mm to 60 mm (6.84 x 10-4 + 8.1 x 10-6L) mm Ring Gauge Master
ASME B89.1.1.6

Optical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND REFERENCE
APPROPRIATE CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
ρ (λ) Spectral Color Values: White Standard Tile
ReflectanceFO CENAM Technical Guide
CIE L: 0 to 100 0.36 Units
CIE a*: -28 to 36 0.26 Units
CIE b*:
-26 to 63 0.24 Units
Spectrophotometers τ: 1 % to 95 % 0.27 % of reading Neutral density Filters,
TransmittanceFO λ: 230 nm to 700 nm 0.5 nm Holmium Oxide Glass
CENAM Technical Guide
Gloss/Specular Ceram Research Gloss and
Reflectance Semi-Gloss Standards
Angle of Incline FO 20º to 92.1º 0.5 Gloss Units ASTM D-523-14
60º to 94.9º 0.5 Gloss Units
85º to 99.8º 0.5 Gloss Units
Eν IlluminanceO 100 lux to 6 000 lux 1 % of reading Luxmeter Minolta CL-200
Light Booth NIST SP 250-37
Eν Light ColorO 60 K to 6 500 K 20 K
Light Booth
Eν Light MetersF 100 lux to 6 000 lux 2 % of reading Luxometer Minolta CL-200
CENAM Technical Guide

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 3 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Chemical
MEASURED INSTRUMENT, RANGE CALIBRATION CALIBRATION
QUANTITY OR GAUGE (AND SPECIFICATION AND MEASUREMENT EQUIPMENT AND REFERENCE
WHERE APPROPRIATE) CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
pH MetersFO 4 pH to 10 pH 0.02 pH SRM NIST Traceable,
(Buffer, 4.01, 7.01, 10.01)
CENAM Technical Guide
Conductivity Meters 84 µS/cm 1 µS/cm SRM NIST Traceable
Fixed pointsFO 1 413 µS/cm 7 µS/cm CENAM Technical
Guide
TurbidimeterFO 0.1 NTU to 100 NTU 0.5 NTU EPA Method 180.1
100 NTU to 800 NTU 5 NTU HACH Standard
Refractive IndexFO 1 ºBrix to 80 ºBrix 0.55 % of reading Sucrose Standards
OIML R-108

Mechanical
MEASURED INSTRUMENT, RANGE OR NOMINAL DEVICE CALIBRATION CALIBRATION
QUANTITY OR GAUGE SIZE AS APPROPRIATE AND MEASUREMENT EQUIPMENT AND REFERENCE
CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
BuretteF 10 mL 14 µL Analytical Balance
25 mL 35 µL AND HR200
CENAM Technical Guide
50 mL 50 µL
Volumetric Pipettes and 1 mL 3.3 µL
PipettesF 5 mL 3.3 µL
10 mL 3.3 µL
25 mL 5.3 µL
Micropipettes and 1 µL 0.023 µL Micro Analytical Balance
PipettesF 2 µL 0.023 µL AND AD-4212B-PT
Analytical Balance
5 µL 0.059 µL AND HR200
10 µL 0.051 µL CENAM Technical Guide
20 µL 0.042 µL
50 µL 0.015 µL
100 µL 0.015 µL
200 µL 0.083 µL
500 µL 0.32 µL
1 000 µL 0.32 µL
2 000 µL 0.32 µL

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 4 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Mechanical
MEASURED INSTRUMENT, RANGE OR NOMINAL DEVICE CALIBRATION CALIBRATION
QUANTITY OR GAUGE SIZE AS APPROPRIATE AND MEASUREMENT EQUIPMENT AND
CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Graduated CylinderF 25 mL 1.2 mL Analytical Balance
50 mL 1.2 mL AND HR200
OHAUS SPX2202
100 mL 1.2 mL CENAM Technical Guide
250 mL 1.2 mL
500 mL 1.2 mL
1 000 mL 1.2 mL
2 000 mL 1.2 mL
Volumetric FlaskF 10 mL 0.02 µL
25 mL 0.02 µL
50 mL 0.02 µL
100 mL 0.03 mL
250 mL 0.07 mL
500 mL 0.08 mL
1 000 mL 0.15 mL
2 000 mL 0.33 mL
ContainersF 10 L 0.34 mL
20 L 0.34 mL
200 L 1.7 mL
Pressure Gauge 1 psi to 1 000 psi 0.6 % of reading Digital Manometer
Pressure TransducerFO Ametek EPC2000
CENAM Technical Guide
1 000 psi to 10 000 psi 0.2 % of reading Digital Manometer
Adittel GP10K
CENAM Technical Guide
0.02 H20 to 280 in H20 0.02 % of reading AMETEC EPC 2000
CENAM Technical Guide
VacuumFO -14 psi to 0 psi 0.02 % of reading
Pressure Gauge
Pressure Transducer
Dynamic Viscosity 0.1 Pa∙s to 30 Pa∙s 0.58 % of reading Cannon Standard OIL
MetersFO NMX-U-038-SCFI
Kinematic Viscosity Ford 121.6 mm²/s 1.4 % of reading Cannon Standard OIL
Cup No.4FO ASTM D4212 − 16

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 5 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Mechanical
MEASURED INSTRUMENT, RANGE OR NOMINAL DEVICE CALIBRATION CALIBRATION
QUANTITY OR GAUGE SIZE AS APPROPRIATE AND MEASUREMENT EQUIPMENT AND
CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Kinematic Viscosity 48.75 mm²/s 1.1 % of reading Cannon Standard OIL
Zahn Cups No. 2FO ASTM D4212 − 16
Kinematic Viscosity 42.37 mm²/s 1.1 % of reading
Zahn Cups No. 3FO
Kinematic Viscosity 32.57 mm²/s 1.1 % of reading
Zahn Cups No. 4FO
Torque Wrench 2 N∙m to 500 N∙m 1 % of reading Torque Transducer
Torque TransducerF CEDAR Mod.
DIS-IP500 500 N∙m
CENAM Technical Guide
Torque Wrench 0.1 N∙m to 15 N∙m 0.5 % of reading Torque Transducer
Torque TransducerFO CEDAR Mod.
IM-CD150M
Indirect Verifications 20 HRC to 30 HRC 0.42 HRC E18-16 Test Blocks
Hardness Tester 30 HRC to 60 HRC 0.39 HRC CENAM Technical Guide
HRCO
60 HRC to 70 HRC 0.38 HRC
Indirect Verifications 40 HRB to 60 HRB 0.38 HRB
Hardness Tester 60 HRB to 80 HRB 0.36 HRB
HRBO
80 HRB to 100 HRB 0.42 HRB
Indirect Verifications 120 HB to 300 HB 1.8 HB
Hardness Tester at 10/1 500 kgf
HBO 300 HB to 600 HB 6 HB
at 10/3 000 kgf
Direct Verification of ASTM D-2240
Durometer Hardness
TesterF
Types A, B, C, D, E, O, & 2.46 mm to 2.54 mm 3.7 µm Head Micrometer
DO
Extension at zero reading

Durometer Indentor
Spring
Types A, B, E & O 0.55 N to 9.05 N 1.4 N Electronic Balance
Types C, D & DO 0.445 N to 44.45 N 1.4 N

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 6 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Mass, Force and Weighting Devices


MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND
APPROPRIATE CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Force Compression and 1 N to 4 906 N 0.3 % of reading Transducer Load Cell
TensionFO 4 906 N to 44 444 N 0.3 % of reading CENAM Technical Guide
44 444 N to 222 222 N 0.1 % of reading
Analytical BalanceFO 1 mg to 200 g (2 x 10-4 + 5.21 x 10-6Wt) g OIML E2 Weights
(Res. = 0.1 mg) CENAM Technical Guide
Precision BalanceFO 0.1 g to 10 000 g (1.16 x 10-2 + 3.35 x 10-6Wt) g OIML F1 and M1
(Res = 0.01 g) Weights
CENAM Technical Guide
ScaleFO 5 kg to 200 kg (1.142 9 + 2.45 x 10-6Wt) g OIML M1 Weights
(Res.= 1 g) CENAM Technical Guide
Weighing DevicesO 200 kg to 10 000 kg (5.85 x 10-1 + 3.1 x 10-5Wt) kg OIML M2 Weights
(Res.= 0.5 kg) CENAM Technical Guide
Mass Class F1, M1 0.1 g 0.05 mg Class E2 and F1Mass
WeightsFO 0.5 g 0.05 mg Micro and
Analytical Balance
1g 0.13 mg Precision Balance
2g 0.16 mg (Res.= 0.01 g)
5g 0.18 mg OIML R-111
10 g 0.22 mg
Mass Class F1, M1 20 g 0.28 mg Class E2 and F1 Mass
WeightsFO 100 g 0.18 mg Analytical Balance
(Res.= 0.01 g) Precision Balance
200 g 0.34 mg OIML R-111
500 g 0.13 mg
Mass Class M1, M2 1 kg 19 mg Class F1 Mass
WeightsFO 2 kg 35 mg Precision Balance
(Res.= 0.01 g) OIML R-111
5 kg 150 mg
Mass Class M2, M3 10 kg 580 mg Class M1 Mass Balance
WeightsFO 20 kg 580 mg OIML R-111

Thermodynamic
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND
APPROPRIATE CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Thermocouple KFO 0 ºC to 500 ºC 0.24 ºC Fluke 724, Dry Well
501 ºC to 1 100 ºC 0.24 ºC NIST 250-35
Thermocouple JFO 0 ºC to 500 ºC 0.24 ºC
501 ºC to 1 100 ºC 0.24 ºC

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 7 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Thermodynamic
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND REFERENCE
APPROPRIATE CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
RTD Pt 100 -80 ºC to 420 ºC 0.11 ºC Fluke 724 with
Dry Well, Freezer, Dry Ice
Bath Comparison
CENAM Technical Guide
Bimetallic ThermometerFO -20 ºC to 500 ºC 1.2 ºC Fluke 724, Dry Well
Temperature ControllersFO 0 ºC to 1 100 °C 0.5 ºC NIST 250-35
IR ThermometerFO 50 ºC to 500 ºC 0.64 ºC Fluke 724 Dry Block
Black Body
(Temperature Generator)
CENAM Technical Guide
Temperature Generation: -20 ºC to 0 ºC 1.4 ºC Fluke 724
Ovens, Furnaces, Muffles, 0 ºC to 25 ºC 1.4 ºC NIST 250-35
Freezers and IncubatorsFO
25 ºC to 100 ºC 1.5 ºC
100 ºC to 450 ºC 1.7 ºC
450 ºC to 900 ºC 1.9 ºC
Relative HumidityFO 11.3 % RH 1.5 % RH Saturated Salt Solution
35 % RH 1.5 % RH OIML R-121
95 % RH 1.5 % RH

Time & Frequency


MEASURED INSTRUMENT, RANGE OR NOMINAL DEVICE CALIBRATION CALIBRATION
QUANTITY OR GAUGE SIZE AS APPROPRIATE AND MEASUREMENT EQUIPMENT AND REFERENCE
CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
StopwatchF 60 s to 86 400 s 16 s/day Direct Comparison
Stopwatch, UTC
CENAM Technical Guide
TachometerF 1 rev/min to 10 000 0.2 % of reading Tachometer
rev/min ASTM-F2046-00

Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL DEVICE CALIBRATION CALIBRATION
QUANTITY OR GAUGE SIZE AS APPROPRIATE AND MEASUREMENT EQUIPMENT AND REFERENCE
CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
Equipment to Measure 0.1 Ω to 1 Ω 0.0026 Ω Process Calibrator
ResistanceFO 1 Ω to 2 Ω 0.002 4 Ω Fluke 743 B
CENAM Technical Guide
2 Ω to 5 Ω 0.002 5 Ω

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 8 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL DEVICE CALIBRATION CALIBRATION
QUANTITY OR GAUGE SIZE AS APPROPRIATE AND MEASUREMENT EQUIPMENT AND
CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Equipment to Measure 5 Ω to 10 Ω 0.002 4 Ω Process Calibrator
ResistanceFO 10 Ω to 20 Ω 0.004 9 Ω Fluke743 B
CENAM Technical Guide
20 Ω to 50 Ω 0.007 Ω
50 Ω to 100 Ω 0.039 Ω
100 Ω to 200 Ω 0.017 Ω
200 Ω to 500 Ω 0.018 Ω
500 Ω to 1 000 Ω 0.9 Ω
Equipment to Output 100 mV to 1 V 0.001 8 V Fluke 5500A
DC VoltageFO 1 V to 10 V 0.002 V Multimeter
CENAM Technical Guide
10 V to 100 V 0.003 V
100 V to 1 000 V 0.048 V
Equipment to Output 0.001 V to 750 V 0.13 V
AC Voltage
At the listed frequencies
50 Hz to 1 kHzFO
Equipment to Output 0.01 A to 3 A 0.001 3 A
AC Current 3 A to 10 A 0.024 A
At the listed frequencies
10 Hz to 900 HzFO
Equipment to output 0.000 01 mA to 9.999 9 mA 0.001 9 mA
DC CurrentFO 10 mA to 99.999 mA 0.006 8 mA
100 mA to 1 A 0.006 9 mA
1 A to 3 A 0.000 7 A
3 A to 10 A 0.006 8 A
Equipment to Output Up to 100 Ω 0.01 % of reading + 0.004 range Agilent Multimeter
ResistanceFO 100 Ω to 1 kΩ 0.01 % of reading + 0.001 range 34401A
Procedure EL-024
1 kΩ to 10 kΩ 0.01 % of reading + 0.001 range Procedure EL-025
10 kΩ to 100 kΩ 0.01 % of reading + 0.001 range
100 kΩ to 1 MΩ 0.01 % of reading + 0.001 range
1 MΩ to 10 MΩ 0.04 % of reading + 0.001 range
10 MΩ to 100 MΩ 0.8 % of reading + 0.01 range

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 9 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:


Electrical
MEASURED RANGE OR NOMINAL DEVICE CALIBRATION CALIBRATION
INSTRUMENT, SIZE AS APPROPRIATE AND MEASUREMENT EQUIPMENT AND
QUANTITY OR GAUGE CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Electrical Current 10 A to 150 A 0.007 9 A Fluke 5500A Multimeter
Derivator (Shunt)FO CENAM Technical
Guide
Equipment to Output 1 kV to 6 kV 0.38 kV Fluke 177 Multimeter
AC High Voltage CENAM Technical
(Hy-Pot)O Guide
Equipment to Output 1 kV to 6 kV 0.1 kV Fluke 177 Multimeter
DC High Voltage Fluke 80K-40
(Hy-Pot)O CENAM Technical
Guide
Equipment to 100 Ω 0.03 Ω Megabox Resistance
Measurement 500 Ω 0.18 Ω CENAM Technical
ResistanceFO Guide
1 kΩ 0.014 kΩ
10 kΩ 0.022 kΩ
100 kΩ 0.02 kΩ
1 MΩ 0.015 MΩ
10 MΩ 0.023 MΩ
100 MΩ 0.022 MΩ
1 GΩ 0.014 MΩ
Temperature -130 °C to 800 °C 0.89 °C Fluke 5500A
Calibration Indication -60 °C to 1 000 °C 0.26 °C Electrical Simulation of
and Control Thermocouple Output
Equipment used with -40 °C to 1 200 °C 0.77 °C CENAM Technical
Thermocouple Type E -140 °C to 130 °C 1.2 °C Guide
Equipment to 0 mV to 329.99 mV 0.006 % of reading + 3 μV Fluke 5500A
Measurement 329.99 mV to 3.299 V 0.005 % of reading + 5 μV Multicalibrator
DC VoltageFO CENAM Technical
3.299 V to 32.999 V 0.005 % of reading + 50 μV Guide
30 V to 329.99 V 0.005 5 % of reading + 500 μV
100 V to 1 020 V 0.005 5 % of reading + 1 500 μV
Equipment to Up to 3.299 99 mA 0.013 % of reading + 0.05 μA
Measurement 3.299 99 mA to 32.999 9 mA 0.01 % of reading + 0.25 μA
DC CurrentFO
32.999 9 mA to 329.999 mA 0.01 % of reading + 3.3 μA
329.999 mA to 2.199 99 A 0.03 % of reading + 44 μA
2.199 99 A to 11 A 0.06 % of reading+ 330 μA
Equipment to Measure 0 Ω to 10.999 Ω 0.012 % of reading + 0.008 Ω Fluke 5500A
ResistanceFO 11 Ω to 32.999 Ω 0.012 % of reading + 0.001 5 Ω Multicalibrator
CENAM Technical
33 Ω to 109.999 Ω 0.009 % of reading + 0.001 5 Ω Guide
110 Ω to 329.999 Ω 0.009 % of reading + 0.001 5 Ω
Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 10 of 19
Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND
APPROPRIATE CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Equipment to Measure 330 Ω to 1.099 99 kΩ 0.009 % of reading + 0.06 Ω Fluke 5500A
ResistanceFO 1.1 kΩ to 3.299 99 kΩ 0.009 % of reading + 0.06 Ω Multicalibrator
CENAM Technical Guide
3.3 kΩ to 10.999 9 kΩ 0.009 % of reading + 0.6 Ω
11 kΩ to 32.999 9 kΩ 0.009 % of reading + 0.6 Ω
33 kΩ to 109.999 kΩ 0.011 % of reading + 6 Ω
110 kΩ to 329.999 kΩ 0.012 % of reading + 6 Ω
330 kΩ to 1.099 99 MΩ 0.015 % of reading + 55 Ω
1.1 MΩ to 3.299 99 MΩ 0.015 % of reading + 55 Ω
3.3 MΩ to 10.999 9 MΩ 0.06 % of reading + 550 Ω
11 MΩ to 32.999 9 MΩ 0.1 % of reading + 550 Ω
33 MΩ to 109.999 MΩ 0.5 % of reading + 5 500 Ω
110 MΩ to 330 MΩ 0.5 % of reading + 16 500 Ω
Equipment to Measure
AC Voltage
At the listed frequenciesFO
10 Hz to 45 Hz 1 mV to 32.999 mV 0.35 % of reading + 20 μV
45 kHz to 10 kHz 1 mV to 32.999 mV 0.15 % of reading + 20 μV
10 kHz to 20 kHz 1 mV to 32.999 mV 0.2 % of reading + 20 μV
20 kHz to 50 kHz 1 mV to 32.999 mV 0.25 % of reading + 20 μV
50 kHz to 100 kHz 1 mV to 32.999 mV 0.35 % of reading + 33 μV
100 kHz to 500 kHz 1 mV to 32.999 mV 1 % of reading + 60 μV
Equipment to Measure
AC Voltage
At the listed frequenciesFO
10 Hz to 45 Hz 33 mV to 329.999 mV 0.25 % of reading + 50 μV
45 kHz to 10 kHz 33 mV to 329.999 mV 0.05 % of reading + 20 μV
10 kHz to 20 kHz 33 mV to 329.999 mV 0.1 % of reading + 20 μV
20 kHz to 50 kHz 33 mV to 329.999 mV 0.16 % of reading + 40 μV
50 kHz to 100 kHz 33 mV to 329.999 mV 0.24 % of reading + 170 μV
100 kHz to 500 kHz 33 mV to 329.999 mV 0.7 % of reading + 330 μV
Equipment to Measure
AC Voltage
At the listed frequenciesFO
10 Hz to 45 Hz 0.33 V to 3.299 99 V 0.15 % of reading + 250 μV
45 kHz to 10 kHz 0.33 V to 3.299 99 V 0.03 % of reading + 60 μV

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 11 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND
APPROPRIATE CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Equipment to Measure Fluke 5500A
AC Voltage Multicalibrator
At the listed frequenciesFO CENAM Technical Guide
10 kHz to 20 kHz 0.33 V to 3.299 99 V 0.08 % of reading + 60 μV
20 kHz to 50 kHz 0.33 V to 3.299 99 V 0.14 % of reading + 300 μV
50 kHz to 100 kHz 0.33 V to 3.299 99 V 0.24 % of reading + 1 700 μV
100 kHz to 500 kHz 0.33 V to 3.299 99 V 0.5 % of reading + 3 300 μV
Equipment to Measure
AC Voltage
At the listed frequenciesFO
10 Hz to 45 Hz 3.3 V to 32.999 9 V 0.15 % of reading + 2 500 μV
45 kHz to 10 kHz 3.3 V to 32.999 9 V 0.04 % of reading + 600 μV
10 kHz to 20 kHz 3.3 V to 32.999 9 V 0.08 % of reading + 2 600 μV
20 kHz to 50 kHz 3.3 V to 32.999 9 V 0.19 % of reading + 5 000 μV
50 kHz to 100 kHz 3.3 V to 32.999 9 V 0.24 % of reading + 17 000 μV
Equipment to Measure
AC Voltage
At the listed frequenciesFO
45 Hz to 1 kHz 33 V to 329.999 9 V 0.05 % of reading + 6 600 μV
1 kHz to 10 kHz 33 V to 329.999 9 V 0.08 % of reading + 15 000 μV
10 kHz to 20 kHz 33 V to 329.999 9 V 0.09 % of reading + 33 000 μV
Equipment to Measure
AC Voltage
At the listed frequenciesFO
45 Hz to 1 kHz 330 V to 1 020 V 0.05 % of reading + 80 000 μV
1 kHz to 5 kHz 330 V to 1 020 V 0.2 % of reading + 100 000 μV
5 kHz to 10 kHz 330 V to 1 020 V 0.2 % of reading + 500 000 μV
Equipment to Measure
AC Current
At the listed frequenciesFO
10 Hz to 20 Hz 0.029 mA to 0.329 99 mA 0.25 % of reading + 0.15 μA
20 Hz to 45 Hz 0.029 mA to 0.329 99 mA 0.13 % of reading + 0.15 μA
45 Hz to 1 kHz 0.029 mA to 0.329 99 mA 0.13 % of reading + 0.25 μA
1 kHz to 5 kHz 0.029 mA to 0.329 99 mA 0.4 % of reading + 0.15 μA
5 kHz to 10 kHz 0.029 mA to 0.329 99 mA 1.3 % of reading + 0.15 μa

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 12 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:

Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND
APPROPRIATE CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Equipment to Measure Fluke 5500A
AC Current Multicalibrator
At the listed frequenciesFO CENAM Technical Guide
10 Hz to 20 Hz 0.33 mA to 3.299 9 mA 0.2 % of reading + 0.3 μA
20 Hz to 45 Hz 0.33 mA to 3.299 9 mA 0.1 % of reading + 0.3 μA
45Hz to 1 kHz 0.33 mA to 3.299 9 mA 0.1 % of reading + 0.3 μA
1 kHz to 5 kHz 0.33 mA to 3.299 9 mA 0.2 % of reading + 0.3 μA
5 kHz to 10 kHz 0.33 mA to 3.299 9 mA 0.6 % of reading + 0.3 μa
Equipment to Measure
AC Current
At the listed frequenciesFO
10 Hz to 20 Hz 3.3 mA to 32.999 9 mA 0.2 % of reading + 3 μA
20 Hz to 45 Hz 3.3 mA to 32.999 9 mA 0.1 % of reading + 3 μA
45Hz to 1 kHz 3.3 mA to 32.999 9 mA 0.09 % of reading + 3 μA
1 kHz to 5 kHz 3.3 mA to 32.999 9 mA 0.2 % of reading + 3 μA
5 kHz to 10 kHz 3.3 mA to 32.999 9 mA 0.6 % of reading + 3 μa
Equipment to Measure
AC Current
At the listed frequenciesFO
10 Hz to 20 Hz 33 mA to 329.999 mA 0.2 % of reading + 30 μA
20 Hz to 45 Hz 33 mA to 329.999 mA 0.1 % of reading + 30 μA
45Hz to 1 kHz 33 mA to 329.999 mA 0.09 % of reading + 30 μA
1 kHz to 5 kHz 33 mA to 329.999 mA 0.2 % of reading + 30 μA
5 kHz to 10 kHz 33 mA to 329.999 mA 0.6 % of reading + 30 μa
Equipment to Measure
AC Current
At the listed frequenciesFO
10 Hz to 45 Hz 0.33 A to 2.199 99 A 0.2 % of reading + 300 μA
45 Hz to 1 kHz 0.33 A to 2.199 99 A 0.1 % of reading + 300 μA
1 kHz to 5 kHz 0.33 A to 2.199 99 A 0.75 % of reading + 300 μA
Equipment to Measure
AC Current
At the listed frequenciesFO
45 Hz to 65 Hz 2.2 A to 11 A 0.06 % of reading + 2 000 μA
65 Hz to 500 Hz 2.2 A to 11 A 0.1 % of reading + 2 000 μA
500 Hz to 1 kHz 2.2 A to 11 A 0.33 % of reading + 2 000 μA

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 13 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:


Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND REFERENCE
APPROPRIATE CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
Equipment to Measure 0.33 nF to 0.4999 nF 0.5 % of reading + 0.01 pF Fluke 5500A Multicalibrator
CapacitanceFO 0.5 nF to 1.0999nF 0.5 % of reading + 0.01 pF Eurament_cg-11
CENAM Technical Guide
1.1 nF to 3.2999 nF 0.5 % of reading + 0.01 pF
3.3 nF to 10.999nF 0.5 % of reading + 0.01 pF
11 nF to 32.999 nF 0.25 % of reading + 0.1 pF
33 nF to 109.99 nF 0.25 % of reading + 0.1 pF
110 nF to 329.99 nF 0.25 % of reading + 0.3 pF
0.33 μF to 1.0999 μF 0.25 % of reading + 1 pF
1.1 μF to 3.2999 μF 0.35 % of reading + 3 pF
3.3 μF to 10.999 μF 0.35 % of reading + 10 nF
11 μF to 32.999 μF 0.4 % of reading + 30 nF
33 μF to 109.99 μF 0.5 % of reading + 100 nF
110 μF to 329.99 μF 0.7 % of reading + 300 nF
0.33 mF to 1.1 mF 1 % of reading + 300 nF
Equipment to Measure 0.01 Hz to 119.99 Hz 0.000 025 Hz
FrequencyFO 120 Hz to 1199.9 Hz 0.000 025 Hz
1.2 kHz to 11.999 kHz 0.016 Hz
12 kHz to 119.99 kHz 0.016 Hz
120 kHz to 1199.9 kHz 0.016 Hz
1.2 MHz to 2 MHz 0.016 Hz
Temperature Calibration 600 °C to 800 °C 0.44 °C Fluke 5500A Multicalibrator
Indication and Control 800 °C to 1 000 °C 0.34 °C Electrical Simulation of
Equipment used with Thermocouple Output
Thermocouple Type BFO 1 000 °C to 1 550 °C 0.3 °C Eurament_cg-11
1 550 °C to 1 820 °C 0.33 °C CENAM Technical Guide
Temperature Calibration 0 °C to 150 °C 0.3 °C
Indication and Control 150 °C to 650 °C 0.26 °C
Equipment used with
Thermocouple Type CFO 650 °C to 1 000 °C 0.31 °C
1 000 °C to 1 800 °C 0.5 °C
1 800 °C to 2 316 °C 0.84 °C
Temperature Calibration -250 °C to -100 °C 0.5 °C
Indication and Control -100 °C to -25 °C 0.16 °C
Equipment used with
Thermocouple Type EFO -25 °C to 350 °C 0.14 °C
350 °C to 650 °C 0.16 °C
650 °C to 1 000 °C 0.21 °C

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 14 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:


Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND REFERENCE
APPROPRIATE CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
Temperature Calibration -210 °C to -100 °C 0.27 °C Fluke 5500A Multicalibrator
Indication and Control -100 °C to -30 °C 0.16 °C Electrical Simulation of
Equipment used with Thermocouple Output
Thermocouple Type JFO -30 °C to 150 °C 0.14 °C Eurament_cg-11
150 °C to 760 °C 0.17 °C CENAM Technical Guide
760 °C to 1 200 °C 0.23 °C
Temperature Calibration -200 °C to -100 °C 0.33 °C
Indication and Control -100 °C to -25 °C 0.18 °C
Equipment used with
Thermocouple Type KFO -25 °C to 120 °C 0.16 °C
120 °C to 1 000 °C 0.26 °C
1 000 °C to 1 372 °C 0.4 °C
Temperature Calibration -200 °C to -100 °C 0.37 °C
Indication and Control -100 °C to 800 °C 0.26 °C
Equipment used with
Thermocouple Type LFO 800 °C to 900 °C 0.17 °C
Temperature Calibration -200 °C to -100 °C 0.4 °C
Indication and Control -100 °C to -25 °C 0.22 °C
Equipment used with
Thermocouple Type NFO -25 °C to 120 °C 0.19 °C
120 °C to 410 °C 0.18 °C
410 °C to 1 300 °C 0.27 °C
Temperature Calibration 0 °C to 250 °C 0.57 °C
Indication and Control 250 °C to 400 °C 0.35 °C
Equipment used with
Thermocouple Type RFO 400 °C to 1 000 °C 0.33 °C
1 000 °C to 1 767 °C 0.4 °C
Temperature Calibration 0 °C to 250 °C 0.47 °C
Indication and Control 250 °C to 1 000 °C 0.36 °C
Equipment used with
Thermocouple Type SFO 1 000 °C to 1 400 °C 0.37 °C
1 400 °C to 1 767 °C 0.46 °C
Temperature Calibration -250 °C to -150 °C 0.63 °C
Indication and Control -150 °C to 0 °C 0.24 °C
Equipment used with
Thermocouple Type TFO 0 °C to 120 °C 0.16 °C
120 °C to 400 °C 0.14 °C
Temperature Calibration -200 °C to 0 °C 0.56 °C
Indication and Control 0 °C to 600 °C 0.27 °C
Equipment used with
Thermocouple Type UFO

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 15 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:


Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND REFERENCE
APPROPRIATE CAPABILITY EXPRESSED STANDARDS USED
AS AN UNCERTAINTY (±)
Temperature Calibration -200 °C to -80 °C 0.05 °C Fluke 5500A Multicalibrator
Indication and Control -80 °C to 0 °C 0.05 °C Electrical Simulation of
Equipment used with RTD RTD Output
Pt 385, 100 ΩFO 0 °C to 100 °C 0.07 °C Eurament_cg-11
100 °C to 300 °C 0.09 °C CENAM Technical Guide
300 °C to 400 °C 0.1 °C
400 °C to 630 °C 0.12 °C
630 °C to 800 °C 0.23 °C
Temperature Calibration -200 °C to -80 °C 0.05 °C
Indication and Control -80 °C to 0 °C 0.05 °C
Equipment used with RTD
Pt 3926, 100 ΩFO 0 °C to 100 °C 0.07 °C
100 °C to 300 °C 0.09 °C
300 °C to 400 °C 0.1 °C
400 °C to 630 °C 0.12 °C
Temperature Calibration -200 °C to -190 °C 0.25 °C
Indication and Control -190 °C to -80 °C 0.04 °C
Equipment used with RTD
Pt 3916, 100 ΩFO -80 °C to 0 °C 0.05 °C
0 °C to 100 °C 0.06 °C
100 °C to 260 °C 0.07 °C
260 °C to 300 °C 0.08 °C
300 °C to 400 °C 0.09 °C
400 °C to 600 °C 0.1 °C
600 °C to 630 °C 0.23 °C
Temperature Calibration -200 °C to -80 °C 0.04 °C
Indication and Control -80 °C to 0 °C 0.04 °C
Equipment used with RTD
Pt 385, 200 ΩFO 0 °C to 100 °C 0.04 °C
100 °C to 260 °C 0.05 °C
260 °C to 300 °C 0.12 °C
300 °C to 400 °C 0.13 °C
400 °C to 600 °C 0.14 °C
600 °C to 630 °C 0.16 °C
Temperature Calibration -200 °C to -80 °C 0.04 °C
Indication and Control -80 °C to 0 °C 0.05 °C
Equipment used with RTD
Pt 385, 500 ΩFO 0 °C to 100 °C 0.05 °C

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 16 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:


Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND
APPROPRIATE CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Temperature Calibration 100 °C to 260 °C 0.06 °C Fluke 5500A
Indication and Control 260 °C to 300 °C 0.08 °C Multicalibrator
Equipment used with Electrical Simulation of
RTD Pt 385, 500 ΩFO 300 °C to 400 °C 0.08 °C RTD Output
400 °C to 600 °C 0.09 °C Eurament_cg-11
600 °C to 630 °C 0.11 °C CENAM Technical Guide
Temperature Calibration -200 °C to -80 °C 0.03 °C
Indication and Control -80 °C to 0 °C 0.03 °C
Equipment used with
RTD Pt 385, 1 000 ΩFO 0 °C to 100 °C 0.04 °C
100 °C to 260 °C 0.05 °C
260 °C to 300 °C 0.06 °C
300 °C to 400 °C 0.07 °C
400 °C to 600 °C 0.07 °C
Temperature Calibration 600 °C to 630 °C 0.23 °C
Indication and Control -80 °C to 0 °C 0.08 °C
Equipment used with
RTD Pt Ni 385, 120 ΩFO 0 °C to 100 °C 0.08 °C
100 °C to 260 °C 0.14 °C
Temperature Calibration -100 °C to 260 °C 0.3 °C
Indication and Control
Equipment used with
RTD Pt Cu 427, 10 ΩFO
Equipment to Output Up to 100 mV 0.005 % of reading + 0.003 5 % range Agilent Multimeter
AC VoltageFO 100 mV to 1 V 0.004 % of reading + 0.000 7 % range 34401A
EL-024, EL-025
1 V to 10 V 0.003 5 % of reading + 0.000 5 % range
10 V to 100 V 0.004 5 % of reading + 0.000 6 % range
100 V to 1 000 V 0.004 5 % of reading + 0.001 % range
Equipment to Output
AC Voltage
At the listed frequenciesFO
3 Hz to 5 Hz Up to 100 mV 1 % of reading + 0.04 range
5 Hz to 10 Hz Up to 100 mV 0.35 % of reading + 0.04 range
10 Hz to 20 kHz Up to 100 mV 0.06 % of reading + 0.04 range
20 kHz to 50 kHz Up to 100 mV 0.12 % of reading + 0.04 range
50 kHz to 100 kHz Up to 100 mV 0.6 % of reading + 0.08 range
100 kHz to 300 kHz Up to 100 mV 4 % of reading + 0.5 range

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 17 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:


Electrical
MEASURED INSTRUMENT, RANGE OR NOMINAL CALIBRATION CALIBRATION
QUANTITY OR GAUGE DEVICE SIZE AS AND MEASUREMENT EQUIPMENT AND
APPROPRIATE CAPABILITY EXPRESSED REFERENCE
AS AN UNCERTAINTY (±) STANDARDS USED
Equipment to Output Agilent Multimeter
AC Voltage 34401A
At the listed frequenciesFO EL-024, EL-025
3 Hz to 5 Hz 1 V to 750 V 1 % of reading + 0.03 range
5 Hz to 10 Hz 1 V to 750 V 0.35 % of reading + 0.03 range
10 Hz to 20 kHz 1 V to 750 V 0.06 % of reading + 0.03 range
20 kHz to 50 kHz 1 V to 750 V 0.12 % of reading + 0.04 range
50 kHz to 100 kHz 1 V to 750 V 0.6 % of reading + 0.08 range
100 kHz to 300 kHz 1 V to 750 V 4 % of reading + 0.5 range
Equipment to Voltage Up to 10 mA 0.05 % of reading + 0.02 range
Direct CurrentFO 10 mA to 100 mA 0.05 % of reading + 0.005 range
100 mA to 1 A 0.1 % of reading + 0.01 range
1 A to 3 A 0.12 % of reading + 0.02 range
Equipment to Output
AC Current
At the listed frequenciesFO
3 Hz to 5 Hz Up to 1 A 1 % of reading + 0.04 range
5 Hz to 10 Hz Up to 1 A 0.3 % of reading + 0.04 range
10 Hz to 5 kHz Up to 1 A 0.1 % of reading + 0.04 range
Equipment to Output
AC Current
At the listed frequenciesFO
3 Hz to 5 Hz Up to 3 A 1.1 % of reading + 0.06 range
5 Hz to 10 Hz Up to 3 A 0.35 % of reading + 0.06 range
10 Hz to 5 kHz Up to 3 A 0.15 % of reading + 0.06 range
Equipment to Output 3 Hz to 5 Hz 0.1 % of reading
FrequencyFO 5 Hz to 10 Hz 0.05 % of reading
10 Hz to 40 Hz 0.03 % of reading
40 Hz to 300 KHz 0.01 % of reading

1. The CMC (Calibration and Measurement Capability) stated for calibrations included on this scope of
accreditation represents the smallest measurement uncertainty attainable by the laboratory when performing
a more or less routine calibration of a nearly ideal device under nearly ideal conditions. It is typically
expressed at a confidence level of 95 % using a coverage factor k (usually equal to 2). The actual
measurement uncertainty associated with a specific calibration performed by the laboratory will typically
be larger than the CMC for the same calibration since capability and performance of the device being
calibrated and the conditions related to the calibration may reasonably be expected to deviate from ideal to
some degree.

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 18 of 19


Certificate of Accreditation: Supplement
Calidad Mx, S.A. de C.V.
Av. López Mateos # 1001, Local 203 Altos, Col. 21 de Enero
Guadalupe, Nuevo León, México C.P. 67160
Contact: Alejandro Lujan Phone: (52) 818-379-2710

Accreditation is granted to the facility to perform the following calibrations:


2. The laboratories range of calibration capability for all disciplines for which they are accredited is the
interval from the smallest calibrated standard to the largest calibrated standard used in performing the
calibration. The low end of this range must be an attainable value for which the laboratory has or has access
to the standard referenced. Verification of an indicated value of zero in the absence of a standard is
common practice in the procedure for many calibrations but by its definition it does not constitute
calibration of zero capacity.
3. The presence of a superscript F means that the laboratory performs calibration of the indicated parameter at
its fixed location. Example: Outside MicrometerF would mean that the laboratory performs this calibration
at its fixed location.
4. The presence of a superscript O means that the laboratory performs calibration of the indicated parameter
onsite at customer locations. Example: Outside MicrometerO would mean that the laboratory performs this
calibration onsite at the customer’s location.
5. The presence of a superscript FO means that the laboratory performs calibration of the indicated parameter
both at its fixed location and onsite at customer locations. Example: Outside MicrometerFO would mean
that the laboratory performs this calibration at its fixed location and onsite at customer locations.
6. Measurement uncertainties obtained for calibrations performed at customer sites can be expected to be
larger than the measurement uncertainties obtained at the laboratories fixed location for similar
calibrations. This is due to the effects of transportation of the standards and equipment and upon
environmental conditions at the customer site which are typically not controlled as closely as at the
laboratories fixed location.
7. The term L represents length in inches or millimeters as appropriate to the uncertainty statement.
8. The term Wt represents weight in pounds or grams (including SI multiple and submultiple units)
appropriate to the uncertainty statement.

Issue 12/2021 This supplement is in conjunction with certificate #L21-791 Page 19 of 19